Don Wauchope2016

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 2

CLB-09266; No.

of pages: 2; 4C:
Clinical Biochemistry xxx (2016) xxx–xxx

Contents lists available at ScienceDirect

Clinical Biochemistry

journal homepage: www.elsevier.com/locate/clinbiochem

Commentary

Lot change for reagents and calibrators


Andrew C. Don-Wauchope ⁎
Department of Pathology and Molecular Medicine, McMaster University, 1280 Main Street West, Hamilton, L8S 4K1, Ontario, Canada
Hamilton Regional Laboratory Medicine Program, 50 Charlton Avenue East, Hamilton, L8N 4A6, Ontario, Canada

A R T IC L E IN F O practice and avoid the commutability limitations of quality control ma-


terials [3]. Quality control (QC) material is another sample type that
Article history: could be considered once the described limitations have been consid-
Received 8 March 2016 ered and tighter targets applied [4]. The patient samples are chosen ac-
Received in revised form 29 March 2016 cording to expected patient values. Typically we would select three low
Accepted 5 April 2016
range patient samples, three mid-range patient samples and three high
Available online xxxx
range patient samples. One additional sample that could be an extreme
Keywords: value that tests either limit of detection or upper linearity is also select-
Lot change ed. These samples would provide a good range of results that should
Laboratory quality evaluate the assay over the full analytical spectrum. In addition to the
Quality control
patient samples we run each QC level in triplicate to estimate QC perfor-
mance when we change lot.
1. Introduction
2.2. Running the samples
Evaluating lot change is an important part of clinical laboratory prac-
tice. The major reason is to maintain a reasonable degree of consistency The QC for the selected assay is reviewed on an instrument running
in results for our users [1]. The CLSI has provided guidance for the assay. If the QC is meeting performance criteria the lot evaluation
performing lot evaluations and this covers all the important compo- can be run. The selected patient samples and QC material are run on
nents of doing lot evaluations [2]. This article will describe a practical the instrument performing the assay. The instrument is then taken
approach to doing lot evaluations that has been developed with team off-line and the new reagent lot loaded and calibrated as per specifica-
of laboratory professionals and technologists in the Hamilton Regional tion and, once ready, the patient samples and QC material is analyzed.
Laboratory Medicine program. We find it a useful and pragmatic ap- Some assays are run in duplicate according to assay methodology but
proach to our lot evaluations for both reagents and calibrators across most are run in singleton.
the range of assays that we offer in our six laboratories.
The CLSI document described the reasons for doing lot evaluations in 2.3. Selecting the acceptance target
detail [2]. As a clinician who works in laboratory medicine, the most im-
portant of these are maintaining accuracy for results and complying This is a crucial step in the process of deciding whether or not a lot is
with local regulatory best practice. As a profession we need to do our acceptable for use. The CLSI makes suggestions about selecting accept-
best to facilitate optimal patient care and one way we can do this is by ability targets [2]. The recommendations are based on the Stockholm
maintaining our assays to be as consistent as possible. The industry criteria [5] and could now be updated to the more recently agreed
who provide the reagents need to participate in this by providing lots criteria [6]. In Hamilton we debated the options available and decided
that are manufactured to meet targets expected by the users of the clin- to base the criteria for accepting lots on the total error allowable (TEa)
ical laboratory [1]. for each analyte. Our ideal target is one third of TEa but if this is not
met then we can consider other options. These include total TEa for
bias and the slope of the regression line between 0.9 and 1.1. Some as-
2. Method description for lot evaluation
says still struggle to meet these requirements and then professional
opinion is required to make a decision based on the nature of the ana-
2.1. Selecting samples for the lot evaluation
lyte, the method being used and the clinical use of the assay.
There are a number of descriptions of different sample types. We
2.4. Analyzing the data
prefer to use individual patient samples as these best reflect clinical

⁎ Core Laboratory, Juravinski Hospital and Cancer Centre, 711 Concession Street,
We use Microsoft Excel spreadsheets to evaluate data as this soft-
Hamilton, L8V 1C3, Ontario, Canada. ware is readily available in our laboratory to all users (example available
E-mail address: donwauc@mcmaster.ca. as a supplement file). The individual and mean differences between

http://dx.doi.org/10.1016/j.clinbiochem.2016.04.003
0009-9120/© 2016 The Canadian Society of Clinical Chemists. Published by Elsevier Inc. All rights reserved.

