Professional Documents
Culture Documents
Berlin 2012
Berlin 2012
http://jic.sagepub.com/
Published by:
http://www.sagepublications.com
Additional services and information for Journal of Intensive Care Medicine can be found at:
Subscriptions: http://jic.sagepub.com/subscriptions
Reprints: http://www.sagepub.com/journalsReprints.nav
Permissions: http://www.sagepub.com/journalsPermissions.nav
What is This?
Abstract
Auto–positive end-expiratory pressure (PEEP) is a common but frequently unrecognized problem in critically ill patients. It has
important physiologic consequences and can cause shock and cardiac arrest. Treatment consists of relieving expiratory airflow
obstruction and reducing minute ventilation delivered by positive pressure ventilation. Sedation and fluid management are impor-
tant adjunctive therapies. This analytic review discusses the prevalence, pathophysiology, and hemodynamic consequences of
auto-PEEP and an approach to its treatment.
Keywords
Auto–positive end-expiratory pressure, dynamic hyperinflation, shock, mechanical ventilation
Received November 3, 2011, and in revised November 29, 2011. Accepted December 13, 2011.
Positive end-expiratory pressure (PEEP) is the distention of the postmortem examination cannot detect its presence.8 There are
lungs with air above atmospheric pressure prior to the onset of a few reports of its prevalence in specific critically ill populations.
inspiration. Clinicians can intentionally apply extrinsic PEEP In one series, 47% of patients on mechanical ventilation had evi-
with mechanical ventilation. This is a positive pressure target dence of at least minimal auto-PEEP.13 In another series, it was
the ventilator maintains in the airway circuit during exhalation. observed in the majority of critically ill patients on mechanical
PEEP can also arise as an intrinsic consequence of a patient’s ventilation. All the patients in this series with chronic obstructive
ventilatory pattern. Pepe and Marini coined the term auto- pulmonary disease had auto-PEEP.14 In a series of patients with
PEEP in 1982 to describe this phenomenon.1 Extrinsic PEEP the acute respiratory distress syndrome (ARDS), most of whom
is a therapy set by a clinician on a ventilator. Auto-PEEP is had stable hemodynamics and high levels of extrinsic PEEP, there
unintentional. It can occur in people with expiratory airflow was a low prevalence of auto-PEEP,15 while others have noted a
obstruction who are breathing without the assistance of higher prevalence.16 Since it is defined as the excess pressure
mechanical ventilation but is more commonly associated with above the set PEEP, the prevalence of auto-PEEP will be lower
positive pressure ventilation.2 By convention, auto-PEEP is with ventilatory strategies that use high levels of extrinsic
defined as the magnitude of the end-expiratory pressure in PEEP.17 Prone ventilation may reduce the prevalence of auto-
excess of the set extrinsic PEEP. High minute ventilation and PEEP in ARDS.18 Trauma is a major risk factor for developing
expiratory airflow obstruction are the most important risk fac- auto-PEEP, likely due to hyperventilation from metabolic acido-
tors for the development of auto-PEEP. Intrinsic and extrinsic sis and shock.13,19 Hyperventilation during cardiopulmonary
PEEP have similar effects on respiratory mechanics, gas resuscitation (CPR) is another common cause.20,21
exchange, and hemodynamics.3 Auto-PEEP is a common cause
of dyspnea, patient–ventilator dysynchrony, and the inability to
trigger mechanical ventilation.4 Since its first description, auto- Pathophysiology of Auto-PEEP
PEEP is increasingly recognized as an important but treatable One form of auto-PEEP is caused by activation of the muscles
cause of shock and cardiac arrest.1,5-12 of exhalation. Normal exhalation during unassisted breathing
Prevalence of Auto-PEEP 1
Division of Pulmonary and Critical Care Medicine, Weill Cornell Medical
The prevalence of auto-PEEP is unknown and is likely more com- College, New York, NY, USA
mon than realized. There are no clear diagnostic criteria and the
Corresponding Author:
threshold for the magnitude of auto-PEEP that has significant car- David Berlin, Weill Cornell Medical College, 1300 York Avenue, New York,
diorespiratory effects varies with different clinical situations. NY 10023, USA
Even when auto-PEEP is the immediate cause of death, Email: berlind@med.cornell.edu
Sedation and, ultimately, neuromuscular blockade can also If shock from auto-PEEP persists, the clinician should
