P/N 16225
Copyright 1999 2000 by METRON. All rights reserved.
METRON:
USA _ FRANCE ________________ NORWAY________________
1345 Monroe NW, Suite 255A 30, rue Paul Claudel Travbaneveien 1
Grand Rapids, MI 49505 91000 Evry, France N-7044 Trondheim, Norway
Phone: (+1) 888 863-8766 Phone: (+33) 1 6078 8899 Phone: (+47) 7382 8500
Fax: (+1) 616 454-3350 Fax: (+33) 1 6078 6839 Fax: (+47) 7391 7009
E-mail: support.us@metron-biomed.com E-mail: : info@metron.fr E-mail: support@metron.no
Disclaimer
METRON provides this publication as is without warranty of any kind, either express or implied,
including but not limited to the implied warranties of merchantability or fitness for any particular pur-
pose. Further, METRON reserves the right to revise this publication and to make changes from time to
time to the content hereof, without obligation to METRON or its local representatives to notify any per-
son of such revision or changes. Some jurisdictions do not allow disclaimers of expressed or implied
warranties in certain transactions; therefore, this statement may not apply to you.
Limited Warranty
METRON warrants that the QA-VTM Ventilator Testing System will substantially conform to pub-
lished specifications and to the documentation, provided that it is used for the purpose for which it was
designed. METRON will, for a period of twelve (12) months from date of purchase, replace or repair
any defective system, if the fault is due to a manufacturing defect. In no event will METRON or its local
representatives be liable for direct, indirect, special, incidental, or consequential damages arising out of
the use of or inability to use the QA-VTM Ventilator Testing System, even if advised of the possibility
of such damages. METRON or its local representatives are not responsible for any costs, loss of profits,
loss of data, or claims by third parties due to use of, or inability to use the QA-VTM Ventilator Testing
System. Neither METRON nor its local representatives will accept, nor be bound by any other form of
guarantee concerning the QA-VTM Ventilator Testing System other than this guarantee. Some jurisdic-
tions do not allow disclaimers of expressed or implied warranties in certain transactions; therefore, this
statement may not apply to you.
ii
TABLE OF CONTENTS
iii
6.3 Low Range Pressure Calibration ............................................................................................ 6-2
6.4 High Range Pressure Calibration ........................................................................................... 6-3
6.5 Temperature Calibration ................................................................................................. 6-3
6.6 Oxygen Calibration ................................................................................................................ 6-4
6.7 Set time and date .................................................................................................................... 6-5
iv
5.1 Physical Setup and Connections ............................................................................................ 5-1
5.2 Leakage Test .......................................................................................................................... 5-2
5.3 Low Pressure Test .................................................................................................................. 5-2
5.4 Flow Test ............................................................................................................................... 5-3
5.5 High Pressure Test ................................................................................................................. 5-4
v
MANUAL REVISION RECORD
This record page is for recording revisions to your QA-VTM User & Service Manual that havehas
been published by METRON AS or itsit’s authorized representatives. We recommend that only the
management or facility representative authorized to process changes and revisions to publications:
make the pen changes or insert the revised pages;
ensure that obsolete pages are withdrawn and either disposed of immediately, or marked as su-
perseded and placed in a superseded document file, and;
enter the information below reflecting that the revisions have been entered.
Rev No Date Entered Reason Signature of Person Entering Change
vi
This page intentionally left blank.
vii
1. Introduction
1.1 Features
The QA-VTM Testing System is a compact, lightweight high perfor-
mance testingperformance-testing device designed to be used by
trained service technicians for verifying the performance of infant,
pediatric, and adult mechanical ventilators. Ventilation parameters,
including breath rate, inspiratory and expiratory timing, flow rate, tid-
al volume, oxygen concentration, and airway pressure can be checked.
Unlike any other tester, QA-VTM makes it easy to add a test load
(e.g., lung simulation) to the test system without corrupting the re-
sults. Flow, high and low range pressure, and leakage rate measure-
ments can also be made independently.
For greater automation and enhanced viewing of the test data, you
may want to utilize the PRO-Soft QA-VTM test software, available
separately. With PRO-Soft you can remotely control the unit, create
and save unique test sequences for each brand or model of ventilator,
perform automated tests, and print the results and/or store them to
disk. PRO-Soft also allows viewing of pressure, volume, and flow
wave formswaveforms. These can be useful in diagnosing problems or
validating a ventilator's own on-screen display.
1.2 Specifications1
1. Inputs
Flow Range: 0.1 to 160 LPM, ± 2% of reading
Pressure
Low range: : - 40.0 to 350 mmHg, 1% of reading
± 1 LSDscale
High range: : 0 to 5000 mmHg, 0.1% of scale
Temperature: +10 to 50 °C, ± 1 °C
Oxygen level: 0 to 100%, ± 5 % of reading
2. System Setup
Reference unit: BTPS, ATPX, STPD, ATPS,
ATPD, and NTPD
Pressure unit: cm H2O, mBar, mmHg, kPa, psi
Flow unit: l/min, l/s, mLl/s
Minute volume: LL, mlL
1-1
Temperature unit: C, F, K
Gas mixture: Air, O2, O2/N2, O2/N2O
Number of samples
before calculation: 1, 2, 4, 8, 16 and 32
Noise: 0 to 9996 mL/min
Detect level: 0 to 9996 mL/min
Resistance unit: cm H2O / l/s, mBar/ l/s
Resistance level: 5, 20 and 50 ohm
Atmos. pressure: 0 to 2000 mmHg
Relative humidity: 10 to 99%
Suppress pressure
peak period: 0 to 100 ms
The unit may be connected to a test lung for compliance setting.
1-2
Weight: 3.5 kg / 7.7 lb.
Dimensions: Depth: 27.0 cm27.0 cm / 11.0/ 11.0 in.
Width: 35.0 cm35.0 cm / 13.5/ 13.5 in.
