Machine Make & Model : Bhabhatron-II (Co-60)

Date : 03.07.2021 Carried out by : RSH/KKS/RRK

Phantom : WP1D
Setup : SSD Depth : 5 cm
Chamber : PTW Farmer type(Sl.no. 011495) Nd,w : 5.375 cGy/nC
Electrometer : PTW Unidose E Field size : 10x10 cm^2
Bias : 400 Time : 1 min

Temperature-Pressure Correction
Temperature (◦C) Pressure (mbar)
kt,p
T T0 P Po
18.65 20 1001.05 1013.2 1.00747701

Polarity Correction
Voltage Meter readings (nC)
(Volts) kpol
M.R.1. M.R.2. AVG.
400 33.52 33.52 33.52
1.00044749
-400 33.55 33.55 33.55

Recombination Correction
Meter Readings (nC)
Voltage (Volts) ks
M.R.1. M.R.2. AVG.
V1 400 33.52 33.52 33.52
V2 200 33.41 33.42 33.415 1.00104853
V1/V2 2 M1/M2 1.0031423

Meter Readings (nC) O/P Ndw*MR*ktp*kp


kq PDD (%) (cGy/min)
M.R.1. M.R.2. M.R.3. M.R.4. M.R.5. AVG. ol*ks
33.52 33.52 33.52 33.52 33.53 33.522 1 78.8 230.7102 181.7996182731
Machine Make & Model : Bhabhatron-II TAW (Co-60)

Date : 03.07.2021 Carried out by : RSH/KKS/RRK

Phantom : WP1D
Setup : SSD Depth : 5 cm
Chamber : Nd,w : 8.08 cGy/nC
Electrometer : SNC PC ELECTROMETER Field size : 10x10 cm^2
Bias : -300 Time : 1 min

Temperature-Pressure Correction
Temperature (◦C) Pressure (mbar)
kt,p
T T0 P Po
18.5 22 999.9 1013.2 1.00128726

Polarity Correction
Voltage Meter readings (nC)
(Volts) kpol
M.R.1. M.R.2. AVG.
300 22.63 22.65 22.64
0.99920621
-300 22.69 22.69 22.676
Calculated Ndw (cGy/nC) =( 〖〖𝑵𝒅𝒘〗 _𝒓𝒆𝒇
"(MR∗Ksat∗kpol∗ktp)" 〗 _𝑹𝒆𝒇 " "
Recombination Correction
)/ 〖 "(MR∗Ksat∗kpol∗ktp)" 〗 _𝑁𝑒𝑤
Meter Readings (nC)
Voltage (Volts) ks
M.R.1. M.R.2. AVG.
V1 -300 22.69 22.69 22.676
MR*ktp*kpol*ks Calculated Nd,w
V2 150 22.59 22.61 22.6 1.001122
V1/V2 -2 M1/M2 1.00336283 22.71263 8.0043

Meter Readings (nC) O/P


kq PDD (%) (cGy/min) % var.
M.R.1. M.R.2. M.R.3. M.R.4. M.R.5. AVG.
22.69 22.67 22.69 22.65 22.68 22.676 1 78.8 232.8909 -0.94
Machine Make & Model : Bhabhatron-II TAW (Co-60)

Date : 03.07.2021 Carried out by : RSH/KKS/RRK

Phantom : WP1D
Setup : SSD Depth : 5 cm
Chamber : SNC 600 C (S.No.241924006) Nd,w : 5.141 cGy/nC
Electrometer : SNC PC ELECTROMETER Field size : 10x10 cm^2
Bias : -300 Time : 1 min

Temperature-Pressure Correction
Temperature (◦C) Pressure (mbar)
kt,p
T T0 P Po
18.5 22 1000.5 1013.2 1.00068678

Polarity Correction
Voltage Meter readings (nC)
(Volts) kpol
M.R.1. M.R.2. AVG.
300 35.12 35.11 35.115
1.0000712
-300 35.11 35.11 35.11

Recombination Correction Calculated Ndw (cGy/nC) =( 〖〖𝑵𝒅𝒘〗 _𝒓𝒆𝒇


Meter Readings (nC) "(MR∗Ksat∗kpol∗ktp)" 〗 _𝑹𝒆𝒇 " "
Voltage (Volts) ks
M.R.1. M.R.2. AVG. )/ 〖 "(MR∗Ksat∗kpol∗ktp)" 〗 _𝑁𝑒𝑤
V1 -300 35.11 35.11 35.11
MR*ktp*kpol*ks Calculated Nd,w
V2 150 35.03 35.03 35.03 1.000762
V1/V2 -2 M1/M2 1.00228376 35.1613798662841 5.170

Meter Readings (nC) O/P


kq PDD (%) (cGy/min) % variation
M.R.1. M.R.2. M.R.3. M.R.4. M.R.5. AVG.
35.11 35.11 35.11 35.1 35.11 35.1080 1 78.8 229.3968 0.57