5 Bipolar Currents Total Safety Concept

for all surgical needs with bipolar cut and with complete self-test, hardware/software diagnosis and total
coagulation. AUTO mode with automatic function and safety monitoring by two 32 bit microprocessors.
Start/ Stop System and impendance sen- Dynamic auto-adjustment of output powers for best performance
sing. while using the lowest possible powers and reducing damaging
side effects on target points (i.e. carbonization or tissue sticking).

Powerful Performance

Output 200W max. in monpolar pure


cut and 140W max. in bipolar pure
cut. Constant output power levels
through the output power control
system make this unit particularly
suitable for cutting under water (e.g.
during TUR).

Up to 2 Twin Foot Pedals


One for monpolar and bipolar use
(REF 34-9330) and one for indepen-
dentbipolar only use (REF 34-9331 -
optional). Water proof (IPX7) with low
voltage, safe medical grade design.

1 Bipolar Output Ideal for Cardio-Surgery

for cables with 2x4mm USA- “Floating type” insulated at high/low


type banana plug connectors frequencies and protected against
(e.g. REF 34-9408) for bipolar the use of defibrillator and a preset
cutting and coagulation with cardio program make this unit an
maximum safety and precision. ideal choice for cardio-surgery.
Model Order REF Order REF
Type 230V 115V

ESU-A 400HT 34-9300 34-9300.US


ESU-X 400HT 34-9302 34-9302.US
ESU-X 300NT 34-9310 34-9310.US
ESU-X 200NT 34-9312 34-9312.US

Technical Specifications ESU HT/NT Series


ESU-A400HT ESU-X 400HT

Currents Output P/P Outout P/P Nominal Frequency(F)- Crest Factor(CF) - Acoustic/Visual
Power Voltage Power Voltage Load Modulation(M)- Duty Cycle (OT) Signal

MONOPOLAR:
Pure 400W 3465V 400W 3465V 3600 F: 435 kHz- CF: 1.6 - M: N/A- OT: N/A Low tone, yellow light
Blend 1 300W 3595V 300W 3595V 3600 F: 435 kHz- CF: 2.3 - M: 29 kHz - OT: 65 % Low tone, yellow ligh
Blend 2 140W 7600V 140W 7600V 6200 F: 435 kHz- CF: 8.1 - M: 19 kHz - OT: 9 % Low tone, yellow light
A. Pure 400W 1470 V 400W 1470 V 3600 F: 435 kHz- CF: 1.6 - M: N/A- OT: N/A Low tone, yellow light
A. Blend 300W 1930 V 300W 1930 V 3600 F: 435 kHz- CF: 2.3 - M: 29 kHz - OT: 65 % Low tone, yellow light
A. Endo 250W 1890V 250W 1890V 3600 Alternating 50% Pure and Blend1 - CF: 2,2 Low tone, yellow light
Fulg 150W 4700V 150W 4700V 345 0 F: 435 kHz- CF: 4.5 - M: 78 kHz - OT: 3,5 % High tone, blue light
Pin Point 250W 3460V 250W 3460V 2500 F: 435 kHz- CF: 2.6 - M: 29 kHz - OT: 50 % High tone, blue light
Soft 280W 3440V 280W 3440V 2500 F: 435 kHz- CF: 2 5 - M: 29 kHz - OT: 56 % High tone, blue light
Spray 140W 7660V 140W 7660V 6300 F: 435 kHz- CF: 8.1 - M: 19 kHz- OT: 9 % High tone, blue light

BIPOLAR:
Pure 160W 850V 160W 850V 310 0 F: 435 kHz- CF: 1.5 - M: N/A- OT: N/A Low tone, yellow light
Blend 130W 1000V 130W 1000V 310 0 F: 435 kHz- CF: 1.8 - M: 29 kHz - OT: 75 % Low tone, yellow light
Micro 130W 450V 130W 450V 100 0 F: 435 kHz- CF: 1.7 - M: N/A- OT: N/A Low tone, yellow light
Precise 130W 760 V 130W 760V 100 0 F: 435 kHz- CF: 1. 7 - M: N/A- OT: N/A Low tone, yellow light
Auto 130W 450V 130W 450V 100 0 F: 435 kHz- CF: 1. 7 - M: N/A- OT: N/A Low tone, yellow light
Macro 130W 760 V 130W 760V 100 0 F: 435 kHz- CF: 1. 7 - M: N/A- OT: N/A High tone, blue light
Sealing 130W 710V 130W 710V 100 0 F: 435 kHz- CF: 1. 7 - M: N/A- OT: N/A High tone, blue light

