Professional Documents
Culture Documents
KO00094Q VC150 Service Manual - ENG
KO00094Q VC150 Service Manual - ENG
KO00094Q VC150 Service Manual - ENG
For technical documentation purposes, the abbreviation GE is used for the legal entity name, GE Medical Systems
Information Technologies, Inc.
Listed below are GE Medical Systems Information Technologies, Inc. trademarks. All other trademarks contained herein
are the property of their respective owners.
GE TruSignal is the property of GE Medical Systems Information Technologies, Inc., a division of General Electric
Corporation. All other product and company names are the property of their respective owners.
Description of NIBP algorithm © GE Medical Systems Information Technologies, Inc., reproduced by permission.
CARESCAPE, DINAMAP, CLASSIC -CUF, CLINI-CUF, Critikon, DINACLICK, DURA-CUF, RADIAL-CUF, SENSA-CUF, SOFT-CUF
Blood Pressure Cuffs, and SuperSTAT are trademarks of GE Medical Systems Information Technologies, Inc.
Welch Allyn® and SureTemp® Plus are registered trademarks of Welch Allyn, Inc.
Masimo rainbow® SET®, RD, LNOP, LNCS, MLNCS and Signal IQ are registered trademarks of Masimo Corporation.
Possession or purchase of this device does not convey any express or implied license to use the device with
unauthorized sensors or cables which would, alone or in combination with this device, fall within the scope of one or
more of the patents relating to this device.
Medtronic, Nellcor, OxiMax, Saturation Pattern Detection and SatSeconds are trademarks of Medtronic.
GE Medical Systems Information Technologies, Inc. is the exclusive distributor of the CARESCAPE VC150.
1 Introduction . . . . . . . . . . . . . . . . . . . . . . 1-1
Manual purpose . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Intended audience . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Additional information on product use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
General use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
Revision history . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
Printed copies of this manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
Conventions used in this manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
Related manuals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
Equipment ID . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Safety information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Responsibility of the manufacturer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-6
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-7
Safety information and product-specific hazards . . . . . . . . . . . . . . . . . . . .1-7
Dangers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-8
Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-8
Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-11
Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-13
CARESCAPETM VC150 vital signs monitor . . . . . . . . . . . . . . . . . . . . . . . . . . 1-15
Product compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-17
Open source license texts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-19
GPL v3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-19
LGPL v2.1 & . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-20
Equipment symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-20
Symbols for Exergen temporal scanner . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-22
Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-24
Contact information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-24
Service policy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-24
Service agreements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-24
Assistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-24
Packing instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-24
Insurance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-25
Repair parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-25
Disposal of product waste . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-25
Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-25
Patient applied parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-25
Packaging material . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-26
Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-26
3 Installation . . . . . . . . . . . . . . . . . . . . . . . 3-1
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Workflow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-2
Installation checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-2
Unpacking the unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-2
Tool required for unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-2
Unpacking the monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-2
First charge of the battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-3
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-4
Monitor on/off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
Turning the monitor on . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-6
Touch screen recalibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-7
Verification of the touch screen recalibration . . . . . . . . . . . . . . . . . . . .3-8
Turning the monitor off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-8
Resetting the monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-8
Changes in settings due to standby/shutdown/reset . . . . . . . . . . . . . . . . .3-9
Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10
Remote alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
Strain relief . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
Configuring the monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
9 Maintenance . . . . . . . . . . . . . . . . . . . . . 9-1
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-2
Preventive maintenance checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-3
Visual inspection checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-5
Connecting to the Service PC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-6
Disconnecting from the service PC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-7
Remote service mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-7
Backing up and importing/restoring settings . . . . . . . . . . . . . . . . . . . . . 9-12
What can be copied . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-12
File import . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-13
Exporting settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-14
Importing settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-14
警告 本维修手册仅提供英文版本。
(ZH-CN) • 如果维修服务提供商需要非英文版本,客户需自行提供翻译服务。
• 未详细阅读和完全理解本维修手册之前,不得进行维修。
• 忽略本警告可能对维修人员,操作员或患者造成触电、机械伤害或其
他形式的伤害。
警告 本維修手冊只提供英文版。
(ZH-TW) • 如果客戶的維修人員有英語以外的其他語言版本需求,則由該客戶負責
提供翻譯服務。
• 除非您已詳閱本維修手冊並了解其內容,否則切勿嘗試對本設備進行維
修。
• 不重視本警告可能導致維修人員、操作人員或病患因電擊、機械因素或
其他因素而受到傷害。
Tento provozní návod existuje pouze v anglickém jazyce.
Vpřípadě, že externí služba zákazníkům potřebuje návod v jiném jazyce, je zajištění
překladu do odpovídajícího jazyka úkolem zákazníka.
VAROVÁNÍ Nesnažte se o údržbu tohoto zařízení, aniž byste si přečetli tento provozní návod a
(CS) pochopili jeho obsah.
V případě nedodržování této varování může dojít k poranění pracovníka prodejního
servisu, obslužného personálu nebo pacientů vlivem elektrického proudu, respektive
vlivem mechanických či jiných rizik.
警告 このサービスマニュアルは英語版しかありません。
(JA) ・サービスを担当される業者が英語以外の言語を要求される場合、翻訳
作業はその業者の責任で行うものとさせていただきます。
・このサービスマニュアルを熟読し、十分に理解した上で装置のサービ
スを行ってください。
・この警告に従わない場合、サービスを担当される方、操作員あるいは
患者が、感電や機械的又はその他の危険により負傷する可能性があり
ます。
Manual purpose
This manual supplies technical information for service representatives and
technical personnel so they can maintain the equipment to the assembly level.
Use it as a guide for maintenance and electrical repairs considered field
repairable. Where necessary, the manual identifies additional sources of
relevant information and technical assistance. While the service manual and
operator manual share some content, the service manual is more technically
oriented and the operator manual is more everyday user oriented. Ordinary
screen and default setup items are not repeated in the service manual. Refer to
the operator's manual for the instructions necessary to operate the equipment
safely in accordance with its function and intended use.
Following checklists provide list of items to check for during initial installation or
preventive maintenance:
Intended audience
This manual is intended for service representatives and technical personnel who
maintain, troubleshoot, or repair this equipment and who have been authorized
and trained by GE.
Intended use
The CARESCAPE VC150 is intended to monitor a single patient's vital signs at the
site of care or during intra-hospital transport.
The optional GE TruSignal pulse oximetry and accessories are indicated for
continuous noninvasive monitoring of functional oxygen saturation (SpO2) and
pulse rate, including monitoring during conditions of clinical patient motion or
low perfusion, with adult, pediatric, and neonatal patients.
The optional Masimo Rainbow SET® Pulse CO-oximetry and accessories are
indicated for the continuous noninvasive monitoring of:
The optional Nellcor™ oximetry and accessories are indicated for the
continuous, noninvasive monitoring of arterial oxygen saturation (SpO2) and
pulse rate of adult, pediatric, and neonatal patients during both motion and non-
motion conditions, and for patients who are well or poorly perfused. The optional
Oximax SPD™ Alert (SPD) feature is intended only for facility-use care of adults to
detect patterns of desaturation indicative of repetitive reductions in airflow
through the upper airway and into the lungs.
The portable device is designed for use in hospitals and hospital-type facilities.
The CARESCAPE VC150 can also be used in satellite areas or alternate care
settings.
"Portable" refers to the ability of the CARESCAPE VC150 to be easily moved by the
caregiver, such as on a roll stand. The CARESCAPE VC150 is not intended to be
used for continuous monitoring during patient transport.
CAUTION
This device is not intended for home use. Federal law (U.S.A.)
restricts this device to sale by or on the order of a physician.
The CARESCAPETM VC150 vital signs monitor is intended to monitor one patient
at a time in a clinical setting with a caregiver present.
WARNINGS
The monitor is not intended for use as critical care monitor.
The monitor is intended for spot check monitoring where caregiver is present by
the patient. The monitor generates alarms and can be used for continuous
monitoring only when applicable option license is installed.
General use
CARESCAPE VC150 vital signs monitors are intended for use in various markets,
from the physician’s office to sub-acute triage and medical/surgical units. The
CARESCAPE VC150 vital signs monitor is intended to monitor one patient at a
time in a clinical setting.
Revision history
Printed copies of this manual
The first letter shown in this revision history table is the first customer-released
version of this document.
Revision Comments
Related manuals
Manual Title
Equipment ID
Every GE device has a unique serial number for identification. A sample of the
information found on a serial number label is shown below.
### ## ## #### # #
A B C D E F
Description
B Year manufactured
E Manufacturing site
F Miscellaneous characteristic
Safety information
The information presented in this section is important for the safety of both the
patient and operator. This section describes how the terms Danger, Warning,
Caution, and Note are used throughout the manual. In addition, standard
equipment symbols are defined.
General
Service personnel must also read and understand the CARESCAPE VC150 Vital
Signs Monitor Operator's Manual, pn KO00065.
This device is intended for use under the direct supervision of a licensed health
care practitioner.
Contact GE for information before connecting any devices to the equipment that
are not recommended in this manual.
Parts and accessories used must meet the requirements of the applicable IEC/
EN 60601 series safety standards, and/or the system configuration must meet
the requirements of the IEC 60601-1 medical electrical systems standard.
Periodically, and whenever the integrity of the device is in doubt, test all
functions.
The use of accessory equipment not complying with the equivalent safety
requirements of this equipment may lead to a reduced level of safety of the
resulting system. Consideration relating to the choice shall include:
use of the accessory in the patient vicinity; and
evidence that the safety certification of the accessory has been performed
in accordance to the appropriate IEC 60601-1 and/or IEC 60601-1
harmonized national standard.
If the device is going to be installed in the USA and use 240V rather than 120V,
the source must be a center-tapped, 240V, single-phase circuit.
DANGER indicates a hazardous situation that, if not avoided, will result in death
or serious injury.
NOTE provides application tips or other useful information to ensure that you
get the most from your equipment.
Dangers
The following dangers apply to this product:
DANGERS
Do not service the battery while the monitor is connected to
external power.
Warnings
The following warnings apply to this product:
WARNINGS
Do not immerse the monitor in water. If the monitor is splashed
with water or becomes wet, wipe it immediately with a dry
cloth.
Arrange the external AC/DC power converter, air hoses, and all
cables carefully so they do not constitute a hazard.
WARNINGS
To avoid the risk of electric shock, this equipment must only be
connected to a supply mains with protective earth.
Use only accessories approved for use with the monitor. Failure
to use recommended accessories may result in inaccurate
readings.
WARNINGS
Keep the Exergen scanner secured when it is not in use.
Prior to using a cleaning agent, read its instructions for use and
adhere to provided safety precautions.
DO NOT use hard or sharp objects to clean the probe well. This
could damage it and cause the unit to not function properly.
DO NOT use steam, heat or gas sterilization on the
thermometer or probe. DO NOT autoclave the probe well.
WARNINGS
When the monitor and the battery have been stored below
room temperature, there is a risk for moisture condensation
and electric shock when the monitor is brought to room
temperature.
Nellcor: If the perfusion index falls below 0.5%, the SpO2 values
may be inaccurate.
Cautions
The following cautions apply to this product:
CAUTIONS
This device is not intended for home use. Federal law (U.S.A.)
restricts this device to sale by or on the order of a physician.
CAUTIONS
The monitor is designed to conform to Electromagnetic
Compatibility (EMC) standard IEC 60601-1-2 and will operate
accurately in conjunction with other medical equipment that
also meets this requirement. To avoid interference problems
affecting the monitor, do not use the monitor in the presence of
equipment that does not conform to these specifications.
Do not exceed a load weighing 5 lbs. (2.5 kg) in the Welch Allyn
accessory compartment on the right side of the monitor.
Be careful not to bend flex cables. A bent section may break the
flex cable.
Notes
The following notes apply to this product:
NOTES
Use only the original cord, or a power cord that has been specified for this
monitor. Refer to CARESCAPE VC150 Supplies and accessories (KO00125).
The monitor is designed so that it can be powered without a battery while its
power supply is connected to mains. However, clinical mode is not possible
without a functional internal Lithium-ion battery specified for
CARESCAPE VC150. If the monitor is on without a battery, a white screen will
cover the parameters in the home screen.
While in configuration mode, all entries stored in the clinical history are
erased when the time and/or date is changed.
The license file contains all purchased licenses. Each feature requires an
individual, purchased license. The license/licenses is/are activated
automatically.
While an NIBP simulator device may be useful to verify that the monitor
responds to oscillometric pulsations, it should not be used as a basis for
assessing the accuracy of measurement. For more information, see
“Appropriate use of NIBP simulators” on page B-2.
All devices are tested and calibrated during manufacturing and are certified
for operation at installation.
The NIBP functional test does not test for measurement accuracy.
NOTES
If the Exergen scanner is not used regularly, remove the battery to prevent
possible damage due to chemical leakage.
If you want to use cleaning agents other than those specifically approved
and listed in the “Cleaning” on page 9-15, you can do it at your own
responsibility. However, bear in mind that using non-approved cleaning
agents voids product warranty on all monitor parts that may come in
contact with a non-approved cleaning agent. If you still want to use a non-
approved cleaning agent, first test it on the corner of the touch screen, not
on the whole screen. Remember to wipe off possible residues or stains with
a cloth dampened with water.
Betadine may discolor the case. Use a 10% solution of bleach to remove the
discoloration. Apply the listed solutions with a dampened sponge, soft
brush, or a cloth, then wipe dry with a clean cloth or towel.
When storing the product for extended periods, the battery must be
disconnected.
NIBP performance and accuracy have only been confirmed using GE hoses
and cuffs.
Use only Exergen probe covers. The size and shape of the probe covers can
affect the performance of the scanner. Inaccurate readings occur unless the
proper probe covers are used. Refer to CARESCAPE VC150 Supplies and
accessories (KO00125) for reorder part numbers.
Any supplied accessories that do not affect EMC compliance are not
included in the list of approved EMC accessories in this manual.
The monitor must be protected by a firewall and must not be exposed to the
internet.
NOTES
This device complies with part 15 of the FCC Rules. Operation is subject to
the following two conditions: (1) This device may not cause harmful
interference, and (2) this device must accept any interference received,
including interference that may cause undesired operation.
This equipment has been tested and found to comply with the limits for a
Class B digital device, pursuant to part 15 of the FCC Rules. These limits are
designed to provide reasonable protection against harmful interference in a
residential installation. This equipment generates, uses and can radiate
radio frequency energy and, if not installed and used in accordance with the
instructions, may cause harmful interference to radio communications.
However, there is no guarantee that interference will not occur in a
particular installation. If this equipment does cause harmful interference to
radio or television reception, which can be determined by turning the
equipment off and on, the user is encouraged to try to correct the
interference by one or more of the following measures:
5150-5250MHz:
Only indoor use is permitted
200 mW e.i.r.p.
Netherlands
5250-5350MHz:
Only indoor use is permitted
1 W e.i.r.p.
Product compliance
The CARESCAPE VC150 Vital Signs Monitor is classified in the following
categories for compliance with IEC 60601-1:
Internally powered or Class I when powered from external supply.
Portable.
For continuous operation.
Not suitable for use in the presence of flammable anesthetics.
Not for use in the presence of an oxygen-enriched atmosphere (oxygen
tent).
Type BF defibrillator-proof applied parts.
IPX1, degree of protection against ingress of water.
Sterilization/disinfection, refer to “Maintenance” on page 9-1.
Software is developed in accordance with IEC 62304.
The VC150 monitor complies with IEC 60601-2-49.
The alarm system is developed in accordance with IEC 60601-1-8.
The CARESCAPE VC150 monitor is a Group 1 Class B device: Group 1
contains all ISM (Industrial, scientific and medical) equipment in which there
is radio-frequency energy that is intentionally generated and/or
conductively coupled and is necessary for the internal functioning of the
equipment itself. Class B equipment is equipment suitable for use in all
establishments.
The SpO2 parameter complies with ISO 80601-2-61.
The NIBP parameter complies with ISO 80601-2-56.
Welch Allyn temperature parameter complies with ISO 80601-2-56 and
ASTM E 1112-00 (Table 1). Exergen temperature parameter complies with EN
12470-5.
Defibrillation protected. When used with the recommended accessories, the
monitor is protected against the effects of defibrillator discharge. If
monitoring is disrupted by the defibrillation, the monitor will recover.
Declaration of Conformity
Hereby, Innokas Yhtymä Oy declares that the radio equipment type IEEE 802.11
a/b/g/n is in compliance with the Directive 2014/53/EU. The full text of the EU
declaration of conformity is available at the following internet address.
http://www.innokasmedical.fi/support
Wireless compliance
This product complies with IEEE 802.11-2007 and 802.11 a/b/g/n protocols
for wireless networking. This product also complies with IEEE 802.11e and
WMM Quality of Service guidelines.
This product supports 802.11n Single-Input, Single-Output (SISO) only.
This product supports 802.11i Personal and Enterprise-grade security and
IEEE 802.11h feature - Dynamic Frequency Selection (DFS).
Le présent appareil est conforme aux CNR d'Industrie Canada applicables aux
appareils radio exempts de licence. L'exploitation est autorisée aux deux
conditions suivantes : (1) l'appareil ne doit pas produire de brouillage, et (2)
l'utilisateur de l'appareil doit accepter tout brouillage radioélectrique subi, même
si le brouillage est susceptible d'en compromettre le fonctionnement.
Dealer No:
DA0083560/12
Complies with
IDA Standards
IDA Dealer license No.
DA107248
This license, commonly known as the GPL, has two versions that are actively and
widely used in many open source communities:
Most GPL version 2 software is actually licensed as "GPL version 2 or any later
version", or words to similar effect, which is commonly understood as a grant
under GPL version 2, section 9 of an "option to follow the terms and conditions . .
. of any later version" of the GPL. In this case, even if you are not the original
licensor of the GPL version 2 software, you can "follow the terms and conditions"
of GPL version 3 instead. This is not commonly done (at least in any explicit way),
however.
There are three versions of the license commonly known as the LGPL:
GNU Library General Public License, version 2 (SPDX short identifier: LGPL-
2.0)
GNU Lesser General Public License, version 2.1 (SPDX short identifier: LGPL-
2.1)
GNU Lesser General Public License, version 3 (SPDX short identifier: LGPL-
3.0)
Equipment symbols
The following symbols are associated with the CARESCAPE VC150 Vital Signs
Monitor.
NOTE
The model of the monitor determines which symbols appear on it.
Power button
YYYY-MM
USB port.
hPa
% Humidity limitations.
Temperature limitations.
YYYY-MM-DD
Measurement On button.
Service
Contact information
Refer to “How To Reach Us…” on page D-1 for contact information for service or
customer support.
Refer to the back page of this manual for manufacturer or distributor contact
information.
Service policy
The warranty for this product is enclosed with the product in the shipping carton.
All repairs on products under warranty must be performed or approved by
product service personnel. Unauthorized repairs will void the warranty. Only
qualified electronics service personnel should repair products not covered by the
warranty.
Service agreements
Check with GE sales organization whether any extended warranty claims are
valid.
Assistance
If the product fails to function properly, or if assistance, service or spare parts
are required, contact Customer Support. Before contacting Customer Support, it
is helpful to attempt to duplicate the problem and to check all accessories to
ensure that they are not the cause of the problem. If you are unable to resolve
the problem after checking these items, contact GE. Prior to calling, please be
prepared to provide:
product name, model number, and serial number
a complete description of the problem
If repair parts or service are necessary, you will also be asked to provide:
the facility's complete name, address, and account number
a purchase order number if the product needs repair or when you order
spare parts
the facility's GE account number, if possible
the appropriate part number for spare or replacement parts
Packing instructions
If you have to return goods for service, follow recommended packing
instructions in “Packing” on page 9-21.
Insurance
Insurance is at the customer's discretion. The shipper must initiate claims for
damage to the product.
Repair parts
To order parts, contact Service Parts at an address or a telephone in section
“How To Reach Us…” on page D-1.
Please allow one working day for confirmation of your order. All orders must
include the following information.
Facility's complete name, address, and phone number
FAX number
Your purchase order number
Your GE account number
Batteries
WARNING
Do not incinerate the battery or expose it to fire or high
temperatures. Doing so may cause the battery to explode.
Other patient applied parts, such as blood pressure cuffs, should be cleaned and
disinfected according to manufacturer’s instructions. Inspect reusable applied
parts for wear, replace as necessary, and dispose of used product as medical
waste according with regional body controlled guidelines.
Packaging material
Retain original packaging materials for future use in storing or shipping the
monitor and accessories. This recommendation includes corrugated shippers
and foam/corrugated spacers.
Monitor
At the end of its service life, the product described in this manual, as well as its
accessories, must be disposed of in compliance with the guidelines regulating
the disposal of such products. If you have questions concerning disposal of the
product, please contact GE or its representatives.
Product availability
NOTES
Due to continual product innovation, design and specifications for these
products are subject to change without notice.
Some of the products mentioned in this manual may not be available in all
countries. Please consult your GE representative for the availability.
SuperSTAT license can be enabled with a license key that is not available in
all countries. Contact your local GE representative about availability.
SuperSTAT license is not currently available in Canada.
Equipment description
The CARESCAPE VC150 Vital Signs Monitor provides a small, portable monitoring
alternative for sub-acute hospital and non-hospital settings. The monitor is for
use on adult, pediatric, or neonatal patients—one at a time. The battery-
operated monitor offers noninvasive determination of systolic blood pressure,
diastolic blood pressure, mean arterial pressure, pulse rate, oxygen saturation,
and temperature. Monitors are available with or without integrated printers as
well as the following parameters and technologies.
The model of the CARESCAPE VC150 Vital Signs Monitor determines which
parameters are in your monitor. Please refer to applicable sections.
Using the CARESCAPE VC150 Vital Signs Monitor, a caregiver can measure,
display, and record patient vital sign data that is derived from each parameter.
The monitor is also capable of alerting the caregiver to changes in the patient’s
condition or when it is unable to effectively monitor the patient’s condition. All of
the main operations of the CARESCAPE VC150 Vital Signs Monitor are only an
icon-touch away. Please review the factory default settings and, where
applicable, enter settings appropriate for your use. CARESCAPE VC150 Vital Signs
Monitor also provides the Service with various data related to the use of the
monitor such as error logs, NIBP DCS (Data Collection System) data, EMR
transaction logs and productivity metrics.
Product configurations
NIBP, Pulse: SuperSTAT, Auscultatory. SuperSTAT algorithm is currently not
available in Canada.
SpO2: GE TruSignal, Nellcor OxiMax, or Masimo rainbow® SET®
Temperature: Welch Allyn SureTemp Plus®, or Exergen TAT-5000S-USB
Each CARESCAPE VC150 Vital Signs Monitor comes with an accessory pack. The
contents of the pack vary according to the model. Unpack the items carefully. If
an accessory is missing or if an item is in a nonworking condition, contact your
local GE representative immediately.
It is recommended that all the packaging be retained in case the monitor must
be returned for service in the future.
Front view
The CARESCAPE VC150 monitor display is a touch screen where screen items are
selected by pressing the touch panel with a fingertip.
CAUTION
Do not scratch the screen or press the screen with a sharp
object. Do not use excessive force on the screen.
Alarm light
Touch screen
Day/night
light sensor
Product identification
label NIBP hose
connection
Right side
A monitor with the Exergen temperature technology configuration setting
cannot perform Welch Allyn temperature measurements and vice versa. If the
monitor has a Welch Allyn frame installed on the right side of the monitor, then
the monitor uses Welch Allyn temperature technology.
If the monitor does not have the Welch Allyn frame installed, then there are two
alternatives.
1. Exergen technology has been ordered and the monitor is configured for use
of Exergen.
or
PATIENT CAREGIVER
PARAMETERS NOTIFICATIONS TIME/DATE
DC connected.
Battery failure.
WLAN is active.
Connecting to WLAN.
No WLAN available.
Notification area
Alarm and technical messages are displayed in the notification area. The alarm
and indicators are active only in the optional monitoring mode.
Parameters
Menu selections for SpO2 settings are different depending upon the purchased
SpO2 technology, additional licenses and selections in the Monitor Setup > SpO2
screen. The screen automatically adjusts to ambient light conditions by toggling
between day and night brightness settings. Default expiration time for
measurement data on screen is 15 minutes. This can be adjusted in the
configuration mode.
Parameters
Parameters
NOTE
If measurement site is changed, the
selected site will be displayed in
next measurement and current site
information will change to dashes.
NOTE
If measurement site is changed, the
selected site will be displayed in
next measurement and current site
information will change to dashes.
Parameters
Graph area
An optional plethysmographic waveform is displayed in the graph area. SpO2
waveforms are not displayed by default. If the monitor is equipped with Masimo
SpO2 technology, an RRa curve can be displayed with the waveform. Waveform
is selected in MonitorSpO2 > SpO2.
Graph area
Graph area
An optional plethysmographic waveform is displayed in the graph area. SpO2
waveforms are not displayed by default. If the monitor is equipped with Masimo
SpO2 technology, an RRa curve can be displayed with the waveform. Waveform
is selected in MonitorSpO2 > SpO2.
