vario TOC cube

Good to know, that children can play here without worries.

2
The advantages for our nature are
evident

Water is the prerequisite for life on our planet. In times of Crucial advantages compared to the classical wet chemical
growing environmental problems the measurement of the Chemical Oxygen Demand (COD) procedures are:
water quality as well as for solids such as soils are vital.
• Decrease of the analysis time from hours to minutes
The measurement of all organic pollutants by means of their • Omission of pollutants like Cr6+ or concentrated
content of organically bound carbon (TOC) serves as the sulphuric acid
analysis method for fast and highly sensitive determination • Completely safe and automatic operation
of the total pollution load. • Additional possibility to measure bound nitrogen (TNb)
instead of the also environmentally problematic
Kjeldahl method

Clean analysis for a clean environment

TOC by experience As the successor of Heraeus, Elementar has a more than 100
years tradition in developing and manufacturing analysers
for the “organic“ elements such as carbon and nitrogen by
means of high temperature digestion.

This concept was already used in 1976 in the first


fully automatic TOC analyser. vario TOC cube already
represents the 3rd generation of high temperature TOC
analysers by Elementar. For each development step
experiences from the predecessor models could be im-
plemented. This top of the line analyser is a totally new
concept. It combines the advanced water analysis with the
elemental analysis of solids based on the special experience
of Elementar for both these areas of instrumental analysis.

Heraeus LiquiTOC 1976

3
A clear concept - superior design

4
The basic concept decides
The measuring principle is based on the high temperature A connected PC calculates the carbon content from the sam-
digestion of the sample in an air/O2 stream at 850°C. For ple volume and/or weight and the detector signal based
special applications with thermally stable solids, digestion on a calibration curve. Inorganic carbon (carbonate) may
temperatures of permanently up to 1200°C or temporarily be automatically determined as TIC after acidification and
1800°C can be achieved. Totally bound carbon is converted purging in the sparger.
into CO2 which is quantitatively determined by means of a
NDIR detector. The advantage of this method as opposed to The TOC is calculated from TC-TIC. A direct non purgable or-
the wet chemical UV/persulfate digestion is the absolute as- ganic carbon (NPOC) determination is also possible after
surance that even stable compounds, particles or salt con- external TIC removal (typically in the sample carousel). The
taining solutions will be completely detected. Additionally sparger may also be used for direct determination of pur-
the high temperature method makes the determination of gable or volatile carbon compounds (VOC or POC). Thus, all
bound nitrogen (TNb) possible. relevant parameters of TOC analysis are accessible with one
instrument and without extensive modifications.
The liquid sample is aspirated via a dosing syringe and in-
jected into the combustion tube where catalytic oxidation The use of NO specific detectors extends the application area
takes place. Alternatively, solid samples are inserted via a to the TNb analysis of bound nitrogen. The total concept is
ball valve. Water vapor is removed in a 3-stage drying pro- enhanced due to short gas and liquid pathways and a mi-
cess, “rock steady” flow rate is maintained in the presence nimum number of valves. This results in shorter analysis
of varying pressure conditions by a flow regulation device times, lower detection limit as well as the ability for mea-
(FRD), resulting in the most accurate measurement of the surement of particle and salt containing samples.
total carbon (TC) content by NDIR detection.

