Professional Documents
Culture Documents
BK91 1310 CPF VED 003 QAC H03 0001 - A NDE Procedure For Centrifugal Pumps
BK91 1310 CPF VED 003 QAC H03 0001 - A NDE Procedure For Centrifugal Pumps
: BK91-1310-CPF-VED-003-
NDE PROCEDURE QAC-H03-0001
Revision A
Date of 22/12/2022
Issue
Page 1 of 69
NDE PROCEDURE
FOR Centrifugal Pumps (API610)
ISSUED FOR
A
REVIEW 22/12/2022 Xu Fengjin Wang Songhua Chen Wei Kong Fanjun
2023/1/13
2 Reference Documents
2.1 All standards, codes and norms refer to the last published edition and addenda, which shall be adopted if
customer is not requiring a former revision.
ASME BPVC Sec. V Article 6 - Liquid Penetrant Examination; Edition 2021
ASNT SNT-TC-1A Personnel Qualification and Certification in Non-destructive Testing, edition
accepted by the ASME BPVC; Edition 2016 with 2018 Addendum
ASME BPVC Sec. VIII Div. 1 – Pressure Vessels; Edition 2021
ASME B31.3; 2020 Process Piping
AWS D1.1; 2020 Structural Steel Welding Standards
ISO 3452-1; 2020 Non-destructive Testing - Penetrant Testing - Part 1: General Principles
ISO 5817; 2014 Welding – Quality levels for imperfections
ISO 17635; 2017 NDT of Welds – General rules for metallic materials
ISO 23277; 2015 NDT of Welds – PT – Acceptance levels
BS EN 1371-1; 2011 Founding – Liquid penetrant testing – Part 1: Sand, gravity die and low pressure
die castings
BS EN 1371-2; 2015 Founding – Liquid penetrant testing – Part 2: Investment castings
BS EN 10228-2; 2016 Steel Forgings – Part 2: Penetrant testing
ISO 9712; 2021 Non-destructive Testing - Qualification and Certification of NDT Personnel
ISO 3059; 2013 Non-destructive Testing - Penetrant Testing and Magnetic Particle Testing -
Viewing Conditions
Sulzer DT-0002 Liquid Penetrant Test Report
2.2 Sulzer documents referenced may be downloaded from https://www.sulzer.com/q-documents
4 Consumables
4.1 The level of contaminants of all penetrant materials used during testing shall be analyzed and certified.
When examining austenitic or duplex stainless steel and titanium, the total content of halogens (chlorine,
fluorine, etc.) shall not exceed 0.1% by weight and when examining nickel based alloys, the content of
Sulphur shall not exceed 0.1% by weight (complying with ASME BPVC Section V, Article 6, Mandatory
Appendix II).
4.2 All consumables (Penetrant, Excess Penetrant Remover, and Developer) used for any one test shall be
from the same manufacturer and the same system.
4.3 Hydrophilic or lipophilic emulsifiers are not applicable when visible penetrant materials are used.
5 Test procedure
5.1 Timing of examination
Testing shall be carried out in the manner defined in the order Quality Plan (Inspection & Test Plan), both
in chronological order and material condition.
5.2 Extent of examination
The extent or scope of the examination shall be as defined in the order Quality Plan (Inspection & Test
Plan). Where welds, or specific areas are being examined, then the examination shall include a minimum
of 25 mm (1 inch) outside the area being tested and in the case of welds must include the entire "Heat
Affected Zone (HAZ)".
5.3 Surface preparation
Satisfactory results are generally achieved on surfaces in the as-welded, as-rolled, as-cast, as-machined
and as-forged condition.
The surface condition shall be adequate to allow accurate interpretation of indications commensurate with
the acceptance criteria specified.
Note: Shot blasting is not recommended prior to liquid penetrant testing. However, if required, it shall be
as light as possible, in order to avoid closing up or sealing possible discontinuities.
5.4 Pre cleaning
Prior to the application of the penetrant, dirt, grease, lint, scale, welding flux, weld splatter, paint, oil, and
other extraneous matter that could obscure surface openings or otherwise interfere with the examination
shall be removed from the surface. Typical cleaning agents which may be used are detergents, organic
solvents, descaling solutions and paint removers. Degreasing and ultrasonic cleaning methods may also
be used. The solvent used during the cleaning shall meet the requirements for contamination control given
in section 4 of this procedure.
5.4.1 Mechanical cleaning
Surface preparation by grinding or machining or other methods shall be necessary when the surfaces are
covered in some parts with slag, scale, weld spatter, etc.
For stainless steel grades of materials only austenitic wire brushes shall be used.
Cutting tools (grinding stones rotary burrs etc.) shall be free from ferrous contaminants.
5.4.2 Chemical cleaning
The surface to be examined and all adjacent areas within at least 25 mm (1 inch) shall be thoroughly
degreased using a suitable solvent.
Drying of the pre-cleaned surfaces shall be accomplished by normal evaporation, forced warm air or by
heated radiators. A minimum drying period of 5 minutes shall elapse before application of the penetrant.
5.5 Penetrant application
The penetrant shall be applied by dipping, spraying (atomizer, spray gun) or by brushing. If spray guns are
used, care must be taken that the compressed air does not contaminate the penetrant, i.e. filters shall be
provided at the air inlet to preclude the entry of dirt, water and oil.
Revision: 10 Date: 03-Mar-2022 Description: Enclosure 1; additional requirements
This document has been prepared & published under the authority of the Technical Quality Working Group Doc No: TP-0001
The temperature of the test surface shall be between 10°C (50°F) and 52°C (125°F). Local heating or
cooling is permitted, if necessary.
The penetration time shall be as specified by the penetrant manufacturer, but it shall be not less than 15
minutes and maximum 60 minutes.
During the penetration time the surface to be examined shall not be allowed to dry. If the penetrant has
dried during the penetration time, the surface shall be cleaned and the penetrant re-applied for the full
penetration time.
5.6 Excess penetrant removal
After the penetration time specified has elapsed, any excess penetrant shall be removed, taking care to
minimize removal of penetrant from surface discontinuities.
Remaining traces of penetrant shall be removed by lightly wiping the surface with clean lint-free cloth or
absorbent paper taking care to minimize removal of penetrant from any discontinuities.
5.6.1 Water washable penetrant
Excess penetrant on the surface being examined shall be removed with water. Penetrant shall be
removed manually by wiping the part surface with a dampened rag until most traces of the penetrant are
removed. The final traces of penetrant shall be removed by water spray where the pressure measured at
the outlet shall not exceed 3.5 bar (50 psi), and the water temperature shall not exceed 43°C (110°F).
5.6.2 Solvent removable penetrant
The surface to be examined shall be cleaned using a dry cloth or absorbent paper until most of the
penetrant has been removed. The remaining traces shall be removed by lightly wiping the surface with
cloth or absorbent paper moistened with solvent. To minimize removal of penetrant from discontinuities,
care shall be taken to avoid use of excess solvent.
Note: Flushing the surface with solvent following the application of the penetrant and prior to developing
is prohibited.
5.7 Surface drying prior to application of the developer
5.7.1 Except when using an aqueous wet developer, the area to be examined shall be dried prior to applying
the developer.
