Professional Documents
Culture Documents
03 MEXA-7000 Daily Operation I1002860000C GZ0000009111C
03 MEXA-7000 Daily Operation I1002860000C GZ0000009111C
MEXA-7000
Daily Operation
Instruction Manual
CODE:GZ0000009111C
Preface
This manual describes the operation of the Motor Exhaust Gas Analyzer, MEXA-7000.
Be sure to read this manual before using the product to ensure proper and safe operation of
the product. Also safely store the manual so it is readily available whenever necessary.
Product specifications and appearance, as well as the contents of this manual are subject to
change without notice.
HORIBA, Ltd. warrants that the Product shall be free from defects in material and
workmanship and agrees to repair or replace free of charge, at option of HORIBA, Ltd., any
malfunctioned or damaged Product attributable to responsibility of HORIBA, Ltd. for a period
of one (1) year from the delivery unless otherwise agreed with a written agreement. In any one
of the following cases, none of the warranties set forth herein shall be extended;
Any malfunction or damage attributable to improper operation
Any malfunction attributable to repair or modification by any person not authorized by
HORIBA, Ltd.
Any malfunction or damage attributable to the use in an environment not specified in this
manual
Any malfunction or damage attributable to violation of the instructions in this manual or
operations in the manner not specified in this manual
Any malfunction or damage attributable to any cause or causes beyond the reasonable
control of HORIBA, Ltd. such as natural disasters
Any deterioration in appearance attributable to corrosion, rust, and so on
Replacement of consumables
HORIBA, LTD. SHALL NOT BE LIABLE FOR ANY DAMAGES RESULTING FROM ANY
MALFUNCTIONS OF THE PRODUCT, ANY ERASURE OF DATA, OR ANY OTHER USES
OF THE PRODUCT.
Trademarks
Company names and brand names are either registered trademarks or trademarks of the
respective companies. (R), (TM) symbols may be omitted in this manual.
Conformable Directive
Installation environment
This product is designed for the following environment.
Overvoltage Category II
Pollution degree 2
Any changes or modifications not expressly approved by the party responsible for compliance
shall void the user's authority to operate the equipment.
Warning
This equipment has been tested and found to comply with the limits for a Class A digital
device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable
protection against harmful interference when the equipment is operated in a commercial
environment. This equipment generates, uses, and can radiate radio frequency energy and, if
not installed and used in accordance with the instruction manual, may cause harmful
interference to radio communications.
Operation of this equipment in a residential area is likely to cause harmful interference in
which case the user will be required to correct the interference at his own expense.
Korea certification
Safety Policy
We arrange warning labels on our products, and describe notes and cautions in this manual.
Make sure to follow these instructions for your safety.
Message
- Type of warning
- Why it is dangerous and
how to avoid injury
- What would happen if the
Pictograph warning is not followed
● Label location
● Labels
Label 1
Label 2
Label 3
Label 4
Label 5
Label 6
Manual Information
Reference
This indicates the part of where to refer the information.
Tip
This indicates reference information.
Contents
1 SYSTEM OVERVIEW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.1 System Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.1.1 Outside view . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.1.2 Gas flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.1.3 Signal flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.2 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
2.2.1 Span gas concentration setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
2.2.2 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
2.3 Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
2.3.1 Measurement (for each line) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
2.3.2 Measurement (simultaneously on multiple lines) . . . . . . . . . . . . . . . . 30
4 TROUBLE SHOOTING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Figures
1 SYSTEM OVERVIEW
This chapter explains the system configuration. For further details, refer to MEXA-7000
installation manual.
1
1 SYSTEM OVERVIEW
*1: These units differ in quantity and availability depending on the customer's specifications.
2
1 SYSTEM OVERVIEW
Note
The flow and unit layout may differ from the customer's actual system.
Note
The flow and unit layout may differ from the customer's actual system.
3
1 SYSTEM OVERVIEW
Note
The flow and unit layout may differ from the customer's actual system.
4
1 SYSTEM OVERVIEW
5
1 SYSTEM OVERVIEW
6
1 SYSTEM OVERVIEW
7
1 SYSTEM OVERVIEW
8
1 SYSTEM OVERVIEW
*1: These components differ in quantity and availability depending on the customer's
specifications.
