DIRECT DIGITIZER REGIUS SIGMA Operation Manual

0197
EN 08
Table of Contents

Introduction. . . . . . . . . . . . . . . . . 5 Chapter 3
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Reader Operation . . . . . . . . . . . 27
Summary of usability specifications
3.1 Power ON . . . . . . . . . . . . . . . . . . . . . . . . . 28
(for IEC/EN 60601-1-6, IEC/EN 62366). . . . . . . . 6
3.2 Image Erase Operation . . . . . . . . . . . . . . 30
Disclaimer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
3.3 Image Read Operation . . . . . . . . . . . . . . . 32
Trademark . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
3.4 Power OFF . . . . . . . . . . . . . . . . . . . . . . . . 34
Terminology Guide. . . . . . . . . . . . . . . . . . . . . . . . 8
Page Layout. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Chapter 4
Chapter 1 Cassette/Plate Operation . . . . . 35
Safety Warnings & 4.1 Cassette/Plate Handling . . . . . . . . . . . . . . 36
4.1.1 Cassette/Plate Basics . . . . . . . . . . . . . 36
Precautions . . . . . . . . . . . . . . . . 11 4.1.2 Removing the Plate From the
1.1 Symbols relating to Safety . . . . . . . . . . . . 12 Cassette. . . . . . . . . . . . . . . . . . . . . . . . 38
1.1.1 Safety Alert Symbol. . . . . . . . . . . . . . . 12 4.1.3 Loading the Plate into the Cassette. . . 40
1.1.2 Warning Notice (Signal Words) . . . . . . 12 4.2 Dealing with a Broken Cassette/Plate . . . 44
1.1.3 Description of Graphic Symbols. . . . . 12
1.2 Safety Precautions . . . . . . . . . . . . . . . . . . 13 Chapter 5
1.2.1 Overall Cautions . . . . . . . . . . . . . . . . . 13 Troubleshooting . . . . . . . . . . . . 45
1.2.2 Electromagnetic Wave Cautions . . . . . 13
5.1 Troubleshooting . . . . . . . . . . . . . . . . . . . . 46
1.2.3 Cautions Relating to the Power
5.1.1 When the Error Lamp is lit or blinking,
Supply and the Power Cable. . . . . . . . 17
but no message is displayed. . . . . . . . 46
1.2.4 Device Installation, Movement,
5.1.2 When a message is displayed. . . . . . . 47
Storage and Connection Cautions. . . . 17
5.1.3 When the Error Lamp is unlit,
1.2.5 Cautions to Be Observed During
and no message is displayed. . . . . . . 53
Usage. . . . . . . . . . . . . . . . . . . . . . . . . 18
5.2 Dealing with Plate Jams . . . . . . . . . . . . . . 54
1.2.6 Cassette/Plate Storage and Usage
5.2.1 Check the Plate Status
Cautions. . . . . . . . . . . . . . . . . . . . . . . . 19
(Do This First) . . . . . . . . . . . . . . . . . . . 54
1.2.7 Device Care and Maintenance
5.2.2 If the Plate is Not Visible From the
Cautions. . . . . . . . . . . . . . . . . . . . . . . . 20
Top Door’s Opening. . . . . . . . . . . . . . . 55
1.2.8 Disposal Cautions . . . . . . . . . . . . . . . . 20
5.2.3 If the Plate’s Front End is Visible

Chapter 2 From the Top Door’s Opening . . . . . . . 57


5.2.4 If the Plate’s Back End is Visible
Product Outline. . . . . . . . . . . . . 21 From the Top Door’s Opening . . . . . . . 58
2.1 Device Summary . . . . . . . . . . . . . . . . . . . 22 5.2.5 If the Plate’s Mid-section is the Only
2.1.1 Functions and Features. . . . . . . . . . . . 22 Portion Visible From the Top Door’s
2.1.2 System Configuration Examples . . . . . 22 Opening. . . . . . . . . . . . . . . . . . . . . . . . 59
2.2 Names of Parts and Functions . . . . . . . . . 23 5.3 Forcibly Removing the Cassette . . . . . . . 61
2.2.1 Front / Left Side / Top. . . . . . . . . . . . . . 23
2.2.2 Rear / Right Side. . . . . . . . . . . . . . . . . 24
2.3 Status Lamp . . . . . . . . . . . . . . . . . . . . . . . 25
2.4 Operation Flow . . . . . . . . . . . . . . . . . . . . . 26

3
Table of Contents

Chapter 6
Maintenance. . . . . . . . . . . . . . . . 63
6.1 A list of Maintenance Items . . . . . . . . . . . 64
6.2 Reader Maintenance . . . . . . . . . . . . . . . . 65
6.2.1 Cassette Insertion Slot Cleaning. . . . . 65
6.2.2 External Cleaning of Device. . . . . . . . 65
6.2.3 Intake Vent, Exhaust Outlet
Cleaning. . . . . . . . . . . . . . . . . . . . . . . . 65
6.2.4 Light Collection Unit Cleaning
(Keeping Them Dust Free) . . . . . . . . . 65
6.2.5 Filter Cleaning . . . . . . . . . . . . . . . . . . . 66
6.2.6 Feeding Roller Cleaning . . . . . . . . . . . 66
6.3 Plate/Cassette Maintenance . . . . . . . . . . 69
6.3.1 Plate Cleaning. . . . . . . . . . . . . . . . . . . 69
6.3.2 Cassette Cleaning. . . . . . . . . . . . . . . . 69

Appendix . . . . . . . . . . . . . . . . . . 71
Specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73

4
Introduction

5
Introduction

Thank you for purchasing the Direct Digitizer REGIUS SIGMA (called “this device” hereinafter).
This device reads exposed X-ray image data saved to a plate within a cassette, then outputs it as diagnosis
image data to the connected console.

In order to operate this device, either the medical image processing workstation ImagePilot or the CS-7
(called “Image processing controller” hereinafter), each of which control reception, processing, and output
of the image data of this device, is necessary. Concerning operation of the Image processing controllers,
please refer to the “Operation Manual” of each Image processing controller.
General X-ray images are the intended diagnostic images of this device. Mammographic images are not
included.

This manual covers cautions to be observed during usage, device operation methods, and cassette/plate
handling methods. Make sure to obtain a thorough understanding of the content within this manual, and
the Image processing controller operation manual, before using this device. Even after reading through
this manual, keep it stored near the device for reference to ensure that the device is used at maximum
efficiency.

*If this manual becomes illegible due to wear and tear, replace it with a new one. (fee required)

Summary of usability specifications (for IEC/EN 60601-1-6, IEC/EN 62366)

1) Medical purposes
• Provision of images for diagnosis of disease and injury.
2) Patient groups
• No patients groups that operate or touch the device.
• No identification of patient groups of read X-ray images.
3) Parts of body or organizations to which the device is mounted or that contact with the device.
• No appropriate points.(Device does not come in contact with the patient.)
4) Operating principle
• Insert the REGIUS cassette (hereinafter called the cassette) with a REGIUS plate (hereinafter
called the plate) inside which latent images were formed by X-ray exposure.
• The plate is pulled out from the opened tray, and while conveying the plate, scan the excitation laser
to the plate.
• Photostimulated luminescence emitted from the plate by the excitation laser is focused and detected,
and detected signals are digitized and output as image data.
• By applying erase light to the plate scanned by the excitation laser, remaining latent images will be
erased.
• The plate is reloaded, which had its latent images erased, into the cassette. Take the cassette out
from the insertion slot.
5) Significant physical characteristics
• Refer to “Specification”(p. 72) in the Appendix.
6) Significant performance characteristics
• Refer to “Specification”(p. 72) in the Appendix.
7) User of this device
• There is no need for special training for using this device. This device is intended for the following users:
Healthy technicians of sound body who fully understand the information contained in this manual
and have specialized knowledge and a license (e.g., doctors or radiological technologists).

6
Introduction

Disclaimer

(1) This manual may not be reproduced in whole or in part without the permission of Konica Minolta, Inc.
(2) The contents of this manual may be subject to change without prior notice.
(3) Konica Minolta, Inc. is not responsible for any claims made for malfunction or damage caused by
installation, relocation, modification, maintenance, and repair made by anyone except Konica Minolta
and contractors designated by Konica Minolta.
(4) Konica Minolta, Inc. is not responsible for any claims made for malfunction or damage to Konica
Minolta products, caused by third-party products not installed by Konica Minolta.
(5) Konica Minolta, Inc. is not responsible for any claims made for malfunction or damage caused by
maintenance or repair using maintenance parts other than those specified by Konica Minolta.
(6) Konica Minolta, Inc. is not responsible for any claims made for malfunction or damage caused by not
observing the precautions and operation methods described in the operation manual.
(7) Konica Minolta, Inc. is not responsible for any claims made for malfunction or damage caused by the
environment, such as power source and installation environment, that is described in the Installation
Requirement or operation manual, that is not suitable for the installation requirements of this device.
(8) Konica Minolta, Inc. is not responsible for any claims for malfunction or damage caused by acts of
nature such as fires, earthquakes, floods, or lightning strikes.
(9) Konica Minolta, Inc. is not responsible for any claims for malfunction or damage caused by using this
device for any purpose other than that specified for this device.
(10) Diagnostic and treatment action is performed under the responsibility of the physician(s).
Konica Minolta, Inc. is not responsible for any diagnostic/treatment conditions or diagnostic/treatment
results.

Trademark

Company names and product names in this manual are trademarks or registered trademarks of their
respective owners.
Please note that ©, ® and ™ marks are omitted hereafter.

Copyright © 2011 - 2018 Konica Minolta, Inc. All Rights Reserved.

7
Introduction

Terminology Guide

Terminology used throughout this manual is defined as follows.


Terminology Definition
Registers plate image read condition settings and requests, manages diagnosis images, etc.
Console The console to be used with this device is the Image processing controller, which is sold sepa-
rately.
Application software that manages the network between this device and its console.
JM (Job Manager) It has been preinstalled in the Image processing controller--the console to be used with this
device.
Reader A device for reading exposed X-ray image data written to the plate. (This device)
A film coated with photostimulable phosphor for storing up X-ray energy.
Plate Its surface is easily-damaged, and accumulated X-ray energy is discharged when exposed to
light, so it is stored within the cassette for handling.
Cassette A case in which the plate is stored for protection and ease of handling.
A container in which the plate is stored, located within the cassette. It also acts as an internal
Tray
cover.

8
Introduction

Page Layout

(1) 3.3 ● Image Read Operation

The following steps explain how to read exposed plate HINT


images with this device. • When inserting the cassette before the ImagePilot is in
image capture standby mode, changing the ImagePi-
(2) CAUTION
• Make sure not to pinch your fingers while inserting the
lot to image capture standby mode will start the image
capture process.
cassette.
• If an X-ray generating device is positioned near this
device, do not emit X-rays while this device is operat- 4 Confirm that the Ready Lamp is blinking,
ing. then remove the cassette.
• When removing the cassette, the device initial-
ization process will begin, and the Busy Lamp
(3) 1 Prepare a cassette with an image reading will be lit.

Chapter 3
plate loaded. • Once the initialization process has ended, the
If there is a memo between the cassette’s memo Busy Lamp will turn off, and the Ready Lamp
(4) clip, remove it. If there is a sticker etc. attached to
the cassette, remove it.
will be lit.

IMPORTANT
CAUTION • Do not insert the cassette during the initialization pro-
• Only cassettes specially made for this device can be cess (while the Busy Lamp is lit). Doing so will cause
used. initialization to end abnormally, and an error will occur.

When an error occurs for a read image


2 Insert the cassette in the cassette inser- If an error occurs for a read image, take the following
(5) tion slot. steps. Do not resume reading with this device or expo-
sure with the image-reading plate until the issue has
CAUTION been resolved.
• Do not insert anything other than the cassette in the
cassette insertion slot. Doing so may cause a malfunc- When a white line error occurs
tion. Condenser unit might be soiled with dust or debris.
If a foreign object has fallen inside of the device, con- Clean the condenser unit.
tact Konica Minolta technical representatives.

(6) Position the back side of the cassette facing up-


REFERENCE
• Refer to “6.2.4 Condenser Unit cleaning (keeping
ward, and slowly slide it into place, straight and all
them dust free)”(p. 61)
the way to the back, in the direction of the arrow.

When other errors occur


The plate might be dirty. Clean the plate.

REFERENCE
• Refer to “6.3.1 Plate cleaning”(p. 63)

Insert so that
center-sections
match.

• Image reading will begin once the ImagePilot’s


status is confirmed.
• The Busy Lamp (blue) will blink while the Image-
Pilot’s status is being confirmed.
• The Busy Lamp is lit during the image reading
process.