Height: 9.5 cm9.5 cm / 4.0/ in4.0 in.
Temperature
Formatted
Requirements:
+15/59 to +35/95 °C / F while operating
0/32 to +50/122 °C / F for storage
Part No: QA-VTM Ventilator Testing System (P.N. 16200)
Standard Accessories: QA-VTM User and Service Manual (P.N. 16225)
QA-VTM Carrying Case (P.N. 16210)
QA-VTM Tubing kit (P.N. 16250)
Optional Accessories: Test lung QA-VTM (P.N. 16240)
PRO-Soft QA-VTM (P.N. 16220)
PRO-Soft QA-VTM User Manual (P.N. 16230)
1-3
This page intentionally left blank.
1-4
2. Installation
2-1
2.2 Setup
Equipment connection is as shown in the typical setup below. For di-
rect communication to printer attach the printer cable to the 25-pin
outlet port on the QA-VTM.
NOTE
Some RS-232C cables are
missing the connection between
the seventh and the eighth wires
in the cable. The cable may still
be called NULLcalled
NULL-modem, but it will not
work with the QA-VTM. Refer to
the PRO-Soft QA-VTM Users
Manual for more information.
2.3 Accessories
1. The Tubing Kit (P.N. 16250) allows you to conduct the QA-
VTM’s testing on a variety of ventilators and other respiratory
care equipment:
Disposable adapter Male – Male adapter Flex Hose Connector
2-2
High Pressure Input SideportSide port Flow-Pressure
Connector Connector
2. The Adult Test Lung Kit (P.N. 16240) allows you to simulate a
load on a variety of ventilators and other respiratory care equip-
ment during testing.
Adult Test Lung Flex Hose Connector
2-3
This page intentionally left blank.
2-4
3. Operating QA-VTM
3-1
4 Flow Port 15 mm ID / 22 mm OD port for attachment of the
breathing circuit or flow tubing.
5. Low Pressure Port Connection port for measuring low range pressure
(-40.0 to 350 mmHg) during ventilation or inde-
pendent pressure measurements. Note: Port
adapters are furnished in Tubing Kit (P/N 16250).
Rear Panel
QA-VTM’s display has four lines in each screen. The top three lines
are used for entry of testing parameters by the operator and for mes-
sages, and test status and results by the QA-VTM. These are con-
3-2
trolled by the F5, F6 and F7 keys, located to the right of the display.
These keys are operational only when you see the leading arrows (“
. . . . . .>”) pointing to the keys.
The bottom line is a menu bar, controlled by function keys F1 through
F4 directly below the display. It is used for system functions, such as
PRINT or TABULATE, and for inter-screen navigation, such as
MORE and GO BACK.
Pressing and holding down one of the 11 alphanumeric keys causes it
to move automatically from character to character. For example,
pressing the “1 ABC” key and holding it down will scroll you auto-
matically through “1”, then “A”, “B” and “C”. Release the key when
the desired alphanumeric character appears. To delete a letter or
number previously entered, use the Delete key.
When all data has been entered on a particular line, press the Enter
key to exit the line. When this is done the data on the line is retained
until changed or the device is turned off. To avoid loss of data, store it
in memory. (See paragraph 3.6 below)
The Set Zero Screen is used to set the zero (0.00) value Ventila-
tor Inlet Pressure and Flow Ports, located to the right of the dis-
play.
3-3
Pressing F5 sets the pressure to zero, and F6 sets the flow to zero.
Upon completion of these adjustments, press MAIN MENU (F4)
3-4
of measurement, and store them for use in all testing, and test results
printouts. There are nine screens in Setup.
NOTE
Press Noise level (F7). Use the alphanumeric keys to enter Formatted: Bullets and Numbering
METRON recommends that the level, in milliliters per minute, that QA-VTM will use in
noise and detect levels be set at
500 and 1000, and adjusted if
erroneous information is being
included in the calculations.
3-5
testing. Press Enter to save. This setting sets the threshold
that filters out noise that is not desired to be incorporated in
the calculations (e.g., breathing circuit vibration noise).
Press No. of Breaths for Calculation (F6) to specify the
number of breaths to be averaged for calculating all displayed
values. Available selections are 1, 2, 4, 8, 16 and 32.
Press Suppress Pressure Peak period (F5) to specify, in
milliseconds, the suppress pressure peak period. Available se-
lections are 10-100 milliseconds, in increments of 10mS.
Note: This setting is used to filter out the pressure spikes that
may be seen at the beginning of inspiration. This spike is not
usually clinically relevant, nor considered the Peak Inspirato-
ry Pressure.
Press Reference unit (F6) to select the gas volume reference
NOTE BTPS
unit that Body temperature,
QA-VTM will useambient pressure,
in testing. saturated
Selections are, in
Most ventilators are calibrated ATPX Ambient temperature, ambient pressure, actual
for BTPS and ATPX conditions. STPD Standard temperature, standard pressure, dry
Consult your ventilator opera-
tor’s manual. ATPS Ambient temperature, ambient pressure, saturated
ATPD Ambient temperature, ambient pressure, dry
NTPD Normal temperature, normal pressure, dry
3-6
(cmH20), millibars (mBar), millimeters of mercury (mmHg), kilopascal (kPa), and pounds per square
inch (psi).
Press MORE (F1) to advance to Setup Screen 3.
3-7
Press Gas Mixture (F7) to specify the gas mixture used in
testing. Selections are, in sequence, Air, 02, O2/N2, and
O2/N20.
Press External Resistance used (F6) to specify the resistance Formatted: Bullets and Numbering
level to be used in testing. Available selections are None, 5,
20, 50, 200 and 500, as these airway resistors are commer-
cially available.
Press Resistance unit (F5) to specify the resistance unit that
QA-VTM will use. Available selections are cmH20 / l/s or
mBar / l/s. Note: Setup is for reference purposes only, should
an external airway resistor be used in the setup.Press either
"Y" or "N" in Acoustic Alarm for failed test (F6) to specify
whether or not this feature is desired.