prevent the patient from triggering the ventilator and help consider adjunctive fluid management. If auto-PEEP impairs
reduce delivered minute ventilation.14,48 Other measures to venous return, fluid resuscitation may help restore the pressure
decrease minute ventilation such as avoiding over-feeding, gradient and recruit stroke volume via the Starling mechan-
correcting metabolic acidosis, and suppressing fever should ism.40 Fluids may be deleterious, however, if auto-PEEP con-
theoretically reduce auto-PEEP.14,45 To avoid dynamic hyper- tributes to right ventricular failure by increasing PVR.44
inflation during CPR, the clinician should provide 2 breaths for Diagnostic studies may help to determine whether or not fluid
every 30 compressions and 8 to 10 breaths/min after endotra- resuscitation will be beneficial. Unfortunately, significant auto-
cheal intubation.21,49,50 PEEP may impair accurate measurement of pulmonary artery
The clinician can directly set a longer expiratory time using occlusion pressure since airway pressure may exceed pulmon-
pressure-targeted time-cycled ventilation (pressure control). ary venous pressure.23 In general, static measures of cardiac
With ventilator modes that cycle to exhalation after delivering filling pressures are poor predictors of whether fluid resuscita-
a targeted volume, setting a higher peak inspiratory flow rate or tion will recruit stroke volume. Variability of stroke volume or
square wave inspiratory flow pattern will shorten inspiratory pulse pressure induced by the positive pressure respiratory
time. Reducing set tidal volume will also allow more time for cycle can predict the response to fluid resuscitation.59 The pres-
exhalation to reduce dynamic hyperinflation.14 However, ence of significant auto-PEEP, however, may affect the
reducing a tidal volume low enough to generate a rapid shallow results of these dynamic tests. The tests rely on the cyclic
breathing pattern can also produce auto-PEEP.51 On pressure- decline of left ventricular stroke volume following an
targeted flow-cycled ventilation (pressure support), increasing inspiratory decline in right-sided venous return. High
expiratory flow sensitivity will facilitate early cycling over to expiratory pressure, especially when accompanied by rela-
exhalation.52 Matching the auto-PEEP by increasing the set tively small tidal volumes, diminishes the stroke volume
extrinsic PEEP may decrease the patient work required to trig- variability during the cardiac cycle. High respiratory rates
ger the ventilator. Since this maneuver has little effect on the commonly associated with auto-PEEP also limit the respira-
total amount of PEEP, however, it has insignificant cardiovas- tory variability of stroke volume. Therefore, the false nega-
cular effects.53-55 There is a theoretical benefit of adding a low tive rate of the pulse pressure and stroke volume variability
level of extrinsic PEEP even in the presence of dynamic hyper- tests are likely higher in the presence of auto-PEEP.60,61
inflation if there is heterogeneous airflow obstruction in ARDS. Stroke volume variability with the respiratory cycle, more-
In this scenario, extrinsic PEEP can stent open slower emptying over, will not reliably predict fluid responsiveness in the
lung units to limit focal dynamic hyperinflation and improve presence of right ventricular failure.62 Auto-PEEP can pre-
impedance to right ventricular outflow.31 Switching to high fre- cipitate acute right heart failure and create false positive
quency jet ventilation appears to have no benefit for patients results of the dynamic tests. All tests of fluid responsiveness
with auto-PEEP.56 that assess the cyclical circulatory effects of passive positive
If the patient is in shock or cardiac arrest due to auto-PEEP, pressure ventilation, including inferior vena cava variability,
the clinician should consider a brief trial of apnea or temporar- are also invalid when there is a strong spontaneous ventila-
ily disconnect the endotracheal or tracheostomy tube from the tory drive.59
ventilator circuit to allow more complete exhalation.8,9,57 Auto-PEEP does not limit the echocardiographic prediction
Obviously, the fraction of inspired oxygen will fall during the of the right ventricular response to fluid resuscitation. If the
maneuver. However, there are multiple reports of full circula- right ventricle is markedly dilated with flattening of intraven-
tory arrest spontaneously reversing by detaching the patient tricular septum and dilation of the inferior vena cava, further
