Professional Documents
Culture Documents
Artigo Coentifico 2
Artigo Coentifico 2
2D GC with Heartcutting
Kevin Mac Namara,a Riccardo Leardib and Andreas Hoffmannc
aIrish Distillers Ltd, Midleton, Co. Cork, Ireland,
bDipartimento di Chimica e Tecnologie Farmaceutiche e Alimentari, Genova, Italy,
cGerstel Gmbh & Co. KG, Mülheim an der Ruhr, Germany.
Bertsch.6 By using controlled carrier gas flows to the injection Modern electronic pneumatic control (EPC)
port and the column intersection point, fractions could be technology is used for the necessary
diverted to the second column or vented to waste. This design
obviated the disadvantages of a valve connection system but
heartcutting pressure equilibrations.
introduced the challenge of stabilizing carrier-gas pressure in
the injector and at the midpoint connector.7 Several consecutive cuts to the second column. With software control
commercial instruments were subsequently introduced to the of all aspects of the 2D analysis the entire process becomes as
market based on the Deans principle. Some were add-on straightforward as a routine sequence to a single column.
pneumatic modules for existing GC units while others were
complete dedicated systems using double-oven configurations Modern GC–GC
for independent temperature control of the two different Build-up of a 2D system: The system described in this section
columns. Gordon and colleagues reviewed several of these has been in use in our laboratory for some years. Initial work
systems in 1986 for both activity and ability to analyse complex on flavour compounds in distilled spirits revealed the obvious
samples.8 All the systems functioned adequately for limitations of single compound capillary GC for identification
heartcutting from complex samples, but the operations of important trace nitrogen and sulphur. Interesting patterns
required were still rather complex and manual, and as such were available from selective detectors but mass spectrometric
these instruments never integrated fully into the routine identification was very difficult because of matrix interference
general chromatographic community. Practically all of these and relative detector sensitivity. In a first application using 2D
systems have ceased to be available commercially. The problem heartcutting we were able to identify several new medium
principally centred on the considerable technical difficulty of boiling sulphur compounds in whiskey.9 Rather than say we
achieving the necessary pressure equilibration for reproducibility logically pre-planned the design of our particular GC–GC
of cutting and retention-time stability. system it is more realistic to say that our current system has
This basic problem can be solved by use of modern evolved to its present format from a learning process based on
electronic mass flow controllers and electronic proportional practical applications and taking advantage of continuing
backpressure regulators with pressure sensing. The following advances in GC design and operation.
sections of this review will attempt to describe such a unit Figure 1 is a schematic of a system offering total flexibility
incorporated into Agilent GC and GC–MS (Agilent for both routine GC and GC–GC. The main oven with MSD
Technologies, Waldbronn, Germany) equipment to produce a and selective detectors already existed and the monitor GC was
system that allows rapid access on demand to 2D capability. added and installed beside the GC–MS. These two systems are
The only intrusive addition to the first GC is a single miniature then connected by the cryotrap system (CTS). The column-
device at the intersection of the two columns that can remain switching device (CSD) is installed in the first oven and linked
in situ in this GC whether or not the system is in 2D mode. to the external pneumatics box. For optimum instrumental
Modern electronic pneumatic control (EPC) technology is flexibility the entire configuration is then run as three separate
used for the necessary heartcutting pressure equilibrations. systems from the Agilent Chemstation. These are described in
When a programmed temperature vaporization injector (PTV) Table 1.
is used for sample introduction powerful techniques such as The Master software (Gerstel GmbH, Mülheim an der Ruhr,
stir-bar sorptive extraction (Twister) and large-volume injection Germany) for sample introduction, cutting and transfer
with solvent venting can be used to productive effect. Fully operations, cryotrapping and remote start of monitor GC and
automated sequences can be programmed for repetitive GC–MS is also run from the Chemstation.
injection to the first column, with single or multiple When GC–GC is required the main GC column is
disconnected from its injector and threaded through the
cryotrap for connection to the exit of the column-switching
Figure 1: Schematic of the GC–GC system. 1 injector for device. The outlet of the precolumn is similarly connected to
either oven; 2 PTV inlet on instrument 1; 3 PTV inlet on the inlet of the column-switching device. Additional permanent
instrument 2 or 3; 4 monitor FID on instrument 1; inlet and outlet lines connect the column-switching device to
5 sulphur chemiluminescent detector; 6 nitrogen the mass flow controller and pressure regulator with sensor
chemiluminescent detector; 7 pneumatics unit for column in the pneumatics box. This will be described in more detail
switching device. later. A final deactivated 0.05 mm i.d. fused-silica line takes a
small split from the column switching device to a flame
ionization detector (FID) for monitoring the eluent from
1 the precolumn.