9
1 SYSTEM OVERVIEW
10
1 SYSTEM OVERVIEW
*1: These components differ in quantity and availability depending on the customer's
specifications.
11
1 SYSTEM OVERVIEW
front panel
inside of panel
12
1 SYSTEM OVERVIEW
13
2 DAILY MEASUREMENT OPERATION
Note
To ensure safety in daily operations, the system must be installed correctly. For further details on
the installation, refer to the installation manual.
If the system power has been completely OFF, finish the system startup and warm-up before
starting the operation. For the startup/warm-up procedures, refer to the installation manual.
If any alarm occurs while the system is running, immediately check the alarm contents and take
appropriate actions. For further details, refer to the maintenance manual and the analyzer
manual.
Make sure to perform regular maintenance (replacement and inspection) for system components/
properties such as the filter and analyzer's calibration curve accuracy. For the detailed
procedures, refer to the maintenance manual and the analyzer manual.
*1: The quantity of elements is only for one filter holder. If the number of filters in the system
are more than two, the corresponding number of elements must be prepared.
*2: The actual replacement interval may vary with the working conditions of your system.
In case of severe filter contamination, replace the filters at shorter intervals.
14
2 DAILY MEASUREMENT OPERATION
Oil catcher
Note
Usually insert the element into the oil catcher in the direction
indicated by the arrow. However, insert the element in the
opposite direction for an EGR line.
Location
Front of SHS
Inspecting/replacing
1. (In case of the system with EGR line) Turn the valve beside the oil catcher to Close.
2. Pull out the holder of the oil catcher and inspect inside element. If the element is
dirty, replace it with a new one.
Note
For the EGR line only, the element must be inserted in the reverse orientation to the arrow shown in
Fig. 15. For the other lines, elements must be inserted in the specified orientation.
3. Restore the oil catcher and turn the valve to original position.
15
2 DAILY MEASUREMENT OPERATION
Dust filter
Location
Front of SHS
Inspecting/replacing
1. Check that the printed letters on the element of the dust filter are readable through
the filter cover, and that the packing (O-ring) is closely attached to the cover.
2. If the element is unclean, remove the cover and replace the element with a new one.
Install the new element with its lettering side facing forward.
Note
The HF-packing should be fitted with the cut side facing the cover.
Note
After the replacement, securely install the cover to prevent leakage. Any leakage may affect the
performance of the analyzer.
16
2 DAILY MEASUREMENT OPERATION
O-ring
Filter holder
Location
Front of OVN
Inspecting/replacing
1. Remove the sample filter holder from OVN.
2. Check the following conditions.
The filter element is set to the filter holder.
The sintered cylinder filter is in its mounting position (OVN body).
The filter element and the sintered cylinder filter are clean.
3. Wash the sintered filter and replace the filter element with a new one, if they are not
clean.
4. Reinstall the holder with clean filters to OVN.
17
2 DAILY MEASUREMENT OPERATION
Location
Heated pre-filter unit
Inspecting/replacing
1. Remove the filter holder from the unit.
2. Check the status of filter element and replace it with a new one if it is unclean.
3. Reinstall the filter holder to the unit.
18
2 DAILY MEASUREMENT OPERATION
*1: To identify the detailed specifications of the span gas needed for your system, refer
to the utility information in the instruction manuals of the respective analyzers.
Operation gas
Table 12 Utility gas (operation gas)
Name Specification Supply pressure Remark
Ozone-source gas O2 purity 99.99% or more 100 kPa 10 kPa For CLA (NO/NOx analyzer)
100 kPa 10 kPa For FIA (THC analyzer)
Fuel gas H2 39.0 to 41.0 vol%/He For FIA and GFA (CH4)
150 kPa 10 kPa
(If both installed)
Burner air *2 High-purity air 100 kPa 10 kPa For FIA and GFA
Carrier gas *3 N2 purity 99.99% or more 100 kPa 10 kPa For MPA (O2 analyzer)
For CLA of atmospheric pressure
Dilution gas *3 N2 purity 99.99% or more 100 kPa 10 kPa
type
*1: If the operation gas is co-used with other systems, a pressure regulator may be
required.
*2: The high-purity air for the FIA burner air can also be used as the zero gas air for FIA
in the CVS line.
