Best Practice & Research Clinical Anaesthesiology

Vol. 19, No. 4, pp. 581–593, 2005


doi:10.1016/j.bpa.2005.08.005
available online at http://www.sciencedirect.com

Supraglottic airway devices

B. Bein* MD
Staff anaesthesiologist
J. Scholz MD
Professor of Anaesthesiology and Intensive Care Medicine
Department of Anaesthesiology and Intensive Care Medicine, University Hospital Schleswig-Holstein, Campus Kiel,
Schwanenweg 21, D-24105 Kiel, Germany

Supraglottic airway devices are developed with increasing frequency following the over-
whelming success of the laryngeal mask airway (LMA). Currently, the LMA, the ProSeal
laryngeal mask airway (PLMA), the laryngeal tube (LT), the laryngeal tube with integrated
suctioning tube (LTS) and the oesophageal tracheal combitube (OTC) are the best evaluated
and most widespread devices. Both the LMA and the PLMA have been shown to be perfectly
suitable for routine anaesthesia and emergency airway management. While the LMA is limited
by the lack of reliable protection against aspiration, the value of the PLMA in this respect
remains undetermined yet. LT and LTS are primarily intended as emergency airway devices, but
have also been successfully used during controlled ventilation in adults. The OTC, though
advocated for emergency as well as routine use, is limited by high airway morbidity and
possible serious complications.

Key words: laryngeal mask airway; laryngeal tube; oesophageal tracheal combitube; airway
device; routine anaesthesia.

For the maintenance of life there is nothing more necessary than a sufficient supply
of fresh air. Robert Hooke, 1667

INTRODUCTION

Since the beginning of modern anaesthesia, airway management has been a challenge.
Following the enthusiasm raised by the first successful ether anaesthesias and the
subsequent widespread use, disastrous complications were reported.1 These could be

* Corresponding author. Tel.: C49 431597 3739; fax: C49 431597 3002.
E-mail addresses: bein@anaesthesie.uni-kiel.de (B. Bein), scholz@anaesthesie.uni-kiel.de (J. Scholz).

1521-6896/$ - see front matter Q 2005 Elsevier Ltd. All rights reserved.
582 B. Bein and J. Scholz

attributed, at least in part, to hypoxia due to airway obstruction. Therefore, a variety of


airway devices were developed to ensure airway patency.2 However, none of these
devices gained general acceptance, and apart from the Guedel airway they are
nowadays only of historical interest. The revolutionary advances of modern anaesthetic
practice are largely perceived by both anaesthesiologists and other specialties as
associated with the development of tracheal intubation. Consequently, for decades
there was only little interest in alternative airway devices, and airway management
focused on providing ventilation by bag mask, laryngoscopy and intubation. This attitude
has changed after the introduction of the laryngeal mask airway, which can be denoted a
milestone in the field of airway management.
Over the past decade, novel options for airway management have evolved and
numerous devices and techniques are currently available. For proper characterization
and classification, a definition is needed which criteria qualify a device as being
‘supraglottic’. Recently it has been suggested to replace the term ‘supraglottic’ by
the term ‘extraglottic’ in order to clarify the relationship between function and
anatomical position.3 Apart from its position relative to the glottis, criteria for
supraglottic airway devices include: appropriate bridging of the oral/pharyngeal space,
low resistance to respiratory gas flow, protection of the respiratory tract from
gastric and nasal secretions, suitability for positive pressure as well as spontaneous
ventilation, and finally lack of adverse events associated with its use.2 More
specifically, for the judgement of any new airway device we should first define what
we would expect from an ‘ideal’ airway device (Table 1) and how these expectations
or criteria are accomplished by the new device.4 The accept–reject profile is of
paramount importance in this respect. This property of a device, first mentioned by
AIJ Brain reporting on the development of the LMA, describes the relative potential
for acceptance or rejection of a foreign body by the oropharynx.5 Although patient-
specific factors may be relevant, this property largely depends on device shape, cuff
position, cuff volume and material and finally decides on the success of a device in
daily clinical routine. For the purpose of this review, we will focus on devices
advocated for routine use, which have already been evaluated based on a sufficient
number of publications allowing for reliable conclusions.

Table 1. Criteria for an ‘ideal’ airway device.

† Efficient bypass of the upper airway for ventilation


† Easy insertion by beginners, steep learning curve
† Efficacy not extremely affected by suboptimal placement
† Stable in use (i.e. suitable for ‘hands-free anaesthesia’)
† Good ‘accept–reject’ profile
† Minimal/no aspiration risk
† Effective upper airway seal allowing for positive pressure ventilation
† Cuff pressure and shape necessary for sealing does not distort/distend the pharynx
† Low airway morbidity
† Good quality (i.e. no device failure due to malfunction)

Modified according to Ref. 4.