NOTE
The default setting in Calculate
Press either "Y" or "N" in Calculate Mass Flow (F5) to
Mass Flow is “Y”, or Yes, as specify whether or not you desire to have QA-VTM calculate
ventilator manufacturers will flow in this way.
reference their flow and volume
settings to atmospheric settings,
regardless of the physical vol- Press MORE (F1) to advance to Setup Screen 5.
ume occupied by the gas in the
patient’s airways and lungs.
3-8
Press Flow unit (F7) to specify the unit of flow to be used in
testing. Available selections are liters per minute (L/Min), li-
ters per second (L/S), and milliliters per second
Formatted: Bullets and Numbering
Press Pressure unit (F6) to specify the unit of pressure used Formatted: Bullets and Numbering
to be used in testing. Selections are centimeters of water
(cmH20), millibars (mBar), millimeters of mercury (mmHg),
kilopascal (kPa), and pounds per square inch (psi).
Press Temperature unit (F5) to specify the temperature unit
to be used in testing. Available selections are Celsius, Fahr-
enheit and Kelvin.
Press MORE (F1) to advance to Setup Screen 6.
Press Operator (F6). Use the alphanumeric keys to enter the Formatted: Bullets and Numbering
test operator’s name or other identifying data. Press Enter to
save.
3-9
Press Establishment (F5). Use the alphanumeric keys to en-
ter the test facility’s name or other identifying data. Press En-
ter to save.
Press Atm pressure (F6) to specify the atmospheric pressure,
in millimeters of mercury (mmHg), to be used in testing.
Press Relative humidity (F5) to specify the relative humidity (10 - 99 percent) to be used in testing. Formatted: Bullets and Numbering
Press MORE (F1) to advance to Setup Screen 7.
3-10
Setup Screen 8. This screen (above) contains the following settings:
Press Adjust O2 measurements to go to the O2 Adjustment Formatted
screen shown below.
Formatted: Bullets and Numbering
Press QA-VTM Serial no. (F7). Use the alphanumeric keys
to enter the QA-VTM’s serial number. Press Enter to save.
QaVtm Application SW rv XX.XX. This shows the current Formatted
software version in the QA-VTM.
Formatted: Bullets and Numbering
If you desire to store the parameters in QA-VTM’s flash
memory, press STORE (F1). If you do not press STORE, Formatted
the QA-VTM will function with the settings you changed un-
til the unit is powered off. Language (F7). This is factory in-
stalled. Default is English.Press Date (F6). Use the alphanu-
meric keys to set, or reset, the system date (DD/MM/YYYY).
Press Enter to save.
NOTE Press Time (F5). Use the alphanumeric keys to set, or reset,
When “24 Hour” shows above the system clock (HH:MM:SS). Press F3 to specify the a 12-
F3, the system is operating on a hour (Am/Pm) or 24-hour (24 hour) clock
12-hour time schedule. When
showing “Am/Pm” above F3, the
system is operating on a 24- Press MORE (F1) to advance to Setup Screen 9.
hour time schedule.
3-11
Press Flush Air (F6) when you are flushing the VTM with air
at 21% O2.
Flush Oxygen (F5) after you have flushed the unit with 100%
O2 for a few minutes.If you desire to store the parameters in
QA-VTM’s flash memory, press STORE (F1).
Setup Screen 190. If you pressed STORE (F1) in Setup Screen 9 this
screen (above) notifies you that the parameters have been stored
in flash memory.
3.6 QA-VTM Memory
Test results for a device under test are compiled when the TAB-
ULATE (F4) is pressed in the various test screens. These can then
be managed within QA-VTM’s memory.
Pressing MEMORY (F2) in the Main Menu advances you to a se-
ries of memory screens. These enable you to store, retrieve, trans-
fer, print and delete tests and test results to and from the QA-VTM
memory. The four memory displays are:
NOTE
If you intend to store all the test
results for the equipment under
test be sure to return to the Main
Menu screen by pressing GO
BACK (F2). Press MEMORY
(F2) and, in Memory Screen 1,
press Store test in memory
(F6).
3-12
The F5 and F6 function keys generate new displays as confirma-
tion of an executed function or error message.
Press GO BACK (F2) to return to the previous memory screen.
Print Test Footer (F6). Selecting this prepares a Report Footer for
printing. The footer is used for operator comments and signature.
Form Feed (F5). Pressing this advances the page, currently being
printed, in the printer.
3-13
3.8 Upgrading QA-VTM
Firmware
To install new firmware, use the following procedure. Note that the
QA-VTM upgrade contains two files: com[xxxx].exe, which establishes
communication between a PC computer and the QA-VTM, and;
[xxxx].a07, which is the actual QA-VTM upgrade file. NOTE:
“[xxxx]” indicates the version of the firmware
1. Prepare the QA-VTM for the software upgrade.
Press and hold the firmware upgrade keys on the keyboard (see
left) at the same time while turning the QA-VTM ON.
The QA-VTM display will show the following:
3-14
4. Test to confirm the upgrade.
Exit the DOS program and unhook the RS-232 transfer NULL
Modem cable from the QA-VTM.
Switch the QA-VTM OFF, thenand then turn it back ON.
Verify that the new firmware version is installed while the QA-VTM
is running its start sequence. The firmware version will be displayed
in Startup Screen Menu 3.
3-15
4. Ventilator Testing
4-1
4. Optional: To simulate a load during testing, attach the Test Lung
(P.N. 16240) to the outlet port at the rear of the QA-VTM (see
above).
NOTE
If you intend to store all the test
results for the equipment under
test be sure to return to the Main
Menu screen by pressing GO
BACK (F2). Press MEMORY
(F2) and, in Memory Screen 1,
press Store test in memory
(F6).
