Operating Manual

FLETA 5
Clinical Centrifuge

Date of Purchase

Serial No.

Place of purchase
Fleta5
Research Use Only

Copyright ⓒ 2018 Hanil Scientific Inc. All rights reserved.

Contact Us
If you have and questions, contact Hanil Scientific Inc. or place of purchase.
+82-2-3472-0727
Inquiry: info@ihanil.com
Order: sales@ihanil.com
Tech. support: techsupport@ihanil.com

The contents in this operating manual are subject to change for device improvement.

Doc.No.: OMF5PDEN1801
CONTENTS

1. General Considerations------------------------------------------------- 3
1.1 Safety -3
1.2 Transport & Storage -4
1.3 Safety label -4
1.4 Electric safety information -4

2. Product description-------------------------------------------------------5
2.1 Structure -5
2.2 Delivery package -6
2.3 Technical specifications -7

3. Installation ---------------------------------------------------------------- 8
3.1 Packing Inspection -8
3.2 Installation -8
3.2.1 Selecting the location -8
3.2.2 Power connetion -9

4. Operation ---------------------------------------------------------------- 10
4.1 Operation panel -10
4.2 Opening and closing the lid -11
4.3 Loading and removal of the rotor -11
4.4 Setting rotor/bucket ID - 12
4.5 Loading tubes - 13
4.6 Setting RPM -14
4.7 Setting RCF-14
4.8 Setting Run Time - 15
4.9 Setting Acceleration Rate - 15
4.10 Setting Deceleration Rate - 16
4.11 Saving a Program - 16
4.12 Calling Up a Program - 17
4.13 Starting/Stopping a Run - 18

5. Maintenance------------------------------------------------------------ 18
5.1 Care instructions -18
5.2 Cleaning -18

6. Troubleshooting--------------------------------------------------------- 19
6.1 General errors -19
6.2 Error codes- 20
6.3 Emergency release -22

7. Rotor ---------------------------------------------------------------------- 23
1. General Considerations
1.1 Safety

Follow precautions and all the safety requirements described on this user manual
to prevent any damage and failure of equipment and loss of lives.
1. The centrifuge should be installed on flat surface to maintain level.
2. Check the voltage to be used, before connecting the centrifuge to the power source.
3. Only use rotors, parts, and accessories provided by Hanil Scientific Inc.
Hanil Scientific Inc. is not responsible for damages of the device and accidents caused by
using parts and accessories not recommended.
4. Do not exceed the maximum rated speed of the rotor or buckets in use.
5. Make sure to prepare necessary safety measures before using samples that are toxic or
radioactive samples or pathogenic or samples or infectious blood.
6. Substances that may generate volatile or explosive vapor can not be centrifuged.
7. The balancing work of samples should done in advance before operation.
8. To ensure safe use of the device, do not expose the device to strong acids, strong bases,
cesium, salt, or alkaline detergents.
8. If the centrifuge is contaminated by toxic or radioactive samples or infectious blood
samples, remove contaminants completely and take needful actions such as ventilation
or isolation of centrifuge.
9. Before operation, rotor and chamber should be dry.
10. Do not attempt to slow or stop the spinning rotor by hand.
11. Only centrifuge with rotor and rotor lid firmly tightened.
12. Do not block vents.
13. When serving the centrifuge, be sure to remove contaminants in advance.
14. Please contact the place of purchase or Hanil Scientific Inc. for product repairs.
15. According to IEC61010-2-020 maintain a 30cm “clearance envelope” around the
centrifuge while the rotor is spinning.
16. Turn the power switch off after using the device.
17. Unplug the power plug before cleaning or left unused for a long period of time.

info@ihanil.com / techsupport@ihanil.com l 3
1. General Considerations
1.2 Transport & Storage

• The device and the accessories • Only lift and transport the
may only be stored in dry rooms device with sufficient number of
helpers.

-Storage- -Transport-
Ambient temperature 5℃~35℃ Ambient temperature -10℃~40℃
Maximum relative humidity 30%~85% Maximum relative humidity 10%~90%
Air pressure 500~1060hpa Air pressure 500~1060hpa

1.3 Safety label attached to a product

Mark indicating danger


Insert tube
symmetrically.
and warning.

Mark indicating a place in


Firmly tighten danger of electric shock.
the rotor lid.

Watch your figers


when close the lid.

1.4 Electric safety information


1.Use a power cord only provided with equipment.
2.Use sockets with a protective earth conductor and suitable power cord.
3.Do not use an extension cord.
4.Do not place anything on the powercable
5.Do not block vents.
6.If you have the following emergencies, shut off the power supply and unplug the
powercord from outlet and contact your place of purchase.

