Professional Documents
Culture Documents
PTW Unidose
PTW Unidose
UNIDOSwebline
Type 10021, Type 10022 and Type 10023
Contents
Operating Manual
Technical Manual
Service Manual
General Information
The product bears the CE-mark "CE-0124" in PTW-Freiburg cannot be held liable for dam-
accordance with the Council Directive ages resulting from the use of accessories or
93/42/EEC about Medical Devices and fulfills consumables from other manufacturers or
the essential requirements of Annex I of this when the user ignores the instructions and in-
directive. formation given in this manual.
The product is a class IIb device (MDD).
The warranty period is 1 (one) year and begins
Product standard on the day of delivery.
The product fully complies with the require- It is unaffected by repairs covered by the war-
ments of the following standards: ranty regulations.
IEC 60731
IEC 61674 PTW-Freiburg only considers itself responsible
IEC 60846 for the safety, reliability, and performance of
IEC 62467-1 the product if the assembly, extension, read-
justment, modification, or repair is carried out
Electrical safety by PTW-Freiburg or by persons authorized by
The product fully complies with the require- PTW-Freiburg, and if the product is used in
ments of the standard IEC 61010-1. compliance with the technical documentation.
Operator safety, specified measuring accura- PTW-Freiburg reserves the right to make mod-
cy, and interference-free operation can only be ifications.
guaranteed if original products and parts are Please contact PTW or your local representa-
used. Furthermore, only the accessories listed tive for the most current information concern-
in this manual are approved by PTW-Freiburg ing the products of interest.
and only they or other accessories whose use
has been expressly permitted by PTW- No part of the technical documentation may be
Freiburg may be used in conjunction with the reproduced without written permission from
product. Safe operation and proper product PTW-Freiburg.
performance are not guaranteed if accessories
or consumables from other manufacturers are
used.
2 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
General Information
D776.131.00/07 en 3
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Contents
Contents
General Information 2
Contents 4
Illustrations 7
Tables 7
Intended Use 8
Safety Information 9
Operating Manual 17
1 Description of the UNIDOSwebline 17
webline
1.1 Features of UNIDOS 17
1.2 Measuring Modes 18
1.3 Operating Elements, Indicators, and Interfaces 19
1.4 Enclosure 22
1.5 Power Supply 23
1.6 Detectors 23
3 Quickstart (Workflow) 29
5 Settings 37
5.1 General Notes 37
5.2 Detector and Detector Database Settings 38
5.3 Settings for Measurements 42
5.4 Basic Device Settings 46
5.5 Miscellaneous Device Settings 47
4 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Contents
6 Measuring 50
6.1 General Notes 50
6.2 Connecting a Detector 50
6.3 Use of an External HV Source 53
6.4 Adjusting Corrections 55
6.5 Performing a Measurement Manually 63
6.6 Performing Measurements in the Autostart Mode 63
6.7 Reading the Measured Value 64
6.8 Evaluation of the Resolution of the Measured Value 65
Technical Manual 79
9 Checking the UNIDOSwebline 79
9.1 Offset Current (Zero Drift) and Zeroing 79
9.2 Long-term Stability 80
9.3 High Voltage and Amplifier 81
9.4 Polarity Effect 84
10 Ethernet Interface 86
10.1 General Notes 86
webline
10.2 Connecting UNIDOS to a Data Network 86
10.3 Communication via VNC 86
10.4 Communication via E-Mail 89
webline
10.5 Controlling UNIDOS via UDP 91
12 Technical Specifications 95
12.1 Measuring Ranges and Resolution 95
webline
12.2 Measuring and Display Unit UNIDOS 96
12.3 Definition of Polarity 100
D776.131.00/07 en 5
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Contents
20 Fuses 115
Literature 117
Index 121
6 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Illustrations
Illustrations
Tables
D776.131.00/07 en 7
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Intended Use
webline
Intended Use UNIDOS can be used with appropriate PTW
detectors (see section 1.6) as a dosimetry system.
a
For laboratory use only
8 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Safety Information
DANGER
This is the safety alert symbol. It is used to
Operation in areas where an explosion hazard
alert the user to potential hazards. Obey all
may occur or in oxygen-enriched atmospheres.
safety messages that follow this symbol to
Explosion Hazard!
avoid any possible bodily injury or equip-
ment damage. The product is not suitable for operation in areas
of risk where an explosion may occur. Explosion
All safety messages consist of the following com- hazards may be caused by the use of combus-
ponents: tible anesthetics, skin-cleansing agents, and
– Safety alert symbol and signal word disinfectants.
– Type of danger Furthermore, the product is not suitable for appli-
– Source of danger cation in oxygen-enriched atmospheres. The
– Consequence atmosphere is considered to be oxygen-enriched
– Measures to prevent hazards. when more than 25 % of oxygen or nitrous oxide
is added to the ambient air.
Signal Words
DANGER
Indicates an imminent hazard. If not avoided, the
hazard will result in death or serious injury.
WARNING
Indicates a hazard. If not avoided, the hazard
can result in death or serious injury.
CAUTION
Indicates a potential hazard. If not avoided, the
hazard could result in minor injury or product /
property damage.
NOTE
Provides useful information to ensure that you
get the most from your equipment.
D776.131.00/07 en 9
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Safety Information
WARNING
DANGER
Improper handling.
Sources of high voltage are sources of risk.
Patient Hazard!
Shock Hazard!
The product is a medical electrical device and
External sources of high voltage may be con-
must only be handled by persons who are trained
nected only if they comply with the requirements
in the use of such equipment and are capable of
of IEC 61010-1. They must not exceed the limit
applying it properly. The operator must be trained
values for current specified for exposed parts in
in the use of the device.
IEC 61010-1.
The output current of the external HV source
must be limited to 2 mA max. as per WARNING
IEC 61010-1.
Electricity is a source of risk, particularly when
The external HV source must be floating. This
the product is not in perfect operating condition
means that neither of the two poles of the ex-
or when it is operated inappropriately.
ternal HV source must be connected to
Shock Hazard!
ground (protective earth conductor).
Strictly observe the following warnings. Failure to
The enclosure, all exposed parts and the
do so may endanger the lives of the patient, us-
connectors of the external HV source must
er, and other persons involved.
have a double insulation according to
IEC 610101-1 with respect to both poles. Before using the product, the user must as-
certain that it is in correct working order and
Only insulated BNC connectors are approved
operating condition.
to connect the external HV source.
Before putting the device into operation, visu-
Before switching from internal to external HV
ally inspect all connection cables and con-
source, set the external high voltage to 0 V.
nectors for signs of damage. Damaged cables
Before switching from external to internal HV
and connectors must be replaced immediate-
source, set the external high voltage to 0 V
ly.
and then turn off and remove the external HV
When disconnecting the device from the pow-
source.
er line, first remove the plug from the wall out-
let, then disconnect the power cord from the
device.
10 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Safety Information
WARNING WARNING
Devices on which moisture condensation has Electricity is a source of risk, particularly when
developed as a result of temperature changes the product is not in perfect operating condition
must not be switched on unless completely or when it is operated inappropriately.
dry. Shock Hazard!
Liquids must not enter the product. If liquids Strictly observe the following warnings. Failure to
have entered the product, it must be thor- do so may endanger the lives of the patient, us-
oughly inspected before being used again. er, and other persons involved.
The product is a device of IEC 61140 protec- Use of peripheral devices: Peripheral devices
tion class I. It may be put into operation only (PC, printer) may only be connected if they meet
when connected to a properly installed power the requirements of the standard UL 60950-1 or
outlet with earthing contact. IEC 60950-1 or of the standard IEC 61010-1.
Extension cords must not be used. These devices may not be operated in the pa-
All devices of a system must be connected to tient environment.
the same electric circuit. Devices which are Devices may be connected to other devices
not connected to the same circuit must be or to parts of systems only if it has been as-
electrically isolated (use RS232 interface with certained that this connection does not impair
galvanic insulation). the safety of the patient, the operator or the
Exclusion of operation in the patient envi- environment.
ronment: Neither the product nor any peripheral If the device specifications do not contain
devices may be operated in the patient environ- information as to connecting the device to
ment (refer to Figure 1). other equipment, you must consult the manu-
Exclusion of operation as device with patient facturer of the other equipment or an expert
contact: The device is not intended for use in about the effects of the connection on the pa-
direct contact with the patient. Neither the device tient, the operator or the environment. Always
nor any peripheral device may have contact to observe the standard IEC 61010-1.
the patient.
Never touch the patient and open connectors of
the device at the same time.
Exclusion of operation as a controlling in-
strument: The device is only designed for use
as a measuring device. The device must not be
used to control radiodiagnostic equipment or
radiotherapy units.
D776.131.00/07 en 11
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Safety Information
WARNING WARNING
webline
The UNIDOS may adversely affect the data Magnetic and electrical fields are capable of in-
transmission of other clients in the network. terfering with the proper performance of the de-
Network Problem! vice.
12 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Safety Information
WARNING CAUTION
Improper handling of the trip output interface. Use of the product without observing the user
Patient Hazard! manual.
The trip output on the back of the device must Bodily Injury! Equipment Damage!
not be used for medical applications and radia- Always use the product in compliance with the
tion interlock systems in particular. user manual. Otherwise the intended protection
can be reduced.
Use the product only in conjunction with the
WARNING products approved for this purpose by PTW
Improper handling of the trip output interface. and/or the products listed in the section "Intend-
Bodily Injury! ed Use".
Any connected electric circuits must not exceed Handle the product with care to avoid equipment
the limit values for voltage and current specified damage.
for exposed parts in IEC 61010-1.
CAUTION
WARNING Improper handling.
Improper handling. Erroneous Measurement!