5.7.2 When using water for penetrant removal, drying may be accomplished using:
• a clean lint-free cloth or absorbent paper
• hot air circulation, provided the surface temperature is not raised above 52°C (125°F)
5.7.3 When using a solvent for penetrant removal, drying may be accomplished by:
• using a clean lint-free cloth or absorbent paper
• normal evaporation
• warm air circulation, provided the surface temperature is not raised above 52°C (125°F)
Drying time may vary with the type of drying used and the size, nature, geometry, and number of parts
being processed. Drying time shall only be as long as necessary to adequately dry the part.
5.8 Application of developer
5.8.1 Only a wet developer shall be used.
The developer shall be applied as soon as possible after penetrant removal and drying. The time interval
shall not exceed 10 minutes.
The developer shall be applied by spraying by means of an atomizer or a spray gun or an aerosol can to
form a uniform thin coating. Prior to applying suspension type wet developer to the surface, the developer
suspension shall be agitated or shaken to ensure adequate dispersion prior to spraying.
5.8.2 If a solvent suspension (atomizer) is used, surfaces shall be dried by normal evaporation.
5.8.3 If a water suspension is used; Aqueous developers shall be applied to the part immediately after the
excess penetrant has been removed and prior to drying. Aqueous developers shall be prepared and
maintained in accordance with the manufacturer’s instructions and applied in such a manner as to ensure
complete, even, part coverage.
Aqueous developers may be applied by spraying, flowing, or immersing the part in a prepared developer
bath.
Immerse the parts only long enough to coat all of the part surfaces with the developer since indications
may leach out if the parts are left in the bath too long. After the parts are removed from the developer
bath, allow the parts to drain.
Drain all excess developer from recesses and trapped sections to eliminate pooling of developer, which
can obscure discontinuities.
Dry the parts in accordance with 5.7. The dried developer coating appears as a translucent or white
coating on the part.
5.8.4 The developed surface shall remain untouched.
6 Viewing Conditions
6.1 The viewing conditions shall be such that glare and reflections are avoided.
6.2 For color contrast penetrant, the test surface shall be inspected under daylight or artificial light with a light
intensity of at least 1076 lux (100 fc) on the surface of the tested part.
6.3 The light meter shall be calibrated according to one of the standards listed in Section 2.1 at least once a
year or whenever the light meter has been repaired.
6.4 The calibration must be documented and the record maintained on file.
6.5 The light intensity in the examination area, where usually PT-examinations take place, shall be verified at
least once a year with calibrated light meters, documented and the record maintained on file.
8 Assessment of Indications
8.1 Due to diversity of order requirements, defect acceptance will be as specified in the order Quality Plan
(Inspection & Test Plan). In the absence of a Quality Plan (Inspection and Test Plan) the Sulzer quality
organization responsible for the order will advise in writing the appropriate acceptance standard.
- Linear indications
For linear indications, the indication must be separated by a distance greater than the length of an acceptable
indication.
- Rounded indications
- Four or more rounded indications in line separated by 2.0 mm (0.06”) or less edge to edge.
- Linear indications
For linear indications, the indication must be separated by a distance greater than the length of an acceptable
indication.
- Rounded indications
- Four or more rounded indications in line separated by 2.0 mm (0,06”) or less edge to edge.
1 2 3
Linear indication
l = length of indication l ≤ 2 mm l ≤ 4 mm l ≤ 8 mm
Non-linear indication
d = major axle diameter d ≤ 4 mm d ≤ 6 mm d ≤ 8 mm
a Acceptance levels 2 and 3 may be specified with suffix “x” which denotes that all linear indications
detected shall be evaluated to level 1.
However, the probability of detection of indications smaller than those denoted by the original
acceptance level can be low. Linear defect such like crack, lack of fusion and lack of penetration is
NOT acceptable regardless of length.
Note: These three (British) European Standards are not included to this procedure.
Statically Cyclically
Loaded Loaded
Discontinuity Category and Inspection Criteria
Nontubular Nontubular
Connections Connections
(1) Crack Prohibition X X
Any crack shall be unacceptable, regardless of size or location.
(2) Weld/Base Metal Fusion
Complete fusion shall exist between adjacent layers of weld
X X
metal and between weld metal and base metal.
(7) Undercut
(A) For material less than 1 in [25 mm] thick, undercut shall not
exceed 1/32 in [1 mm], with the following exception: undercut
shall not exceed 1/16 in [2 mm] for any accumulated length up X
to 2 in [50 mm] in any 12 in [300 mm].
For material equal to or greater than 1 in [25 mm] thick,
undercut shall not exceed 1/16 in [2 mm] for any length of
weld.
(B) In primary members, undercut shall be no more than 0.01
in [0.25 mm] deep when the weld is transverse to tensile stress
under any design loading condition. Undercut shall be no more X
than 1/32 in [1 mm] deep for all other cases.
(8) Porosity
(A) CJP groove welds in butt joints transverse to the direction of
computed tensile stress shall have no visible piping porosity.
X
For all other groove welds and for fillet welds, the sum of the
visible piping porosity 1/32 in [1 mm] or greater in diameter shall
not exceed 3/8 in [10 mm] in any linear inch of weld and shall
not exceed 3/4 in [20 mm] in any 12 in [300 mm] length of weld.
(B) The frequency of piping porosity in fillet welds shall not
exceed one in each 4 in [100 mm] of weld length and the
maximum diameter shall not exceed 3/32 in [2.5 mm].
X
Exception: for fillet welds connecting stiffeners to web, the sum
of the diameters of piping porosity shall not exceed 3/8 in [10
mm] in any linear inch [25 mm] of weld and shall not exceed
3/4 in [20 mm] in any 12 in [300 mm] length of weld.
(C) CJP groove welds in butt joints transverse to the direction of
computed tensile stress shall have no piping porosity.
For all other groove welds, the frequency of piping porosity shall X
not exceed one in 4 in [100 mm] of length and the maximum
diameter shall not exceed 3/32 in [2.5 mm].
Note: An “X” indicates applicability for the connection type; a shaded area indicates non-applicability
Acceptance criteria for ASTM A514, A517, and A709 Grade HPS 100W [HPS 690W] steels shall be based on
inspection performed not less than 48 hours after completion of the weld.
Sulzer Test Procedure
Technical Quality
Sulzer Pumps
Non-destructive Testing – Ultrasonic Testing (UT) of ferritic and martensitic material for Page:
pump shafts 1 of 2
1.0 Scope
1.1 This procedure details the requirements for ultrasonic testing (UT) of ferritic and martensitic material for pump
shafts.
Revision: 04 Date: 25 May 2018 Amendment: Major rev. - Split in a ferritic / martensitic and an austenitic / duplex version.
This document has been prepared & published under the authority of the Technical Quality Working Group Doc No: TP 0005
6.0 Reporting
6.1 For each examination and for each item tested shall be completed a report. This report has to include at least
following information:
a) Identification of the object tested
b) Surface condition
c) Test procedure (identification and revision)
d) Extent of examination
e) Restrictions which limit the effectiveness of the examination
f) Equipment used (ultrasonic test instrument, probes, calibration and reference blocks, couplant)
g) Scanning diagram showing extent of coverage
h) Results (of the recording indications: size, depth, position and classification of defects, eventually with
drawing)
i) Date of examination
j) Operators' name, qualification (Level) and signature
6.2 Components shall be clearly identified as being either acceptable or rejectable. Non-conforming areas on
components shall be marked on the components themselves, or where this is not possible on an NDE defect
map.
Revision: 04 Date: 25 May 2018 Amendment: Major rev. - Split in a ferritic / martensitic and an austenitic / duplex version.