4-2
4.3 Other Parameter
Testing
As another option, when all template parameters are not required, in-
dividual ventilator parameters can be tested. Parameter selections are:
Rate Volume Pressure Environment
Time Flow CPAP Measurements
1. Prepare the equipment as discussed in paragraph 4.1 above.
2. From the Main Menu scroll through Test Mode (F6) until the
Ventilator Test appears.
4-3
Time Parameter
Volume Parameter
Flow Parameter
Pressure Parameter
Environment Parameter
4-4
5. Other Testing
5-1
5.2 Leakage Test
1. From the Main Menu scroll through Test Mode (F6) until the
Leakage Test appears. Press START (F4). The screen changes
to the Leakage Test screen, and testing begins. (See below)
5-2
2. Test measurement results are shown on the screen. The pressure
unit used in the measurements is as specified during Setup (see
Setup Screen 2). Available selections are centimeters of water
(cmH20), millibars (mBar), millimeters of mercury (mmHg),
kilopascal (kPa), and pounds per square inch (psi).
3. To reset the displayed values and restart testing, press RESET
(F1). To readjust the flow and pressure to 0O press Set Zero
(F7).
4. You can then either print the test results by selecting PRINT (F3),
or store them to memory by pressing TABULATE (F4).
5. Return to the Main Menu screen by pressing GO BACK (F2). If
you intend to store all the tests results for the equipment under
test be sure to press MEMORY (F2) and, in Memory Screen 1,
press Store test in memory (F6).
5-3
4. You can then either print the test results by selecting PRINT (F3),
or store them to memory by pressing TABULATE (F4).
5. Return to the Main Menu screen by pressing GO BACK (F2). If
you intend to store all the tests results for the equipment under
test be sure to press MEMORY (F2) and, in Memory Screen 2,
press Store test in memory (F6).
5-4
6.Testing and Maintenance Formatted: Bullets and Numbering
WARNING!
HIGH VOLTAGES ARE CAPABLE
OF CAUSING DEATH!
USE EXTREME CAUTION WHEN PERFORMING TESTS AND CAL-
IBRATION. USE ONLY INSULATED TOOLS WHEN THE UNIT IS
PLUGGED IN, AND THE CASE HOUSING IS OFF.
6.2Testing Procedures
1. Test Measurement Board before Mounting. The ultrasonic
sensors are of slightly different characteristics, and have to be
matched before mounting in the tube.
a.Connect the board to be tested to the power in the test setup. Formatted: Bullets and Numbering
b.Check voltage levels.
c.Check output from the regulators. ( +12V , +5V, -12V ).
5-1
b.Connect channel 2 to TP703, channel 3 to TP705 and channel 4
to TP706.
c.Move the trigger point about 1ms so that the signals on chan-
nels 2-4 are visible. The signals must look like the figure in
must look like the figure in chapter 8. The signal on TP706
must be stable, and not jump back and forth to different posi-
tions.
11-Dec-98
0 35 19
5 s
2.00 V
5 s
.2 V DC
.5 V DC 250 MS/s
50 mV DC DC 3.64 V
.5 V AC AUTO
g.Plot 02. This plot shows receiver signal on TP700. This is typi-
cal response from an ultrasonic transducer in gas.
11-Dec-98
0 36 57
.1 ms
1.00 V
.1 ms 5.75 ms
.2 V DC
.1 V DC 50 MS/s
50 mV DC DC 3.64 V
.5 V AC AUTO
5-2
h.Plot 03. This plot shows the receiver signal after an incomplete
envelope detection. TP701 and TP702 show the same signal.
11-Dec-98
0 39 52
20 s
1.00 V
20 s 848 s
.2 V DC
.1 V DC 250 MS/s
50 mV DC DC 3.64 V
.5 V AC STOPPED
i.Plot 04. This plot shows the receiver signal on TP703 after pass- Formatted: Bullets and Numbering
ing through a comparator. The signal goes to the D-flip-flop.
flop.
11-Dec-98
0 40 46
20 s
1.00 V
20 s 848 s
.2 V DC
.1 V DC 250 MS/s
50 mV DC DC 3.64 V
.5 V AC STOPPED
5-3
j.Plot 05. The plot shows the signal on TP704. This is source to
the zero level trigger.
11-Dec-98
0 41 12
20 s
1.00 V
20 s 848 s
.2 V DC
.1 V DC 250 MS/s
50 mV DC DC 3.64 V
.5 V AC STOPPED
k.Plot 06. The Plot shows the signal on TP705. This is the signal
from plot 05 after the comparator. The signal goes to the
clock input on the D-flip-flop.
11-Dec-98
0 41 34
20 s
1.00 V
20 s 848 s
.2 V DC
.1 V DC 250 MS/s
50 mV DC DC 3.64 V
.5 V AC STOPPED
l.Plot 07. The plot shows the inputs and the output on the D-flip-
flip-flop. Channel 2 is the data input on TP703. Channel 3 is
5-4
put on TP703. Channel 3 is the clock signal on TP706. Chan-
nel 4 is the signal on TP706 to the counter.
11-Dec-98
0 53 10
20 s
5.0 V
20 s
5.0 V
20 s
5.0 V
20 s 878 s
.5 V DC
.5 V DC 250 MS/s
.5 V DC DC 3.6 V
.5 V AC AUTO
m.Plot 08. The plot shows the transmit pulse on TP600 and re-
ceiver pulse on TP706.
11-Dec-98
0 56 37
.2 ms
5.0 V
.2 ms
10.0 V
.2 ms
1 V DC
.5 V DC 25 MS/s
.5 V DC DC 3.6 V
.5 V AC AUTO
n.Plot 09. The plot is the same as plot 07, but with a different time
base. Here we can see the transmit pulse on channel 1. Chan-
nel 2 shows signal on TP703, channel 3 shows signal on
TP705 and channel 4 shows signal on TP706.