-Unusual noises or smell from the equipment.


-Damage or wear of a power cord.
-Breakdown of circuit breaker, fuse or safety device.
-If you spill liquid on the equipment.
-If theequipment has been damaged.

4 l www.ihanil.com
2. Product description
2.1 Structure




⑦ ⑧

①Door:
Door protects the inside of the chamber.
②Door latch:
Fix the door.
③RPM measuring window:
Visual inspection for rotor stop or for a speed check using a stroboscope.
④Chamber:
Where the rotor is loaded.
⑤ Operating panel
Set desired run conditions.
⑥Power switch
On/off the centrifuge
⑦Emergency lid open tool
Active the emergency release(on the bottom of the device)
⑧Fuse case
If the wiring is short-circuited, overloaded, or short-circuited, turn off the power.
⑨ Power connector
The power cord is connected.

info@ihanil.com / techsupport@ihanil.com l 5
2. Product description
2.2 Delivery package

①Mainbody
②Power Cable
③Operating manual
④Twrench, for rotor locking
⑤Optional items

6 l www.ihanil.com
2. Product description
2.3 Technical Specifications

Max. RPM 5,500 rpm

Max. RCF 6,230 xg

Max. Capacity 32 x 15 mL

ACC/DEC
9/10 steps
ramps

Acc. time to
20/80 sec
ACC/DEC max. speed

Dec. time
from max. 26/82 sec
speed

Time control < 100 hr, continuous

Program memory 100

Noise level ≤ 65 dB

Imbalance cutoff Yes

Dimension (W x D x H, mm) 570 x 475 x 325

Weight without rotor 37 kg

700 VA(Usual)
Power requirement (VA)
1079VA(Max.)

Power input (V, Hz) 210~240 V, 50/60 Hz (110V optional)

Cat. No. FL-5

info@ihanil.com / techsupport@ihanil.com l 7
3. Installation
3.1 Packing Inspection

•Check packing conditions carefully, before unpacking.


•Contact Hani lScientific Inc. immediately if damages found.
•Check the delivery for completeness.

•You can get contact details on packing boxes and the back of the manual.

3.2 Installation
3.2.1 Selecting the location

Installation on hard and flat ground.


• Centrifuge should be installed on hard and flat place.
• If the centrifuge is installed in an inclined place, the shaft may be bent due to the
weight of the rotor.

Goodventailation.
• For air circulation and safety, maintain a 30cm“clearance envelope”around the
centrifuge while the rotor is spinning.

Constant temperature/humidity
• Centrifuge equipped with the sensitive electronic software which is fragile with
humidity and temperature.
•Must avoid direct ray or heater and be put in the ambience of controlled temp.and air.

Avoid the corrosive gas


• Install the centrifuge in a place where corrosive gas is not generated.
• Sulfur dioxide gas and chlorine gas may cause corrosion.

Leveling
• The shaft should be put exactly vertical on the hotizontaliy flat ground by
the leveling tool

8 l www.ihanil.com
3. Installation
3.1 Packing Inspection
3.2.2 Power Connection

• Connect the device to voltage sources which correspond to the electrical


requirements on the label attached to the device.
• Use sockets with a protective earth conductor and provided powercord.

1. Connect the power cable to the power socket on the back side of the product and plug
the cord into the outlet.
2. Switch the centrifuge on using the power switch.

info@ihanil.com / techsupport@ihanil.com l 9
4. Operation
4.1 Operation Panel

① ③ ④ ⑤ ②

⑥ ⑦ ⑧ ⑨ ⑩
⑪ ⑫ ⑬

No. Button Function

Turn the dial in the setting mode


① ‘MODE’ dial
to move cursor to set the RPM / RCF / Time / Prog./Acc/Dec.

② ‘Enter’ dial Set the value.

③ Display-‘RPM’ ‘RCF’ RPM/ RCF value is displayed.

Centrifugation time is displayed.


④ Display-‘Time’
(Set to 0:0:0 , for continuous centrifugation

⑤ Display-‘Prog’ Program No. is displayed.

⑥ Display-‘Acc’ Acceleration rate is displayed.

⑦ Display-‘Dec’ Deceleraterate is displayed.

⑧ Display-‘Rotor’ Rotor Id. is displayed.

⑨ Display-‘Radius’ Radius is displayed.

Display-
⑩ Current status of the device is displayed.
‘Ready’ ‘Check’ ‘Set’

⑪ ‘Start’ Start the centrifugation.

⑫ ‘Stop’ Abort the centrifugation.


‘Lid’ Open the lid.

10 www.ihanil.com
4. Operation
4.2 Opening and closing the lid

•The lid can only be opened when the centrifuge is switched on.
•Do not reach with your fingers between the housing and lid.
•Close the lid completely before operation.