Supraglottic airway devices 583

DEVICES AND TECHNIQUES

Laryngeal mask airway (LMA)

The LMA is an ingenious supraglottic airway device that was developed by AIJ Brain and
first described in 1983.6 The LMA emerged from a research project aimed to evaluate
more comfortable and less invasive alternatives to the established facemask and
tracheal tube, respectively.5 It consists of a curved tube, which is attached to the lumen
of a small, ellipsoid bowl with an inflatable cuff designed to provide a seal around the
laryngeal inlet. Two elastic bars are positioned over the bowl aperture to prevent
obstruction by the epiglottis. The device is available in the appropriate size to fit the
larynx of neonates up to large adults. After ensuring an adequate anaesthetic depth and
lubrication of the dorsal part of the mask the recommended standard procedure for
device insertion is with the mask aperture facing the base of the tongue and the cuff tip
pressed against the posterior pharyngeal wall. The index finger of the dominant hand is
used to guide the LMA in the hypopharynx until a resistance is felt which means that the
tip of the mask has reached the upper oesophageal sphincter. Alternative techniques
have been described, especially for use in children, which rely on introducing the device
upside down, followed by rotation.7 The cuff is then inflated with air until an effective
airway is established allowing appropriate ventilation of the lungs without relevant
leakage. The recommended cuff volume differs according to the LMA size used (for a
size 4 LMA for example !30 ml). Typical first time insertion success ranges between 90
and 95%8,9, and time to achieve an effective airway is approximately 30 seconds.10
Airway leak pressure mostly lies between 20 and 25 cm H2O.10,11 Malposition of the
LMA resulting in low airway leak pressure often is erroneously treated with cuff over
inflation. This, however, does not solve the problem but puts oropharyngeal mucosa at
risk of damage provoked by huge intracuff pressure transmitted to tissue. The
manufacturer, therefore, has recommended an upper intracuff pressure of 60 cm H2O.
It has been shown, that even at recommended cuff volumes, intracuff pressure may
result in cuff pressures largely above recommended values.12 This may be of clinical
significance, since in daily clinical practice, recommended intracuff pressure often is
ignored and cuff over inflation unnoticed, especially during short procedures with
spontaneously breathing patients.13 A reasonable approach is to inflate the cuff with the
lowest volume resulting in an effective airway followed by cuff pressure control. Cuff
pressure exceeding the recommended value should be treated by adjusting device
position (for example by gently pushing or pulling or a jaw thrust manoeuvre) or a
reinsertion attempt.
The LMA has an impressive performance, both regarding its application in daily
clinical routine and number of publications.8,9,14 More than 1500 publications deal with
the LMA in various patient populations including neonates and children, during a variety
of surgical procedures and different anaesthetic agents. At present, the LMA has been
used in approximately 150 million patients without an apparent death directly
associated with it. Currently, about 30–60% of all general anaesthetic procedures are
performed with a LMA. It may serve primarily as a substitute for the facemask, thereby
allowing for ‘hands free’ anaesthesia. Importantly, the LMA must not been used in
patients at an increased risk of aspiration regardless of the underlying reason. Further
contraindications comprise a decreased lung and/or thoracic compliance, an increased
airway resistance (chronic obstructive lung disease, acute bronchospasm), glottic or
subglottic airway obstruction, limited mouth opening and any oropharyngeal pathology.
584 B. Bein and J. Scholz

Compared with the classical facemask, the LMA offers several advantages.15 After
proper placement, ‘hands free’ anaesthesia is possible. Low flow anaesthesia may be
performed, and waste of anaesthetics as well as pollution of the operating theatre
endangering the personnel are decreased. Since oxygenation is easier accomplished
with the LMA compared with the facemask particularly in case of difficult patient
anatomy and at the beginning of resident training, a reduced incidence of hypoxia with
the LMA has been reported.16 With the LMA in correct position, the laryngeal inlet may
be visualized which allows for tracheal intubation in case of a difficult laryngoscopy.17
Compared with the tracheal tube, insertion of the LMA is not as painful as direct
laryngoscopy followed by inflating a cuff in the irritable tracheal lumen. Therefore, the
sympathetic activation is less pronounced which may be especially important in patients
with coronary artery disease.18,19 The incidence of post anaesthesia airway morbidity is
reduced, and coughing and retching are less often during emergence from
anaesthesia.20,21 Muscle relaxants are not applied, and thus neuromuscular monitoring
is not necessary and a residual block at the end of the surgical procedure not present.
Since the LMA has a favourable accept–reject profile compared with the tracheal tube,
the consumption of analgesics and anaesthetics is decreased.15 Therefore, the LMA is
cost-effective compared to both face mask and tracheal tube.22