Hazards to Persons! Erroneous Measure- Malfunction!
ment! The range indicator (LOW AUTO SIG) for the
If damages or malfunctions are identified, the Autostart function will operate correctly only for
device must be repaired before it is used again. conventional signal shapes (i.e. almost
squarewave signal over time) occurring for linear
accelerators when an almost constant dose rate
WARNING is turned on and off.
Hazards originating from other system compo- For all other signal shapes (slow transients) and
nents. when in doubt the accuracy of the Autostart func-
tion must be checked by comparing the results
Patient Hazard! Equipment Damage!
with those of manual measurements. For this
Observe the safety information provided in the purpose perform multiple measurements of the
user manuals of the system components. same signal in the Autostart mode and manually
with the MEAS key and compare the average
values and standard deviation of both methods.
D776.131.00/07 en 13
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Safety Information
CAUTION NOTE
Operation under inadequate ambient conditions. Please observe the user manuals of all connect-
Equipment Damage! ed devices!
Before connecting the device to the power line,
make sure that line voltage and frequency are
NOTE
equivalent to the ratings indicated on the name-
plate of the device. Set up the device so that the operator has a
clear, unobstructed view of the control panel.
Set up the device in a location which affords
sufficient ventilation.
Always observe the ambient conditions as indi- NOTE
cated in the "Technical Specifications".
Do not position the device so that it is difficult to
operate the disconnecting device.
CAUTION
Improper handling of cables and connectors.
Equipment Damage!
To avoid damage to the cables and connectors,
observe the following points.
Do not kink the cables. Observe the minimum
bend radius of the cables. Never exert pres-
sure on the cables. Do not step on the cables.
Do not roll trolleys over the cables. Do not pull
or twist the cables.
Disconnect the cables from the devices during
transport and when the devices are not in
use. Do not let the cables hang down unre-
strained.
The connectors must always be clean. Do not
allow the connectors to lie on the floor. Al-
ways protect connectors when pulling them
through cable conduits.
14 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Safety Information
Symbol Description
Caution!
Please refer to the warning in the
user manual!
Caution!
Possibility of electric shock
(Connection cable to external HV
source)
Reference number
Serial number
D776.131.00/07 en 15
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Safety Information
16 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Description of the Device
Operating Manual
D776.131.00/07 en 17
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Description of the Device
Dose Rate
Dose rate or current measurements by measuring
the voltage drop at a resistor. Three resistors are
accessible for the measuring ranges Low, Medium
and High.
Dose
18 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Description of the Device
1 2
webline
Figure 2: Front view of the UNIDOS
1 TFT-Display ---
6 Power key
D776.131.00/07 en 19
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Description of the Device
4 5 6 7 8
2
3
9 10 11
webline
Figure 3: Rear view of the UNIDOS
1 On-off switch I
0
5 Desiccator DESICCANT
6 Type label ---
20 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Description of the Device
DANGER WARNING
Sources of high voltage are sources of risk. Improper handling of the trip output interface.
Shock Hazard! Patient Hazard!
External sources of high voltage may be con- The trip output on the back of the device must
nected only if they comply with the requirements not be used for medical applications and radia-
of IEC 61010-1. They must not exceed the limit tion interlock systems in particular.
values for current specified for exposed parts in
IEC 61010-1.
The output current of the external HV source WARNING
must be limited to 2 mA max. as per Improper handling of the trip output interface.
IEC 61010-1. Bodily Injury!
The external HV source must be floating. This Any connected electric circuits must not exceed
means that neither of the two poles of the ex- the limit values for voltage and current specified
ternal HV source must be connected to for exposed parts in IEC 61010-1.
ground (protective earth conductor).
The enclosure, all exposed parts and the
connectors of the external HV source must WARNING
have a double insulation according to The UNIDOS
webline
may adversely affect the data
IEC 610101-1 with respect to both poles. transmission of other clients in the network.
Only insulated BNC connectors are approved Network Problem!
to connect the external HV source. webline
Via the Ethernet interface, UNIDOS may be
Before switching from internal to external HV connected to data networks only if these net-
source, set the external high voltage to 0 V. works do not determine the safety of the patient,
Before switching from external to internal HV the operator, or other persons present, i.e., only
source, set the external high voltage to 0 V to class A networks as defined in Annex H of
and then turn off and remove the external HV IEC 60601-1: 2005.
source. Only system administrators who are capable of
understanding, assessing, and controlling the
webline
risk arising from the connection of UNIDOS
to the data network are authorized to establish
the connection.
It must also be ensured that any future risk re-
sulting from changes to the data network (e.g.,
connection of devices affecting the system safety
or changes to the network configuration, etc.) will
be understood, assessed, and controlled.
D776.131.00/07 en 21
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Description of the Device
1.4 Enclosure
NOTE The UNIDOS
webline
is equipped with an enclosure
When an external HV source is used, the con- made of aluminum.
stancy of this source determines the quality of
the measuring results. In order to get an ideal view of the display the
webline
Through the detector capacitance, changes in UNIDOS can be set sloped by means of the
the HV source’s output voltage will induce charge carrying handle.
shifts and measuring errors.
NOTE
When using an external HV source, you have to
set longer measuring times, in order to minimize
the influence of noise as necessary. Keep in
mind that the noise depends on the polarity of
the HV. PTW recommends setting a positive HV
webline
when using a UNIDOS with the connecting
system M (T10021) and setting a negative HV
webline
when using a UNIDOS with the connecting
system TNC (T10022) or BNT (T10023).
NOTE
webline
With UNIDOS using the connecting system
M (T10021) the internal conductor of the external
HV source is connected to the chamber wall.
webline
With UNIDOS using the connecting system
TNC (T10022) or BNT (T10023) the shield of the
external HV source is connected to the chamber
wall.
22 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Description of the Device
webline
The UNIDOS is powered via a wide range
power supply which automatically adapts to the In this user manual, the type numbers will be men-
voltage of the power line. Therefore, it is not neces- tioned with the identifiers M, N, and W only if not all
webline
sary to switch the UNIDOS to the respective connecting systems are referred to but only a par-
line voltages. ticular one.
webline
The UNIDOS is equipped with rechargeable
webline
NiMH batteries. When the UNIDOS is turned
on and not connected to the power line, it will be WARNING
powered from the internal, rechargeable batteries.
Improper handling.
The battery mode symbol is shown on the
display. Shock Hazard!
Use only detectors which are adapted to the
webline webline
The UNIDOS is shipped with brandnew re- connecting system of your UNIDOS (M sys-
chargeable batteries. Before setup, these new re- tem, TNC system or BNT system). Please ob-
chargeable batteries have to be inserted into the serve that it is not possible to use an adapter
webline
UNIDOS and have to be charged (refer to cable from the M connecting system to another
section 2). Due to their technology rechargeable connecting system without accessible parts of
batteries show a limited durability. the plug or the detector conducting high voltage.
CAUTION
Use of the product with other detectors.
Bodily Injury! Equipment Damage!
Use the product only with detectors specified by
PTW for this purpose. The user is responsible for
the application of other detectors.
D776.131.00/07 en 23
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Description of the Device
webline
The PTW ionization chamber type 30010 with a UNIDOS (T10021, T10022, T10023) can be
volume of 0.6 cm³ is a typical detector for use with operated with the following PTW detectors (refer to
webline
UNIIDOS . The check device type 48012 is Table 2):
available as a radioactive check device for the PTW
ionization chamber type 30010. This user manual
refers in general to that chamber and that check
device.
a
The DIADOS diagnostic detector (T60004) is available with the BNT connecting system only.
24 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Description of the Device
NOTE
The connecting systems TNC and BNT are elec-
trically identical and can be adapted to each oth-
er.
Only original PTW cables may be used as
interconnecting cable between detector and
webline
UNIDOS .
The complete length of the cables used is limited
to 100 m.
D776.131.00/07 en 25
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Putting the Device into Operation
WARNING
webline
Before putting the UNIDOS into operation you
Improper handling of rechargeable batteries.
should insert the provided rechargeable batteries
Bodily injury! into the device and charge them.
If electrolyte leaks from the battery, avoid contact
with skin or clothes. Immediately clean skin or Therefore remove the lid from the battery com-
clothes that have come in contact with electrolyte partment on the back of the device.
with water. In case of eye contact, immediately
rinse eyes with water and consult doctor. CAUTION
Do not open rechargeable batteries. Mounting the rechargeable batteries with errone-
ous polarity.
Equipment damage!
WARNING
Please absolutely pay attention that all
Improper handling of rechargeable batteries. rechargeable batteries are mounted in correct
Risk of Fire or Skin Burns! direction (refer to Figure 4). Otherwise the
webline
Never short-circuit the rechargeable batteries. UNIDOS will be damaged.
If all rechargeable batteries are mounted in erro-
neous direction, you will furthermore hear a
CAUTION warning signal.
Improper handling of rechargeable batteries.
Equipment damage!
Insert the rechargeable batteries. For the num-
Do not open rechargeable batteries. Never ex- ber of batteries and the battery polarity please
pose rechargeable batteries to water. Store and refer to Figure 4.
use the rechargeable batteries at the ambient
conditions (temperature, relative humidity) speci- Close the battery compartment.
fied for the device.
26 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Putting the Device into Operation
NOTE
New rechargeable batteries will reach their full
capacity only after several charging and dis-
charging cycles.
webline
Then the UNIDOS will be ready for opera-
tion.
D776.131.00/07 en 27
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Putting the Device into Operation
webline
The UNIDOS commences in shutdown
mode.
28 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Quickstart (Workflow)
3 Quickstart (Workflow)
3. Await the stabilization time of 15 min 8. Adjust correction mode and correction factors,
section 6.4
4. Perform a zeroing: ZERO key
Enter detector parameters in the detector da- 10. Hold measurement: HOLD key
tabase, section 5.2 (If you have bought de-
webline
tectors together with the UNIDOS , the 11. Reset measured values and measuring
parameters of these detectors are already time: RESET key
entered in the database.)