Please cite this article as: A.C. Don-Wauchope, Lot change for reagents and calibrators, Clin Biochem (2016), http://dx.doi.org/10.1016/
j.clinbiochem.2016.04.003
2 A.C. Don-Wauchope / Clinical Biochemistry xxx (2016) xxx–xxx

specimens for the old and new lots are calculated, individual and mean targets. The alternative target is then recorded on the approval form. If
percentage bias from the old to the new lot are calculated, a linear re- assays still do not meet criteria then further troubleshooting is initiated.
gression line between the old (x) and new (y) results is derived. Each
individual patient sample is reviewed for bias and it is expected that
4. Conclusion
all 10 meet the target criteria. The linear regression equation is
reviewed to make sure it falls within 10% of equivalence. The QC results
There is variability in professional practice for evaluating lot chang-
are reviewed to see if the results fall within 2 standard deviations of the
es. It is however an important competent of laboratory practice and
current mean.
each of us applying professional judgment should be able to back up
our decisions with data interpretation. The CLSI document is a good
3. Discussion
source of information to help us in this regard. Our approach, while it
does not entirely follow the CLSI guide, is a practical and useful
Preparing this protocol did present some challenges as we had a
alternative.
range of different practices in each of the laboratories. There was no
real authoritative guidance at the time we consolidated the laboratory
approach. Since then the CLSI guidance has been released and this Acknowledgments
does provide guidance and confirms some of the choices we made.
The difference of opinion between professional staff members added The Clinical Laboratory staff of the Hamilton Regional Laboratory
to the discussion. The two major areas that we had to debate were the Medicine program including Drs. Stephen Hill, Pete Kavsak and Cynthia
type and number of samples and the criteria for acceptance. Balion; and technologists from our laboratories including Lynn Ford,
For some the use of QC material was adequate and the introduction Lorna Clark, and John Beattie who all contributed to the development
of patient material is an unnecessary and labor intensive step. For others of a combined protocol for our core laboratories and Pat Hudecki who
the number of samples suggested was too few and for some it was too produced early versions of the spreadsheet that we adapted for use. In-
many. Typically we find our approach does give us good and useful in- stitute for Quality Management in Healthcare who encouraged me to
formation about assay performance. We can detect both consistent present on this subject for their annual symposium on two occasions.
and proportional bias. We are able to provide our suppliers with details
about why we are not happy with a lot change and demonstrate to our
Appendix A. Supplementary data
proficiency testing scheme that our lot changes meet our criteria. I have
carefully reviewed the CLSI method for calculating the number of pa-
Supplementary data to this article can be found online at http://dx.
tient samples and followed the guideline to work out the details for a
doi.org/10.1016/j.clinbiochem.2016.04.003.
few assays [2]. Each required a significant amount of time and the sam-
ples required would be more challenging to identify for the laboratory
technologist. References
The criteria of acceptance likewise generated some difference of
[1] R. Bais, D. Chesher, More on lot-to-lot changes, Clin. Chem. 60 (2) (Mar 2014)
opinion. Previous practice had included review of bias, slope and QC 413–414.
material performance. The debate around contribution of analytical im- [2] CLSI, User Evaluation of Between-Reagent Lot Variation. Approved Guideline. CLSI
precision and biological variation to reagent and calibrator continues. Document AP26-A, Clinical and Laboratory Standards Institute, Wayne, PA, 2013.
[3] W.G. Miller, A. Erek, T.D. Cunningham, O. Oladipo, M.G. Scott, R.E. Johnson,
Some would like to base the acceptance of the lot on analytical perfor-
Commutability limitations influence quality control results with different reagent
mance of the assay while others prefer using biological variation as lots, Clin. Chem. 57 (1) (Feb 2011) 76–83.
the base for acceptance. As this is a matter of professional opinion we [4] M.C. Cho, S.Y. Kim, T.D. Jeong, W. Lee, S. Chun, W.K. Min, Statistical validation of re-
agent lot change in the clinical chemistry laboratory can confer insights on good clin-
compromised on the one third of TEa (based on biological variation)
ical laboratory practice, Ann. Clin. Biochem. 51 (6) (Oct 20 2014) 688–694.
which is a pragmatic approach to analytical performance to achieve a bi- [5] D. Kenny, C.G. Fraser, P.H. Petersen, A. Kallner, Consensus agreement, Scand. J. Clin.
ological variation target [6]. It must be noted that many of the reagent Lab. Invest. 59 (7) (1999) 585.
manufacturers use wider limits of acceptance for lot changes. If the [6] S. Sandberg, C.G. Fraser, A.R. Horvath, R. Jansen, G. Jones, W. Oosterhuis, et al., Defin-
ing analytical performance specifications: consensus statement from the 1st strategic
assay does not meet the one third of TEa criteria each professional mem- conference of the European Federation of Clinical Chemistry and Laboratory Medi-
ber of staff can use their knowledge of the assay and apply alternative cine, Clin. Chem. Lab. Med. 53 (6) (Jun 2015) 833–835.

Please cite this article as: A.C. Don-Wauchope, Lot change for reagents and calibrators, Clin Biochem (2016), http://dx.doi.org/10.1016/
j.clinbiochem.2016.04.003

You might also like