Laryngeal mask airway Proseal (PLMA)

The PLMA is a meanwhile well known and established modified laryngeal mask which
has been improved especially with respect to airway seal and protection against
aspiration (Figure 1).23 Briefly, compared to the classic LMA, the PLMA features a new
dorsal cuff designed to push the ventral cuff into the periglottic tissue to form a better
seal than the classic LMA. Additionally, a drainage tube has been added, which allows

Figure 1. Laryngeal mask airway Proseal. A gastric catheter has been inserted in the drain tube.
Supraglottic airway devices 585

passage of a %18 Charriere gastric tube for emptying the stomach thus protecting
against regurgitation. The airway tube has been wire-reinforced to reduce stiffness of
the double-lumen configuration. The PLMA does not have the mask aperture bars since
the drainage tube is thought to serve for this purpose. The PLMA is marketed with a
special guiding handle to facilitate placement. Alternatively, placement may be
accomplished as described for the classic LMA or with an elastic bougie inserted in
the tube aimed at guiding the PLMA through the oropharyngeal inlet.24 The PLMA was
primarily designed to improve airway seal thus offering protection against aspiration.
This issue was addressed in a cadaver study comparing the LMA and the PLMA during
increasing oesophageal pressure.25 In this study, the correctly placed PLMA allowed
fluid to bypass the pharynx, and no fluid was seen above or below the cuff. Comparing
the efficacy of seal after clamping the drainage tube, the PLMA attenuated liquid flow
between oesophagus and pharynx significantly more than the LMA. It should be
emphasized, however, that protection against aspiration depends of an optimal PLMA
position on the laryngeal inlet. In a large multicenter trial comparing the PLMA with the
LMA regarding efficacy of seal and time to achieve an effective airway, in 4 patients in the
PLMA group (2%) the upper oesophageal sphincter could be visualized within the mask
bowl.10 In another 3 patients, folding over of the drainage tube occurred which
prevents fluid from being drained effectively. Therefore, the PLMA does not offer a
reliable protection against aspiration in all patients. Similar to the cadaver study, the
PLMA formed a more effective seal compared with the LMA, but time to achieve an
effective airway was shorter and fibre optic determined anatomic position better with
the LMA. Since handling is more complex and the PLMA is more expensive, its future
value has yet to be determined.

Intubating laryngeal mask (ILMA)

The LMA has been used successfully for the management of the difficult airway.17 In
1997, a modified LMA was developed, the intubating laryngeal mask airway (ILMA,
FastTrache). The ILMA, specifically designed to allow for tracheal intubation through
the device, is apparently not intended for routine use as an airway device for ventilation,
but rather for management of difficult tracheal intubation. Since the device is described
in detail in another chapter of this issue (Emergency Airway Management), it is not
further discussed here.

Laryngeal mask disposable

There is an increasing demand for disposable airway devices, since the process of
washing and preparation for repeat use is time consuming. More important, it has been
shown that even repeat autoclaving does not remove protein deposits from the
reusable LMA, thus potentially allowing disease transmission through residual biological
debris.26 On the other hand, it is conceivable that different materials used in the
reusable and disposable device may influence the accept–reject profile. In a recent case
report, the airway tube became completely separated from the distal mask during use
of a disposable LMA Uniquee, which raised concerns regarding device quality.27
However, in three recent trials comparing disposable laryngeal mask airways by
different manufacturers with the classic LMA, the disposable devices performed
comparable with respect to ease of insertion, fibre optic view, incidence of blood on the
mask at the time of removal, and the incidence of postoperative sore throat.28–30
586 B. Bein and J. Scholz

Interestingly, cuff pressure of the disposable LMAs did not increase during anaesthesia
with nitrous oxide, while there was a large increase with the classic LMA. Thus,
disposable devices apparently offer a good laryngeal seal and similar clinical
performance.

