E411 PM Checklist

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 2

cobas e411 Preventive Maintenance Check List

Customer Name : Type : Yearly


Serial No : Case ID :
PM Communication No
Date : FSE Name :
To Clean/ Check Remarks (if any) Clean/Check
Cover, Housing Not Done
Chassis Clean
Filter MD 6 Clean
S/R Probe Clean
Sipper probe Clean
Valve Body Not Done
DI water and waste tank Clean
S/R - LLD 2.51
Sipper - LLD Voltage (V) 3.49
Pressure sensor 3.1
Detection Unit
Incubator 27.96
Temperature (degree C)
Reagent disk 17.86
System Reagent 28.01
Gripper finger Clean
Spring for Gripper finger Clean
Light barriers Clean
Reagent disk, BarcodeReader window Clean
Incubator Clean
Module (Fans, PCB's, Power supplies) Clean
Mixer paddle Clean
Mixer speed Check
Mechanism check Not Applicable
To Replace Remarks (if any) Replace
Tube B for measuring cell Yes
Tube B for sipper Yes
Tube B for sipper syringe No
Pinch valve tubings Yes
Seal piece P for S/R syringe Yes
Seal piece S for sipper syringe Yes
Tube 510 Yes
Tube 465 Yes
Nozzle seal pipetter Yes
Nozzle seal sipper Yes
Packings for syringes Yes
Mixer belt Yes
Measuring Cell If Required, Depending on Instrument No
performance.
Spring for Gripper finger Yes
Checks Remarks (if any) Attach Print Out
System Volume Check (Mandatory) No
PMT-HV Adjustment (Mandatory) Yes
APC Check (Optional) No
Blank Cell Calibration (Mandatory) Yes
QC Check(Few Parameters)(Mandatory) Yes

Customer Name :
FSE Signature :
Seal & Signature :