PT. QUANTUM ANALYTICS TEKNIKATAMA DOC NO.

: QAT-SR/HR/21/05

TEMPORARY REPORT DATE : April 1st, 2021


TEKNIKATAMA

CUSTOMER DETAILS
Customer : PT South Pacific Viscose Lenzing Device Location : NGBC Stack
Contact Name : M. Irfan Ardiansyah Service Engineer : Salahuddin & Hermin
Department : Head of Section Instrument Date of Work : March 29th – April 1st, 2021

PRODUCT DETAILS
Manufacture : SICK
Model No. : GM32-P1124 (Cross Duct Type)
SN / Tag. (if any) : 1641 8032

FINDINGS

1. Optical surface of the Sender/Receiver and Reflector is slightly dirty.

Fig. 1. Optical surface of Sender/Receiver and Reflector

2. 4-Quadrant lens was found broken and has been white stain due to exposured by UV ray.

Fig. 2. 4-Quadrant lens was broken


3. There are some alarms found before verification test with test gases is carried out.
These alarms are Measurement value out of range, Absorption out of range, and Spectral evaluation
(BadModel)

Fig. 3. Alarms logbook

ACTIONS

1. Cleaning the optical surface with an optical cleaning cloth and purge it with clean air (nitrogen).

Fig. 4. Optical surface of Sender/Receiver and Reflector after cleaning


2. Replace the broken 4-Quadrant lens with new one.

Fig. 5. 4-Quadrant lens


RESULT

1. The optical surface has been cleaned.


2. 4-Quadrant lens has been replaced and the performance of UV lamp increased from 64 % to 92 %.
3. Optical alignment after replace the 4-Quadrant lens is good.
4. Appear “absorption out of range” that still can not be resolved. We need further communication with
factory to resolve the issue.

Purwakarta, April 1st, 2021

Carried out by :

Salahuddin & Hermin


Service & Maint. Dept.