Laryngeal tube (LT)

The LT is a reusable, single-lumen, silicon tube with an oropharyngeal and oesophageal


low-pressure cuff, and a ventilation outlet between these cuffs (Figure 2). The device is
available in different sizes for use in neonates up to adults. The proximal tube is
equipped with a standardized 15 mm tracheal tube connector for attaching a bag valve
or breathing circuit. The colour of this connector differs according to LT size. The distal
lumen is positioned in the oesophageal inlet and equipped with a low pressure cuff to
prevent regurgitation and aspiration. A second, pharyngeal cuff seals the airway
proximally, and air is insufflated between these cuffs thus entering the glottic aperture.
After positioning the patient’s head in a neutral position, the LT is placed into the
oropharynx until a distinct resistance is felt. Following placement, both cuffs are inflated
via a common connecting tube using a pressure gauge up to an intracuff pressure of 60–
80 cm H2O. Alternatively (especially in an emergency) a special syringe with coloured
ring marks may be used to inflate a volume of 70 ml. It has been reported, that the
insufflated cuff volume closely matches the resulting in vivo cuff pressure. Nitrous oxide
may increase LT cuff pressure, and thus it is advisable to monitor and adjust intracuff
pressure during anaesthesia to minimize possible ischaemic damage to the
oropharynx.31 Since the diameter of the connecting tubes of the proximal and distal

Figure 2. Laryngeal tube (below) and laryngeal tube S (above). A gastric catheter has been introduced in the
drain tube.
Supraglottic airway devices 587

cuff are different, the oropharyngeal cuff is inflated first to accommodate in the
oropharyngeal cavity, and the distal cuff is filled thereafter.
The LT, although primarily designed as a device for emergency and out of hospital
use, has also been investigated during routine surgical procedures.11,32–38 During
controlled ventilation, it performed comparable to the LMA in most studies with
respect to insertion time, oropharyngeal leak pressure and peak airway pressure. One
study reported a decreased postoperative pharyngeal morbidity39, which is unlikely
given the higher mucosal pressures applied.40 In contrast, during spontaneous
ventilation, the LT was found inferior to the LMA in all three studies currently
available, mostly due to repeated manipulations necessary to maintain a patent
airway.36,38,41 Compared to the PLMA, the LT was found inferior in several clinical
aspects, since efficacy of ventilation was higher and number of airway interventions and
airway obstruction was significantly less frequent for the PLMA.33,42
The LT has also been studied in children and was found suitable for that purpose by
the authors.43,44 A failure rate between 5–12% and the need for subsequent
manipulations in 20–35% of patients, however, are significantly worse compared to
the results obtained in adults. Recently, a disposable LT has been developed, but
available data are limited. In conclusion, the LT is a suitable device for controlled
ventilation in adults, but both LMA and PLMA are superior in most technical aspects of
airway management.
LTS #4

Figure 3. Laryngeal tube S. The proximal cuff is positioned in the hypopharynx, while the distal cuff seals the
oesophageal inlet. Gastric drainage is accomplished by a separate lumen.
588 B. Bein and J. Scholz

Since sealing the oesophagus in case of excessive vomiting may be dangerous, the LT
has been modified and equipped with a suctioning tube (LTS).

Laryngeal tube S (LTS)

The device was developed based on the LT, and the main difference between both is
rooted in the second lumen of the LTS, allowing for free gastric drainage and evacuation
of stomach contents by an oesophageal catheter, but not for ventilation. The second
tube is suitable for insertion of a gastric tube up to a diameter of 14 Charriere (Figures 2
and 3). Technique of insertion and handling are as described previously for the LT. After
its initial description in 200245, several studies have compared the LTS with the
PLMA.12,35,46,47 There seem to be some advantages for the PLMA in terms of patient
comfort as reflected by a lower incidence of sore throat and dysphagia.12,46 While three
out of four studies indicated that the PLMA and LTS are similar with respect to both
physiologic and clinical function (insertion success and time, oropharyngeal leak
pressure, peak and plateau airway pressures, ability to pass a gastric tube), the study by
Cook and co-workers found the LTS inferior to the PLMA regarding these measures.35
On the other hand, the laryngeal tube S may offer an advantage in patients at an
increased risk of aspiration (non-fasted, gastro oesophageal reflux disease, emergency
cases) due to the safe airway seal together with the possibility of free gastric drainage. In
a former study comparing the LMA and LT using a model rubber pharynx, the LT
showed significantly increased storage capacities for regurgitated fluids.48 The LTS and
the PLMA have not been compared in this respect yet. From the data presented it can
be concluded that the LTS can be used to establish a safe and effective airway in
mechanically ventilated anaesthetized adult patients.