Graph area
Main Menu
The main menu bar contains icons to navigate within the user interface.
Main menu
The home icon is used to close the active monitor setup menu
screen and revert the monitor to the main screen where the
patient’s measurements are displayed.
NOTE
Alarm Setup is available only when the optional monitoring
mode is selected.
Operating modes
The monitor has the following modes of operation:
Clinical mode
Spot-check mode
Monitoring mode (optional)
Configuration mode with three levels of access:
Monitor Setup for settings by any user.
Default Setup (password-protected area) for settings by someone at
the hospital or care unit that has the clinical skills and authority to set
up default settings for the monitor.
Service Mode (password-protected area) for additional configuration,
calibration and maintenance of the monitor.
CAUTION
Service Mode is intended for use by qualified and trained service
personnel only.
Clinical mode
The clinical mode starts right after the
monitor is turned on. When the home
screen appears and two beeps are
heard, vital signs for the patient can be
monitored.
NOTE
If Query for duplicate measurements
feature is enabled by service and you
take multiple measurements before
Measurement Expiration Time is over,
the monitor asks whether to save a
new snapshot or overwrite the
previous measurement. If you select
overwrite, the previous measurement
value is discarded.
Monitor configuration
The monitor has three types of settings:
1. The current settings, which can be adjusted by the regular user during care
session. Settings revert to factory defaults each time a patient is changed.
2. The current default settings, to which the monitor returns upon starting a
new patient. These defaults can be changed only in the configuration mode.
3. Factory defaults for alarm settings on the monitor when it is first received by
the customer. Service can revert alarm settings to factory defaults if
necessary.
All users can access Alarm Setup, Monitor Setup and Patient & Records to
configure or customize user-preferred default settings for measurements in
clinical mode.
When a menu icon is selected, the related menu screen will be displayed, the
menu item will be highlighted and other menu items dimmed. That indicates you
are in that section.
Alarm setup
The alarm limit adjustment allows you to change upper and lower alarm limit
settings used on individual parameter items while monitoring a patient. If a
parameter alarm limit has been set OFF in the configuration mode, color of the
alarm limit on the Alarm Setup screen is displayed as gray and cannot be
adjusted. Also, you can select spot-check or the optional monitoring mode here.
Refer to “Alarm limit setup” in CARESCAPE VC150 Operator’s Manual for
instructions. Depending on the purchased licenses for options, the screen items
will vary.
NOTES
There are no alarm limits and no clinical alarm messages for temperature
measurement.
All changes are temporary and return to the default configuration settings
when the monitor is turned off or a new patient is admitted. To permanently
change the alarm settings refer to “Default Setup” in CARESCAPE VC150
Operator’s Manual.
Alarm setup displays only parameters that are selected in the Monitor
setup or in the Default Setup.
WARNINGS
Monitors located in the same clinical area may contain
different alarm default settings, which can result in a potential
hazard.
Backspace icon.
Dash icon.
Monitor Setup
Monitor Setup consists of four tabs that all users are allowed to configure. These
configurations are only for the current patient, except for the Audible & Visual
settings. To make these changes permanent, these must be changed in the
configuration mode.
Audible & Visual is used to configure alarm and display brightness and
sound settings. Refer to “Audible & Visual” in CARESCAPE VC150 Operator’s
Manual for instructions.
NIBP is used to select cuff position, target inflation pressures and patient
position. Refer to “NIBP” in CARESCAPE VC150 Operator’s Manual for
instructions.
Default Setup
Password protected Default Setup is used for a more advanced configuration of
some general items General, Alarm Defaults, Visual Settings and
Measurement settings. It is intended for someone at the hospital or care unit
that has the clinical skills and authority to configure default settings for the
monitor. Availability of configurable features depends on what parameter-
related options have been ordered. Changes applied in Default Setup remain set
even if the monitor is switched off. Refer to “Default Setup” in CARESCAPE VC150
Operator’s Manual for instructions.
WARNING
Remember to press the home icon to save settings and exit
Default Setup. Failure to do so allows change of settings by
unauthorized users.
Service Mode
Service Mode is used for advanced configuration of the software and calibration
of monitor hardware. If Service Mode is entered, the monitor will prompt to
cancel measurements and discharge the patient because the monitor has to be
restarted after changes in Service Mode.
CAUTION
Service Mode is intended for use by qualified and trained
service personnel only.
WARNING
Remember to press the home icon to save settings and exit the
Service Mode. Failure to do so allows change of settings by
unauthorized users.
NOTE
Refer to “Patient and caregiver data” in the Operator’s Manual KO00065
(English) or a localized VC150 Operator’s Manual for more information.
Printout
The printer is an optional feature of the monitor. If your monitor contains a
printer, each time a printout is started the following information is printed.
1 Contents of the printout can be selected in the Patient & Records screen.
2 Item Name
The monitor is a portable unit that receives power from an internal rechargeable
Lithium-Ion battery.
When the On/Off button is pressed, the motherboard is brought out of a sleep
mode and the power regulators are turned on. The power regulators provide
conditioned power from the battery. The external DC source is used to charge
the battery and to provide power for the monitor. Once the monitor is powered,
a self-test is performed. The self-test automatically tests the main functions of
the monitor. If the self-test fails, the user cannot access the clinical mode. The
monitor can be powered up with external DC source and used without a battery
but the clinical mode cannot be used and an alarm message is displayed in the
notification area.
Under normal operating conditions, the monitor is ready to record the patient
vital signs using three external attachments: the temperature probe, SpO2
sensor, and NIBP cuff. The monitor can be connected to a hospital network with
a wireless connection. USB devices can be connected to USB-A slots on the rear
side of the monitor. A PC can be connected to a USB-B slot on the left side of the
monitor
NOTES
Prior to each use, inspect the power supply cord to ensure proper
connection and condition.
Be sure to unplug the monitor before transport.
SpO2
The SpO2 probe has a built-in sensor. When the SpO2 sensor is attached to the
SpO2 connector and patient, the probe senses both pulse rate and oxygen
saturation. The analog signals are routed to the SpO2 module for GE TruSignal,
Nellcor, or Masimo. The analog signals are analyzed on the SpO2 module. The
results are digitized and sent to the motherboard via opto couplers. The couplers
provide patient isolation as well as serial data interface. The motherboard
temporarily stores the data and routes it to the display and/or printer.
A reset signal to the SpO2 module is also provided so that power-up sequencing
is correct. If the SpO2 circuit quits communicating with the motherboard, the
motherboard will attempt to reset the SpO2 module.
NOTE
SuperSTAT algorithm is currently not available in Canada. NIBP
determination for neonatal population cannot be done. Auscultatory
method cannot be used for neonatal patients.
When the cuff and hose are attached to the monitor and a NIBP determination is
initiated, the pump inflates the cuff. Pressure transducers PT1 and PT2 monitor
the pressure information. The pneumatic manifold has one valve, which is used
to deflate the cuff. Valve control is through the NIBP board. When the cuff
pressure data acquisition for the determination is complete, the processor on
the NIBP board calculates the systolic NIBP, the diastolic NIBP, the mean arterial
pressure (MAP), and the pulse rate. The results are then transferred to the
motherboard, which will display the results on the monitor screen. The user can
also print the results.
The pneumatics are controlled by the NIBP processor. The NIBP processor
monitors pressure information from PT2. If an over-inflation condition occurs,
the OVERPRESSURE signal is routed to the pneumatics to release the air
pressure. The motherboard also generates a technical alarm condition with the
word Overpressure appearing below the diastolic value. A NIBP Overpressure
will appear within 10 seconds in the notification area.
When NIBP measurements are performed: 1) the monitor will use a previous
NIBP value for adaptive target inflation pressure as long as this is displayed on
the screen. 2) the NIBP values are displayed for a maximum of 30 minutes or
until another determination is initiated. When the values on the screen expire or
the patient is discharged, the adaptive target pressure will be automatically
cleared.
For subsequent single, auto, or STAT determinations taken before data of the
previous determination of the same patient has expired, as few as four pressure
steps may be necessary to complete the determination process. When
employing fewer pressure steps, the system uses the stored information from
the previous blood pressure determination to decide the best pressure steps to
take. The algorithm measures the consistency of pulse size to tell if the
oscillations taken at a step are good and if more steps are needed.
The first determination settles at an initial target pressure of 135 mmHg (adult
mode) and 100 mmHg (neonate mode), depending on initial target pressure
preset. To allow for rapid settling of cuff pressure, the monitor will momentarily
inflate to a higher pressure then immediately deflate to the target pressure. After
inflating the cuff, the NIBP parameter begins to deflate. The oscillations versus
cuff pressure are measured to determine the mean pressure and calculate the
systolic and diastolic pressures.
During an NIBP determination, the parameter deflates the cuff one step each
time it detects two pulsations of relatively equal amplitude. The time between
deflation steps depends on the frequency of these matched pulses (pulse rate of
the patient). However, if the monitor is unable to find any pulse within several
seconds, it will deflate to the next step. The process of finding two matched
pulses at each step provides artifact rejection due to patient movement and
greatly enhances the accuracy of the monitor. The figure shows a full
determination sequence for an adult patient.
At each step the microprocessor stores cuff pressure, the matched pulse
amplitude, and the time between successive pulses. The stepped deflation and
matched pulse detection continues until diastolic pressure is determined or total
cuff pressure falls below 8 mmHg. The parameter then deflates the cuff (to zero
detected pressure), analyzes the stored data, and updates the screen.
The operating cycle is composed of four parts: inflation time, deflation time,
evaluation time, and wait time. Wait time, which varies from mode to mode, is
affected by the cycle time (auto mode) or operator intervention (manual mode).
The figure shows the basic operating cycle for an NIBP determination.
Inflation
Time
CUFF PRESSURE
Wait Time
Evaluation Time
Deflation Time
Determination Time
Cycle Time
TIME
Systolic search
NOTE
Arrhythmias will increase the time required by the NIBP parameter to
determine a blood pressure.
If systolic pressure is not found, the SuperSTAT algorithm can search at cuff
pressures higher than the initial target pressure. The algorithm will inflate above
the initial target pressure to obtain more data in the systolic region. The pressure
is limited to the maximum allowed for the selected patient type.
The SuperSTAT algorithm evaluates the data obtained during the determination,
and the prior determination if it is available, to determine if additional data is
needed to complete the determination. It can then selectively pump to a single
cuff pressure to obtain the data it needs and then return to the existing deflation
sequence. This search process makes SuperSTAT more efficient.
Accuracy of the SuperSTAT NIBP measurements was validated against the intra-
arterial method. Do not use the auscultatory method to verify the accuracy of
the SuperSTAT NIBP parameter. The auscultatory method (using the cuff and
stethoscope) determines the systolic and diastolic pressures from sounds that
occur during cuff deflation. Mean arterial pressure cannot be determined by the
auscultation method. The oscillometric method used with all DINAMAP
technologies determines systolic, mean and diastolic pressures from the
oscillation pattern that occurs in the cuff during deflation.
At each step the microprocessor stores cuff pressure, the matched pulse
amplitude, and the time between successive pulses. The stepped deflation and
matched pulse detection continues until diastolic pressure is determined or total
cuff pressure falls below 7 mmHg. The monitor then deflates the cuff (to zero
detected pressure), analyzes the stored data, and updates the screen.
The operating cycle is composed of four parts: inflation time, deflation time,
evaluation time, and wait time. Wait time, which varies from mode to mode, is
affected by the cycle time (auto mode) or operator intervention (manual mode).
The figure shows the basic operating cycle.
Systolic search
If systolic pressure is not found, the NIBP parameter can search at cuff pressures
higher than the initial target pressure. The parameter will inflate the cuff above
the initial target pressure to get more data in the systolic region. The pressure is
limited to the maximum allowed for the selected patient type.
In manual mode, if a previous valid systolic pressure is displayed and less than 2
minutes old, and the new systolic pressure oscillations are compared with the
previous valid determination and the monitor "thinks" that the systolic was not
obtained, the monitor will inflate the cuff to a pressure above the immediately
preceding inflation.
NOTE
Auscultatory method cannot be used for neonatal patients. SuperSTAT
algorithm, using intra-arterial pressure reference, is currently not available
in Canada. NIBP determination for neonatal population cannot be done.
NOTES
For instructions on how to change the NIBP algorithm, refer to “Changing
the NIBP configuration” on page 5-11.
NOTE
For neonatal determinations the SuperSTAT algorithm is always used.
SuperSTAT algorithm is not currently available in Canada. NIBP
determination for neonatal population cannot be done without it.
Temperature
The monitor uses the following technologies to measure patient temperature:
Welch Allyn or Exergen.
When the monitor is powered on, the monitor automatically calibrates the
temperature circuit to account for ambient room temperature.
NOTE
If large changes occur in the ambient temperature, the temperature system
can be recalibrated by cycling power using the On/Off button.
The monitor frequently verifies the signal for temperature and if it is out of
tolerance, the monitor will issue an alarm.
Refer to the operator's manual for complete instructions for use. For more
information on the theory of operation, scientific, educational and technical
information, access the following website: www.TAThermometry.org.
Functional description
The CARESCAPE VC150 monitor operation consists of many blocks that are
described in the diagram below.
Motherboard
NIBP module
SpO2 module (optional)
USB board
Isolation board (optional)
Printer (optional)
Welch Allyn module (optional)
The following paragraphs describe the functional interface relationship.
Motherboard
The motherboard has a main processor and a supplementary processor that
both run specifically assigned subprocesses such as patient parameter interface
(PPI) devices, user interface, power, printer, USB board, real time clock, audio
circuit, and host communication. If the processor system detects a failure, it
displays an error message related to the condition and releases the cuff if the
failure is related to NIBP operation.
The motherboard uses the battery supply to generate regulated DC power that
is routed to the USB and isolation board for isolation. The external DC input is
used to charge the battery via charging circuitry on the motherboard.
The SpO2 module filters and amplifies SpO2 data and then routes it to the
motherboard that processes the data and sends the results to the screen. If a
snapshot is created, the data can also be printed.
NOTE
Motherboard labeling with initial software versions (at the time of delivery
from the manufacturer) is only for the quality purposes of the manufacturer.
The labeling does not state the current software version in use!
NIBP module
The NIBP board consists of a pump, a deflate valve, and a dump valve. The
pneumatics inflate/deflate the cuff during NIBP determinations. During normal
operation the pneumatics are controlled by the NIBP board. If an overpressure
condition occurs, the NIBP processor provides the appropriate control signals to
ensure a safe condition, where the cuff vents to ambient atmosphere pressure.
SpO2 module
The monitor can be configured for use with either a GE TruSignal, Nellcor, or
Masimo SpO2 module. The SpO2 module provides continuous readings of oxygen
saturation and pulse rate. The isolation board provides power and data
communications between the SpO2 module and the main processor. Patient
data from a sensor is routed from the SpO2 module to motherboard.
USB board
USB board contains a DC-input connector, three USB-A ports and a reset-switch.
These can be accessed underneath the USB door at the back of the monitor.
Isolation board
Isolation board provides isolated power for the SpO2 module and Welch Allyn.
The information is routed to the motherboard via an optically coupled
electrically isolated serial connection.
Printer (optional)
The printer receives power from the motherboard and communicates with the
main processor to print a graphic hard copy of the patient’s vital sign values. The
printer communicates with the supplementary processor, which then
communicates with the main processor.
The printer has a built-in sensor to monitor the printer paper presence. When the
printer is out of paper, it sends a signal to the main processor via the
supplementary processor.
The optical switch monitors whether the Welch Allyn temperature probe is
inserted in the probe well or not. The motherboard powers the switch via the
isolation board.
Connectivity
With the purchase of option-specific licenses and wireless or Hostcomm
connection (Hostcomm is currently reserved for manufacturing and production
use only, it is not available for clinical use.) to hospital network, the monitor can
send patient data and information entered by a caregiver to the hospital EMR.
Only EMR queries and sending out patient data is possible. The monitor cannot
edit anything in the EMR. The monitor can interact with an LDAP server to
validate the caregiver identity.
WARNING
Wireless devices cannot be fully protected against malicious
attacks from an intruder who has penetrated the hospital
wireless/network infrastructure. Consider carefully the benefits
vs. risks of using the wireless function.
CARESCAPE
VC150 ADT
WLAN EMR
NTP
Non-GE
Router Gateway
LDAP
NOTE
Connection between Router and HIS (EMR, ADT & LDAP) does not always
require a separate gateway. The WLAN and router belong to the existing
hospital infrastructure.
Connections
CAUTION
Auxiliary equipment connected to the CARESCAPE VC150 vital
signs monitor will result in the formation of an electromedical
system and thus must comply with the requirements of IEC
60601-1. All USB ports except USB-B are non-isolated and
should be connected to equipment conforming to IEC-60601-1
or configured to comply with IEC 60601-1 only. Do not connect
unapproved devices to the monitor.
USB-A USB-B
USB
Power cord
Remote alarm
connection
When a USB-A cable is connected, the monitor tries to find an external USB
device. When a USB-B cable is connected, the monitor can be connected to a PC.
NOTES
The primary method of asserting alarms is always locally by the monitor
itself. The remote alarm should only be considered a supplemental method
of asserting alarms.
Battery
The CARESCAPE VC150 uses a 6-pack Lithium-ion battery (3 cells in series and 2
cells in parallel). The battery pack includes a protection circuit for overcurrent,
overvoltage and low voltage as well as an NTC resistor for temperature
information to the charging circuit on the motherboard. In order to guarantee
battery availability, there are two batteries available for the CARESCAPE VC150
that have the same capacity, but different nominal voltage. The voltage is
marked on the battery and it must be selected in software during the initial
installation. Check CARESCAPE VC150 Supplies and accessories for battery FRU
order codes.
Introduction
This section describes how to unpack and install the CARESCAPE VC150 monitor.
Workflow
A recommended workflow is the following:
Installation checklist
Unpack the monitor. Refer to “Unpacking the unit” on page 3-2.
Attach the battery. Refer to “First charge of the battery” on page 3-3.
Start up the monitor. Refer to “Turning the monitor on” on page 3-6.
Calibrate touchscreen. Refer to “Touch screen recalibration” on page 3-
7.
Verify the touchscreen calibration. Refer to “Verification of the touch
screen recalibration” on page 3-8.
Proceed to “Monitor configuration checklist” on page 4-2.
While unpacking all items from their individual packages, identify the contents of
all shipping materials and perform a visual check for any damage incurred
during transportation. Refer to “Visual inspection checklist” on page 9-5 to check
what the shipment should contain at a minimum. The contents may vary
depending on the purchased technology options. If an item has sustained
damage, contact customer service immediately.
4. Remove the monitor from the box and the plastic bag.
5. Place the monitor on a steady table, face down. Ensure that the surface will
not scratch the touch screen. Use a soft, lint-free cloth if necessary.
6. If the monitor has been stored in cold ambient temperature, moisture may
condense inside the monitor. Let the monitor and the battery warm up to
room temperature before connecting the battery and turning on the
monitor.
7. Fill in the shipment checklist below. Contents of a CARESCAPE VC150
shipment depends on the country of destination. A typical shipment must
contain at least the following:
Customer Date
Manual kit
NOTES
Before taking the monitor into clinical use for the first time, the battery must
be charged for at least 8 hours to optimize battery performance. This initial
8-hour battery charge is very important for correct performance of the
monitor.
The battery needs at least three charging cycles to determine its charging
capacity.
The battery pack should be charged at room temperature: 5°C to 40°C (41°F
to 104°F).
You can charge or top off the battery pack at any time. You should not wait
until the battery is fully discharged.
Procedure
1. Remove the bottom plate and pull the battery out from the battery
compartment as instructed in “Battery replacement” on page 7-19.
2. Check what the battery voltage is and write it on the line below. Use this
information to select for the corresponding Battery type in “Calibration” on
page 4-17.
NOTE
The monitor is delivered either with 11.1V or 10.8V battery. Also if the battery
is ordered separately, the voltage may be either of these two.
CAUTION
To ensure the optimal battery performance, select the Battery
type (voltage) after every new battery installation.
3. Plug in the battery cable, and place the battery in the monitor as instructed
in “Battery replacement” on page 7-19.
4. Inspect the power supply cord to ensure proper connection and condition.
5. Plug the AC cord into the AC mains input on the external power supply.
6. Plug the supply DC output underneath the USB door.
NOTE
Use only the original power adapter, or a power adapter that has been
specified for this monitor. Refer to the CARESCAPE VC150 supplies and
accessories document (KO00125).
Charging the battery pack for 8 hours before first use, or after prolonged
periods of storage, calibrates the battery charging circuitry with the charge
status of the battery.
If the monitor is idle for extended periods, it should be fully charged at least
once a month to ensure optimum performance.
The battery pack should be charged before use, because a battery loses
charge when left in storage. Charging is done automatically by the monitor
when the external DC power is connected.
To prolong the life of the battery, keep the monitor connected to a DC power
supply whenever possible. Do not allow the battery to become completely
discharged.
A fully charged battery will power the monitor for up to 8 hours in a heavy-
use scenario and up to 11 hours in a light-use scenario “Monitor battery
specifications” on page A-18 for details).
To ensure full charge cycles, replace only with the specified battery.
When the monitor is operating on battery power and no battery alarm is active,
a green battery icon indicates the charge level as a percentage.
WARNING
If the monitor is to be stored for longer than one month, first
charge the battery and then remove it and store it separately
from the monitor.
Monitor on/off
Turning the monitor on
NOTE
For safety reasons, the monitor is designed so that it can be powered
without a battery while its power supply is connected to mains. However,
clinical mode is not possible without a functional internal Lithium-ion
battery specified for CARESCAPE VC150.
WARNING
Inspect the device for damage before use.
1. Push the power On/Off button on the left side of the monitor to turn it on.
During power-up, the monitor displays the software version on a white
power-up screen and grayish blue borders. The monitor will also flash the
alarm light at the top. After the monitor has completed the start-up
sequence, there will be two short tones and the opening screen changes to
the home screen.
NOTES
If the monitor fails to sound the start-up tones, do not use the monitor. This
indicates problems with the audible alarm circuit. Potential system alarms
cannot be heard. Refer to Audio not available in “Host messages” on
page 6-13.
If the power-up screen does not change to the home screen within 2
minutes, refer to “Monitor does not turn on” on page 6-32.
The monitor does not start up if the battery charge level is too low.
2. If you notice that selections do not accurately track the position of your
finger, it is possible to re-calibrate the touch screen on this device during
power-up. Refer to “Touch screen recalibration” on page 3-7.
NOTE
While starting up, the monitor will display the current software version.
3. You can recalibrate the touch screen during first 10 minutes after you have
turned the monitor ON. Press power button 4 times within 5 seconds. The
screen will turn black and then return to clinical view with countdown timer.
Calibration is started automatically if the user does not cancel it while the
countdown is progressing. When the recalibration is complete, the monitor
will restart.
4. You can start the recalibration procedure when a blinking rectangle appears
in the top left corner (A).
A D NOTE
If necessary, press the On/Off button to exit the recalibration procedure.
5. Press the rectangle to select it. Hold your finger down for a moment. If the
B C software registered the selection, the rectangle moves to the next location.
6. Select the rectangle at each location (B, C, D and E).
NOTE
Do not slide your finger on the screen. Selections are made by pressing the
screen with your fingertip and then lifting the finger away from the screen.
1. Insert an Allen key into the 3 mm hole under the USB door, right above the
power cord plug.
Reset switch
2. Search and push the switch until the monitor shuts down.
NOTE
The following actions will take place in each of the three cases mentioned
above:
NOTES
When the Reset switch is pressed, the monitor shuts down in an
uncontrollable manner. Alarm light and power LEDs might remain lit. After
pressing the Reset, press the Power button to start up the monitor.
If the standby lasts less than 5 seconds, the patient is discharged, but the
caregiver remains logged on. If the standby lasts more than 5 seconds, also
the caregiver is discharged.
Connections
The monitor provides connections for NIBP, SpO2, Welch Allyn (WA) temperature,
a Medical USB port (not currently available), a USB-B port for connection to PC,
three regular USB connections and power. The USB-B port signals are isolated.
The Exergen is connected to USB-A port.
NIBP SpO2
USB-A USB-B
USB
Power cord
Remote alarm
connection (nurse call)
Remote alarm
CAUTION
Auxiliary equipment connected to the CARESCAPE VC150 Vital
Signs Monitor will result in the formation of an electromedical
system and thus, must comply with the requirements of IEC
60601-1 and be configured to comply with IEC 60601-1 only.
If external alarm control is required, the remote alarm cable should always be
used. The primary method of asserting alarms is always locally by the monitor
itself. The remote alarm should only be considered a supplemental method of
asserting alarms.
When a high or medium priority alarm condition is displayed on the monitor, the
remote alarm signal becomes active immediately.
Strain relief
Ask the customer whether strain relief is needed for USB-A. If yes, then wrap a
cable tie on the cable, pull it tightly on the cable and cut off the remaining part of
the cable tie. When the hatch is secured with the thumb screw, the hatch will
then lock the cable tie.
NOTES
Do not use real patient information when debugging/testing connectivity
settings!
