Download as pdf or txt
Download as pdf or txt
You are on page 1of 5

CLINICAL SCIENCE

Central and Peripheral Corneal Endothelial Cell Analysis


With Slit-Scanning Wide-Field Contact Specular
Microscopy: Agreement With Noncontact Specular
Microscopy
Jinhee Lee, MD, Yosai Mori, MD, PhD, Miyuki Ogata, Keiichiro Minami, PhD, and
Kazunori Miyata, MD, PhD

Key Words: corneal endothelial cell, peripheral, slit-scanning,


Purpose: The prospective case series aimed to examine the cataract
agreement between the use of a slit-scanning contact specular
microscope and a noncontact specular microscope in corneal (Cornea 2019;38:1137–1141)
endothelial cell (CEC) analysis and to evaluate the differences
between the central and peripheral regions in normal corneas.
Methods: After confirming normal corneal endothelium with slit-
lamp microscopy, CEC images of 56 eyes of 56 cataractous patients
T he specular microscope has been used for evaluating the
morphology and function of corneal endothelial cells
(CECs) noninvasively, and the endothelial cell density
were analyzed in the central and 4 peripheral regions using a slit- (ECD), the percentage of hexagonal shape cells (HEX), and
scanning contact specular microscope. A noncontact specular coefficient of variation (CV) are determined for analysis.1
microscope was used for the analysis in the central region. The Conventionally, a noncontact specular microscope is com-
endothelial cell density (ECD), the percentage of hexagonal shape prehensively used for the analysis of the central CECs,1,2
cells (HEX), and the coefficient of variation (CV) in the central which are essential for corneal clarity. However, continuous
region were compared. Differences between central and peripheral effects on the peripheral CECs, which are observed after
CECs were also evaluated. implantations of an anterior supported intraocular lens3,4 or
Results: The mean ECD was 2778 cell/mm2 and was not shunt devices5,6 in the anterior chamber, result in an
different from the results using the noncontact specular micro- accelerated decrease of the central CECs. Although the
scope (2736 cell/mm2, P = 0.051). There was a significant evaluation of peripheral CECs is important, it is difficult
correlation (P , 0.001, R2 = 0.72). The analysis of HEX resulted and impractical to examine peripheral CECs using conven-
in larger values with the slit-scanning contact microscope tional noncontact specular microscopes. A contact specular
(53.13% vs. 48.89%, P , 0.001), whereas there was no microscope2,7–9 and confocal microscope10,11 have been used
difference in the CV (38.48 vs. 38.04, P = 0.56). On comparing for peripheral CEC analysis; however, these require experi-
the central and peripheral regions, there was no significant enced skill and are not suitable for use in clinical practice.
difference in the ECD, whereas significant differences were Recently, a slit-scanning wide-field contact specular
found in the superior region in the HEX and CV (P , 0.001) and
microscope has been developed.12 By using contact and
in the nasal region in CV (P = 0.023). confocal microscopy image capturing, wide-field and high-
quality CEC images can be analyzed. Peripheral regions are
Conclusions: The analysis of ECD with the use of the slit-scanning also observable by placing the cone at the place of interest.
contact specular microscope did not differ from the noncontact Nevertheless, the agreements with conventional noncontact
specular microscope, and the results demonstrated no difference specular microscopes have not been examined. The objectives
between the central and peripheral ECD. of this prospective and observational case series study were to
examine the agreement between the slit-scanning wide-field
contact specular microscope and a noncontact specular
microscope in the central CEC analysis and to evaluate the
analysis in the peripheral regions in normal corneas.

Received for publication January 28, 2019; revision received March 14, 2019;
accepted March 16, 2019. Published online ahead of print May 3, 2019.
From the Miyata Eye Hospital, Miyazaki, Japan. MATERIALS AND METHODS
The authors have no funding or conflicts of interest to disclose. This study was approved by the institutional review
Correspondence: Kazunori Miyata, MD, PhD, Miyata Eye Hospital, 6-3
Kurahara-cho, Miyakonojo, Miyazaki, 885-0051, Japan (e-mail: kmiyata@ board of Miyata Eye Hospital and was performed according
miyata-med.ne.jp). to the tenets of the Declaration of Helsinki. Written informed
Copyright © 2019 Wolters Kluwer Health, Inc. All rights reserved. consent was obtained from the patients before examinations.