27

Number Element Explanation Icon


(1) Title Shows the title of the explained content. -

DANGER
Used to highlight important information that must be known or
Warning Notice followed during operation.
(2)
(Signal Words) Refer to "1.1.2 Warning Notice (Signal Words)"(p. 12) for details on WARNING
each type.
CAUTION
(3) Steps of operation Numbered steps explaining the process of operation. -

(4) Hint Information that may be useful during operation. HINT


Important notes for operating the machine. Details should be
(5) Note
carefully noted.
IMPORTANT

(6) Reference Indicates a section of the manual for reference as needed. REFERENCE

(Pages indicated here are for explanatory purposes, and may differ from the actual pages.)

9
10
Chapter
1
Safety Warnings & Precautions
An explanation of the warnings and precautions
to follow in order to safely use this device.

11
1.1 Symbols relating to Safety

1.1.1 Safety Alert Symbol 1.1.3 Description of Graphic


Symbols

A triangle indicates a danger against which


you should take precaution.
e.g.)
This is the “safety alert symbol.” This symbol alerts This symbol warns against “Danger of
you to matters and/or operation potentially hazardous Electrical Shocks”.
to yourself and other people. Read these messages A diagonal line indicates a prohibited course
and follow the instructions carefully. of action.
e.g.)
This symbol indicates you “Do Not
Disassemble”.
1.1.2 Warning Notice (Signal A solid circle indicates an imperative course
Words) of action.
e.g.)

Signal words indicate the degree of potential hazards This symbol indicates you must “Remove
Plug from Power Outlet”.
in the use of the product. Signal words include the
following three types, which are used according to
risk of damage caused by danger and the severity of Graphic
Explanation
damage. symbols

DANGER
Indicates a imminently hazardous situation which, if Indicates items that cannot be touched.
not avoided, will result in death or serious injury.

WARNING Indicates that AC power is being used.


Indicates a potentially hazardous situation which, if not
avoided, could result in death or serious injury.
Indicates the off position for the main power
CAUTION
switch of this device.
Indicates a potentially hazardous situation which, if
not avoided, could result in minor or moderate injury.
It may also be used to indicate hazardous situation Indicates the on position for the main power
where only physical damage is likely to occur. switch of this device.

Indicates that a plate is jammed inside the


device.

This CE mark on this product indicates


that this product is in conformity with
the applicable requirements set out in
the Directive 93/42/EEC
(Medical Device Directive) and in
0197 Directive 2011/65/EU (RoHS Directive).
0197 indicates the identification number
of the notified body responsible only for
implementation of the Directive 93/42/
EEC (Medical Device Directive).

12
1.2 Safety Precautions

Chapter 1
1.2.1 Overall Cautions 1.2.2 Electromagnetic Wave
Cautions
WARNING
• Though this device is classified as Class 1 Laser CAUTION
Product according to IEC 60825-1:1993+A1: (1) EMC Statement - The unit has been tested and
1997+A2:2001, IEC 60825-1:2007, it includes found to comply with the limits for medical de-
a laser unit with a maximum output of 130 mW vices to the IEC 60601-1-2:2007. These limits
(Class 3B). are designed to provide reasonable protection
Under no circumstances should external panels be against harmful interference in a typical medical
removed. installation. The unit generates uses and can ra-
Removal of panels may result in exposure to lasers diate radio frequency energy and, if not installed
or electrocution. and used in accordance with the instructions,
Device System Laser Unit Specifications: may cause harmful interference to other devices
Class: 3B in the vicinity. However, there is no guarantee
Medium: Semiconductor laser that interference will not occur in a particular in-
Wavelength: 659 nm stallation. If the unit does cause harmful interfer-
Max Output: 130 mW ence to other devices, which can be determined
• In the event of detection of smoke, unusual odors by turning the equipment off and on, the user is
or sounds, the device’s power supply circuit break- encouraged to try to correct the interference by
er should be turned off immediately and the power one or more of the following measures:
plug disconnected from the power outlet. Then con- -- Reorient or relocate the receiving device.
tact Konica Minolta technical representatives. Con- -- Increase the separation between the equipment.
tinued use under such circumstances may result in -- Connect the equipment into an outlet on a
fire. circuit different from that to which the other
• Do not let water, liquids, bits of metal, or other such device(s) are connected.
materials enter inside of the device. Doing so may -- Contact Konica Minolta technical representa-
cause fire, electric shock, or other damage. tives.
In the event of accidental introduction of such (2) EMC Statement - Take precautions against the
foreign objects or substances into the device, the unit especially regarding EMC. Install and put
power should be switched off immediately and the into service according to the EMC information
power plug disconnected from the power outlet. provided in the manual.
Then contact Konica Minolta technical representa- (3) EMC Statement - Do not use mobile phones or
tives. pocket pagers in the vicinity of the unit. Use of
• Do not disassemble or modify this device. Not only mobile phones or pocket pagers near the unit
will this cause the device to operate abnormally, but can cause errors in operation due to electromag-
may also cause fire or electric shock. netic wave interference: such devices should off
If the device is in need of repair, contact your Koni- in the vicinity of the unit.
ca Minolta technical representatives. (4) EMC Statement - Cable List.
-- AC Power Cable (3 m)
-- USB Cable (5 m, USB2.0 High-speed mode
compatible)
(5) EMC Statement - The use of accessories, trans-
ducers and cables other than those specified,
with the exception of transducers and cables
sold by Konica Minolta, Inc. as for internal com-
ponents, may result in increased emissions or
decreased immunity of the unit.
(6) EMC Statement - Do not use adjacent to or
stacked with other equipment. If adjacent or
stacked use is necessary, confirm normal opera-
tion in the configuration in which the unit will be
used.

13
1.2 Safety Precautions

CAUTION
(7) EMC Statement

Guidance and manufactureís declaration - electromagnetic emissions


The unit is intended for use in the electromagnetic environment specified below.
The customer or the user of the unit should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment - guidance
The unit uses RF energy only for its internal function.
RF emissions
Group 1 Therefore, its RF emissions are very low and are not likely to
CISPR 11
cause any interference in nearby electronic equipment.

RF emissions
Class B
CISPR 11

The unit is suitable for use in all establishments other than


Harmonic emissions
Class A domestic and those directly connected to the public lowvoltage
IEC 61000-3-2
power supply network that supplies buildings for domestic purposes.
Voltage fluctuations/
flicker emissions Complies
IEC 61000-3-3

(8) EMC Statement

Guidance and manufacture’s declaration - electromagnetic immunity


The unit is intended for use in the electromagnetic environment specified below.
The customer or the user of the unit should assure that it is used in such an environment.
Electromagnetic environment -
Immunity test IEC 60601 test level Compliance level
guidance
Electrostatic discharge +/- 6 kV contact +/- 6 kV contact Floors should be wood, concrete or
(ESD) ceramic tile. If floors are covered with
+/- 8 kV air +/- 8 kV air
IEC 61000-4-2 synthetic material, the relative humid-
+/- 2 kV for power sup- +/- 2 kV for power sup- ity should be at least 30 %. Mains
Electrical fast transient/
ply lines ply lines power quality should be that of a typi-
burst
+/- 1 kV for input/output +/- 1 kV for input/output cal commercial or hospital environ-
IEC 61000-4-4
lines lines ment.
+/- 1 kV differential mode +/- 1 kV differential mode Mains power quality should be that of
Surge
a typical commercial or hospital envi-
IEC 61000-4-5 +/- 2 kV common mode +/- 2 kV common mode
ronment.
<5% UT (>95% dip in UT) <5% UT (>95% dip in UT) Mains power quality should be that
for 0.5 cycle for 0.5 cycle of a typical commercial or hospital
Voltage dips, short in-
40% UT (60% dip in UT) 40% UT (60% dip in UT) environment. If the user of the unit
terruptions and voltage
for 5 cycles for 5 cycles requires continued operation during
variations on power sup-
70% UT (30% dip in UT) 70% UT (30% dip in UT) power mains interruptions, it is recom-
ply input lines
for 25 cycles for 25 cycles mended that the unit be powered from
IEC 61000-4-11
<5% UT (<95% dip in UT) <5% UT (<95% dip in UT) an uninterruptedly power supply or a
for 5 sec for 5 sec battery.
Power frequency magnetic fields
Power frequency (50/60
should be at levels characteristic of a
Hz) magnetic field 3 A/m 3 A/m
typical location in a typical commercial
IEC 61000-4-8
or hospital environment.
[NOTE] UT is the a.c. mains voltage prior to application of the test level.

14
1.2 Safety Precautions

CAUTION
(9) EMC Statement

Chapter 1
Guidance and manufacture’s declaration - electromagnetic immunity
The unit is intended for use in the electromagnetic environment specified below.
The customer or the user of the unit should assure that it is used in such an environment.
IEC 60601 Compliance
Immunity test Electromagnetic environment - guidance
test level level
Portable and mobile RF communications
equipment should be used no closer to any
part of the unit, including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency
of the transmitter.

Recommended separation distance


d=[1.2]√P

d=[1.2]√P 80 MHz to 800 MHz

Conducted RF 3 Vms d=[2.3]√P 800 MHz to 2.5 GHz


[3] V
IEC 61000-4-6 150 kHz to 80 MHz
where P is the maximum output power rating
Radiated RF 3 V/m of the S transmitter in watts (W) according to
[3] V/m the transmitter manufacturer and d is the rec-
IEC 61000-4-3 80 MHz to 2.5 GHz
ommended separation distance in meters (m).

Field strengths from fixed RF transmitters, as


determined by an electromagnetic site survey,
a
should be less than the compliance level in
each frequency range.b

Interference may occur in the vicinity of equip-


ment marked with the following symbol:

[NOTE] At 80 MHz and 800 MHz, the higher frequency range applies.
[NOTE] These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mo-
bile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the unit is used execeeds the appli-
cable RF compliance level above, the unit should be observed to verify normal operation. If abnormal performance
is observed, additional measures may necessary, such as reorienting or relocation the unit.
b Over the frequency range 150 kHz to 80 MHz, field strength should be less than [3] V/m.

15
1.2 Safety Precautions

CAUTION
(10) EMC Statement

Recommended separation distance between portable and mobile RF communications equipment and the
unit.
The unit is intended for use in and electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of the unit can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the unit as recommended below, ac-
cording to the maximum output power of the communications equipment.
Separation distance according to frequency of transmitter
Rated maximum output
m
power of the transmitter
150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
W
d=[1.2]√P d=[1.2]√P d=[2.3]√P
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 8
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in me-
tres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum
output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
[NOTE] At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
[NOTE] These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.

16
1.2 Safety Precautions

1.2.3 Cautions Relating to the 1.2.4 Device Installation,


Power Supply and the Movement, Storage and

Chapter 1
Power Cable Connection Cautions

WARNING CAUTION
• Under no circumstances should the ground be con- • Contact Konica Minolta technical representatives
nected to a gas pipe: this could cause a gas explo- when installing or moving the device.
sion. • When storing the device, make sure to fulfill the
• The following points should be strictly observed in “Environment” outlined in “Specifications.”
order to avoid the risk of electrocution or fire. • Use the USB cable specified by our company for
-- Always use the power cable provided with the device connection. Also, take the following precau-
device. tions when handling and connecting a USB cable.
-- Do not use the supplied power cable for other -- Do not use a damaged cable.
devices as the cable is for this device only. -- Be careful to ensure that no dirt or dust accumu-
-- Use a rated outlet for the power supply. lates on the connector.
-- Use a power supply with a ground wire included. -- Make sure the cable is firmly and properly con-
-- Do not use an extension cord. nected to the device.
-- Do not use multiple outlet extension plugs. -- Do not tug on the cable.
-- Make sure the power supply is firmly and prop- -- Do not twist the cable.
erly plugged into the outlet. -- Do not bundle the cable.
-- When the device is to be left unoperated for ex- -- Do not fix the cable in place using, for example,
tended periods of time, the power plug should be a stapler.
disconnected from the power outlet. -- Do not allow the cable to become trapped be-
-- Make sure that the device is not exposed to wa- tween two objects.
ter or other liquids. -- Do not place objects on the cable.
-- Never hold the power plug with wet hands. -- The cable should not be used near any heat gen-
-- Be careful to ensure that no dirt or dust accumu- erating devices.
lates on the power plug. -- Do not carry out modifications, repairs or other
processes on the cable.
• This device is not to be used in the presence of pa-
CAUTION tients at hospitals, clinics etc.