Select detector, section 5.2.2
Adjust system parameters, sections 5.4 and
5.5
– date and time
– device name*
– language*
– interfaces*
1a 1b 2 1c 1d 3
6
5
7a 7b 8 7c
webline
Figure 5: UNIDOS measuring window
1 Status line with 6 Display area for hints and error messages con-
a measurement status cerning measured values
30 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Basic Principles of Operation
The indication of the measured values and the bar Editable menu items appear in blue print. A blue bar
graph as well as the end of bar graph scale can be cursor identifies the selected menu item.
adjusted in setup menu (refer to section 5.3.2).
The correction parameters are only displayed, if in A function is initiated by activation of the corre-
setup menu the corrections are enabled (refer to sponding control key (key to the right of the function
section 5.3.3) and if the indication of measured key).
value is set to radiological units (refer to sec- In the example above, this means:
tion 5.3.2). you change the value with INTEG
you close the menu window with ZERO .
3 4
webline
Figure 6: UNIDOS menu window (example)
D776.131.00/07 en 31
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Basic Principles of Operation
32 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Basic Principles of Operation
4.2.4 Power Key and Operation Modes Functions of the Power LED
webline
UNIDOS is turned on and off with the Power Mode LED
webline
key on the front panel. If UNIDOS needs to be
disconnected from the power line (e.g., when it will shutdown – mains power on
not be used for some time), turn it off with the on-off shutdown – battery power off
switch on the back.
standby – mains power blinking slowly
These are the operating modes of the device: standby - battery power blinking slowly,
brief impulses
OFF or shutdown
All assemblies of the device are switched off. standby – battery power, blinking fast
The shutdown mode is activated by pressing the battery very low
Power key for about 2 seconds.
NOTE
Standby
webline
The backlighting of the display is turned off and During battery power operation UNIDOS will
all keys except the Power key are inactive. All be completely turned off when the batteries are
other assemblies and interfaces are still on, al- discharged.
lowing the device to be controlled via the inter-
faces.
In this operating mode the operating tempera-
ture of the measuring amplifier is maintained
and there will be no warm-up time when the de-
vice is activated from standby.
Push the Power key briefly to activate the
standby mode.
ON
All assemblies are switched on. Normal operat-
ing mode. Press the Power key briefly to switch
webline
UNIDOS from standby or shutdown to the
normal mode.
webline
When UNIDOS is connected to power, i.e., the
power cord is connected and the on-off switch is
turned on, it will first enter the shutdown mode.
D776.131.00/07 en 33
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Basic Principles of Operation
4.3.1 General Notes You are in the measuring window or in the menu
where you want to edit a value.
webline
All editable UNIDOS values appear in blue
print. The user rights determine the scope of edita- Select and activate the numeric value to edit by
ble values. rotating and pressing the navigation knob.
We distinguish between three kinds of values: An entry window as shown below will appear:
numeric values
A numeric value may consist of multiple numeric
fields, i.e., date, time or calibration factor (con-
sisting of the factor and the exponent)
NOTE
The program has preset ranges for parameters
that are entered in numeric form (e.g. the tem-
perature can be entered between + 10° C and
+ 40° C). The program will not accept values
outside this range.
Figure 7: Entry window for adjustment of numeric
values
alphanumeric values (Example: Calibration factor)
values from option lists Position the cursor on a digit to change. Press
Help text explain each item of the option list. the HOLD key to move the cursor to the right
and the RESET key to move it to the left.
NOTE
Rotate the navigation knob until the desired digit
When controlling the program via VNC (see sec-
is displayed.
tion 10.3) data can be entered by means of the
PC keyboard and the mouse. In the case of values that consist of multiple
fields, press the navigation knob to advance to
the next field.
34 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Basic Principles of Operation
4.3.3 Editing Alphanumeric Values To toggle between capital and small letters
(CAPS function) press the MEAS key with the
You are in the measuring window or in the menu following options:
where you want to edit a value.
CAPS Abc function:
Select and activate the alphanumeric value to The character following a blank is shown in
edit by rotating and pressing the navigation capital letters and all other characters are
knob. shown in small letters.
CAPS ABC function:
An entry window as shown below will appear: All following characters are shown in capital
letters.
CAPS abc function:
All following characters are shown in small
letters.
To delete a character:
press the HOLD key to delete the character
on the left side of the cursor.
Backspace function
D776.131.00/07 en 35
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Basic Principles of Operation
36 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Settings
5 Settings
5.1 General Notes The setup main menu is divided into the following
areas:
webline
The UNIDOS allows you to adjust certain de-
Detector:
vice and detector characteristics with the help of the
Detector and detector database settings
setup menu. These settings are saved non-volatile
in the device, so they remain unchanged even after Measurements:
turning the device off. Settings for measuring and displaying the
measuring results
Press the SETUP key to display the setup System:
menu.
basic device settings
Miscellaneous:
miscellaneous device settings
WARNING
Improper handling.
Loss of data!
webline
Do not switch off the UNIDOS immediately
after changing parameters in the setup menu.
webline
UNIDOS needs about 1 s to store the
changed data internally.
Figure 10: Setup menu
NOTE
For changing individual values please refer to
section 4.3 – Changing Parameters.
D776.131.00/07 en 37
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Settings
38 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Settings
NOTE
The operation voltage as well as the rated range
of use of high voltage (HVmin ... HVmax) can be
found in the user manuals or the data sheets of
the detectors.
D776.131.00/07 en 39
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Settings
Radiation quality to which the calibration factor Serial number of the source
refers
Serial number of adapter
Calibration laboratory
Check reading kp,0
Calibration date
Check date
Calibration temperature
Date for next calibration of the detector
NOTE
The data of the radioactive check device can be
NOTE found in the calibration certificate.
The calibration data of the detector can be found
in the calibration certificate.
The following data for detector test can be stored in
the database in each detector data record:
Furthermore the following data of the radioactive
Lower limit for detector test result
check device can be stored in the database in each
detector data record: Upper limit for detector test result
NOTE
If No check device is selected, the following input
boxes for the check device are not shown.
Half-life period
14 137
The half-life periods of the isotopes C, Cs
90
and Sr are fixed. For user-defined isotopes,
however, the half-life period can be adjusted.
40 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Settings
Activate the Edit detector database menu item. Select by means of the navigation knob the de-
tector to delete and press the RESET key.
Editing a Detector
Copying a Detector
D776.131.00/07 en 41
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Settings
5.3 Settings for Measurements 5.3.2 Adjusting the Display of the Measuring
Window
42 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Settings
Indication of measured value in measuring win- Final scale value of the bar graph
dow in the measuring quantities At 0.0, the final scale value of the bar graph cor-
- Dose resp. charge responds to the end of the measuring range. In
- Dose rate resp. current or the measuring window the final scale value ap-
- Dose and dose rate resp. charge and current pear on the right of the bar graph (see Figure 5).
D776.131.00/07 en 43
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Settings
Activate the Enable/disable corrections menu Activate the Enable/disable statistics menu item.
item.
NOTE
If corrections are disabled the corresponding
data are not shown in the measuring window
(refer to Figure 5).
44 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Settings
5.3.6 Adjusting Alarm Thresholds Dose measurement threshold to switch the trip
output interface
Activate the Set alarm thresholds menu item.
Enable or disable alarm monitoring, i.e. sending
of alarm e-mails
Dose measurement threshold
D776.131.00/07 en 45
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Settings
Device name
Type number of the device
Hardware revision
Firmware version
Figure 20: Basic device settings area of the setup
Date of last device calibration
menu
Due date of next device calibration
Date and time of last device test
5.4.1 Date and Time
Result of last device test
Activate the Set system date/time menu item.
Measuring range in which the device test is exe-
Adjust the system date and the time. cuted (not changeable)
Zeroing
5.4.2 Device Name Detector test to check the whole measuring sys-
tem including the detector
Activate the Set device name menu item.
Device test to check the device without detector
Adjust the desired device name. You can enter connected
each user-defined name.
46 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Settings
NOTE
5.5.1 Language Settings changed in the Configure interfaces
menu are not accepted until a restart of
Activate the Set language menu item. UNIDOS
webline
.
D776.131.00/07 en 47
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Settings
NOTE
For detailed information on the Ethernet inter-
face, VNC and utilization of e-mail please refer to
section 10.
For more information on setting the IP addresses
and the POP3 mailbox please contact your net-
work administrator.
48 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Settings
Activate the Change passwords menu item. Activate the Set brightness (mains operation)
menu item or the Set brightness (battery opera-
Select by means of the navigation knob the us- tion) menu item.
er, Expert or Administrator, whose password you
like to change and confirm with the INTEG key. A slider is displayed to adjust the display bright-
ness in mains operation mode respectively bat-
Enter the existing password and confirm with tery operation mode.
the INTEG key.
Select by means of the navigation knob the de-
Enter the new password and confirm with sired brightness and confirm with
the INTEG key. the INTEG key.
Repeat the new password and confirm with If the SETUP key is held depressed longer than
the INTEG key. 5 s, the brightness settings for mains and battery
operation will be reset to the defaults.
The password is changed.
NOTE
The Expert user is only authorized to change the
expert password. The Administrator can change
the expert and the administrator password.
NOTE
If you have forgotten one of your passwords,
please contact your PTW distributor.
D776.131.00/07 en 49
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Measuring
6 Measuring
webline
After switching on the UNIDOS needs a stabi- NOTE
lization time of 15 min. At the end of this period the
If you use an external HV source you first have to
offset current is sufficiently stabilized to carry out
remove the connected detector and then to con-
measurements with highest accuracy.
nect the external HV source according to sec-
tion 6.3.