•Opening the lid


Press the ( ) lid key.

•Closing the lid


Put the lid on and press the edge of the lid.

4.3 Loading and removal of the rotor

•Only use rotors, parts, and accessories provided by Hanil Scientific Inc.
•Do not use scatched or cracked rotor in use.

Loading

① Clean the motor shaft and the chamber.


- When loading the rotor, the temperature should be 10 to 30 degrees.

② Load the rotor vertically onto the motor shaft.


③ Insert the wrench into the center hole of the rotor
and turn it clockwise to tighten the rotor bolt.
④ Check the rotor to make sure it loaded firmly.

Unloading

① To remove the rotor,


insert the wrench into the center hole of the rotor
and turn it counterclockwise to loosen the bolt.

info@ihanil.com / techsupport@ihanil.com l 11
4. Operation
4.4 Setting rotor/bucket ID

• Be sure to set rotor id. or bucket id. correctly before use

1. Press the Enter knob.


2. Turn the Mode knob to move the cursor to rotor ID mode.
3. Turn the Enter knob until the desired rotor/bucket ID displayed.
4. Press the Enter button.
5. Do you save? appears. Turn the mode knob to select [YES]
6. Press the Enter button.

Rotor/Bucket ID

Rotor Bucket ID.

HBTM 3

HBT50-1
2
HBT50-2
HSR-4S
HBT50c-1 6

HBT15-8 1

HBT15c-4 5

HSM-2 - 4

12 www.ihanil.com
4. Operation
4.5 Loading tubes

• Only use tubes provided or approved by Hanil Scientific Inc.


• Always use the same type of tube.
• Tubes should be loaded symmetrically.
• Do not exceed the maximum rated speed of the tube.
• Same volume of sample should be put on tubes.
• Check symmetric loading by balncing tubes with scales.

① Check the maximum load for each tube.


② Put tubes into rotor holes.
③ Tubes located opposite each other must be the same type and contain the same quantity.

Correct arrangement

Wrong arrangement

info@ihanil.com / techsupport@ihanil.com l 13
4. Operation
4.6 Setting RPM

1. Press the Enter knob.

2. Turn the Mode knob to select RPM setting mode.


▶ RPM setting mode: RPM flashes.

3. Turn the Enter knob until the desired rpm is displayed.

4. Press the Enter knob to save the value.

5. Do you save? appears.


Turn the Mode knob to select [Yes]

6. Press the Enter knob.

4.7 Setting RCF

1. Press the Mode knob in stand-by mode.

2. Press the Enter knob.

3. Turn the Mode knob to select RCF setting mode.


▶ RCF setting mode: xg flashes.

4. Turn the Enter knob until the desired RCF is displayed.

5. Press the Enter knob to save the value.

6. Do you save? appears.


Turn the Mode knob to select [Yes]

7. Press the Enter knob.

14 www.ihanil.com
4. Operation
4.8 Setting Run Time

1. Press the Enter knob.

2. Turn the Mode knob to select Time setting mode.


▶ Time setting mode: 00:00:00 flashes.

3. Turn the Enter knob until the desired run time is displayed.

4. Press the Enter knob to save the value.

5. Do you save? appears.


Turn the Mode knob to select [Yes]

6. Press the Enter knob.

4.9 Setting Acceleration Rate

1. Press the Enter knob.

2. Turn the Mode knob to select ACC setting mode.


▶ ACC setting mode: ACC flashes.

3. Turn the Enter knob until the desired acceleration rate is displayed.

4. Press the Enter knob to save the value.

5. Do you save? appears.


Turn the Mode knob to select [Yes]

6. Press the Enter knob.

info@ihanil.com / techsupport@ihanil.com l 15
4. Operation
4.10 Setting Deceleration Rate

1. Press the Enter knob.

2. Turn the Mode knob to select DEC setting mode.


▶ DEC setting mode: DEC flashes.

3. Turn the Enter knob until the desired deceleration rate is displayed.

4. Press the Enter knob to save the value.

5. Do you save? appears.


Turn the Mode knob to select [Yes]

6. Press the Enter knob.

4.11 Saving a Program

1. Press the Enter knob until desired program number is displayed.


▶ You can see that the program number is changed on the PROG display.

2. Set the desired Speed/Time/ACC/DEC.

3. Press the Enter knob to save the value.

4. Do you save? appears.


Turn the Mode knob to select [Yes]

6. Press the Enter knob.

16 www.ihanil.com
4. Operation
4.12 Calling Up a Program

1. Turn the Enter knob until desired program number is displayed.


▶ You can see that the program number is changed on the PROG display.