Figure 4. Oesophageal tracheal combitube. A gastric catheter has been introduced in the distal tube.
Supraglottic airway devices 589

Oesophageal-tracheal combitube (OTC)

The OTC, first described in 1987 by Frass and co-workers, has been introduced into
clinical practice as an emergency airway device.49 The OTC is a double-lumen, double-
cuffed tube that allows ventilation independent of its position, either in the oesophagus
or the trachea (Figure 4). This unique feature is achieved by two independent tubes
fused together both equipped with the standardized 15 mm tracheal tube connector. It
comprises a small volume tracheooesophageal and a large volume oropharyngeal cuff,
both equipped with a separate connecting tube for inflation. Between both cuffs, there
are eight small ventilation outlets allowing ventilation if the distal end of the OTC is
placed in the oesophagus. The OTC is available in two different sizes, 37 French and 41
French. The 37 French OTC is recommended for patients up to 180 cm height, while
the larger OTC is intended for use in taller patients.50 For insertion, the patient’s head
should be in the neutral position. The thumb of the non-dominant hand subsequently
opens the mouth, puts the tongue forward and performs a chin lift together with the
index finger. The OTC is then advanced along the tongue until the teeth are positioned
between the ring marks. The proximal cuff is inflated with 85 ml (100 ml), and the distal
cuff with 10 ml (15 ml). In most cases (90–95%), the distal lumen is positioned in the
oesophagus (which means that the OTC is extraglottic). Lung ventilation through the
proximal lumen and, therefore, oesophageal position of the distal lumen has to be
assured by capnography, chest movement and chest auscultation during ventilation of
the proximal lumen.
The OTC is primarily intended for use in an emergency and/or out of hospital
situation. In this environment it has proven useful and, therefore, has been implemented
in the ASA difficult airway algorithm.51 Moreover, some investigators suggest the OTC
for routine anaesthesia.50,52 The use of the OTC in daily routine, however, is highly
controversial.12,53–55 Apart from serious complications, (i.e. oesophageal or pharyngeal
injury56,57) a significantly increased airway morbidity and stress response were
reported comparing the OTC with both the LMA and the tracheal tube during routine
surgery12,18,58, which may be explained by the unphysiologically high cuff pressures
applied.12,59 Consequently, in its current application the OTC cannot be recommended
for this purpose urging calls for a modified design.55
According to the use of a device in daily clinical routine, a different proportion of
anaesthesiologists will become familiar with alternative airway devices.60 LMA, PLMA,
LT and LTS all show acceptable times for device insertion, airway leak pressures and
postanaesthesia airway morbidity. In contrast, the OTC has been shown to be inferior
in most aspects of practical use such as time to successful ventilation, number of failed
attempts, and handling as well as in terms of patient satisfaction and comfort. The high
cuff pressures necessary may be harmful for the oropharyngeal mucosa if applied for a
longer period of time. Postanaesthesia airway morbidity has gained widespread
attention, especially in a health environment where cost containment is essential and
patient satisfaction of high priority. A high incidence of sore throat and dysphagia as
reported with the OTC, may therefore discourage anaesthesiologists from using a
device probably causing patient dissatisfaction and increased length of hospital stay. This
has important ramifications also for OTC use in an emergency. Since appropriate
training is required, a device not suitable for daily clinical routine will have a
disadvantage compared with a device in which training may be accomplished during
regular working hours.
590 B. Bein and J. Scholz

Research agenda

† the ProSeal laryngeal mask airway should be thorougly evaluated regarding its
behavior during regurgitation of gastric fluids
† the impact of different materials on the accept–reject profile of an airway
device warrants further investigation
† for better comparison of airway devices, standardized criteria should be
defined which have to be adressed in clinical studies
† the cost benefit ratio for tracheal intubation versus use of a supraglottic airway
should be evaluated in large scale trials

FUTURE DEVELOPMENTS

Today, new airway devices are introduced with ever-increasing frequency. PAXpresse,
Laryvente, CobraPLAe, Airway Management Devicee, ELISHAe and others have
recently been developed. Every device claims an ‘edge’ over established or novel
competitors, in terms of either technical aspects or costs or both. However,
anaesthetists are unlikely to have the opportunity to use all of them and, more
important, do not have the time to become proficient in all of them. Therefore, it is a
reasonable approach to develop expertise in one or two tried and tested and
universally applicable techniques. Given appropriate training, the type of device is not
that important for clinical acceptable outcome. It should be emphasized, however, that
every new device first has to prove to be at least equivalent to the current clinical gold
standard, the LMA, with respect to physiologic properties and airway morbidity.