5-5
11-Dec-98
0 58 56
.2 ms
5.0 V
.2 ms
5.0 V
.2 ms
5.0 V
.2 ms
10.0 V
.2 ms
1 V DC
.5 V DC 25 MS/s
.5 V DC DC 3.6 V
.5 V AC AUTO
o.Plot 10. The plot shows signals on TP600 (Transmitter) and on Formatted: Bullets and Numbering
on TP700. We can see that we get a signal on receiver im-
er immediately after the transmit signal. This is from the
switch, and is filtered away in software. After about 1 ms,
the receive signal appears. This is used to stop the counter.
11-Dec-98
1 00 54
.2 ms
1.00 V
.2 ms
5.0 V
.2 ms
.5 V DC
.5 V DC 25 MS/s
.1 V DC DC 3.6 V
.5 V AC AUTO
5-6
4. Test of Pressure Measurement Signals – Low Range
a.Let the pressure sensor input be open. Measure voltage on Formatted: Bullets and Numbering
TP800 in respect to GND. Adjust RP801 to the value on the
digital multimeter is 0.55 V.
b.Connect the pressure reference to the input. Apply 0.500 Bar.
Adjust RP800 to the value on the digital multimeter is 4.95
V.
c.Remove pressure, and check that the voltage is 0.55V. If not,
repeat step 1 to 3.
d.Apply –0.054 Bar and check that the voltage is > 0.00V.
voltage is > 0.00V.
5.Test of Pressure Measurement Signals – High Range
5-7
8. Reprogramming of Software in Measurement Board. If the
measuring processor board has lost the program or you need to
install a new program please follow this procedure.
a.Switch the QA-VTM unit off Formatted: Bullets and Numbering
6.3Troubleshooting Tech-
niques
1. Erroneous or Missing Flow Measurement In case of erroneous
or missing flow measurements, use the following procedures to
locate the error. All procedures assume that the QA-VTM unit
has been switched ON for about five minutes.
a.Ultrasonic sensor system errors. Check the two software cor- Formatted: Bullets and Numbering
rection LEDs, located on the left side of the top edge of the
measuring processor card. If one or both LEDs are lit, or
flashing, it indicates that the software cannot find reliable
measurement data.
b.Wrong trigger level. Locate the trigger level adjustment pot
level adjustment pot (close to the temperature sensor con-
nector) and try to adjust to a position where the LEDs are
neither flashing nor lit. Apply airflow through the QA-VTM
VTM and check out that the LED's are dark within airflow
from 0 to 160 L/M. Note that some flashing in the area 100-
100-160 l/M must be allowed. If too much activity by the
too much activity by the LED's try to adjust the pot. After
successful adjustment the unit must be recalibrated.
c.Sensor Error. Check that ultrasonic sensors and cables are not
damaged.
d.Circuit Error. Use the described test procedures to check that
the measurement logic circuit is OK.
2.Erroneous or Missing Pressure Measurement In case of wrong
wrong or missing pressure measurements, check the following
5-8
microcontroller AD-converter ref. voltage test points.
test points.
a.Check that pin 51 on microcontroller measures 5V and is sta-
ble.
b.Check the zero and gain adjustments by measuring the voltage
on test pins described in the test procedures.
c.Check the pressure sensor by measuring the output voltage di-
rectly from the sensor while applying different pressures to
the sensor.
3.Erroneous or Missing Oxygen Measurement In case of wrong or
ment In case of wrong or missing oxygen measurements check
oxygen measurements check the following microcontroller AD-
converter ref. voltage test points.
5-9
6.5O2 Fuel Cell Formatted: Bullets and Numbering
Replacement
CAUTION!
5-10
7.6. CaliALIBRATINGbrating QA-VTM
WARNING!
HIGH VOLTAGES ARE CAPABLE
OF CAUSING DEATH!
USE EXTREME CAUTION WHEN PERFORMING TESTS AND CAL-
IBRATION. USE ONLY INSULATED TOOLS WHEN THE UNIT IS
PLUGGED IN, AND THE CASE HOUSING IS OFF.
U
S
7.16.2 FFlow Calibration
E Formatted: Bullets and Numbering
1. Enter the Calibration Menu Screen.
E
2. Press MORE (F1) or GO BACK (F2)Xuntil you are at the correct
T
parameter. (FLOW) R
E
3. Close the flow inlet with a finger, so that
M
the airflow through the
sensor tube is zero. E
C
A
U
T
I
O
N
6-1
W
H
E
N
P
4. Press F7 to set zero level.
5. Apply a known airflow between 60 and 90 l/m through the tube.
Enter this value as the reference value by pressing F6, dial the
numbers representing the value and end with the enter key.
(10000 ml/10/s = 60 l/m).
6. When your calibrated airflow is steady at this level, press F5 to
calibrate the unit.
7. After successful calibration, restart QA-VTM by switching the
unit off and then on. Check/adjust zero level and make measure-
ments according to the following table. Fill in the QA-VTM read-
ings in the measurement protocol.
Flow sensor
Reference Measured value Tolerance
0 L/M 0.0 0 – 160 L/M +/- 32 % of full s
calereading
10 L/M 9.5
50 L/M 49.7
100 L/M 100.8
150 L/M 151.5
6-2
-54 mBar -54.0 -54 – 500 mBar +/- 1 % of full
scalereading
0 mBar 0.0
100 mBar 100.0
250 mBar 250.0
500 mBar 500.0
6-3
6. When your calibrated temperature is steady, press F5 to calibrate
the unit.
7. Press F5 to enter to calibrate the unit.
8. After a successful calibration, restart the QA-VTM by switching
the unit off and then on. Make measurements according to the
following table and fill in the QA-VTM readings in the measure-
ment protocol.
Temperature sensor
Reference Measured value Tolerance
22.8 ºC 22.7 ºC 10-50 ºC +/- 21 % of full scaledegree
*The temperature sensor has been calibrated for the value shown above.
No further measuring points are provided. Stability of reading has been
verified by repeated measurements.
a. a. Unplug the connector from the back of the O2 fuel cell. Formatted: Bullets and Numbering
b.