This document has been prepared & published under the authority of the Technical Quality Working Group Doc No: TP 0005
1.0 Scope
1.1 This procedure details the requirements for the radiographic examination (RT) of welded joints with wall
thickness up to 101 mm, using X-ray or gamma ray source, class T3 or better industrial radiographic films
and wire-type IQI`s, based on the ISO requirements.
Test category B techniques are used when test category A techniques are too insensitive.
The choice of radiographic technique must be agreed between the contracting parties.
Fig. A1 - Maximum X-ray tube voltage of X-ray sources up to 1 000 kV as a function of the irradiated
thickness and material
Revision: 07 Date: 19 Oct 2022 Amendment: Completely revised
This document has been prepared & published under the authority of the Technical Quality Working Group Doc No: TP 0006
Legend
U voltage of the X-ray source, kV w radiated thickness, mm
1 Copper and nickel and their alloys 2 Steel
3 Titanium and its alloys 4 Aluminum and its alloys
NOTE The calculations for the curves in Fig. A1 are described in Enclosure 3.
For some applications where a change in thickness occurs over the radiographed area of the test object, a
change in tube voltage may be made. However, it should be considered that an excessively high tube
voltage will result in a loss of defect detection sensitivity. For copper, nickel and their alloys, the increase in
tube voltage shall not exceed 60 kV. For steel, the increase shall not exceed 50 kV; for titanium and its
alloys, the increase shall not exceed 40 kV; and for aluminum and its alloys, the increase shall not exceed
30 kV. These increases should be applied in proportion to the differences between weld and base metal
thickness.
Permissible ranges of transmitted thickness for gamma-ray sources are given in Table 1.
Safety Requirements
Transmitted thickness
w
Source mm
Test Category A Test Category B
Tm 170 w≤5 w≤5
Yb 169 1 ≤ w ≤ 15 2 ≤ w ≤ 12
Se 75 10 ≤ w ≤ 40 14 ≤ w ≤ 40
Ir 192 20 ≤ w ≤ 100 20 ≤ w ≤ 90
Co 60 40 ≤ w ≤ 200 60 ≤ w ≤ 150
Table 1 - Ranges of transmitted thickness of the test object for gamma-ray sources for steel, copper and
nickel-based alloys.
5.3 Other equipment used during the RT and evaluation processes shall be calibrated and checked at intervals
not to exceed:
- Daily check (or check prior to use) of densitometer and temperature sensors in developing equipment.
- Quarterly calibration of densitometer
- Annually calibration of step wedge film
6 Technique
6.1 Exposure arrangements
6.1.1 General
If possible, radiographic techniques according to 6.1.2 to 6.1.9 (Fig. 1 to Fig. 19) must be used. The films
must be placed as close as possible to the test specimen.
The ellipse technique (double wall/double image) according to Fig. 11 should only be used for De ≤ 100
mm, wall thicknesses t ≤ 8 mm and weld widths ≤ De/4. Two images offset by 90° are sufficient for t/De <
0.12, otherwise three ellipse images are required. The distance between the two depicted welds must
correspond approximately to one weld width.
If it is not possible to perform an elliptical test for De ≤ 100 mm, the perpendicular technique according to
6.1.7 (Fig. 12) may be used. In this case, three uptakes offset by 120° or 60° from each other are required,
depending on the spatial conditions around the pipe.
For the imaging arrangements according to Fig. 13 and Fig. 14, the inclination of the beam must be as
small as possible and directed in such a way that superimposition of the two images is avoided. The
distance beam source - test object, f', for the technique according to Fig. 13 and Fig. 14 must be as small
as possible according to 11.1. The BPK must be placed close to the film together with a lead letter "F" on
the side facing the film.
Appendix A specifies the minimum number of radiographs required to obtain a sufficiently complete
radiographic inspection over the entire circumference of a butt weld in a pipe.
Legend:
S Source
f Distance between source and object
b Distance between source-side of object to film
t Nominal thickness
F Film
Fig. 1 – Exposure arrangement for plane welds with the radiation source on one side and the film on the
opposite side
NOTE If the distance b in Figure 1 is less than 1.2 t, then the nominal thickness t may be used for b and f
can be considered as the distance from the radiation source to the surface of the base metal.
6.1.3 Single-wall radiography of curved test objects with the radiation source outside the test object (see Figure 2
to Figure 4).
Fig. 2 – Exposure arrangement for curved test objects with the radiation source outside and the film inside
NOTE If the distance b in Figure 2 is less than 1.2 t, then the nominal thickness t may be used for b and f
can be considered as the distance from the radiation source to the surface of the base material.
Fig. 3 – Exposure arrangement for inserted welds with the radiation source on the outside and the film on
the inside
Fig. 4 - Arrangement for recording surface-mounted welds with the radiation source outside and the film
inside
6.1.4 Single-wall radiography of curved test specimens with the radiation source inside the test specimen for
central exposures (see Fig. 5 to Fig. 7)
Fig. 5 – Exposure arrangement for welds with centrally located source (central projection) and film
outside
Fig. 6 – Exposure arrangement for inserted welds with the radiation source arranged on the centre axis of
the tube perpendicular to the centre of the weld and film on the outside
Fig. 7 – Exposure arrangement for surface-mounted welds with the radiation source arranged on the centre
axis of the tube perpendicular to the centre of the weld and film on the outside
6.1.5 Single-wall radiography of curved test objects with eccentric radiation source inside the test object (see Fig.
8 to Fig. 10)
Fig. 8 – Exposure arrangement for welds with the radiation source eccentric inside the test object and film
outside
Fig. 9 – Exposure arrangement for inserted welds with eccentric radiation source inside the test object and
film outside
Fig. 10 – Exposure arrangement for surface-mounted welds with eccentric radiation source inside the test
object and film on the outside
6.1.6 Double-wall double-image transmission technique (DW-DI) for tubes with elliptical technique and radiation
source and film outside the test object (see Fig. 11)
Fig. 11 – Exposure arrangement for both tube walls with the ellipse technique
NOTE The distance beam source - test object can be determined on the basis of the perpendicular
distance f', calculated from b'.
6.1.7 Double-wall double-image transmission technique (DW-DI) with perpendicular technique and radiation
source and film outside the test object (see Fig. 12)
Fig. 12 – Exposure arrangement for both tube walls with the perpendicular technique
6.1.8 Double-wall single-image radiographic technique (DW-SI) of curved test specimens for evaluating the wall
near the film (see Fig. 13 to Fig. 16)
Fig. 13 – Exposure arrangement for curved test specimens with radiation source outside for assessing the
wall close to the film with IQI applied film-side
Fig. 14 – Exposure arrangement for curved test specimens with the radiation source located directly on the
surface on the outside for assessing the wall near the film with IQI applied film-side
Fig. 15 – Exposure arrangement for tubes with longitudinal weld seam with radiation source outside for
assessing the wall close to the film with IQI applied film-side
Fig. 16 – Exposure arrangement for inserted weld seams with radiation source outside for assessing the
wall close to the film with IQI applied film-side
6.1.9 Radiography of test specimens with different material thicknesses (see Fig. 17 to Fig. 19)
Legend
1 Wall thickness compensation
Fig. 17 – Exposure arrangement for fillet welds with inclined film arrangement
Fig. 18 – Exposure arrangement for taking fillet welds with vertical film
Markers shall be placed alongside but clear of the outer edges of the weld to identify the position of the
area under examination; see Enclosure 2.