Cornea  Volume 38, Number 9, September 2019 www.corneajrnl.com | 1137

Copyright © 2019 Wolters Kluwer Health, Inc. Unauthorized reproduction of this article is prohibited.
Lee et al Cornea  Volume 38, Number 9, September 2019

Cataractous patients who visited Miyata Eye Hospital from sharply identified was selected. After tracing individual CEC
September 2017 to February 2018 were enrolled. Normal borders automatically, the examiner corrected improper
corneal endothelium was confirmed in a slit-lamp microscopy tracing manually in an area of 0.60 · 0.40 mm. Then, the
examination. Patients with the history of ocular surgery, ECD, HEX, and CV were obtained using CellChek software.
trauma, contact lens wear, corneal endothelial dysfunction, or For the noncontact measurement, the same procedure and
diseases affecting the CEC were excluded. For bilateral analysis were conducted for an image with sharper CEC
cataract patients, the eye first examined was selected for borders. The numbers of CEC used for analysis were also
analysis. The sample size required for examining the linear recorded for the 2 measurements.
correlation between 2 measurements in the central region was
31 eyes for a significance level of 0.05 and detection power of
0.9 when the R2 was 0.25. The study was planned to exceed Statistical Analysis
the calculated sample size. The agreement between the 2 specular microscopes was
examined with the central ECD, HEX, and CV. After
confirming their normality using the Shapiro–Wilk test, the
CEC Analysis differences and correlations were examined using a paired t
CEC images were obtained using a slit-scanning test and linear regression analysis, respectively.
contact specular microscope, CellChek C (version 1.0.0.5; The intraobserver test was examined for ECD, HEX,
Konan Medical, Nishinomiya, Japan), and noncontact spec- and CV using the intraclass correlation coefficient (ICC).
ular microscope, CellChek SL (version 30.00; Konan Med- Bland–Altman analysis13 was also used for evaluating the
ical), on the same day. The former specular microscope used 95% limits of agreement. In the slit-scanning contact specular
alternative confocal microscopy optics, which was similar to microscope measurements, ECD, HEX, and CV were com-
the conventional confocal microscope except for the use of pared between the center and the peripheral regions using
a slit-scanning technology, for enabling wider field observa- a paired t test with the Holm correction.
tion of the cornea. After administering topical anesthesia Statistical analyses were performed using R version
using 0.4% oxybuprocaine, the probe with an applanation 3.4.1. P values less than 0.05 were considered statistically
diameter of 4 mm was placed at the apex of the cornea for significant.
capturing sequential images of the central CEC (width of
0.65 mm and a height of 0.48 mm). In the same manner,
images at 4 peripheral regions (superior, nasal, inferior, and RESULTS
temporal) were captured. The distance from the center to the Fifty-six eyes of 56 patients (mean age: 61.2 years)
peripheral regions was approximately 5 mm. The noncontact were included in the study. Table 1 shows the demographic
specular microscope was also used for obtaining the central data of the subjects. For the sample size, the power in
CEC image (width of 0.24 mm and a height of 0.40 mm) by detecting the correlation between the 2 measurements was
targeting the corneal apex for making the CEC images within anticipated to be over 0.99. The number of CECs used for
the relevant images with the slit-scanning contact specular analysis with the slit-scanning contact specular microscope
microscope. The image capturing process was repeated until was 4.1 times more than that with the noncontact specular
acceptable quality images were confirmed. Figure 1 shows the microscope, although the CEC image size was 2.5 times
typical CEC images obtained using the 2 specular micro- larger. In 5 of 56 images (8.9%), the ECDs were calculated
scopes: the image sizes were 3.25 times different. with CECs being less than 100 in the noncontact
From the captured CEC images, the ECD, HEX, and measurement.
CV of individual cell areas were analyzed. The analysis was Table 2 shows the analysis results of the central CEC
performed by a single experienced examiner (M.O.) for mini- images. The mean ECD measured with the slit-scanning
mizing the variance of the analyzed results.1 From the contact specular microscope was not significantly different
sequential images, an image in which all CECs could be from that measured with the noncontact specular microscope
(P = 0.051, paired t test), and the mean difference (1.61%)
was neglectable. Significant and strong correlation, high ICC,
and the 95% limits of agreement indicated the agreement in
the central ECD analyses. The mean HEX was significantly
larger with the slit-scanning contact specular microscope (P
, 0.001). There was no significant difference in the mean CV
between the 2 specular microscopes (P = 0.56). The R2 and
ICC in the HEX and CV were relatively low.
Figure 2 shows the ECD, HEX, and CV measured with
the slit-scanning contact specular microscope at the central
and peripheral regions, respectively. In the ECD, there was no
FIGURE 1. Typical images obtained using a slit-scanning significant difference between them (P = 1.00, paired t test
contact (left) and noncontact (right) specular microscope in with Holm correction). The HEXs in the superior region were
the central cornea. The image sizes were 0.65 · 0.48 and 0.24 significantly lower than those in the central region (P ,
· 0.40 mm, respectively. 0.001). The CVs in the superior and nasal regions were

