Error Drain and Pressure
Error Drain and Pressure
Error Drain and Pressure
-Make sure the vacuum -210 mBar is stable in pressure syringe with Barflex.
-Go to Main Screen > Maintenance > Advanced services > Motors > Syringes.
-Check the valve 30 , tubing and T connector connected to it for blockage or loose connection.
-Go to “Disable Draining Alarm― option in General settings and select “No―.
- Run Auto-control.
-Enable the pressure sensor-->Restart the instrument-->check the pressure sensor 200 +/-5 mBar and 0+/-5 mBar in pressure sensor adjustment menu -->Auto control.