Spek:
Weight-> 7,5 kg
Consumption-> 100 VA
Printer
Thermal printer
High power instrument with the same outlet can interfere with the instrument
3. clock test
New Electrode-> if problem or drift is happened, feed fresh serum to activate electrode for 30
minutes, then check again
Shut down:
2. beyond 24 hours:
Remove reagent pack from its housing and cover the protective insert
Setiap kalibrasi 3 times if it’s failed. If the 3rd times is failed, the system will stop and ask for
recalibration.
Extra flushing neede after whole blood & urine test to avoid blockage
Urine samples must be diluted 1:4 with CAL A -> only K, Na, Cl for Urine sample
Feed fresh serum instead of deproteinizer under deproteinize when activating new electrodes
Conditioning-> only when Na & pH have problem. Take 5 minutes. 2 PT clibration performed
automatically after exiting conditioning
Perform cleaning when electrode has abnormal problem or membrane is dyed. It can also help to
clean electrodes and liquid pathway to prevent them from blockage or clot. After exiting, 2 PT
calibration will start automatically. Default 5 hari.
Adjust sample volume every 3 months to copensate the aged pump tube
The instrument can transfer data to computer via RS232 serial cable
Sterilization with hydrogen peroxide 2%, never use organic solution or alcohol.:
1. probe
Daily maintenance:
3. do 2 PT calibration to check the instrument. After calibration is passed, test a sample to check
2. take out reference electrode and shake it for several times to avoid crytal formation
7. clean liquid pathway by injectig DI water with a syringe. Avoid hard pushing from breaking the
membrane.
Monthy maintenance:
1. check the solution level of K, Na, Cl, Li electrodes. Replace if less than 2/3
3. Take out reference electrode, remove crystal if there is too much. Refill if the solution is not
enough
1. Replace refill solution for K, Na, Cl, Ca, Li, and pH electrodes
Every 6 months:
As Necessary:
2. replace O-rings