3.8 Connect external safety devices
Caution: Safety and warning devices for X-ray systems and radiation protection rooms
must be installed in accordance with local regulations.
Note: In the case of multi-generator systems the safety devices are attached not inside
the power module but to the distributor box (see section 3.7.8).
Títulos relacionados
https://es.scribd.com/document/389459791/Y-MG103-605-v01r04-en-20066589-4-pdf# 23/66
6/2/23, 14:43 Y MG103-605 - v01r04 - en - 20066589 PDF | PDF | Electromagnetic Compatibility | High Voltage
32 Y.MG103 … 605
20066589 – v01r04 – en
YXLON International
ALL RIGHTS RESERVED
de 88
3.8.1 Indicator lights for 230V systems
Cooler phase 1
Cooler phase 0
Mains
Phase 0
Mains
Mains
Phase 1
Mains
Door contact II
Door contact I
Pre-warning contact
Loading 230V~ 1A max.
Títulos relacionados
https://es.scribd.com/document/389459791/Y-MG103-605-v01r04-en-20066589-4-pdf# 24/66
6/2/23, 14:43 Y MG103-605 - v01r04 - en - 20066589 PDF | PDF | Electromagnetic Compatibility | High Voltage
Y.MG103 … 605 33
Technology with Passion
20066589 – v01r04 – en
YXLON International
ALL RIGHTS RESERVED
de 88
3.8.1.2 Internal flashing indicator light (in the control unit)
The internal flashing light in the control unit (24V / 1W) is electronically monitored. It flash-
es when X-ray radiation is on.
Note: If national regulations require red warning indicators for ‘X-ray On’, then the yellow
warning light cover should be replaced by the red warning light cover supplied with the
system.
3.8.1.3 External flashing indicator light with internal voltage supply of 230V~
► Remove all jumpers connected to 205 and 220!
► Connect the jumper and the light as illustrated in
fig. 15.
► Plug jumper W1 (on circuit board E3 or E4, see fig.
15/16) across contacts 2-3 (as it is when shipped).
Light 1
(230V~; flashes; 40 - 60W)
Note: If no light is being connected, the plug-in jumper should be plugged across contacts
1-2.
Títulos relacionados
https://es.scribd.com/document/389459791/Y-MG103-605-v01r04-en-20066589-4-pdf# 25/66
6/2/23, 14:43 Y MG103-605 - v01r04 - en - 20066589 PDF | PDF | Electromagnetic Compatibility | High Voltage
34 Y.MG103 … 605
20066589 – v01r04 – en
YXLON International
ALL RIGHTS RESERVED
de 88
3.8.1.4 External flashing indicator light with external voltage supply
► Remove all jumpers connected to 205 and 220!
► Connect the light and external voltage as illustrat-
ed in fig. 19.
► Plug jumper W1 (on circuit board E3 or E4, see fig.
15/18) across contacts 2-3 (as it is when shipped).
24V~ - 230V~
external light voltage
Light 1
I min. = 0.1A
I max. = 0.8A
P max. = 75W
Note: If no light is being connected, the plug-in jumper should be plugged across contacts
1-2.
3.8.1.5 External solid warning light with internal voltage supply of 230V~
► Connect the light as shown in fig. 21.
► Open the solder jumpers between X3 and X4 (on
circuit board E3 or E4, see fig. 18). Light monitor-
ing is now activated. Light 2
Solid light 230V~
40 - 60 W
Note: If no light is being connected, the solder bridge X3 and X4 must be linked (as
shipped).
Títulos relacionados
https://es.scribd.com/document/389459791/Y-MG103-605-v01r04-en-20066589-4-pdf# 26/66
6/2/23, 14:43 Y MG103-605 - v01r04 - en - 20066589 PDF | PDF | Electromagnetic Compatibility | High Voltage
Y.MG103 … 605 35
Technology with Passion
20066589 – v01r04 – en
YXLON International
ALL RIGHTS RESERVED
de 88
3.8.1.6 Fuses for 230V indicator lights
The fuses for warning lights F1 and F2 are located as follows:
‐ In the MGP40: on circuit board E3. Rating: 0.8A inert.
‐ In the MGP41: on circuit board E4. Rating: 0.8A inert.
‐ In the MGP41-600: on circuit board E4. Rating: 0.8A inert.
Cooler phase 0
Mains
Phase 0
Mains
Mains
Phase 1
Cargue
Warning sus documentos para descargar
light 2 (solid)
Safety circuit
Automatic systems
(supplied bridged)
Cargue para descargar
Door contact II
O
Door contact I
Comienza
Locktu prueba GRATUITA
mechanism magnet de 30 días
Loading 230V~ 1A max.
Prewarning contact
Loading 230V~ 1A max.
Note: Warning lights L1, L2 and L3 get connected directly to the circuit board at plug X6.
Títulos relacionados
https://es.scribd.com/document/389459791/Y-MG103-605-v01r04-en-20066589-4-pdf# 27/66
6/2/23, 14:43 Y MG103-605 - v01r04 - en - 20066589 PDF | PDF | Electromagnetic Compatibility | High Voltage
36 Y.MG103 … 605
20066589 – v01r04 – en