7.1 Film Overlap
The separate radiographs of a welded joint shall overlap to show common markers ensuring that no portion
of the joint remains unexamined.
8.0 Identification of Radiographs
Each section of the weld radiograph shall have suitable symbols affixed to identify:
▪ contract order number
▪ the pipe reference number and/or drawing number
▪ the weld reference number
▪ the section of the weld
▪ date of exposure
▪ IQI location marker
The weld reference number for radiographs of weld repairs shall be clearly identified (e.g. suffixed with the
letter "R" and the repair sequence i.e. "R1", "R2", etc.).
The photographic density of the film shall be as defined in Table 1, below. The density of the film shall be
checked using a calibrated densitometer.
The wire IQI used shall be placed on the side of the test object facing the radiation source, in the center of
the test area to be evaluated (area of interest), on the base metal next to the weld. The identification marks
and, if used, the lead mark "F" shall not be located in the weld test area to be evaluated (HAZ), unless the
geometric configuration does not permit otherwise. The IQI shall be located close to the surface of the test
object; it must lie in an area of uniform thickness characterized by uniform optical density of the film.
If a wire IQI is used, the wires shall be oriented perpendicular to the weld and its location shall ensure that
at least 10 mm of the wire length is in an area of uniform optical density, this area usually being in the base
metal adjacent to the weld. For radiographs in accordance with 6.1.6 and 6.1.7 (Figure 11 and Figure 12),
the IQI with the wires should be placed transverse to the pipe axis; the projection of the wires should not
project into the image of the weld. For outer diameters of the pipe smaller than 50 mm, the visible wire
length may be shorter than 10 mm. In this case, the visible wire length shall be ≥ 20% of the outer diameter
of the pipe.
For single wall radiography techniques in 6.1.4 and 6.1.5 (Figure 5 through Figure 10), the IQI used may be
placed on either the source- or the film-side. If the IQIs cannot be placed on the source-side, they shall be
placed on the film-side and the image quality shall be determined at least once under the same conditions
by a comparison exposure with a IQI on the source-side and a second IQI on the film-side.
In the case of double-wall radiography techniques according to 6.1.6 and 6.1.7 (Figure 11 to Figure 12), the
BPK used shall be placed on the source-side. By agreement between the contracting parties, the IQI may
be arranged on the film-side.
In the case of double wall radiography techniques according to 6.1.8 (Figure 13 to Figure 16), the IQI used
may be arranged on the film-side. If the IQI is placed on the film-side.
If the IQI is placed on the film-side, the letter "F" shall be placed close to the IQI and visible on the
radiographic image and this shall be indicated in the test report.
If it has been ensured that radiographic images of similar test objects and areas are produced with identical
recording and processing techniques and no differences in the image quality number are likely, it is not
necessary to prove the image quality for each radiographic image. The extent of image quality verification
should be agreed between the contracting parties.
For radiographs of pipes with a centrally located radiation source, at least three IQI should be distributed
evenly around the circumference. The films with the images of the IQIs are then considered representative
of the entire circumference.
Values from the manufacturer may be used if they comply with these documents.
If the focal spot size is described by two dimensions, the larger of the two must be used.
The distance f or f' must be chosen so that the ratio of this distance to the focal spot size d, i.e. f/d or f'/d,
does not fall below the values given by equation (1) and equation (2):
For simplicity, only f and b are used in these equations. The equations also apply to f' and b', as shown in
Fig. 11, 12, Fig. 13 and 14.
If the distance b < 1.2t, the dimension b in equation (1) and equation (2) and in Figure 20 must be replaced
by the nominal thickness t.
To determine the distance source - object fmin or f'min, the ladder diagram in Fig. 20 can be used. The ladder
diagram is based on equation (1) and equation (2).
If planar irregularities shall be verified for test category A, the minimum distance fmin or f'min shall be the
same as for test category B, thus reducing the geometric uncertainty by a factor of 2.
In critical technical applications on crack-sensitive materials, a more sensitive radiographic technique than
test category B must be used.
Legend
Fig. 20 - Ladder diagram for determining the minimum distance between source and test object fmin as a
function of the distance between test object and film b and the size d of the source.
If the ellipse technique according to 6.1.6 (Fig. 11) or the perpendicular technique according to 6.1.7 (Fig.
12) is used, the outer diameter Do of the tube must be used as b or b' in equation (1) and equation (2) and
in Fig. 20.
If the radiation source is outside the test object and the film is on the other side [technique according to
6.1.8 (Fig. 13 to Fig. 16) as double-wall radiation/single-image evaluation], the minimum distance radiation
source - test object is determined only by the wall thickness, i.e. not by the tube diameter.
Where possible, the double wall radiography technique (see 6.1.8, Figure 13 to Figure 16) should
preferably be avoided by locating the radiation source within the test object to be radiographed (see 6.1.4
and 6.1.5, Figure 5 to Figure 10) to achieve a more suitable alignment of the test. The reduction of the
minimum radiation source - test object distance should not exceed 20%, provided that the IQI requirements
are met.
If the radiation source is centered inside the test object and the film is outside (technique according to
6.1.4, Figure 5), this percentage value may be increased, provided the IQI requirements are met. However,
the reduction of the minimum distance source - test object must not exceed 50 %. A further reduction of the
minimum distance may be agreed between the contracting parties, provided that the IQI requirements are
met.
Radiographic film shall be ISO 11699-1 Class T3 (ASTM E 1815 Type II) or better and have a
maximum fog density of 0.3.
Table 10 X-ray source voltage, U, film system classes and metallic screens for radiography of steel,
copper and nickel-based alloys.
All screen shall be clean and free from scratches, grease, dirt and all that can interfere with the
interpretation of the film.
Preferred vacuum packed films shall be used.
14.0 Processing
The film shall be processed manually or automatically in accordance with requirements of the film system
used. The film or the processing solution or both shall be agitated during development. The development
time and temperature shall be in accordance with the film manufacturer’s recommendations. The
radiographs shall be free from processing imperfections or defects, which would interfere with
interpretation.
Films shall be processed in accordance with the ISO 11699-2 Standard Guide for Controlling the Quality of
Industrial Radiographic Film Processing. The storage durability of the radiographic film shall be 25 years
minimum.
15.0 Viewing
The radiographs shall be examined by diffused light in a darkened room with the illuminated area that will
not cause troublesome reflections, shadows or glare on the radiograph image. Ambient light in the viewing
area shall not exceed 30 Lux (lx).
Viewing equipment for interpretation shall include a light source sufficient for the essential IQI to be visible
for the specified density range.
The films shall be free from all blemishes, mechanical damage or defects, which would interfere with
interpretation.
See attachment 1;
1. ISO 10675-1 Non-destructive testing of welds – Acceptance levels for radiographic testing –
Part 1: Steel, nickel, titanium and their alloys
17.0 Reporting
For each examination and for each item tested shall be completed a report. This report has to include at
least following information:
▪ Identification of the object tested
▪ Material thickness
▪ Welding process used
▪ Test procedure (identification and revision)
▪ Extent of examination
▪ Restrictions which limit the effectiveness of the examination
▪ Equipment used (X-ray tube, manufacturer, kV used, focus size and angle of beam to film)
▪ Radiographic technique (exposure time, source to film distance and object to film distance)
▪ Film brand and type, screen and filter details
▪ Density range
▪ Type of IQI and sensitivity
▪ Processing technique
▪ Acceptance criteria
▪ Interpretation, results (type of defects)
▪ Date of examination and date of evaluation, if different
▪ Operators'/film viewers' name, qualification (Level) and signature
▪ Dimensional map of marker placement (if used)
▪ Number of radiographs (exposures)
▪ Number of film in each film holder or cassette
▪ Single- or double-wall exposure
▪ Single- or double-wall viewing
▪ A listing of each radiograph location
Components shall be clearly identified as being either acceptable or rejectable. Non-conforming areas on
components shall be marked on the components themselves, or where this is not possible on an NDE
defect map.
