IS 1828 (Part 1) :2005

1s07500-1:1999

7
F;

Fi

F’ F True force

Figure 1 — Schematic diagram for the determination of reversibility

6.5 Assessment of the force indicator

6.5.1 Relative accuracy error

The relative accuracy error expressed as a percentage of the mean true force, ~, is given by the equation:

fi-~
q= ~x loo (lo)
F

For the particular case of the calibration being carried out with a constant true force, the relative accuracy error is
given by the equation:

j+– F
q= -X loo (11)
F

6.5.2 Relative repeatability error

The relative repeatability error, b, for each discrete force, is the difference between the highest and lowest
measured values with respect to the average. It is given by the equation:

b= ‘Max
I “Fmi”
F
Xloo (12)

For the particular case of the calibration carried out with a constant true force, the relative repeatability error is given
by the equation:

~max - ~mifl
b= Xloo (13)