CONCLUSIONS

Both the LMA and the PLMA are established supraglottic airway devices with a large
body of evidence proving safety and efficacy. The PLMA may be advantageous with
respect to the efficacy of seal and protection against aspiration. Disposable LMAs
perform comparable to the reusable devices. The relatively new LT and LTS both are
suitable for routine use. Compared with the LMA and PLMA, they may be inferior with
respect to airway morbidity and technical aspects of ventilation while offering a better
protection against aspiration. The OTC, designed for emergency use, should not be
used in routine anaesthesia due to high airway morbidity. For the performance of all
devices it is important to realize that efficacy of seal may vary dependent on the

Practice points

† the laryngeal mask airway is the best evaluated supragottic airway device so far
† though new airway devices are developed in increasingly numbers, it is
reasonable to become proficient with only a few of them
† compared to the laryngeal mask airway, the laryngeal tube is inferior in terms of
airway morbidity and technical aspects of ventilation
† the oesophageal tracheal combitube is not suitable for routine anaesthesia due
to its high airway morbidity
Supraglottic airway devices 591

individual patient’s laryngopharyngeal anatomy. Thus, applying fixed cuff volumes may
not be recommended, may contribute largely to airway morbidity and should be
replaced also in the instruction manuals by insufflating the lowest cuff volumes resulting
in an acceptable airway leak pressure guided by meticulous cuff pressure monitoring.
Furthermore, intracuff pressure monitoring should become routine practice with any
extraglottic airway device .

REFERENCES

1. Whalen FX, Bacon DR & Smith HM. Inhaled anesthetics: an historical overview. Best Practice and Research.
Clinical Anaesthesiology 2005; 19: 323–330.
2. McIntyre JW. Oropharyngeal and nasopharyngeal airways: I (1880–1995). Canadian Journal of Anaesthesia
1996; 43: 629–635.
3. Brimacombe J & Keller C. The Elisha airway device: supraglottic and infraglottic, or simply extraglottic?
Anesthesia and Analgesia 2005; 100: 603.
4. Charters P. Airway devices: where now and where to? British Journal of Anaesthesia 2000; 85: 504–505.
5. Brain AI. The development of the Laryngeal Mask—a brief history of the invention, early clinical studies
and experimental work from which the Laryngeal Mask evolved. European Journal of Anaesthesiology.
Supplement 1991; 4: 5–17.
6. Brain AI. The laryngeal mask—a new concept in airway management. British Journal of Anaesthesia 1983;
55: 801–805.
7. Nakayama S, Osaka Y & Yamashita M. The rotational technique with a partially inflated laryngeal mask
airway improves the ease of insertion in children. Paediatric Anaesthesia 2002; 12: 416–419.
8. Brimacombe J. Analysis of 1500 laryngeal mask uses by one anaesthetist in adults undergoing routine
anaesthesia. Anaesthesia 1996; 51: 76–80.
9. Verghese C & Brimacombe JR. Survey of laryngeal mask airway usage in 11,910 patients: safety and efficacy
for conventional and nonconventional usage. Anesthesia and Analgesia 1996; 82: 129–133.
10. Brimacombe J, Keller C, Fullekrug B et al. A multicenter study comparing the ProSeal and Classic laryngeal
mask airway in anesthetized, nonparalyzed patients. Anesthesiology 2002; 96: 289–295.
11. Asai T, Kawashima A, Hidaka I & Kawachi S. The laryngeal tube compared with the laryngeal mask:
insertion, gas leak pressure and gastric insufflation. British Journal of Anaesthesia 2002; 89: 729–732.
12. Bein B, Carstensen S, Gleim M et al. A comparison of the proseal laryngeal mask airway, the laryngeal tube
S and the oesophageal-tracheal combitube during routine surgical procedures. European Journal of
Anaesthesiology 2005; 22: 341–346.
13. Matioc AA & Arndt G. The laryngeal tube and pharyngeal mucosal pressure. Canadian Journal of
Anaesthesia 2003; 50: 525–526.
14. Lopez-Gil M, Brimacombe J & Alvarez M. Safety and efficacy of the laryngeal mask airway. A prospective
survey of 1400 children. Anaesthesia 1996; 51: 969–972.
15. Brimacombe J. The advantages of the LMA over the tracheal tube or facemask: a meta-analysis. Canadian
Journal of Anaesthesia 1995; 42: 1017–1023.
16. Johnston DF, Wrigley SR, Robb PJ & Jones HE. The laryngeal mask airway in paediatric anaesthesia.
Anaesthesia 1990; 45: 924–927.
17. Benumof JL. Use of the laryngeal mask airway to facilitate fiberscope-aided tracheal intubation. Anesthesia
and Analgesia 1992; 74: 313–315.
18. Oczenski W, Krenn H, Dahaba AA et al. Hemodynamic and catecholamine stress responses to insertion
of the Combitube, laryngeal mask airway or tracheal intubation. Anesthesia and Analgesia 1999; 88: 1389–
1394.
19. Wood ML & Forrest ET. The haemodynamic response to the insertion of the laryngeal mask airway: a
comparison with laryngoscopy and tracheal intubation. Acta Anaesthesiologica Scandinavica 1994; 38: 510–
513.
20. Joshi GP, Inagaki Y, White PF et al. Use of the laryngeal mask airway as an alternative to the tracheal tube
during ambulatory anesthesia. Anesthesia and Analgesia 1997; 85: 573–577.
592 B. Bein and J. Scholz