NOTE
webline
If UNIDOS at operating state temperature is
switched off for a short time a stabilization time of The measuring window is displayed.
at least 15 min has to be observed.
Usually there is no reason to switch off the
webline
UNIDOS at operating state temperature for
a short time, because the power cable may be
webline
disconnected even while the UNIDOS is
switched on. The built-in batteries take over the
power supply without any delay.
webline
Figure 23: UNIDOS measuring window
50 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Measuring
Figure 24: Select detector menu window The measuring window is displayed.
Select and activate by means of the navigation Select and activate by means of the navigation
knob the desired detector. knob the HV menu item.
The measuring window is again displayed and The HV (High voltage and polarity) input window
shows the most important parameters of the se- is displayed.
lected detector.
webline
UNIDOS checks automatically for which
quantity the detector is calibrated, provides the
appropriate calibration factors and sets the high
voltage to the value stored in the database.
NOTE
If the new operating voltage is higher than the
currently selected high voltage, a message to
this effect will be displayed and you can cancel
the selection of a detector. This prevents that the
voltage applied to the connected detector is too Figure 25: HV (High voltage and polarity) input
high. window
D776.131.00/07 en 51
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Measuring
52 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Measuring
D776.131.00/07 en 53
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Measuring
54 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Measuring
NOTE
CAUTION
In order to enter corrections they have to be acti-
Improper operation. vated in setup menu (refer to section 5.3.3) and
Equipment damage! the display have to be set to radiological units
Before removing the external HV source and (refer to section 5.3.2).
activating the internal HV source the external
voltage has to be set to 0 V in order to avoid high webline
UNIDOS offers two air density correction
potential jumps.
methods:
Switch the external HV source off. Air density correction kP/km by measuring the
check reading when using a radioactive check
Remove the external HV source from the device
webline
UNIDOS . or the correction by means of a single user-defined
correction factor, refer to section 6.4.4.
Remove the currently connected detector from
webline
the UNIDOS .
NOTE
Select the desired detector for internal high volt- Air density correction T&P is only available for
age according to section 6.2.1. vented detectors.
NOTE
The correction factors as well as the correction
date and time remain stored for approximately
webline
30 min after switching off the UNIDOS .
D776.131.00/07 en 55
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Measuring
56 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Measuring
NOTE
The air pressure that must be entered is the real
air pressure at the measuring point and it is not
the air pressure corrected to sea level.
(273.2 T) P0
k TP
(273.2 T0 ) P
T temperature in measuring volume [°C]
P air pressure at the measuring point [hPa]
T0 reference temperature 20 °C or 22 °C
P0 reference air pressure 1013.25 hPa
webline
The UNIDOS has to be adjusted to the calibra-
tion temperature 20 °C or 22 °C according to tem-
perature noted on the detector calibration certifi-
cate. Adjust the calibration temperature in the Edit
current detector menu (refer to section 5.2.3).
D776.131.00/07 en 57
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Measuring
6.4.3 Air Density Correction kp/km by Adjust the measuring range Low with
Measuring the Check Reading the RANGE key.
NOTE
webline
UNIDOS calculates from the value k p,0 en-
tered in Detector edit menu the value kp for the Figure 28: Correction input window
actual date according to the radioactive decay.
Select and activate by means of the navigation
knob the correction mode kp/km.
58 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Measuring
webline
UNIDOS calculates the check reading k p to The kp/km Correction input window is displayed.
expect and displays it, along with the menu item
km, in the quickinfo window of the measuring
window. If the program is unable to calculate the
check reading kp, a message to this effect will be
displayed together with instructions for correct-
ing the problem.
NOTE
Take care that the serial numbers of the detector,
the check device and of the adaptor itself corre-
spond to the serial numbers indicated in the cali-
bration certificate of the detector. Take care that
webline
the UNIDOS is adjusted to these parame-
ters (refer to section 5.2.3).
Figure 31: km measurement message window
D776.131.00/07 en 59
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Measuring
NOTE
The correction factor kp/km is only valid for the
currently selected detector and is only saved for
this detector.
If another detector is selected and the first one is
reselected afterwards, the pertinent correction
factor kp/km will automatically be displayed again.
If another correction mode is selected for this
detector or if the correction is completely turned
off and the correction mode kp/km reselected
afterwards, another km measurement must be
performed to determine the correction factor
kp/km anew.
60 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Measuring
6.4.4 Correction by a Single User-defined Select and activate by means of the navigation
Correction Factor knob the CorrFactor input field in the measuring
window.
Instead of an air density correction, a user-defined
correction factor can be used to correct the measur- Enter the desired correction factor (refer also to
ing value. section 6.4.5) and activate it.
D776.131.00/07 en 61
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Measuring
webline
6.4.5 Additional Corrections UNIDOS will calculate the product of all correc-
tion factors CorrFactor and sends this value to the
In addition to the air density correction factor you measuring window when the correction menu is
can take into account other correction factors. closed. All measured values will be multiplied with
Select and activate by means of the navigation the CorrFactor before being displayed.
knob the CorrFactor input field in the measuring
window. NOTE
The correction factors k Q, kPol, kS and k are only
The Correction input window is displayed. valid for the currently selected detector and are
only saved for this detector.
If another detector is selected and the first one is
reselected afterwards, the set correction factors
will automatically be displayed again.
If the correction is turned off and on again for this
detector, the correction factors k Q, kPol, kS and k
will have to be set anew.
62 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Measuring
Autoreset
When Autoreset is also enabled, all measured val-
ues will be reset to zero when radiation is reapplied.
After the reset, measurement of the new dose or
charge will start again.
D776.131.00/07 en 63
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Measuring
The Autostart and Autoreset functions as well as 6.7 Reading the Measured Value
the sensitivity for the Autostart function can be se-
lected in the setup menu (Menu Enable/Disable The measured value is displayed without a mes-
AutoStart, refer to section 5.3.4). sage beside and the bar graph is blue:
64 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Measuring
To improve the resolution, select the next The message LOW SIGNAL described in section
lower measuring range with 6.7 indicates that the resolution of the displayed
the RANGE key. measuring value is below 0.5 %. Please note that
the resolution is assessed on the basis of the corre-
If in autostart mode the message LOW AUTO sponding current or charge measurement. Conse-
SIG appears beside the measured value and the quently, a dose or dose rate reading in radiological
bar graph is yellow, this indicates that the meas- measuring quantities may appear to be indicated
ured value is not adequately resolved to obtain with a resolution far better than 0.5 %.
repeatable measuring results.
For the lower range limits of charge and current
To improve the resolution, select the next with a minimum resolution of 0.5 %, refer to Table 3
lower measuring range with in section 12.1. For readouts in radiological units
the RANGE key. the lower range limit depends on the detector used
(calibration factor).
If the measuring range is exceeded, the mes-
sage OVERLOAD appears beside the measured Example:
value and the bar graph changes to red. An ionization chamber with a volume of 0.6 cm³ and
a calibration factor of about 5 10 Gy/C is exposed
7
Select the next higher measuring range with
the RANGE key. to a dose rate that is equivalent to a current of
0.200 pA. The indication in radiological units is
0.560 Gy/min.
webline
Since the resolution of UNIDOS in the
measuring range Low is 0.001 pA, the current of
0.200 pA is measured with a resolution of 0.5 %.
The message LOW SIGNAL will therefore not be
displayed.
If the dose rate is now slightly reduced, the mes-
sage LOW SIGNAL will appear, because to current
drops to a value below 0.200 pA and will the resolu-
tion of 0.5 % no longer applies. The dose rate indi-
cation in radiological units (e.g. 0.500 mGy/min)
should not make the user believe that the resolution
of the measurement is far better than 0.5 %.
D776.131.00/07 en 65
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Measuring
Example:
Let us assume a resolution of 0.01 mGy for the
dose and of 0.001 mGy/s for the dose rate.
The numeric limit for a 0.5 % resolution of the dose
is 200 * 0.01 mGy = 2 mGy.
If a dose rate of 0.1 mGy/s (= 100 digits, i.e., reso-
lution of 1 %) is integrated over a period of 20 s, the
dose indication will be 2 mGy.
Nevertheless the message LOW SIGNAL will be
displayed, because the integration is based on a
dose rate with a resolution of only 1 %.
66 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Measured Value Statistics
σx
cv 100%
x
N N
2
N x
i1
i
2
( x )
i1
i
2
σx
Figure 36: Measuring window in statistics mode N(N 1)
D776.131.00/07 en 67
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Measured Value Statistics
68 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Measured Value Statistics
Select in setup menu the Signal-controlled sta- Select in setup menu the Time-controlled statis-
tistics mode and the number of measurements tics mode, the number of measurements, the
(refer to section 5.3.5). The Signal-controlled measuring time and the time for measuring
statistics mode can also be selected by selecting pause (refer to section 5.3.5). The time-
and activating by means of the navigation knob controlled statistics mode can also be selected
the statistics menu item in the measuring win- by selecting and activating by means of the nav-
dow. igation knob the statistics menu item in the
measuring window.
In the measuring window the Series
<commence> menu item will be displayed and In the measuring window the Series <com-
webline
UNIDOS will be switched to measuring mence> menu item will be displayed.
mode AutoStart with AutoReset.
Select and activate by means of the navigation
Select and activate by means of the navigation knob the Series <commence> menu item.
knob the Series <commence> menu item.
webline
UNIDOS will perform measurements ac-
webline
As soon as UNIDOS detects applied radia- cording to the adjusted measuring and pause
tion a measurement will be started automatical- and will add the measured values to the statis-
ly. When the signal is removed, the measured tics list.
value of dose or charge is added to the statistics
list. When radiation is reapplied, the measured The statistics may include at most 100 meas-
value will be reset to zero and a new measure- ured values.
ment will be started. If the number of measurements is set to 0
and the statistics list includes 100 values, the
The statistics may include at most 100 meas-
oldest value will be deleted adding a new
ured values.
value.