Formatted
b. Unscrew fuel cell from the manifold and dispose of it.
a. Remove new fuel cell from it packaging. Verify presence of
gasket and screw it into place. (**CAUTION** DO NOT
CROSS THREAD OR OVER TIGHTEN.)
c. d.
b.d. Reconnect fuel cell cable. Formatted: Bullets and Numbering
6-4
Oxygen sensor
Reference Measured value Tolerance
0 % Sensor exposed 0.0 0-100 % O2 +/- 5 % of full
to 100% N scalereading
21 % Sensor ex- 21.0
posed to air
*The measuring circuit has been tested with a 0V input Voltage and a 0%
reading has been verified in the display instead of use of Nitrogen
Press Date (F6). Use the alphanumeric keys to set, or reset, the
NOTE system date (DD/MM/YYYY). Do not use separator characters.
When “24 Hour” shows above Press Enter to save.
F3, the system is operating on a
12-hour time schedule. When Press Time (F5). Use the alphanumeric keys to set, or reset, the
showing “AMm/PmM” above
F3, the system is operating on a
system clock (HH:MM:SS). Do not use separator characters.
24-hour time schedule. Press F3 to specify a 12-hour (AM/PM) or 24-hour (24 hour)
clock.
6-5
This page intentionally left blank.
6-6
8.7. Component Functions and Parts Formatted: Bullets and Numbering
7-1
about 4 V. R714 and R715 give the circuit some hysteresis to avoid
switching on noise. D705 and D706 are limiter diodes.
The signal from U702-B is now a square pulse, and is fed to the data
input on U704. This is a D-flip-flop. In this circuit the signal is syn-
chronized with the zero crossings from the receiver transducer.
The other part of the receiver input signal goes to the zero detector.
The signal is fed U703-A, and is amplified. R719 and R720 set gain.
D707 and D708 are limiting diodes.
From here the signal is fed to U703-B, which is used as a comparator.
R724 and R725 set the reference level. This is a few mV. R722 and
R723 gives the comparator some hysteresis. D709 and D710 are lim-
iting diodes.
The signal is now a square pulse, with a frequency of about 40kHz,
and is fed to the clock input U704.
U704 is a D flip-flop. The data input comes from the level detector,
and the clock input from the zero detector. In addition to this, a nega-
tive pulse from the microcontroller, CLEAR, to the reset input, is
used to reset the output before the start of the measurement cycle. The
output on the D-flip-flop is set by the data input, and synchronized by
the clock input.
The digital signal goes to U705, an AND gate. The other input to the
AND gate is the enable signal (REC_EN) from the microcontroller.
This enable signal must be logic H to stop the counter in the digital
part of the flow sensor. Enable is set high just before the received sig-
nal enters the gate. This to eliminateprevent noise from stopping the
counter.
The flow sensor digital part is measuring the time, from sending a
pulse on the transmitter to the moment the received pulse is detected.
It consists of a 4 bit counter, U500, with surrounding circuits. U500 is
clocked by an oscillator, with clock frequency of 32 MHz. U502 is a
quad SR-latch, that is used to convert pulses to levels. U501 is a 4 bit
counter, that is used to stretch the pulse from the 32 MHz clock, so
that the pulse width is long enough so that the microcontroller can de-
tect it. The time delay between transmitting a pulse, and receiving the
pulse, is about 0.8 to 1.0 ms. With the 32 MHz clock, we will need 15
bits resolution. U500 represents the four least significant bits. The
eleven most significant bits are handled by the microcontroller.
Before each measurement sequence, the counters are initialized,
U500 is cleared, and U502-A set high. This leads to that ENP on
U500 is high, U502-B is reset, which leads to that ENT on U500 goes
high. The init pulse comes from the microcontroller. The microcon-
troller then sends out a pulse T (Transmit pulse). This goes to the
transmitter and counter. This then goes to the reset input on U502-B,
7-2
and sets this high. The output goes high, and will stay high. This is fed
to ENT on U500, and the counter starts.
When U500 has counted to 0F Hex (1111), the RCO (overflow/carry)
signal activates. This pulse is too short for the microcontroller to de-
tect it. It is fed to U501. RCO is inverted in U503-A, and is used to
clear U501. The RCO signal is inverted twice, because of timing, and
fed to the set input U502-C. The output of this goes high and is fed to
the microcontroller. At the same time, the signal from the output of
U502-C is fed to the ENT and ENP on U501, and starts the counter.
When the counter has counted to 04 Hex (0100), the QC output on
U501 goes high. This is sent to the reset input of U502-C and the out-
put goes high, and stops U501. The signal to the microcontroller goes
high. At this moment, the microcontroller has a long enough pulse to
be detected. This happens each time U500 outputs a RCO pulse.
The time from pulse T is sent ,sent, and to pulse R (received pulse) is
received is in a time interval decided by the physical dimensions of
the sensor tube and the measurement. This means that we have a min-
imum time delay. Before this minimum time runs out, it will not be
possible to stop the counter. This is done by using the AND gate
U705-A. Input to this gate is pulse R, and the enable signal REC_EN.
This enable signal is set HIGH, just before the minimum time has
elapsed. The clear signal into D flip-flop (U705) is set just before
REC_EN goes HIGH.
When pulse R comes to the reset input on U502-A, the output goes
HIGH, and ENP on U500 goes HIGH and stops the counter. At the
same time, the signal goes to the microcontroller and tells it that the
sequence is over. The microcontroller reads the four bits from the
counter. Together with the eleven bits from the internal counter, it has
a binary number representing the time delay from sending a pulse to
the pulse is received. The measurement is done in both directions. In
this way, we get a time difference, which is used to calculate flow.
The measurement is repeated each 5 ms.