1138 | www.corneajrnl.com Copyright  2019 Wolters Kluwer Health, Inc. All rights reserved.

Copyright © 2019 Wolters Kluwer Health, Inc. Unauthorized reproduction of this article is prohibited.
Cornea  Volume 38, Number 9, September 2019 Central and Peripheral Corneal Endothelial Cell

endothelium, it was hard to identify the sampling differences


TABLE 1. Demographic Data of Subjects
from the CEC images.
Specular Microscope Slit-Scanning Contact Noncontact The ECDs in the peripheral regions were not signifi-
N 56 eyes of 56 patients cantly different from those in the central regions. These
Age, yr 61.2 6 9.3 [26–74] results coincided with the previous results.7,12,15 However,
Gender Male: 23, female: 33 the distribution of ECD has not been confirmed yet. In some
Eye Right: 22, left: 34 studies, higher ECD was observed in the peripheral
Keratometry, diopter 44.00 6 1.79 [40.50–48.75] region.8,16 It is anticipated that higher peripheral ECDs in
Image size, mm 0.65 · 0.48 0.24 · 0.40 normal corneas would be advantageous for Descemet mem-
No. CECs used for analysis 550 6 84 [343–726] 133 6 20 [77–164] brane endothelial keratoplasty and Descemet stripping auto-
Mean 6 SD [range].
mated endothelial keratoplasty.17 Wiffen et al18 reported that
ECD was lower in the peripheral region. The methodology
used in the previous studies has not been established: various
instruments, such as the use of contact specular micros-
copy,7,8,12,16,18 confocal microscopy,15 and the analysis area
significantly higher than in the central region (P , 0.001, sizes8,16 and ECD analysis methods, were used previously.
0.023, respectively). Further assessments of peripheral ECD using advanced and
wider field specular microscope are important.
The current results showed lower HEX and higher CV
of the superior region. Peripheral morphology in more than 4
DISCUSSION regions had rarely been evaluated.19 Previous reports showed
In the current study, the central ECD, HEX, and CV that HEX of the inferior region15 was lower and CV of the
were measured with slit-scanning contact and noncontact temporal region was higher than the center region,18 whereas
specular microscopes in cataract eyes. The central ECD Amann et al8 showed that CV was not different between the
values were in good agreement with each other, which central and peripheral regions. In the current study, both
coincided with the previous comparison between noncon- HEX and CV of the central region did not agree well
tact specular and conventional contact microscopes.2,9 The between the 2 measurements. The previous studies ad-
comparison between noncontact and confocal measure- dressed that the morphology of the CEC could alter with
ments by Kitzmann et al14 showed reproducibility in the the devices used.19,20 Hence, it was hard to compare these
ECD analysis. Together with the previous findings, it was with the previous results.
demonstrated that measurement using the slit-scanning There were some limitations to this study. First, the
contact specular microscope was not different from con- specular microscope examinations were taken once for each
ventional ECD examination. Furthermore, the number of measurement. The SD of ECD was 1.6% to 1.8% in
CECs used for analysis increased with the use of the slit- measurements using the noncontact specular microscope.21
scan contact measurement. The mean ECD indicated that With the high correlation, ICC suggested no difference in
about 84.5% of CECs in the wide-field images were used the central ECD analysis; however, interobserver testing is
for analysis. On the other hand, only 50% of CECs were required for confirming the repeatability. Next, in the
used in the noncontact measurement. The confocal optics current comparison for normal endothelium, it was hard to
and wide-field measurement could analyze more CECs with confirm that both CEC images were overlapped. The image
higher resolution. Further assessments are necessary to positions in the central cornea examination would be close
confirm this. by aiming the probe position at the corneal apex, and the
In the analysis of the HEX and CV, agreement was not images of the slit-scanning contact specular microscope
obtained. One of the factors might be the differences in the were sufficiently larger. Further evaluation of endothelium
quality and size of CEC images, which were superior in the with abnormal morphology would be required. In addition,
slit-scanning contact microscope compared with the non- interobserver testing is necessary for confirming the repeat-
contact specular microscope (Fig. 1). Although CEC images ability of the newly developed specular microscope. Last,
of the slit-scanning contact microscope could include the the exclusion criteria consisted of diseases that may affect
smaller images of the noncontact microscope, there was a risk CECs. Subjects who are primary angle-closure suspects
of sampling differences. As the current study analyzed normal with iridotrabecular contact,22,23 suspended debris in the

