Rangkuman Manual

You might also like

Download as docx, pdf, or txt
Download as docx, pdf, or txt
You are on page 1of 4

Rangkuman Manual

Wholeblood, serum, plasma: potassium, sodium, chloride

Spek:

Dimensi-> 300 x 260 x 360 mm

Weight-> 7,5 kg

Power supply-> AC100-240V

Consumption-> 100 VA

Fuse-> 2xF3.15L 250 VAC

Sample volume-> typical 120 mikroliter minimum 65 mikroliter

Sample type whole blood, serum, plasma, urine

Screen: LCD 240 x 64 pixel

Printer

Thermal printer

240 x 128 pixel

Full graphics 8 dots/mm

Printing speed 15mm/s

Paper width 57,5 mm

Reference electrode-> once a week must be shaken

Don’t mix up reference electrode refill solution with ise electrodes

Replace ISE solution every 3 months

Ungrounded can cause drift problem on all electrodes

High power instrument with the same outlet can interfere with the instrument

After power on-> self check:

1. battery and sample sensor check

2. cal A & cal B respirated to check measuring chamber

3. clock test

If error -> power off the instrument and start troubleshooting


Lanjut ke Auto Calibration:

1. Cal A -> cek kestabilan elektroda

2. Cal B-> shows the concentration f Cal B

New Electrode-> if problem or drift is happened, feed fresh serum to activate electrode for 30
minutes, then check again

Shut down:

1. within 24 hours-> switch down directly

2. beyond 24 hours:

Perform Stop Use (Service>Stop Use)

Remove reagent pack from its housing and cover the protective insert

Remove electrde from measuring chamber and place it into box

Release pump tube

Place reagent pack in a safe place

Clean the surface of instrument and place it into carton box

Bisa LIS/ enggak? -> Tanya Teknisi

Setiap kalibrasi 3 times if it’s failed. If the 3rd times is failed, the system will stop and ask for
recalibration.

Extra flushing neede after whole blood & urine test to avoid blockage

Urine samples must be diluted 1:4 with CAL A -> only K, Na, Cl for Urine sample

Deproteinase-> Default every 60 samples or every 10 days. Ganti>> service>>Maintenance Interval

Pakai deproteinizer. Last for 30 minutes

2 PT calibration is performed automatically after exiting deproteinize

Feed fresh serum instead of deproteinizer under deproteinize when activating new electrodes

Conditioning-> only when Na & pH have problem. Take 5 minutes. 2 PT clibration performed
automatically after exiting conditioning

Perform cleaning when electrode has abnormal problem or membrane is dyed. It can also help to
clean electrodes and liquid pathway to prevent them from blockage or clot. After exiting, 2 PT
calibration will start automatically. Default 5 hari.

Adjust sample volume every 3 months to copensate the aged pump tube
The instrument can transfer data to computer via RS232 serial cable

Maintenance -> from manual for 20-30 samples per day

Sterilization with hydrogen peroxide 2%, never use organic solution or alcohol.:

1. probe

2. liquid flow path

3. display and key

Daily maintenance:

1. clean the probe

2. clean the surface of the instrument

3. do 2 PT calibration to check the instrument. After calibration is passed, test a sample to check

4. near day off, deproteinize or cleaning to maintain electrode if necessary

Weekly maintenance:-> at the end of routine measurements:

1. sterilize exterior and interior surface and aspiration probe tie-in

2. take out reference electrode and shake it for several times to avoid crytal formation

3. cleaning the pathway to prevent blockage

4. check the multiplexer if there is any leakage

5. test middle level QC, adjust intercept and slope accordingly

6. perform deproteinize and cleaning to aintain electrodes

7. clean liquid pathway by injectig DI water with a syringe. Avoid hard pushing from breaking the
membrane.

Monthy maintenance:

1. check the solution level of K, Na, Cl, Li electrodes. Replace if less than 2/3

2. Correct coefficient factors (slope and intercept) with high/middle/low level QC

3. Take out reference electrode, remove crystal if there is too much. Refill if the solution is not
enough

4. clean all the pathway to avoid blockage


Every 3 months:

1. Replace refill solution for K, Na, Cl, Ca, Li, and pH electrodes

2. adjust sample volume of aspiration if it is not enough

Every 6 months:

1. replace pump tube

2. Check –ring of electrode

3. check connection of tube

4. check the movement of multiplexer

As Necessary:

1. replace printer paper

2. replace O-rings

3. Perform conditioningif Na nad pH electrode has problem

4. Perform deproteinize if K, Cl, Ca, Li, and pH has problem

5. Perform cleaning if deproteinize is invalid