*3: Zero gas N2 is commonly used for MPA carrier N2 and DLA dilution N2.
Other gases
Table 13 Utility gas (other gases)
Name Specification Supply pressure Remark
Compressed air: 10 kPa to 60 kPa For SHS in main cabinet
Moisture
Saturated at -2°
C or less
Purge gas Dust
Removed by filter of 5 m 400 kPa to 900 kPa For OVN
or less
Oil 0.1 mg/m3 or less
Check gas See "Span gas" in Table 11
19
2 DAILY MEASUREMENT OPERATION
Note
The analyzers not selected here will not be warmed up even when their lines are set to [STBY]. Make
sure to select all necessary analyzers.
20
2 DAILY MEASUREMENT OPERATION
Procedure
1. Press the A-SEL button in the MEXA Utility area (Fig. 19).
(Or, select the A-SEL button and select [A-SEL] from the menu that appears.)
The Analyzer select screen is displayed (Fig. 20).
2. Select the line for setting.
Line button: Line to be set
3. Enable or disable the control for each analyzer/component.
Each analyzer: Control [ON]/[OFF]
4. Press the OK button.
The system saves the modifications and closes the Analyzer select screen.
To cancel the modifications, press the CANCEL button.
Tip
The analyzers set to ON here can be controlled from the STBY/PAUSE switching screen and by
the Line control button in the LINE Command area.
The analyzers set to OFF are only controllable from the relevant component button. The
component names on the buttons for the analyzers in the OFF state will be displayed in gray
character.
21
2 DAILY MEASUREMENT OPERATION
Warm-up
Warm up the analyzers in the lines in use.
Note
Before starting warm-up of the line containing CLA (NO/NOx analyzer) by setting the line to STBY,
verify that the ozone-source O2 is supplied to the system. Setting the line to [STBY] without O2 supply
may result in malfunctions of the ozone generator.
22
2 DAILY MEASUREMENT OPERATION
Procedure
1. Verify that the power of system and MCU is ON, and the RD LED and SD LED on IFC
are blinking.
2. Open the main valves of the operation gases and calibration gases for the
analyzers.
3. Press the Stand-By button at the upper right corner of the MCU screen (Fig. 21).
The STBY/PAUSE switching screen is displayed (Fig. 22).
4. Press the Line button and select lines to be warmed up.
[All lines]: Control all lines
Line name (e.g. [Tailpipe]): Control on each line
5. Press the STAND-BY button.
Warm-up for the specified lines starts.
6. Wait until the warm-up finishes.
The required warm-up time is normally 1 hour. If the system is equipped with a heated
(OVN or 19-inch) FIA, however, 2 hours will be required.
Tip
A typical warm-up time after the system enters the STBY mode is as follows
If the line warm-up starts immediately after system startup: 2 hours or more
If the system has been in the PAUSE mode
(system power ON) for 2 hours or more: 1 hour to 2 hours
The warm-up time required for individual units is typically 2 hours for analyzers or 1 hour for
sampling units.
The unit status in the STBY and PAUSE modes is as shown below.
23
2 DAILY MEASUREMENT OPERATION
2.2 Calibration
The analyzers must be calibrated before every measurement. The calibration procedure is
outlined below
Note
Fully warm up the analyzers before calibration. The system typically requires 1 hour to 2 hours
for warm-up after the line is set to STBY mode.
Always perform calibration before every measurement in order to ensure sufficient accuracy.
Procedure
1. Press the Line button at the top of the Data display area and select one of the
lines from the menu.
The analyzer component buttons, output values, and range buttons for the selected line
are displayed.
2. Press the relevant component button and select [Set span] from the menu.
The Span gas set screen is displayed (Fig. 23).
3. Select the Span box for the relevant range and enter the span concentration by the
keypad.
4. Press the OK button.
The system saves the modifications and closes the Span gas set screen.
To cancel the modifications, press the CANCEL button.
5. Repeat the steps from 1. to 4. for each analyzer.
Tip
The standard (HORIBA Std.) systems allow the users to input the concentration for one range
only. If a range is given a concentration value except 0, the concentration values for the
remaining ranges are automatically reset to 0.