VC150 uses IHE PCD-01 profile (HL7 version 2.6) and IHE PDQ profile (HL7
version 2.5). To use other HL7 versions, CSG is required to make mapping
into target system version.
NOTE
Look for the following log entries:
3. Return to checklist if you are configuring the monitor with the checklist.
Service mode
Various general, region-specific and advanced software items (System info,
Maintenance, Regional Settings, Connectivity, Print) can be configured in
Service Mode.
WARNING
Remember to press the home icon to save settings and exit the
Service Setup. Failure to do so allows change of settings by
unauthorized users.
Enable licenses
Licenses are purchased for an individual monitor with unique serial number. The
monitor uses the license info to enable a license-based feature or features. The
licenses are enabled by entering the license key manually or, alternatively, by
importing a license file to the monitor.
NOTE
Record the license codes or license files for possible later use. For example,
in the event of motherboard replacement, the licenses need to be
reinstalled.
SuperSTAT NIBP
determination (STAT)
Manual entry
1. Select a text field in Monitor Setup > Advanced > Login Service Mode >
System Info > Licenses.
3. Select Confirm to confirm the license key and to close the keyboard screen.
5. Enter other license keys for other features that are to be activated.
6. Select the home icon to exit the service mode and restart the monitor. If
there are other configuration items to work on, you can complete those
before exiting the service mode.
Verify licenses
1. Enter Monitor Setup > Service mode > System Info.
2. Scroll down to Licenses and verify that licensed features have been enabled
(marked with OK).
Disabling a license
If a feature must be disabled, unselect the feature’s checkbox in the Monitor
Setup > Advanced > Login Service Mode > System Info > Licenses. The OK text
will change to red INVALID text to indicate that the feature is disabled.
System Info
Some hospital-specific items can be set in Monitor Information subsection of
the Monitor Setup > Advanced > Login Service Mode > System Info.
These settings affect all applicable monitor screens. Purchased licenses and the
technologies used affect the availability of screen items.
Description
System info provides few configuration items and information on various parts
of the monitor.
Monitor information
1. Enter Monitor Setup > Advanced > Login Service Mode > System Info.
2. Scroll down to Monitor Information and configure items like Care Area,
Hospital Policy for Patient Name and Hostcomm Unit Address, if required.
Maintenance
1. Enter Monitor Setup > Advanced > Login Service Mode > Maintenance to
configure various advanced software settings.
2. Scroll down to applicable subsection(s) and configure required items.
3. When you have completed all configuration items in Maintenance, return to
“Monitor configuration checklist” on page 4-2 to continue the configuration
or select the home icon to exit the Service Mode.
Software update
1. Enter Monitor Setup > Advanced > Login Service Mode > Maintenance.
2. Scroll down to Software Update and start the software update.
Local history
1. Enter Monitor Setup > Advanced > Login Service Mode > Maintenance.
2. Scroll down to Local History and remove the local history.
1. Enter Monitor Setup > Advanced > Login Service Mode > Maintenance.
2. Scroll down to Day/Night mode and configure items like Day Mode Low
Limit and Night Mode Upper Limit.
Day Mode Low Limit When in the day mode, the VC150 will
switch to the night mode if the sensor
reading drops below Day Mode Low
Day/Night mode
Limit.
Night Mode Upper Limit When in the night mode, the VC150
will switch to the day mode if the
sensor reading exceeds Night Mode
Upper Limit.
Workflow
The service user can enable or disable login requirements, prevent snapshots for
unidentified patient and allow editing of snapshots.
1. Enter Monitor Setup > Advanced > Login Service Mode > Maintenance.
2. Scroll down to Workflow and configure items like Day Mode Low Limit and
Night Mode Upper Limit.
NOTE
Before enabling this setting,
check whether hospital policy
allows editing measured values.
Measurement settings
1. Enter Monitor Setup > Advanced > Login Service Mode > Maintenance.
2. Scroll down to Measurement settings and configure all applicable
measurement related items.
NOTE
If the unit of measurement is changed and the old patient history contains
measurement data with other unit of measurement, new measurement
data will not be stored in local history. Delete local history. Refer to “Clearing
local history” on page 4-61.
Maintenance action
Description
icons
page 4-15.
1. Clear the local history as instructed in “Clearing local history” on page 4-61.
2. Change the unit of measurement in the Monitor Setup > Advanced > Login
Service Mode/Maintenance/Measurement settings.
3. Select the home icon to exit the Service Mode.
4. Check that correct unit of measurement is displayed on the right end of the
Patient menu.
5. Return to “Monitor configuration checklist” on page 4-2 to continue the
configuration.
Calibration
1. Enter Monitor Setup > Advanced > Login Service Mode > Maintenance.
2. Scroll down to Calibration and select calibration related items like NIBP
Calibration, Nurse Call Test, Reset battery gauge or Battery Type..
NOTE
It is important to select correct
battery voltage. If a wrong
voltage is selected, this affects
Calibration
NOTE
If Auto_Shutdown is set to None, the monitor will remain in
continuous standby. After one week of standby, the monitor will
reboot itself after 15 minutes of standby and return to standby after
the reboot is complete. If the user tries to activate the monitor
during this time, it might be in middle of restart sequence and not
exit the standby mode correct.
Logging
1. Enter Monitor Setup > Advanced > Login Service Mode > Maintenance.
2. Scroll down to Logging and configure logging related items like NIBP DCS
Data or NIBP Collection Mode.
Delete NIBP DCS data Select Delete to delete all NIBP DCS
data.
DCS data
When DCS Mode is On, there are different ways to access the DCS data: Store,
Passthrough and Cache. If DCS Mode is selected as Off, nothing happens.
Store
Store is the most common option where DCS begins upon NIBP measurement
and will collect data for 400 seconds or until the user switches off the DCS Mode.
The data will be packed into a single file named as title_date_time.dcs.file format
i.e. gehc_nibp_17OCT2013_120842.dcs.gz. The file is exported along with other
log files as instructed in “Exporting error logs” on page 6-3. DCS shall continue to
operate until the collection time expires. A new session is started next time the
user initiates a NIBP measurement.
Passthrough
When Passthrough is selected and a PC is connected to the monitor USB-A port
with a VC150 DCS cable (pn 2081723-001), an external collection software
controls the DCS. Nothing is saved on the monitor. If the USB line breaks up, the
data will be lost.
Cache
Cache is similar to Passthrough, but Cache also stores collected data on the
monitor. If USB connection to the external control breaks up, the monitor
software falls to Store and collects the data to the end of the collection time. The
file is exported along with other log files as instructed in “Exporting error logs” on
page 6-3.
NOTES
When DCS is operating in Cache or Store mode ensure that USB export is
not active. If it is active, the monitor cannot write DCS files.
When using Passthrough or Cache mode, the USB cable must not be
attached before the monitor is started.
EMR log
1. Enter Monitor Setup > Advanced > Login Service Mode > Maintenance.
2. Scroll down to EMR log and configure items like EMR Transaction Log Filter,
EMR Transaction Log Start Time, EMR Transaction End Time or EMR
Transaction Log.
Time
Passwords
1. Enter Monitor Setup > Advanced > Login Service Mode > Maintenance.
2. Scroll down to Passwords and change Default Setup password and Service
Mode password.
WARNING
Do not set the Default Setup password too easy to guess or
empty if there is a risk that unauthorized users could change
clinically important settings such as alarm limits, alarm delays,
or target inflation pressures.
Passwords
Service Mode password Factory default for password is:
B1xm3D1cAL
Barcode
1. Enter Monitor Setup > Advanced > Login Service Mode > Maintenance.
2. Scroll down to Barcode and configure Barcode shortcut, if necessary.
Remove Patient Prefix & If enabled, the prefix and postfix are
Postfix stripped away from the read barcode
before it is populated to the
Identification text field.
Remove Patient Prefix & If enabled, the prefix and postfix are
Barcode
Postfix stripped away from the read barcode
before it is populated to the
Identification text field.
Regional Settings
1. Enter Monitor Setup > Advanced > Login Service Mode > Regional Settings
to configure region-specific settings.
2. Scroll down to required subsection and configure items like Date Format,
Time Format, NTP Server Address, Time Zone, Area, Time Zone, Location,
Powerline Frequency (Masimo Only), Language and Keyboard.
NOTE
If monitor time zone is configured correctly, the system will automatically
perform daylight saving time adjustments. NTP connection is not required
for the DST changes as the SW handles this locally according the DST rules
for the given time zone.
Central European Time, 1 hour east from Greenwich, Central European Summer
Time, daylight saving starts on the last Sunday in March at 2 AM and ends on the
last Sunday in October at 3 AM.
NOTE
The day must be between 0 and 6. Day 0 is Sunday.
For more information and recent changes to the libc TZ-variable, refer to the
TZ-Variable documentation found in:
http://www.gnu.org/software/libc/manual/html_node/TZ-Variable.html
Connectivity
Various connectivity-specific settings can be set in subsections of the Monitor
Setup > Advanced > Login Service Mode > Connectivity screen.
WARNING
Wireless devices cannot be fully protected against malicious
attacks from an intruder who has penetrated the hospital
wireless/network infrastructure. Consider carefully the benefits
vs. risks of using the wireless function.
NOTE
Obtain required connectivity information from the hospital IT. Refer to
subsection tables.
Connectivity Licenses
If connectivity licenses were purchased, GE service will provide license files for
import via USB. Following connectivity licenses are available.
Licenses
CERNER
CSG PDQ PCD-01 VITALSLINK LDAP
(iBUS protocol)
Connectivity diagram
The diagram below provides an overview on VC150 connectivity. Configurable
items are covered in subsequent sections.
NTP
Date & time
Patient query
ADT Patient demographics
Caregiver &
patient Patient and
Caregiver query HL7 message
identification measurement data EMR
LDAP Caregiver validation
EMR encoding
Measurement data
NIBP
SQL database SpO2
Temp
VC150 monitor
Required tools
ap_scan=1 AP scanning/selection
Each network (usually AP's sharing the same SSID) is configured as a separate
block in configuration file. The network blocks are in preferred order.
Example:
network={
ssid="simple"
password Password string for EAP. This field can include either the
plaintext password (using ASCII or hex string) or a
NtPasswordHash (16-byte MD4 hash of password) in
hash:<32 hex digits> format.
ca_path= /opt/VSM/Storage/Certificates/ Directory path for CA certificate files (PEM). This path
may contain multiple CA certificates in OpenSSL format.
ca_cert2 File path to CA certificate file. This file can have one or
more trusted CA certificates.
EAP-FAST variables
pac_file File path for the PAC entries. wpa_supplicant will need
to be able to create this file and write updates to it
when PAC is being provisioned or refreshed. These are
located in /opt/VSM/Storage/Certificates/.
ctrl_interface=DIR=/var/run/wpa_supplicant
network={
ssid="example"
key_mgmt=WPA-EAP
eap=PEAP
identity="user@example.com"
password="foobar"
ca_cert="/opt/VSM/Storage/Certificates/ca.pem"
phase1="peaplabel=1"
phase2="auth=MSCHAPV2"
priority=10
Certificates
1. Enter Monitor Setup > Advanced > Login Service Mode > Connectivity.
2. Scroll down to Certificates subsection and select it.
3. Select USB Import Certificates.
4. Connect USB-B cable between computer and VC150, when instructed. Wait
until computer recognizes VC150 as a mass storage device.
5. Import PEM or DER formatted certificates to VC150.
6. Disconnect from Service PC. Refer to “Disconnecting from the service PC” on
page 9-7.
NOTE
Trusted CA root certificates are
imported from here.
Private keys
1. Enter Monitor Setup > Advanced > Login Service Mode > Connectivity.
2. Scroll down to Private Keys subsection and select it.
3. Select USB Import Private Keys.
4. Connect USB-B cable between computer and VC150, when instructed. Wait
until computer recognizes VC150 as a mass storage device.
5. Import PEM or DER formatted private key file(s) to VC150.
6. Disconnect from Service PC. Refer to “Disconnecting from the service PC” on
page 9-7.
7. Restart the monitor. Refer to “Monitor on/off” on page 3-6.
Related UI section
Private Keys/USB Import Private Private key import. Import PEM or DER formatted private
Keys keys through USB.
SSID input is limited to 0-32 octets. If SSID name cannot be input using
monitor keyboard (SSID containing non UTF8 encoded characters), wpa
custom profile must be used.
1. Enter SSID Profile 1/2/3/4 subsection of the Monitor Setup > Advanced tab
> Login Service Mode > Connectivity.
2. Enter SSID.
3. Select proper Encryption and insert Passphrase.
4. Select proper Frequency Band.
5. Enter RTS and Fragmentation Threshold.
6. Set WMM QoS, Enabled or Disabled.
7. Set DHCP, Enabled or Disabled.
8. Set any additional parameters if necessary.
NOTE
If you want to manually configure IP Address, Netmask, Gateway Address
and DNS Address, set DHCP to Disabled.
Encryption Selection for encryption method. Select None, WEP, WPA-PSK, WPA-
EAP or WPA Custom.
WEP Key (WEP) WEP encryption key. Obtain input from hospital IT and
enter it here or make a selection.
WEP Key Format (WEP) Selection of Hex or ASCII format for
WEP key.
WPA Supplicant Custom WPA supplicant custom configuration Select WPA supplicant custom
Configuration File (WPA file, if WPA Custom is selected. configuration file.
Custom)
WPA2 User Certificate (WPA- User certificate for WPA2, if WPA-EAP Select WPA2 user certificate.
EAP) is selected.
WPA2 CA Certificate (WPA- CA certificate for WPA2, if WPA-EAP is Select WPA2 CA certificate.
EAP) selected.
WPA2 Private Key (WPA-EAP) Private key for WPA2, if WPA-EAP is Select WPA2 private key.
selected.
WPA2 Private Key Password Password for WPA2 private key. Obtain input from hospital IT and
(WPA-EAP) enter it here or make a selection.
NOTE
For safety reasons the password
must have at least 8 characters.
If a hospital uses a shorter
password, a special method
must be used to enter the
password.
Gateway Address
DNS Address
NOTES
If selection for regulatory domain cannot be seen, the selection is set at the
manufacturer or locked up with a license.
Wifi diagnostics
Wifi Diagnostics is a tool that helps service user to verify that wireless
connection is established properly. Wifi diagnostics can be accessed at Monitor
Setup > Advanced tab > Login Service Mode > Connectivity > Wifi Diagnostics.
check with the counters that packages are sent and received.
check the status of the current network connection with ifconfig.
use Wireless info to check the connection quality between the VC150
and the access point
enable Extra logging by checking the check box underneath the
Wireless Info
NOTE
Extra logging information can be viewed only when the log is exported to
Service PC. It cannot be viewed from Maintenance > Error log!
The numerical value of the Transmit power (TX power) may be shown in
Wireless info even if Wifi is not connected.
Host to Ping Test packet ping to verify connection Obtain input from hospital IT and
perform tests or checks.
Ping Test to determine host connection
Extra Logging Selection for debugging the Wifi When Extra logging checkbox is
selected, the monitor will
immediately start logging all EMR
messages. When Apply is selected,
monitor will begin logging the WPA
supplicant output also. Logging will
end when monitor is restarted.
WARNING
When Extra logging is on, all EMR traffic will be logged. Do
not send/receive real patient information when Extra
logging is on, only dummy test patient information should
be used.
Definition
PDQ = Patient Demographics Query = Sending queries and Retrieving patient
demographics from hospital server.
Procedure
1. Enter PDQ Query Configuration subsection of the Monitor Setup >
Advanced tab > Login Service Mode > Connectivity.
2. Protocols available:
PDQ, used with IHE PDQ compatible database servers.
iBus, used with the Cerner server
Picklist, when Carescape Gateway is used
Server address M M M
Outbound port
Inbound port O
Use SSL M M
User name O
Password
Receiving Application O
Receiving Facility
Receiving Authority
Sending Application
Sending Facility
patCtx M
patIssuer
IdentCtx O
orgId
barcodeOrgId
base M
patientPath
barcodePath
Require Authenticated
caregiver
Protocol Selection of the protocol to use. This Select None, PDQ, iBus or Picklist.
affects the available configurable
items. If an item is available for a
certain protocol, it is listed, e.g., Use
SSL (PQD, iBus).
Server Address Input for server address. Obtain input from hospital IT and
enter it here.
Outbound Port Input for outbound port.
NOTE
This is really insecure and should not be setup unless dual port
PDQ Query Configuration
configuration is needed.
Use SSL (PDQ, iBus) Use of SSL. Select whether SSL is enabled or
disabled.
Username (iBus) Username used to establish a Obtain input from hospital IT and
connection. enter it here.
Query List Size Limit for retrieved query results. Set required limit.
Receiving Application (PDQ) Name of the application that receives Obtain HL7 standard input or
the packet data. contexts from hospital IT and enter it
here.
Receiving Facility (PDQ) Name of the facility that receives the
packet data.
patCtx (iBus) Patient context parameter if iBus Obtain iBus installation input from
protocol installed. hospital IT and enter it here.
Query Barcode Input Selection for automatic patient Select Enable if you want to enable
search. automatic search.
Select Disable if automatic search is
not required.
NOTE
If enabled, barcode shortcut must not be set as Off in Service mode >
Maintenance.
WARNING
If Require Authenticated Caregiver option is not selected,
Patient Demographic Query can be made by anyone who
has access to the monitor. Patient confidentiality may be
compromised.
Allow Manual Patient Search Manual input for patient search data. Select Enable if you want to enable
manual patient search.
Select Disable if manual patient
search is not required. If disabled,
scanning the barcode is the only
method for searching a patient.
Definition
EMR= Electronic Medical Record = Sending measurement data to hospital server.
Procedure
1. Enter EMR Outbound configuration subsection of the Monitor Setup >
Advanced tab > Login Service Mode > Connectivity.
2. VC150 can only send data to the EMR server. Once data is sent to EMR, it
cannot be edited anymore. Protocols available:
PCD-01 works with IHE PCD-01 compatible database servers
Can also be configured to work with other HL7 supporting servers
PCD-01 iBus
Protocol M=Mandatory selection,
O=Optional selection
Server address M M
Outbound port M M
Inbound port O
SSL M M
Retry limits M M
Patient records M M
Caregiver identity M M
PCD-01 iBus
Protocol M=Mandatory selection,
O=Optional selection
Send priority M M
Receiving Application O
Receiving Facility O
Sending Application O
Sending Facility M
Assigning authority O
Username (iBus) O
Password (iBus) O
base (iBus) M
chartdocPath (iBus) M
useDeviceTimeStamp (iBus) M
EncounterDescriptor (iBus) M
CLASS (iBus) O
Server Address Input for server address. Obtain input from hospital IT and
enter it here.
Outbound Port Input for outbound port.
Retry Limit Number of times the monitor tries to Set the number of retry attempts.
send patient information. For example,
EMR Outbound Configuration
Patient Records Selection of whether patient can be Unidentified means that identifier
unidentified or must be identified or field can be empty. Identified means
verified before his/her data can be sent that ID field should have some
to the EMR. identifier. Verified means that it has
been checked with PDQ of LDAP.
Caregiver Identity Selection of whether the caregiver can
be unidentified or must be identified or
verified (requires LDAP license).
Caregiver settings
1. Enter Caregiver subsection of the Monitor Setup > Advanced tab > Login
Service Mode > Connectivity.
VC150 can use two only send data to the EMR server. Once data is sent to EMR, it
cannot be edited anymore. Protocols available:
LDAP iBus
Protocol M=Mandatory selection,
O=Optional selection
Server address M M
Outbound port M M
Use SSL M M
Authentication Method M
Search Password O
Search Depth M
Search Filter M
LDAP Timeout M
Username O
Password O
Require password M
orgId M
base M
personnelPath M
loginBasicPath O
Server Address Input for server address. Obtain input from hospital IT and
enter it here.
Outbound Port Input for outbound port.
Procedure
1. Enter Hospital EMR encoding subsection of the Monitor Setup > Advanced
tab > Login Service Mode > Connectivity.
2. Configure the hospital EMR encoding according to the hospital policy. Refer
to “EMR encoding examples/PCD-01” on page 4-47 or “EMR encoding
examples/iBus” on page 4-49, if necessary.
3. Return to “Connectivity configuration checklist” on page 4-3 to continue the
configuration.
Description
VC150 sends data to an EMR server in a certain format. Some EMR servers will
accept measurement data only when it is sent in a specific format. With
Hospital EMR Encoding, the service user can configure formatting of the
parameter, so that the EMR server will accept the measurement data in transfer.
Once the service user changes the format of a parameter, all other parameters
will be ignored during measurement data transfer to the EMR server. The service
user must remember to change the encoding for every parameter that must be
transferred to the EMR.
The default EMR encoding uses MDC codes for the parameters and it does not
include all of the parameters/values that the device can measure. Refer to EMR
encoding examples on service manual “EMR encoding examples/PCD-01” on
page 4-47.
NOTE
The IT department of the hospital must provide the required EMR encoding
settings.
SpO2 and many other items Parameter mapping from Cerner or Check with hospital IT for names and
PCD-01. formatting. You can insert
identifications in HL7 format. Refer to
“EMR encoding examples/PCD-01”
on page 4-47.
NOTE
Once the service user changes a
Hospital EMR Encoding
NOTE
There are no default MDC-values
for Masimo rainbow -
parameters. If Masimo rainbow
parameters need to be sent,
MDC encodings must be added
manually.
PCD-01
SpO2 150456^MDC_PULS_OXIM_SAT_O2^MDC
Perfusion_Index 150488^MDC_BLD_PERF_INDEX^MDC
Pulse_Rate 149530^MDC_PULS_OXIM_PULS_RATE^MDC
NIBP_Pulse_Rate 149546^MDC_PULS_RATE_NON_INV^MDC
Systolic 150021^MDC_PRESS_BLD_NONINV_SYS^MDC
Mean 150023^MDC_PRESS_BLD_NONINV_MEAN^MDC
Diastolic 150022^MDC_PRESS_BLD_NONINV_DIA^MDC
Temperature 150344^MDC_TEMP^MDC
Temperature_Monitored 150344^MDC_TEMP^MDC
Respiration_Rate 151562^MDC_RESP_RATE^MDC
mmHg mm[Hg]^mm[Hg]^UCUM
percent %^%^UCUM
beatsPerMin {pulse}/min^{pulse}/min^UCUM
breathsPerMin {breath}/min^{breath}/min^UCUM
C Cel^Cel^UCUM
F [degF]^[degF]^UCUM
kPa kPa^kPa^UCUM
unitless {unitless}^{unitless}^UCUM
None 0^MDC_UNKNOWN^MDC
EarLobe 132592^MDC_HEAD_EAR^MDC
Nose 0^MDC_UNKNOWN^MDC
Finger 132820^MDC_UPEXT_FINGER^MDC
Toe 0^MDC_UNKNOWN^MDC
Other 0^MDC_UNKNOWN^MDC
Forehead 132592^MDC_HEAD_EAR^MDC
Axillary 132812^MDC_UPEXT_AXILLA^MDC
Rectal 132740^MDC_TRUNK_BUTTOCK^MDC
Oral 132612^MDC_HEAD_MOUTH^MDC
LeftArm 132853^MDC_UPEXT_ARM_UPPER_L^MDC
RightArm 132854^MDC_UPEXT_ARM_UPPER_R^MDC
LeftLeg 132677^MDC_LOEXT_LEG_L^MDC
RightLeg 132678^MDC_LOEXT_LEG_R^MDC
iBus
SpO2 SPO2
Pulse_Rate PULSE_RATE
NIBP_Pulse_Rate PULSE_RATE
Systolic SYSTOLIC
Mean MEAN
Diastolic DIASTOLIC
Temperature TEMPERATURE
Temperature_Monitored TEMPERATURE
Respiration_Rate RESPIRATORY_RATE
mmHg MMHG
percent PERCENT
beatsPerMin BEATSPERMIN
breathsPerMin BREATHSPERMIN
C C
F F
kPa KPA
unitless UNITLESS
Cuff_Type CUFF_TYPE
Cuff_TypeNM
M_EWS M_EWS
M_RRNM
M_RR_Unit
M_RR_UnitNM
M_Pain_score
M_Pain_scoreNM
M_Pain_score_Unit
M_Pain_score_UnitNM
M_Inputs
M_InputsNM
M_Inputs_Unit
M_Inputs_UnitNM
M_Outputs
M_OutputsNM
M_Outputs_Unit
M_Outputs_UnitNM
M_Level_of_consciousness
M_Level_of_consciousnessNM
M_Level_of_consciousness_Unit
M_Level_of_consciousness_UnitNM
M_Free_Text_1NM
M_Free_Text_2
M_Free_Text_2NM
M_Free_Text_3
M_Free_Text_3NM
M_Custom_1NM
M_Custom_1_Unit
M_Custom_1_UnitNM
M_Custom_2
M_Custom_2NM
M_Custom_2_Unit
M_Custom_2_UnitNM
M_Custom_3
M_Custom_3NM
M_Custom_3_Unit
M_Custom_3_UnitNM
M_Custom_4
M_Custom_4NM
M_Custom_4_Unit
M_Custom_4_UnitNM
M_Custom_5
M_Custom_5NM
M_Custom_5_Unit
M_Custom_5_UnitNM
M_Custom_11
M_Custom_11NM
M_Custom_11_Unit
M_Custom_11_UnitNM
M_Custom_12
M_Custom_12NM
M_Custom_12_Unit
M_Custom_12_UnitNM
M_Custom_13
M_Custom_13NM
M_Custom_13_Unit
M_Custom_13_UnitNM
M_Custom_14NM
M_Custom_14_Unit
M_Custom_14_UnitNM
M_Custom_15
M_Custom_15NM
M_Custom_15_Unit
M_Custom_15_UnitNM
M_Custom_16
M_Custom_16NM
M_Custom_16_Unit
M_Custom_16_UnitNM
M_Custom_17
M_Custom_17NM
M_Custom_17_Unit
M_Custom_17_UnitNM
M_Custom_18
M_Custom_18NM
M_Custom_18_Unit
M_Custom_18_UnitNM
M_Custom_19
M_Custom_19NM
M_Custom_19_Unit
M_Custom_19_UnitNM
M_Custom_20
M_Custom_20NM
M_Custom_20_Unit
M_Custom_20_UnitNM
Web server
1. Enter Monitor Setup > Advanced > Login Service Mode > Connectivity.
2. Scroll down to Web Server subsection.
3. Configure Enabled, Port and Secure to enable the remote service.
Port Web server port number. Obtain input from hospital IT and
enter it here.