21. Akhtar TM, McMurray P, Kerr WJ & Kenny GN. A comparison of laryngeal mask airway with tracheal
tube for intra-ocular ophthalmic surgery. Anaesthesia 1992; 47: 668–671.
22. Macario A, Chang PC, Stempel DB & Brock-Utne JG. A cost analysis of the laryngeal mask airway for
elective surgery in adult outpatients. Anesthesiology 1995; 83: 250–257.
23. Brain AI, Verghese C, Strube PJ & The LMA. ‘ProSeal’—a laryngeal mask with an oesophageal vent. British
Journal of Anaesthesia 2000; 84: 650–654.
24. Howath A, Brimacombe J & Keller C. Gum-elastic bougie-guided insertion of the ProSeal laryngeal mask
airway: a new technique. Anaesthesia and Intensive Care 2002; 30: 624–627.
25. Keller C, Brimacombe J, Kleinsasser A & Loeckinger A. Does the ProSeal laryngeal mask airway prevent
aspiration of regurgitated fluid? Anesthesia and Analgesia 2000; 91: 1017–1020.
26. Bannon L, Brimacombe J, Nixon T & Keller C. Repeat autoclaving does not remove protein deposits from
the classic laryngeal mask airway. European Journal of Anaesthesiology 2005; 22: 515–517.
27. Spielman FJ. Complete separation of the tube from the mask during removal of a disposable laryngeal
mask airway. Canadian Journal of Anaesthesia 2002; 49: 990–992.
28. Brimacombe J, Keller C, Morris R & Mecklem D. A comparison of the disposable versus the reusable
laryngeal mask airway in paralyzed adult patients. Anesthesia and Analgesia 1998; 87: 921–924.
29. Cao MM, Webb T & Bjorksten AR. Comparison of disposable and reusable laryngeal mask airways in
spontaneously ventilating adult patients. Anaesthesia and Intensive Care 2004; 32: 530–534.
30. Van Zundert AA, Fonck K, Al-Shaikh B & Mortier E. Comparison of the LMA-classic with the new
disposable soft seal laryngeal mask in spontaneously breathing adult patients. Anesthesiology 2003; 99:
1066–1071.
31. Asai T & Shingu K. Time-related cuff pressures of the laryngeal tube with and without the use of nitrous
oxide. Anesthesia and Analgesia 2004; 98: 1803–1806.
32. Dorges V, Ocker H, Wenzel V & Schmucker P. The laryngeal tube: a new simple airway device. Anesthesia
and Analgesia 2000; 90: 1220–1222.
33. Brimacombe J, Keller C & Brimacombe L. A comparison of the laryngeal mask airway ProSeal and the
laryngeal tube airway in paralyzed anesthetized adult patients undergoing pressure-controlled ventilation.
Anesthesia and Analgesia 2002; 95: 770–776.
34. Cook TM, McCormick B & Asai T. Randomised comparison of the laryngeal tube and the laryngeal mask
during anaesthesia with controlled ventilation. Anaesthesia 2003; 58: 508.
35. Cook TM & Cranshaw J. Randomized crossover comparison of ProSeal(R) Laryngeal Mask Airway with
Laryngeal Tube Sonda(R) during anaesthesia with controlled ventilation. British Journal of Anaesthesia 2005;
95: 261–266.
36. Chiu CL, Murugasu J & Chan L. The use of modified VBM laryngeal tube compared to Laryngeal Mask
Airway during spontaneous ventilation. Anaesthesia and Intensive Care 2003; 31: 187–192.
37. Asai T, Murao K & Shingu K. Efficacy of the laryngeal tube during intermittent positive-pressure
ventilation. Anaesthesia 2000; 55: 1099–1102.
38. Miller DM, Youkhana I & Pearce AC. The laryngeal mask and VBM laryngeal tube compared during
spontaneous ventilation: a pilot study. European Journal of Anaesthesiology 2001; 18: 593–598.
39. Wrobel M, Grundmann U, Wilhelm W, Wagner S & Larsen R. Laryngeal tube versus laryngeal mask
airway in anaesthetised non-paralysed patients. A comparison of handling and postoperative morbidity.
Anaesthesist 2004; 53: 702–708.
40. Brimacombe J, Keller C, Roth W & Loeckinger A. Large cuff volumes impede posterior pharyngeal
mucosal perfusion with the laryngeal tube airway. Canadian Journal of Anaesthesia 2002; 49: 1084–1087.
41. Figueredo E, Martinez M & Pintanel T. A comparison of the ProSeal laryngeal mask and the laryngeal tube
in spontaneously breathing anesthetized patients. Anesthesia and Analgesia 2003; 96: 600–605.
42. Cook TM, McKinstry C, Hardy R & Twigg S. Randomized crossover comparison of the ProSeal laryngeal
mask airway with the laryngeal tube during anaesthesia with controlled ventilation. British Journal of
Anaesthesia 2003; 91: 678–683.
43. Richebe P, Semjen F, Cros AM & Maurette P. Clinical assessment of the laryngeal tube in pediatric
anesthesia. Paediatric Anaesthesia 2005; 15: 391–396.
44. Genzwuerker HV, Hohl E & Rapp HJ. Ventilation with the laryngeal tube in pediatric patients undergoing
elective ambulatory surgery. Paediatric Anaesthesia 2005; 15: 385–390.
45. Doerges V, Ocker H, Wenzel Vet al. The laryngeal tube S: a modified simple airway device. Anesthesia and
Analgesia 2003; 96: 618–621.
Supraglottic airway devices 593