from the ventilator circuit. This has been called the ‘‘Lazarus fluid resuscitation will not improve stroke volume.43 Auto-
effect’’ and is thought to be a mechanism of ‘‘auto-resuscita- PEEP is also unlikely to affect the ability of a passive leg raise
tion.’’ In 1 case, a young woman with status asthmaticus had study to predict stroke volume response to a fluid bolus. Diag-
cardiac arrest each time the positive pressure ventilation was nostic passive leg raise has been validated in multiple settings
resumed and auto-resuscitated each time her physicians including high inflation pressures.63,64
detached her from the circuit to allow declaration of death.11 Exogenous catecholamines may be useful to support a circu-
If severe consequences of auto-PEEP develop, the clinician lation failing from auto-PEEP. Alpha-receptor-mediated veno-
may need to limit the patient’s minute ventilation and permit constriction may mobilize blood from venous capacitance
hypercapnia. This may be particularly important in cases of sta- vessels and increase cardiac output from augmented venous
tus asthmaticus.47 Since both hypercapnia and dynamic hyper- return. In the setting of right ventricular failure from auto-
inflation can raise PVR, the clinician must monitor the PEEP-induced elevation of PVR, catecholamine vasopressors
hemodynamic consequences of changing minute ventilation.58 can augment systemic arterial pressure to restore the perfusion
Even if the ventilator changes relieve auto-PEEP, the hemody- gradient to the right ventricle and augment its contractility.42,44
namic instability will persist if the extrinsic PEEP is set too There is currently no automated diagnostic testing that can
high. Moreover, a prolonged and high-pressure inspiratory reliably diagnose and manage significant auto-PEEP. Only the
phase may limit the hemodynamic benefit of reducing auto- presence of a trained clinician can mitigate the severe hemody-
PEEP. namic consequences of this common problem.
32. Motley HL, Werko L, Cournard A. Influence of intermittent positive 50. Travers AH, Rea TD, Bobrow BJ, et al. Part 4: CPR overview:
pressure breathing on cardiac output. Fed Proc. 1947;6(1):167-168. 2010 American Heart Association guidelines for cardiopulmon-
33. Buda A, Pinsky M, Ingels N, Daughters G, Stinson EB, Alderman ary resuscitation and emergency cardiovascular care. Circulation.
EL. Effect of intrathoracic pressure on left ventricular perfor- 2010;122(18 suppl 3):S676-S684.
mance. N Engl J Med. 1980;302(4):236-236. 51. Vieillard-Baron A, Prin S, Augarde R, et al. Increasing respiratory
34. Jardin F, Farcot JC, Boisante L, Curien N, Margairaz A, Bourdarias rate to improve CO2 clearance during mechanical ventilation is
JP. Influence of positive end-expiratory pressure on left ventricular not a panacea in acute respiratory failure. Crit Care Med. 2002;
performance. N Engl J Med. 1981;304(7):387-392. 30(7):1407-1412.
35. Pinsky MR. Heart-lung interactions. Curr Opinion Crit Care. 52. Hess DR. Ventilator waveforms and the physiology of pressure
2007;13(5):528-531. support ventilation. Respir Care. 2005;50(2):166-183.
36. Pinsky M. Effect of mechanical ventilation on heart-lung interac- 53. Brochard L. Intrinsic (or auto-) PEEP during controlled mechan-
tions. In: Tobin MF, ed Principles and Practice of Mechanical ical ventilation. Intens Care Med. 2002;28(10):1376-1378.
Ventilation. 2nd ed. New York, NY: McGraw Hill; 2006:729-757. 54. Tobin MJ, Lodato RF. PEEP, Auto-PEEP, and waterfalls. Chest.
37. Guyton AC, Jones CE, Coleman TG. Cardiac Output and Its Reg- 1989;96(3):449-451.
ulation. 2nd ed. Philadelphia, PA: Saunders; 1973. 55. Baigorri F, Demonte A, Blanch L, et al. Hemodynamic
38. Vieillard-Baron A, Chergui K, Rabiller A, et al. Superior vena responses to external counterbalancing of auto-positive end
caval collapsibility as a gauge of volume status in ventilated expiratory pressure in mechanically ventilated patients with
septic patients. Intens Care Med. 2004;30(9):1734-1739. chronic obstructive pulmonary-disease. Crit Care Med. 1994;
39. Jellinek H, Krafft P, Fitzgerald RD, Schwarz S, Pinsky MR. Right 22(11):1782-1791.
atrial pressure predicts hemodynamic response to apneic positive 56. Conti G, Bufi M, Rocco M, et al. Auto-PEEP and dynamic hyper-
airway pressure. Crit Care Med. 2000;28(3):672-678. inflation in COPD patients during controlled mechanical ventila-
40. van den Berg PCM, Jansen JRC, Pinsky MR. Effect of positive tion and high frequency jet ventilation. Intens Care Med. 1990;
pressure on venous return in volume-loaded cardiac surgical 16(2):81-84.