The special (STD special) systems and the Federal systems allow the users to input the
concentration for several specified ranges (up to the number of SGS SPAN ports). To change the
selection of available ranges, you must call up another screen. (Refer to the MCU operation
manual.)
24
2 DAILY MEASUREMENT OPERATION
2.2.2 Calibration
For calibration, the following methods are available. Typically, the by-line methods are used.
By-line calibration (automatic sensitivity calibration)
25
2 DAILY MEASUREMENT OPERATION
26
2 DAILY MEASUREMENT OPERATION
27
2 DAILY MEASUREMENT OPERATION
28
2 DAILY MEASUREMENT OPERATION
2.3 Measurement
The system performs measurement for each line or for multiple lines at once.
Measurement for each line page 29
Simultaneous measurement for multiple lines page 30
Note
To ensure sufficient measurement accuracy, perform sensitivity calibration before every
measurement cycle.
During measurement, always check that no alarm occurs.
Make sure to purge the lines replace the filters, at appropriate timings during measurement and
after measurement.
29
2 DAILY MEASUREMENT OPERATION
Tip
You can set up the X and Y axes in the trend chart display.
(Refer to " 3.2.4 Time trend chart " (page 42))
In case of the system that support several sample ports for one line, such as gasoline or diesel
and exhaust or EGR, set the Ope./Sample switching button to [Sample] and select the sample
port.
(Refer to " 3.3.8 Sample port selection " (page 49))
The sequence time for the purge is changeable.
(Refer to " 3.3.6 Analyzer sequence time setting " (page 48))
30
2 DAILY MEASUREMENT OPERATION
Note
Power switches of all analyzers including the in-ANR, 19-inch, and OVN type analyzers must be
always ON. If the analyzers are turned OFF, the system will require longer time for a next startup and
warm-up cycle. In addition, the communication with IFC and MCU may fail depending on the timing.
31
3 MCU BASIC OPERATION
Note
When shutting down MCU, always follow the specified procedure below. Turning OFF MCU with
an inappropriate method may adversely affect the system startup from then on.
When you are going to stop the system for a short time from half day to a few days, do not turn
OFF MCU but leave the analyzers in the PAUSE status.
Procedure
1. Finish the measurement.
2. Press the Menu button on the Setup area and select [Utility].
3. From all buttons in the SYSTEM Utility area, press the Shut down button or the
Restart button.
Shut down button: Quit the software (MCU power OFF)
Restart button: Quit and restart the software
4. Confirm the displayed message and press the OK (or CANCEL) button.
32
3 MCU BASIC OPERATION
Note
For daily measurement operations, use the Normal user level.
Understand that the Service user level is dedicated only for HORIBA service engineers.
Procedure
1. Press the HORIBA button on the Title bar and select [User level] from the menu.
2. From the menu that appears next, select any of the user levels below.
[Normal]: Daily operation user
[Supervisor]: Maintenance user
3. If your are changing your user level from [Normal] to [Supervisor], the Enter
current password screen is displayed (Fig. 25).
Enter your password using the keyboard and press the OK button.
Tip
If the user level is [Supervisor], a string "Supervisor" is displayed in the center of the Title bar.
If your are selecting the Normal user level, no password is required.
33
3 MCU BASIC OPERATION
Procedure
1. Press the HORIBA button on the Title bar and select [Hard copy] from the menu.
2. From the menu that appears next, select [Printer setup].
The Printer Setup screen is displayed (Fig. 26).
3. Change the settings as necessary.
[Output Device]: Device type (Printer/USB Memory)
[Output Color]: Printer type (Monochrome/Color)
[Reverse Option]: Reverse screen color (ON/OFF)
[Gray Scale]: Monochrome gradation (ON/OFF)
[Printing mode]: Paper orientation (Landscape/Portrait)
[Printing scale]: Printout scale (Scale 100%/Scale 70%/Scale 30%)
4. Press the OK button. The system saves the modifications and closes the Printer
Setup screen.
Tip
The Gray Scale function may not be supported by some printers.
34
3 MCU BASIC OPERATION
Procedure
1. Press the HORIBA button on the Title bar and select [Select color] from the menu.
The Set warning color screen is displayed (Fig. 27).
2. Press the Color button and select one of the colors.
Color button: Letter color in which alarm-triggering item is displayed
(Red/White/Pink)
3. Press the OK button. The system saves the modifications and closes the Set
warning color screen.