If the number of measurements is set to 0
If the number of measurements is set to a
and the statistics list includes 100 values, the
value unequal to 0 and the statistics list in-
oldest value will be deleted adding a new
cludes the adjusted number of values, the se-
value.
ries of measurements is finished.
If the number of measurements is set to a
value unequal to 0 and the statistics list in- Select and activate by means of the navigation
cludes the adjusted number of values, the knob the Series <finish> menu item for interrupt-
complete list will be deleted with the next en- ing or finishing the series of measurements.
try and that entry will be the first of the list
(restart of series of measurements). The
mean value and the relative standard devia-
tion will be automatically recalculated and
displayed
D776.131.00/07 en 69
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Measured Value Statistics
Select and activate by means of the navigation Press the INTEG key.
knob the List <edit> menu item in the measuring
The statistics list will be deleted after an
window.
acknowledgement.
The Edit statistics menu window will be dis-
The Edit statistics menu window is closed auto-
played:
matically after deleting all values.
70 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Error Messages and Problems
8.1 Error Messages on the window for warnings and error messages occur-
Display ring during measurement, zeroing, etc., e.g.:
D776.131.00/07 en 71
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Error Messages and Problems
The battery symbol Check the power cord and the connection
appears during to the power line. Make sure that the
mains operation on-off switch on the back is switched on.
LOW BAT The charge level of the batteries is approach- Connect the device to the power line and
webline
(with yellow back- ing the low limit. The UNIDOS is still fully switch it on by means of the on-off switch
ground) ready for operation and measurement for on the back of the device (mains opera-
about 15 minutes. After expiration of this peri- tion). In mains operation the batteries of
webline webline
od, the UNIDOS is switched off (shut- the UNIDOS are recharged. The
down) and must be connected to the power charge time is about 2 hours, overcharg-
line. ing is not possible.
72 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Error Messages and Problems
OVERLOAD The measuring range is exceeded. The input Switch to the next higher measuring
(with red back- amplifier is overloaded. This message is dis- range with the RANGE key.
ground) played separately for dose or charge respec-
tively dose rate or current.
NOTE
During dose rate or current measurement OVERLOAD is only displayed as long as
the overload continues. During a dose or charge measurement OVERLOAD is dis-
played from the beginning of the overload up to the end of the measurement (also if
the dose or current is again within the measuring range in the meantime). Thus,
integrated measured values gained under wrong conditions, are indicated.
After a strong overload of the measuring range Low it can take up to one minute till
the amplifier returns to the initial state again. This time can be shortened considera-
bly by switching to the measuring range High for a few seconds.
VNC The message VNC appears when The message disappears automatically
webline
UNIDOS communicates via VNC. when the VNC communication is termi-
This message is intended as an nated.
information to users communicating
directly.
D776.131.00/07 en 73
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Error Messages and Problems
Zeroing not possible. The waiting period of 5 min was not Observe the waiting period of 5 min and
appears after pressing observed. repeat the zeroing.
the ZERO key. The detector was irradiated or moved Repeat the zeroing.
during zeroing.
The detector or the connection cable is Try to carry out a zeroing without cable
faulty. and detector connected.
The amplifier was exposed to high hu- Replace the desiccant (refer to sec-
midity for some time, the desiccant is tion 17).
used up.
Error in the system test. The measuring range was exceeded. Refer to OVERLOAD message, sec-
Test is canceled. appears tion 8.1.3.
during device or detector
test.
webline
Error message from POP 3 UNIDOS is unable to check the Check the network cable and the con-
server. E-mail could not be POP3 mailbox for e-mails, because the nection to the network.
retrieved. communication with the network was Check the settings in the Configure in-
interrupted or the wrong parameters terfaces menu and correct them, if re-
were entered for the e-mail check. quired.
After switching on the The non-volatile memory (EEPROM) of In this case resetting is not possible.
webline webline webline
UNIDOS shows the the UNIDOS contains invalid data. Please send the UNIDOS back to
message Error while load- the factory. The typical life time of the
ing system settings. Please EEPROM is about 10 years (refer also
contact your PTW repre- to section 19).
sentative. and further con-
trolling is not possible.
webline
Hardware Error The communication with the measuring Restart the UNIDOS .
Communication with the module is interrupted. Measurements If this message still appears after re-
measuring unit is interrupt- are no longer possible. starting several times you have to send
webline
ed. Measuring value dis- After exiting the message window the the UNIDOS back to the factory.
play is deactivated. Restart message Communication error with
the instrument (Shutdown). measuring unit. appears in the measur-
ing window.
webline webline
Hardware error. The UNIDOS has detected an error Please send the UNIDOS back to
The electrical measuring in the memory area for the electrical the factory for inspection or recalibra-
unit is not calibrated. calibration data. tion.
Measuring value display is After exiting the message window the
deactivated. message Electrical measuring unit un-
calibrated. appears in the measuring
window.
74 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Error Messages and Problems
kp/km out of range 0.5 - 2.0 System date is not set correctly. Check the system date in the setup
appears after measuring km menu (refer to section 5.4.1)
with the radioactive check During calibration another check device, Compare the serial numbers of
device (refer to sec- another detector or another adapter was - check device
tion 6.4.3) used as during km measurement. - detector
- adapter for ionization chamber (only
in the case of flexible chambers)
with the serial numbers in the calibration
certificate.
The data of the detector or the check Compare all detector and device check
device are not set correctly in the detec- data entered in the Edit detector menu
tor data record. (refer to section 5.2.3) with the parame-
ters in the calibration certificate.
The detector or the extension cord is Perform a km-measurement without ex-
faulty. tension cable.
Check the detector by means of a detec-
tor test (refer to section 9.3.1).
kp/km measuring not possi- The kp/km measurement was aborted by see HV ERROR or OVERLOAD mes-
ble. Measurement can- the measuring module. sages, section 8.1.3.
celed. The measuring range was exceeded or
a high voltage error has occurred
D776.131.00/07 en 75
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Error Messages and Problems
Fault Remedy
The indicated measured value is always Check whether the entered calibration factor for the selected de-
zero. tector is greater than zero.
Use another detector or another extension cable, if necessary.
webline
The indicated measured value is negative. The current fed into the UNIDOS is negative. The sign of the
current depends on the polarity of the high voltage.
You have the impression that the meas- Check the type and serial numbers of the detectors connected and
webline webline
ured value shown by the UNIDOS is compare them with the data stored in UNIDOS (Edit detector
wrong or inaccurate. menu, refer to section 5.2.3).
Check whether the message LOW SIGNAL appears beside the
measured value. This indicates insufficient resolution of the meas-
ured value (refer to section 6.7 and 6.8).
webline
Check whether the UNIDOS shows the correct high voltage.
Compare measuring quantity, unit and calibration factor displayed
webline
on the UNIDOS with the data given in the calibration certifi-
cate.
Check the correction factors (refer to section 6.4).
Check by a blank measurement (i.e. with a non-irradiated detector
webline
the offset current of the UNIDOS . If required, carry out a zero-
ing with the ZERO key.
Check whether the high voltage is applied to the detector accord-
ing to section 9.3.
76 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Error Messages and Problems
D776.131.00/07 en 77
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Operating Manual – Error Messages and Problems
78 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Technical Manual – Checking the UNIDOSwebline
Technical Manual
9.1 Offset Current (Zero Drift) For checking the offset current (zero drift) initiate a
and Zeroing charge or dose measurement over the intended
measuring time or check the current or dose rate.
webline Compare the offset current (zero drift) with the
The offset current (zero drift) of the UNIDOS
checked with a detector not irradiated is caused by measured values which you expect when the detec-
leakage currents tor is irradiated.
in the detector cable, the extension cable, or Press the ZERO key.
cable connections.
The zeroing is executed and takes about 74 s.
Typical values for the leakage current of the ampli- During zeroing the following message window
fier (offset current) can be taken from the technical with progress indication is displayed.
data (section 12). Typical values for the leakage
current of a PTW ionization chamber type 30010
can be found in the appropriate user manual.
Preconditions:
The UNIDOS
webline
has to be switched on at least
15 min. Then the device is operational.
The detector has to be connected with the ex-
tension cable used for the measurements.
NOTE
During the zeroing neither the detector nor the
detector or extension cable should be moved.
D776.131.00/07 en 79
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Technical Manual – Checking the UNIDOSwebline
Abort the zeroing be pressing the RESET key. 9.2 Long-term Stability
In this case the electrical settings of the
webline webline
UNIDOS do not change. The long-term stability of the UNIDOS and the
detector used can be checked with a radioactive
If an automatic zeroing cannot be executed suc- check device.
cessfully:
Preconditions:
Connect the detector without extension cable
The UNIDOS
webline webline
directly to the UNIDOS and check whether has to be switched on at least
the zeroing now works. If yes, then the exten- 15 min.
sion cable is faulty.
The detector has to be connected with the ex-
Check, whether an automatic zeroing is possible tension cable used for the measurements.
without an extension cable or a detector con-
With high voltage applied to the detector a set-
nected. If yes, then the detector is faulty or is
ting time of 15 min has to be awaited.
showing isolation errors, e.g. due to too much
humidity. Air density correction has to be performed by
entering temperature and air pressure.
The amplifier was exposed to high humidity for
some time, the desiccant is used up. Replace Check the long-term stability as follows:
the desiccant (see section 17).
Check the offset current according to sec-
tion 9.1.
