CT Test Unit
A133506
Quick Start Guide

Version 1.4
September 2022
P/N A11 001358
Legal Notices
The information in this document is subject to change without notice.

Chroma ATE INC. makes no warranty of any kind with regard to this manual,
including, but not limited to, the implied warranties of merchantability and fitness for
a particular purpose. Chroma ATE INC. shall not be held liable for errors contained
herein or direct, indirect, special, incidental or consequential damages in connection
with the furnishing, performance, or use of this material.

CHROMA ATE INC.


88 Wenmao Rd., Guishan Dist., Taoyuan City 333001, Taiwan

Copyright Notices. Copyright 2012 Chroma ATE INC., all rights reserved.
Reproduction, adaptation, or translation of this document without prior written
permission is prohibited, except as allowed under the copyright laws.

ii
Warranty
All of Chroma’s instruments are warranted against defects in material and
workmanship for a period of one year from date of shipment. Chroma agrees to
repair or replace any assembly or component found to be defective, under normal
use during this period. Chroma’s obligation under this warranty is limited solely to
repairing any such instrument, which in Chroma’s sole opinion proves to be
defective within the scope of the warranty when returned to the factory or to an
authorized service center. Purchaser is responsible for the shipping and cost of the
service item to Chroma factory or service center. Shipment should not be made
without prior authorization by Chroma.

This warranty does not apply to any products repaired or altered by persons not
authorized by Chroma, or not in accordance with instructions furnished by Chroma.
If the instrument is defective as a result of misuse, improper repair, or abnormal
conditions or operations, repairs will be billed at cost.

Chroma assumes no responsibility for its product being used in a hazardous or


dangerous manner either alone or in conjunction with other equipment. High voltage
used in some instruments may be dangerous if misused. Special disclaimers apply
to these instruments. Chroma assumes no liability for secondary charges or
consequential damages and in any event, Chroma’s liability for breach of warranty
under any contract or otherwise, shall not exceed the purchase price of the specific
instrument shipped and against which a claim is made.

Any recommendations made by Chroma regarding the use of its products are
based upon tests believed to be reliable; Chroma makes no warranty of the results
to be obtained. This warranty is in lieu of all other warranties, expressed or implied,
and no representative or person is authorized to represent or assume for Chroma
any liability in connection with the sale of our products other than set forth herein.

Chroma undertakes no liability for any special or consequential damages, or any of


our customer’s compensation responsibilities to third parties, arising from use or
access of our product.

CHROMA ATE INC.


88 Wenmao Rd., Guishan Dist.,
Taoyuan City 333001, Taiwan
Tel: 886-3-327-9999
Fax: 886-3-327-8898
e-mail: info@chromaate.com

www.chromaate.com

iii
Material Contents Declaration
The recycling label shown on the product indicates the Hazardous Substances contained in
the product as the table listed below.

: See <Table 1>.

: See <Table 2>.

<Table 1>
Hazardous Substances
Lead Mercury Cadmium Hexavalent Polybrominated Selected Phthalates
Chromium Biphenyls/ Group
Part Name
Polybromodiphenyl
Ethers
Pb Hg Cd Cr6+ PBB/PBDE DEHP/BBP/DBP/DIBP
PCBA O O O O O O
CHASSIS O O O O O O
ACCESSORY O O O O O O
PACKAGE O O O O O O
“O” indicates that the level of the specified chemical substance is less than the threshold level specified in
the standards of SJ/T-11363-2006, EU Directive 2011/65/EU, and 2015/863/EU.
“” indicates that the level of the specified chemical substance exceeds the threshold level specified in the
standards of SJ/T-11363-2006, EU Directive 2011/65/EU, and 2015/863/EU.

Remarks:
1. The CE marking on product is a declaration of product compliance with EU Directive 2011/65/EU, and
2015/863/EU.
2. This product is complied with EU REACH regulation and no SVHC in use.