• In order to avoid the risk of electrocution or fire due


to damage to the power cable, the following cau-
tions should be strictly observed:
-- Do not tug on the power cable.
-- Do not twist the power cable.
-- Do not bundle power cables.
-- Do not fix the power cable in place using, for ex-
ample, a stapler.
-- Do not allow the power cable to become trapped
between two objects.
-- Do not place objects on the power cable.
-- The power cable should not be used near any
heat generating devices.
-- Do not carry out modifications, repairs or other
processes on the power cable.
A damaged power cable should be immediately re-
placed by a power cable of identical standard.
For details of replacement, contact Konica Minolta
technical representatives.

17
1.2 Safety Precautions

1.2.5 Cautions to Be Observed CAUTION


During Usage • Make sure the device is not set to a mode in which
image deletion occurs during image reading.
• To prevent image errors during image reading, en-
sure the device is not shaken or moved in any way.
WARNING
• If an X-ray generating device is positioned near this
• Do not use this device when lightning may strike. device, do not emit X-rays while this device is oper-
Doing so may cause fire or electric shock. ating. Doing so may cause an image error.
Make sure to unplug the power supply as well. If • Use of radio communication devices near the de-
thunder is heard while the power cable is still con- vice can cause errors in operation due to electro-
nected, do not touch either the unit itself or any of magnetic wave interference: such devices should
the cables. be switched off in the vicinity of the device.
• Do not use this device if damaged. Doing so may Devices which generate electromagnetic waves,
cause fire or electric shock. such as microwave therapy equipment, also may
Should repairs become necessary, contact Konica cause errors in operation of the device or in images
Minolta technical representatives. the device produces.
Contact Konica Minolta technical representatives.
• Do not drop anything other than cassettes insertion
CAUTION and ejection slot.
• Do not use the device with the air intake and vents
• Do not sit or climb on the device. There is risk of blocked or covers in place. Heat may accumulate
injury due to the fall or damage to the device. inside the device, which could cause damage to the
• Do not place objects on top of this device. There device.
is a risk of any objects placed on top of the device Make sure to allow space as shown below when
toppling and causing damage. using the device.
• When using this device, make sure to fulfill the “En- 0 mm or higher *
vironment” covered in “Specifications.” (130 mm or higher) 0 mm or higher *
Also, make sure the device is not installed in an
area under one or more of the following conditions.
Doing so may interfere with device functionality or 486 mm or
cause a fire. higher
-- Dusty
-- Direct sunlight
-- Unstable
-- Unlevel 0 mm or higher *
-- Vibration (130 mm or higher) 310 mm or higher
-- Poor ventilation * Distance from the device’s protruding portion.
-- Oily smoke or steam Value within () is the distance needed when
-- Extreme light/dark contrast opening a door for maintenance or troubleshooting.
• Make sure not to pinch your fingers while inserting
• If an error occurs for a read image, do not resume
the cassette.
reading with this device or exposure with the image-
• Do not use the device on its back, tilted or upside
reading plate until measures have been taken from
down.
this manual to solve the problem.
• Note the following instructions during device opera-
• When dealing with a plate jam (when the plate is
tion.
stuck inside), follow the process as outlined in this
-- Do not open the device door or cover.
manual to remove the plate. Ignoring the process
-- Do not turn OFF the power supply circuit breaker.
described in this manual and trying to remove the
-- Do not unplug the power cable from the device
plate forcefully may damage the plate or guide.
or outlet.
• Do not touch the yellow filter inside of this device.
If any of the actions above are performed, the ma-
• This is not an explosion-proof device, so do not use
chine will stop operating. Also, if performed during
it near flammable or explosive gases. Doing so may
image reading, images may be lost from the plate.
result in fire.

18
1.2 Safety Precautions

1.2.6 Cassette/Plate Storage CAUTION


and Usage Cautions • Note the following instructions when handling a

Chapter 1
plate.
-- Do not damage or soil it.
WARNING -- Do not bend it.
-- Do not leave the plate under a the fluorescent
• When the plate is damaged causing the fluorescent lamp or sunlight for an extended period of time.
substance to be exposed to the open air, the fol- -- Do not store where any sort of radiation is pres-
lowing cautions should be noted. ent.
-- Immediately stop using the device. -- Use and store in an environment between 10 to
-- If your eyes become contaminated, avoid rub- 30 degrees C, 15 to 80% RH (non condensing).
bing; quickly rinse your eyes with water, and a -- When storing a cassette with the plate loaded,
doctor should be consulted. make sure the tray (plate storage device) is com-
-- If the fluorescent medium is swallowed, a doctor pletely closed.
should be consulted after carrying out the emer- -- Stop using a damaged plate and replace it with a
gency measures. new one.
-- If the medium comes into contact with the skin, -- Stop using a plate of deteriorated quality, and re-
wash the affected area thoroughly with water. place it with a new one.
If the actions above are not taken, bodily damage • To maintain image quality, erase the plate’s im-
may occur. ages every morning before starting examinations.
• A plate with phosphor exposed must be tightly A plate in one of the following conditions may influ-
sealed in a plastic bag for disposal. Also, make ence image exposure, so be sure to erase plate im-
sure to wear rubber gloves when handling the ages before performing exposure.
plate. -- A cleaned plate
• When using the cassette to perform exposure, -- A plate unattended for a long period of time
contaminants (such as blood, bodily fluid, etc.) or -- A plate after multiple exposure or high-dose ex-
dust, debris, etc. may have adhered to its surface. posure is performed
Always clean off the cassette’s surface for each -- A plate after jam repairs are performed
examination and patient before performing image • Note the following instructions when handling a
reading with the device. cassette.
• If a contaminant (such as blood, bodily fluid, etc.) -- Do not damage or soil it.
touches the surface of the cassette, add a small -- Do not bend it.
amount of absolute ethanol or isopropyl alcohol to -- Do not expose it to heavy shock.
a nonwoven cloth, and use it to wipe off the con- -- Do not overload it.
taminant. -- Do not let water, liquids, pins, clips, or other such
• Use only the chemicals listed below for cassette objects, debris or dust inside.
cleaning: -- Keep it away from sterilization equipment, or
-- Absolute ethanol other such devices which use ultraviolet rays.
-- Isopropyl alcohol -- Do not leave in sunshine for a prolonged period.
• When using absolute ethanol or isopropyl alcohol -- Do not stack cassettes. Secure by placing the
for cleaning, make sure to follow the instructions blue corner block underneath and standing it
provided by each pharmaceutical company before there.
use. -- During use or storage, keep away from areas of
high temperature, high humidity, direct sunlight,
any form of radiation, and exposure to water.
CAUTION -- Use and store in an environment between 10 to
30 degrees C, 15 to 80% RH (non condensing).
• Make sure to hold the cassette firmly while han-
-- Stop using a damaged cassette and replace it
dling so as not to drop it. Dropping the cassette on
with a new one.
the foot may cause injury.
• When carrying a cassette, make sure its tray (plate
• Make sure not to pinch your fingers while loading
storage) is closed tightly.
the plate into the cassette.
• When inserting a cassette into the device, make
• Only use the cassette and plate specified by our
sure its tray (plate storage) is closed tightly.
company with this device.

19
1.2 Safety Precautions

CAUTION 1.2.7 Device Care and


• Do not disassemble or modify the cassette. Contact Maintenance Cautions
Konica Minolta technical representatives when a
repair should be required.
• During exposure, make sure the back side of the CAUTION
cassette (with the lever, etc.) is facing away from
the patient. • When taking care of or maintaining the device, fol-
Make sure the cassette’s blue corner block is fac- low the process as outlined in this manual.
ing downward, or facing to the right with respect to • Perform maintenance periodically.
the upward direction when output is desired, when • When opening the device door or cover during
performing exposure. cleaning or troubleshooting, make sure that no de-
• Note the following instructions when loading a plate bris, dust, or foreign materials enter.
into the cassette.
-- When the tray (plate storage) is pushed out with
the plate release lever, do not pull it out further by
force. Pulling it out by force may damage internal
components beyond repair. 1.2.8 Disposal Cautions
-- Make sure the right type of plate is combined
with the cassette.
-- Check the plate’s dimensions to make sure it is CAUTION
loaded in the right direction.
-- After loading the plate, make sure the tray is • Please follow the rules and regulations of your rel-
closed securely. evant authorities in the disposal of this product, ac-
cessories, products sold separately, consumables,
media and their packing materials.

For EU member states only


This symbol means: Do not dispose of this product
together with your household waste!

Please refer to the information of your local


community or contact our dealers regarding the
proper handling of end-of-life electric and electronic
equipments.

Recycling of this product will help to conserve


natural resources and prevent potential negative
consequences for the environment and human
health caused by inappropriate waste handling.

20
Chapter
2
Product Outline
An explanation of the device’s features and functions.

21
2.1 Device Summary

An explanation of the device’s functions and features, along with system configuration examples using this device.

2.1.1 Functions and Features

This device (a reader) is used for reading X-ray image data taken on an exclusive plate.
The exclusive plate is bent like film, and this device reads images from it as they are fed in. Therefore, this device is
miniaturized so that it can be installed on a table.
Receiving and all setting conditions for read image data is performed by a console called the “Image processing con-
troller,” which is sold separately.

CAUTION
• This device can only use an exclusive cassette and plate combination. The cassette and plate are designed specifically for
use with this device, so they cannot be used with other readers.

2.1.2 System Configuration Examples

System configuration examples using this device are provided below.

1-to-1 Connection

ImagePilot/CS-7 REGIUS SIGMA This is the most basic configuration. One Im-
age processing controller console is connect-
ed to this device via USB cable (included).
When using the CS-7 as the Image process-
USB2.0 ing controller, only a 1-to-1 connection sys-
tem configuration is possible.

LAN Cable HUB

M-to-1 Connection

This system configuration is made up of


REGIUS SIGMA multiple ImagePilot consoles, connected to a
single device. The device is connected to a
ImagePilot client via USB cable (included).
USB2.0 (Multiple devices cannot be connected in a
single system.)
ImagePilot
Client HUB
IMPORTANT
LAN Cable
For externally • Please use ImagePilot user tools with an
connected devices ImagePilot server, even when the REGIUS
such as printer, etc. SIGMA is connected to an ImagePilot client.

LAN Cable
ImagePilot Server

22
2.2 Names of Parts and Functions

2.2.1 Front / Left Side / Top

Chapter 2
(6) Intake Vent (1) Top Door

Internal (2) Cassette


Area Insertion Slot
(3) Status Lamp
(4) Erase Button
(5) Exhaust Outlet
(10) Light Collection Unit
Cleaning Handle
(11) Unjamming Dial (6) Intake Vent
(Bottom Surface)

(7) Power Supply


Circuit Breaker

(9) Left Door (8) USB Cover

Number Name Function Reference


“5.2 Dealing with Plate
This can be opened to unjam a plate, or forcibly remove a Jams”(p. 54)
(1) Top Door
loaded cassette. “5.3 Forcibly Removing
the Cassette”(p. 61)
“3.2 Image Erase
Operation”(p. 30)
(2) Cassette Insertion Slot The cassette is inserted here.
“3.3 Image Read
Operation”(p. 32)
(3) Status Lamp Indicates device status and errors. “2.3 Status Lamp”(p. 25)
“3.2 Image Erase
(4) Erase Button Pressing this button will erase image data from the plate.
Operation”(p. 30)
(5) Exhaust Outlet Air is emitted through here from the inside to keep it cool. “6.2.3 Intake Vent, Ex-
haust Outlet Cleaning”(p.
(6) Intake Vent Takes in air from the outside to keep the inside cool. 65)
Turns the device power ON/OFF.

(7)
Power Supply Circuit CAUTION “3.1 Power ON”(p. 28)
Breaker • Do not turn the power OFF during operation except in “3.4 Power OFF”(p. 34)
case of an emergency.

Removing the cover reveals a USB port for connecting the Im-
(8) USB Cover -
age processing controller and USB cable (included).
This can be opened to unjam the plate, or to clean light collec-
(9) Left Door -
tion unit.
“6.2.4 Light Collection Unit
Light Collection unit
(10) Used to clean light collection unit. Cleaning (Keeping Them
Cleaning Handle
Dust Free)”(p. 65)
“5.2 Dealing with Plate
(11) Unjamming Dial Used to unjam the plate.
Jams”(p. 54)

23
2.2 Names of Parts and Functions

2.2.2 Rear / Right Side

(1) Power Supply Port

(2) Rating Plate


(4) Exhaust Outlet

(3) Rear Door

Number Name Function Reference


(1) Power Supply Port The power cable (included) connects here. -
(2) Rating Plate The device’s rating is indicated here. -
“5.2 Dealing with Plate
(3) Rear Door Can be opened to unjam the plate.
Jams”(p. 54)
“6.2.3 Intake Vent,
(4) Exhaust Outlet Air is emitted through here from the inside to keep it cool. Exhaust Outlet
Cleaning”(p. 65)

24
2.3 Status Lamp

Device status can be checked according to how the Status Lamp is lit.