H3PO4

Air

Sample Sparger Syringe Combustion Drying Detection

simplified functional diagram

5
6
Superior - the sum of the Good

Variable sample feeding High temperature digestion with matrix separation

For liquid samples the well proven digitally cont- A high combustion temperature is vital for a quantitative
rolled syringe aspirates all aliquots of sample required oxidation of bound carbon to CO2 and a precondition for the
for the number of replicates from the sampler vial all at digestion of stable compounds and particles. The vario TOC
once. The transfer tubing is rinsed with the sample its- cube can be operated at a permanent furnace temperature
elf, and then each aliquot is sequentially injected into of up to 1200°C. In solids mode the combustion enthalpy of
the combustion tube. This procedure eliminates repli- the tin capsules results in a temporary temperature increase
cate variability due to atmospheric CO2 after the sep- of up to 1800°C. This allows the analysis of even refractory
tum is pierced. Variable injection volumes, from 50 μl to samples.
2 ml, allow multi-point calibrations from just one calibra-
tion solution over a very large dynamic range. Analyses of Quartz ash crucibles are used for the separation and collec-
particulate containing samples benefit from a single cera- tion of combustion residues. The combustion takes place in
mic manifold valve and short tubing paths. the crucible, and non volatile or incombustible materials,
such as salts, remain behind. This is the first use of ash
For solid sampling, Elementar has incorporated its well pro- crucibles for matrix separation in TOC determinations (pa-
ven solid sampling technology utilised in organic elemen- tent applied for). Thus, contamination of the catalyst can
tal analysis. The samples are weighed into tin capsules and be prevented and the residues can be easily removed from
automatically transferred to the combustion tube via the the crucible.
patented ball valve of Heraeus. This allows for virtually any
type of sample to be analysed, from just a few μl of a liquid, NDIR detection of the next generation
large particulate containing samples, semi-solids to actual
solid samples. Automatic sample feeding of up to 120 sam- A completely new high performance detector for CO2 is used
ples is possible, with analysis times of only 3-4 minutes. in the vario TOC cube. The actual realisation of the non
dispersive infrared (NDIR) principle means a minimisation of
Highly stable operation and results spectral interferences and a vastly low-noise characteristic.
A dynamic detection range for TOC from ppb into the per-
Long term stability and high sample throughput are impe- centage range is achieved by the 24 bit digital resolution of
rative if TOC analysers are to be effectively used for scree- the measuring signal without the need for measuring range
ning purposes. The vario TOC sets new standards in this re- gain adjustments. Unique is the possibility to also mea-
gard. Even with large injection volumes the highly effective sure TNb by means of NDIR detection. For highest demands
3-stage drying system is working almost maintenance free. regarding sensitivity a CLD is available.
Fluctuations in the gas stream caused by sample injection
and evaporation are controlled with a flow regulation de-
vice (FRC). The ability to separate sample matrices from ana-
lysed gases results in long uninterrupted measuring periods
even for highly contaminated samples.

7
8
Due to the highly developed NDIR
detector the vario TOC cube may be
regarded as the most sensitive
high temperature TOC analyser
worldwide.

Technology of the future

High performance components User friendly software

High temperature components which are subject to corrosi- The Windows® based software leaves nothing to be de-
on, such as the sample feeder arm, and combustion tubes sired in terms of functionality, storage capacity, automatic
are fabricated from high-alloy tempered steel for long life. optimisation or error diagnostics. Interface between the
Stepper motors utilise robust and accurate harmonic drive analyser and PC is via standard USB connection. Commu-
assemblies. nications with LIMS and data export to MS Excel® are also
integrated as standard features.
The latest 24 bit micro processor generation guarantees fas-
test signal processing and the CAN-bus technology derived The look-and-feel of the software equates Windows®
from German automotive engineering guarantees absolute standard. Data input and output is carried out in an MS
safety in the internal data communication. Excel® like data sheet. This makes the software intuitive us-
able.
Power is everything
Remote diagnosis over the internet is particularly effective
The internal power supply of the vario TOC cube allows as all aspects of the system are available via the software.
operation with supply voltages ranging from 100-240 VAC Protection of electronic data storage can be assured via the
50/60 Hz. The vario TOC cube corresponds completely with 21 CFR part 11 components which are available.
the EC low voltage directive assuring the highest level of
electrical safety.

9
LIquid Measurements TOC, TIC, TC [mg/l]* Solids Measurements TC [%]*

Pure water River sediment (Elbe)


Measurements 10 TOC 0.069 ± 0.006 8.496 ± 0.136 Measurements 5
Quantity 2 ml Quantity 20 mg

Drinking water (Hanau) Susp. sediment (Elbe)


Measurements 10 TOC 0.634 ± 0.009 5.396 ± 0.088 Measurements 5
Quantity 0.6 ml Quantity 20 mg

Well water Soil std. 2 (0.732% C)


TIC 20.04 ± 0.25
Measurements 5 0.714 ± 0.010 Measurements 5
Quantity 0.8 ml
TC 21.53 ± 0.12 Quantity 20 mg

River water (Elbe) Soil std. 1 (3.5% C)


Measurements 5 TOC 39.10 ± 0.40 3.514 ± 0.039 Measurements 5
Quantity 0.8 ml Quantity 20 mg

Municipal waste water Filter dubst


Measurements 5 TOC 27.96 ± 1.32 2.157 ± 0.040 Measurements 3
Quantity 0.8 ml Quantity 20 mg