4.13 Starting or Stopping a Run

Starting a run
1. Check that all parameters are correct and the lid is closed.
2. Press the Start button.

Stopping a run
1. Press the Stop button.

info@ihanil.com / techsupport@ihanil.com l 17
5. Maintenance
5.1 Care instructions
• The following procedures should be performed regularly.

①Regularly inspect the rotor chamber for check the motor shaft is nomal.
②Rotate the shaft with your hand to make sure it turns smoothly
③Use the stopwatch to check that the time setting is correct

5.2 Cleaning

• Before cleaning the centrifuge, be sure to switch off the device and
disconnect the power cord.

• Oustside of the device


①Clean the outside of equipment with a soft and dry cloth.
②Do not use aggressive chemicals on the device such as alchol, benzene, acetone
or phenol.
③Use the stopwatch to check that the time setting is correct
▶If the device is contaminated, use a mild cleaning fluid to clean.
④Make sure do not scratch the surface of equipment when cleaning it.
▶Do not use a metal sponge.
▶If the device is rusted, remove it with a mild detergent and wipe it with a dry cloth.

• Chamber
①If the rotor chamber is not dry, wipe moisture from the chamber with a dry cloth.
②If the rotor chamber is dirty, wipe moisture from the chamber with a dry cloth.

• Rotor
①To prevent corrosion, take out the rotor from the rotor chamber.
②If any sample is spilt inside the rotor, wash and dry the rotor well.

• Disposal
In case of product is to be disposed of, the local wastes laws and regulations are
to be observed.

18 www.ihanil.com
6. Troubleshooting
6.1 General errors

Problem Recommended Action

Check the power cord connection.


Power failure
Check the power fuse of the device.

Device cannot be started Check the lid is closed completely.

Lid cannot be opened Press the ‘Lid open button’.

Remove the dirt at the door latch and


Lid cannot be closed close the lid.
Checkthe lid latch is not damaged.

Check the device whether it is installed


on the hard and flat place

Unusual noise and vibration


Reload the rotor symmetrically.
Reload the tubes symmetrically.
Tighten the clamping of the rotor with
wrench by turning clockwise

info@ihanil.com / techsupport@ihanil.com l 19
6. Troubleshooting
6.2 Error code

Error Error
Cause Recommended action
code message

E1 Imbalance - When an imbalance is - Check weight-balances of samples and then


error detected turn off and on the device for checking.

E2 Over speed - over speed than set speed - Shut off the power supply, and then, turn on
error the power switch again to check the device.

- If the error code shows continuously although


you try to operate again, please call please call
Field Service Engineer.

E3 Motor - Motor overheat - Shut off the power supply, and then, turn on
overheat the power switch again to check the device.

E4 Motor -Motor operation error - Check that the rotor is rotating through the
operation RPM measuring window
error - Shut off the power supply, and then, turn on
the power switch again to check the device.

E5 Lid - Lid open during operation - Shut off the power supply, and contact Hanil
open(Operat Scientific Inc.
ng)

E7 System error - System error - Contact Hanil Scientific Inc.

E11 Motor Motor temperature sensor - Shut off the power supply, and then, turn on
temperature error the power switch again to check the device.
sensor error

* If none of these solutions work, please send an email to techsupport@ihanil.com.

20 www.ihanil.com
6. Troubleshooting
6.2 Error code

Error Error
Cause Recommended action
code message

E12 Hi voltage - Hi voltage - Shut off the power supply, and then, turn on
error the power switch again to check the device.

E13 Low voltage - Low Voltage - Contact Hanil Scientific Inc.


error

E15 RPM Senser - RPM Sensor error - Check if the id disc is on the bottom of the
error - Rotor identification error rotor

E17 Communicati - Occurs when Main and IO - Contact Hanil Scientific Inc.
on error board are disconnected

E20 Lid_1_sensor -Lid-1 Sensor failure - Contact Hanil Scientific Inc.


error

info@ihanil.com / techsupport@ihanil.com l 21
6. Troubleshooting
6.3 Emergency lid release

In the event of a power failure or the centrifuge lid cannot be opened,


active the emergency release.

① Turn off the power switch.


② Wait for the rotor to stop completely.
③ Find the “emergency release key” at the bottom the decvice.
④ Pull down the “emergency release key” and the door will open.

22 www.ihanil.com
7. Rotor
7.1 Rotors and adaptors

Recommended RPM :
4,500

info@ihanil.com / techsupport@ihanil.com l 23
Hanil Scientific Inc.
16 Arayukro, Gimpo 10136, Rep. of KOREA
T. +82-2-3452-8965
info@ihanil.com
www.ihanil.com