PROCUREMENT OF PORTABLE INDUSTRIAL X-RAY MACHINE

(01 LINE ITEMS)

(AS PER FOLLOWING SPECIFICATIONS)

Part No Description U/I Qty Req

Portable Industrial
MG 161L Ea 01
X-Ray Machine
TECHNICAL SPECIFICATIONS REQUIRED FOR PORTABLE INDUSTRIAL
X-RAY MACHINE MG 161L (or equivalent)
Operational /
5 KV to 200 KV (1600W @ max. 10 mA) or equivalent specification
Power Rating
Current Range 0.5 MA to 10 MA (0.1 mA interval)
Up to 10 min (Adjustable in seconds from 1 sec to 10 minutes) or
Time
equivalent
Tube Head Directional metal ceramic with Focal spot 0.4 to 1.5 mm
Cable Lengths 10 m Long (minimum) with cable folding stowing and protecting cover
0.4 to 1.5 mm “Should be synchronized with our existing X-Ray System
Focal Spot i.e. (CRX Vision and DXR 250C-W) and supported for IPS and Digital
detector with certificate”
Cooling System Water Cooled or Air Cooled
Must be portable type and; for water cooled machines must be mounted
Mobility on trolley / transport cart (including dust cover, document compartment
and non-inflatable wheels)
Input Power Input:- 230 ± 10 VAC, 50 / 60 Hz Max 16A
Requirement / Output:- DC voltage operation with constant potential high dose output
output Operation (low ripple)
Operating
- 20ºC to +60ºC (or equivalent specification)
Temperature
Operating manual, Technical manual, Maintenance manual (including
Publications error codes), Spare parts manual and troubleshooting diagram (in
English), Training Videos CDs , USB
Warranty / 03 Years
Calibration
Standard accessories and Diaphragm / Collimator attachments, Laser /
Telescope Centering device, Remote control (with standard cable
Accessories / length), Safe lights (Blinking + Flash lamp), High quality storage boxes
Utilities (flight cases), 3 or 4 legs Carrying cradle (all accessories are as per
(or equivalent applicability), Inspection stand and High-performance mobile
specification) workstation (high definition display), configured with latest version with
test bench. These accessories / utilities are mandatory for operational
utilization.
Standard tool kit for maintenance (all necessary tools) and administrator
Tool Kit
kit including interface cables, CD-ROM, Bulb extracting tool and USB’s.
Warm-up
(or equivalent Fully automatic, based on real time clock
specification)
Removable Control Module (Minimum 20m cable length with cable
Control Panel
folding stowing)
Additional Functions Change of focus, film focus distance correction extended warm-up,
(or equivalent Comprehensive tube data base two slots for activated tubes, power mode.(for
specification) water cooled machine)
Spare Parts All necessary spares for 03 year to be provided (control / power cards, Bulb
Package etc.)
Max 145 Kg (With 10M cable and compact dimension) or equivalent (for water
Weight cooled)
Max 27 kg (for air cooled machine) or equivalent
Protection class IP65 or equivalent
Duty Cycle 100% for 1hr operation time
Parameter Continuous online display of device temperatures, supply voltage and power
Monitoring rating (KV, mA, Time)
Separate Stabilizer and surge protector for compatible power rating for on-site
Power Protection
installation along with requisite fitting.
Certifications CE Conformity, NFC 74100.