Device certificate
1. If you need to secure the remote service connection, enter Monitor Setup >
Advanced > Login Service Mode > Connectivity.
2. Scroll down to Device Certificate subsection and select it.
3. Configure Certificate, Private Key, Private Key Password and SSL Mode.
Certificate Selection of the certificate that server uses to Obtain input from
prove its own identity, so that the client can verify hospital IT and enter it
that the server is what it claims to be. here.
LDAP filter configuration depends on the local HIS and may vary to a great
extent and sophistication. In this section various configuration examples are
listed to provide an overview on LDAP filter configuration.
Search Manual field DN (Distinguished Name) account with access rights to perform
Distinguished searches. A DN both uniquely identifies an entry and describes its
Name position in the DIT (Directory information Tree). A DN is much like an
absolute path on a file system, except for path direction. File system
paths usually start with the root of the file system and descend the
tree from left to right. LDAP DNs ascend the tree from left to right. For
example, the DN
uid=john.doe,ou=People,dc=example,dc=com represents
an entry that is immediately subordinate to
ou=People,dc=example,dc=com which is itself immediately
subordinate to the entry dc=example,dc=com.
Search Password Manual Field Password for the DN. Can be empty if not required by LDAP server.
Search Scope Base Manual Field. If Search Authentication Method is selected, the search base is DN.
Default: %1 This specifies the search area constraints.
Search Depth LDAP Scope Base* Indicates that only the entry specified as the search base should be
considered. None of its subordinates will be considered.
LDAP Scope Onelevel Indicates that only the immediate children of the entry specified as
the search base should be considered. The base entry itself should not
be considered, nor any descendants of the immediate children of the
base entry.
LDAP Scope Subtree Indicates that the entry specified as the search base, and all of its
subordinates to any depth, should be considered.
LDAP Scope Indicates that the entry specified by the search base should not be
Subordinate considered, but all of its subordinates to any depth should be
considered.
Search Filter Manual Field. Filter for search results. User defined value and accepts AND, OR or
Default: "uid=%1" NOT statements. One wild card can be defined by %1.
30 sec*
60 sec
120 sec
Automatically Enabled A referral is a type of LDAP response that indicates that the server
Follow Referrals could not process the requested operation, but suggests that the
request might succeed if you try it somewhere else.
Disabled*
Source:
https://www.ldap.com/the-ldap-search-operation
https://www.ldap.com/basic-ldap-concepts
Equality = Returns entries containing attribute values that exactly match the
specified value. For example, cn=Bob Johnson
Substring =string* string Returns entries containing attributes containing the specified substring.
For example, cn=Bob* cn=*Johnson cn=*John* cn=B*John. The
asterisk (*) indicates zero (0) or more characters.
Greater than or >= Returns entries containing attributes that are greater than or equal to the
equal to specified value. For example, buildingname >= alpha.
Less than or <= Returns entries containing attributes that are less than or equal to the
equal to specified value. For example, buildingname <= alpha.
Presence =* Returns entries containing one or more values for the specified attribute.
For example, cn=* telephonenumber=* manager=*.
Approximate ~= Returns entries containing the specified attribute with a value that is
approximately equal to the value specified in the search filter. For
example, cn~=suret l~=san fransico could return cn=sarette
l=san francisco.
Source: https://www.centos.org/docs/5/html/CDS/ag/8.0/
Finding_Directory_Entries-LDAP_Search_Filters.html
AND & All specified filters must be true for the statement
to be true. For example, (&(filter)(filter)(filter)...).
Source: https://www.centos.org/docs/5/html/CDS/ag/8.0/
Finding_Directory_Entries-LDAP_Search_Filters.html
LDAP Filter example 1 (does not work with Microsoft Active Directory)
(&(sn=%1)(|(ou:dn:=Users)(ou:dn:=Personnel)(ou:dn:=Staff)))
(&(sAMAccountName=%1)(objectClass=person)(!(objectClass=organizationalP
erson)))
(&(sAMAccountName=%1)(objectClass=person)(objectClass=organizationalPe
rson)(memberOf=cn=People,dc=vsmtesting,dc=local))
5. Enter Distinguished Name in a format that the HIS uses. An example below.
9. When you have entered required data and made required selections, exit
the service mode to return to the clinical mode.
Record Settings
1. Enter Monitor Setup > Advanced > Login Service Mode > Record Settings.
2. Configure output-related items such as PDF Print and Paper.
3. Configure factory default manual fields such as Respiration Rate, Pain
Score, Consciousness, Inputs, Outputs, Oxygen delivery, Oxygen Flow
Rate, O2.
4. Configure Custom field name 9-20 and Free fields. Custom fields are
customizable areas for additional manually entered information and may
have customized drop-down options. Free fields provides more writing
space.
NOTES
Number of accepted decimals is based on decimals entered into the limits,
i.e. the upper limit 10.0 will accept values with one decimal. Both upper and
lower limits should have the same amount of digits.
Unit indicates the selected unit of measurement for numeric field. Only unit
of measurement for numeric field can be customized. Setting in Hospital
EMR encoding units must match with the units set up in EMR.
feature.
Show (On/Number/List/Off)
When Number is selected, the field with numeric value is displayed. The
field value can only be numeric. Min and Max indicate minimum and
maximum values for the field. Unit indicates the display unit for the field
value.
WARNING
The line frequency mode for Masimo SpO2 must be set
according to each country’s electrical power utilities
implementation, and it must be checked and reset any time the
monitor is set to or reverts to factory default settings. If the LF
mode is set incorrectly, the susceptibility to ambient light is
increased and low perfusion performance may be affected,
resulting in potentially inaccurate readings.
1. Enter Monitor Setup > Advanced > Login Service Mode > Maintenance >
Local History and select Remove.
2. Press Confirm to remove all measurements.
WARNING
Make sure you want to do this before you select it. The action
will take place immediately after selecting it. There is no way to
restore patient entries that have been erroneously deleted.
3. Select the home icon to save the measurement settings and exit the Service
Mode. The monitor will restart.
Adding a caregiver
Refer to VC150 Operator’s Manual, section “Patient Data/Adding a caregiver”.
Providing identification
1. Refer to VC150 Operator’s Manual, section “Patient Data/Providing
identification”.
2. Return to “Connectivity verification checklist” on page 4-3 to continue with
the connectivity verification.
Snapshot output
Snapshot merge
Refer to VC150 Operator’s Manual, section “Patient Data/Snapshot merge”.
Snapshot output
1. For details on Export to PDF, Print and Send, refer to VC150 Operator’s
Manual, section “Patient Data/Snapshot output”.
2. Return to “Connectivity verification checklist” on page 4-3 to continue with
the connectivity verification.
Introduction
This chapter provides various checklists for service work performed on the unit.
NIBP
1. Perform an NIBP determination on yourself.
a. Connect an appropriate air hose and an appropriate adult cuff together.
Refer to CARESCAPE VC150 Operator’s Manual for instructions.
b. Connect the end of the air hose to the NIBP connector on the left side of
the monitor.
c. Attach the adult cuff to the upper part of your arm.
d. Select the Inflate icon on the monitor to begin a single determination.
2. Verify that a blood pressure determination completes and the results are
displayed on the monitor.
NOTE
While an NIBP simulator device may be useful to verify that the monitor
responds to oscillometric pulsations, it should not be used as a basis for
assessing the accuracy of measurement. For more information, see
“Appropriate use of NIBP simulators” on page B-2.
Temperature
Welch Allyn
NOTES
The accuracy of predictive temperature measurements made in a water
bath will not be representative of performance with actual patients.
Exergen
1. Connect the Exergen temporal scanner’s into any USB port at the back of
the monitor.
2. Take a temperature reading on yourself (not on a heat generator).
3. Verify that the temperature displays on both the Exergen temporal scanner
and in the Temp parameter box on the monitor. The temperature will be
displayed if Show temperature is selected in the Monitor setup >
Temperature.
EWS
If you want to test operation of the EWS feature, perform the following:
NOTE
All devices are tested and calibrated during manufacturing and are certified
for operation at installation.
To adequately test the safety and integrity of the monitor, the following test
equipment is recommended:
24Vdc power supply
IEC 60601-1 safety tester
Digital manometer (with range to 350 mmHg)
Stopwatch/timer (capable of measuring seconds)
Adult NIBP cuff, hose, inflation bulb, and associated tubing. SuperSTAT
algorithm is currently not available in Canada.
NIBP calibration kit
For Welch Allyn SureTemp Plus® calibration (if SureTemp Plus® option is
installed):
Calibration key, SureTemp Plus® (PN 06138-000)
Calibration tester, SureTemp Plus® 9600+ (PN 01802-110)
For Exergen calibration (if Exergen option is installed):
Calibration verification kit (EX 129003)
SpO2 cable (for the appropriate SpO2 type, if SpO2 is installed)
Printer paper (if printer is installed)
WARNING
Calibration equipment should always be kept dry and free of
particulate matter. Moisture or foreign substances introduced
to the pneumatic system will likely cause damage to the
monitor and/or the accessories.
NOTE
This test is written so that a knowledgeable technician who is familiar with
the monitor and the test equipment will be able to follow the test procedure.
Setup
1. Connect manometer to unit as shown.
2. Use a ‘T’ connection to connect an inflation bulb into the pneumatic setup.
3. Consult the following diagram for pneumatic setup guidelines.
NOTE
Use adult BP cuff for the calibration. Use a mandrel with diameter of
approximately 9 cm.
1. Turn the unit on and select Start in Monitor Setup > Advanced > Service
Mode > Maintenance/Calibration/NIBP Calibration.
2. Close the valve on the inflation bulb. When Module Status displays ready,
select Calibrate.
3. Using the inflation bulb, inflate the system to 210 mmHg.
4. Allow the system to stabilize for about 60 seconds (it is normal to see some
decrease in pressure at this point). Re-inflate the system to 210 mmHg.
5. Start the stopwatch and record the pressure value. Observe the Pressure
Sensor 1 and 2 screens, not the Pressure Sensor Difference screen.
6. After 60 seconds record the pressure value again.
The leakage rate is the difference between the first and second readings.
7. Record the leakage rate on the “Installation and checkout form, CARESCAPE
VC150” on page 5-30.
8. Select Close to cancel and exit the calibration pop up menu.
9. Select the home icon to exit the service mode.
1. Turn the unit on and enter Monitor Setup > Advanced > Service Mode >
Maintenance and select Calibration/NIBP Calibration.
2. Open the valve on the inflation bulb and remove all pressure from the
system (manometer reads zero).
3. When Module Status displays ready, select Adult Overpressure test.
4. Close the valve on the inflation bulb and inflate the cuff, hose and pressure
indicator setup to 200 mmHg.
5. Record the pressure reading in the cell on the Pressure Sensor 1 or 2 of the
NIBP Calibration screen, next to the Adult Overpressure button. Verify that
the digital manometer displays the same reading.
6. Use the valve on the bulb to reduce the pressure to 150 mmHg.
7. Record the pressure reading. Verify that the digital manometer displays the
same reading.
8. Use the valve on the bulb to reduce the pressure to 100 mmHg.
9. Record the pressure reading. Verify that the digital manometer displays the
same reading.
10. Use the valve on the bulb to reduce the pressure to 50 mmHg.
11. Record the pressure reading. Verify that the digital manometer displays the
same reading.
12. Open the bulb to release pressure.
13. Record all test results on “Installation and checkout form, CARESCAPE
VC150” on page 5-30.
14. If readings in the Pressure Sensor 1 or 2 and the manometer are different,
refer to “Pressure transducer calibration” on page 5-7 for instructions.
15. Select Close to cancel and select the home icon to exit the service mode.
1. Turn the monitor on and enter Monitor Setup > Advanced > Service Mode >
Maintenance and select Calibration/NIBP Calibration.
2. Open the valve on bulb to open the pressure system to atmosphere.
3. Verify the manometer reads zero.
4. When Module Status displays ready, select Calibrate.
5. Close the valve on bulb and slowly inflate the pressure to 200 mmHg (using
the manometer as reference).
6. Stop inflation when the pressure on the manometer display reads exactly
200 mmHg.
7. Select Save Calibration.
NOTE
If the calibration was successful, Module Status displays saved and
Pressure sensors 1 and 2 value will initialize to 200 accordingly. If the
calibration was unsuccessful (caused by e.g. too high rate of change 5
mmHg) complete is displayed. Then return to step 1 and restart the
calibration procedure.
Verification
Refer to “Pressure transducer verification” on page 5-6 for instructions.
1. Remove the cuff and hose from the monitor, restrict airflow from cuff hose
port.
2. Select the inflate icon on the home screen to begin NIBP determination.
3. Verify that an overpressure alarm message is displayed on the screen and
an audible alarm sounds.
4. Remove the air restriction.
5. Select the overpressure icon to acknowledge the overpressure condition.
6. Select the inflation icon to verify that the NIBP starts again.
7. Verify that the overpressure alarm message is cleared from the screen.
8. Confirm all test results are recorded on the “Installation and checkout form,
CARESCAPE VC150” on page 5-30.
3. Use the inflation bulb to inflate close to 300 mmHg. Slowly inflate (1 to 2
mmHg/sec) until the valve opens and pressure is released. Observe the
Pressure Sensor 1 and 2 screens, not the Pressure Sensor Difference
screen.
4. Record and verify the pressure at which the valve opens on the “Installation
and checkout form, CARESCAPE VC150” on page 5-30.
6. Use the inflation bulb to inflate close to 150 mmHg. Slowly inflate (1 to 2
mmHg/sec) until the valve opens and pressure is released. Observe the
Pressure Sensor 1 and 2 screens, not the Pressure Sensor Difference
screen. Perform this only where neonatal measurement is allowed by
local regulations.
7. Record and verify the pressure at which the valve opens on the “Installation
and checkout form, CARESCAPE VC150” on page 5-30.
8. Select Close to cancel and exit the calibration pop up menu.
9. Select the home icon to exit the service mode. The monitor will restart.
1. Attach an adult cuff and hose to a simulator (be sure to select the correct
cuff size).
2. Select the inflation icon on the home screen to begin a determination.
3. Record the Systolic, Diastolic, MAP and pulse rate values from the monitor
display.
4. Wait 1 minute, then select the auto mode icon.
5. Select STAT in NIBP Settings/Auto Mode.
6. Select Confirm. A STAT determination starts.
7. Record the Systolic, Diastolic, MAP and pulse rate values from the monitor
display.
8. Select the red STAT icon. This ends the STAT mode.
9. Wait 1 minute, then select the auto mode icon.
10. Select Cycle in NIBP Settings/Auto Mode.
11. Select a cycle in the drop-down list.
12. Select Confirm.
13. Verify that the alarm silence indicator (bell) is lit (in the optional monitoring
mode only).
14. Record the Systolic, Diastolic, MAP and pulse rate values from the monitor
display.
15. Select the red Cycle icon. This ends the Cycle mode.
16. Confirm all test results are recorded on the “Installation and checkout form,
CARESCAPE VC150” on page 5-30.
NOTE
SuperSTAT selection is not displayed if SuperSTAT license is inactive.
SuperSTAT is available depending on regional regulations. SuperSTAT
algorithm is currently not available in Canada.
NOTE
Clear local history before changing the unit of measurement. Refer to
“Clearing local history” on page 4-61.
4. Select the home icon to save the measurement settings and exit the service
mode. The monitor will restart.
Temperature
Perform this only if the monitor is equipped with the temperature parameter.
1. Turn the monitor on. Make sure the temperature probe is properly stored in
the probe well.
2. Adjust Temperature Simulator to 96.8° F (36.0° C) and wait until the
temperature is reached.
3. Place Welch Allyn temperature probe into simulator.
4. If monitor unit of measurement for temperature is set to ° C, switch the
measurement units to ° F. Refer to “Unit of measurement for Welch Allyn” on
page 4-15 for instructions.
5. Measure the temperature. When predictive measurement ends and the
snail icon appears on screen, select the snail icon to put the temperature
measurement into monitor mode.
6. Record and verify that the reading on the temperature display is 96.8° F
±0.2° F.
7. Set the simulator to 101.3° F (38.5° C) and repeat step #6.
8. Record and verify that the reading on the temperature display is 101.3° F
±0.2° F.
9. Set the simulator to 105.8° F (41.0° C) and repeat step #6.
10. Record and verify that the reading on the temperature display is 105.8° F
±0.2° F.
11. If necessary, switch the measurement units back to °C. Refer to “Unit of
measurement for Welch Allyn” on page 4-15 for instructions.
12. Confirm all test results are recorded on the “Installation and checkout form,
CARESCAPE VC150” on page 5-30.
13. Calibration is complete. If the monitor does not pass the calibration
verification, contact GE Technical Support.
NOTE
The Calibration Key test does not test the calibration of the probe. Such a
test requires the use of Welch Allyn Model 9600 Calibration Tester (PN
01802-110).
1. Remove any probe currently connected to monitor. Remove the probe from
the probe well.
2. Connect the SureTemp Plus® (PN 06138-000) calibration key to monitor.
Insert the probe well to monitor.
3. Place the thermometry probe to probe well and then remove it.
4. Wait until a temperature value is shown on screen. The temperature should
be 36.3 +/- 0.1ºC or 97.3 +/- 0.2ºF.
5. Remove the SureTemp® Plus Cal-Key.
6. Connect the original probe back to the monitor.
If the reading from the calibration key is not within specified range or you are
having other problems with the use of the calibration key, perform calibration
verification using the Calibration tester, SureTemp Plus® 9600+ (PN 01802-110)
and a temperature probe. If problem persists, contact GE Technical Support.
NOTE
If calibration is needed, return the scanner to Exergen for repair.
NOTE
Comparisons between the CM and the instrument being tested should
always be conducted under the same conditions.
1. Turn on the verifier device, using either a 9-volt battery or the power supply.
Make sure the red LED is illuminated.
Power On LED
On/Off switch
Battery
compartment
Power supply
jack
Portable black
body verifier
6. Record and verify that the difference between the temperature reading of
the CM and the instrument being verified is ±0.2°C (0.4°F).
If the readings differ by more than the acceptable field limits, repeat this
procedure. If the readings still differ by more than the acceptable limits, the
device fails.
7. Confirm that test results are recorded on the “Installation and checkout
form, CARESCAPE VC150” on page 5-30.
External DC verification
1. Plug the power supply into the monitor.
2. Verify that the Charging indicator is illuminated.
3. Confirm all test results are recorded on the “Installation and checkout form,
CARESCAPE VC150” on page 5-30.
SpO2
Perform the following steps only if the monitor is equipped with the SpO2
parameter.
1. Connect the appropriate SpO2 probe and connect appropriate SpO2 sensor
to the SpO2 connector. Use extension cable if necessary. Place the probe on
your finger.
2. Verify that the unit displays:
Pulse value
Saturation value
Signal strength asterisks
3. Remove the sensor from your finger. Dashes ‘--’ will be displayed in
parameter area, an SpO2 Sensor Off Patient alarm message will be
displayed on the notification area and an alarm sound is signaled (in
monitoring mode) through the speaker.
4. Select the Silence icon.
5. Verify that the sound is silenced.
6. Select the alarm message to acknowledge it.
7. Verify the alarm message is cleared.
8. Reapply the SpO2 sensor to your finger.
9. Verify that the unit displays:
Pulse value
Saturation value
Signal strength asterisks
10. Confirm that all test results are recorded on the “Installation and checkout
form, CARESCAPE VC150” on page 5-30.
Perform the following steps only if the monitor is equipped with a printer.
1. Check whether the printer has a paper roll inside. If not, load paper into the
print mechanism.
NOTE
Use only printer paper from GE. Refer to CARESCAPE VC150 Supplies
and accessories (KO00125) for printer accessories.
a. With the monitor powered on, turn it so that the side with the printer is
facing you.
b. While grasping the side of the monitor, lift the printer door open by
placing your thumb in the indented area and pulling. The printer door
will pop open.
c. Place the roll of paper into the compartment so that the end of the
paper comes off the right side of the roll (paper is wound around the roll
clockwise). Place the roll of paper in the holding bracket that is
integrated in the door of the printer, making sure the paper extends out
of the printer cavity for at least 5 cm (two inches).
2. Firmly press the door to close it.
3. Select Monitor Setup > Advanced > Service Mode > Maintenance/Print
Last 50 Errors to check the printer output. You can also print a snapshot, if
available.
4. Verify that the printer outputs a record and the print quality is good.
5. Confirm that test results are recorded on the “Installation and checkout
form, CARESCAPE VC150” on page 5-30.
Safety testing
Electrical safety tests
Electrical safety tests provide a method of determining if potential electrical
health hazards to the patient or operator of the device exist.
Requirements
Qualified personnel must perform all safety tests presented in this document:
NOTES
If local authorities require safety tests before the device can be placed in
clinical use, the hospital biomed is responsible to perform tests listed in
“Electrical safety tests” on page 5-18.
The manufacturer has performed these electrical safety tests for patient
monitor and acquisition modules during final inspection. You do not have to
perform the electrical safety tests during installation checkout, if the patient
monitor was manufactured less than 24 months ago. Check the serial
number label for manufacturing week and year (refer to “Equipment ID” on
page 1-6).
Test equipment
The test equipment needed to perform electrical safety tests is listed below.
Item Specification
Perform electrical safety tests using an electrical safety analyzer per IEC 60601-
1, UL 60601-1, or CSA C22.2 No. 601. The schematics in the section provide a
general understanding of the test equipment. Actual configuration of test
equipment may vary.
Failure of the power cord strain relief is very common. Often users of the
equipment pull on the power cord itself, rather than the power cord plug, to
unplug the patient monitor from a wall receptacle.
Inspect the power cord for wear or damage regularly. If damage is
suspected, test for continuity through each conductor of the power cord
connector.
Verify line, neutral, and earth conductors are properly connected to the
power cord plug and are not short-circuited. Replace the power cord as
necessary with a regulatory-approved cord for the country of use.
WARNING
Use only AC power cords recommended or manufactured by
the manufacturer.
Perform tests in both normal condition (NC) and in single fault condition (SFC),
where Neutral conductor is open. Perform the test with normal and reverse
polarity.
Acceptance criteria NC
All readings must be less than or equal to 500 µA.
Metal bar of the printer frame or metal plate of the dummy printer cover (if
the monitor is not equipped with the printer)
CAUTION
The metal bar must not hit the writing head of the printer.
PC USB connector frame (use the PC-USB safety test cable of the VC150
Safety Test Connectors kit, GE p/n 2085615-001)
Remote alarm connector frame (use the Nurse Call safety test cable of the
VC150 Safety Test Connectors kit, GE p/n 2085615-001)
Perform tests in both normal condition (NC) and in single fault condition (SFC),
where one of the supply conductors is open at a time. Perform the test with
normal and reverse polarity.
Leakage Tester
LINE
AC MAINS
NORM
Power Cord
NEUTRAL LINE
Open
EARTH NEUTRAL
Closed
EARTH
RVS
Device
Open
Under
Closed
Test
Printer/Enclosure
MD*
NOTES
*The MD represents the network and voltage measuring instrument and its
frequency characteristics per IEC 60601-1.
15. Remove the AC mains voltage from the device under test.
If the measured reading is greater than the acceptance criteria below, the
device under test fails. Contact GE Service.
Acceptance criteria NC
All readings must be less than or equal to 100 µA.
This procedure measures the leakage current from the “Device Under Test” input
connector to ground.
Perform tests in both normal condition (NC) and in single fault condition (SFC),
where one of the supply conductors is open at a time. Perform the test with
normal and reverse polarity.
Leakage Tester
LINE
AC MAINS
NORM
Power Cord
NEUTRAL LINE
Open
EARTH NEUTRAL
Closed
EARTH
RVS
Device
Open
Under
Closed Test
Patient connector
MD*
NOTES
*The MD represents the network and voltage measuring instrument and its
frequency characteristics per IEC 60601-1.