7-3
8.37.3 Temperature measurement Formatted: Bullets and Numbering
To monitor temperature, the sensor LM35 is used. This sensor outputs
a voltage level proportional to the Celsius scale. 0 degrees C gives 0
mV output, and 50 degrees C gives 500 mV output. LM35 is connect-
ed to the board by a three-wire cable connected to J830. The signal is
fed to aan amplifier, with a gain of 20 dB. This is done in U830, with
surrounding circuits. From the amplifier, the signal goes to the ADC
on the microcontroller. The level is in the 0-5 V range.
7-4
The internal timer/counter is used, together with the external tim-
er/counter, to measure the time delay between transmitting and
receiving a pulse through the flow tube. The time delay is repre-
sented by a 15-bit binary number, where the internal tim-
er/counter counts the eleven most significant bits, and the exter-
nal timer/counter counts the four least significant bits. After each
measurement sequence is complete, the four bits form the exter-
nal counter is transferred to the micro controller through port G
[0.3].
The on-chip ADC is used for measurement of pressure, tempera-
ture and oxygen level in the air passing through the unit. The ana-
log DC levels from pressure, temperature and oxygen are fed into
AN <2..0> on the ADC. The ADC has a 5.0 V reference on VRH
and GND on VRL.
Communication with the front processor board is done by the
synchronous serial interface SPI. The SPI bus consists of four
signals: MISO, MOSI, SCK and SS. The front CPU acts as mas-
ter, and the measurement board is slave.
A Motorola HC 16 microcontroller controls the measuring pro-
cessor board. The board has 256 Kb of Ram and 512 Kb of Flash
ROM used for software and calibration data.
The processor board calculates the following parameters and pre-
sents the data for the main processor, 100 times per second.
Airflow in 1/10 ml/S.
High range pressure in cmH2O.
Low range pressure in mmH2O.
O2 level in 1/10 %.
Temperature in 1/10 Degrees C.
2. Main Processor Board
The processor system is divided into five function blocks: (1)
CPU; (2) memory; (3) display and printer interface; (4) serial and
keypad interface, and; (5) keypad. The CPU and keypad board
are located behind the front panel. The unit comprises a proces-
sor system, a display, control components and a software con-
trolled SIP port for communication with the measuring sensor
board. The processor controls the measurement process in the
QA-VTM. Measured values are transferred digitally from the
measurement board before the data stream is processed and dis-
played by the processor board.
a. The CPU comprises a Motorola 68HC11 operating at 8
MHz, which gives a BUS frequency of 2 MHz. The UART
in µP is used only during boot-up. A Maxim µP supervisory
7-5
circuit, with a 4.65 volt reset, is used to monitor the 12 Volt
power, and ensure battery power to RAM when the unit is
switched off. U1080 and DS2404 are timer circuits. These
provide the system with real time clock and calendar infor-
mation. U1160 is an address decoder. U1120 is an address
latch for the multiplex address bus. µP has a 64 KB address
range, and U1130, together with U1150, are used as a bank-
switch to address the process circuit's RAM and flash-ROM.
b. The memory component comprises 128 KB RAM, and either
a 28F200 or a 28F400 Flash-ROM for storing programs.
The CPU module is programmable, and may be repro-
grammed with new software supplied by Metron. Transis-
tors in the memory circuit are used to supply the flash-ROM
with a 12 volt programming voltage.
c. The printer and display interface comprises PD71055, PIO
that is used to write to display and to the Centronics output.
Data to Centronics goes via LS05, which has an open col-
lector output. A strobe pulse to the Centronics port is estab-
lished through U1410.
d. Keypad scanning and the RS-232 interface are established by
U1300, DUART. The keypad is a standard keypad matrix.
U1310, MAX238 is an RS-232 transmitter/receiver between
DUART and the RS-232 port.
3. Power Supply. The measuring processor board and the main
processor board are both supplied from the same power supply.
The power supply has the following electrical characteristics.
Output Voltage Minimum load Maximum Load Peak Load
+5 V 0.4 A 4A 7A
+15 V 0.2 A 2A 3A
-15 V 0A 0.5 A 1A
Please note that both the measuring processor board and the main
processor uses 5V, +12 V and -12V, and that the +15V and -15V
voltages are only used as input for the +12V and -12 V regula-
tors, located on the measuring processor board.
7-6
PART DESCRIPTION QTY. COMPS.
NAME/NO.
1620004 HEX INVERTERS SO14 1 U503
1620008 QUAD 2 INPUT AND GATE SO14 1 U705
1620074 DUAL D-TYPE FLIP-FLOPS SO14 1 U704
7812 POS. 12V REG. TO220 2 REG100-101
7912 NEG. 12V REG. TO220 1 REG1
ADM202E RS232 DRIVER/RECEIVER SO16 1 U400
AM29F010 FLASH EPROM 128Kx8 4 U300-303
BAS16 SWITCHING DIODE SOT23 19 D700-710, D800-803, D830-831,
D860-861
C100nF 0805 100nF 10% 0805 64 C103-104, C202-209, C300-305,
C400-404, C500-503, C600-601,
C604, C700-701, C705-706, C708-
709, C712-717, C722-723, C800-801,
C803, C805-809, C811-812, C814-
817, C830-832, C860-864
C100pF 0805 100pF 10% 0805 2 C602-603
C1nF 0805 1nF 10% 0805 5 C703-704, C711, C802, C810
C22pF-0805 22pF 10% 0805 2 C200-201
C330pF- 330pF 10% 0805 1 C702
0805
1561 33uF 16V TANTAL 7343 12 C100-102, C105-111, C804, C813
C470pF- 470pF 10% 0805 5 C710, C718-721
0805
C47pF-0805 47pF 10% 0805 2 C504, C707
CON FRICTION LOCK MOLEX 6410 1X3 2 J830, J860
MOL6410
3MS
CON1MS SOLDER TAIL 1x1 MALE STRAIGHT 4 TP100-103
CON3.96FRI FRICTION LOCK HEADER 3.96 1X6 2 J102-103
CTLH1X6
MS
CONBOXH2 BOXHEADER 2x5 MALE 1 J101
X5MS
CONBOXH2 BOXHEADER 2x8 MALE 1 J104
X8MS
CONSMB- SMB COAX ANGLE 2 J600-601
COAXMA
CON- SOLDER TAIL 2x20 MALE STRAIGHT 1 J100
STIFTL2X20
MS 2mm
DSS306- EMI-FILTER 100V/6A 3 FL100-102
91FZ103N10
0
7-7
PART DESCRIPTION QTY. COMPS.