Disposal
Do not dispose of electrical appliances as unsorted municipal waste, use separate collection
facilities. Contact your local government for information regarding the collection systems
available. If electrical appliances are disposed of in landfills or dumps, hazardous substances
can leak into the groundwater and get into the food chain, damaging your health and well-
being. When replacing old appliances with new one, the retailer is legally obligated to take
back your old appliances for disposal at least for free of charge.

iv
<Table 2>
Hazardous Substances
Lead Mercury Cadmium Hexavalent Polybrominated Selected Phthalates
Chromium Biphenyls/ Group
Part Name
Polybromodiphenyl
Ethers
Pb Hg Cd Cr6+ PBB/PBDE DEHP/BBP/DBP/DIBP
PCBA  O O O O O
CHASSIS  O O O O O
ACCESSORY  O O O O O
PACKAGE O O O O O O
“O” indicates that the level of the specified chemical substance is less than the threshold level specified in
the standards of SJ/T-11363-2006, EU Directive 2011/65/EU, and 2015/863/EU.

“” indicates that the level of the specified chemical substance exceeds the threshold level specified in the
standards of SJ/T-11363-2006, EU Directive 2011/65/EU, and 2015/863/EU.

1. Chroma is not fully transitioned to lead-free solder assembly at this moment; however, most of the
components used are RoHS compliant.
2. The environment-friendly usage period of the product is assumed under the operating environment
specified in each product’s specification.
3. This product is complied with EU REACH regulation and no SVHC in use.

Disposal
Do not dispose of electrical appliances as unsorted municipal waste, use separate collection
facilities. Contact your local government for information regarding the collection systems
available. If electrical appliances are disposed of in landfills or dumps, hazardous substances
can leak into the groundwater and get into the food chain, damaging your health and well-
being. When replacing old appliances with new one, the retailer is legally obligated to take
back your old appliances for disposal at least for free of charge.

v
Safety Summary
The following general safety precautions must be observed during all phases of
operation, service, and repair of this product. Failure to comply with these
precautions or specific WARNINGS given elsewhere in this manual will violate
safety standards of design, manufacture, and intended use of the instrument.
Chroma assumes no liability for the customer’s failure to comply with these
requirements.

BEFORE APPLYING POWER


Verify that the power is set to match the rated input of this
device.

PROTECTIVE GROUNDING
Make sure to connect the protective grounding to prevent an
electric shock before turning on the power.

NECESSITY OF PROTECTIVE GROUNDING


Never cut off the internal or external protective grounding
wire, or disconnect the wiring of protective grounding
terminal. Doing so will cause a potential shock hazard that
may bring injury to a person.

FUSES
Only fuses with the required rated current, voltage, and
specified type (normal blow, time delay, etc.) should be
used. Do not use repaired fuses or short-circuited fuse
holders. To do so could cause a shock or fire hazard.

DO NOT OPERATE IN AN EXPLOSIVE ATMOSPHERE


Do not operate the instrument in the presence of flammable
gases or fumes.

DO NOT REMOVE THE COVER OF THE INSTRUMENT


Operating personnel must not remove the cover of the
instrument. Component replacement and internal adjustment
can be done only by qualified service personnel.

vi
Safety Symbols

DANGER – High voltage.

Explanation: To avoid injury, death of personnel, or


damage to the instrument, the operator must refer to the
explanation in the manual.

High temperature: This symbol indicates the temperature is


hazardous. Do not touch to avoid personal injury.

Protective grounding terminal: This symbol indicates that


the terminal must be connected to ground before operation
of the equipment to protect against electrical shock in case
of a fault.

Earth (ground) terminal

Frame or chassis: To identify a frame or chassis terminal.

Alternating Current (AC)

Direct Current (DC) / Alternating Current (AC)

Direct Current (DC)

Push-on/Push-off power switch

The WARNING sign highlights an essential operating or


maintenance procedure, practice, condition, statement, etc.,
WARNING
which if not strictly observed, could result in injury to, or
death of, personnel or long term health hazards.

The CAUTION sign highlights an essential operating or


maintenance procedure, practice, condition, statement, etc.,
CAUTION
which if not strictly observed, could result in damage to, or
destruction of, equipment.