(1)
Ready

Chapter 2
(2)
Busy

(3)
Error

(4)
Erase

Erase Button

Lamp
Number Name (Color) Device Status
Status
Ready Lamp Lit Indicates that operation can begin.
(1)
(Green) Blinking Indicates that operation has ended, and the cassette can be removed.
Operation is in progress.

Lit CAUTION
• Do not touch this device or the table it is positioned on during operation.

Busy Lamp
(2) Blinking Preparing for operation.
(Blue)
Operation is halted.

Unlit HINT
• Turn OFF the power when in this state.

Error Lamp Lit Indicates a warning. Operation may continue in some cases, or be halted in others.
(3)
(Yellow) Blinking An error has occurred. Operation is halted and cannot resume.
An image is being erased.

CAUTION
• Do not touch this device or the table it is positioned on during operation.
Lit

HINT
Erase Lamp • The Erase Lamp will also be lit while pressing the Erase Button.
(4)
(Orange)

The device is in image erase mode.

Blinking
CAUTION
• Loading the cassette in this situation will trigger the image erase process. Before
performing the image reading process, make sure the Erase Lamp is unlit.

REFERENCE
• If the Erase Lamp is lit or blinking, refer to "5.1 Troubleshooting"(p. 46) for troubleshooting.
• Refer to “3.2 Image Erase Operation”(p. 30) for details on Image Erase Mode.

25
2.4 Operation Flow

The plate inside of the device to be used for the image reading process is described as follows. (The charts below
are shown from the left side of the device.)

1 Load the cassette.


Cassette
HINT
• Load the cassette when the Ready Lamp (green) is lit.

Plate

Tray opens
2 The plate is pulled in once the Image processing
controller’s status is confirmed.
HINT
• The Busy Lamp (blue) will blink while the Image processing
controller’s status is being confirmed.
• The Busy Lamp is lit during the image reading process.

3 Image reading takes place while the plate is fed


to the bottom of the device.

Light Collection Unit


Images are read.

4 Once image reading has completed, the plate is


fed back into the cassette as images are erased.

Erase Unit
Images are erased.

5 Once the plate has gone back inside the cas-


sette, the cassette can be removed.
HINT
• When the image reading process ends, the Busy Lamp (blue)
will turn off, and the Ready Lamp (green) will start blinking.
When the cassette is removed, the Ready Lamp will turn on.

26
Chapter
3
Reader Operation
An explanation on how to operate this device.

27
3.1 Power ON

An explanation on how to start up this device. Starting Up the SIGMA


The relation between SIGMA startup and Image pro-
1 Make sure the power cable is connected. cessing controller startup will be described.
This device can be turned ON/OFF regardless of the
2 Turn ON the device’s power supply circuit Image processing controller’s startup status.
breaker.
If only performing the image erase process
The Image processing controller does not have to be
Ready Lamp
started up. The image erase process can also be per-
formed while the Image processing controller is not
connected.
• If the Image processing controller has not started up, or
is not connected to this device, the Error Lamp (yellow)
will be lit.

If performing the image read process


Power Supply
Circuit Breaker The Image processing controller must be started up.
Also, this device must be connected to the Image pro-
cessing controller via the included USB cable.
• The device will start up, and initialization will be- • The Image processing controller can be started up be-
gin. fore or after this device. Make sure the USB cable is
• Once in a usable status, the Ready Lamp (green) connected properly before startup.
will be lit.

HINT
• All lamps will be lit immediately after turning on the
device. If any lamps are unlit, contact Konica Minolta
technical representatives.
• To maintain image quality, erasing of the plate is rec-
ommended before beginning the examination.

28
3.1 Power ON

ImagePilot displays for this device CS-7 displays for this device

Icon & Progress Display Icon & Progress Display


An icon for this device is located in the task tray of the An icon for this device is located in the task tray of the
ImagePilot it is connected to. CS-7 it is connected to.
The icon is displayed as shown below according to The icon is displayed as shown below according to
connection status. connection and reading status.

When disconnected, or when


the device has not started When connected normally. When connected normally. This device is processing.
up.

Chapter 3
Right clicking the icon will display a menu, and menu
item is explained as follows.
Menu items Content When disconnected, or
when the device has not Error condition.
Toggle the progress display ON/
started up.
OFF. When a check mark is dis-
played to the left of the item name,
it is ON. Further, when this device is processing, a progress
Display Progress display dialog box is displayed on the screen.
When turned ON, progress status
will be displayed in a dialog box
above the task tray during device
operation.

Example: Progress dialog box


• You can move the progress display dialog box to
wherever you like on the screen by dragging it.

Message Display
If a connection between the device and CS-7 is
• Click [X] to close the dialog box. If you want to display established, or if a network error or communication
the dialog again, turn “Progress display” OFF at the error occurs, a message will be displayed in the dialog
menu, then back ON. box at the bottom of the screen.
Example:
Message Display
If a connection between the device and ImagePilot is
established, or if a network error or communication er-
ror occurs, a message will be displayed in the dialog
box above the task tray.
Example: • For errors other than network errors or communication
errors, an error message will be displayed in the regular
CS-7 dialog box.

• For errors other than network errors or communication


errors, an error message will be displayed in the regular
ImagePilot dialog box.

29
3.2 Image Erase Operation

Plate images can be erased by performing the HINT


following steps. • The Erase Lamp will be lit while pressing the erase
button.
CAUTION • In order to prevent operational errors from occurring, if
• Make sure not to pinch your fingers while inserting the the cassette is not inserted within 5 seconds from the
cassette. beginning of Image Erase Mode, the device will return
to Image Read Mode.

HINT
• If only performing the image erase process, the Image 5 Insert the cassette in the cassette inser-
processing controller does not have to be started up. tion slot.

CAUTION
1 Prepare a cassette with an image erasing
• Do not insert anything other than the cassette in the
plate loaded.
cassette insertion slot. Doing so may cause a malfunc-
If there is a memo between the cassette’s memo
tion.
clip, remove it. If there is a sticker etc. attached to
If a foreign object has fallen inside of the device, con-
the cassette, remove it.
tact Konica Minolta technical representatives.

CAUTION
• Only cassettes specially made for this device can be Position the back side of the cassette facing up-
used. ward, and slowly slide it into place, straight and all
the way to the back, in the direction of the arrow.

2 Confirm that the Image processing con-


troller is not in image capture standby
mode (that the CR “Modality” button has
not been clicked).

REFERENCE
• For details, refer to the Image processing controller
Operation Manual.

Insert so that
3 Confirm that the device’s Ready Lamp center-sections
(green) is lit. match.

• The image erase process will begin.


4 Hold down the erase button for at least 2 • The Erase Lamp (orange) is lit while images are
seconds. being erased.
• The device will go into Image Erase Mode, and
HINT
the Erase Lamp (orange) will start blinking.
• If the cassette was inserted before pressing the erase
button, holding down the erase button for at least 5
seconds will start up the image erase process.

Erase Button

Erase Lamp

30
3.2 Image Erase Operation

6 Wait for the image erase process to


complete.

CAUTION
• Do not touch the device, the inserted cassette, or the
platform/table the device is set on while images are
being erased.

• When the image erase process has completed,


the Ready Lamp (green) will start blinking to in-
dicate that the cassette can be removed.

7 Confirm that the Ready Lamp is blinking,

Chapter 3
then remove the cassette.
• When removing the cassette, the device initial-
ization process will begin, and the Busy Lamp
will be lit.
• Once the initialization process has ended, the
Busy Lamp will turn off, and the Ready Lamp
will be lit.

IMPORTANT
• Do not insert the cassette during the initialization pro-
cess (while the Busy Lamp is lit). Doing so will cause
initialization to end abnormally, and an error will occur.

HINT
When an image remains after erasing once
• When the image erase process was performed once
for an image exposed at high dosage, etc., but was
not completely erased, the second image erase pro-
cess will be performed.
There are two ways to execute the second erase pro-
cess: “Auto Run” and “Display message then se­lect
run”. The setting can be changed by the user tool.
Refer to the Image processing controller Operation
Manual.

When responding to a message to perform the


second erase process
• If you need to perform the second erase process, the
Error Lamp (yellow) will light up, and an image re-
erase request message will be displayed in the Image
processing controller.
• Clicking the [Continue] button at the message will start
the image erase process.
• Clicking the [Eject] button will eject the cassette, and
the second erase process will not be performed. When
doing this, make sure to perform the image erase pro-
cess at a later time.
• Even when selecting the message display option, the
second erase process will be performed automatical­ly
if Image processing controller has not started up or is
not connected to the device.

31
3.3 Image Read Operation

The following steps explain how to read exposed plate 3 Confirm that the device’s Ready Lamp
images with this device. (green) is the only lamp that is lit.

CAUTION CAUTION
• Make sure not to pinch your fingers while inserting the • Confirm that the Erase Lamp (orange) is not blinking.
cassette. If it is blinking, it is in Image Erase Mode. In this case,
• If an X-ray generating device is positioned near this press the erase button, or wait up to 5 seconds with-
device, do not emit X-rays while this device is operat- out inserting the cassette, and the device will return to
ing. image read mode.

1 Prepare a cassette with an image reading 4 Insert the cassette in the cassette inser-
plate loaded. tion slot.
If there is a memo between the cassette’s memo
clip, remove it. If there is a sticker etc. attached to CAUTION
the cassette, remove it. • Do not insert anything other than the cassette in the
cassette insertion slot. Doing so may cause a malfunc-
CAUTION tion.
• Only cassettes specially made for this device can be If a foreign object has fallen inside of the device, con-
used. tact Konica Minolta technical representatives.
• Make sure no dust or debris is on the cassette’s front
surface. If any is found, be sure to clean it off.
Position the back side of the cassette facing up-
ward, and slowly slide it into place, straight and all
the way to the back, in the direction of the arrow.

If any foreign materials are on this surface,


the plate may not eject properly, and images
may be taken unevenly.
Insert so that
center-sections
match.
2 Put the Image processing controller in im-
age capture standby mode (click the CR • Image reading will begin once the Image pro-
“Modality” button). cessing controller’s status is confirmed.
• The Busy Lamp (blue) will blink while the Im-
REFERENCE age processing controller’s status is being con-
firmed.
• For details, refer to the Image processing controller
• The Busy Lamp is lit during the image reading
Operation Manual.
process.

HINT
• When inserting the cassette before the Image pro-
cessing controller is in image capture standby mode,
changing the Image processing controller to image
capture standby mode will start the image capture pro-
cess.

32
3.3 Image Read Operation

5 Until the image reading process has ended, HINT


and the Busy Lamp (blue) has turned off, When an image remains after being erased once
do not touch the device or cause it to • After reading an image, the image erase process will
vibrate. be performed. When the image erase process was
performed once for an image exposed at high dosage,
etc., but was not completely erased, the second image
erase process will be performed.
There are two ways to execute the second erase pro-
cess: “Auto Run” and “Display message then se­lect
run”. The setting can be changed by the user tool.
Refer to the Image processing controller Operation
Manual.

When responding to a message to perform the

Chapter 3
second erase process
• If you need to perform the second erase process, the
Error Lamp (yellow) will light up, and an image re-
CAUTION erase request message will be displayed in the Image
processing controller.
• Do not touch the device, the inserted cassette, or the
• Clicking the [Continue] button at the message will start
platform/table the device is set on during the image
the image erase process.
reading process. If other device(s) are placed on the
• Clicking the [Eject] button will eject the cassette, and
same platform, make sure their vibration does not
the second erase process will not be performed. When
interfere with this device. Vibration may cause the im-
doing this, make sure to perform the image erase pro-
age to shake.
cess at a later time.
• Do not open or close any device doors during the im-
age reading process. Doing so will cause the device to
stop working, which may result in a plate jam or image
loss.
When an error occurs for a read image
If an error occurs for a read image, take the following
steps. Do not resume reading with this device or expo-
• When the image reading process and image sure with the image-reading plate until the issue has
erase process end, the Ready Lamp (green) will been resolved.
start blinking, which indicates that the cassette
can be removed. When a white line error occurs
Light collection unit might be soiled with dust or debris.
6 Confirm that the Ready Lamp is blinking, Clean the light collection unit.
then remove the cassette.
• When removing the cassette, the device initial- REFERENCE
ization process will begin, and the Busy Lamp
• Refer to "6.2.4 Light Collection Unit Cleaning (Keeping
will be lit.
Them Dust Free)"(p. 65)
• Once the initialization process has ended, the
Busy Lamp will turn off, and the Ready Lamp
will be lit. When other errors occur
The plate might be dirty. Clean the plate.
IMPORTANT
HINT
• Do not insert the cassette during the initialization pro-
cess (while the Busy Lamp is lit). Doing so will cause • White streaks may form roughly 2mm from the edge of
initialization to end abnormally, and an error will occur. the image, but this is not considered abnormal.