TABLE 2. Analysis Results of Central CECs


Slit-Scanning Contact, Noncontact, 95% Limits
Mean 6 SD [Range] Mean 6 SD [Range] P Correlation ICC of Agreement
ECD 2778 6 300 cells/mm2 [2104 to 3426] 2736 6 242 cells/mm2 [2203 to 3425] 0.051 R2 = 0.72 (P , 0.001) 0.82 2271 to 357 cells/mm2
HEX 53.13% 6 5.44% [41 to 65] 48.89% 6 7.40% [35 to 72] ,0.001 R2 = 0.093 (P = 0.022) 0.24 211.07% to 19.54%
CV 38.48 6 5.50 [35 to 73] 38.04 6 5.17 [35 to 74] 0.56 R2 = 0.18 (P = 0.0012) 0.42 211.71 to 10.82

Copyright  2019 Wolters Kluwer Health, Inc. All rights reserved. www.corneajrnl.com | 1139

Copyright © 2019 Wolters Kluwer Health, Inc. Unauthorized reproduction of this article is prohibited.
Lee et al Cornea  Volume 38, Number 9, September 2019

microscope for ECD analysis. There was no difference


between the central and peripheral ECD in patients with
cataract. The current results demonstrated that the slit-
scanning contact specular microscope examination would be
effective and advanced for central and peripheral CEC
examinations.