The Notice sign highlights an essential operating or


maintenance procedure, condition, or statement.

vii
CT Test Unit A133506 Quick Start Guide

Table of Contents
1. Overview ........................................................................................... 1
1.1 Introduction ..................................................................................... 1
1.2 Checking before Use ...................................................................... 1
1.3 Ambient Environment ..................................................................... 3
1.4 Cleaning ......................................................................................... 3
2. Specifications ................................................................................... 4
2.1 Common Specifications.................................................................. 4
2.2 Electrical Specifications.................................................................. 4
3. Description of Panel ........................................................................ 5
3.1 Front Panel ..................................................................................... 5
3.2 Side View ....................................................................................... 6
4. Operation .......................................................................................... 9
4.1 Preparation before Test.................................................................. 9
4.2 Connecting Test Device ................................................................. 9
4.2.1 Standard Connection .............................................................. 9
4.3 Panel Display & Operation ........................................................... 11

ix
CT Test Unit A133506 Quick Start Guide

Overview
1.1 Introduction
This Quick Start Guide covers the description and operation of A133506.
A133506 is a CT Test Unit that has current transformer coils, inductance,
quality factor and DC resistance measurement functions when connected
with 13350D and 13350M-200K.It can judge the test result for PASS or
FAIL based on the inputted specification from the operation panel, also
display the measured results of current transformer coils, inductance,
quality factor and DC resistance on 13350D.

1.2 Checking before Use


Before shipment, this device was inspected and found to be free of
mechanical and electrical defects. As soon as the device is unpacked,
inspect for any damage that may have occurred in transit. Save all packing
materials in case the device has to be returned. If damage is found, please
file claim with carrier immediately. Do not return the device to Chroma
without prior approval.

Please refer to Table 1-1 and Table 1-2 to check the following options to
match the one came along with the shipment upon receipt.

Table 1-1 Standard Accessory


Model Item Qty Remark
Connecting the A133506 panel
High current test rod 1
for measuring coil ratio
BNC terminal to alligator Connecting HCUR and HPOT
2
clip wire (red) connectors for measurement
BNC terminal to alligator Connecting LCUR and LPOT
2
clip wire (black) connectors for measurement
A133506
R type terminal to 1 Connecting SEC.(+) terminal for
alligator clip wire (red) measuring coil ratio
R type terminal to 1 Connecting SEC.(-) terminal for
alligator clip wire (black) measuring coil ratio
External connector of 1 Connecting the referenced
referenced sample sample on A133506

1
CT Test Unit A133506 Quick Start Guide

Connector of external 1 Connecting the external load to


load A133506

Table 1-2 Option Accessory


Item Qty Remark
A110234 High frequency Connecting HCUR, HPOT, LCUR and LPOT
1
testing wire connector for measurement

Note Only item name is required when ordering any of the accessories.

2
CT Test Unit A133506 Quick Start Guide

1.3 Ambient Environment


(1) Do not use the test unit in a
dusty or vibrating location. Do
not expose it to sunlight or
corrosive gas and be sure
that the ambient temperature
is 0C ~ +40C and the
relative humidity is between
20% ~ 80%.

(2) The test unit has been


carefully designed to reduce
the noise from the AC power
source. However, it should be
used in an environment with
lowest noise as possible. If
noise is inevitable, please
install a power filter.

(3) The test unit should be stored


within the temperature range
of -40C ~ +85C. If the scan
box is not going to be used
for a long time, please store it
in its original box or a similar
package and keep it away
from direct sunlight and
humidity in order to maintain
its accuracy when in use
again.

1.4 Cleaning
Remove all connected wires and cables on the device before cleaning. Use
a dry cloth to clean the chassis. As to the rear panel of test unit that
connects the internal circuit board, to avoid damaging the device due short-
circuited internally, it is forbidden to use any damp cloth for cleaning.

3
CT Test Unit A133506 Quick Start Guide

Specifications
2.1 Common Specifications
Power Supply 5VDC (provided by 13350M-200K)
Interface GOLDENFISH (connect with 13350M-200K)
Operating
0°C - 40°C
Temperature
Storage
-40°C - 85°C
Temperature
120 x 65.6 x 280 mm / 4.73 x 2.58 x 11.03 in.
Dimension (exclude exterior parts)
(WxHxD) 145.8 x 113.7 x280 mm / 5.74 x 4.48 x 11.03
in. (exclude exterior parts)
Weight Approx. 1.3 kg / 2.87 lbs

2.2 Electrical Specifications


 Parameter Determination: L, DCR, Turn, Q
 Frequency Determination: 20Hz to 200KHz
 Voltage Determination: 10mV to 2Vrms, 10mV each section
 Coil Raito: Frequency 20Hz to 200KHz, test voltage 0.1V to 10Vrms

CAUTION As this test unit accepts 50V induced voltage only, the
Vs for setting coil ratio should not larger than 50V.