NAME/NO.
HM628128B HM628128BLT-7 2 U304-305
LT-7
LM4040 5V REFERENCE LM4040 CIM3-5.0 3 Z100-101, Z200
CIM3-5.0
LS S269 LED SOT23 3 D100, D200-201
MAX394CW QUAD ANALOG SWTCH SPDT SO20-30 1 U601
P
1670025 MC68HC16Z1CF16 QFP132 0.635mm 1 U200
MPX2700DP PRESSURE SENSOR 2 MT800-801
POT- POTMETER BI 25P 10K 1 RP700
BI25P10K
POTMUL- POTMETER MULTITURN 10K 6 RP800-803, RP860-861, R100K-0805
TI10K
100K 1% 100K 1% 0.1W 08056 6R70 R703, R816, R838, R863-864, R878
0.1W 0805 3,
R816
,
R838
,
R863
-864,
R878
R100R-0805 100R 1% 0.1W 0805 2 R814, R828
R10K-0805 10K 1% 0.1W 0805 63 R202-220, R222-234, R600-603,
R713, R718, R721, R726, R728,
R732, R803-806, R808, R815, R820-
823, R829, R834-836, R862, R865-
868, R871, R877
R10M-0805 10M 1% 0.1W 0805 1 R201
R1K-0805 1K 1% 0.1W 0805 15 R221, R235-236, R734-743, R800,
R817
R1M-0805 1M 1% 0.1W 0805 1 R708
R20K-0805 20K 1% 0.1W 0805 2 R729, R731
R240R-0805 240R 1% 0.1W 0805 1 R500
R2K-0805 2K 1% 0.1W 0805 6 R100-105
R2K7-0805 2K7 1% 0.1W 0805 2 R702, R705
R330K-0805 330K 1% 0.1W 0805 1 R200
R33K-0805 33K 1% 0.1W 0805 2 R711-712
R390K-0805 390K 1% 0.1W 0805 1 R723
R3K9-0805 3K9 1% 0.1W 0805 2 R700, R717
R470R-0805 470R 1% 0.1W 0805 4 R716, R727, R730, R733
R47K-0805 47K 1% 0.1W 0805 28 R701, R704, R706-707, R709-710,
R715, R720, R724-725, R801-802,
R807, R809-810, R818, R824-825,
7-8
PART DESCRIPTION QTY. COMPS.
NAME/NO.
R831, R833, R837, R839, R860-861,
R869-870, R872-873
R4K7-0805 4K7 1% 0.1W 0805 5 R714, R719, R722, R830, R832
R560R-0805 560R 1% 0.1W 0805 1 R300
RNC-0805 NC 1% 0.1W 0805 9 R811-813, R819, R826-827, R874-
876
SCO-060 OSCILLATOR 1 Y500
SWITCH SWITCH 12x12mm 1 SW200
12x12mm
1620110 OPAMP TL071 SO8 3 U600, U700, U830
1620072 DUAL OPAMP TL072 SO8 4 U701-703, U706
TL074 QUAD OPAMP TL074 SO14 3 U800-801, U860
TP SOLDER TAIL 1x1 MALE STRAIGHT 12 TP600, TP700-706, TP800-801,
TP830, TP860
7-9
This page intentionally left blank.
7-10
Appendix A - Menu Maps
A-1
A-2
A-3
A-4
Appendix B - Schematics
B-1
This page intentionally left blank.
B-2
USA FRANCE NORWAY
1345 Monroe NW, Suite 255A 30, rue Paul Claudel Travbaneveien 1
Grand Rapids, MI 49505 91000 Evry, France N-7044 Trondheim, Norway
Phone: (+1) 888 863-8766 Phone: (+33) 1 6078 8899 Phone: (+47) 7382 8500
Fax: (+1) 616 454-3350 Fax: (+33) 1 6078 6839 Fax: (+47) 7391 7009
E-mail: support.us@metron-biomed.com E-mail: info@metron.fr E-mail: support@metron.no
E-mail:
Date:
Error Report
Product: Version: Serial no.:
(METRON AS internally)
Comments:
C-1
USA FRANCE NORWAY
1345 Monroe NW, Suite 255A 30, rue Paul Claudel Travbaneveien 1
Grand Rapids, MI 49505 91000 Evry, France N-7044 Trondheim, Norway
Phone: (+1) 888 863-8766 Phone: (+33) 1 6078 8899 Phone: (+47) 7382 8500
Fax: (+1) 616 454-3350 Fax: (+33) 1 6078 6839 Fax: (+47) 7391 7009
E-mail: support.us@metron-biomed.com E-mail: info@metron.fr E-mail: support@metron.no
From: Phone:
(name)
Address: Fax:
E-mail:
Date:
Improvement Suggestion
Product: Version:
(METRON AS internally)
Comments:
D-1
USA FRANCE NORWAY
1345 Monroe NW, Suite 255A 30, rue Paul Claudel Travbaneveien 1
Grand Rapids, MI 49505 91000 Evry, France N-7044 Trondheim, Norway
Phone: (+1) 888 863-8766 Phone: (+33) 1 6078 8899 Phone: (+47) 7382 8500
Fax: (+1) 616 454-3350 Fax: (+33) 1 6078 6839 Fax: (+47) 7391 7009
E-mail: support.us@metron-biomed.com E-mail: info@metron.fr E-mail: support@metron.no
C-1