Place analyzer on stable surface, provide enough room for reagents. You’ll OPEN CLOSED
need 0.8m x 0.6m to safely place the analyzer and reagents. Waste container
can be placed on the ground
Watch out for proper positioning of Make sure to use a vial with a cap.
OPEN sample vial to avoid damage to Minimal sample volume: 300µL
Follow Guide in reagent tube package analyzer
Power on
Results
Push the power button on the right front panel 4ml iHClean 2 minutes
of the analyzer. Open Cleaning + +
OPEN vial Automatic FastBlank
0.1mL iHClean 1 minute
Gestures: Closed Cleaning + +
Symbol in bottom of screen indicates CLOSED vial Automatic FastBlank
Tap new results available in database.
(quick touch and release over a certain point) Tap symbol to jump to latest result Rinsing - 1 minute
Hold (touching a certain point and holding your
finger there for 1-2 seconds before releasing it) Manage results 4ml iHClean 2 minutes
RBC cleaning + +
Swipe (quickly moving your finger along the
OPEN vial Automatic FastBlank
screen without lifting your finger) Tap-hold over the screen’s database area
Scan QR codes
Print results
Tap-hold over a random point of the screen (requires compatible printer connected)
Cleaning the outside
Tap “Read QR” button. Camera will be activated. - Wipe the outside of the analyzer clean using a damp cloth with some
Send as email attachment
detergent.
(requires internet connection)
- Clean with a cloth dampened with clean water, wipe dry
Hold QR code from reagent (or reagent pack) in front of camera,
until system recognizes and acknowledges the code. - Do not use alcohol based cleaning material
Date and Time (System Settings) Units (Customize) Daily – for overnight power off
Enter target values
(Sampling / Calibration) Hold (touching a certain point of the screen and
holding your finger there for 1-2 seconds before
releasing it)
Run calibrator sample 3-7 times Remove any sampling vial from top opening
Tap “Shutdown” button.
The process takes about 2 minutes
Analyzer powers off automatically
Verify LJ charts
(Swipe right)