Perform this test once for SpO2 and once for temperature.
This test applies only to the Welch Allyn temperature option; it is not
intended for the Exergen temperature option.
Perform first AC leakage current measurements.
17. Repeat the steps in this procedure to measure the other patient connector.
If the measured reading is greater than the acceptance criteria below, the
device under test fails. Contact GE Service.
Acceptance criteria NC
All AC leakage current readings must be less than or equal to 100 µA.
All DC leakage current readings must be less than or equal to 10 µA.
This procedure measures the leakage current from a mains voltage source into
the “Device Under Test” input connector.
Leakage Tester
LINE
AC MAINS
NORM Power Cord
NEUTRAL LINE
EARTH NEUTRAL
EARTH
RVS
**
Device
Under
Closed Test
Patient connector
(Keep cable length as
MD* short as possible.)
NOTES
*The MD represents the network and voltage measuring instrument and its
frequency characteristics per IEC 60601-1.
**Per IEC 60601-1, the resistance to protect the circuitry and the person
performing the test. The resistance must be low enough to accept currents
higher than the allowable values of the leakage current to be measured.
Perform this test once for SpO2 and once for temperature.
This test applies only to the Welch Allyn temperature option; it is not
intended for the Exergen temperature option.
WARNING
Shock hazard. The following step causes high voltage at the
patient connector. Do not touch the patient connector.
Acceptance criteria
All readings must be less than or equal to 5 mA.
Perform tests in both normal condition (NC) and in single fault condition (SFC), at
which one of the supply conductors is open at a time. Perform the test with
normal and reverse polarity.
Leakage Tester
LINE
AC MAINS
NORM
Power Cord
NEUTRAL LINE
Open
EARTH NEUTRAL
Closed
EARTH
RVS
Device
Open
Under
Patient connector
Closed
(SpO2) Test
MD*
Patient connector
(TEMP)
NOTES
*The MD represents the network and voltage measuring instrument and its
frequency characteristics per IEC 60601-1.
This test applies only to the Welch Allyn temperature option; it is not
intended for the Exergen temperature option.
Perform first AC leakage current measurements.
6. Read the current leakage indicated on the tester and record the results on
the “Installation and checkout form, CARESCAPE VC150” on page 5-30.
7. Change leakage tester switches to:
Polarity – NORMAL
Neutral – CLOSED
GND (Earth) – OPEN
8. Read the current leakage indicated on the tester and record the results on
the “Installation and checkout form, CARESCAPE VC150” on page 5-30.
9. Change leakage tester switches to:
Polarity – REVERSED
Neutral – CLOSED
GND (Earth) – OPEN
10. Read the current leakage indicated on the tester and record the results on
the “Installation and checkout form, CARESCAPE VC150” on page 5-30.
11. Change leakage tester switches to:
Polarity – REVERSED
Neutral – OPEN
GND (Earth) – CLOSED
12. Read the current leakage indicated on the tester and record the results on
the “Installation and checkout form, CARESCAPE VC150” on page 5-30.
13. Change leakage tester switches to:
Polarity – REVERSED
Neutral – CLOSED
GND (Earth) – CLOSED
14. Read the current leakage indicated on the tester and record the results on
the “Installation and checkout form, CARESCAPE VC150” on page 5-30.
15. Remove the AC mains voltage from the device under test.
16. Repeat the steps in this procedure for DC leakage current measurement.
If the measured reading is greater than the acceptance criteria below, the
device under test fails. Contact GE Service.
Acceptance criteria NC
All AC leakage current readings must be less than or equal to 100 µA.
All DC leakage current readings must be less than or equal to 10 µA.
This procedure measures the leakage current from the "Device Under Test"
patient connectors to ground. Patient connectors must be connected together.
Perform tests in both normal condition (NC) and in single fault condition (SFC),
where one of the supply conductors is open at a time. Perform the test with
normal and reverse polarity.
Leakage Tester
AC MAINS
LINE NORM
Power Cord
NEUTRAL LINE
Open
EARTH NEUTRAL
Closed
EARTH
RVS
Device
Open
Patient connector Under
Closed (SPO2) Test
MD*
Patient connector
(TEMP)
NOTES
*The MD represents the network and voltage measuring instrument and its
frequency characteristics per IEC 60601-1.
This test applies only to the Welch Allyn temperature option; it is not
intended for the Exergen temperature option.
Perform first AC leakage current measurements.
6. Read the current leakage indicated on the tester and record the results on
the “Installation and checkout form, CARESCAPE VC150” on page 5-30.
7. Change leakage tester switches to:
Polarity – NORMAL
Neutral – CLOSED
GND (Earth) – OPEN
8. Read the current leakage indicated on the tester and record the results on
the “Installation and checkout form, CARESCAPE VC150” on page 5-30.
9. Change leakage tester switches to:
Polarity – REVERSED
Neutral – CLOSED
GND (Earth) – OPEN
10. Read the current leakage indicated on the tester and record the results on
the “Installation and checkout form, CARESCAPE VC150” on page 5-30.
11. Change leakage tester switches to:
Polarity – REVERSED
Neutral – OPEN
GND (Earth) – CLOSED
12. Read the current leakage indicated on the tester and record the results on
the “Installation and checkout form, CARESCAPE VC150” on page 5-30.
13. Change leakage tester switches to:
Polarity – REVERSED
Neutral – CLOSED
GND (Earth) – CLOSED
14. Read the current leakage indicated on the tester and record the results on
the “Installation and checkout form, CARESCAPE VC150” on page 5-30.
15. Remove the AC mains voltage from the device under test.
16. Repeat the steps in this procedure for DC leakage current measurements.
If the measured reading is greater than the acceptance criteria below, the
device under test fails. Contact GE Service.
Acceptance criteria NC
All AC leakage current readings must be less than or equal to 500 µA.
All DC leakage current readings must be less than or equal to 50 µA.
Results
Measurement limits must comply with 3rd edition of IEC 60601-1.
Service
OK = Test OK
Mechanical and electrical checks N.A. = Test not applicable
Fail = Test failed
“Pressure transducer verification” on page 5-6 Min Max Actual OK N.A. Fail
“NIBP overpressure verification” on page 5-8 Min Max Actual OK N.A. Fail
Safety testing - Earth leakage Current Min Max Actual OK N.A. Fail
SpO2 patient auxiliary current test AC Min Max Actual OK N.A. Fail
SpO2 patient auxiliary current test DC Min Max Actual OK N.A. Fail
Temp patient auxiliary current test AC Min Max Actual OK N.A. Fail
Temp patient auxiliary current test DC Min Max Actual OK N.A. Fail
Total patient leakage current test AC Min Max Actual OK N.A. Fail
Total patient leakage current test DC Min Max Actual OK N.A. Fail
Overview
The symptoms and solutions in this chapter represent only a few of the
problems that you may encounter and are not intended to cover every possible
problem that may occur.
System failures
A technical error message is displayed when the monitor has a depleted battery,
or a hardware or software failure. When a system failure is encountered, an
alarm message is displayed on the screen and the error is logged in the error log.
If the power is lost or the monitor was booted, a note is added to the error log.
Alarm conditions
When responding to a monitor alarm, always check the patient first and then
check the monitor, cuff, hose and sensors. Select the alarm message to
acknowledge the alarm condition.
Logs
Error log, productivity metrics and EMR transaction log help out with debugging
or obtaining information on performance. Examine log files to find out
discrepancies or errors. If there is another monitor that operates without
problems, you can try comparing logs of two monitors to determine differences
in settings.
Error log
An error log contains information on errors.
The log is a “rolling” list and minimum of 50 latest error log entries can be found
in Monitor Setup > Service Mode > Maintenance screen. Scroll down to Logging
> Error Log and select View.
You can print an error log with up to 50 error condition entries. In addition, error
log can be exported to PC as instructed in “Backing up and importing/restoring
settings” on page 9-12. The error log is saved until the monitor experiences a
memory loss, then all entries are deleted.
Productivity metrics
The monitor provides productivity metrics log that can be used to find out for
example the following:
caregiver ID
time of login and logout
NTP related messages
changes in date and time
transferred data to EMR
Factory Factory
Alarm condition Alarm limit range
default limit default priority
NIBP Systolic High (adult/pediatric) 35 to 290 or 4.7 - 38.7 kPa 200 mmHg or 26.6 kPa Medium
NIBP Systolic High (neonate) 35 to 140 or 4.7 - 18.7 kPa 100 mmHg or 13.3 kPa
NIBP Systolic Low (adult/pediatric) 30 to 285 or 4.0 - 38.0 kPa 80 mmHg or 10.6 kPa
NIBP Systolic Low (neonate) 30 to 135 or 4.0 - 18.0 kPa 40 mmHg or 5.3 kPa
NIBP Diastolic High (adult/pediatric) 15 to 220 or 2.0 - 29.3 kPa 120 mmHg or 16 kPa
NIBP Diastolic High (neonate) 15 to 110 or 1.0 - 14.7 kPa 60 mmHg or 8 kPa
NIBP Diastolic Low (adult/pediatric) 10 to 215 or 1.3 - 28.7 kPa 30 mmHg or 4 kPa
NIBP Diastolic Low (neonate) 10 to 105 or 1.3 - 14.0 kPa 20 mmHg or 2.6 kPa
Factory Factory
Alarm condition Alarm limit range
default limit default priority
SpOC High2 (Masimo) 2.0 - 34.0 mL O2/dl blood 25.0 O2/dl blood
SpOC Low2 (Masimo) 1.0 - 33.0 O2/dl blood 10.0 O2/dl blood
Factory Factory
Alarm condition Alarm limit range
default limit default priority
To change alarm default settings, refer to Refer to “Alarm limit setup” in CARESCAPE VC150 Operator’s Manual for
instructions.
B-USB does not work Faulty USB cable. Try with another USB cable.
Barcode scanner does not work Refer to "USB accessory does not work" section of this table.
Monitor does not shut down Refer to flowchart “Monitor does not shut down” on page 6-36.
Monitor does not start without Refer to flowchart “Monitor does not start without power source” on page 6-
power source 31.
Monitor does not turn on Refer to flowchart “Monitor does not turn on” on page 6-32.
Welch Allyn:
Refer to flowchart “Welch Allyn temp not available” on page 6-35.
Refer to “Suretemp+ messages” on page 6-27.
Temp errors
Exergen:
Refer to “Exergen messages” on page 6-12"
Refer to "USB accessory does not work" section of this table.
Touch function does not work Refer to flowchart “Touch function does not work” on page 6-33.
NOTE
NOTE When the service user changes
When debugging WLAN/EMR, do format for a parameter, every other
not use real patient information! parameter will be ignored when
measurement data is sent to EMR.
Service user must remember to
change the encoding for every
parameter that needs to be sent.
Refer to “Hospital EMR encoding” on
page 4-46.
USB board cable or USB board 1. Test if the device works in another
broken USB port.
2. Measure USB signals from the
back panel USB connectors and the
USB accessory does not work
test points on the USB board. Refer to
“Back panel USBs” on page 6-48.
4. If measurements indicates error
and device does not work in another
USB port, replace cable between the
USB board and the Motherboard
5. If the previous step does not fix the
problem, replace USB board
6. If still not fixed, replace the
motherboard.
GE TruSignal V2 messages
SpO2 parameter
Notification area box Error Log Message / Description Service User Action
(GE TruSignal V2)
Measurement conditions
SpO2 Bad Bad Placement Poor sensor placement. Refer to CARESCAPE VC150 user
Placement manual for correct
CARESCAPE VC150 vital sign
determination.
SpO2 Faulty Faulty probe Sensor Failure. Reboot the monitor. Replace the
probe sensor.
Refer to logs for details.
SpO2 Low signal Low signal Low signal quality. Refer to CARESCAPE VC150 user
quality quality manual for correct
CARESCAPE VC150 vital sign
SpO2 Sensor Off Sensor Off Sensor off patient. determination.
Patient Patient
If the problem persists, after
SpO2 Pulse Pulse Search Waiting the algorithm to provide a
essential service procedures,
Search valid SpO2 value.
replace the sensor.
SpO2 Measurement Data update period has exceeded Reposition the sensor. Replace the
Measurement Delayed 30 seconds. sensor if the problem persists.
Delayed Reboot the monitor. If the problem
persists, contact the manufacturer.
Hardware codes
SpO2 Not Not Available Refer to Error Log for details. Check isolation and flex connector
Available contacts. Check operating voltages
from Isolation board. Refer to
“Isolation interface” on page 6-45.
Exergen messages
Exergen messages
TEMP parameter External Thermometry
Notification area Service User Action
box Measurement status
Temp Battery Battery Low Battery Low Error Replace the Exergen battery as instructed in
Low “Changing the Exergen temperature unit
battery” on page 7-70.
Temp Battery Battery Empty Battery Fatal Error
Empty
Temp Measurement High Target Refer to CARESCAPE VC150 user manual for
Measurement Too High correct CARESCAPE VC150 vital sign
Too High determination.
Temp Measurement Low Target If the problem persists after essential service
Measurement Too Low procedures, replace the Exergen hand held
Too Low device.
Temp Ambient Ambient high High Ambient
high
Exergen troubleshooting
Exergen temperature scanner Loose connection to USB port. If the scanner appears to be working
appears to work properly, but does Scanner battery cover not properly, but will not communicate
not communicate with the monitor. secured. with the monitor:
1. Verify that the connector is
Loss of electrical contact
properly connected.
between the scanner and the
2. Verify that the scanner battery
monitor (e.g., corrosion on the
cover is secured.
connector inside the scanner).
If there is still no communication with
The monitor is not configured to the monitor:
interface with an Exergen 1. Plug and unplug the connector
scanner. several times.
2. Unplug the monitor and remove
the battery.
3. Rub off any corrosion on the
connector inside the scanner.
4. Check if the monitor is configured
to interface with an Exergen
scanner.
Host messages
SpO2 Not Available Oximetry OEM module initialization SYSTEM Check the system and module
(Not Available in the not finished within 30 seconds after version are compatible. Reboot the
SpO2 parameter box) the application start up. unit.
Temp Not Available Thermometry OEM module SYSTEM Check the system and module
(Not Available in the initialization not finished within 30 version are compatible. Reboot the
thermometry seconds after the application start unit.
parameter box) up.
If the problem persists, refer to
Refer to Error Log for details. flowchart “Welch Allyn temp not
available” on page 6-35 and to
module specific messages and user
actions.
NIBP Not Available NIBP OEM module initialization not SYSTEM Check the system and module
(Not Available in the finished within 30 seconds after the version compatibility. Reboot the
NIBP parameter box) application start up. unit. Refer to the module specific
messages and user actions.
Refer to Error Log for details.
If the problem persists, replace
2067981-008 VC150 FRU NIBP
module.
Battery Low Blue, low priority alarm indicates that BATTERY If the alarm occurs earlier than
the battery level is low and expected, check the battery type. If
approximately 10 NIBP the type is wrong or the problem still
measurements can be taken. persists, drain the battery, clear the
battery monitoring algorithm and
Red, high priority alarm indicates charge cycle the battery to adapt the
that the battery level is low and the algorithm.
monitor will shut down within 5
minutes. If problem still persist replace
2067981-027 VC150 FRU Battery.
System did not shut The application did not close cleanly SYSTEM If the problem persists frequently,
down correctly before the latest restart, e.g. remove the monitor from clinical
depleted battery use.
Patient ID required Attempted to send an EMR record EMR Check connectivity settings and
without a valid patient ID. Authentication method settings.
Patient ID requires Validated patient ID is required for EMR Check connectivity settings and
validation sending an EMR record, but the Authentication method settings.
patient ID is not validated.
Caregiver ID required Attempted to send an EMR record EMR Check connectivity settings and
without a valid caregiver ID. Authentication method settings.
Caregiver ID requires Validated caregiver ID is required for EMR Check connectivity settings and
validation sending an EMR record, but the Authentication method settings.
caregiver ID is not validated.
Invalid EMR Encoding Hospital specific EMR settings are EMR Check settings of the local history
not valid. and EMR.
Send to EMR failed Sending an EMR record failed due to EMR Check EMR settings or WLAN range.
network or other error.
Invalid SSL Authenticating Patient Data Query EMR Check SSL settings, ensure the
configuration failed due to an SSL error. certificate is valid and up-to-date.
Cannot connect to Authenticating Patient Data Query EMR Check connectivity settings and
remote server failed due to connection error to Authentication method settings.
remote host.
Invalid LDAP Search Filter (if LDAP LDAP Refer to Error log.
Search method is selected)
Invalid LDAP Search Base (if LDAP LDAP Refer to Error log.
Search method is selected)
PM expired N days ago Preventive maintenance date SYSTEM Perform the preventive maintenace
passed N days ago. as instructed in “Preventive
maintenance checklist” on page 9-3.
PM in N days Preventive maintenance date will SYSTEM Make sure the needed FRU's are
pass after N days. available when the preventative
maintenance is due.
Local History Storing patient data to Local History LOCAL If storage notifications are not
Unavailable database failed, possibly because HISTORY reported the storage problem is
the storage space is full. most likely due to mass storage
device wear or system memory
Refer to Error Log for details. exhaust due to excess amount of
entries in the local history. Send or
export the patient data and reboot
the unit in order to free the
resources.
Local History Unit Storing a physiological LOCAL Check the unit settings on Exergen
Error measurement to the Local History HISTORY temporal scanner or SureTemp+
database failed because the unit of module, NIBP module and for SpHb if
measurement has changed, only enabled.
allow one unit per type of
measurement in the DB.
Local History Reset Local History database is reset. LOCAL Informative message, no action
HISTORY needed. Appears when Local history
is removed from Service Mode.
Failed to update time Updating the system time from NTP TIME Check connectivity settings, WLAN
server failed, possibly because of a settings and NTP server settings.
network error.
Password has been Setup password has been changed. SECURITY Informative message, no action
changed needed. Appears when Default
Setup or Service Setup password
has been changed.
Passwords do not Changing setup password failed SECURITY If the password is correct for sure, try
match because the given passwords do not the default one in case the
match. passwords has been accidentally
reset. Try accessing the service
settings through the remote
management and check the
localization settings and import
known settings.
System will restart The user has entered Default Setup SECURITY If the message is encountered
after setting changes or Service mode and the monitor will unexpectedly, reboot the monitor.
restart if settings are changed.
Patient will be User is about to change the system SECURITY If the message is encountered
discharged if time is time, warn about discharging patient unexpectedly, reboot the monitor.
changed if the system time changes.
Audio not available Audio cannot be played through the SYSTEM If the message is encountered
speaker inside the enclosure, unexpectedly, reboot the monitor.
possibly because of a hardware
problem.
USB not available All functionality that relies on USB SYSTEM If the message is encountered
communication is unavailable, unexpectedly, reboot the monitor.
possibly because of a hardware
problem. If the problem persists, replace the
USB board and/or the motherboard.
Refer to Error Log for details.
USB <port number> Connected USB device doesn't SYSTEM Replace the USB device.
overcurrent operate as it is drawing excessive
amount of power. If the problem persists, replace the
USB board and/or the motherboard.
Refer to Error Log for details.
Settings changed After importing new settings SYSTEM Informative message, no actions
remotely, medium priority alarm required. If problem persists, refer to
signaled in the first start-up. “Troubleshooting flow charts” on
page 6-31.
Storage 75% full Permanent storage partition volume STORAGE Disable DCS feature and remove
is nearly full. local history. Refer to NIBP DCS
Data in “Logging” on page 4-18
and “Clearing local history” on
page 4-61.
Storage 90% full Permanent storage partition volume STORAGE
Export logs and DCS files to free
is nearly full.
up storage space. Refer to
VC150 Operator Manual
KO00065, section “Snapshot”
and “Exporting error logs” on
page 6-3.
Storage 100% full Permanent storage partition volume STORAGE Disable DCS feature. Refer to
is full. NIBP DCS Data in “Logging” on
page 4-18.
Delete DCS data. Refer to NIBP
DCS Data in “Logging” on
page 4-18.
Remove local history. Refer to
NIBP DCS Data in “Clearing local
history” on page 4-61.
If problem persists, remove the
monitor from clinical use and
return it to the manufacturer.
Printer Temperature Printer temperature is too high. PRINTER Open the printer door and let the
printer cool down for 5 minutes. If
problem persists, refer to the
“Cannot print” section of the
“Technical error conditions” on
page 6-7.
Printer out of paper Printer is out of paper. PRINTER Add paper to the printer. If problem
persists, refer to the “Cannot print”
section of the “Technical error
conditions” on page 6-7.
Printer error Printer has encountered an error. PRINTER Reboot the monitor. Check that the
printer paper is in place and do a test
print. Service Mode > Maintenance
> Logging > Print Last 50 Errors.
Internal bus MSP communication is interrupted. SYSTEM Shut down the monitor. Unplug the
communication error Displayed also as a full screen battery connector. Wait 10 seconds
Please contact service message. and then reattach the battery.
Charge battery for four hours and
Refer to Error Log for details. then start the monitor.
Battery not connected Displayed only as a full screen SYSTEM Refer to flowchart “Monitor does not
message when battery is detached start without power source” on
or broken. page 6-31.
Invalid Identification Incorrect ID due to missing or SYSTEM Make sure the used identification
incomplete ID or missing password. parameters are valid.
Caregiver Authentication of the caregiver failed SYSTEM Make sure the used identification
authentication failed due to wrong/invalid password or parameters are valid.
unknown user name.
Programming Message indicates that the SpO2 SYSTEM Informative message, no action
Complete programming is complete. required.
Programming Not Message is displayed when external SYSTEM Set the Programming Allowed
Allowed programming dongle is plugged in parameter to Yes from the Service
the SpO2 connector, but the Mode > Measurement Settings
Programming Allowed parameter is
not changed to Yes
Programming Failed Message indicates that the SYSTEM Set the Programming Allowed
programming has failed. This can parameter to Yes from the Service
happen e.g. if the programming Mode > Measurement Settings
dongle is detached from the SpO2
connector during the programming. Connect the programming dongle to
the SpO2 connector and do not
Refer to “Masimo messages” on detach it during the programming.
page 6-20
Masimo messages
Parameter box
Notification area Description Service User Action
(Masimo)
RRa Faulty Probe Faulty Probe in RRa Defective sensor or Replace the sensor.
parameter box adhesive sensor
RRa Interference Interference Patient interference Refer to CARESCAPE VC150 user manual
Detected Detected in RRa detected for correct CARESCAPE VC150 vital sign
parameter box determination.
RRa No Cable No Cable No cable connected Refer to CARESCAPE VC150 user manual
Connected Connected in RRa for correct CARESCAPE VC150 vital sign
parameter box determination.
RRa No Sensor No Sensor No sensor or adhesive Refer to CARESCAPE VC150 user manual
Connected Connected in RRa sensor connected for correct CARESCAPE VC150 vital sign
parameter box determination.
RRa Not Available Not Available in Refer to Error log for Refer to CARESCAPE VC150 user manual
RRa parameter box details. for correct CARESCAPE VC150 vital sign
determination.
Refer to Productivity
Metrics for details.
Parameter box
Notification area Description Service User Action
(Masimo)
RRa Replace Sensor Replace Sensor in Sensor or adhesive Refer to CARESCAPE VC150 user manual
RRa parameter box sensor life expired for correct CARESCAPE VC150 vital sign
Unrecognized sensor determination.
RRa Sensor Off Sensor Off Patient Sensor is off patient Refer to CARESCAPE VC150 user manual
Patient in RRa parameter for correct CARESCAPE VC150 vital sign
box determination.
SpHb Not Available N/A Sensor lifetime expired Replace the SpHb sensor.
SpO2 Demo Mode Demo Mode in SpO2 Demo mode. Replace 2067981-013 FRU VC150 SpO2
parameter box Masimo Assy.
Refer to Error Log for
details.
SpO2 Faulty Cable Faulty Cable in Defective cable Replace the cable.
SpO2 parameter box Sensor/cable fault
SpO2 Faulty Probe Faulty Probe in Defective sensor or Replace the sensor.
SpO2 parameter box adhesive sensor
SpO2 Interference Interference Interference detected. Refer to CARESCAPE VC150 user manual
Detected Detected in SpO2 for correct CARESCAPE VC150 vital sign
parameter box determination.
SpO2 No Cable No Cable No cable connected Check the sensor and patient cable
Connected Connected in SpO2 connection. Try with a different sensor.
parameter box
If the problem persists, replace
2067981-016 VC150 FRU SpO2 Masimo
accessories.
Parameter box
Notification area Description Service User Action
(Masimo)
SpO2 No Sensor No Sensor No sensor or adhesive Refer to CARESCAPE VC150 user manual
Connected Connected in SpO2 sensor connected for correct CARESCAPE VC150 vital sign
parameter box determination.
Refer to Productivity
Metrics for details. Check the sensor and patient cable
connection. Try with a different sensor.
SpO2 Not Available Not Available in Refer to Error log for If a replacement SpO2 module was
SpO2 parameter box details. installed, update the monitor software
to the latest version. Refer to “Software
upgrade” on page 10-2.