6 5 7 2 4
3 Table 1: Description of the instruments used to establish the
different two dimensional configurations.
Instrument 1 Agilent 6890 GC with FID and Gerstel PTV injector.
www.lcgceurope.com 15
Mac Namara, Leardi and Hoffmann
The advantages from this instrumental design are: the pressure at the column switching device will be the head
• The entire configuration does not constitute a dedicated pressure on the main column after the selected compounds
GC–GC system. Each instrument is self-contained and can be are transferred. In addition, the commencement of ballistic
run entirely separately for specific applications when heartcutting heating of the cryotrap is the start signal for the chosen
operations are not required. For heartcutting the Master analysis on the main column. Therefore, this combination of
software links all individual modules in an orderly sequence. “column head pressure” (pressure at the column switching
• The only connections required are those of the pre- and device) and “injection” (heating of the cold trap) can be used
main columns to the column switching device. The to establish a retention-time locked analysis for heartcuts on the
connections are made with Graphpack technology ensuring a main column. We applied this procedure to organophosphorus
leak-free coupling. (The most recent commercial addition to pesticides in a complex lemon oil and the modus operandi will
the GC–GC market is an add-on Deans unit but this requires be explained in detail in the applications section.
connection to three micro-volume tees). • Total automation is possible with all aspects of the 2D
• The cryotrap can also be ballistically heated after trapping analysis. The selected cut time windows together with the
selected compounds and therefore acts as a proxy injector for relevant cryotrap cool and heat times are programmed to the
the transfer of selected compounds to the main column. Master software. A sequence with data files is established for
• The EPC principle used for cutting and venting means that each of the chosen instrument pairs and the injection
sequence starts all other sequences.
• Sequences can be multiple injections of different samples
Figure 2: (a) Pneumatics and the column switching device in with similar cuts for each sample or multiple injections of the
heartcutting condition, countercurrent flow-off; same sample with different consecutive cuts transferred to
(b) Pneumatics and the column switching device in venting the main column. In this way a complex sample can be
condition, countercurrent flow-on. totally profiled by automated heartcutting. An application
example to a whiskey extract will be shown.
Pneumatics and the column-switching device: Figures 2(a) and
(a)
2(b) show the principle and the simplicity of the column-
PTV-inlet switching device. The external pneumatics unit contains the
(CIS 4) mass flow controller and proportional valve plus sensor, with
corresponding connections in and out of the column-switching
Precolumn device. Figure 2(a) can be defined as the zero condition. The
mass flow controller is off and the carrier gas flow to both
FID columns is supplied through the injector. A digital pressure
reading at the column-switching device is available at the
kPa Chemstation. This zero condition applies when a cut is to be
Countercurrent transferred to the main column. For venting of all unwanted
Countercurrent mL flow-in
flow-out eluent from the precolumn, a countercurrent flow greater than
Cooled TRAP
the column flows is established across the column switching
(CTS)
device by the mass flow controller. This condition is depicted in
Main column Figure 2(b) (for capillary operation this can be 10 mL/min).
The outlet proportional valve then acts as a programmable
restrictor and quickly re-establishes the zero condition pressure
at the column-switching device through the software.
This combination of mass flow control and programmable
restriction guarantees extremely rapid pressure re-equilibration
at the mid-point connector after transfer of cuts. The entire
(b)
operation is software driven and user friendly. After all compounds
PTV-inlet of interest have been transferred to the cold trap the switching
(CIS 4) pneumatics will be in the non-zero condition The mass flow
controller supplies both the countercurrent venting flow for
Precolumn precolumn eluent and the main carrier gas flow to the column
switching device. The Master software now allows a suitable
FID pressure to be established at the switching device for
subsequent application of retention-time locking technology.
kPa
Countercurrent Applications
Countercurrent mL flow-in Organophosphorus pesticides in lemon oil: Natural essential oils
flow-out Cooled TRAP represent a very complex matrix and offer very difficult
(CTS) challenges for detection of possible pesticide residues. Element-
Main column
specific detectors are useful for indicating possible positives but
spectral confirmation on any single column is severely
compromised by extensive matrix coelution.