Customer shall be notified if any examined item is rejectable.
The minimum number of fixtures required is shown in Fig. B.1 to Fig. B.4, which apply to all pipes except
those tested with double-wall/double-image techniques according to 6.1.6 and 6.1.7 (Fig. 11 and Fig.
12).
If the deviation of the wall thickness Δt/t of the weld to be tested does not exceed 20% (test category A)
when using a single image, Figure B.3 and Figure B.4 are applied. This technique is only recommended
if the probability of transverse cracks is low or if the weld is tested for such irregularities using other non-
destructive testing methods.
If Δt/t is 10% or less (test category B), Figure B.1 and Figure B.2 are applied. In this case, it is likely that
transverse cracks will also be detected.
If the test object is tested for individual transverse cracks, the required minimum number of radiographs
must be higher than the values in Figure B.1 to Figure B.4.
Figure B.1 - Minimum number of exposures N for single-wall radiography with external radiation source at a
maximum permissible increase in the radiographed thickness Δt/t = 10 % (test category B) as a function
of the ratios t/De and De/f, due to oblique radiography in the assessable areas with sufficient optical density
Legend
1 Area of the pipe wall (inaccessible)
Fig. B.2 - Minimum number of exposures N for eccentric radiography with internal radiation source as well
as for double-wall radiography with a maximum permissible increase in the radiographed thickness Δt/t =
10 % (test category B) as a function of the ratios t/De and De/SFD, due to the oblique radiography in the
assessable areas with sufficient optical density
Figure B.3 - Minimum number of exposures N for single-wall radiography with external radiation source at a
maximum permissible increase in the radiographed thickness Δt/t = 20 % (test category A) as a function
of the ratios t/De and De/f, due to oblique radiography in the assessable areas with sufficient optical density
Legend
1 Area of the pipe wall (inaccessible)
Fig. A.4 - Minimum number of exposures N for eccentric radiography with internal radiation source as well
as for double-wall radiography with a maximum permissible increase in the radiographed thickness Δt/t =
20 % (test category A) as a function of the ratios t/De and De/SFD, due to the oblique radiography in the
assessable areas with sufficient optical density
YES t
YES YES
Mf
X X
NO
t Mf
plates, longitudinal welds only if additional coverage: X =
D2
NO NO
YES YES
YES YES
NO
NO NO
YES YES
curved components from the inside curved components from the outside
with source-film distance greater than radius
Table C.1 - Approximate formulas for calculating maximum tube stresses as used in
Figure A1 used
ISO 10675-1
No. Irregularity Class 3 a) Class 2 a) Class 1
1 crack (100)
longitudinal (101) Not acceptable Not acceptable Not acceptable
transverse (102)
crater (104)
2a uniformly distributed A ≤ 2,5 % A ≤ 1,5 % A≤1%
porosity and gas pores d ≤ 0,4s, max. 5 mm d ≤ 0,3s, max. 4 mm d ≤ 0,2s, max. 3 mm
(2012, 2011)
L = 100 mm L = 100 mm L = 100 mm
single-layer
2b uniformly distributed A≤5% A≤3% A≤2%
porosity and gas pores d ≤ 0,4s, max. 5 mm d ≤ 0,3s, max. 4 mm d ≤ 0,2s, max. 3 mm
(2012, 2011)
L = 100 mm L = 100 mm L = 100 mm
multi-layer
3 b) clustered (localized) dA ≤ wp, max. 25 dA ≤ wp, max. 20 dA ≤ wp/2, max. 15
porosity (2013) mm mm mm
d ≤ 0,4s, max. 5 mm d ≤ 0,3s, max. 4 mm d ≤ 0,2s, max. 3 mm
4 c) linear porosity (2014) l ≤ s, max. 75 mm l ≤ s, max. 50 mm l ≤ s, max. 25 mm
d ≤ 0,4s, max. 4 mm d ≤ 0,3s, max. 3 mm d ≤ 0,2s, max. 2 mm
L = 100 mm L = 100 mm L = 100 mm
d)
5 elongated cavity (2015) h < 0,4s, max. 4 mm h < 0,3s, max. 3 mm h < 0,2s, max. 2 mm
worm-hole (2016) Σl ≤ s, max. 75 mm Σl ≤ s, max. 50 mm Σl ≤ s, max. 25 mm
L = 100 mm L = 100 mm L = 100 mm
e)
6 shrinkage cavity h < 0,4s, max. 4 mm
(except crater pipes) l ≤ 25 mm Not acceptable Not acceptable
(202)
7 crater pipes (2024) h ≤ 0,2t, max. 2 mm Not acceptable Not acceptable
l ≤ 0,2t, max. 2 mm
ISO 10675-1
No. Irregularity Class 3 a) Class 2 a) Class 1
13 crater crack (104) Not acceptable Not acceptable Not acceptable
14a continuous undercut soft transition soft transition soft transition
(5011) intermittent required required required
undercut (5012) h ≤ 0,2t, max. 1 mm h ≤ 0,1t, max. 0,5 h ≤ 0,05t, max. 0,5
t > 3 mm mm mm
b)
14b continuous undercut soft transition soft transition
(5011) intermittent required required Not acceptable
undercut (5012) l ≤ 25 mm, h ≤ 0,2t l ≤ 25 mm, h ≤ 0,1t
0,5 mm ≤ t ≤ 3 mm
15a b) shrinkage grooves (5013) soft transition soft transition soft transition
t > 3 mm required required required
l ≤ 25 mm, l ≤ 25 mm, l ≤ 25 mm,
h ≤ 0,2t, max. 2 mm h ≤ 0,1t, max. 1 mm h ≤ 0,05t, max. 0,5
mm
15b b) shrinkage grooves (5013) soft transition soft transition
0,5 mm ≤ t ≤ 3 mm required required Not acceptable
h ≤ 0,2 mm + 0,1t l ≤ 25 mm, h ≤ 0,1t
19b b) root concavity (515) h ≤ 0,2 mm + 0,1t l ≤ 25 mm, h ≤ 0,1t Not acceptable
0,5 mm ≤ t ≤ 3 mm
20 poor restart (517) Permissible, the limit
t ≥ 0,5 mm depends on the type Not acceptable Not acceptable
of irregularity
21a b) Sagging (509) l ≤ 25 mm, h ≤ 0,25t, l ≤ 25 mm, h ≤ 0,1t, l ≤ 25 mm, h ≤ 0,05t,
incompletely filled groove max. 2 mm max. 1 mm max. 0,5 mm
(511)
t > 3 mm
21b b) Sagging (509) l ≤ 25 mm, h ≤ 0,25t l ≤ 25 mm, h ≤ 0,1t Not acceptable
incompletely filled groove
(511)
0,5 mm ≤ t ≤ 3 mm
22a linear misalignment (507) h ≤ 0,2 mm + 0,25t h ≤ 0,2 mm + 0,15t h ≤ 0,2 mm + 0,1t
0,5 mm ≤ t ≤ 3 mm
22b linear misalignment (507) h ≤ 0,25t, max. 5 mm h ≤ 0,15t, max. 4 h ≤ 0,1t, max. 3 mm
Longitudinal welds mm
t > 3 mm
22c linear misalignment (507) h ≤ 0,5t, max. 4 mm h ≤ 0,5t, max. 3 mm h ≤ 0,5t, max. 2 mm
circular welds
t ≥ 0,5 mm
NOTE The allowability limits correspond to the limits specified for visual inspection. These irregularities are
usually evaluated by means of visual inspection.
a) The admissibility limits 3 and 2 may be followed by an X indicating that all displays over 25 mm are
not allowed.
b) If the length of the weld is less than 100 mm, the maximum length of the indicators must not exceed
25% of the length of the weld.