Electroplat. waste water Langbeinite w/ additive


Measurements 10 TOC 3.776 ± 0.133 34.23 ± 0.0034 Measurements 3
Quantity 0.8 ml Quantity 200 mg

Sea water (Red Sea)


Measurements 10 TOC 0.0858 ± 0.0013 Nitrogen Measurements TNb [mg/l]*
Quantity 0.5 ml

Brine Mix standard NH4/NO3


TOC 1.006 ± 0.022 100.0 ± 0.3
Measurements 20 Detector CLD Measurements 6
Quantity 0.2 ml Quantity 50 µl

Cold brine (Bad Soden/Salm) Sea water (Valencia)


TIC 112.0 ± 4.6 1.84 ± 0.25
Measurements 5 Measurements 8
Quantity 0.25 ml
TC 116.3 ± 2.9 Detector CLD
Quantity 0.5 ml

Hydrochloric acid (3.2%) Nitric acid (approx. 5%)


TOC 1.379 ± 0.033 10.455 ± 450
Measurements 20 Detector NDIR Measurements 12
Quantity 0.05 ml Quantity 0.1 ml

Industrial waste water Standard (100 ppb)


Measurements 5 TOC 41.84 ± 0.81 0.091 ± 0.013
Detector NDIR Measurements 8
Quantity 0.25 ml Quantity 2 ml

* Element contents in [%] incl. standard deviation in [%]


10
Even if it gets really tough

Pure water Highly concentrated solutions

In connection with an injection volume of 2 ml the highly Brine is a saturated (36%) salt solution from processes like
sensitive NDIR detector allows the analysis of water quali- chlorine alkaline electrolysis. After a 1:1 dilution it is pos-
ties for pharmaceutical applications. A standard deviation sible to measure 200 samples in series and to determine
of 6 ppb (3 ppb trace) for a concentration of 70 ppb signi- safely even 1 ppm TOC. Natural brine with 9.5% dissolved
ficantly exceeds the requirements of the Pharmacopoeia. mineral content is one of the most concentrated in Germa-
ny and has a natural high turbidity. Even diluted mineral
Drinking water acids can be safely measured with vario TOC cube.

Even with high carbonate contents a very low TOC value Solids
is precisely measurable after acidification and purging of
the TIC. Comparable conditions can also be found for well The integrated sampler in connection with the high tem-
water. perature combustion allows TC analyses with tin capsules
of up to 120 samples. Samples can range from true solids to
Surface water heavy slurries to pasty semi-solids. This technique can also
be used for liquid samples in which only a limited quantity
Water from rivers and lakes do not represent any difficulties exists (even several μl’s). Carbon can even be determined
for the vario TOC cube. Particles and suspended solids which from thermally stable carbide ceramics with a decomposi-
are commonly present in these samples can be measured tion temperature of 1800°C with a 100% recovery.
directly without any problems.
Nitrogen measurements
Waste water
vario TOC makes it possible to determine the chemically
The large dynamic measuring range offers advantages for bound nitrogen (TNb) which is gaining more interest in
strongly fluctuating TOC concentrations in municipal waste water analysis.
water. Even waste water from circuit board processing with
high electrolyte contents such as sulphates are measured There are 3 detectors available for the detection of the
correctly due to the matrix separation and interference- formed NO. Using a NDIR detector, for example, a stan-
free CO2 detection. dard deviation of 5 ppb at 100 ppb (CLD) as well as a large
concentration range up to more than 50,000 ppm can be
Sea water attained.

The salt content of the oceans is approx. 3.5% whereas the


TOC content lies in the lower ppm range. The vario TOC cube
is able to perform over 100 precise TOC measurements be-
fore the separated salt has to be simply removed from the
matrix separator. A special TIC only mode is very useful for
sea water analysis.

