Lux Level Test Certificate Form
Lux Level Test Certificate Form
PURPOSE
CUSTOMER DATA
Customer Plant
Area of Test Date of Test
INSTRUMENT USED FOR TESTING
Instrument Name Make
Model Serial Number
Calibrated On Calibration Due On
No of Locations 9
OBTAINED TEST RESULTS
Location ACCEPTANCE
Room Name Obtained Light Intensity Redaings in Lux Result
Number CRITERIA
L1 42 Average
L2 42
44
L3 45
The average Light
L4 42 Standard Deviation
Intensity Level should
L5 42 be within 400 Lux and
Pass
4 Relative Standard
L6 40 Deviation should not
exceed 15%
Relative Standard
L7 48
Deviation in %
L8 46
9%
L9 52
The above test Confirms the Noise Level Test in the specified Clean Air Equipment/ Cleanroom as per
INFERENCE:
the specified requirement of 65 dBA and the Relative Standard Deviation is within 15%