The following discussion will show in detail the elaboration
(essentially software operations) of a 2D procedure for
identification of several organophosphorus (OP) pesticides in • In theory, 160.4 Kpa will be the pressure required at the
lemon essential oil. Six compounds were cut after a 1 µL column switching device when retention-time locked analysis
splitless injection to a DB-17MS precolumn and then profiled is required for compounds transferred to the main column.
under retention-time locking GC–MS after transfer to a HP5-MS However, this pressure should now be retuned as the
main column. Detailed information on retention-time locking characteristics of the “ new injector ” (column switching
concepts and techniques can be found in relevant Agilent device and cold trap) will be slightly different. The compounds
publications.10–12 The following are the initial operations are already focused in the “liner”, as represented by the first
required to set up this procedure in GC–GC mode. The length of the second column.
sequence of operations is quite logical and constitutes a • Cap the injector in the main oven (Instrument 2) and
general-purpose template, which is applicable whether single or establish 160.4 Kpa as a blank head pressure in a GC–MS
multiple cuts are desired. method. This instrument is then ready for a retention-time
• With the HP5-MS column installed in the main oven only locked run but with injection from the column switching
(Instrument 2, see Figure 1), run the five injections required device and the cold trap.
for retention-time locking set-up. In this instance, this • Run the locking compound on the DB17-MS with transfer
calculated a locked inlet pressure of 160.4 Kpa for elution of to the HP5-MS. Relock in the software and read the new
chlorpyrifos methyl (locking compound) at 16.593 min in headpressure required for the expected locking compound
constant pressure mode. time. For this column combination 140.7 Kpa was indicated
• Disconnect the HP5-MS from the inlet in the main oven. and this is now the required locking head pressure at the
Thread the column through the cryotrap and reconnect to column-switching device. When heartcuts are analysed on
the exit of the column-switching device in the pre-oven. The the main column under these conditions they can be
DB-17MS precolumn is installed from the pre-oven inlet to
the entry of the column switching. The lines to and from the
pneumatics box, together with the line to the monitor FID Figure 3: (a) Scan TIC after injection of lemon oil with
are always in place and so the two columns are the only 10 mg/L OP pesticides to HP5-MS column only; (b) Scan TIC
connections required for full 2D operation. after injection of lemon oil with 1 and 10 mg/L OP pesticides
to DB-17MS precolumn with three heartcut transfers to
HP5-MS column.
Table 2 : Total parameter set for GC–GC–MS analysis of
pesticides in lemon oil. (a)
A. Sample Lemon oil spiked with 1 and 10 ppm 2.5107
(MS Scan) or 50 ppb (MS-Sim)
Organophosphorus pesticides mix A 2107
(Sigma-Aldrich)
1.5107
B. Injection 1 µL splitless to PTV injector (CIS 4 –
Gerstel GmbH) 60 °C, 10 °C/s to 300 °C 1107
10 min. 1 min splitless time.
Carrier gas: helium. 5106
www.lcgceurope.com 17
Mac Namara, Leardi and Hoffmann
automatically screened using the Agilent 567-compound The heartcut entries are for the three compounds methyl
retention time locked pesticide mass spectral library.12 parathion, fenchlorphos and chlorpyrifos after their retention
• A final advantageous step can now be incorporated to times were established on the first column. The reproducible
compensate for pressure changes resulting from the narrow heartcut windows (0.3 min) are possible because of the
temperature program in the first oven. With the column retention time stability on the precolumn.
switching pneumatics in zero condition (mass flow controller Figure 3 compares the scan TIC traces for direct injection to
off) set the first oven at its starting temperature and adjust the HP5 column only (no heartcutting) of the lemon oil spiked
the inlet head pressure until 140.7 Kpa is achieved at the with 10 mg/L (ppm) of the OP mix, with a similar TIC trace
column switching device. Set the first oven to its final but after three heartcuts from the DB17 precolumn.
programmed temperature and again adjust the inlet pressure The 2D procedure was at both 10 and 1 mg/L spiking
to the new higher value to achieve 140.7 Kpa at the levels. The analysis parameters are as outlined in Table 2. What
switching device. Using these initial and final head pressure is immediately apparent is the degree of matrix removal with
values include a pressure program in the GC method with a heartcutting, which allows easy detection of the three
rate, and initial and final times, to match the corresponding compounds in the MS. Both the direct analysis and the
oven temperature program. In this way the column-switching heartcut analysis to the HP5 column were run under retention-
device will always be at 140.7 Kpa during the precolumn time locking conditions, but the direct analysis only reported
run and for chosen heartcuts at any point in this run. After fenchlorphos when the data file was screened. At the 1 mg/L
cuts the EPC pneumatics instantly re-establish 140.7 Kpa level no compounds were found without initial heartcutting.