REFERENCE
• Refer to "6.3.1 Plate Cleaning"(p. 69)

33
3.4 Power OFF

An explanation on how to shut down this device.

1 Confirm that no cassette is inserted in the


device, and that the Busy Lamp (blue) is
unlit.

2 Turn OFF the power supply circuit breaker.

Busy Lamp

Power Supply
Circuit Breaker

CAUTION
• Unplug the power supply when not in use for a long
period of time.

34
Chapter
4
Cassette/Plate Operation
An explanation on cassette and plate handling.

35
4.1 Cassette/Plate Handling

The following is a description for handling the cassette and plate.


Read "1.2.6 Cassette/Plate Storage and Usage Cautions"(p. 19) carefully for correct handling procedures.

4.1.1 Cassette/Plate Basics

Names and Structure of Cassette Parts

Cassette Front Plate


Surface
(1) Tray

With tray open

(2) Cassette Lock Release Buttons


(3) Cassette Lock Release Levers

(4) Plate Release Lever


Cassette Back
Surface

(5) Memo Clip


(6) Cassette Label
(9) Cassette Type (7) The plate’s “first day” label
(8) Corner Blocks

Number Name Function


(1) Tray Secures the plate.
(2) Cassette lock release button Used to remove the plate.
Used to remove the plate.
(3) Cassette lock release lever
Not included with some cassette types.
(4) Plate release lever Used to remove the plate.
A memo with the patient’s name, etc. can be held here.
(5) Memo clip
Not included with some cassette types.
(6) Cassette label Indicates the cassette’s first day of use.
A label included with the plate, on which the first day of use should be entered before
(7) The plate’s “first day” label
attaching it.
(8) Corner blocks (blue) Used to confirm the cassette’s direction during exposure.
(9) Cassette type Indicates the type of cassette.

36
4.1 Cassette/Plate Handling

Entering the first day of use Cassette direction during exposure


When first using the cassette and plate, please enter
the date on each label. CAUTION
If the date is not entered on these labels, warranties Make sure the back surface of the cassette (with the
may not be applied. lever, etc.) is facing away from the patient when per-
• The labels come with a protective film, so make sure to forming exposure.
attach it after entering the date. Make sure the cassette’s blue corner block is facing
• Make sure to firmly attach the plate’s “first day” label to downward, or facing to the left with respect to the up-
the cassette so it cannot peel off. ward direction when output is desired, when perform-
• When replacing the plate with a new one, make sure ing exposure.
to enter the new plate’s first day of use on its “first day”
label before attaching it. ● When exposure is done from
vertical position
Erasing images on a plate

CAUTION X-ray radiation

To maintain image quality, erase the plate’s images ev-


ery morning before starting examinations.
A plate in one of the following conditions may influence

Chapter 4
image exposure, so be sure to erase plate images be-
fore performing exposure. Blue corner blocks
• A cleaned plate are downward
• A plate unattended for a long period of time
● When exposure is done from
• A plate after multiple exposure or high-dose exposure is
horizontal position
performed
• A plate after jam repairs are performed
Also, if an image erase request message is displayed on X-ray radiation
the Image processing controller’s screen, make sure to
perform image erasing.

Image area Blue corner blocks


The scope of space guaranteed for image area is as are facing to the
follows for the cassette. left

1mm
Cassette/Plate Maintenance

IMPORTANT
To use the cassette and plate safely, periodic mainte-
2mm Image Area 2mm
nance is required.
• Refer to “6.3.1 Plate Cleaning”(p. 69) and “6.3.2 Cas-
sette Cleaning”(p. 69) for cleaning methods.

1mm

37
4.1 Cassette/Plate Handling

If the cassette has cassette lock release le-


4.1.2 Removing the Plate From vers
the Cassette Cassette Lock Release
Levers

CAUTION
• Make sure your hands are clean before handling the
plate to prevent it from getting dirty. And refrain from
touching the plate unnecessarily.

1 Use a flat table that is wider than the cas- Attach the opening tools for cassette’s protruding
portions to the cassette lock release levers, then
sette when performing this task.
slide the tool to insert the tool’s two notches into
the cassette lock release buttons.
2 Turn the cassette so that its back surface
is facing upward, then place it on the ta- Cassette Lock
ble. Release Levers

3 Attach the opening tools for cassette to


the cassette.
Setup methods will differ depending on whether or Notch
not cassette lock release levers are present.
• A cassette with cassette lock release levers
14”×17”
14”×14”
11”×14”
10”×12” Opening tools for cassette
24×30cm
• A cassette with no cassette lock release levers
8”×10”
18×24cm
15×30cm

Cassette Lock Release


Buttons

38
4.1 Cassette/Plate Handling

If the cassette does not have cassette lock 4 With the opening tools for cassette at-
release levers tached, move the plate release lever out-
ward.

Plate Release Lever

Slide the tool’s two notches into the cassette lock


release buttons.

The plate will be pushed to the outside along with


its tray.

Chapter 4
CAUTION
When the tray is pushed out with the plate release le-
ver, do not pull it out further by force. Pulling it out by
force may damage internal components beyond repair.
Opening tools
for cassette Cassette Lock Release
Buttons

Tray

39
4.1 Cassette/Plate Handling

5 With the opening tools for cassette still


attached, turn the cassette over so that its 4.1.3 Loading the Plate into the
front surface is facing upward. Cassette
Front Surface

CAUTION
• Only use a plate specially made for this device.
• Make sure your hands are clean before handling the
Cassette Opening tools for cassette
plate to prevent it from getting dirty. Do not scratch or
bend the plate.
• Use a cassette that matches the plate size.
• Make sure there is no dirt or debris on the plate before
Front Surface loading it.

1 Use a flat table that is wider than the cas-


sette when performing this task.
• If the cassette’s tray is closed, go to step 2.
• If the cassette’s tray is open, go to step 6. If a
opening tools for cassette is attached to the cas-
sette, make sure to remove it.

2 Turn the cassette so that its back surface


6 Push the tray down slightly, and remove is facing upward, then place it on the table.
the plate from it.
The plate is secured to its tray by two notches.
3 Attach the opening tools for cassette to
the cassette.
Setup methods will differ depending on whether or
not cassette lock release levers are present.
• A cassette with cassette lock release levers
14”×17”
14”×14”
11”×14”
10”×12”
24×30cm
• A cassette with no cassette lock release levers
8”×10”
Notches 18×24cm
15×30cm

7 Pull the plate out of the cassette.

Plate

8 Remove the opening tools for cassette.

40
4.1 Cassette/Plate Handling

If the cassette has cassette lock release le- If the cassette does not have cassette lock
vers release levers
Cassette Lock Release
Levers

Slide the tool’s two notches into the cassette lock


Attach the opening tools for cassette’s protruding release buttons.
portions to the cassette lock release levers, then
slide the tool to insert the tool’s two notches into
the cassette lock release buttons.

Cassette Lock
Release Levers

Chapter 4
Notch

Opening tools
for cassette Cassette Lock Release
Buttons
Opening tools for cassette

Cassette Lock Release


Buttons

41
4.1 Cassette/Plate Handling

4 With the opening tools for cassette attached, 7 Reposition the cassette so that its front
move the plate release lever outward. surface is facing upward, and the tray can
move past the outer edge of the table.
Plate Release Lever

Front Surface

The tray will be pushed out.


8 Put the plate inside of the cassette, and
CAUTION secure it to the tray’s two notches.
When the tray is pushed out with the plate release le-
ver, do not pull it out further by force. Pulling it out by CAUTION
force may damage internal components beyond repair. • Make sure the plate is facing in the same direction as
the cassette.
The plate’s white surface is considered the front.
• Check the plate’s length and width and make sure it is
loaded in the correct direction.

Plate (white surface up)

Tray

5 Remove the opening tools for cassette. Tray

6 Hold the cassette so that the tray comes


down and shake it a bit.
This action is performed to make sure no foreign
matter is inside the cassette, and to remove any
dust or debris that may be inside.

Notches

42
4.1 Cassette/Plate Handling

9 Push the tray and plate into the cassette.


A click can be heard once the tray is inserted com-
pletely.

10 Make sure the tray is completely closed.

Chapter 4
Make sure there is no space
between the tray and cassette.

• If there is space between the tray and cassette,


open the tray and reload the plate.

11 Perform the image erase process.


CAUTION
• After the plate has been loaded, make sure to perform
the image erase process.
Refer to “3.2 Image Erase Operation”(p. 30)

43
4.2 Dealing with a Broken Cassette/Plate

CAUTION
• If the plate’s front surface is significantly scratched, or the protective film on its phosphor side has been scratched or bro-
ken, immediately stop using it, and replace it with a new plate.
• If the cassette’s shape is altered due to being dropped etc., this may cause problems to exposed images, so immediately
stop using it and replace it with a new cassette.
• If an image is damaged (flawed), clean the plate and the cassette’s interior.
Make sure there are no scratches, debris, or foreign objects on the plate’s surface, refer to "6.3.1 Plate Cleaning"(p. 69),
and then start cleaning.
Make sure a plate is not loaded in the cassette, then perform the following steps to make sure there is no dust or debris
inside.
-- Open the tray, place the tray’s opening face down, then lightly tap the cassette to remove any dust or debris.
-- Use a vacuum, etc. to suck out any dust or debris that might remain inside the tray’s opening.
Perform a test exposure after cleaning, and if a diagnosis problem occurs, immediately stop using it and replace it with a
new one.
• If scratches, dirt, altered shape, discoloration, reduction in sensitivity, etc. lead to a reduction in diagnosis image quality,
or interfere with device operation, immediately stop using it and replace it with a new one.

44
Chapter
5
Troubleshooting
An explanation on how to deal with errors when they occur.

45
5.1 Troubleshooting

IMPORTANT
• If a problem occurs during the image read process, reexposure of the image is required after discarding it and performing
the image erase process for the plate.

If a warning or error occurs for this device, the Error Lamp (yellow) will light up or start blinking. Also, a message will
be displayed on the Image processing controller.
Messages will either be displayed in the center of the screen, or on the task tray.
Check the following for reference, and take appropriate action.

Error Lamp status Reference


The device’s Error Lamp is lit or blinking. Message is not dis-
played on the Image processing controller. “5.1.1 When the Error Lamp is lit or blinking, but no mes-
(Including when the Image processing controller has not start- sage is displayed”(p. 46)
ed up or is starting up.)
Message is displayed on the Image processing controller. “5.1.2 When a message is displayed”(p. 47)
The device is not operating normally, but the Error Lamp is
“5.1.3 When the Error Lamp is unlit, and no message is
unlit. Message is not displayed on the Image processing con-
displayed”(p. 53)
troller.

If taking measures as described here does not solve the problem, contact Konica Minolta technical representatives.

5.1.1 When the Error Lamp is lit or blinking, but no message is dis-
played

When the Error Lamp is lit


If this device is not connected to the Image processing controller, the Error Lamp will light up.
The Error Lamp will also light up when connected, even when the Image processing controller has not started up, or
when the Image processing controller is starting up but not connected.

When only performing the image erase process


The device can be used while the Error Lamp is lit.

When performing the image read process


Make sure this device is properly connected to the Image processing controller, then start up the Image processing
controller. Once this device has been connected to the Image processing controller, the Error Lamp will turn off.

When the Error Lamp is blinking


An error has occurred for this device. Take the following action.
1. Refer to “5.2 Dealing with Plate Jams”(p. 54) while performing the image erase or image read process, and remove the
plate and cassette from this device.
2. Turn the device power OFF.
3. Confirm that the Image processing controller and this device are connected, then start up the Image processing con-
troller. (If already started up, it does not have to be restarted.)
4. Turn the device power ON, and make sure that no errors have occurred.

46
5.1 Troubleshooting

5.1.2 When a message is displayed

Messages will be displayed either in the center of the screen, or on the task tray.