80 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Technical Manual – Checking the UNIDOSwebline
Enter the temperature and air pressure for air 9.3 High Voltage and Amplifier
density correction. The thermometer of the
check device provides the required temperature. Using the system test feature which comprises the
detector test and the device test, you can verify that
Start an integration measurement for an integra- the correct high voltage is applied to the detector
tion time of at least 60 s by pressing webline
and that the UNIDOS amplifier is operating
the INTEG key. The integration time can be correctly.
adjusted in the setup menu (refer to section
5.3.1). Both tests are performed in the Low measuring
range. The test results are indicated in electrical
Provided the radioactive decay is correctly con- units.
sidered, the standard deviation of the measuring
values for 1 year should be less than 0.5 %. Repeat the system tests, i.e. the detector test and
the device test, at regular intervals which are suita-
ble for you. Record the test results.
NOTE
Using the optional electrical check device
UNITEST, you can check all amplifier stages
(Low, Medium and High) for electrical constancy.
For more information on the UNITEST and on
webline
the expected indication value at UNIDOS
please refer to the UNITEST user manual.
D776.131.00/07 en 81
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Technical Manual – Checking the UNIDOSwebline
During system tests the detector must not be 9.3.1 Detector Test
irradiated.
NOTE
You can perform a detector test only if the detec-
tor settings match the following conditions:
Perform the system tests as follows:
– HVmax ≠ 0 V
– HV source = internal
Adjust the UNIDOSwebline to the detector to be
checked.
Start the detector test by pressing
Connect the detector without extension cable to
webline the RESET key.
the UNIDOS .
A message window is displayed indicating the
NOTE detector used and the high voltages adjusted for
Polarity and absolute value of high voltage are this detector.
not important for this test.
Check whether the correct detector is connected
and confirm by pressing the INTEG key.
Check the offset current according to sec-
webline
tion 9.1. UNIDOS sets automatically all parameters
needed for this detector, performs a measure-
Press the SETUP key to display the setup ment and displays the test result.
menu.
The result of the detector test is stored in the
Select and activate the Perform system test detector data record together with date and time
menu item (refer also to section 5.4.4). (refer to section 5.2.3).
The System test input window is displayed. Exit the detector test by pressing
the ZERO key.
82 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Technical Manual – Checking the UNIDOSwebline
Start the device test by pressing Limits are preset for both the device test and the
the INTEG key. detector test. The test results must not exceed
these limits.
A message window pops up prompting you to
PTW-Freiburg determines the limits for the de-
remove the detector and showing the high volt-
vice test. These values cannot be changed by
age selected for the device test.
the user.
Disconnect the detector and confirm by pressing The user, however, is free to determine the lim-
the INTEG key. its for the detector test in the detector’s data
webline
record (see section 5.2.3). For determination of
UNIDOS sets automatically all parameters the limits for the detector test please refer to
needed for the device test, performs a meas- section 9.3.4.
urement and displays the test result.
If the test results are outside these limits, a mes-
The result of the device test is stored in the sys- sage will inform you that the test failed and no test
tem information together with date and time (re- results will be saved. These are some of the poten-
fer to section 5.4.3). tial reasons:
Press the ZERO key twice to exit the system The UNIDOS
webline
high voltage source or the
test and the setup menu. amplifier is defective. The device must be re-
turned to PTW-Freiburg for repair.
webline
UNIDOS returns to the measuring window
If only the detector test failed, the connected
with the parameters and high voltage used be-
detector is damaged (e.g. broken electrode).
fore.
D776.131.00/07 en 83
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Technical Manual – Checking the UNIDOSwebline
9.3.4 Determining the Limits for the Detector 9.4 Polarity Effect
Test
The polarity effect leads to a different reading at
The test results of the detector test depend heavily positive and negative high voltages. The effect de-
on the detector used and the connecting system. pends on the radiation quality and energy.
For this reason PTW-Freiburg cannot predefine the
limits for the detector test but the user has to de- Preconditions:
termine them. At this PTW-Freiburg recommends
The UNIDOS
webline
proceeding as follows: has to be switched on at least
webline
15 min. The UNIDOS must not be switched
Prior to the first use of a new detector, perform off until the complete measuring procedure has
several detector tests and note each test result. been finished.
This is to determine the margin of deviation of The polarity effect can be measured with
webline
the detector test. UNIDOS as follows:
84 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Technical Manual – Checking the UNIDOSwebline
D776.131.00/07 en 85
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Technical Manual – Ethernet Interface
10 Ethernet Interface
86 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Technical Manual – Ethernet Interface
webline
10.3.2 Connecting to UNIDOS via VNC WARNING
In the Setup menu under Configure interfaces Improper handling.
(see section 5.5.2) activate the VNC server. Erroneous Measurement!
PTW-Freiburg is not in a position to guarantee
Download the UNIDOS webline VNC Viewer the trouble-free performance of VNC viewers
from the PTW-Freiburg homepage and install it other than the UNIDOS webline VNC Viewer.
on your PC.
Initiate the communication by clicking the Con- Parameters are selected with the cursor
nect button. keys and instead of rotating the navigation
knob.
If all parameters were correctly entered, the Selections are confirmed and activated with the
display will now show a window with the current ENTER key instead of pressing the navigation
webline
screen image of UNIDOS . knob.
D776.131.00/07 en 87
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Technical Manual – Ethernet Interface
1 2 3 4
5 6 7
webline
1 Display of the UNIDOS measuring window in the VNC viewer
2 Device control keys in the VNC viewer
3 Range control key in the VNC viewer
4 Displays the setup menu in the VNC viewer
webline
5 Field for entry of the UNIDOS IP address
6 Key to establish/terminate the VNC connection
7 Key to close the VNC viewer
88 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Technical Manual – Ethernet Interface
10.4.1 General Notes In the subject line of your e-mail, send one of the
following commands (STATUS, LIBRARY,
webline webline
E-mail communication with UNIDOS can be SYSTEMTEST or SERVICE) to UNIDOS .
done by the following e-mail classes:
Please contact your system administrator to obtain The subject line of the answer e-mail always reads:
the POP3 and SMTP server addresses and the "device name, COMMAND"
data for the POP3 mailbox.
The answer e-mail always begins with "UNIDOS
Before you can send alarm e-mails, the following webline COMMAND date and time" and general
parameters need to be set in the Setup menu under system information, such as device name, type
Set alarm thresholds (also refer to section 5.3.6): number, serial number, firmware version, etc.
activate the alarm monitoring
The following commands can be sent to
set alarm thresholds for dose and dose rate UNIDOS
webline
via e-mail:
set alarm e-mail address
STATUS
webline
UNIDOS will return the following information in
the answer e-mail:
correction data
such as correction status, mode, factors, etc.
D776.131.00/07 en 89
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Technical Manual – Ethernet Interface
all data of the selected detector and the set high 10.4.3 Alarm E-Mails
voltage
all interface settings WARNING
Improper handling.
important system settings
such as user, language, date and time, etc. Erroneous Measurement!
The alarm e-mail function is not suitable for ap-
The passwords will not be sent.
plications relevant for system safety, because the
LIBRARY permanent availability of the network and timely
UNIDOS
webline
will send all parameters of all stored arrival of alarm mail message at the addresses
detector data records in the answer e-mail. cannot be guaranteed.
STATISTICS
webline When one of the alarm thresholds is exceeded,
UNIDOS will return the following information in webline
UNIDOS will automatically send an alarm
the answer e-mail:
e-mail to the set alarm e-mail address.
settings for statistics such as statistic mode,
measuring time, time for measuring pause and The following faults will also trigger an alarm e-mail:
number of measurements – HV ERROR
– OVERLOAD
results of statistics such as number of measured
values, mean value and relative standard devia-
The subject line of an alarm e-mail always reads:
tion
device name, ALARM
all measured values of the statistics as well as
the appropriate information such as unit, adjust- The alarm e-mail always begins with "UNIDOS
ed high voltage, date and time of the measure- webline ALARM date and time" and general system
ment, integration time and error messages information, such as device name, type number,
serial number, firmware version, etc.
SYSTEMTEST
UNIDOS
webline
will send all data of the device and An alarm e-mail provides the following additional
detector test, such as test results, date and time of information:
the tests, etc. in the answer e-mail. alarm monitoring data
such as cause of alarm, set alarm threshold, etc.
SERVICE
UNIDOS
webline
will return the following information in data of the current measurement
the answer e-mail: such as measurement status, range, values,
time, etc.
supply voltages, test voltages
90 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Technical Manual – Ethernet Interface
NOTE
webline
The UDP port number for UNIDOS is 8123.
NOTE
webline
Commands for controlling UNIDOS via UDP
or the RS232 interface are described in a sepa-
rate user manual.
D776.131.00/07 en 91
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Technical Manual – Trip Output (Relay Output)
webline
The UNIDOS is equipped on the back with a Configuration
relay output TRIP OUTPUT functioning as a two-
way contact. The trip output may be used to switch Connector: 9 pin D-Sub- plug (with pins)
an external signal when the absolute value of the Trip output: (refer to Figure 41)
measured dose is exceeding the preset value. 2 two-way contacts, activated togeth-
er (pins 3, 7 and 8 are not assigned)
WARNING outputs are electrically isolated
Improper handling of the trip output interface. plug housing is connected to the
Patient Hazard! earthed conductor (PE)
WARNING
Improper handling of the trip output interface. 2
Bodily Injury! 6
92 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Technical Manual – Trip Output (Relay Output)
In order to be able to use the trip output you have to When the device is switched off, after switching
set the following parameters in the Setup menu on the device and after a reset of the device the
under Set alarm thresholds (also refer to sec- relay contacts are switched as shown in Fig-
tion 5.3.6): ure 41 (contacts between pin 5 and 9 resp. pin 2
and 6 are closed).
activate the trip output
set the switch off level trip dose for dose meas- When starting a measurement with
urement the MEAS key or the INTEG key the relay
contacts will be switched over (contacts between
NOTE pin 9 and 4 resp. pin 6 and 1 are closed).
