CHAPTER 1_2a:
INSTRUMENT
CALIBRATION
Zero shift calibration error - This error Span shift calibration error- This error’s
affects all calibration points equally, effect is unequal at different points
creating the same percentage of error throughout the range moving the
across the entire range moving the “span” adjustment
“zero” adjustment
CALIBRATION – TYPICAL ERRORS
Linearity calibration error - This type of error does Hysteresis calibration error - Occurs when the
not directly relate to a shift in either zero (b) or instrument responds differently to an increasing
span (m) because the input compared to a decreasing input cannot
slope-intercept equation only describes straight be rectified by simply making calibration
lines The behavior of a linearity adjustment is adjustments to the instrument –
unique to each model of instrument, and so you one must usually replace defective components
must consult the manufacturer’s documentation or correct coupling problems within the instrument
mechanism.