Message displayed in the center of the screen

Message Appropriate action


• If you are not going to perform an image read, click the
Study is not registered. To eject the cassette, click on [Eject]
[Eject] button and remove the cassette.
button.
• If you are going to perform an image read, click the [OK]
In order to perform image retrieval, click on [OK] button to
button to close the dialog, then click the [CR] button on
close this dialog, and then click on [CR] button.
the Image processing controller.
Study is not registered. In order to perform image retrieval,
Click the [OK] button to close the dialog, then click the [CR]
click on [OK] button to close this dialog, and then click on [CR]
button on the Image processing controller.
button.
If [CR] is displayed in the message’s device name column,
an error has occurred for this device.
Check the error information column, refer to the “Error In-
Device error occurred. [Device name] [Error Information]
formation List”(p. 48), and take the appropriate action. After
taking action, click the [Yes] button to close the message
dialog.
If [CR] is displayed in the message’s device name column,

Chapter 5
a warning has occurred for this device.
Device warning occurred. [Device name] [Warning Informa- Check the warning information column, refer to the “Warn-
tion] ing Information List”(p. 47), and take the appropriate action.
After taking action, click the [Yes] button to close the mes-
sage dialog.
• To re-erase an image, click the [Continue] button.
• If you do not wish to re-erase an image, click the [Eject]
button to remove the cassette. By pressing the [Eject]
Need Re-erasing by over X-ray exposure.
button, you can eject the cassette without performing the
erase process a second time. When doing this, make
sure to perform the image erase process at a later time.

Warning Information List

Warning Information Appropriate action


Refer to “5.3 Forcibly Removing the Cassette”(p. 61), eject
the cassette, and then check the tray’s status (refer to the
"10 Make sure the tray is completely closed."(p. 43) steps for
Failed to take the plate out of the cassette. Open upper door
“4.1.3 Loading the Plate into the Cassette”(p. 40).
and remove cassette.
If the tray is closed, start the reading process again.
If the tray was not closed, reload the plate, then erase the
plate’s image via image erase mode.
Cassette insertion is not proper enough. Push it moreover. Press the cassette in all the way.
Refer to “5.3 Forcibly Removing the Cassette”(p. 61), eject
Cannot start cassette process because of overheating. Open the cassette, then restart the reading process after waiting
upper door and remove cassette Try it again after a while. a bit.
(Images from the plate within the cassette can be reread.)

47
5.1 Troubleshooting

Warning Information Appropriate action


The LED used to erase the image on the plate has ex-
ceeded its lifespan. This may lead to a malfunction during
Erase lamp life had exceeded. Please call service to get lamp
the image erase process.
changed.
Contact Konica Minolta technical representatives. and re-
place the LED.
Cleaning of a roller is required. The roller is dirty. Please clean the roller.

Error Information List


Error information is displayed as follows.
Error Occurs [ (message) ERR_CODE = (code)].

Error Information
Error Appropriate action
Message
Code
Initialization failed. Cassette is set in the reader.
7000 Please open the door and remove the Cassette
and Plate.
7001 Initialization failed. Plate still might be inside the Refer to “5.2 Dealing with Plate Jams”(p. 54) and remove
7002 reader. Please open the door and check inside. the plate and cassette from this device.
Initialization failed. Cassette is set in the reader. Use image erase mode to erase the image from the re-
7003
Please open the door and remove the Cassette moved plate.
- 7005
and Plate.
Initialization failed. Plate still might be inside the
7006
reader. Please open the door and check inside.
7007 Initialization failed. Sensor Error was detected Open and close the top door. The device will be re-initial-
-7016 during initialization. ized, so check to see if the error occurs again.
Error was detected while processing the cassette. Refer to “5.2 Dealing with Plate Jams”(p. 54) and remove
Please open the door and remove the Cassette the plate and cassette from this device.
7017
and Plate. After removal, erase the cassette in Use image erase mode to erase the image from the re-
ERASE MODE. moved plate.
7018 Initialization failed. Sensor Error was detected Open and close the top door. The device will be re-initial-
7019 during initialization. ized, so check to see if the error occurs again.
Initialization failed. Cassette was detected. Please Refer to “5.3 Forcibly Removing the Cassette”(p. 61) and
7020
open the top door and remove the Cassette. remove the cassette from this device.
Initialization failed. Sensor Error was detected Open and close the top door. The device will be re-initial-
7021
during initialization. ized, so check to see if the error occurs again.
Initialization failed. Sensor Error was detected
7023 Contact Konica Minolta technical representatives.
during initialization.
Refer to “5.3 Forcibly Removing the Cassette”(p. 61), re-
7100
Error was detected during Cassette insertion. move the cassette, then restart the image reading process.
- 7106
(Images from the plate within the cassette can be reread.)
Error was detected while processing the cassette. Refer to “5.2 Dealing with Plate Jams”(p. 54) and remove
7107 Please open the door and remove the Cassette the plate and cassette from this device.
- 7109 and Plate. After removal, erase the cassette in Use image erase mode to erase the image from the re-
ERASE MODE. moved plate.
Refer to “5.2 Dealing with Plate Jams”(p. 54) and remove
the plate and cassette from this device.
Cassette might not contain the plate. Please open
If the plate is not inside the device or cassette, check to
7110 the door and remove the cassette. Confirm the
see if the wrong cassette was inserted.
plate is in the cassette.
If the plate is inside the device, use the image erase mode
to erase the plate’s image.

48
5.1 Troubleshooting

Error Information
Error Appropriate action
Message
Code
Error was detected while processing the cassette. Refer to “5.2 Dealing with Plate Jams”(p. 54) and remove
7111 Please open the door and remove the Cassette the plate and cassette from this device.
- 7117 and Plate. After removal, erase the cassette in Use image erase mode to erase the image from the re-
ERASE MODE. moved plate.
After opening the top door, refer to the “4 Open the rear
door.”(p. 56) process from “5.2.2 If the Plate is Not Visible
Critical error has occurred while processing the
From the Top Door’s Opening” step 4 to remove the plate
cassette. Please contact service. Please open the
7118 and cassette.
top door and remove the Cassette. And open the
Once this error occurs, it cannot be solved by turning the
back door and remove the Plate.
device power OFF and ON again. Then contact Konica Mi-
nolta technical representatives.
Error was detected while processing the cassette. Refer to “5.2 Dealing with Plate Jams”(p. 54) and remove
7119 Please open the door and remove the Cassette the plate and cassette from this device.
- 7124 and Plate. After removal, erase the cassette in Use image erase mode to erase the image from the re-
ERASE MODE. moved plate.
After opening the top door, refer to the “4 Open the rear
door.”(p. 56) process from “5.2.2 If the Plate is Not Visible
Critical error has occurred while processing the
From the Top Door’s Opening” step 4 to remove the plate
cassette. Please contact service. Please open the
7125 and cassette.
top door and remove the Cassette. And open the
When a device error is not resolved even when you have

Chapter 5
back door and remove the Plate.
taken measures to deal with a plate jam, contact Konica
Minolta technical representatives.
Error was detected while processing the cassette. Refer to “5.2 Dealing with Plate Jams”(p. 54) and remove
7126 Please open the door and remove the Cassette the plate and cassette from this device.
- 7129 and Plate. After removal, erase the cassette in Use image erase mode to erase the image from the re-
ERASE MODE. moved plate.
Refer to “5.3 Forcibly Removing the Cassette”(p. 61) and
Cassette Size error was detected during cassette
remove the cassette from this device.
7130 insertion. Please confirm whether the cassette
Confirm the cassette size and the cassette type is compat-
size is supported.
ible with this device.
Error was detected while processing the cassette. Refer to “5.2 Dealing with Plate Jams”(p. 54) and remove
7131 Please open the door and remove the Cassette the plate and cassette from this device.
- 7135 and Plate. After removal, erase the cassette in Use image erase mode to erase the image from the re-
ERASE MODE. moved plate.
Refer to “5.2 Dealing with Plate Jams”(p. 54) and remove
Cassette size and the Plate size might not match.
the plate and cassette from this device.
Please open the door and remove the Cassette.
Confirm that the cassette and plate sizes match up cor-
7136 Confirm the size of the cassette and plate. Erase
rectly.
the cassette in ERASE MODE before using the
Load the removed plate into a cassette of the correct size,
cassette again.
then use image erase mode to erase the plate’s image.
7138 Error was detected while processing the cassette. Refer to “5.2 Dealing with Plate Jams”(p. 54) and remove
7139 Please open the door and remove the Cassette the plate and cassette from this device.
7143 and Plate. After removal, erase the cassette in Use image erase mode to erase the image from the re-
- 7153 ERASE MODE. moved plate.
Error was detected while processing the cassette. Refer to “5.2 Dealing with Plate Jams”(p. 54) and remove
Please open the door and remove the Cassette the cassette and plate from this device. Use image erase
7154
and Plate. After removal, erase the cassette in mode to erase the image from the removed plate.
ERASE MODE. Moreover, please clean a roller. Further, clean the roller.

49
5.1 Troubleshooting

Error Information
Error Appropriate action
Message
Code
7400 Initialization failed. Reader error was detected Open and close the top door. The device will be re-initial-
- 7409 during initialization. ized, so check to see if the error occurs again.
An abnormality was detected in the image-erasing device.
Now that image reading is complete, wait for the image to
7410 be transmitted. After confirming that transmission is com-
Erasure error was detected.
7411 plete, refer to “5.3 Forcibly Removing the Cassette”(p. 61)
and remove the cassette.
Then contact Konica Minolta technical representatives.
Refer to “5.2 Dealing with Plate Jams”(p. 54) and remove
7412
Reader error was detected. the plate and cassette from this device.
- 7414
Then contact Konica Minolta technical representatives.
Initialization failed. Reader error was detected Open and close the top door. The device will be re-initial-
7415
during initialization. ized, so check to see if the error occurs again.
Refer to “5.2 Dealing with Plate Jams”(p. 54) and remove
7416
Reader error was detected. the plate and cassette from this device.
- 7419
Then contact Konica Minolta technical representatives.
Initialization failed. Reader error was detected Open and close the top door. The device will be re-initial-
7420
during initialization. ized, so check to see if the error occurs again.
Refer to “5.2 Dealing with Plate Jams”(p. 54) and remove
7421
Reader error was detected. the plate and cassette from this device.
7422
Then contact Konica Minolta technical representatives.
7650
Reader error was detected. Contact Konica Minolta technical representatives.
7651
7700 Initialization failed. Reader error was detected Open and close the top door. The device will be re-initial-
7701 during initialization. ized, so check to see if the error occurs again.
Once this error occurs, it cannot be solved by turning the
7710
Detected Memory error. device power OFF and ON again. Contact Konica Minolta
- 7746
technical representatives.
An abnormality was detected in the transmission data.
7750
Detected Data error. Confirm transmission data and communication cable. Then
- 7759
turn the device power OFF and ON.
An abnormality was detected in the cassette size param-
eters.
7800 Parameters which cannot be processed in the
Refer to “5.2 Dealing with Plate Jams”(p. 54) and remove
7801 Reader Settings are specified
the plate and cassette from this device.
Then contact Konica Minolta technical representatives.
Refer to “5.2 Dealing with Plate Jams”(p. 54) and remove
7802 Reader error was detected. the plate and cassette from this device.
Then contact Konica Minolta technical representatives.
Examination requirement(s) have been made which can-
Parameters which cannot be processed in the
7803 not be fulfilled by this device. Review current examination
Reader Settings are specified
requiment setting.
7810 Top door open Close the top door.
7811 Back door open Close the rear door.
7812 Left side door open Close the left door.
Program data of service call reader corrupted.
7999 Contact Konica Minolta technical representatives.
Need to recover the program.

50
5.1 Troubleshooting

Error Information
Error Appropriate action
Message
Code
Due to a reader power failure during image capture, image
Reader is not in the status of sending image. Im-
27000 data has been lost.
age transmission failed.
Re-expose the image and perform the read process again.

Messages displayed on the task tray.

Warning

Error
Message Appropriate action
Code
Message displayed when a reader transmitting image data
is connected to another Image processing controller.

If reader connection is changed


Reconnect to the Image processing controller used during
the initial image reading process. When returning to the
proper connection, retransmission will begin. (However,
This reader is in process. Connect to the former this will only occur when Image processing controller and
PC. reader operations are uninterrupted. If power is turned
27100
To terminate process, turn of the reader. The im- OFF, or operations are halted, image data will not be re-

Chapter 5
age will be deleted. transmitted.)