46. Roth H, Genzwuerker HV, Rothhaas A et al. The ProSeal laryngeal mask airway and the laryngeal tube
Suction for ventilation in gynaecological patients undergoing laparoscopic surgery. European Journal of
Anaesthesiology 2005; 22: 117–122.
47. Gaitini LA, Vaida SJ, Somri M et al. A randomized controlled trial comparing the ProSeal Laryngeal Mask
Airway with the laryngeal tube suction in mechanically ventilated patients. Anesthesiology 2004; 101: 316–
320.
48. Miller DM & Light D. Storage capacities of the laryngeal mask and laryngeal tube compared and their
relevance to aspiration risk during positive pressure ventilation. Anesthesia and Analgesia 2003; 96: 1821–
1822.
49. Frass M, Frenzer R, Zdrahal F et al. The esophageal tracheal combitube: preliminary results with a new
airway for CPR. Annals of Emergency Medicine 1987; 16: 768–772.
50. Gaitini LA, Vaida SJ, Mostafa S et al. The Combitube in elective surgery: a report of 200 cases.
Anesthesiology 2001; 94: 79–82.
51. Practice guidelines for management of the difficult airway: an updated report by the American Society of
Anesthesiologists Task Force on Management of the Difficult Airway. Anesthesiology 2003; 98: 1269–1277.
52. Urtubia RM, Aguila CM & Cumsille MA. Combitube: a study for proper use. Anesthesia and Analgesia 2000;
90: 958–962.
53. Krafft P, Hartmann T, Agro F et al. Is it unethical to use the Combitube in elective surgery patients?
Anesthesiology 2003; 98: 1022.
54. Mandal NG. Combitube and similar devices should undergo long-term safety evaluation before their
routine use in clinical practice. Anesthesiology 2001; 95: 1041–1042.
55. Urtubia RM & Medina J. The Combitube should be redesigned for anaesthetic use. Anaesthesia 2001; 56:
275.
56. Stoppacher R, Teggatz JR & Jentzen JM. Esophageal and pharyngeal injuries associated with the use of the
esophageal–tracheal Combitube. Journal of Forensic Science 2004; 49: 586–591.
57. Klein H, Williamson M, Sue-Ling HM et al. Esophageal rupture associated with the use of the Combitube.
Anesthesia and Analgesia 1997; 85: 937–939.
58. Oczenski W, Krenn H, Dahaba AA et al. Complications following the use of the Combitube, tracheal tube
and laryngeal mask airway. Anaesthesia 1999; 54: 1161–1165.
59. Keller C, Brimacombe J, Boehler M et al. The influence of cuff volume and anatomic location on
pharyngeal, esophageal, and tracheal mucosal pressures with the esophageal tracheal combitube.
Anesthesiology 2002; 96: 1074–1077.
60. Stringer KR, Bajenov S & Yentis SM. Training in airway management. Anaesthesia 2002; 57: 967–983.