The trip output cannot be enabled in Autostart
During measurement the relay contacts will be
mode.
switched over (contacts between pin 5 and 9
resp. pin 2 and 6 are closed) when the absolute
value of the measured dose is exceeding the
switch of level.
D776.131.00/07 en 93
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Technical Manual – Trip Output (Relay Output)
Switching Delay
CAUTION
Improper handling.
Erroneous Measurement!
Due to the mentioned switching delay the abso-
lute amount of the applied dose may be larger
than the switch off level.
94 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Technical Manual – Technical Specifications
12 Technical Specifications
Only values for which tolerances or limits are specified are guaranteed. Values for which no tolerances are
specified are for information only.
Meas. Lower limit of the measuring range Upper limit of the Digital resolution
range measuring range
Table 3: Digital resolution and measuring ranges for charge and current measurement
The dose and dose rate values are rounded values. They refer to a PTW ionization chamber with a volume of
0.6 cm³ and a typical calibration factor of 5 10 Gy/C, calibrated for absorbed dose to water.
7
D776.131.00/07 en 95
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Technical Manual – Technical Specifications
a
For laboratory use only
96 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Technical Manual – Technical Specifications
webline
Device designation Dose and dose rate measuring device PTW-UNIDOS ,
type 10021 (M connecting system),
type 10022 (TNC connecting system) and
type 10023 (BNT connecting system)
Sampling rate with integration via 1 kHz
dose rate or current
Max. measuring time for dose meas- 100 h
urement
Adjustable integration time (1 ... 9999) s
High voltage (0 ... 400) V, adjustable in 1 V steps
Max. deviation of high voltage 1V
Accuracy of the ratio of two high 2%
voltages
Limitation of the current supplied by typ. 200 µA by semiconductor;
b
high voltage max. 300µA by resistor
Measuring ranges Refer to Table 3, section 12.1
0.25 %
c
Repeatability
Measuring range LOW: 1 fA
b
Offset current (zero drift)
Measuring range MED: 50 fA
Measuring range HIGH: 5 pA
Zeroing automatically in 74 s
Non-Linearity in both measuring 0.25 %
bd
modes
< 0.1 % per year
b
Long term stability
b
Response time < 1.5 s
Rated range of use
Temperature (+ 10 ... + 40) °C
Rel. humidity (20 ... 80) %, no condensation (max. 20 g/m³ absolute humidity)
Atmospheric pressure (700 ... 1060) hPa
Maximum operating altitude 3000 m above sea level
Ambient conditions for transport and
storage (according to ISTA 2A)
Temperature (- 29 ... + 60) C
Rel. humidity (15 ... 85) %
Atmospheric pressure (700 ... 1060) hPa
b
The output current is limited and thus accessible according to IEC 61010-1 (not HAZARDOUS LIVE).
c
according to IEC60731
d
Expanded standard uncertainty (k=2) of the electrical calibration is 0.5% or 32 fA, whichever is greater.
D776.131.00/07 en 97
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Technical Manual – Technical Specifications
webline
Device designation Dose and dose rate measuring device PTW-UNIDOS ,
type 10021 (M connecting system),
type 10022 (TNC connecting system) and
type 10023 (BNT connecting system)
Reference conditions
Temperature + 20 C
Rel. humidity 50 %
Stray radiation 0 µSv/h
Display 4-digit
Stabilization time 15 min
Dimensions (H W D) 152 mm 257 mm 262 mm
5.98 in x 10.12 in x 10.31 in
Weight 5.9 kg / 13.0 lbs
Lifetime of non-volatile memory typically 10 years
Power supply (100 - 240) V; (50 / 60) Hz
6 pieces rechargeable NiMH batteries (AA size) for operation inde-
pendent of mains for about 2 hours, PTW Order no. L389028
Power consumption during mains approx. 23 W
operation
Desiccator PTW Order no. L137004
External high voltage input ± 1000 V DC, floating (refer to section 1.3, page 21)
Ethernet interface IEEE 802.3, 10Mbit/s, TCP/IP protocol
98 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Technical Manual – Technical Specifications
D776.131.00/07 en 99
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Technical Manual – Technical Specifications
Connecting system M
The collecting electrode and the guard electrode
are connected to ground potential.
For positive high voltage (HV), the outer electrode
(= outer shield = external chamber wall) is positive
with respect to ground potential.
100 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Technical Manual – Electromagnetic Compatibility (EMC) according to IEC 60601-1-2
The product is intended for use in the electromagnetic environment specified below.
The customer or the user of the product should assure that it is used in such an environment.
RF emissions Group 1 The product uses RF energy only for its internal function.
CISPR 11 Therefore, its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment.
RF emissions Class B The product is suitable for use in all establishments, in-
CISPR 11 cluding domestic establishments and those directly con-
Harmonic emissions Class A nected to the public low-voltage power supply network
a
IEC 61000-3-2 that supplies buildings used for domestic purposes.
a
The product fulfils the requirements of CISPR 11 Class B. However, it is not intended for use in domestic establishments.
D776.131.00/07 en 101
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Technical Manual – Electromagnetic Compatibility (EMC) according to IEC 60601-1-2
The product is intended for use in the electromagnetic environment specified below.
The customer or the user of the product should assure that it is used in such an environment.
±1 kV ± 1 kV
for input/output for input/output
lines lines
Surges ± 1 kV ± 1 kV Mains power quality should be that of a
IEC 61000-4-5 line(s) to line(s) line(s) to line(s) typical commercial or hospital environ-
(symmetrical) ment.
± 2 kV ± 2 kV
line(s) to earth line(s) to earth
(unsymmetrical)
102 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Technical Manual – Electromagnetic Compatibility (EMC) according to IEC 60601-1-2
<5 % UT 0 % UT
(> 95 % dip in UT) (> 95 % dip in UT)
for 5 s for 5 s
b
Power frequency 3 A/m 30 A/m Power frequency magnetic fields should
(50/60 Hz) be at levels characteristic of a typical lo-
magnetic field cation in a typical commercial or hospital
IEC 61000-4-8 environment.
NOTE: UT is the a.c. mains voltage prior to application of the test level.
b
Test level from standard IEC 61326-1 was applied.
D776.131.00/07 en 103
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Technical Manual – Electromagnetic Compatibility (EMC) according to IEC 60601-1-2
The product is intended for use in the electromagnetic environment specified below.
The customer or the user of the product should assure that it is used in such an environment.
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by ab-
sorption and reflection from structures, objects and people.
c
Test levels from standard IEC 61326-1 were applied.
d
Calculation with test level from standard IEC 60601-1-2
104 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Technical Manual – Electromagnetic Compatibility (EMC) according to IEC 60601-1-2
a)
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which
the product is used exceeds the applicable RF compliance level above, the product should be observed to
verify normal operation. If abnormal performance is observed, additional measures may be necessary,
such as re-orienting or relocating the product.
b)
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [U 1] V/m = 10 V/m.
D776.131.00/07 en 105
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Technical Manual – Electromagnetic Compatibility (EMC) according to IEC 60601-1-2
The product is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the product can help prevent electromagnetic interference by main-
taining a minimum distance between portable and mobile RF communications equipment (transmitters) and
the product as recommended below, according to the maximum output power of the communications equip-
ment.
150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
3,5 3,5 7
d= P = 1.2 P d= P = 1.2 P d= P = 2.3 P
U1 E1 E1
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d
in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is
the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by ab-
sorption and reflection from structures, objects and people.
e
Calculation with test level from standard IEC 60601-1-2
106 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Technical Manual – Accessories and Spare Parts
D776.131.00/07 en 107
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Technical Manual – Accessories and Spare Parts
108 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Service Manual – Cleaning
Service Manual
WARNING 15 Cleaning
Improper execution of service tasks.
Hazards to Persons! Equipment Damage! General Information on Cleaning
Ensure that the service tasks described below
are only performed by specialized staff. WARNING
In addition, ensure that all other service tasks are Liquids have entered the device.
only performed by PTW-Freiburg or by a person
Shock Hazard, Erroneous Measurement!
authorized by PTW-Freiburg.
The device must be thoroughly inspected by
service personnel before being used again.
WARNING
Improper execution of service tasks.
CAUTION
Hazards to Persons! Equipment Damage!
Disinfection of the surface with disinfectants on a
Before carrying out any service tasks on the phenol base or peroxide compounds.
webline
UNIDOS , de-energize the device.
Equipment Damage!
webline
Do not open the housing of the UNIDOS .
Do not use disinfectants on a phenol base or
peroxide compounds to disinfect the surface.
D776.131.00/07 en 109
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Service Manual – Preventive Maintenance
webline
Device surfaces of the UNIDOS 16 Preventive Maintenance
Turn the UNIDOSwebline off and definitely pull out
the mains plug before you begin any treatment In the following you will find a list of the mainte-
of the device surfaces. nance measures and inspections necessary for the
product.
Wipe the UNIDOSwebline down. Liquids must not
be allowed to enter the product. Do not spray
liquids onto the product and connectors. Check before each use
The inner surfaces of the connector sockets Before each use, visually inspect the device and
must not be cleaned. the cables for signs of mechanical damage.
The detector plugs must not be soiled. The elimina- Only regularly inspected devices are fail-safe. To
tion of dust is best carried out with a soft brush (not preserve the functional security and operational
by blowing!). safety, a safety inspection is to be executed at
regular intervals according to national regulations.
110 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Service Manual – Desiccator Cartridge
Inspections must also be performed after repairs or If the relative humidity in the amplifier is less than
after each influence which may have changed the 40 %, the desiccant in the desiccator window is
behavior of the device. blue. The desiccant turns pink when the relative
humidity exceeds 40 % and should be replaced as
soon as possible.
Calibration
The desiccant has a minimum useful life of 5 years.