If the Image processing controller is restarted while the im-


age read process is taking place
Image data will not be retransmitted. Refer to “5.2 Dealing
with Plate Jams”(p. 54), remove the plate and cassette, then
turn the reader OFF and back ON. Use image erase mode
to erase the image from the removed plate.
Reconnect to the originally operating reader. When return-
ing to the proper connection, retransmission will begin.
(However, this will only occur when Image processing con-
troller and reader operations are uninterrupted. If power is
turned OFF, or operations are halted, image data will not
be retransmitted.)
The reader was switched. To complete image
27101 Restart the Image processing controller PC when connec-
reading, connect to the former reader.
tion with former reader cannot be established. Image data
will not be retransmitted.
Then refer to “5.2 Dealing with Plate Jams”(p. 54) remove
the plate and cassette, and turn the reader OFF and back
ON. Use image erase mode to erase the image from the
removed plate.
Message displayed when multiple readers are connected
to a single system. Readers not detected by the Image
Another reader is working in the system.
processing controller cannot be used.
This reader is unusable now.
27105 To use a connected reader, disconnect USB cables from
Stop another reader if you want to
all readers, then connect only to the reader you want to
use this reader.
use. (Before disconnecting, make sure the readers are not
in the middle of a process.)

51
5.1 Troubleshooting

Error
Message Appropriate action
Code
USB was disconnected during image transmis-
sion. Confirm USB cable connection. Once connection is estab-
27110
Connect USB. lished, image data transmission will resume.
Image transmission will start after connection.

Error

Error
Message Appropriate action
Code
If the Image processing controller server has not started
up yet, wait for it to do so.
If the Image processing controller server has started up,
make sure network devices (HUB, cables, etc.) are work-
Connection failed.
ing properly between the Image processing controller
24000 Check all LAN connections.
client connected to the reader, and the Image processing
Reboot the system if it is not recovered.
controller server. (Also remove and reconnect the cable
connector.)
If the error is not fixed, restart the Image processing con-
troller PC.
Make sure network devices (HUB, cables, etc.) are work-
ing properly between the Image processing controller
Connection with server PC failed. client connected to the reader, and the Image processing
25000 Check all LAN connections with server. controller server. (Also remove and reconnect the cable
Reboot the system if it is not recovered. connector.)
If the error is not fixed, restart the Image processing con-
troller PC.
Make sure network devices (HUB, cables, etc.) are work-
ing properly between the Image processing controller
client connected to the reader, and the Image processing
controller server. (Also remove and reconnect the cable
connector.)
Data transmission failed by LAN disconnection.
By returning to the proper connection status, image data
25200 Check all LAN connections with server.
will be retransmitted. (However, this will only occur when
Re-send data after connection recovery
Image processing controller and reader operations are
uninterrupted. If power is turned OFF, or operations are
halted, image data will not be retransmitted.)
If the error is not fixed, restart the Image processing con-
troller PC. (Image data will be lost in this case.)
Make sure network devices (HUB, cables, etc.) are work-
ing properly between the Image processing controller
Data transmission failed by LAN disconnection. client connected to the reader, and the Image processing
25300 Check all LAN connections with server. controller server. (Also remove and reconnect the cable
Reboot the system if it is not recovered. connector.)
If the error is not fixed, restart the Image processing con-
troller PC.

52
5.1 Troubleshooting

5.1.3 When the Error Lamp is unlit, and no message is displayed

If the unit is not operating normally, but the Error Lamp is not lit, and no message is displayed, take action according
to the following chart.

Device Status Appropriate action


If a cassette has been inserted to perform image reading, but The Image processing controller is not in image capture
the Busy Lamp continues to blink, the reading process will not standby mode.
start. Click the Image processing controller’s [CR] button.

Chapter 5

53
5.2 Dealing with Plate Jams

If a plate jam occurs, check the plate status inside the device as outlined in "5.2.1 Check the Plate Status (Do This
First)"(p. 54), then remove it according to any of the following procedures:
• “5.2.2 If the Plate is Not Visible From the Top Door’s Opening”(p. 55)
• “5.2.3 If the Plate’s Front End is Visible From the Top Door’s Opening”(p. 57)
• “5.2.4 If the Plate’s Back End is Visible From the Top Door’s Opening”(p. 58)
• “5.2.5 If the Plate’s Mid-section is the Only Portion Visible From the Top Door’s Opening”(p. 59).
After removing the plate, check its condition. If its shape has been altered, replace it with a new one
If there are no problems, load it in the cassette and perform the image erase process.

CAUTION
• Make sure to perform this process with clean hands, so as not to get the plate, guide, roller, etc. dirty inside of the device.
• Do not touch the yellow filter inside of this device.
• Do not let foreign objects, dust or debris get inside of the device.
• Follow the process described in this manual to remove the plate. Attempting to remove the plate by means other than as de-
scribed here could scratch the plate or guide.

5.2.1 Check the Plate Status (Do This First)

1 Open the top door, and check the plate status through the door’s opening.

Top Door

Select one of the 4 patterns below which applies to the plate’s current status, then skip to the appropriate refer-
ence section.

The plate is not visible from the opening. The back end of the plate is visible from the
opening.
REFERENCE REFERENCE
• Refer to “5.2.2 If the Plate • Refer to “5.2.4 If the Plate’s
is Not Visible From the Top Back End is Visible From the
Door’s Opening”(p. 55) Top Door’s Opening”(p. 58)
The plate is in any position Plate’s back end

The front end of the plate is visible from the Only the mid-section of the plate is visible from
opening. the opening. (The front and back ends cannot be
seen.)
REFERENCE REFERENCE
• Refer to “5.2.3 If the Plate’s • Refer to “5.2.5 If the Plate’s
Front End is Visible From Mid-section is the Only Por-
the Top Door’s Opening”(p. 57) tion Visible From the Top
Plate’s front end Plate’s front end Plate’s back end Door’s Opening”(p. 59)

54
5.2 Dealing with plate jams

3 If the plate comes out from the bottom,


5.2.2 If the Plate is Not Visible perform the following.
From the Top Door’s 1. Rotate the dial further to the right, and
Opening pull out the plate from the top door’s
opening.

1 Open the left door. IMPORTANT


• Make sure the plate does not hit or rub against the top

2 Rotate the dial 3 times to the right, and check door or any other component.
• Continue to turn the dial until the plate is completely
whether or not the plate’s back end comes
out of the device.
out from the bottom.

Dial

Chapter 5
If the plate comes out from the bottom, perform step
3. If the plate does not come out from the bottom, per- Dial
form step 4.

2. Hold the lever up, and remove the cas-


sette.

Lever

3. Check the inside of the device, then


close all of its doors.
• Before closing the top door,
make sure that no dust is on the yellow portion
inside of the device. If dust is present, wipe it off
with a dry, soft cloth.

55
5.2 Dealing with plate jams

4. Load the plate into the cassette, and 3. Rotate the dial 3 times to the left.
perform the image erase process.

REFERENCE
• Loading the plate
Refer to “4.1.3 Loading the Plate into the Cassette”(p.
40)
• Erasing images
Refer to “3.2 Image Erase Operation”(p. 30)

4 If the plate does not come out from the


bottom, perform the following.
Dial
1. Pull up the lever under the top door,
hold the lever up, and remove the cas- 4. Open the rear door.
sette.

IMPORTANT
• There is a chance that the plate is still inside of the
cassette, be careful to pull out the cassette very slowly
so as not to scratch the plate.

Lever

Rear Door

5. Pull the lever up.


• The guide will fold down to the bottom.

2. Check whether a plate is in the cas-


sette. If a plate is in the cassette, reload
the plate into the cassette and erase
the image.
• If no plate is in the cassette, go to the next pro-
cedure. Lever Guide
• If there is a plate with the image to be erased,
close all doors before the image erase process
is performed.

REFERENCE
• Removing the plate
Refer to “4.1.2 Removing the Plate From the
Cassette”(p. 38)
• Loading the plate
Refer to “4.1.3 Loading the Plate into the Cassette”(p.
40)
• Erasing images
Refer to “3.2 Image Erase Operation”(p. 30)

56
5.2 Dealing with plate jams

6. Remove the plate at the bottom from


the rear door’s opening. 5.2.3 If the Plate’s Front End
is Visible From the Top
IMPORTANT
• Make sure the plate does not hit or rub against the
Door’s Opening
rear door or any other component.

IMPORTANT
• When the plate’s tip extends to the guide, refer to
“5.2.5 If the Plate’s Mid-section is the Only Portion
Visible From the Top Door’s Opening”(p. 59) to eject the
plate.

7. Push the guide back into place until a


click is heard, indicating that it has at-
tached to the hooks on each side.

IMPORTANT
• Closing the rear door while the guide is still down
could cause the rear door to break.
Guide Plate’s tip

Chapter 5
1 Hold the lever up, and remove the cas-
sette.

IMPORTANT
• Pull out the cassette slowly so as not to scratch the
plate.

Lever
Guide

8. Check the inside of the device, then


close all of its doors.
• Before closing the top door,
make sure that no dust is on the yellow portion
inside of the device. If dust is present, wipe it off
with a dry, soft cloth.

9. Load the plate into the cassette, and


perform the image erase process.

REFERENCE
• Loading the plate
Refer to “4.1.3 Loading the Plate into the Cassette”(p. 2 Check the inside of the device, then close
40) the top door.
• Erasing images Make sure that no dust is on the yellow portion
Refer to “3.2 Image Erase Operation”(p. 30) inside of the device. If dust is present, wipe it off
with a dry, soft cloth.

57
5.2 Dealing with plate jams

3 Load the plate into the cassette, and perform


the image erase process. 5.2.4 If the Plate’s Back End
is Visible From the Top
REFERENCE
Door’s Opening
• Loading the plate
Refer to “4.1.3 Loading the Plate into the Cassette”(p.
40)
• Erasing images 1 Open the left door.
Refer to “3.2 Image Erase Operation”(p. 30)

2 While rotating the dial to the right, pull out


the plate from the top door’s opening.

IMPORTANT
• Make sure the plate does not hit or rub against the top
door or any other component.
• Continue to turn the dial until the plate is completely
out of the device.

Dial

3 Hold the lever up, and remove the cas-


sette.

Lever

58
5.2 Dealing with plate jams

4 Check the inside of the device, then close


all of its doors. 5.2.5 If the Plate’s Mid-section is
• Before closing the top door, the Only Portion Visible From
make sure that no dust is on the yellow portion
inside of the device. If dust is present, wipe it off
the Top Door’s Opening
with a dry, soft cloth.

5 Load the plate into the cassette, and per- 1 Open the left door.
form the image erase process.
2 Rotate the dial to the left to feed the plate
REFERENCE out of the device until its back end is vis-
• Loading the plate ible.
Refer to “4.1.3 Loading the Plate into the Cassette”(p.
40)
• Erasing images
Refer to “3.2 Image Erase Operation”(p. 30)

Dial

Chapter 5
• If the plate will not eject
If the plate does not come to the roller, you can-
not eject it using the dial. Push the plate to the
roller by hand to eject.

Plate

59
5.2 Dealing with plate jams

3 When the back end comes out of the 6 Load the plate into the cassette, and per-
opening, pull out the plate from the open- form the image erase process.
ing while rotating the dial to the right.
REFERENCE
IMPORTANT • Loading the plate
• Make sure the plate does not hit or rub against the top Refer to “4.1.3 Loading the Plate into the Cassette”(p.
door. 40)
• Continue to turn the dial until the plate is completely • Erasing images
out of the device. Refer to “3.2 Image Erase Operation”(p. 30)

Dial

4 Hold the lever up, and remove the cas-


sette.

Lever

5 Check the inside of the device, then close


all of its doors.
• Before closing the top door,
make sure that no dust is on the yellow portion
inside of the device. If dust is present, wipe it off
with a dry, soft cloth.

60
5.3 Forcibly Removing the Cassette

The following steps describe a method of forcibly re-


moving the cassette, if the cassette is accidentally in-
serted before turning the power ON, etc.

1 Open the top door.

CAUTION
• Do not let foreign objects, dust or debris get inside of
the device.

Top Door

Chapter 5
2 Hold the lever up, and remove the cas-
sette.

Lever

3 Check the inside of the device, then close


the top door.
Make sure that no dust is on the yellow portion
inside of the device. If dust is present, wipe it off
with a dry, soft cloth.

61
62
Chapter
6
Maintenance
An explanation of maintenance items
that must be followed for safe device operation.

63
6.1 A list of Maintenance Items

Maintenance items and periods are as follows.