Tip
This function changes the letter color of alarm-triggering items such as over-scale measurements,
which is set to red by default. It is useful especially when alarm-triggering items are illegible on a
screen hard copy.
35
3 MCU BASIC OPERATION
Note
When any alarm is issued, immediately confirm the contents and take appropriate actions. For further
details, refer to the troubleshooting in the Maintenance manual.
Procedure
1. Press the ALARM button on the Title bar. The Alarm screen is displayed (Fig. 28).
2. Confirm the display contents (alarm contents, line issuing the alarm, causal unit,
time of alarm occurrence).
The buttons available in this screen provide the following functions.
History button: Call up the Alarm history (past alarms) screen
Clear button: Delete the alarm contents from the Alarm screen
Alarm contents (selection):
Call up the Maintenance screen for the relevant analyzer
3. Press the OK button. The system saves the modifications and closes the Alarm
screen.
Tip
Pressing the History button displays the alarm history. To reset the display contents, press the
File clear button at the Supervisor user level.
The Clear button can only delete the alarm history. The actual alarm is not cleared.
For the details on the Maintenance screen for analyzer, refer to the MCU operation manual.
36
3 MCU BASIC OPERATION
37
3 MCU BASIC OPERATION
38
3 MCU BASIC OPERATION
39
3 MCU BASIC OPERATION
Gas control/calibration
Procedure
1. Press the Component button for the analyzer in the Data display area.
2. From the menu that appears, select the gas control.
[Reset]: Without gas sampling
[Zero]: Supplying zero gas
[Span]: Supplying span gas
[Cal.]: Automatic sensitivity calibration or zero/span calibration
Tip
If you select [Cal.] in Reset status, the automatic sensitivity calibration (zero calibration + span
calibration) is initiated. If you select [Cal.] in Zero (or Span) status, the zero calibration (span
calibration) is initiated at the moment.
The zero calibration for the standard (HORIBA Std.) system is always performed at the lowest
range.
40
3 MCU BASIC OPERATION
41
3 MCU BASIC OPERATION
Procedure
1. Press the View button in the Setup area and select [1 line/Chart] or [2 lines/Chart]
from the menu.
(Refer to " 3.2.1 Selection of data display format " (page 38))
The Time trend chart is displayed in the left half of the screen (Fig. 29).
2. Press the C button located at the upper left corner.
3. From the menu that appears, select size of the chart.
[Half screen]: Display half of the chart in the left half of the area
[Full screen]: Display the entire chart (no numeral data display)
42
3 MCU BASIC OPERATION
Procedure
1. Following the procedure in " Displaying trend chart" (page 42), call up the trend
chart.
2. Press C button and select [Config] from the menu.
The Trend chart signal setting screen is displayed (Fig. 30).
3. Using the table in the upper half of the screen, specify the following items.
[Include]: (check box) ON: Displayed, OFF: Not displayed
[Line#/Style]: Color of plot
[Line]: Gas line
[Component]: Component to be displayed
[Full scale in auto range]:
Display full scale in auto range mode (fixed)
4. Press the number button in [Graphs] to specify the number of charts to be
displayed.
Number button: Number of charts to be displayed (1 to 3)
5. Press the Graph # button and specify the Y-axis range for each chart.
Graph # button: Y-axis scale for chart # (-10% to 100% of data full scale)
6. Using the scroll bar, specify the upper limit for X-axis.
[X-axis]: X-axis range for the all charts (100 seconds to 1800 seconds)
7. Press the OK button. The system saves the modifications and closes the Trend
chart signal setting screen.
Tip
To specify A/F, _, and the EGR ratio (optional) for chat display, select [Other] for [Line].
If the auto range is not adopted, the full scale of the data is linked to that displayed on the Range
button.
If you select multiple charts, the components to be displayed are automatically distributed to the
charts.
43
3 MCU BASIC OPERATION
Note
When starting to use your MEXA-7000 system, first set the status of the line to be used to [STBY].
The analyzer warm-up is started.
When stopping the system for a short time, such as the end of the day's work, leave the system in
PAUSE status.
Procedure
1. Press the Stand-By button in the Setup area. The STBY/PAUSE switching screen is
displayed.