Figure B.1 to Figure B.9 show different area fractions (in %) of irregularities in a 60 mm × 20 mm area. These
should support the evaluation of irregularities in radiographs.
Legend
A1 Clustered area of indication 1
A2 Clustered area of indication 2
dA1 Diameter of A1
dA2 Diameter of A2
D Distance between indications
wp Width of weld seam
L Test length of 100 mm each
Legend
A1 Clustered area of indication 1
A2 Clustered area of indication 2
dA1 Diameter of A1
dA2 Diameter of A2
dAc Diameter of the envelope around the entire clustered area
D Distance between indications
wp Width of weld seam
L Test length of 100 mm each
The total pore area in the clustered (localized) porosity is represented by a circle of diameter dAc enclosing all
pores.
For all pores in this circle, the conditions for a single pore must be met. A permissible pore area must be localized.
If D is less than dA1 or dA2, whichever is smaller, the entire pore area is represented by a circle of diameter dAc,
where dAc = dA1 + dA2 + D.
Systematic clustered (localized) porosity are not allowed. dA corresponds to dA1, dA2 or dAc. whichever is applicable.
Legend
d1 Diameter of pore of indication 1
d2 Diameter of pore of indication 2
D Distance between indications
l Length of indication
L Test length of 100 mm each
wp Width of weld seam
Legend
d1 Diameter of the pore of indication 1
d2 Diameter of the pore of indication 2
D Distance between indications
l Length of indication
L Test length of 100 mm each
wp Width of weld seam
The sum of the different pore areas must be calculated with respect to the evaluation area L × wp (see Figure C.3).
If D is smaller than the smaller diameter of one of the adjacent pores, the enveloping area of both pores shall be
considered in the sum of the irregularities (see Figure C.4).
Revision: 07 Date: 19 Oct 2022 Amendment: Completely revised
This document has been prepared & published under the authority of the Technical Quality Working Group Doc No: TP 0006
Legend
l1 Length of the cavity of indication 1
l2 Length of the cavity of indication 2
l3 Length of the cavity of indication 3
h1 Height of the cavity of indication 1
h2 Height of the cavity of indication 2
h3 Height of the cavity of indication 3
D Distance between indications
L Test length of 100 mm each
wp Width of weld seam
Legend
l1 Length of the cavity of indication 1
l2 Length of the cavity of indication 2
l3 Length of the cavity of indication 3
h1 Height of the cavity of indication 1
h2 Height of the cavity of indication 2
h3 Height of the cavity of indication 3
D Distance between indications
L Test length of 100 mm each
wp Width of weld seam
The sum of the length of the indicators Σl must be determined for each test length L (see Figure C.5).
If D is less than the shorter length of an adjacent irregularity, the total length over both irregularities shall be
included in the sum of irregularities (see Figure C.6).
1.0 Scope
This procedure details the requirements for the radiographic examination of castings using a X-ray or
gamma ray source, industrial radiographic films and wire-type IQI`s.
Revision: 7 Date: 12 Aug 2020 Amendment: Completely rewritten for ASME BPVC 2019 update.
This document has been prepared & published under the authority of the Technical Quality Working Group Doc No: TQ TP-0007
5.0 Source
Radioactive Isotopes (Co-60, Ir.192 and SE-75), or X-Ray tubes can be used.
An X-ray or gamma ray source of adequate power to meet the film quality requirements as specified in
§ 11 shall be used.
X Radiation Sources Selection of appropriate X-ray voltage and current levels is dependent upon
variables regarding the specimen being examined (material type and thickness) and economically
permissible exposure time.
The suitability of these X-ray parameters shall be demonstrated by attainment of required penetrame-
ter (IQI) sensitivity and compliance with all other requirements stipulated herein. Guide E94 contains
provisions concerning exposure calculations and charts for the use of X-ray sources.
Gamma Radiation Sources Isotope sources, when used, shall be capable of demonstrating the re-
quired radiographic sensitivity.
The radiation energy employed for any radiographic technique shall achieve the density and IQI image
requirements of the ASME BPV Code Section V, Article 2.
5.2 Other equipment used during the RT and evaluation processes shall be calibrated and checked at
intervals not to exceed:
- Daily check (or check prior to use) of densitometer and temperature sensors in developing equipment.
- Quarterly calibration of densitometer
- Annual calibration of step wedge film
- Annual calibration of film viewer
6.0 Visual examination
Casting surfaces shall be visually inspected for defects that may impair the quality of the radiograph pro-
duced.
7.0 Method of test
All radiography shall be performed in accordance with ASME V, T-270.
The direction of the central beam of radiation shall be centred on the area of interest.
Revision: 7 Date: 12 Aug 2020 Amendment: Completely rewritten for ASME BPVC 2019 update.
This document has been prepared & published under the authority of the Technical Quality Working Group Doc No: TQ TP-0007
For composite viewing of multiple film exposures, each film of the composite set shall have a minimum
density of 1.0.
The maximum density shall be 4.0 for either single or composite viewing. A tolerance of 0.05 in density is
allowed for variations between densitometer readings.
The density of the film shall be checked using a calibrated densitometer.
Revision: 7 Date: 12 Aug 2020 Amendment: Completely rewritten for ASME BPVC 2019 update.
This document has been prepared & published under the authority of the Technical Quality Working Group Doc No: TQ TP-0007
Revision: 7 Date: 12 Aug 2020 Amendment: Completely rewritten for ASME BPVC 2019 update.
This document has been prepared & published under the authority of the Technical Quality Working Group Doc No: TQ TP-0007
Revision: 7 Date: 12 Aug 2020 Amendment: Completely rewritten for ASME BPVC 2019 update.
This document has been prepared & published under the authority of the Technical Quality Working Group Doc No: TQ TP-0007
Any group IQI may be used for any material with a higher group number, provided the applicable quality
level is maintained.
9.5 For cylindrical or flat castings where more than one film holder is used for an exposure, at least one
penetrameter (IQI) image shall appear on each radiograph. For cylindrical shapes, where a panoramic
type source of radiation is placed in the center of the cylinder and a complete or partial circumference is
radiographed using at least four overlapped film holders, at least three penetrameters (IQI) shall be used.
On partial circumference exposures, a penetrameter (IQI) shall be placed at each end of the length of the
image to be evaluated on the radiograph with the intermediate penetrameters (IQI) placed at equal divi-
sions of the length covered. For full circumferential coverage, three penetrameters (IQI) spaced 120°
apart shall be used, even when using a single length of roll film.
When an array of individual castings in a circle is radiographed, the requirements of a) or b), or both, shall
prevail for each casting.