ESU·X 300NT ESU-X 200NT

Currents Output P/P Output P/P Nominal Frequency(F)- Crest Factor(CF) - Acoustic/Visual
Power Voltage Power Voltage Load Modulation(M)- Duty Cycle (OT) Signal

MONO POLAR:
Pure 350W 3455 V 200W 3455 V 3600 F: 435 kHz- CF: 1.6 - M: N/A- OT: NIA Low tone, yellow light
Blend 1 300W 3595 V 200W 3530V 3600 F: 435 kHz- CF: 2.3 - M: 29 kHz - OT: 65 % Low tone, yellow ligh
Blend 2 140W 7600V 140W 7600V 6200 F: 435 kHz- CF: 8.1 - M: 19 kHz - OT: 9 % Low tone, yellow light
Endo 220W 1720 V 200W 1700V 3600 Alternating 50% Pure and Blend1 - CF: 2,2 Low tone, yellow light
Fulg 150W 4700V 150W 4700V 345 0 F: 435 kHz- CF: 4.5 - M: 78 kHz - OT: 3,5 % High tone, blue light
Pin Point 250W 3460 V 200W 3450V 2500 F: 435 kHz- CF: 2.6 - M: 29 kHz - OT: 50 % High tone, blue light
Soft 280W 3440 V 200W 3435 V 2500 F: 435 kHz- CF: 2.5 - M: 29 kHz - OT: 56 % High tone, blue light
Spray 140W 7600V 140W 7600V 6300 F: 435 kHz- CF: 8 1 - M: 19 kHz - OT: 9 % High tone, blue light

BIPOLAR:
Pure 140W 840 V 140W 840V 3100 F: 435 kHz- CF: 1.5 - M: NIA- OT: N/A Low tone, yellow light
Blend 120W 990V 120W 990V 3100 F: 435 kHz- CF: 1.8- M: 29 kHz - OT: 75 % Low tone, yellow light
Micro 120W 440 V 120W 440V 100 0 F: 435 kHz- CF: 1.7 - M: N/A- OT: N/A Low tone, yellow light
Auto 120W 440 V 120W 440 V 1000 F: 435 kHz- CF: 1.7 - M: N/A- OT: N/A Low tone, yellow light
Macro 120W 750V 120W 750V 1000 F: 435 kHz- CF: 1.7- M: NIA- OT: N/A High tone, blue light

Mains Power: 230/115Volt+/- 10%/ Hz : 50/60 Phases: alternate single-phase. Electrical input: at max. power: 828 VA- at switching on : 840VA- <
in stand by: 100VA. Mains switch: bipolar with green light, positioned on rear. Supply cable: detachable with standard SCHUKO plug (different plugs �
inquiry), length 3m. Fuses: 2 fuses in the appliance inlet. Protective earth resistance: Lower than 0,1 ohm . Equipotential connection : by DIN 42801 �
inti. plug. Dimensions: WxDxH 38cm x 35cm x 16cm, Weight :15 kgs (ESU-A400HT: 16kgs). Installation: units can be used on a trolley, shelf or
hanging arm. Special conditions/facilities for the installation: None (no water, gas, compressed air, etc. required). Electric supply must meet the appli-
ffi
0
cable safety standards. Loss of heat in the environment: None. Limits of environmental temperature / humidity during the functioning or the store: z
standard IEC limits - temp from 10° to 40°C I humid. from 35 to 75% I press. from 700 to 1060 hPa. Cooling: by convection without fan. Twin foot- <ii
-o
switch pedals: electrical type, water -proof (IPX7) for operating rooms, with low, medical, voltage 12VDC Argon gas section (only for model ESU-A O
400HT): Gas supply: by 1/2 cylinders with 511 (cylinder full= pressure 200 ATM) with pressure reducer or by external central pipeline - Connection for �
gas supply: Quick-action coupling connector - Max working pressure and maximum gas flow: 1atm, 15I/min - Adjustment of gas flow: from 1 to 15
I/min, setting by push-buttons and displayed on progressive LED-bar. Warranty 2 years. Errors, technical alterations and all rights reserved.
§