SpO2 - SpO2 Only - SpO2 Only Mode in SpO2 only -mode. OEM manufacturer specific mode,
Mode SpO2 parameter box contact OEM manufacturer support
Refer to Productivity organisation.
Metrics for details.
SpO2 Perfusion Low Perfusion Low in Low perfusion index. Refer to CARESCAPE VC150 user manual
SpO2 parameter box for correct CARESCAPE VC150 vital sign
determination.
Programming Failed N/A Application error Refer to “Host messages” on page 6-13.
Communication error
FLASH error Check power supply and replace if
Programming timeout, necessary. Check the upgrade tool
issued by the driver. compability, MX-5 board type and MX-5
board mode.
Refer to Error log for
details. If the problem persists, replace
2067981-013 FRU VC150 SpO2 Masimo
Assy.
Parameter box
Notification area Description Service User Action
(Masimo)
SpO2 Pulse Search Pulse Search in Pulse search. Refer to CARESCAPE VC150 user manual
SpO2 parameter box for correct CARESCAPE VC150 vital sign
determination.
SpO2 Replace Cable Replace Cable in Cable life expired Replace the cable.
SpO2 parameter box Unrecognized cable
Refer to Productivity
Metrics for details.
SpO2 Replace Sensor Replace Sensor in Unrecognized sensor Replace the sensor.
SpO2 parameter box or adhesive sensor
Sensor or adhesive
sensor life expired
Refer to Productivity
Metrics for details.
SpO2 Sensor Off Sensor Off Patient Sensor is off patient Refer to CARESCAPE VC150 user manual
Patient in SpO2 parameter for correct CARESCAPE VC150 vital sign
box determination.
Time remaining on N/A Shows the remaining No action required. After specified time
SpHb sensor lifetime of the SpHb replace the sensor.
sensor.
Nellcor messages
SpO2 Not Available Not Available Module malfunction If a replacement SpO2 module was
installed, update the monitor software
Refer to Error Log for to the latest version. Refer to “Software
details. upgrade” on page 10-2.
SpO2 Faulty probe Faulty probe Defective or incompative Replace the sensor.
sensor
If the problem persists, replace
2067981-015 VC150 FRU SpO2 Nellcor
Assy
Measurement conditions
SpO2 Interference Interference Refer to Productivity Refer to CARESCAPE VC150 user manual
Detected Detected Metrics for details. for correct CARESCAPE VC150 vital sign
determination.
SpO2 Sensor Off Sensor Off Patient Refer to Productivity Refer to CARESCAPE VC150 user manual
Patient Metrics for details. for correct CARESCAPE VC150 vital sign
determination.
SpO2 Pulse Search Pulse Search Refer to Productivity Refer to CARESCAPE VC150 user manual
Metrics for details. for correct CARESCAPE VC150 vital sign
determination.
SpO2 Pulse Timeout Pulse Timeout Refer to Productivity Refer to CARESCAPE VC150 user manual
Metrics for details. for correct CARESCAPE VC150 vital sign
determination.
NIBP messages
Fatal Codes
NIBP Not Available Not Available Refer to Error Log for Refer to essential service procedures.
details.
If the problem persists, replace
2067981-008 VC150 FRU NIBP module.
NIBP Pneumatic leak Pneumatic leak Pump-up timeout, Refer to CARESCAPE VC150 user manual
possible pneumatic leak for correct hose and cuff placement.
NIBP Unable to Zero Unable to Zero Pressure too great Refer to CARESCAPE VC150 user manual
between measurements for correct NIBP accessory usage and
vital sign measurement procedures.
NIBP Level Timeout Level Timeout Maximum allowed time Refer to CARESCAPE VC150 user manual
of 1 minute reached for a for correct hose and cuff placement.
single cuff pressure.
If the problem persists, replace Cuff and
Refer to Productivity hose. If problem still persists, replace
Metrics and Error Log 2067981-008 VC150 FRU NIBP module.
for details.
NIBP Total Timeout Total Timeout Length of determination Refer to CARESCAPE VC150 user manual
has exceeded 2 minutes for correct hose and cuff placement.
for an adult/pediatric, or
85 seconds for a If the problem persists, replace Cuff and
neonatal hose. If problem still persists, replace
determination. 2067981-008 VC150 FRU NIBP module.
Refer to Productivity
Metrics for details.
Suretemp+ messages
Temp Ambient Ambient High Ambient temperature too high. Check internal temperature of the
High probe well. Verify ambient
temperature does not exceed
Temp Ambient Ambient Low Ambient temperature too low. operating specifications.
Low
Temp Faulty Faulty Probe Refer to Productivity Metrics and Try another predict cycle. If the
Probe Error Log for details. problem persists, replace the
probe.
Temp Interference Battery or power-supply voltage Verify that supply voltage conforms
Interference Detected exceeds algorithm maximum value. to specifications.
Detected
Refer to Productivity Metrics and
Error Log for details.
Temp Measurement Measured temperature too high. Check the patient’s condition. Refer
Measurement Too High to CARESCAPE VC150 user manual
Too High for correct CARESCAPE VC150 vital
sign determination.
Temp Measurement Measured temperature too low. Check the patient’s condition. Refer
Measurement Too Low to CARESCAPE VC150 user manual
Too Low
for correct CARESCAPE VC150 vital
sign determination.
Temp No No A predictive measurement could not Check the patient’s condition. Refer
Determination Determination be completed. to CARESCAPE VC150 user manual
for correct CARESCAPE VC150 vital
Refer to Productivity Metrics and
Error Log for details. sign determination.
Temp Not Not Available Refer to Error Log for details. Reboot the monitor. Verify that
Available supply voltage is correct.
Temp Probe Too Probe Too Hot Temperature conditions are such Correct probe tip conditions.
Hot that the probe is located above If the problem persists, replace the
112F/43.3C. probe. If the problem still persists,
replace the module.
Temp Probe Well Probe Well Probe well missing or not installed Reattach the probe well and reboot
Missing Missing properly. the monitor. If the problem persists,
replace 2067981-024 VC150 FRU
Welch Allyn Frame Assy.
Temp Replace Replace Sensor Probe not characterized/calibrated. Replace temperature probe.
Sensor
Refer to Productivity Metrics and
Error Log for details.
Temp Sensor Off Sensor Off Probe is off patient or poorly placed. Refer to CARESCAPE VC150 user
Patient Patient manual for correct probe
placement.
Err Processing Error Restart scanner. If unit is defective, it will have to be returned. Contact
GE Technical Support.
Battery low
When battery charge for at least 10 NIBP measurements remains
Battery icon will change to red when low priority battery alarm is triggered.
The monitor continues to operate normally.
A low priority alarm is displayed.
An orange LED is displayed above the On/Off-button.
NOTE
It is strongly recommended that you plug the monitor into external DC
power when the low battery alarm is active.
When battery charge for 5 minutes or less of unit operation time remains
A red battery indicator is displayed on the screen and an orange LED above
the On/Off-button starts to blink slowly.
A high priority alarm message is displayed in the notification area.
The user is not able to initiate:
any new NIBP determinations of any type
any printouts
NOTE
Monitor cannot be restarted when battery charge is critically low. This
means that if a high priority Low battery alarm is active and monitor is shut
down, it cannot be restarted before the charger is plugged in. This is to
protect the battery pack.
6HUYLFHPRGH!
0DLQWHQDQFHEDWWHU\ 12
FKDUJHYROWDJHDQG
FXUUHQWRN"
<(6
%DWWHU\SK\VLFDOO\LQ 12
JRRGFRQGLWLRQ"
<(6
%DWWHU\FRQQHFWLRQRN" 5HSODFHWKHEDWWHU\
12
<(6
0HDVXUHGEDWWHU\YROWDJH 12
FRUUHFW"
<(6
5HDWWDFKEDWWHU\
FRQQHFWRU
)XVH)RQWKHPRWKHU 5HSODFH)IXVHHTXLYDOHQW
12
ERDUGRN" WR7$
<(6
0HDVXUHGPRWKHUERDUG
12 5HSODFHWKHFKDUJHUFDEOH
EDWWHU\YROWDJHFRUUHFW"
<(6
3UREOHPVROYHG
'RHVWKHSRZHUOHGWXUQ
RQZKHQWKHSRZHU 12
EXWWRQLVSUHVVHG"
0HDVXUHSRZHU
<(6 PHPEUDQH
3RZHUEXWWRQ
PDOIXQFWLRQLQJ"
%ODQNVFUHHQ"
<(6 <(6
5HDWWDFKWKHGLVSOD\ 12
FDEOHV
0RQLWRU2."
5HSODFHWKHKDQGOHVWULS
12 DVV\
0HDVXUH/&'
5HSODFHWKH
EDFNOLJKW
PRWKHU 12
YROWDJHV
ERDUG
9ROWDJHVRN"
<(6
5HSODFHWKHGLVSOD\FDEOH
0RQLWRU2."
12
<(6
5HSODFHWKHIURQWFRYHU
0RQLWRU2."
12
5HSODFHWKHPRWKHUERDUG
3UREOHPVROYHG
5HERRWDQGUHFDOLEUDWH
2."
12
7RXFKIOH[QRWFRQQHFWHGRU
FRQQHFWHGLPSURSHUO\"
12
0HDVXUHYROWDJHVIURPWKH
WRXFKFRQQHFWRURQWKH <(6
5HSODFHWKHPRWKHU
<(6 PRWKHUERDUGZKLOHWRXFKLQJ
ERDUG
WKHGLVSOD\
9DOXHV2."
12
5HSODFHWKHIURQW
FRYHU
3UREOHPVROYHG
:HOFK$OO\QRSHUDWHV2."
12127,167$//('
&KHFNFDEOHEHWZHHQ
WKHLVRODWLRQERDUGDQG
12
WKHPRWKHUERDUG
&DEOH2." <(6
0HDVXUHYROWDJHVRQ
<(6 WKHLVRODWLRQERDUG
5HSODFHWKH 9ROWDJHVRN"
LVRODWLRQERDUG
FDEOH
12
5HSODFHWKH6S2
PRGXOH
5HSODFHWKH
LVRODWLRQERDUG
3UREOHPVROYHG
:HOFK$OO\QFDEOH
6S2ZRUNV" <(6
FRQQHFWHG2."
12127,167$//('
&KHFNFDEOHEHWZHHQ
WKHLVRODWLRQERDUGDQG
12
WKHPRWKHUERDUG
&DEOH2." <(6
5HSODFHWKH
LVRODWLRQERDUG
FDEOH
0HDVXUHYROWDJHVRQ
<(6 WKHLVRODWLRQERDUG
9ROWDJHVRN"
12
5HSODFHWKH:HOFK
12
$OO\QPRGXOH
5HSODFHWKH
LVRODWLRQERDUG
3UREOHPVROYHG
3UHVVDQGKROG5HVHW
EXWWRQ
3UREOHPVROYHG"
12
8QSOXJPDLQVSRZHUUHPRYH
WKHEDWWHU\7KHQUHDWWDFK
EDWWHU\DQGSOXJLQWKHPDLQV
SRZHU6WDUWXSWKHPRQLWRU
$IWHUSUHYLRXVSURFHGXUHVFDQ
WKHPRQLWRUFDQEHVKXWGRZQ
SURSHUO\"
12
<(6
0HDVXUHWKHSRZHU
5HSODFHWKHPRWKHU
<(6 PHPEUDQH
ERDUG
3RZHUEXWWRQRN"
12
5HSODFHWKH
KDQGOHVWULS
3UREOHPVROYHG
Fuses
The monitor contains replaceable T3.5A blow-up fuses F3 on the motherboard
and F1 on the USB board. The fuses protect the low voltage DC input and the
battery. A FRU kit “VC150 FRU Cables” contains two T3.5A fuses. A FRU kit “VC150
FRU Fuses” contains 20 pieces of T3.5A fuses.
Motherboard
F3
USB board
F1
The battery
The battery packet contains irreplaceable overcurrent and temperature
protectors. A battery protection management protects battery cells from
excessive discharge.
Fuse replacement
1. Disconnect power cable.
2. Disconnect battery instructed in “Battery replacement” on page 7-19.
3. Use pliers to remove the fuse.
Board tests
NOTE
These are not routine operations and should be performed only when
troubleshooting the device.
Check the tables below for typical voltages. Use a digital multimeter (DMM) to
measure voltages. If measured voltage is different from the values in the tables,
refer to “Troubleshooting” on page 6-1 to figure out the problem. If there is no
reason provided, then the board is faulty and must be replaced.
Motherboard voltages
The CARESCAPE VC150 motherboard design includes several voltages and
power supply points. As all negative poles are connected to common ground
(GND), there are no negative voltages. The battery and/or battery charger act as
the main power supply for CARESCAPE VC150 and the source for all voltages on
the monitor. The battery voltage is lower when the battery charge level is
partially used up. The voltage is higher during charging.
Power Management IC D30 generates five voltages for the main processor, the
memory circuits (flash and RAM memories), part of the WLAN, and the audio
circuit power supply. The starting order of the voltages is very important for the
main processor.
Motherboard
Test point/connector
Area (top) Voltage
pin
Motherboard
Test point/connector
Area (top) Voltage
pin
TP36 +5V_Continuous
TP100 +3.3V_Continuous
GND
GND
Power membrane
Power membrane is connected to the motherboard.
Power membrane
Printer interface
Printer lines are mostly direct lines between a supplementary controller and the
printer unit. Line signals are digital and periodical pulses. Some voltages can be
checked with DMM.
Printer interface
NIBP interface
The NIBP connection has 2 serial buses (4 lines), boot enable and reset lines with
digital and periodical line signals. If you do not find signals at test points on the
NIBP board, the cable has broken or there is a bad connection.
Main processor of the motherboard provides power for the NIBP board. Default
state is enabled.
NIBP interface
TST7 GND
TST7
If you want to measure line resistance between test points on the motherboard
and NIBP board, perform the following:
Resistance
TP115 TST2
Isolation interface
Isolation interface connects the motherboard to the isolation board, which is
then connected to SpO2 and Welch Allyn temperature measurement modules.
The isolation board isolates communication and control lines between the
motherboard and measurement modules and generates isolated voltages for
modules. Connection between the isolation board and the motherboard consists
of serial buses for the SpO2 module and temperature module (4 lines), +5V and
+3.3V power supplies for the isolation board, SpO2 and temperature reset lines,
SpO2 and temperature power enable lines, temperature wakeup line and
ground.
The RX and TX lines are active only if there is traffic; check them with an
oscilloscope. You cannot measure most test points on the isolation board due to
the transparent isolation barrier that must not be removed. The barrier is needed
to prevent contact with the isolation area on the board. Measure isolated SpO2
voltages and data lines at test points you can access without removing the
isolation barrier.
The GND line connects ground on the motherboard with the primary ground on
isolation boards. The secondary GND on the isolation board provides ground for
SpO2 and temperature modules. They are isolated from the primary ground on
the isolation board.
These voltages also have isolated grounds. When measuring isolated voltages,
remember to connect negative pole of the test equipment probe to ground of
that voltage, not to the monitor ground.
Isolation interface
Isolation interface
TP2
TP15 TP3
TP16
TP21
TP4
TP30 TP14
TP22
The USB isolator circuit is powered from two sources. The PC-connector end is
powered by connected equipment. This voltage is in the PC USB connector pin 1.
If a device is connected to the PC USB connector you can measure voltage
between the connector's pins 1 and 4 (pin 4 is PC USB GND). The motherboard
end is powered by the motherboard.
TP10 TP11
TP8 TP12
TP5
USB board:
USB1: TP4 D+
USB1: TP3 D-
USB2: TP8 D+
USB2: TP7 D-
USB3: TP12 D+
USB3: TP11 D-
Reset line's test point Motherboard: TP131 +3.3Vdc (0V when reset
USB board: TP13. button is pressed)
Reset
F1
TP14
TP8
TP15 TP3
TP12 TP7
TP4
TP17
TP11
LCD backlight
A backlight connector on the motherboard has four signals: +12V (J12/1), LCD
ON (J12/2), Backlight PWM (J12/3) and GND (J12/4).
LCD interface
LCD connector J20 has 20 pins. There are 4 pairs of color signals (8 lines), one
pair for clock signal and two signals from the main processor I/O. Measure these
signals with an oscilloscope. Pins 1 and 2 of the connector are +3.3V and pins 7,
10, 13 and 16 are GND. If there are any display problems, first check the cable
between the motherboard and display. Change the cable first before changing
the display or the motherboard.
LDC interface
Test point/connector
Area (top) Voltage
pin
Description
This chapter provides instructions for disassembly and reassembly of the
monitor.
NOTE
GE supports troubleshooting and repair to the FRU level only. GE does not
make available schematics, assembly drawings, or bills of material beyond
what is provided in this manual. Attempting repair on a module or factory-
sealed component is not recommended.
FRU list
The following table offers details on each of the corresponding bubble numbers
that appear on the exploded assembly drawing (located after this table). Photos
of each FRU follow.
NOTES
If you need to replace a faulty motherboard or back cover, it can be done
only by GE Service Depot. Contact GE Service Depot.
If the customer wants to change the SpO2 technology, only GE Service can
do it. The SpO2 technology can be changed only if any of the SpO2
technologies exists.
NOTE
Contains 20 fuses.
NOTE
Contains 5 connector covers.
FRU photos
1.
3.
5. 2.
4.
6.
7.
8. 9.
10.
11.
12.
contains
CAUTIONS
Internal electronic components are susceptible to damage by
electrostatic discharge. To avoid damage when disassembling
the monitor, observe the standard precautions and procedures
for handling static-sensitive components.
NOTE
Update monitor software if you replace any of the OEM modules (NIBP
module, Welch Allyn module or any SpO2 module).
The following guidelines may not guarantee a 100% static-free workstation, but
can greatly reduce the potential for failure of any electronic assemblies being
serviced:
Discharge any static charge you may have built up before handling
assemblies containing semiconductors.
A grounded, antistatic wristband (3M part number 2046 or equivalent) or
heel strap should be worn at all times while handling or repairing
assemblies containing semiconductors.
Use properly grounded soldering and test equipment.
Use a static-free work surface (3M part number 8210 or equivalent) while
handling or working on assemblies containing semiconductors.
Torques
Main screws Metric screw thread M3mm, Torx T10 0.5 Nm 0.05 Nm [4.43 in-lb 0.44]
Bottom Plate screws Metric screw thread M3 mm, Allen key 0.5 Nm 0.05 Nm [4.43 in-lb 0.44]
2.5 mm
Handle screws, Board Frame screws, Self tapping PT 3.5 mm, Torx T10 1.05 Nm 0.1 Nm [9.29 in-lb 0.88]
Welch Allyn Frame, USB Frame,
Printer housing screws
NIBP module screw Self tapping PT 3.0 mm, Torx T10 0.28 Nm 0.03 Nm [2.48 in-lb
0.26]
Screw for Masimo metal spacer nut Metric screw thread M3 x 4 mm, Torx 0.5 Nm ± 0.05 Nm [4.43 in-lb ± 0.44]
T10
Masimo ground cable Metric screw thread M3 x 6 mm, Torx 0.5 Nm ± 0.05 Nm [4.43 in-lb ± 0.44]
T10
Welch Allyn module screws, Printer Self tapping PT 2.2 mm, Torx T6 0.21 Nm 0.03 Nm [1.86 in-lb
screw 0.26]
Speaker module screws Self tapping PT 1.6 mm, Torx Plus 6 IP 0.13 Nm 0.03 Nm [1.15 in-lb
0.26]
Battery replacement
Tools necessary:
DANGERS
ELECTRICAL SHOCK HAZARD — Before replacing the battery,
disconnect the monitor from the DC power supply. Failure to do
this may result in serious personal and/or device damage.
NOTE
Procure replacement batteries from your GE representative.
Disassembly
1. Unplug the monitor from the DC power source.
2. Place the monitor on a clean, steady surface, face down. Ensure that the
surface will not scratch the touch screen. Use a soft, lint-free cloth if
necessary.
3. Remove four screws with the 2.5 mm Allen key.
6. Push the locking pin down, take a firm grip on the connector and unplug the
battery. Pay attention to how the plug was positioned (locking pin facing up)
in the socket. This affects polarity.
Reassembly
1. Plug in a replacement battery. Ensure that the plug is positioned as in step 6.
The locking pin must be at the top of the plug. Plug position affects polarity.
WARNING
Prior to plugging in the replacement battery pack, perform a
visual check on it to ensure it is physically intact and without
leaks.
WARNING
When reconnecting the battery, ensure the battery maintains
the correct polarity.
2. Insert the replacement battery into the compartment. Ensure that the
battery cord is not squeezed or covering the screw hole.
3. Secure the bottom plate. Use the 2.5 mm Allen key. Apply 0.5 Nm ±0.05 Nm
[4.43 in-lb ± 0.44] torque.
NOTE
Plate side dimensions are not equal. If screws do not go easily, the plate may
be 90 degrees turned.
4. Insert the external DC power converter plug into the external DC power
socket and plug into an AC outlet. The new battery must be charged at least
for 8 hours.
5. Verify the configuration settings prior to returning the monitor to clinical use.
NOTE
Recycle the old battery according to local regulations on material recycling.
Do not dispose of in regular trash!
Monitor disassembly
Tools required:
4. Turn the monitor to an upright position and pull at edges on both sides of
the front panel.
WARNINGS
Be careful not to tip the front cover assembly over.
5. Hold monitor pieces together and turn the unit so that it rests on the table
with the touch screen facing up.
6. Open the front cover assembly. Tilt the front cover as pictured below to
avoid damage. Do not strain the display flex cable (arrowed in the image).
WARNING
Display flex cable connector is a sensitive part. Do not bend the
flex cable in vain. Never pull the flex cable. Always get a hold on
the connector when unplugging the flex cable.
1
2
3. Touch flex cable. Slide the connector latch out about for 2 mm. Then
release the touch flex cable.
NOTE
Contact surfaces of the power membrane FFC must face the mother board.
9. Release the snap on the right and lift up the board a bit. Then release the left
snap.
11. If the monitor is equipped with the printer, turn the monitor in an upright
position. Hold the motherboard with your right hand fingers. Release the
printer flex cable with left thumb and index finger.
NIBP module
12. Release the NIBP module screw. Use Torx T10 screwdriver.
14. Gently release another snap feature with a screwdriver. Avoid causing
damage to NIBP board or its components.
15. Remove the manifold hose from a pressure sensor. Remove the valve hose
from another pressure sensor.
16. The opaque plastic NIBP manifold is quite tight and must not be pulled
directly from the NIBP board. If pulled directly from the board, the deflation
valve may sustain serious damage. The manifold must be loosened first.
Place a screwdriver between the manifold and blue colored ends of the NIBP
deflation valve. Then gently pull the manifold off the NIBP board.
18. Remove the pump hose (A). Pay attention to the order the hoses are
mounted onto the manifold.
19. Remove the manifold hose (B) and the valve hose (C) from the NIBP hose
connector inlets.
C
A
B
C
A
23. Push the back cover backwards (dotted arrow) and middle frame with
handle bar towards you (solid arrow).
24. Pull the board and printer door off the back cover. Ensure that the printer
cover passes through the opening smoothly. Watch out for cables.
NOTE
Remove the printer or dummy printer cable from the USB board connector,
25. Remove printer assembly screws in the following order and remove the
printer assembly. Use a Torx T10 screwdriver.
3
2
31. After removing the dump valve, guide the valve and pump connector
through the opening in the board frame.
32. Rotate the assembly and loosen the lock latches a bit to remove the NIBP
pump clamp. Arrows indicate release direction.
33. Release the NIBP clamp. Remove the pump assembly from the recess. The
pump hose assembly is now free.
Side strips
34. Open the left slide strip snap feature with a screwdriver.
36. Open the right slide strip snap feature with a screwdriver.
37. Remove the right strip assembly. The USB FFC slides through an opening in
the board frame.
Handle
39. Hold the handle and remove the screws at the top. Use a Torx T10
screwdriver.
41. Remove the drain slope insulation tapes and the drain hose. Pull out the
drain slope assembly.
Ventilation plates
42. Pull out the two ventilation plates. There is one on each side.
USB frame
43. Remove the USB frame screws (2). Use a Torx T10 screwdriver. Then remove
the USB frame assembly.
Remote alarm
44. Remove the nurse call (remote alarm) cable 4J00199X. Release the HEX nut
with the socket wrench and 10 mm socket with a slot for a cable. To avoid
twisting the cable around and damaging it, turn the socket for 1/4 round
clockwise. Pull the socket off the HEX nut, rotate 1/4 round counter-
clockwise then place on the HEX nut again. Repeat until the HEX nut is
released. Rotate the nut with your fingers to remove it.
Alternative method
You can also use a standard 10 mm socket, guide the nurse call cable through
the socket and loosen the hex nut without using a socket wrench.
45. Pull the cable out through a hole in the back cover frame.
49. If Masimo SpO2 module is in use, remove ground cable from the isolation
board. Use a Torx T10 screwdriver.