If the required penetrameter (IQI) sensitivity does not show on any one film in a multiple film technique
(see §10.9), but does show in composite (superimposed) film viewing, interpretation shall be permitted
only by composite film viewing for the respective area.
9.6 When it is not practicable to place the penetrameter(s) (IQI) on the casting, a shim or separate block
conforming to the requirements of §9.7 may be used.
The penetrameter (IQI) shall be no closer to the film than the source side of that part of the casting being
radiographed in the current view.
The radiographic density measured adjacent to the penetrameter (IQI) through the body of the shim or
separate block shall not exceed the density measured in the area of interest by more than 15 %. The
density may be lighter than the area of interest density, provided acceptable quality level is obtained and
the density requirements of §8 are met.
Revision: 7 Date: 12 Aug 2020 Amendment: Completely rewritten for ASME BPVC 2019 update.
This document has been prepared & published under the authority of the Technical Quality Working Group Doc No: TQ TP-0007
9.7 Shims and Separate Blocks Shims or separate blocks made of the same or radiographically similar ma-
terials (as defined in Method E1025) may be used to facilitate penetrameter positioning. There is no re-
striction on shim or separate block thickness provided the penetrameter and area-of-interest density tol-
erance requirements of §8.2 are met.
The shim or separate block shall be placed at the corner of the film holder or close to that part of the area
of interest that is furthest from the central beam. This is the worst case position from a beam angle
standpoint that a discontinuity would be in.
The shim or separate block dimensions shall exceed the penetrameter (IQI) dimensions such that the
outline of at least three sides of the penetrameter (IQI) image shall be visible on the radiograph.
9.8 Film Side Penetrameter (IQI) In the case where the penetrameter (IQI) cannot be physically placed on
the source side and the use of a separate block technique is not practical, penetrameters (IQI) placed on
the film side may be used. The applicable job order or contract shall dictate the requirements for film side
radiographic quality level (see table below).
9.9 Additional IQI Selection Requirements.
The thickness on which the IQI is based is the single-wall thickness.
(a) Casting Areas Prior to Finish Machining.
The IQI shall be based on a thickness that does not exceed the finished thickness by more than 20% or
1/4 in. (6 mm), whichever is greater. In no case shall an IQI size be based on a thickness greater than the
thickness being radiographed.
(b) Casting Areas That Will Remain in the As-Cast Condition.
The IQI shall be based on the thickness being radiographed.
The IQI selection and essential hole, or wire size shall comply with ASME V, table T-276.
Revision: 7 Date: 12 Aug 2020 Amendment: Completely rewritten for ASME BPVC 2019 update.
This document has been prepared & published under the authority of the Technical Quality Working Group Doc No: TQ TP-0007
Revision: 7 Date: 12 Aug 2020 Amendment: Completely rewritten for ASME BPVC 2019 update.
This document has been prepared & published under the authority of the Technical Quality Working Group Doc No: TQ TP-0007
11.2 Screens
Source Screens (Front- / Back-)
Material Thickness
(mm)
X-Ray Tube ≤ 150 kV Pb none - 0.15
150 – 250 kV Pb 0.02 - 0.15
250 – 320 kV Pb 0.02 - 0.2
320 – 450 kV Pb 0.1 - 0.2
Se 75 Pb 0.1 - 0.2
Ir 192 Pb 0.1 - 0.2
Co 60 Pb 0.1 - 0.5
Fe, or Cu 0.25 - 0.7
≤ 2 MeV Pb 0.1 - 0.5
Fe, or Cu 0.25 - 0.7
11.3 Multiple Film Techniques
Two or more films of equal or different speeds in the same cassette are allowed, provided prescribed
quality level and density requirements are met (see §8 & §9).
12.0 Interception of unwanted and scattered radiation
As a check on back-scattered radiation a lead symbol "B" with minimum dimensions of 11 mm in height
and 1,5 mm thickness shall be attached to the back of each film holder.
The lead symbol “B” shall be placed in a location so that it would appear within an area on the radio-
graph that meets the requirements of §8.
If a light image appears on a darker background of the radiograph, protection from backscatter is insuf-
ficient and the radiograph shall be considered unacceptable.
A dark image of the “B” on a lighter background is not cause for rejection.
13.0 Processing
The film shall be processed manually or automatically in accordance with requirements of the film sys-
tem used. The film or the processing solution or both shall be agitated during development. The devel-
opment time and temperature shall be in accordance with the film manufacturer’s recommendations.
All radiographs shall be free from mechanical, chemical, or other blemishes to the extent that they do
not mask and are not confused with the image of any discontinuity in the area of interest of the object
being radiographed. Such blemishes include, but are not limited to:
(a) fogging;
(b) processing defects such as streaks, watermarks, or chemical stains;
(c) scratches, finger marks, crimps, dirtiness, static marks, smudges, or tears;
(d) false indications due to defective screens.
Revision: 7 Date: 12 Aug 2020 Amendment: Completely rewritten for ASME BPVC 2019 update.
This document has been prepared & published under the authority of the Technical Quality Working Group Doc No: TQ TP-0007
14.0 Viewing
The radiographs shall be examined by diffused light in a darkened room with an illuminated area that
will not cause troublesome reflections, shadows or glare on the radiograph image.
Viewing equipment for interpretation shall include a light source sufficient for the essential penetrame-
ter to be visible for the specified density range.
The viewing conditions shall be such that light from around the outer edge of the radiograph or coming
through low-density portions of the radiograph does not interfere with interpretation.
The maximum achievable brightness (cd/m2) shall be determined as per ASTM E 1390 and the ag-
gregate value (maximum viewable density) shall be permanently marked on each film viewer.
15.0 Reporting
For each examination and for each item tested shall be completed a report. This report has to include
at least following information:
Identification of the object tested
Material thickness
Test procedure (identification and revision)
Extent of examination
Restrictions which limit the effectiveness of the examination
Equipment used (X-ray tube or isotope, focal spot size of X-ray or source size)
Radiographic technique (sketch showing the geometric arrangement of the radiograph, to-
gether with the location markers and angle of beam to film)
Film brand and type, screen and filter details
Density range
Type of IQI and sensitivity
Processing technique
Acceptance criteria
Interpretation, results (type of defects)
Date of examination
Operators'/film viewers' name, qualification and signature
Components shall be clearly identified as being either acceptable or rejectable. Non-conforming areas
on components shall be marked on the components themselves, or where this is not possible on an
NDE defect map.
16.0 Assessment of indications
Due to diversity of project requirements defect acceptance will be as specified in the project Quality
Plan (Inspection&Test Plan). In the absence of a Quality Plan (Inspection&Test Plan) the Sulzer quality
organisation responsible for the project will advise in writing the appropriate acceptance standard.
Revision: 7 Date: 12 Aug 2020 Amendment: Completely rewritten for ASME BPVC 2019 update.
This document has been prepared & published under the authority of the Technical Quality Working Group Doc No: TQ TP-0007
Revision: 7 Date: 12 Aug 2020 Amendment: Completely rewritten for ASME BPVC 2019 update.
This document has been prepared & published under the authority of the Technical Quality Working Group Doc No: TQ TP-0007
Revision: 7 Date: 12 Aug 2020 Amendment: Completely rewritten for ASME BPVC 2019 update.
This document has been prepared & published under the authority of the Technical Quality Working Group Doc No: TQ TP-0007
Revision: 7 Date: 12 Aug 2020 Amendment: Completely rewritten for ASME BPVC 2019 update.