and the end result is very precise retention time Figure 4 shows the mass spectra of these compounds from the
reproducibility on the first column and absence of pressure 10 mg/L 2D analysis.
disturbance during heartcutting. Lower detection limits using retention-time locking can be
These parameters are summarized in Table 2, together with achieved by setting up a GC/MS-SIM for the compounds of
Master software entries for injection, pneumatic options, interest.13 The same parameters as in Table 2 are used but now
heartcutting and cold trap refocusing. the cutting is extended to six compounds — six individual 18 s
cuts from the precolumn. Figure 5 shows the main column
SIM trace after heartcutting for the lemon oil as is and the
Figure 4: Mass spectra of (a) methyl parathion, lemon oil spiked with 50 µg/L (ppb) of the OP mix. The
(b) fenchlorphos and (c) chlorpyrifos at 10 mg/L after
lemon oil itself contains low µg/L amounts of disulphoton,
heartcutting from lemon oil.
methyl parathion and chlorpyrifos.
Table 3 shows the retention-time locked screener report for
(a) the six compounds in the 50 µg/L spike sample.
109
1.2105
O O
125
1105 P Figure 5: (a) SIM trace for six pesticides after heartcutting
263
O S
8104 from lemon oil; (b) SIM trace for six pesticides spiked at
6104 50 µg/L after heartcutting from lemon oil.
4104 79 N
63 93
2104 47 O O (a)
200 233
Cl
2000
125
Cl
Abundance
1105 Cl
109 12 14 16 18 20 22 24
79 93
47 63 167 270 (b)
Table 3 : GC–MS retention-time locked screener report for lemon oil spiked with six pesticides at 50 µg/L each after direct
injection to a DB-17MS precolumn and heartcutting to a HP5-MS main column.
Target Qualifiers
Compound Status ExpRT Delta m/z Resp. Out of Range XCR
86 Ethoprophos ✕ 10.742 0.058 158 28076 0.82
170 Disulphoton ? 14.551 0.045 88 53845 89,97,142 0.47
217 Methyl parathion ? 16.594 0.059 263 40648 109,125 0.37
236 Fenchlorphos ✕ 17.330 0.045 285 60596 0.95
278 Chlorpyrifos ? 19.234 0.007 197 92456 97 0.70
383 Prothiofos ? 23.752 0.003 309 26145 162,113 0.77
124
2105 9 10 11
68 95 105
Abundance
(b)
40 60 80 100 120 140
(b) 10000
124
1.6106 8000
1.4106 6000
1.2106
1106 O 4000 2
N
8105 1
N 2000
6105
94 151
4105 81 9 10 11
2105 166
Time (min)
40 60 80 100 120 140 160
m/z Peaks: 1 IPMP, 2 IBMP.
www.lcgceurope.com 19
Mac Namara, Leardi and Hoffmann
these values can predict the extraction efficiency for a compound Cold trapping may not be necessary in all instances, but if it is
from an aqueous matrix to a PDMS-coated device. For the needed and not available then the analyst must rely on phase-ratio
pyrazines this was calculated as 60% recovery from a 10 cc refocusing (thick film columns) or use of PLOT columns.
sample for a standard 10 mm Twister, making them suitable These introduce other complications such as much longer
candidates for this technique. elution times for semivolatile compounds and are best used
For this analysis, the same strategy in terms of pressure and only if a dedicated analysis with one cut is required. With cold
flow programming was used as for the pesticide application, trapping available all eventualities are catered for and maximum
except in this instance a more polar column (Supelcowax-10, flexibility is attained.
Supelco Inc. Bellefonte, Pennsylvania, USA) could be used as Figure 7 shows the second column SIM trace for a wine and
precolumn to maximize the selectivity difference for the a 10 ng/L standard of the pyrazines after SBSE and subsequent
compounds of interest. This integration of SBSE and GC–GC transfer of two cuts from the polar precolumn to the HP5-MS
now means that that the sample preparation step is the main column. This analysis used the combination of
combination of the Twister enrichment and the first dimension Instrument 1 and Instrument 2 (Table 1).
separation. In SBSE the glass-lined and PDMS-coated Figure 8 shows the same analysis for the wine only but now
magnetic stir bar is simply dropped into 10 cc of sample and using the combination of Instrument 1 and Instrument 3,
stirred for 1 h. 50 to 100 samples can be treated simultaneously
and all stir bars can then be analysed in a sequence using a
modified PAL autosampler (Gerstel GmbH). Figure 9: (a) Monitor FID trace after injection of flavour
For each analysis all aspects of the subsequent GC–GC extract to precolumn. Cut region at 16–18 min for transfer to
separation will also be sequence automated, thus the entire second column indicated; (b,c,d) Corresponding MS, N & S
procedure is a good example of the applicability of modern traces for heartcut run on second column.
system development for complex routine automated analysis.