REFERENCES
1. McCarey BE, Edelhauser HF, Lynn MJ. Review of corneal endothelial
specular microscopy for FDA clinical trials of refractive procedures,
surgical devices, and new intraocular drugs and solutions. Cornea. 2008;
27:1–16.
2. Módis L, Jr, Langenbucher A, Seitz B. Corneal endothelial cell density
and pachymetry measured by contact and noncontact specular micros-
copy. J Cataract Refract Surg. 2002;28:1763–1769.
3. Huang D, Schallhorn SC, Sugar A, et al. Phakic intraocular
lens implantation for the correction of myopia: a report by the
American Academy of Ophthalmology. Ophthalmology. 2009;116:
2244–2258.
4. Coli AF, Price FW, Jr, Whitson WE. Intraocular lens exchange for
anterior chamber intraocular lens-induced corneal endothelial damage.
Ophthalmology. 1993;100:384–393.
5. Koo EB, Hou J, Keenan JD, et al. Effects of glaucoma tube surgery on
corneal endothelial cells: a review. Eye Contact Lens. 2016;42:
221–224.
6. Casini G, Loiudice P, Pellegrini M, et al. Trabeculectomy versus EX-
PRESS shunt versus Ahmed valve implant: short-term effects on corneal
endothelial cells. Am J Ophthalmol. 2015;160:1185-1190.e1.
7. Blackwell WL, Gravenstein N, Kaufman HE. Comparison of central
corneal endothelial cell numbers with peripheral areas. Am J Ophthalmol.
1977;84:473–476.
8. Amann J, Holley GP, Lee SB, et al. Increased endothelial cell density in
the paracentral and peripheral regions of the human cornea. Am J
Ophthalmol. 2003;135:584–590.
9. Isager P, Hjortdal JO, Guo S, et al. Comparison of endothelial cell
density estimated by contact and non-contact specular microscopy. Acta
Ophthalmol Scand. 2000;78:42–44.
10. Syed ZA, Tran JA, Jurkunas UV. Peripheral endothelial cell count is
a predictor of disease severity in advanced Fuchs endothelial corneal
dystrophy. Cornea. 2017;36:1166–1171.
11. Huang P, Tepelus T, Vickers LA, et al. Quantitative analysis of depth,
distribution, and density of cysts in Acanthamoeba keratitis using
confocal microscopy. Cornea. 2017;36:927–932.
12. Tanaka H, Okumura N, Koizumi N, et al. Panoramic view of human
corneal endothelial cell layer observed by a prototype slit-scanning
wide-field contact specular microscope. Br J Ophthalmol. 2017;101:
655–659.
13. Bland JM, Altman DG. Measuring agreement in method comparison
FIGURE 2. Comparison of ECD, HEX, and CV in the central studies. Stat Methods Med Res. 1999;8:135–160.
and the peripheral regions by using a paired t test with the 14. Kitzmann AS, Winter EJ, Nau CB, et al. Comparison of corneal
Holm correction. There was no significant difference in ECD endothelial cell images from a noncontact specular microscope and
a scanning confocal microscope. Cornea. 2005;24:980–984.
between these (P = 1.00). The HEX in the superior region were 15. Zheng T, Le Q, Hong J, et al. Comparison of human corneal cell density
significantly lower than those in the central regions (P , by age and corneal location: an in vivo confocal microscopy study. BMC
0.001). The CVs in the superior and nasal regions were sig- Ophthalmol. 2016;16:109.
nificantly higher than those in the central region (P , 0.001, 16. Sturrock GD, Sherrard ES, Rice NS. Specular microscopy of the corneal
0.023, respectively). endothelium. Br J Ophthalmol. 1978;62:809–814.
17. Van den Bogerd B, Dhubhghaill SN, Koppen C, et al. A review of the
evidence for in vivo corneal endothelial regeneration. Surv Ophthalmol.
2018;63:149–165.
aqueous humor and breakdown of the blood–aqueous bar- 18. Wiffen SJ, Hodge DO, Bourne WM. The effect of contact lens wear on
the central and peripheral corneal endothelium. Cornea. 2000;19:
rier after laser peripheral iridotomy,24,25 and exfoliation 47–51.
syndrome,26 which might differentiate the CEC and its 19. Goukon H, Kamiya K, Shimizu K, et al. Comparison of corneal
distribution from normal eyes, might be included. Addi- endothelial cell density and morphology after posterior chamber phakic
tional examinations would be necessary to diagnose corneal intraocular lens implantation with and without a central hole. Br J
Ophthalmol. 2017;101:1461–1465.
disorders before recruiting subjects. 20. Luft N, Hirnschall N, Schuschitz S, et al. Comparison of 4 specular
In conclusion, the slit-scanning contact specular micro- microscopes in healthy eyes and eyes with cornea guttata or corneal
scope was not different from the noncontact specular grafts. Cornea. 2015;34:381–386.

1140 | www.corneajrnl.com Copyright  2019 Wolters Kluwer Health, Inc. All rights reserved.

Copyright © 2019 Wolters Kluwer Health, Inc. Unauthorized reproduction of this article is prohibited.
Cornea  Volume 38, Number 9, September 2019 Central and Peripheral Corneal Endothelial Cell

21. Jones SS, Azar RG, Cristol SM, et al. Effects of laser in situ 24. Kerr Muir MG, Sherrard ES. Damage to the corneal endothe-
keratomileusis (LASIK) on the corneal endothelium. Am J Ophthalmol. lium during Nd/YAG photodisruption. Br J Ophthalmol. 1985;69:
1998;125:465–471. 77–85.
22. Wang PX, Koh VT, Loon SC. Laser iridotomy and the corneal 25. Higashihara H, Sotozono C, Yokoi N, et al. The blood-aqueous barrier
endothelium: a systemic review. Acta Ophthalmol. 2014;92:604–616. breakdown in eyes with endothelial decompensation after argon laser
23. Varadaraj V, Sengupta S, Palaniswamy K, et al. Evaluation of angle iridotomy. Br J Ophthalmol. 2011;95:1032–1034.
closure as a risk factor for reduced corneal endothelial cell density. J 26. Miyake K, Matsuda M, Inaba M. Corneal endothelial changes in
Glaucoma. 2017;26:566–570. pseudoexfoliation syndrome. Am J Ophthalmol. 1989;108:49–52.

Copyright  2019 Wolters Kluwer Health, Inc. All rights reserved. www.corneajrnl.com | 1141

Copyright © 2019 Wolters Kluwer Health, Inc. Unauthorized reproduction of this article is prohibited.

You might also like