2. Press Line button and select the line/lines to be controlled from the menu.
[All lines]: Control on all lines
Line name (e.g. "Tailpipe"): Individual control on respective lines
3. Press the STAND-BY or PAUSE button.
The screen closes and the status of the selected line changes.
STAND-BY button: To warm-up/stand-by (STBY) mode
PAUSE button: To temporary stop (PAUSE) mode
Tip
The background of Component buttons for the THC/CH4 analyzer and NO/NOx analyzer are displayed
in dark color when the relevant lines are in the PAUSE mode. This means that the hydrogen flame and
ozone generator are OFF. If you switch the lines to [STBY] and therefore the hydrogen flame and
ozone generator turn ON, the relevant buttons are displayed in light color.
These buttons blink during the ON/OFF sequence of hydrogen flame or ozone generator.
44
3 MCU BASIC OPERATION
45
3 MCU BASIC OPERATION
46
3 MCU BASIC OPERATION
47
3 MCU BASIC OPERATION
Procedure
1. Press the Timer button in the MEXA Utility area and select [Function timer] from the
menu that appears.
The Function timer screen is displayed (Fig. 31).
2. Select objects of setting. Set the following items.
Line: Line to be set
Function: Sequence to be set
[Cal.]: Sequence time in automatic sensitivity calibration
[Zero]: Zero gas (1st) (0 second to 120 seconds)
[Span]: Span gas (0 second to 120 seconds)
[Re-zero] Zero gas (2nd) (0 second to 120 seconds)
[Read]: Data integration time (6 seconds to 60 seconds)
[Purge]: Line purge time (0 minute to 120 minutes)
3. Check or change the time for each sequence.
[Waiting time (T1): Gas stabilizing time for each sequence
(waiting time until data collection)
4. Press the OK button. The system saves the modifications and closes the Function
time screen.
Tip
Of all settings displayed in the Function time screen, [Cal.] and [Purge] are changeable to the Normal
users. Other items require a Supervisor qualification to be modified. The Normal users are only
allowed to confirm the values of these settings.
48
3 MCU BASIC OPERATION
49
4 TROUBLE SHOOTING
4 TROUBLE SHOOTING
This chapter explains how to handle the troubles that may occur in daily operation.
Trouble Situation Cause Action Reference
Check connection of
Entire system power
No power supplied power line to system and
cannot be ON
breakers.
4.1
Unit power SW is
Power cannot be Turn ON unit SW
OFF
turned ON Power for particular
unit cannot be ON Incorrect wiring, or poor
contact of wires and Check wiring.
connectors
PC or display power is not Check power line and
Nothing displayed
ON power SW.
Computer not start up A disk is in floppy disk Remove disk and reboot
correctly drive system.
Host computer is
Disable external control. 3.1.4
performing external control
4.2
MCU will not Close unnecessary
Sub screens or menus
start up/operate screens/menus by
remain open
CANCEL buttons
Software not operable
User is attempting to
operate buttons that are Quit measurement
not allowed due to ([Measure]) or change 3.1.2
sampling status or user [User level]
level.
Replace pump diaphragm/
Sample pump deteriorated
valve.
Flow rate indicator Filter plugged Clean or replace filter.
does not move during Incorrect piping, leakage Check sample piping and
measurement from filter leakage.
4.3 Analyzer is out of line Enable analyzer control to
No gas flow 3.3.3
control be used by A- SEL button.
Inappropriate gas supply Check zero gas/span gas
Flow rate indicator pressure supply pressures.
does not move during Check connections of
calibration Incorrect piping, leakage calibration gas piping and
leakage.
No calibration gas flow Refer to 4.3
Span gas concentration Set gas concentration in
3.2.3
data missing or incorrect Span gas set screen.
(System with SGS)
4.4 Select relevant range and
Range selection for 3.2.3
Sensitivity retry.
calibration not correct
calibration not Others
performed correctly Time setting for sensitivity
Modify setting of Function
calibration sequence 3.3.6
time.
inappropriate
Wait until warm-up
Insufficient warm-up 2.1.3
completed.
50
4 TROUBLE SHOOTING
51
2 Miyanohigashi, Kisshoin Minami-ku, Kyoto 601-8510 Japan
http://www.horiba.com