The exact calibration factor and supplementary data
Desiccator cartridges can be obtained from PTW
are included in the calibration certificate.
under the ordering number L137004.
webline
We recommend having the UNIDOS calibrated
Follow these steps to replace the desiccator car-
every 2 years at PTW-Freiburg or by specially
tridge:
trained staff on site.
Switch off the UNIDOSwebline with the Power key
D776.131.00/07 en 111
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Service Manual – Rechargeable NiMH Batteries
WARNING
Improper handling of rechargeable batteries.
Risk of Fire or Skin Burns!
Never short-circuit the rechargeable batteries.
CAUTION
Improper handling of rechargeable batteries.
Equipment damage!
Do not open rechargeable batteries. Never ex-
pose rechargeable batteries to water. Store and
use the rechargeable batteries at the ambient
conditions (temperature, relative humidity) speci-
fied for the device.
112 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Service Manual – Rechargeable NiMH Batteries
webline
The UNIDOS is equipped with rechargeable Follow these steps to check the charge level of the
NiMH batteries for a mains-independent operation rechargeable batteries:
of 2 hours. In battery operation mode the
UNIDOS
webline
is fully functional as long as the Disconnect UNIDOSwebline from the power line by
display with the measurement indication is working. disconnecting the power cord or with the on-off
When the rechargeable batteries are discharged, switch on the back.
the internal power supply is switched off and the
display goes out. About 15 minutes before the When you see the battery symbol, the charge
message LOW BAT appears in the upper left corner level is OK.
of the display.
If the message LOW BAT appears in the upper
The rechargeable batteries of the UNIDOS
webline
are left corner of the measuring window, the batter-
recharged while the UNIDOS
webline
is connected to ies are almost discharged and need to be re-
the power line and is switched on by the on-off charged.
switch on the back of the device. The charge time is
about 2 hours and is indicated by the battery CAUTION
symbol with changing charge. Improper handling of rechargeable batteries.
Equipment damage!
CAUTION webline
Do not operate the UNIDOS with disposable
Improper handling. batteries. Use only the special rechargeable
Malfunction! NiMH batteries.
webline
The indication of the charge level is correct only If the UNIDOS will not be operated on bat-
when charging was not interrupted, e.g., by dis- tery power for some time (several months), re-
webline
connection of UNIDOS from the power line. move the rechargeable batteries.
D776.131.00/07 en 113
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Service Manual – Rechargeable NiMH Batteries
CAUTION
Mounting the rechargeable batteries with errone-
ous polarity.
Equipment damage!
Please absolutely pay attention that all
rechargeable batteries are mounted in correct
direction (refer to Figure 43). Otherwise the
webline
UNIDOS will be damaged.
If all rechargeable batteries are mounted in erro-
neous direction, you will furthermore hear a
warning signal.
114 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Service Manual – Non-Volatile Memory
webline webline
The UNIDOS is equipped with a non-volatile The UNIDOS uses 2 microfuses
memory (EEPROM) for the storage of all calibra- (T 2 AL, 250 V) which are accessible from the back
tion factors and detector data. Upon power-up, the of the device.
system checks the memory contents for correct-
ness. If the memory contains invalid data, the fol- To replace the fuses proceed as follows:
lowing message appears when you switch on the
UNIDOS
webline
Error while loading system settings. Disconnect the UNIDOSwebline from the power
Please contact your PTW representative. line and switch it off.
At the end of its service life, the EEPROM may lose Disconnect the power cord from the
webline
the data. The component has a typical life of at UNIDOS .
webline
least 10 years. The UNIDOS must be returned
Lift the locking clip on the fuse holder using a
to the factory for replacement of the EEPROM.
small screwdriver and remove the fuse holder.
D776.131.00/07 en 115
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Service Manual – Disposal of the Product
webline
The typical lifetime of the UNIDOS is about
10 years.
webline
At the end of the product life the UNIDOS
components must be disposed of in compliance
with the applicable waste control regulations. The
different materials must be separated and recycled
as appropriate.
webline
UNIDOS does not contain hazardous materi-
als.
116 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Literature
Literature
D776.131.00/07 en 117
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Literature
[22] ISTA 2A
ISTA 2 Series Partial Simulation Performance
Test Procedure – Part 2A: Packaged-Products 150 lb
(68 kg) or less
118 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Appendix: China Electronic Industry Standard Compliance
The following product pollution control information is provided according to SJ/T11364-2006 Marking for Control
of Pollution caused by Electronic Information Products.
根据SJ/T11364-2006标准规定的《电子信息产品污染控制标识要求》特提供以下有关产品污染控制方面的信息。
This symbol indicates the product contains hazardous materials in excess of the limits estab-
lished by the Chinese standard SJ/T11363-2006 Requirements for Concentration Limits for Cer-
tain Hazardous Substances in Electronic Information Products. The number in the symbol is the
Environment-friendly Use Period (EFUP), which indicates the period during which the toxic or
hazardous substances or elements contained in electronic information products will not leak or
mutate under normal operating conditions so that the use of such electronic information products
will not result in any severe environmental pollution, any bodily injury or damage to any assets.
The unit of the period is "Year".
In order to maintain the declared EFUP, the product shall be operated normally according to the
instructions and environmental conditions as defined in the product manual, and periodic mainte-
nance schedules specified in Product Maintenance Procedures shall be followed strictly.
Consumables or certain parts may have their own label with an EFUP value less than the prod-
uct. Periodic replacement of those consumables or parts to maintain the declared EFUP shall be
done in accordance with the Product Maintenance Procedures.
This product must not be disposed of as unsorted municipal waste, and must be collected sepa-
rately and handled properly after decommissioning.
该标识表明本产品含有的有毒有害物质超出SJ/T11363-2006标准规定的《电子信息产品中有毒有
害物质的限量要求》。标识中的数字为本产品的环保使用期限(EFUP),表明在正常使用的条件
下,电子信息产品内含的有毒有害物质或元素不会发生外泄或突变,用户使用该电子信息产品不会
对环境造成任何严重污染或对人身、财产造成任何严重损害的期限。单位为年。
为保证所声明的环保使用期限,应按产品手册中所规定的指示和环境条件进行正常使用,并严格遵
守产品维护程序中规定的定期维护和保养日程。
产品中的耗件或某些零部件可能具有单独的标识,其环保使用期限有可能短于产品本身的环保使用
期限。应按产品维护程序定期更换这些耗件或零部件,以保证所声明的整个产品的环保使用期限。
本产品在使用寿命结束后不可作为普通生活垃圾处理,必须另行收集并作妥善处理。
D776.131.00/07 en 119
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Appendix: China Electronic Industry Standard Compliance
Display
X X X X X X
显示设备
Others
X O X X X X
其它
O: Indicates that this toxic or hazardous substance contained in all of the homogeneous materials for this part
is below the limit requirement in SJ/T11363-2006.
X: Indicates that this toxic or hazardous substance contained in at least one of the homogeneous materials
used for this part is above the limit requirement in SJ/T11363-2006.
Data listed in the table represents best information available at the time of publication.
This table shows where these substances may be found in the supply chain of PTW electronic information
products, as of the date of sale of the enclosed product. Note that some of the component types listed
above may or may not be a part of the enclosed product.
O: 表示该有毒有害物质在该部件所有均质材料中的含量均在SJ/T11363-2006 标准规定的限量要求以下。
X: 表示该有毒有害物质至少在该部件的某一均质材料中的含量超出SJ/T11363-2006 标准规定的限量要求。
此表所列数据为直至发布时所能获得的最全面信息。
此表表明自本产品销售之日起在PTW电子信息产品的供应链上何处可能找到以上所述有毒有害物质。请注
意,以上所列的部件类型中的一些可能不属于本产品。
120 D776.131.00/07 en
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Index
Index
D776.131.00/07 en 121
webline
UNIDOS
Type 10021, Type 10022 and Type 10023
Index
N U
Navigation knob with confirmation key ........................... 32 UDP ............................................................................... 91
Non-volatile memory .................................................... 115 Unit
electrical .................................................................... 42
O radiological .......................................................... 39, 42
User rights ............................................................... 17, 48
Offset current ................................... 58, 63, 76, 79, 80, 82
Operating elements ................................................. 19, 32 V
Operation
Basic principles .......................................................... 30 Version number ............................................................. 28
Problems.................................................................... 77 VNC ............................................................................... 86
P W
Passwords ..................................................................... 49 Waiting period ............................................................... 74
Polarity ........................................................................... 51 Workflow........................................................................ 29
Definition .................................................................. 100
Polarity effect ................................................................. 84 Z
Power key ...................................................................... 33
Power supply ................................................................. 23 Zeroing ................................ 32, 58, 74, 76, 79, 80, 82, 84
Putting into operation ..................................................... 26
Q
Quickinfo window
statistics column......................................................... 68
Quickstart....................................................................... 29
R
Radiation quality ............................................................ 40
Radioactive check device .................................. 40, 58, 80
Isotope ....................................................................... 40
Serial number ............................................................ 40
Rechargeable batteries ................................................ 113
Relay output................................................................... 92
Repair .......................................................................... 110
RS232 interface ............................................................. 47
S
Safety information ............................................................9
Safety inspections........................................................ 110
Settings .......................................................................... 37
General ...................................................................... 37
Miscellaneous device settings ................................... 47
Setup menu ................................................................... 37
Spare Parts .................................................................. 107
Stabilization time............................................................ 50
Statistics .................................................................. 44, 67
Statistics mode
manual ....................................................................... 68
signal-controlled ......................................................... 69
time-controlled ........................................................... 69
System test .................................................................... 46
T
Tables ..............................................................................7
Technical Inspection of the Measuring System ............ 111
Technical specifications ................................................. 95
Time ............................................................................... 46
Trip output...................................................................... 92
122 D776.131.00/07 en