This document has been prepared & published under the authority of the Technical Quality Working Group Doc No: TQ TP-0007
Revision: 7 Date: 12 Aug 2020 Amendment: Completely rewritten for ASME BPVC 2019 update.
This document has been prepared & published under the authority of the Technical Quality Working Group Doc No: TQ TP-0007
For castings having radiographed thicknesses from 51 mm to 114 mm, the radiographs shall be com-
pared to those in ASTM E186, Standard Reference Radiographs for Heavy‐Walled 50.8 to 114 mm
Steel Castings. The maximum acceptable severity levels for imperfections shall be as follows:
A – Gas porosity 2
B – Sand and slag inclusions 2
C – Shrinkage
Type 1 1
Type 2 2
Type 3 3
D - Cracks 0
E – Hot tears 0
F - Inserts 0
Revision: 7 Date: 12 Aug 2020 Amendment: Completely rewritten for ASME BPVC 2019 update.
This document has been prepared & published under the authority of the Technical Quality Working Group Doc No: TQ TP-0007
(2) All surfaces including machined gasket seating surfaces shall be examined by the magnetic particle
or the liquid penetrant method. When the casting specification requires heat treatment, these examina-
tions shall be conducted after that heat treatment.
(3) Surface indications determined by magnetic particle examination shall be compared with those
indicated in ASTM E125, Standard Reference Photographs for Magnetic Particle Indications on
Ferrous Castings, and shall be removed if they exceed the following limits:
Type Degree
(4) Surface indications determined by liquid penetrant examination are unacceptable if they exceed the
following limits:
(-a) all cracks and hot tears;
(-b) any group of more than six linear indications other than those in (a) above in any rectangular area
of 38 mm × 150 mm or less or any circular area having a diameter of 88 mm or less, these areas being
taken in the most unfavorable location relative to the indications being evaluated;
(-c) other linear indications
more than 6 mm long for thicknesses up to 19 mm inclusive,
more than one‐third of the thickness in length for thicknesses from 19 mm to 57 mm,
and more than 19 mm long for thicknesses over 57 mm
(aligned acceptable imperfections separated from one another by a distance equal to the length of the
longer imperfection are acceptable);
(-d) all indications of nonlinear imperfections which have any dimension exceeding 5 mm.
(5)When more than one casting of a particular design is produced, each of the first five shall be exam-
ined to the full extent prescribed herein. When more than five castings are being produced, examina-
tions as prescribed shall be performed on the first five and on one additional casting for each additional
five castings produced. If any of these additional castings proves to be unacceptable, each of the re-
maining four castings of that group shall be examined fully.
Revision: 7 Date: 12 Aug 2020 Amendment: Completely rewritten for ASME BPVC 2019 update.
This document has been prepared & published under the authority of the Technical Quality Working Group Doc No: TQ TP-0007
(b) All castings having maximum body thickness 114 mm and greater and castings of lesser
thickness which are intended for severe service applications shall be examined as follows.
(1) Each casting shall be subjected to 100% visual examination and to complete surface examination
by either the magnetic particle or the liquid penetrant method. When the casting specification requires
heat treatment, these examinations shall be conducted after that heat treatment. Acceptability limits for
surface imperfections shall be as given in (a)(3) and (a)(4) above.
(2) All parts of castings up to 300 mm in thickness shall be subjected to radiographic examination and
the radiographs compared to those given in ASTM E280, Standard Reference Radiographs for
Heavy‐Walled 114‐mm to 305‐mm Steel Castings. The maximum acceptable severity levels for im-
perfections shall be as follows:
A – Gas porosity 2
B – Sand and slag inclusions 2
C – Shrinkage
Type 1 2
Type 2 2
Type 3 2
D - Cracks 0
E – Hot tears 0
F - Inserts 0
(3) For castings having a maximum thickness in excess of 12 in.(300 mm), all thicknesses which are
less than 12 in. (300 mm) shall be examined radiographically in accordance with the preceding para-
graph. All parts of such castings having thicknesses in excess of 12 in. (300 mm) shall be examined
ultrasonically in accordance with Section V, Article 5.
Any imperfections which do not produce indications exceeding 20% of the straight beam back reflec-
tion or do not reduce the height of the back reflection by more than 30% during a total movement of the
transducer of 2 in. (50 mm) in any direction shall be considered acceptable. Imperfections exceeding
these limits shall be repaired unless proved to be acceptable by other examination methods.
Revision: 7 Date: 12 Aug 2020 Amendment: Completely rewritten for ASME BPVC 2019 update.
This document has been prepared & published under the authority of the Technical Quality Working Group Doc No: TQ TP-0007
For castings having radiographed thicknesses from 51 mm to 114 mm, the radiographs shall be com-
pared to those in ASTM E186, Standard Reference Radiographs for Heavy‐Walled 50.8 to 114 mm
Steel Castings. The maximum acceptable severity levels for imperfections shall be as follows:
A – Gas porosity 2
B – Sand and slag inclusions 2
C – Shrinkage
Type 1 2
Type 2 2
Type 3 2
D - Cracks 0
E – Hot tears 0
F - Inserts 0
Revision: 7 Date: 12 Aug 2020 Amendment: Completely rewritten for ASME BPVC 2019 update.
This document has been prepared & published under the authority of the Technical Quality Working Group Doc No: TQ TP-0007
(2) All surfaces of each casting shall be examined by the magnetic particle or the liquid penetrant
method. When the casting specification requires heat treatment, these examinations shall be conduct-
ed after that heat treatment.
(3) Surface indications shall be judged acceptability in accordance with MSS SP-53 Table 1, using the
ASTM E125, Standard Reference Photographs as a reference for surface flaws:
Maximum acceptable indications are as follows:
Linear indications
8 mm long for thicknesses up to 13 mm inclusive,
13 mm long for thicknesses from 13 mm to 25 mm,
18 mm long for thicknesses over 25 mm
Aligned linear indications separated from one another by a distance greater than the length of an ac-
ceptable indication.
Rounded indications
8 mm dia. for thickness up to 13 mm inclusive,
13 mm long for thicknesses over 13 mm
Four or more rounded indications in a line separated by 2 mm or less edge to edge are unacceptable
(b) All castings having maximum body thickness 114 mm and greater and castings of lesser
thickness which are intended for severe service applications shall be examined as follows.
(1) Each casting shall be subjected to 100% visual examination and to complete surface examination
by either the magnetic particle or the liquid penetrant method. When the casting specification requires
heat treatment, these examinations shall be conducted after that heat treatment. Acceptability limits for
surface imperfections shall be as given in (a)(3) and (a)(4) above.
(2) All parts of castings up to 300 mm in thickness shall be subjected to radiographic examination and
the radiographs compared to those given in ASTM E280, Standard Reference Radiographs for
Heavy‐Walled 114‐mm to 305‐mm Steel Castings. The maximum acceptable severity levels for im-
perfections shall be as follows:
A – Gas porosity 2
B – Sand and slag inclusions 2
C – Shrinkage
Type 1 2
Type 2 2
Type 3 2
D - Cracks 0
E – Hot tears 0
F - Inserts 0
Revision: 7 Date: 12 Aug 2020 Amendment: Completely rewritten for ASME BPVC 2019 update.
This document has been prepared & published under the authority of the Technical Quality Working Group Doc No: TQ TP-0007