4
CT Test Unit A133506 Quick Start Guide

Description of Panel
3.1 Front Panel
The front panel of A133506 CT Test Unit is shown in Figure 3-1 and the
parts numbered from (1) - (4) are described below.

Figure 3-1 Outline of Front Panel

(1) High Current Side : It is the test rod connected to sense the high
Test Rod current side of current transformer.
(2) NG LED : When on, it indicates the measured value
exceeds the set range and the result is Fail.
5
CT Test Unit A133506 Quick Start Guide

(3) START Key : Using this key to trigger when trigger the
measurement externally or manually.
(4) GOOD LED : When on, it indicates the measured value is
within the set range and the result is PASS.

3.2 Side View


In the following three figures, Figure 3-2 shows the side of connecting
13350M-200K, Figure 3-3 shows the load area to set internal/external DUT
while
Figure 3-4 starting from the left shows the DUT measuring area, CT Test
Unit setting area for internal/external reference sample and large current
side area for measuring coils. Each part is numbered from (1) to (12) and
described below.

Figure 3-2 Side View of Left

6
CT Test Unit A133506 Quick Start Guide

Figure 3-3 View of Bottom

Figure 3-4 Side View of Right

(1) Goldfish Connector : It is a 46-pin goldfish connector to connect


(Large) 13350M-200K.
(2) High Current Side : Same as (1) on the front panel which is a test
Test Rod rod connected to sense the high current side
of current transformer.
7
CT Test Unit A133506 Quick Start Guide

(3) Goldfish Connector : It is a 30-pin goldfish connector to connect


(Small) 13350M-200K.
(4) External Load : It connects the load resistor of current
Connect transformer connected to the external DUT.
(5) Load Switch : It controls the DUT of current transformer to
use internal or external load.
(6) LCUR Connector : It is the main signal path for low potential.
(DUT)
(7) LPOT Connector : It is the signal path for low potential
measurement. (DUT)
(8) HPOT Connector : It is the signal path for high potential
measurement. (DUT)
(9) HCUR Connector : It is the main signal path for high potential.
(DUT)
(10) Reference Sample : It controls the reference sample of current
Switch transformer for internal or external use.
(11) Reference Sample : It is the connector to connect the reference
Connector sample of current transformer for external
use.
(12) High Current Side : It is connector for current transformer to
Connector connect high current side.

8
CT Test Unit A133506 Quick Start Guide

Operation
4.1 Preparation before Test
Before connecting the 13350M Series, be sure that there is no foreign
object attached to the goldfish connector on the left side panel and ensure
the power of 13350M is OFF.

4.2 Connecting Test Device

4.2.1 Standard Connection


A133506 CT Test Unit is able to use the following two types of connections
(using internal referenced sample current transformer and built-in DUT
load) as shown in Figure 4-1 and Figure 4-2. The yellow lines indicate the
goldfish connector, the blue line indicates the high current side path, the
black fill area indicates the current transformer while the red lines are HCUR
and HPOT and the black lines are LCUR and LPOT.

9
CT Test Unit A133506 Quick Start Guide

13350D

13350M-200K

A133506

Figure 4-1 A133506 Connection (Using Test Rod)

10
CT Test Unit A133506 Quick Start Guide

13350D

13350M-200K

A133506

Figure 4-2 A133506 Connection (Using External Terminal)

4.3 Panel Display & Operation


When connecting A133506 to 13350M-200K and 13350D, turn on the
13350M-200K power the 13350D will prompt for system confirmation and
link for loading. The panel will display the initial settings or the status set
since last LOCK automatically.

When the 13350M-200K is powered off, besides the status locked before
powered off, the rest of the settings will return to the initial state
automatically during next power on.

Refer to Automatic Transformer Tester 13350 User’s Manual for detail


operation. Here it is suggested to select 106X for CONST Rsou.