50. Release the connector latch and remove the Welch Allyn FFC.
WARNING
Be very careful and use gloves when you are handling the
isolation board. The isolation board assembly is a hi-pot tested
critical-to-safety component. The assembly must be free of
grease and dust. Do not touch or remove the transparent part
with your finger or anything else that might affect it.
52. If SpO2 module replacement is needed, release the 4 snaps to remove SpO2
connector housing. Do not break the snaps.
53. Pull the SpO2 unit carefully out through the opening. Note the orientation
and be careful to avoid damage.
55. Release the Welch Allyn frame snaps (2). Pull out the Welch Allyn module.
56. Remove the Welch Allyn module FFC carefully through the back cover hole.
Avoid damage.
USB door
57. Turn the back cover around and release the thumb screw on the USB door.
58. Push the USB door as indicated by an arrow. At the same time press right
side hinge in the direction indicated by the arrow until the hinge pin comes
out of the back cover recess. Now the disassembly is complete.
Monitor reassembly
Tools required
A screwdriver with adjustable torque or an electronic screwdriver with
adjustable torque.
A screw torque calibrator for the electronic screwdriver, if the electronic
screwdriver is used.
A wrench for the Torx T10 screws
A wrench for the Torx Plus 6 IP screws
A medium sized slotted screwdriver
10 mm socket (for nurse call cable)
A socket wrench
Pliers
Wire cutters
Isolation tape
Required parts
Required new field replacement units (FRU kits)
USB door
1. Replace the USB door.
NOTE
Pay attention to the orientation of the NIBP hose connector. The pin gate
recess of the connector must be facing upwards.
Alternative method
You can also use a standard 10 mm socket, guide the nurse call cable through
the socket and secure the hex nut without using a socket wrench.
5. Insert the hose into the opening at the bottom of the frame.
NOTE
The label has to cover all unused holes. The label is a single-use item. Once it
is attached on the cover plate, you cannot change the positioning.
7. If Welch Allyn module is being used, carefully insert the Welch Allyn FFC
module cable (A) and align the module with a stand-off hole (B) and grip hole
(C).
8. Check that the cable passes through the opening smoothly and that the
Welch Allyn module is aligned with the stand-off hole and the snaps are
locked.
9. Then secure the module. Use self tapping PT 3.5 mm, Torx T10 screws. Apply
1.05 Nm ± 0.1 Nm [9.29 in-lb ± 0.88] torque.
USB frame
10. Replace the USB frame. Secure the frame with 2 PT 3.5 Torx T10 screws.
Apply 1.05 Nm ± 0.1 Nm [9.29 in-lb ± 0.88] torque.
SpO2 Board
Each SpO2 technology has its own board and connector cable combination that
must always be used.
WARNING
Never mix different technology boards and connectors! This
will result in erroneous operation of the SpO2 measurement.
Masimo board is secured onto isolation board with 3 stand-offs, two M3 screws
(Torx IP 10) and a spacer nut. GE TruSignal and Nellcor boards are secured onto
isolation board with 2 stand-offs.
11. Insert stand-offs on the board you are going to install. The stand-offs for
GE TruSignal or Nellcor boards are installed the same way.
12. If the Masimo SpO2 module is in use, insert three standoffs and secure a
metal spacer nut with a M3 x 4 mm screw. Use a Torx T10 screwdriver. Apply
0.5 Nm ± 0.05 Nm [4.43 in-lb ± 0.44] torque.
NOTE
Make sure the spacer nut is oriented as in the picture below.
13. Check that the stand-offs and the connector of the SpO2 module aligns with
the pins and connector of the correct technology marked on the isolation
board. Stand-offs are not reusable.
14. Each SpO2 technology has its own connector cable. Insert the correct
connector cable through the opening. Note connector orientation! Tilt the
large blue connector a bit so that you guide the wider section of the cable
through the opening first. Then tilt the connector back and insert normally.
15. Guide the smaller connector of the flex cable behind the drain hose.
CAUTION
It is dangerous to use the wrong SpO2 sensor. The sticker must
indicate what SpO2 technology is used.
17. Connect the flex cable to the SpO2 board. Ensure that the SpO2 flex
connector aligns with the correct pins and the correct row.
18. If the Masimo SpO2 module is in use, connect the ground cable of the SpO2
flex to the isolation board as in picture. Use M3 x 6 mm screw and Torx T10
screwdriver. Apply 0.5 Nm ± 0.05 Nm [4.43 in-lb ± 0.44] torque
Isolation board
19. Plug the Welch Allyn FFC into the isolation board connector. Slide the
connector latch in to secure the plug. The FFC contact surface must face the
board surface.
20. Align the isolation board with the frame stand-offs and replace the board.
Ensure that all 4 stand-offs lock up. Verify that the SpO2 flex connector is
properly attached.
Ventilation plates
21. Slide ventilation plates in place.
Handle
22. Replace the handle. Secure it with two self tapping PT 3.5 mm, Torx T10
screws and apply 1.05 Nm ± 0.1 Nm [9.29 in-lb ± 0.88] torque.
Side strips
23. Push in 3 snap features to install the right side strip.
24. Guide the Medical USB FFC through the opening as in the image below. Then
push the left side strip assembly in place. Three snap features lock it in
place.
25. Attach the 2 shorter handle gaskets on the back side of frame. Refer to
“Board frame assembly exploded view” on page 8-6 for more details.
26. Attach the 2 longer handle gaskets on the front side of the frame.
28. Place the pump in the correct position in the recess. Position the hose, filter
and back pressure valve carefully. NIBP does not work if any of the hoses is
bent or kinked.
29. Secure the pump clamp with 3 snap features. Ensure the hose is not kinked
and the assembly is secured. Guide the pump hose through the opening in
the board frame.
30. Insert the NIBP valve. Ensure the snap features secure it.
31. Replace the speaker module. Secure with 4 self tapping PT 1.6 mm, Torx Plus
6 IP screws. Apply 0.13 Nm 0.03 Nm [1.15 in-lb 0.26] torque.
NOTE
For correct printer alignment, it is very important to follow the correct order
when securing the screws.
33. Insert the printer flex through the opening on the board frame.
NOTE
Do not guide the printer/dummy printer ground cable through the board
frame.
34. Secure the printer assembly in correct order as numbered in the image
below. Use a Torx T10 screwdriver. Apply 1.05 Nm ± 0.1 Nm [9.29 in-lb ±
0.88] torque.
1
2
Main assembly
36. Plug ground cable of the printer/dummy printer to USB board connector.
37. Leave the printer door open and guide it through the opening in the back
cover assembly.
39. Connect the handle assembly and back cover assembly. Check that cables
are not tangled up or squeezed. Close the printer door.
40. Secure 2 board frame screws with a Torx T10 screwdriver. Apply 1.05 Nm ±
0.1 Nm [9.29 in-lb ± 0.88] torque.
NOTE
If these two screws are released and secured many times and excessive
torque has been used, it may wear out screw thread in the plastic screw
housing. If you cannot secure these screws, you may leave these out. These
two screws are not critical to product safety or integrity.
42. Guide valve hose as in the image below. Check that it aligns smoothly on all
turns. Correct place for hose intersection is important so that the hose will
not be blocked at later stages of reassembly.
43. Connect the valve hose to the upper inlet of the NIBP connector. The correct
place for hose intersection is important so that the hose will not be blocked
at later stages of reassembly.
NOTE
NIBP hose should not be twisted when attached. Twisted hose may result in
bent hose and blockage.
44. Connect the manifold hose to the lower inlet of the NIBP connector.
NOTE
NIBP hose should not be twisted when attached. Twisted hose may result in
bent hose and blockage.
45. Align the 3 small pins on the deflation valve with the 3 manifold holes (on
tube side of manifold). Connect the manifold hose to the deflation valve.
46. Connect the pump hose (A) to the manifold inlet next to the manifold hose
inlet. Connect the valve hose (B) to the pressure sensor.
47. Connect the NIBP pump cable. Secure the manifold with NIBP board screw.
Use a self tapping PT 3.0 mm and a Torx T10 screwdriver. Apply 0.28 Nm ±
0.03 Nm [2.48 in-lb ± 0.26] torque. Then check that hoses are not squeezed
or kinked.
NOTE
NIBP will not work if any of the hoses is kinked or squeezed.
48. Align the NIBP with board frame guiding pins. Press the right side snap into
place first. Then press the top side snap into place. Then look underneath
the NIBP module and check that the hoses are not squeezed or kinked. NIBP
will not work if any of the hoses is kinked or squeezed.
Motherboard
49. Turn the monitor to the upright position. Hold the motherboard with your
right hand fingers. Ensure that the printer flex connector is tight. Take a firm
grip on the flex and push the contacts into the connector (the silver colored
contacts should not remain visible once the flex is connected).
CAUTION
Be careful not to bend the flex cable. A bent section may break
the flex cable.
50. Connect the power membrane FFC to the power button connector. Contact
surfaces of the power membrane FFC must face the mother board. Close
the latch.
51. Connect the USB FFC to the connector on the motherboard. Close the latch.
Check that FFC is connected correctly.
52. Connect isolation board cable 4J00198x to the isolation connector on the
motherboard. Note that the connector must be oriented as in the picture
below!
53. The black stripe on the cable must face down.
54. Connect charge power cable 4J00200x to the battery connector on the
motherboard.
55. Connect the main power cable 4J00197x to the board charger connector on
the motherboard.
56. Connect the nurse call cable 4J00199x to the nurse call connector on the
motherboard.
57. Connect the USB FFC to the USB board connector on the motherboard.
Close the latch. Check that FFC is secured correctly.
58. Connect the speaker module cable to the speaker connector on the
motherboard.
59. Press the motherboard towards the frame until it locks up. Note that the
power membrane should be aligned as in the image below.
60. Connect the NIBP module cable 4J00196x. Check that the connectors are at
the correct pins and row.
61. Connect the 4J00195x display power cable onto the backlight connector on
the motherboard.
NOTE
The display flex connector cannot withstand rough handling. Handle with
care.
63. Connect the touch flex cable. Ensure that the flex contact surface faces the
board surface. Slide the connector latch in to secure the cable.
64. Press the monitor frames to lock the front cover assembly with snap
features. Close the printer door.
65. Secure the front cover assembly with 2 screws at the top and 2 screws at
the bottom. Use Torx T10 screws. Apply 0.5 Nm ± 0.05 Nm [4.43 in-lb ± 0.44]
torque.
67. Refer to “FRU replacement checklist” on page 7-2 for FRU replacement
checkout procedures.
NOTE
If the Exergen scanner is not used regularly, remove the battery to prevent
possible damage due to chemical leakage.
Procedure
1. Unplug the scanner cable from the monitor USB port.
2. To replace the 9V battery, use a bent paper clip or similar tool and insert it
directly into the battery cover ejection pinhole on the side of the
thermometer (A).
3. Apply pressure to spring release the battery cover (B).
4. Remove the cover.
E
A
C
B
Verification
Check the LED display. The low battery error message should not be displayed.
Perform a temperature measurement to determine whether the Exergen unit is
powered correctly. Refer to “Disposal of product waste” on page 9-22 for battery
disposal instructions.
NOTE
SpO2 technology can be changed only if any of the SpO2 technologies exist.
5 Motherboard 1
6 NIBP module 1
9 Battery pack 1
Numbers in this table refer to the main assembly exploded view images. Refer to
“FRU kits” in CARESCAPE VC150 Supplies and accessories (KO00125) for
availability.
10
13
1 8 6 7 5 4
9
11
8-3
8-4
Note tubes in correct order.
- Pump tube to be lowest.
- Valve hose in the middle.
- Valve hose NIBP end tube is at top.
Guide all back cover cables through this opening:
- Isolation Cable
- Medical USB FFC
Connect the pump and valve - Nurse Call Cable
connectors to the NIBP board. - Charge Power Cable
Ensure the printer door is opened when the Board
frame is being assembled on back cover.
Use 10 IP torx. Plug the Valve Hose tube into NIBP board sensor.
Use 0.28 Nm +/- 0.03 Nm Attach the tubes to the slots.
torque for screw assembly.
KO00094Q
KO00094Q
Read the plug code on a cable and relative connector code on the main board.
Plug all Back Cover cables into connectors on mother board. Connection order:
1. Printer flex to board connector
2. Isolation Cable
3. Medical USB FFC
4. Nurse Call Cable
5. Charge Power Cable 7
6. Main Power Cable
10
Plug the power membrane
flex into this connector.
Observe the orientation
Connectors are on PWB side.
Use 0.5 Nm ± 0.05 Nm
(4.43 in-lb ± 0.44)
Printer connector torque for screw assembly.
Use Torx 10 IP.
8-5
Exploded view drawings: Main assembly exploded view
Find
Description Quantity
number
1 Board frame 1
3 Handle 1
8 Pump clamp 1
Numbers in this table refer to the board frame assembly exploded view images.
Refer to “FRU kits” in CARESCAPE VC150 Supplies and accessories (KO00125) for
availability.
Guide and insert the charger power cable connector into the opening.
11
13
10
12
8
1
15
16
3
9
2
6
14
13
7
KO00094Q
KO00094Q
Secure the printer housing alignment by screwing in correect order.
Torque 1.05 Nm ± 0.1 Nm (9.29 in-lb ± 0.88).
0.1 ± 0.3
0.1 ± 0.3
3rd screw
0.1 ± 0.3
0.1 ± 0.3
2nd screw
Attach the clamp housing over the pump Attach the printer door
Ensure that Printer flex is in the opening and
valve and filter. Secure with 3 snaps. with hinge pins.
visible to the front side.
8-9
Exploded view drawings: Main assembly exploded view
Find
Description Quantity
number
1 Back cover 1
2 Drain slope 1
15 Foot 4
19 Printer gasket 1
22 NIBP label 1
23 SpO2 label 1
Numbers in this table refer to the back cover assembly exploded view images.
Refer to “FRU kits” in CARESCAPE VC150 Supplies and accessories (KO00125) for
availability.
17
2
19
25
18
8
6
9
13
14
23
22 12
7 3
24
26
1
25 4
27
8-11
8-12
19 20
B
and 2:1
sket m.
n t h e ga d 0.8 m
e
wee exce
bet not
s t ance must
Di edge
er
cov Tape label inside the polished area.
the
A Observe orientation.
2:1 Text must be readable at this location.
A
A
B A-A B-B
Tape label inside the polished area.
Observe orientation.
Plug the V150 GND Cable on Text must be readable at this location.
USB Board tap connector.
KO00094Q
KO00094Q
Dummy SPO2 (when no SPO2).
8-13
Exploded view drawings: Main assembly exploded view
Introduction
This chapter provides information on various maintenance procedures to help
keep the monitor uptime as high as possible. Most of the instructions apply to
the daily caregiver, but Service can also perform these procedures to remove
stains or other dirt incurred on the monitor or monitor parts during maintenance
work.
WARNINGS
Failure on the part of the responsible individual, hospital, or
institution using this equipment to implement a satisfactory
maintenance schedule may cause undue equipment failure
and possible health hazards.
NOTE
GE does not, in any manner, assume the responsibility for performing the
recommended maintenance schedule, unless an Equipment Maintenance
Agreement exists. The sole responsibility rests with the individuals, hospitals,
or institutions utilizing the device.
Customer
Service
NOTE
The manufacturer has
performed these electrical
safety tests for patient monitor
and acquisition modules
during final inspection. You do
not have to perform the
electrical safety tests during
installation checkout, if the
patient monitor was
manufactured less than 24
months ago. Check the serial
number label for
manufacturing week and year
(refer to “Equipment ID” on
page 1-6).
Customer
Service
Printer output test. Refer to “Printer Every time the monitor has
output test” on page 5-17. been partly or completely
disassembled.
Long term storage. Refer to “Extended Every time the monitor is stored for
storage” on page 9-20. an extended period of time.
Customer
Service
Date
Inspection item Notes Schedule
performed
Any signs of physical damage to Do not use the monitor if damage is Every time the
the monitor case determined. Replace any damaged monitor is under
the display screen parts of the unit. service,
the power membrane button
Dirt or stains on the monitor or the Advise the customer about the need
accessories. to clean the monitor regularly.
2. Plug in USB-B cable to USB-B connector on the monitor and to a regular USB
connector on a service PC.
3. The monitor will appear as a USB mass storage device CARESCAPE VC150
on the service PC.
WARNING
The monitor shall not be used for clinical measurements during
remote service session.
NOTES
Refresh the browser screen in case the upload page does not become
visible.
For security reasons, the remote session will automatically terminate after 5
minutes of inactivity, except when software update or file transfer is in
progress. Make sure that you have pressed the Commit changes button
after uploading the settings.
1. If Web Server has not been setup before, enter Monitor Setup > Service
Mode > Connectivity > Web Server and set the desired Port number, and
Secure as Enabled or Disabled. If Secure is set as Enabled, go to Service
Mode > Connectivity > Device Certificate to set up parameters for a secure
web server connection.
NOTES
Turning security disabled is insecure and should be used only in secure
network environment.
The monitor can be accessed remotely if both the monitor and the service
PC are connected to the same network. Firefox version 29.0 or newer should
be used for the remote service.
2. Select Start Remote Service in Monitor Setup > Advanced. If the IP address
is not shown on the Remote service active window, the Web server settings
are incorrect.
3. Launch a web browser and enter the monitor IP address in the address field.
Then press the Enter key on the PC keyboard.
4. Enter the service mode password and select Login. When a remote
connection has been established correctly, the CARESCAPE VC150 Remote
Service Mode screen appears.
5. Monitor Information is used to check for serial number and other monitor-
specific items.
6. Software is used to check for the software version. If you need to update the
software, select Start software update process, browse the service PC for
the software update package and select the package to update the
software.
NOTES
Do not transfer different parameter and alarm default settings without
clinician's consent. When alarm default settings are changed, make sure
that the clinicians are aware of the new settings.
Refer to “Log file samples” on page 6-4 for log file samples.
10. Preventive maintenance (PM) displays the date of the last preventive
maintenance and has options to set how often the PM is performed and
when the monitor starts to remind about the upcoming PM.
11. When all remote service work is complete, press Log out from the top-left of
the screen.
NOTE
The browser should be set up so that it will not store browsing history after
you close the browser. Refer to browser help for instructions.
vc150-settings-********.vc150.enc
file (asterisks denote a date when the settings were exported) that can be copied
to service PC. This file can be imported to the monitor and used to restore
settings, if necessary. The same file can also be imported to another
CARESCAPE VC150 monitor (even if it uses another software version), but some
monitor-specific settings are not copied to another CARESCAPE VC150 monitor.
In case there are multiple *.vc150.enc files in incoming settings transfer storage,
the first file will be selected (sorted by name, in ascending order).
It is possible to import older vc150-setting files to the monitor. New setting files
are encrypted (.enc)
NOTES
All clinical settings are backed up and restored.
File import
The following kinds of files can be imported to the CARESCAPE VC150 monitor:
Settings
Licenses
EWS setting files
Private keys (connectivity related)
Certificates (connectivity related)
The format for the imported file name must adhere to the following
requirements:
UTF-8
characters a-z, underline ’_’ and dash ‘-’ can be used
no special characters in file names, nor accents nor punctuation
maximum of 32 characters
ASCII 7-bit (for license keys)
When the monitor is connected to a service PC, it will show up as a USB
mass storage device with CARESCAPE VC150 as part of the drive name. The
drive partition that appears when the monitor is used as a USB mass
storage device has FAT file system. The monitor has no FAT nor NTFS or
operating system requirements for service PC as long as the system
recognizes a USB mass storage.
Exporting settings
1. Connect to the service PC as instructed in “Connecting to the Service PC” on
page 9-6.
2. Select Export All Settings with USB to export all settings from the monitor to
the PC or Export EWS Settings with USB to export EWS settings from the
monitor to the PC.
3. Disconnect as instructed in “Disconnecting from the service PC” on page 9-
7.
Importing settings
1. Refer to “What can be copied” on page 9-12 to determine what you can
import to a monitor with different serial number.
NOTE
Even though you may import settings with a file, you may need to perform
some additional configuration. Perform all steps in “Monitor configuration
checklist” on page 4-2 to ensure the configuration is complete.
NOTE
All settings file will include EWS settings, if these were configured in the
original devices before exporting the settings. If you want to import EWS
settings separately e.g in case the monitor will have different EWS settings
than the original, import first all settings and then EWS settings. New EWS
Settings will overwrite the previous ones, but other configuration settings
remain the same.
Cleaning
List of approved cleaning agents
Use only these cleaning agents to clean the monitor unit or the Welch Allyn
probe well. Periodically, GE evaluates additional cleaning agents for
compatibility with the monitor. If your cleaning agent is not listed in this section,
contact your GE representative to determine if additional information is
available.
Isopropyl alcohol,
max 55% (vol)
NOTES
The touch screen surface is made of PET (polyethylene terephthalate) and
must not be cleaned with any solvent or any alkaline cleaning agent.
If you want to use other cleaning agents than those specifically approved
and listed in this section, you can do it at your own risk. If you still want to
use a non-approved cleaning agent, first test it on the corner of the touch
screen, not on the whole screen. Remember to wipe off possible residues or
stains with a cloth dampened with water.
NOTE
Never autoclave or steam clean the monitor, cuffs, or accessories.
Cleaning schedule
To prevent cross-contamination, clean exterior surfaces of the monitor, monitor
accessories, and reusable sensors on a regular basis in compliance with your
institution's infection control unit and/or biomedical department's local policy.
Procedure
WARNINGS
Always shut down the monitor and disconnect it from the electrical
network before any cleaning procedure.
Prior to using a cleaning agent, read the instructions for use and adhere
to provided safety precautions.
NOTE
Clean the monitor and sensors according to local policies.
NOTE
Exergen temporal scanner only: Alcohol-based cleaning agents can be used
on the scanner’s lens, probe head and metal neck only.
General
Before reuse, the cuff and the air hose must be thoroughly cleaned with a
cleaning agent listed in “List of approved cleaning agents” on page -15.
WARNINGS
Consult the manufacturer for specific cleaning or disinfection
instructions of the cuffs and air hoses. The user has the responsibility to
validate any deviations from the recommended method of cleaning and
disinfection.
1. Visually inspect the thermometer probe for any physical damage that might
cause future product failure.
2. Clean the probe and the probe well according to instructions below. Use an
alcohol-based cleaning agent on the Welch Allyn metal probe or plastic
section of the probe only.
WARNING
Do not immerse or soak the probe in any type of fluid. Do not use steam,
heat or gas sterilization on the probe. Do not autoclave the probe.
CAUTION
Never immerse any monitor accessories.
WARNING
Do not use hard or sharp objects to clean the probe well. This could
damage it and cause the unit to not function properly. Do not use
steam, heat or gas sterilization on the thermometer or probe. Do not
autoclave the probe well.
9. Periodically clean the probe’s surface by wiping it with a soft cloth, sponge
or soft brush dampened with a cleaning agent listed above for Welch Allyn.
Then wipe dry with a clean cloth or towel.
Short-term storage
The monitor and its accessories shall be stored in a clean and stable place with
no weight applied on them.
Keep the monitor connected to an external DC power source when not in use to
ensure maximum battery charge. As long as the monitor remains connected to
an external DC power source, the monitor will charge the battery whenever
software determines it necessary.
Batteries should always be connected and fully charged before being placed in
short-term storage. It is recommended that batteries should not be left in
storage more than 2 weeks without a full recharge. When the battery will no
longer hold a charge, it must be replaced.
NOTE
If the battery is replaced, a battery discharged, or a patient discharged, the
monitor will clear user settings and revert to the default settings set in the
configuration mode.
Extended storage
If it becomes necessary to store the monitor for an extended period of time,
remove all attached accessories.
NOTE
When storing the product for extended periods, the battery must be
disconnected.
When storing the monitor for extended periods, perform the following:
1. Check that the battery charge level is between 35% - 100%. If the below
35%, then charge the battery until the charge level 35% or higher.
2. Disconnect the battery. Store the battery inside the monitor.
3. Use original packing material to pack up the monitor and the batteries for
extended storage. If the original packing material has been lost, refer to
“Unpacking the unit” on page 3-2.
WARNING
When the monitor and the battery has been stored in under
room temperature, there is a risk for moisture condensation
and electric shock when the monitor is brought to room
temperature.
NOTE
If the battery is replaced, a battery discharged, or a patient discharged, the
monitor will clear user settings and revert to the default settings set in the
configuration mode.
Storage temperature
When packaged according to instructions in the Service Manual, the device
survives a storage temperature range of – 20° C to + 50° C (– 4° F to + 122° F).
Packing
Minimum requirements
Original CARESCAPE VC150 shipping box and material, if available
Packing tape
A marker pen
Procedure
1. Remove all hoses, cables, sensors, and power cords from the monitor before
packing.
2. Clean the device and parts as instructed in “Cleaning” on page 9-15.
3. Check the that product codes on the main label on the monitor and the
shipping box match. If these do not match, cross out the code on the
shipping box.
4. Disconnect the battery and replace it in the monitor as instructed in “Battery
replacement” on page 7-19.
5. Secure USB and power connector hatch.
6. Pack the monitor in a bag and insert the bag and the product inside the
monitor box. Use the original shipping carton and packing materials, if
available. If not, refer to “Missing original packing material” on page 9-22.