patients. J Appl Physiol. 2002;92(3):1223-1231. 57. Tobin MJ, Charles AG, Patrick FJ. Fighting the Ventilator. In:
41. Schmitt JM, Vieillard-Baron A, Augarde R, Prin S, Page B, Jardin Tobin MJ, ed Principles and Practice of Mechanical Ventilation.
F. Positive end-expiratory pressure titration in acute respiratory 2nd ed. New York, NY: McGraw-Hill; 2006:1121-1136.
distress syndrome patients: impact on right ventricular outflow 58. Mekontso Dessap A, Charron C, Devaquet J, et al. Impact of acute
impedance evaluated by pulmonary artery Doppler flow velocity hypercapnia and augmented positive end-expiratory pressure on
measurements. Crit Care Med. 2001;29(6):1154-1158. right ventricle function in severe acute respiratory distress syn-
42. Guyton AC, Lindsey AW, Gilluly JJ. The limits of right ventricu- drome. Intens Care Med. 2009;35(11):1850-1858.
lar compensation following an acute increase in pulmonary circu- 59. Durairaj L, Schmidt GA. Fluid therapy in resuscitated sepsis: less
lation resistance. Circ Res. 1954;2(4):326-332. is more. Chest. 2008;133(1):252-263.
43. Vieillard-Baron A, Prin S, Chergui K, Dubourg O, Jardin F. Echo- 60. Huang CC, Fu JY, Hu HC, et al. Prediction of fluid responsive-
Doppler demonstration of acute cor pulmonale at the bedside in ness in acute respiratory distress syndrome patients ventilated
the medical intensive care unit. Am J Respir Crit Care Med. with low tidal volume and high positive end-expiratory pressure.
2002;166(10):1310-1319. Crit Care Med. 2008;36(10):2810-2816.
44. Mebazaa A, Karpati P, Renaud E, Algotsson L. Acute right ven- 61. De Backer D, Taccone FS, Holsten R, Ibrahimi F, Vincent JL.
tricular failure-from pathophysiology to new treatments. Intens Influence of respiratory rate on stroke volume variation in
Care Med. 2004;30(2):185-196. mechanically ventilated patients. Anesthesiology. 2009;110(5):
45. Stewart TE, Slutsky AS. Occult, occult auto-PEEP in status asth- 1092-1097.
maticus. Crit Care Med. 1996;24(3):379-380. 62. Mahjoub Y, Pila C, Friggeri A, et al. Assessing fluid
46. Leatherman JW, McArthur C, Shapiro RS. Effect of prolongation responsiveness in critically ill patients: false-positive pulse
of expiratory time on dynamic hyperinflation in mechanically pressure variation is detected by Doppler echocardiographic
ventilated patients with severe asthma. Crit Care Med. 2004; evaluation of the right ventricle. Crit Care Med. 2009;37(9):
32(7):1542-1545. 2570-2575.
47. Oddo M, Feihl F, Schaller MD, Perret C. Management of mechan- 63. Cavallaro F, Sandroni C, Marano C, et al. Diagnostic accuracy of
ical ventilation in acute severe asthma: practical aspects. Intens passive leg raising for prediction of fluid responsiveness in adults:
Care Med. 2006;32(4):501-510. systematic review and meta-analysis of clinical studies. Intens
48. Tobin MJ, Jubran A, Laghi F. Patient-ventilator interaction. Am J Care Med. 2010;36(9):1475-1483.
Respir Crit Care Med. 2001;163(5):1059-1063. 64. Fougeres E, Teboul JL, Richard C, Osman D, Chemla D, Monnet
49. Aufderheide TP, Sigurdsson G, Pirrallo RG, et al. X. Hemodynamic impact of a positive end-expiratory pressure
Hyperventilation-induced hypotension during cardiopulmonary setting in acute respiratory distress syndrome: importance of the
resuscitation. Circulation. 2004;109(16):1960-1965. volume status. Crit Care Med. 2010;38(3):802-807.