11
12
Variability with a compact shape Selection of solid sample feeding
vario TOC cube represents a complete line of TOC analysers Different modules enable the measurement of samples in
which can be configured for the individual user require- tin foils or capsules, either in single mode or with an auto-
ments. The modular system is based on a basic instrument matic sampler for 80 or 120 samples. This is suitable for solid
which already includes the main functions: samples but can also be used for heavy particulate contai-
ning samples, viscous liquids, semi-solids or liquid samples
• Automatic dosing and distribution of liquid samples of very small quantity. Sample as large as 300 mg can be
• Sparger for stripping of TIC as CO2 or for POC determi- accomodated in the 120 position sampler, and up to 1 g in
nation the 80 position sampler.
• High temperature combustion for oxidation of
TC/NPOC or POC (VOC) to CO2 Selection of the detections
• Measuring gas drying and gas flow regulation Three different NDIR detectors are available for carbon
• CO2 detection in the carrier gas stream determination.
• Power, control and regulation electronics • A standard wide range NDIR detector is suitable for
• Instrument operation, control and measuring value almost all applications
• processing is done via a connected PC or laptop • For lowest concentrations in the ppb area an extreme-
ly sensitive IR detector can be installed.

vario TOCcube
vario TOC cube--der
literally
Name variable
ist Programm
The content of bound nitrogen (TNb) can be optionally
In connection with the basic instrument there are a variety measured with 3 different detection types:
of modules available to achieve various applications.
1. Electrochemical cell (EC) as the most economic detec-
Selection of liquid sample feeding tion type
Different modules allow the measurement of liquids either 2. Chemiluminescence detector (CLD) according to DIN
in single mode or with an automatic sample feeder for eit- 38409 H27
her 32, 50 or 80 samples. The sampler for 32 samples is es- 3. Non dispersive IR detector for largest concentration
pecially suitable for 40 ml septum vials according to EPA. range from 2 to more than 50,000 ppm.
The sample can be stirred in each of the operation modes.
Air purging for stripping TIC or POC (VOC) is also possible. All For low TOC concentrations in the ppb range a sensitive
types of samplers will be integrated in the instrument and and powerful IR detector is available.
do not require any additional footprint. Random access to
selected sample positions is also a feature of vario TOC auto For environmental analysis the vario TOC select is an
samplers. economic, tailormade alternative.

13
Analysis method: Catalytic high temperature combustion at 850°-950°C (max. 1200°C)
Oxidation from C to CO2 which is quantitatively determined with a NDIR
Optional detection of the formed NO with NDIR, CLD or EC

Measuring parameters: TC, TOC, DOC, NPOC, TIC, POC, optional TNb

Norms and standards: ISO 8245, EPA 415.1, European standard acc. to EN 1484, ENV 12260,
European Pharmacopoeia 6.0, vol.1 (20244) 2008, p. 71,
U.S.Pharmacopoeia-NF, USP 30 (643) 2007; DIN 38409

Safety standards: CE-label, IEC 1010, EMV low voltage directives

Measuring range*: C: 0 – 60,000 mg/l in liquids (0 - 25,000 mg/l for high sensitive version)
0 - 100% in solids
N: 0 – 1000 mg/l (CLD)
0 - 50,000 mg/l (NDIR)
0 - 700 mg/l (ECD)

Detection limits* (SD): C: 6 µg/l (3 ppb for high sensitive version)
N: 8 µg/l (CLD)
2 mg/l (IR)
0.1 mg/l (ECD)

Precision: < 1% at > 5 mg/l C

Analysis time*: 3 - 4 min per parameter

Sample quantity*: liquid: < 50 – 2,000 μl variable


solid: up to 1 g soil or 10 mg C absolute (optional up to 50 mg C)

Calibration: Automatic multi point calibration

* depending on sample type and analysis mode


14
Specification vario TOC cube

Particle size*: exceeds the requirements EN 1484 for 100 μm cellulose standard,
with capsule technique no upper limit

Autosampler: • liquid samples manual


• solid samples manual
• automatic sampler for liquid samples with
32 (40 ml), 50 (12 ml) or 80 (6 ml) positions
all positions can be stirred
• automatic sampler for solid or particulates containing samples with
120 or 80 positions depending on sample size

Gases: Synthetic air (hydrocarbon and CO2 free) or oxygen (4.5), 500 ml/min

Instrument control: Operation and control via PC under Windows®;


All instrument functions are digitally controllable, the comprehensive operation software
includes e.g. automatic leak test, extensive error diagnosis, monitoring of the maintenance
cycles, sleep-/wake up function, statistical evaluation and almost unlimited memory
capacity for analysis data incl. graphics.
Integration in data networks like LIMS and the possibility of remote control and diagnosis
via the internet.
In full compliance with 21 CFR Part 11 (option).

Electrical connections: 100/110/200/230 VAC, 50/60 Hz, 1.8 kW

Dimensions: 48 × 55 × 57 cm (W × D × H)

Weight: approx. 60 kg

* depending on sample type and analysis mode


15