Professional Documents
Culture Documents
Medical Electronics 2 Marks With Answers
Medical Electronics 2 Marks With Answers
UNIT I
ELECTRO-PHYSIOLOGY AND BIO-POTENTIAL RECORDING
2. What are the basic bioelectrical signals parameters used and what is its frequency
range?
Electrocardiography (ECG) 0.05 to 120 Hz
Electroencephalography (EEG) 0.1 to 100 Hz
Electromyography (EMG) 5 to 2000 Hz
7. Define VCG:
Vectrocardiography is the technique of analyzing the electrical activity of the heart by
obtaining ECGs along three axes at right angles to one another and displaying any two of these
ECGs as a vector display on an X, Y oscilloscope. This display is known as a vector cardiogram.
8. Define PCG:
The Phonocardiograph is an instrument used for recording the sounds connected with
the pumping action of the heart.
These sounds provide an indication of the heart rate and its rhythm city.
EC2021-MEDICAL ELECTRONICS - 2 MARK QUESTIONS & ANSWERS
12. What are the electrodes are the used for ECGs?
Limb electrodes
Floating electrodes
Progelled disposable electrodes
Paste less electrodes
13. What are the electrodes are the used for EEG and EMG?
The Electrodes Used For EEG recording are the chloride silver discs.
The electrode for EMG work are usually needle type. Needle types are used in clinical
EMG, Neurography and electrophysiological investigation of the muscle tissues.
14. What are the properties of the gel which are used in the pregelled disposal electrodes?
Stay moist for intended shelf life and during use. This is controlled by including
humectants in the gel.
Provide low electrolyte skin impedance by having ionic salts and surfactanas.
Cause minimum skin irritation, for which gels should have a PH range of 3.5 to 9.
15. What are the precautions should be taken to achieve good ECG display in the presence
of electro surgery interference?
The electro surgery return plates should be directly under the surgical site, as for as
possible.
The ECG electrodes should be placed at the maximum possible distance away from the
surgical site.
Only shielded ECG patient cables and electrodes leads must be used.
EC2021-MEDICAL ELECTRONICS - 2 MARK QUESTIONS & ANSWERS
The changes of output with the same value of input but with a different history of input
variation.
26. What are the applications in the biomedical field of instrumentation amplifier?
Extremely high input impedance
Low bias and offset currents
Less performance deterioration if source impedance changes
Very high CMRR
High slew rate
Low power consumption
EC2021-MEDICAL ELECTRONICS - 2 MARK QUESTIONS & ANSWERS
UNIT II
BIO-CHEMICAL AND NON ELECTRICAL PARAMETER MEASUREMENT
2. Name the few physical principles based on which blood flow meters are worked?
Faradays law of electromagnetic induction.
Doppler shift method
Nuclear resonance method.
7. Define cardiac output measurement? Give the reason for decreasing the cardiac output?
It is the quantity of blood delivered by the heart to the Aorta per minute.
It is the major determination of oxygen delivery to the tissues.
Due to the low blood pressure cardiac output is usually decreases.
8. What are the different types of methods used to measure the cardiac output?
Indicator dilution technique
Dye dilution technique
Thermal dilution technique
Impedance technique
Ultra sound method.
EC2021-MEDICAL ELECTRONICS - 2 MARK QUESTIONS & ANSWERS
23. What are the types of tests under the single breath category?
Tests that measure expired volume only
Tests that measure expired volume in a unit time
Tests that measure expired volume/Time.
EC2021-MEDICAL ELECTRONICS - 2 MARK QUESTIONS & ANSWERS
UNIT III
ASSIST DEVICES AND BIO-TELEMETRY
1. Define pacemaker.
Pacemaker is an electrical pulse generator for starting and maintaining the normal heart
beat.
7. Define defibrillator.
Ventricular Fibrillation can be converted in to a more efficient rhythm by applying a
high energy shock to the heart.
This sudden surge across the heart causes all muscle fibers to contract simultaneously.
This instrument for administering the shock is called as a defibrillator.
UNIT IV
RADIOLOGICAL EQUIPMENTS
Radiography Fluoroscopy
X ray image is developed by X ray image is developed by
photosensitive film photoelectric effect and fluorescence
principle.
A wide range of contrast can be Contrast can be increased by
obtained. introducing electronic image
intensifiers.
Patient is not exposed to X rays Patient is exposed to X rays during
during examinations of the X ray examinations of the X ray image
image
Patient dose is low Patient dose is high
Movement of organs cannot be Movement of organs can be observed
observed
Since some X rays are lost by means of bucky grid, the density of the image in the film
will be reduced and the image resolution is also reduced.
Therefore to improve the density and resolution of the image ,the image intensifiers are
used.
6. State the applications of X rays.
Skeletal structure
Respiratory organs
Bronchial carcinoma
Circulatory organ
Digestive organs
body as cross sectional images in any direction and can easily discriminates between healthy and
deceased tissues.
17. What are the biological effects occur due to NMR components.
Heating due to RF power.
Static magnetic field
Electric current induction due to rapid change in magnetic field.
UNIT V
RECENT TRENDS IN MEDICAL INSTRUMENTATION
5. Define telemedicine.
It is the application of tele communications and computer technology to deliver health
care from one location to other.
It involves the use of modern information to deliver timely health service to those in
need by the electronics transmission.
12. What are the diagnostic statements observed from the ECG records.
Morphological statement based on ECG waveshapes.
Rhythm statement- concerned with the site and the rate of cardiac pacemaker.
20. Define:Microshock
A physiological response to a current applied to the surface of the heart that results in
unwanted stimulation like muscle contractions or tissue injury is called microshock.