Use of intermediate cold trapping for cuts transferred to the (a)
cut
main column allows the use of standard capillary columns.
8105
Figure 8: Second column nitrogen trace for methoxy pyrazines 6105 Monitor FID
in wine after stir-bar sorptive extraction and heartcutting. 4105
2 2105
Abundance
1.5106 5 10 15 20 25 30 35
(b)
Response
8106
6106 MS trace
1106
4106
1
2106
9 10 11 10 20
Time (min) (c)
1.4107
Peak identification: 1 IPMP, 2 IBMP. 1.2107
1107 Nitrogen trace
8106
6106
Table 4 : Description of the different sequences needed for 4106
total profiling of a sample by GC–GC. 2106
Response
1.2108
(a)
1108 (b)
8107 8106
Abundance
6107
6106
4107 Sulphur traces
2107 4106
0
2106
26
22 15
18
Precolumn
14 10 7.0 8.0
10 5
cut time (min) 6 Time (min)
0
Analytical column
retention time (min) (b)
2.2107
(c) 100
2107
1.5107
5106
0
60
12
10 10
8 8
6 6
4 4
Precolumn 2 2 40
cut time (min) 0
Analytical column 200 400
retention time (min)
Scan number
www.lcgceurope.com 21
Mac Namara, Leardi and Hoffmann
Conclusions
Automated 2D gas chromatography with heartcutting is
available as a powerful routine tool for complex sample analysis.
Most of the modules required to construct a system are
probably available in candidate laboratories. Modern
instrument control networking ensures that the system
components are always available for individual one-dimensional
analysis. In fact, successful GC–GC application is simply a
sequence extension of the individual operations.
The all-important pressure control and manipulation at the
column intersection is achieved with modern EPC pneumatics.
Integration of cryofocusing into the system allows maximum
flexibility and will cater for the entire volatility range of
compounds. The same cryofocusing unit can be used for sample
introduction with stir-bar sorptive extraction or large-volume
injection with solvent venting. Combination of these low
sample preparation techniques and the automated features of
modern GC–GC allow the build-up of very powerful systems
capable of processing large numbers of samples. A further
developmental stage is the use of short 0.1 mm i.d. capillary
columns for lower total analysis time.
References
1. J. Giddings, Multidimensional Chromatography, Techniques and
Applications, (Marcel Dekker Inc. 1990).
2. R. Smits, LC•GC Int., 7(9), 505–508 (1994).
3. H.-J. de Geus, J. de Boer and U.A.Th. Brinkman, Trends in Analytical
Chemistry, 15(5), (1996), 168–178.
4. P. Schoenmakers, P. Marriott and J. Beens, LC•GC Eur., 16(6), 335–339
(2003).
5. D. Deans, Chromatographia, 1, 18–22 (1968).
6. W.J. Bertsch, HRC&CC, 1, 85–90 (1978).
7. K. Himberg and E. Sippola, J. Microcolumn Separations, 1, 271–277
(1989).
8. B.M. Gordon, C.E. Rix and M.F. Borgerding, Journal of Chromatographic
Science, 23, 51–60 (1985).
9. K. MacNamara, Elaboration et Connaissance des Spiriteux, R. Cantagrel
Ed. (Lavoisier-Tec et Doc, Paris, 1993).
10. V. Giarrocco, B. Quimby and M. Klee, Agilent Technologies, Applications
Brief, 5966-2469E. www.agilent.com/chem
11. H. Prest and P. Cormia, Agilent Technologies, Applications Brief
5968–3797E. www.agilent.com/chem.
12. P. Wylie and B. Quimby, Agilent Technologies, Applications Brief
5967–5860E. www.agilent.com/chem
13. H. Prest and D. Peterson, Agilent Technologies, Applications Brief
5988–4188EN. www.agilent.com/chem
14. O.P.H Augustyn, A. Rapp and C.J. van Wyk, S. Afr. J. Enol. Viticult., 3,
53–60 (1982).
15. R. Harris et al., Vitis, 26, 201–207 (1987).
16. C. Sala et al., Journal of Chromatography A, 953, 1–6 (2002).
17. E. Baltussen et al., Journal of Microcolumn Separations, 11, 737–747
(1999).
18. M. Adahchour et al., Chromatographia, 55, 361–367 (2002).