Maintenance items Maintenance period Reference


Plate Plate cleaning Once per week “6.3.1 Plate Cleaning”(p. 69)
Cassette Cassette cleaning Once per week “6.3.2 Cassette Cleaning”(p. 69)
Reader Cassette insertion slot cleaning Once per week “6.2.1 Cassette Insertion Slot Cleaning”(p. 65)
External cleaning of device Once per month “6.2.2 External Cleaning of Device”(p. 65)
Intake vent, Exhaust outlet cleaning Once per month “6.2.3 Intake Vent, Exhaust Outlet Cleaning”(p. 65)
Light collection unit cleaning (keeping When a white line “6.2.4 Light Collection Unit Cleaning (Keeping
them dust free) error occurs Them Dust Free)”(p. 65)
Filter cleaning When an error occurs “6.2.5 Filter Cleaning”(p. 66)
Feeding roller cleaning Once per month “6.2.6 Feeding Roller Cleaning”(p. 66)
If any dust or dirt is apparent, perform cleaning regardless of the maintenance period.

HINT
• To use this device for a long time, periodical maintenance and cleaning of each unit are recommended. Contact Konica Mi-
nolta technical representatives. for details.

64
6.2 Reader Maintenance

6.2.1 Cassette Insertion Slot 6.2.4 Light Collection Unit


Cleaning Cleaning (Keeping Them
Dust Free)
CAUTION If a white line error occurs for a read image, light col-
• Cleaning must be performed after the device power lection unit might contain dust or debris. Follow the
is turned OFF, and the power supply has been un- steps below to clean light collection unit.
plugged.
When doing this, clean the power supply’s plug of any Cleaning methods
dust or debris. 1 Make sure the Busy Lamp is not lit (indi-
cating that the reader is not operating).
Cleaning methods
1 Use a dry, soft cloth to lightly wipe the 2 Open the left door.
cassette insertion slot free of dust.
• If the dirt will not come off, wipe it with a slightly 3 Move the light collection unit cleaning
damp cloth, then wipe it dry with a dry cloth. handle 3 times.

Light Collection Unit


6.2.2 External Cleaning of Cleaning Handle
Left
Device Door

CAUTION
• Cleaning must be performed after the device power

Chapter 6
is turned OFF, and the power supply has been un-
plugged.
When doing this, clean the power supply’s plug of any
dust or debris.
This will activate a cleaning brush inside of the de-
vice, which will free the light collection unit of dust.
Cleaning methods
1 Use a dry cloth to lightly wipe the device’s 4 Check that the light collection unit clean-
external area free of dust. ing handle has returned to its original po-
• If the dirt will not come off, wipe it with a slightly sition.
damp cloth, then wipe it dry with a dry cloth.

5 Close the left door.

6.2.3 Intake Vent, Exhaust 6 Prepare a test image, perform image read,
Outlet Cleaning and check whether or not white lines have
been generated.
Cleaning methods
1 Use a vacuum to remove dust from the in-
take vent and exhaust outlet.

65
6.2 Reader maintenance

6.2.5 Filter Cleaning 6.2.6 Feeding Roller Cleaning

Use the included cleaning tool to clean the feeding


CAUTION roller.
• Cleaning must be performed after the device power Be sure to clean the feeding roller when messages
is turned OFF, and the power supply has been such as “Roller cleaning required” are displayed on the
unplugged. image processing controller.
When doing this, clean the power supply’s plug of any
dust or debris. CAUTION
• Be sure to read the procedure before using the clean-
Cleaning methods ing tool.

1
If you use the cleaning tool without reading how to
Open the top door.
clean the feeding roller, the cleaning tool may not
perform satisfactorily.
CAUTION Please be sure to read and fully understand how to
• Do not let foreign objects, dust or debris get inside of clean the feeding roller before using the cleaning tool.
the device.

Using the cleaning tool for the first time

1 Remove the protective sheet that is af-


fixed to the adhesive sheet on the back of
the cleaning tool.
Top Door After that, begin cleaning starting from Step 4 of
the “Cleaning methods”.

2 Make sure there is no dust on the yellow


portion inside of the device (the filter). If
dust is present, wipe it off with a dry, soft
cloth.

Filter Protective sheet


Adhesive sheet

3 Close the top door.

66
6.2 Reader maintenance

Cleaning methods 4 Turn OFF the power supply circuit breaker


of this device.
1 Remove any dust from the entire cleaning
tool. 5 Use a dry, soft cloth to lightly wipe the
cassette insertion slot free of dust.
2 Wash the adhesive sheet with water.
To wash the adhesive sheet with water, rub it
clean with your fingers while running water over it.
Wipe off dust

Power Supply
Circuit Breaker

Wipe off dust 6 Insert the cleaning tool into the cassette
Adhesive sheet insertion slot of this device.
• Insert it to the back until you can no longer see

3 Dry the adhesive sheet for a while.


the arrow-type guide of the cleaning tool.
Dry the adhesive sheet naturally until no drops of
water are left on it.

Chapter 6
Insert it to here

CAUTION
• Do not wipe the adhesive sheet with a cloth, paper, or
the like.
There is a risk that fibers or dust may adhere to the
adhesive sheet.
To dry the adhesive sheet, expose it to the air and dry
it naturally.

67
6.2 Reader maintenance

7 Pull the cleaning tool out until it is in a po-


sition where you can see the entire arrow-
type guide.

8 Repeat Steps 6 and 7 about ten times.

Pull it out to here

CAUTION
• When inserting the cleaning tool into the cassette in-
sertion slot, take care not to get your fingers or other
things caught in it.

CAUTION
• Do not leave the back of the cleaning tool exposed to
direct sunlight.
There is a risk that the adhesive sheet may soften.
Store it in a place where it is not exposed to direct
sunlight.
• Do not store the cleaning tool in a place where there is
a lot of dust.
There is a risk that dust may adhere to the adhesive
sheet.
Store the cleaning tool in a plastic bag or the like.

Sunlight

Dust

68
6.3 Plate/Cassette Maintenance

4 After cleaning, load the plate into the cas-


6.3.1 Plate Cleaning sette and perform the image erase pro-
cess.
Cleaning methods
CAUTION
CAUTION • If absolute ethanol was used to clean the plate, make
sure the plate is completely dry before loading it into
• When using absolute ethanol, make sure to follow the
the cassette.
handling precautions provided by the pharmaceutical
• After cleaning the plate, make sure to perform the im-
company.
age erase process.
• Make sure your hands are clean before handling the
plate to prevent it from getting dirty. And refrain from
touching the plate unnecessarily.
REFERENCE

1 Secure an area for the plate to be set on • Loading the plate


Refer to “4.1.3 Loading the Plate into the Cassette”(p.
for cleaning.
40)
Make sure the area meets the following require-
• Erasing images
ments.
Refer to "3.2 Image Erase Operation"(p. 30)
• It is flat.
• It is wider than the plate.
• There is no dust or debris present.
• It meets the area requirements outlined in
“1.2.6 Cassette/Plate Storage and Usage
6.3.2 Cassette Cleaning
Cautions”(p. 19).
When cleaning the back surface of the plate, lay
out a clean, dry cloth to set its front surface on (face CAUTION
down) for protection. • When using absolute ethanol or isopropyl alcohol for

Chapter 6
cleaning, make sure to follow the handling precautions
2 Remove the plate from the cassette. provided by each pharmaceutical company.

REFERENCE Cleaning methods


• Refer to “4.1.2 Removing the Plate From the 1 Dirt on the cassette’s surface should be
Cassette”(p. 38) lightly wiped off with a cloth slightly damp
with absolute ethanol or isopropyl alco-
3 Lightly wipe the plate’s front and back hol.
surfaces with a soft, nonwoven dry cloth,
then use a blower to clean away fiber par- CAUTION
ticles, etc. • Using too much absolute ethanol or isopropyl alcohol
• If the dirt cannot be removed by a dry cloth, could alter the cassette’s color.
use a soft, nonwoven cloth slightly damp with
absolute ethanol to wipe it clean. Do not use too
much absolute ethanol. After cleaning, fully dry
the plate.

CAUTION
• Note that wiping with too much force could leave
scratches or alter its shape.
• Do not damp the cloth with water to clean off dirt.

69
70
Appendix
A list of device specifications, etc.

71
Specification

REGIUS SIGMA
Item Content
Model name DIRECT DIGITIZER REGIUS SIGMA
Remarks CE marking, it means conformity with the Medical Devices Directive 93/42/EEC.
REGIUS CASSETTE RC-300
Cassette size
14”×17”, 14”×14”, 11”×14”, 10”×12”, 8”×10”, 24×30cm, 18×24cm, 15×30cm
Readable plate REGIUS PLATE FP-1S
Sampling pitch Two types (87.5 μm/175 μm)
Maximum resolution 4,020×4,892 (14”×17”, 87.5 μm reading mode)
Digital gradation level 4,096 steps (12 bits)
Processing capability (note1) Up to 45 plates/hour (14”×17”/175 μm)
Cassette feed/Load time (note 2) Approximately 80 seconds (14”×17”/175 μm)
Slots Insertion/ejection slot ×1
Dimensions 510 (W) × 610 (D) × 355 (H) mm
Weight Approximately 28 kg
Supply condition AC100 - 240V ±10%(50/60Hz)/100VA or less
Maximum electric power
100VA or less
consumption

Temperature Humidity
80%RH

15 to 80% RH
Operating conditions 10 to 30 °C
(No condensation)
Environment 15%RH

90%RH

90% RH or less
Storage -20 to 60 °C
(No condensation) 0%RH

Noise level 53 dB or less (Average value during operation)


Life time (note 3) 6 years
Operation mode Continuous operation
Power cable, USB cable (USB2.0 High-speed mode compatible), Opening tools for cassette,
Accessories
Operation manual, Cleaning Tool
Products sold separately REGIUS PLATE FP-1S, REGIUS CASSETTE RC-300, ImagePilot, CS-7
note 1:
• Processing Capability is when the arrived X-ray dose at the plate is 7 mR or less. Feed/Load Time increase with X-ray dose if
the greatest doses on plate are over 7 mR.
• Processing capability differs depending on X-ray dose, connection environment and operation method.
note 2: Cassette Feed/Load Time: Time required between cassette insertion and cassette exit.
note 3: Life time is valid only if precautions are followed and the product is operated correctly.

72
Index

B M
Busy Lamp..............................................................25 maintenance.....................................................20, 63
memo clip................................................................36
C movement...............................................................17

cassette...........................8, 19, 30, 32, 36, 44, 61, 69


O
cassette insertion slot.................................30, 32, 65
Cassette Insertion Slot............................................23 Opening tools for Cassette...............................38, 40
cassette label....................................................36, 37
cleaning.............................................................20, 64 P
console......................................................................8
plate................................8, 19, 36, 38, 40, 44, 54, 69
corner block......................................................36, 37
plate’s “first day” label.......................................36, 37
power supply circuit breaker.................13, 23, 28, 34
D
power supply port....................................................24
dial........................................................23, 55, 58, 59
disposal...................................................................20 R
reader........................................................................8
E
Ready Lamp............................................................25
Erase Button.....................................................23, 30
Erase Lamp.............................................................25 S
Erase unit................................................................26
Safety Alert Symbol.................................................12
Error..................................................................48, 52
specification............................................................72
Error Lamp..............................................................25
Status Lamp......................................................23, 25
exhaust outlet..........................................................65
system configuration examples...............................22
Exhaust Outlet..................................................23, 24

U
F

Appendix
unjamming dial........................................................23
filter...................................................................18, 66
forcibly removing the cassette................................61
W
I Warning.............................................................47, 51

image erase mode..................................................30


image erasing...................................................26, 30
image reading...................................................26, 32
installation...............................................................17
intake vent.........................................................23, 65

J
jamming..................................................................54
JM.............................................................................8

L
Light Collection Unit..........................................26, 65
Light Collection Unit Cleaning Handle..............23, 65

73
74
Konica Minolta Healthcare Americas, Inc. Konica Minolta Business Solutions (Canada)
411 Newark-Pompton Turnpike, Wayne, NJ 07470, Ltd.
U.S.A. 5875 Explorer Drive, Mississauga,
TEL. 973-633-1500 Ontario, L4W 0E1, Canada
TEL. 905-890-6600
EU Authorized Representative:
Konica Minolta Business Solutions Konica Minolta Medical & Graphic (SHANGHAI)
Europe GmbH Co., Ltd.
Hoogoorddreef 9, 1101 BA Amsterdam, Room1602, Tower B, Hongwell International Plaza,
The Netherlands 1602 West Zhongshan Road, Xuhui District,
TEL. +31-20-658-4100 Shanghai 200235 China
TEL. 021-6422-2626

Konica Minolta Healthcare India Private Ltd.


Office No.201, 2nd Floor, Atrium 2,
Next to Courtyard Marriott Hotel, Andheri Kurla Road,
Chakala, Andheri (East), Mumbai - 400093, India
TEL. +91-22-61916900

A1P6BA01EN08
2018-10-31