7. Secure the shipping box with packing tape.
8. If accessories must be sent or stored with the monitor, add these in a
separate plastic bag and add this bag into the larger box that already has
the monitor box inside.
9. Verify that the printed arrows on both boxes point upwards, to the same
direction.
10. Observe the environment conditions detailed in the “Product specifications”
on page A-2.
11. It is recommended that all returned goods be insured. Claims for loss or
damage to the product must be initiated by the sender.
1. Place the product in a plastic bag and tie or tape the bag to prevent loose
particles or materials from entering openings such as hose ports.
2. Use a sturdy corrugated container to ship the product; tape securely to seal
the container for shipping.
3. Pack with at least 10 cm (4 inches) of padding on all sides of the product.
Batteries
The Li-Ion rechargeable battery can be recycled. Do not puncture or place the
battery in a trash compactor. Do not incinerate the battery or expose it to fire or
high temperatures. Dispose of these materials according to local or national
regulations.
Introduction
This chapter provides instructions on how to upgrade CARESCAPE VC150
monitor software and recover the monitor software if necessary.
Software upgrade
CARESCAPE VC150 software can be upgraded with USB connection or hospital
WLAN. Software upgrade from version 1.6.21 (or older) to version 1.7.x can only
be done with USB cable.
WARNINGS
Connect the external DC power supply of the monitor to mains
before starting the software upgrade. Keep the monitor
connected to DC power throughout the software upgrade.
Upgrade procedure
1. When the files are sent and settings backed up, select Start button in the
Monitor Setup > Advanced > Service Mode > Maintenance/Software
Update.
7. After the USB cable has been removed, the monitor will continue with the
upgrade. Wait until the upgrade is complete and the monitor reboots
automatically. If the MSP software was updated during the software
upgrade, the monitor will not restart automatically. It must be restarted
manually.
NOTES
If you update the software from version 1.6.21 (or older) to version 1.7.x, you
need to do it twice. Follow instructions on the monitor screen.
After MSP upgrade, restart the monitor manually when the update screen
with white background changes to dark blank screen.
NOTE
The monitor may reboot for second software update and display a message
Refurbishing internal storage. The clinical view will appear normally after a
while.
Verification
If the software starts up as usual and no error messages appear on screen, the
software update was successful and the monitor can be used.
1. Check the software version at Monitor Setup > Advanced > Service Mode >
Monitor Information > Software version.
2. After the software update, proceed to “Software update completed” on
page 10-5.
Required tools
Software package
A personal computer connected to hospital WLAN
A CARESCAPE VC150 monitor connected to hospital WLAN
Procedure
1. Launch Remote Service Mode as instructed in “Remote service mode” on
page 9-7 and select Start software update process.
NOTE
If the upload screen does not appear automatically after selecting Start
software process, select Refresh button on the web browser.
2. Follow instructions.
3. When the upgrade-file (.bin) has been uploaded, the monitor will continue
with the upgrade. Wait until the upgrade is complete and the monitor
reboots automatically. If the MSP software was updated during the software
upgrade, the monitor will not restart automatically. It must be restarted
manually.
NOTE
After MSP upgrade the monitor will shut down and behave unpredictably.
This means that the alarm light and the LCD back light might remain lit.
Restart the monitor manually when the update screen with white
background changes to dark blank screen.
Verification
If the software starts up as usual and no error messages appear on screen, the
software update was successful and the monitor can be used.
1. Check the software version at Monitor Setup > Advanced > Monitor
Information > Software version.
2. After the software update, proceed to “Software update completed” on
page 10-5.
1. Enter Monitor setup > Advanced and verify that the correct software
version is displayed in Software version field.
2. Perform Parameter-level functional testing as instructed in “Parameter-level
functional testing” on page 5-2.
3. Configure the monitor. Refer to “Monitor configuration” on page 4-1.
4. Back up new configuration settings. Refer to “Backing up and importing/
restoring settings” on page 9-12.
NOTE
Depending on Configurator Tool selections, importing EWS settings may
override some monitor settings.
Import procedure
1. When the files are sent and settings backed up, select Start button in the
Monitor Setup > Advanced > Service Mode > Settings and Logs.
2. Select Import EWS Settings with USB to start the import process.
3. Connect USB A-B cable as instructed on screen.
NOTE
If you need to import EWS settings separately, it is important to import these
in correct order: 1) all settings 2) EWS settings.
Product specifications
Mechanical
Dimensions
Power requirements
Monitor
Protection against electrical shock Internally powered or Class I when powered from specified external
medical power supply.
Fuses The monitor contains two T3.5A replaceable fuses on the motherboard
(F3) and USB board (F1). The battery packet contains overcurrent and
temperature protection. The fuses protect the low voltage DC input and
the battery.
Environmental
Storage/transport
Printer specifications
Paper type The paper roll used by the printer must be compatible with GE PN 770137.
NIBP specifications
Blood pressure accuracy (Auscultatory) Complies with ISO 81060-2 standard Criterion 1 (mean error 5 mmHg,
standard deviation 8 mmHg).
Blood pressure accuracy (SuperSTAT) Complies with ISO 81060-2 standard Criterion 1 (mean error 5 mmHg,
SuperSTAT algorithm is currently not standard deviation 8 mmHg).
available in Canada. NIBP determination
for neonatal population cannot be done.
BP range (Auscultatory)
BP range (SuperSTAT) SuperSTAT algorithm is not currently available in Canada. NIBP determination for neonatal
population cannot be done without it.
NOTES
NIBP performance and accuracy have only been confirmed using GE hoses and cuffs.
The SuperSTAT algorithm is used for all neonate NIBP determinations. SuperSTAT algorithm is not currently
available in Canada. NIBP determination for neonatal population cannot be done without it.
Measurement range
SpO2 0 to 100%
Saturation Accuracy
Accuracy, Arms (root mean square of paired values; previously represented by ± 1 SD)
Measurement range
*Test methods used to establish SpO2 accuracy: GE Trusignal Technology with OxyTip+/GE TruSignal sensors have
been validated for no motion and motion accuracy in controlled hypoxia studies with healthy non-smoking adult
volunteers over the specified SpO2 range. SpO2 readings were compared to SaO2 values of drawn blood samples
measured by CO-oximetry. Subjects comprised both adult men and women and spanned a range of skin
pigmentation. GE TruSignal technology with OxyTip+/GE TruSignal sensors have been validated for low perfusion SpO2
accuracy over the specified range in a bench top testing against BioTek Index 2 patient simulator with 0.3% signal
amplitude.
**GE TruSignal Technology with GE OXY-AF (equivalent to TS-AF) and GE OXY-SE (equivalent to TS-SE) sensors have
been clinically validated for neonatal accuracy. Test subjects included 28 neonates and 1 infant (15 females and 14
males) with ages ranging from newborn to 37 days, weights from 560g to 3060g and skin tones from light to dark.
Accuracy (Arms) of the collected convenience samples was 2.7 for the OXY-AF sensor (52 data points were collected in
the range of 87%-100%) and 2.7 for the OXY-SE sensor (53 data points were collected in the range of 81%-100%).
***GE TruSignal Technology with GE Oxy-AF (equivalent to TS-AF) and GE Oxy-AP (equivalent to TS-AP) sensors have
been validated for motion accuracy in controlled hypoxia studies with healthy non-smoking adult volunteers over the
specified saturation SpO2 range(s). The following motion types were used: mechanically induced 3 Hz tapping motion
at an amplitude of 1-2 cm, patient induced non-repetitive rubbing motion, and patient induced non-repetitive hand
motion in supine position. Pulse oximeter SpO2 readings were compared to SaO2 values of drawn blood samples
measured by CO-oximetry. Subjects comprised both adult men and women and spanned a range of skin
pigmentations.
****GE TruSignal technology with OxyTip+/GE TruSignal sensors have been validated for pulse rate accuracy over the
specified range in bench top testing against a patient simulator. Accuracy was calculated as the root-mean-square
(rms) difference between paired pulse rate data recorded with the pulse oximeter equipment and with the patient
simulator.
NOTE
Accuracy may vary for some sensors; always check the instructions for the sensor.
*The sensors were clinically tested for accuracy with the following
sensors:
*The sensors were clinically tested for accuracy with the following
sensors:
OXY-E (equivalent to OXY-E-UN, TS-E-D, TS-E2-GE, TS-E4-GE)
OXY-SE (equivalent to OXY-SE-3, TS-SE-3)
OXY-F (equivalent to OXY-F-UN, TS-F-D, TS-F2-GE, TS-F4-GE, TS-SA-D, TS-SA4-
GE)
OXY-W (equivalent to OXY-W-UN, TS-W-D)
OXY-AP (equivalent to OXY-AP-10, OXY-AP-25, TS-AP-10, TS-AP-25)
OXY-AF (equivalent to OXY-AF-10, TS-AF-10, TS-AF-25)
*The sensors were clinically tested for neonatal accuracy with the
following sensors:
OXY-SE (equivalent to OXY-SE-3, TS-SE-3)
OXY-AF (equivalent to OXY-AF-10, TS-AF-10, TS-AF-25)
Because SpO2 measurements are statistically distributed, only about 2/3 of the
measurements can be expected to fall within ±1 Arms of the value measured
by a CO-oximeter.
NOTE
Wavelength of SpO2 probe LEDs Infrared: 880 to 900 nm (nominal)
This information may be useful
Red 650 to 670 nm (nominal)
to clinicians, such as those
performing photodynamic
Maximum energy of SpO2 probe LEDs Infrared LED 42 µJ/pulse therapy.
Red LED 62 µJ/pulse
Measurement range
1 to 100%
SpO2
20 to 300 bpm
Pulse rate
0.03 to 20%
Perfusion range
Averaging Time
During normal measurement conditions in the Normal mode, the SpO2 averaging time is six (6) to seven (7) seconds or
approximately three (3) seconds in Fast mode.
Sensor Model Type LoSAT Range 60% to 80% Standard Range 70% to 100%
The root mean square of the difference (rmsd) for set SpO2 and displayed SpO2 was 0.727. Seventy data points were
taken in the test in the 70 to 100% saturation range. The root mean square of the difference (rmsd) for set PR and
displayed PR was 1.581. Seventy data points were taken in the test in the 40 to 250 BPM pulse rate. All data for this
test is stored at Nellcor/Covidien.
*Adult specifications are shown for OXIMAX MAX-A and MAX-N sensors with the N-600.
Saturation accuracy will vary by sensor type. This variation equals plus or minus one standard deviation. Plus or minus
one standard deviation encompasses 68% of the population. Accuracy is based on deep hypoxia studies on healthy
adult volunteer subjects. Arterial blood samples were analyzed simultaneously on multiple CO-oximeters.
1 Subjects used to validate SpO2 measurement accuracies were healthy and recruited from the local population.
Comprised of both men and women, subjects spanned a range of skin pigmentations and ranged in age from 18-50
years old. Accuracy specifications are based on controlled hypoxia studies with healthy non-smoking adult volunteers
over the specified saturation SpO2 range(s). Pulse oximeter SpO2 readings were compared to SaO2 values of drawn
blood samples measured by hemoximetry. All accuracies are expressed as ±1 SD. Pulse oximeter equipment
measurements are statistically distributed; about two-thirds of pulse oximeter measurements can be expected to fall
in this accuracy (Arms) range. Because scatter and bias of pulse oximeter SpO2 and blood SaO2 comparison commonly
increase as the saturation decreases, and accuracy specifications are calculated from data spanning the stated
range, different accuracy values may result when describing partially overlapping ranges.
2Clinical functionality of the MAX-N has been demonstrated on a population of hospitalized neonate patients. The
observed SpO2 accuracy was 2.5% in a study of 42 patients with ages of 1 to 23 days, weight from 750 to 4,100
grams, and 63 observations made spanning a range of 85% to 99% SaO2.
3Clinical functionality has been demonstrated in a population of hospitalized neonate patients. The observed SpO2
accuracy was 3.0% in a study of 57 patients with ages of 24 to 40 weeks, weight from 710 to 5,000 grams, and 185
observations made spanning a range of 63% to 99% SaO2.
4 Specification applies to N-600x oximeter performance. Reading accuracy in the presence of low perfusion (detected
IR pulse modulation amplitude 0.03% - 1.5%) was validated using signals supplied by a patient simulator. SpO2 and
pulse rate values were varied across the monitoring range over a range of weak signal conditions and compared to
the known true saturation and pulse rate of the input signals.
SatSeconds OFF
Measurement range
Resolution
SpO2 1%
SpCO 1%
SpMet 0.1%
Saturation accuracy
SpCO - adult/pediatric/infants 1 - 40 ± 3%
SpMet 1 - 15 ± 1%
1SpO
2, SpCO and SpMet accuracy was determined by testing on healthy adult volunteers in the range of 60-100%
SpO2, 0-40% SpCO, and 0-15% SpMet against a laboratory CO-oximeter. SpO2 and SpMet accuracy was determined
on 16 neonatal NICU patients ranging in age from 7-135 days old and weighing between 0.5-4.25 kg. Seventy-nine
(79) data samples were collected over a range of 70-100% SaO2 and 0.5-2.5% MetHb with a resultant accuracy of
2.9% SpO2 and 0.9% SpMet.
2The Masimo sensors have been validated for no motion accuracy in human blood studies on healthy adult male and
female volunteers with light to dark skin pigmentation in induced hypoxia studies in the range or 70-100% SpO₂
against a laboratory CO-oximeter and EGG monitor. This variation equals plus or minus one standard deviation. Plus
or minus one standard deviation encompasses 68% of the population.
3The Masimo sensors have been validated for motion accuracy in human blood studies on healthy adult male and
female volunteers with light to dark skin pigmentation in induced hypoxia studies in the range or 70-100% SpO₂
against a laboratory CO-oximeter and EGG monitor. This variation equals plus or minus one standard deviation which
encompasses 68% of the population.
4The Masimo SET® Technology has been validated for low perfusion accuracy in bench top testing against a Biotek
Index 2 simulator and Masimo‘s simulator with signal strengths of greater than 0.02% and transmission of greater
than 5% for saturations ranging from 70 to 100%. This variation equals plus or minus one standard deviation which
encompasses 68% of the population.
5The Masimo sensors have been validated for pulse rate accuracy for the range of 25-240 bmp in bench top testing
against a Biotek Index 2 simulator. This variation equals plus or minus one standard deviation which encompasses
68% of the population.
6SpHb accuracy has been validated on healthy adult male and female volunteers and on surgical patients with light to
dark skin pigmentation in the range of 8-17 g/dl SpHb against a laboratory CO-oximeter. This variation equals plus or
minus one standard deviation which encompasses 68% of the population. The SpHb accuracy has not been validated
with motion or low perfusion.
Sensor accuracy specified when used with Masimo SET® MX board using a Masimo patient cable for LNOP sensors,
the LNCS sensors, or the M-LNCS sensors. Numbers represent Arms (rms error compared to the reference). Because
pulse oximeter measurements are statistically distributed, only about two-thirds of the measurements can be
expected to fall within a range of ±Arms compared to the reference value. SpO2 accuracy is specified from 70% to
100%. Pulse Rate accuracy is specified from 25 to 240 bpm.
Desat occurrences 2
Desat threshold 2%
PI delta range 50
FastSat Off
NOTE
If large changes occur in the ambient temperature, the temperature
system can be recalibrated by cycling the monitor’s power using the
On/Off button.
Clinical accuracy of predictive mode Results of clinical accuracy validation according to ISO 80601-2-56
NOTE
Use only Welch Allyn probe covers on the Welch Allyn temperature probes. The size, shape, and thermal
characteristics of the probe covers can affect the performance of the probe. Inaccurate readings or retention
problems may occur unless Welch Allyn temperature probes and Welch Allyn probe covers are used. Refer to the
CARESCAPE VC150 supplies and accessories document for reorder part numbers.
Units of measure °Celsius (C) or °Fahrenheit (F). Defined by purchased Exergen scanner.
Range
Arterial head balance range for body 34.5° to 43° C (94° to 110° F)
temperature1
1 Automatically applied when temperature is within normal body temperature range, otherwise reads surface
temperature.
NOTE
Use only Exergen probe covers. The size and shape of the probe covers can affect the performance of the
scanner. Inaccurate readings occur unless the proper probe covers are used. Refer to the CARESCAPE VC150
supplies and accessories document (KO00125) for reorder part numbers.
The above battery run times are valid for a new battery and depend
heavily on display brightness settings. After approximately 300 full charge
and discharge cycles, the capacity of the battery is reduced to 70 percent
of its original rating.
Charge time Approximately 4 hours from full discharge when the monitor is off.
Approximately 4 hours when the monitor on.
Connectivity specifications
Wireless specifications
WLAN bands Capable of communicating on 2.4 GHz and 5.1 GHz bands.
Wireless specifications
Number of SSID profiles Up to 4 SSID profiles that service can define and configure.
EMC Compliance Complies with IEC Publication 60601-1-2 Medical Electrical Equipment,
Electromagnetic Compatibility Requirements and Tests and CISPR 11
(Group 1, Class B) for radiated and conducted emissions.
RF information displayed Radio frequency (i.e., the frequency associated with the channel number)
Transmit power in terms of dBm
Signal level (RSSI) in terms of dBm
Wireless specifications
5GHz:
5180 to 5320MHz
5500 to 5700MHz
5745 to 5825MHz
Alarm specifications
Alarm delays and remote alarm There is no discernible delay in alarms due to electronics or software
algorithms. All priority alarms are triggered immediately unless a delay
has been configured. Refer to SpO2 section for different types of alarm
delays available depending on SpO2 technology.
All medium and high-priority alarms are asserted to the remote alarm
interface within 0.5 sec unless alarms are silenced for the two-minute
period.
Patent information
Nellcor patents
Covidien LP. US Patents: 5,485,847; 5,676,141; 5,743,263; 6,035,223;6,226,539;
6,411,833; 6,463,310; 6,591,123; 6,708,049; 7,016,715; 7,039,538;
7,120,479;7,120,480; 7,142,142; 7,162,288; 7,190,985;7,194,293; 7,209,774;
7,212,847; 7,400,919.
Masimo patents
Patents: www.masimo.com/patents.htm
Exergen patents
For patent information, please visit www.exergen.com/patents
The exact terms of the licenses, disclaimers, acknowledgements, and notices are
reproduced in the materials provided with this product. Innokas offers to provide
you with the source code as defined in the applicable license.
This offer is valid for a period of three (3) years from the date of the distribution of
this product by Innokas.
By submitting a request, you give your consent that Innokas (or third parties on
behalf and under direct authority of Innokas) will process your personal data.
The processing will be done for the purpose of the request and the undertakings
related to it. The main reason for storing this data is to prove compliance with
the license terms. The data processing will be done in compliance with Innokas
guidelines and applicable legislation.
Unlike the transducers/electrodes used in these devices, the NIBP cuff has two
functions. In addition to sensing the pressure pulses in the cuff, the cuff occludes
and then releases the patient's artery to create the conditions that allow blood
pressure to be measured.
An "artificial arm" would need to test both the sensing and occluding functions of
the cuff, and mimic the nonlinear dynamics of the artery to provide an effective
clinical simulation. While this has been attempted (2), there are no effective
"arms" available.
Commercial NIBP simulators do attempt to test both functions of the NIBP cuff.
Pressure signals are generated by the simulator in response to the inflation and
deflation cycles of the monitor. While the cuff may be in the system, it is
wrapped on a mandrel. The ability of the cuff to transducer pressure signals or
to occlude the artery is not tested.
There are further limitations to the pressure pulses used by simulators. During
the deflation of the cuff, the shape of the generated pressure oscillations
changes as the cuff goes from systolic to diastolic pressures. This is due to the
fact that the artery is only open when the arterial pressure is above cuff
pressure. As can be seen in Figures 1A-1C, the shape of the oscillation changes
as the cuff pressure changes, and the artery opens. Commercial NIBP simulators
use one waveform shape at all pressure levels, which is simply scaled to reflect
the oscillometric envelope.
"Since the CuffLink produces the same response independent of the inflate/
deflate cycle or the algorithm used by the monitor, we offer the term "Target
Value" as an approximation of the patient's actual blood pressure" (3)
"Neither the monitor or the NIBP Pump 2 is broken. Some monitors were
designed to give readings close to those obtained by the Auscultatory method of
blood pressure determinations. Other monitors have been designed to agree
with Invasive blood pressure readings. It is well known that Invasive and
Auscultatory NIBP readings on the same subject can be quite different" (4
Simulators can also be used to test for leaks and conduct static pressure
calibration of NIBP monitors, and as part of preventive maintenance programs.
Summary
The accuracy of an NIBP monitor can only be determined by comparison to a
clinical blood pressure reference. NIBP simulators are useful for certain types of
testing, but should not be used for accuracy testing.
References
1. FDA, CDRH, Noninvasive Blood Pressure (NIBP) Monitor Guidance, March 10,
1997
2. Mieke, S, Substitute of simulators for human subjects; Blood Press Monit,
October 1, 1997; 2(5): 251-256
3. DNI Nevada CuffLink Noninvasive Blood Pressure Analyzer, Operating and
Service Manual; Revision E, 11/97
4. NIBP Pump 2 Noninvasive Blood Pressure Simulator and Tester, Operations
Manual; Revision C, January 2003
WARNINGS
Use of known RF sources, such as cell/portable phones, or
other radio frequency (RF) emitting equipment near the system
may cause unexpected or adverse operation of this device/
system. Consult qualified personnel regarding device/system
configuration. Recommended separation distance between
ordinary mobile phone and CARESCAPETM VC150 is 3.3 m.
CAUTIONS
The equipment or system should not be used adjacent to, or
stacked with, other equipment. If adjacent or stacked use is
necessary, the equipment or system should be tested to verify
normal operation in the configuration in which it is being used.
NOTE
Medical electrical equipment requires special electromagnetic compatibility
(EMC) precautions which must be considered when installing and putting
this equipment into operation. Refer to VC150 Service Manual (KO00206),
section C: Electromagnetic Compatibility (EMC).
Electromagnetic environment –
Emissions test Compliance
guidance
Electromagnetic immunity
Guidance and manufacturer´s declaration - electromagnetic emissions
The CARESCAPE VC150 is intended for use in the electromagnetic environment specified below. The customer or the
user of the CARESCAPE VC150 should assure that it is used in such an environment.
Electromagnetic environment –
Immunity test IEC 60601 test level Compliance level
guidance
Electrical Fast ± 2 kV for power supply ± 2 kV for power supply Mains power should be that of a
Transient/Burst lines lines typical commercial or hospital
IEC 61000-4-4 environment.
±1 kV for input/output Not applicable
lines
Voltage dips, short <5% UT (>95% dip in UT <5% UT (>95% dip in UT) for Mains power quality should be that
interruptions and for 0.5 cycle 0.5 cycle of a typical commercial or hospital
voltage variations on environment. If user of the
power supply input <40% UT (>60% dip in UT) <40% UT (>60% dip in UT) CARESCAPE VC150 requires
lines for 5 cycles for 5 cycles continued operation during power
IEC 61000-4-11 mains interruptions, it is
<70% UT (>30% dip in UT) <70% UT (>30% dip in UT) recommended that the
for 25 cycles for 25 cycles CARESCAPE VC150 be powered
from an uninterruptible power
<5% UT (>95% dip in UT) <5% UT (>95% dip in UT) for supply or battery.
for 5 s 5s
Electromagnetic immunity
Guidance and manufacturer´s declaration - electromagnetic emissions
The CARESCAPE VC150 is intended for use in the electromagnetic environment specified below. The customer or the
user of the CARESCAPE VC150 should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment – guidance
d = 1.2 P
NOTES
At 80 MHz and 800 MHz, the higher frequency range applies.
These guidelines may not apply in all situations. Electromagnetic propagation is affected by reflection from
structures, objects, and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which the
CARESCAPE VC150 is used exceeds the applicable RF compliance level above, the CARESCAPE VC150 should
be observed to verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as reorienting or relocating the CARESCAPE VC150.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
The CARESCAPE VC150 is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the CARESCAPE VC150 can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the
CARESCAPE VC150 as recommended below, according to the maximum output power of the communications
equipment.
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTES
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
CAUTION
The use of accessories, transducers and cables other than
those specified in the CARESCAPE VC150 Supplies and
accessories document (KO00125) may result in increased
emissions or decreased immunity performance of the
equipment or system. Use only accessories listed in the
CARESCAPE VC150 Supplies and accessories document.
The monitor complies with EMC standards when used with cables, transducers,
and accessories listed in the CARESCAPE VC150 Supplies and accessories
document.
For other product information please contact one of the offices listed on the
back cover of the manual.
Make sure you have the following information available before calling:
Ordering Manuals
When ordering additional operator manuals, be sure to include the software
version of the product.
Asia Headquarters
GE Medical Systems
Information Technologies Asia; GE (China)
Co., Ltd.
No1 Huatuo Road,
Zhangjiang Hi-tech Park Pudong
Shanghai P.R.China 201203
Tel: + 86 21 5257 4650
Fax: + 86 21 5208 2008