::: Application Report

Determination of the
density of LPG with an
Anton Paar density meter
Relevant for:
Petroleum industry
Chemical industry

LPGs characterized by their density Convenient and safe LPG density The DMA HP can measure under
measurement using Anton Paar pressures from 0 to 700 bar. The DMA HPM
Liquefied petroleum gases (LPG) are by- can be used at pressures from 0 to 1400
products in petroleum refining and natural equipment
bar. Both instruments can be used for
gas processing. They are a mixture of The density or relative density of LPGs is density measurements at temperatures
various gaseous substances, such as measured using an external density from -10 to 200 °C and show high
propane, butane, and others, kept liquefied measuring cell DMA HP with a density repeatabilities up to 0.00001g/cm3.
by applying pressure and/or lowering the meter DMA 4100/4500/5000 as the
temperature. LPGs are used as fuel, evaluation unit or using an external density
starting material in the chemical industry, measuring cell DMA HPM with an mPDS Reliable data processing
propellant gas, refrigerant, and in other 2000V3 evaluation unit. The DMA HP/
similar applications. The evaluation unit DMA 4100, DMA 4500
DMA HPM and the sample container are
or DMA 5000 gives you the digital results in
connected via hoses, tubes and valves,
The density of LPGs is an important feature g/cm3 (true density), D20/20, D20/04
for quality control in their processing and making up a closed system that guarantees
(relative density), SG, and/or API values
transporting. ultimate safety conditions.
converted to 15 °C or 60 °F.
LPG density measured using the There are two options for measurement: The data can be transmitted via RS232
hydrometer method Either you apply helium pressure on the interface to IBM-compatible computers, or
filled in sample or you measure at the vapor printed out directly. Densities of other
According to ASTM D1657, DIN 51 757 and pressure of the sample. The measuring cell liquids can be measured according to
ISO 3993, the density of LPG is measured temperature is controlled by a Peltier ASTM D4052, D5002 and DIN 51 757.
by a pressure hydrometer. thermostat (DMA HP) or by external means
(DMA HPM).
A transparent pressure cylinder with a
pressure hydrometer inside is filled with Easy adjustment to the required
sample until the hydrometer floats freely. measuring conditions
The hydrometer reading and the To waste The adjustment is carried out using air/
temperature reading are used to calculate water or nitrogen/n-pentane at measuring
the density at a reference temperature (e.g. temperature and measuring pressure. Any
15 °C). The pressure in the measurement other pure organic liquid with known
cylinder is given by the vapor pressure of Gas density/temperature/pressure relationship
the liquid at measuring temperature. DMA HP/HPM can also be used. Density standards with
The method is restricted to pressures up to Sample inlet density certificates are commercially
14 bar. One measurement takes approx. 30 available for temperatures from 15 °C to
minutes. Repeatabilities of 0.001 g/cm3 50 °C and pressures up to 40 bar.
and reproducibilities of 0.003 g/cm3 can be Setup for filling samples using overpressure
achieved.
The LPG density results obtained from the
DMA HP/DMA HPM comply with the
Calculating LPG density from the requirements of ASTM D4052 and
sample composition DIN 51 757.
According to DIN 51 618 or ASTM D2598-
02, the density of LPG can be calculated Optimized density measurements
from the composition of the sample.
Commonly, the composition is determined Small sample volumes of 2 to 3 mL are
by gas chromatography (GC) using sufficient. One measurement takes only
standard method DIN 51 619. approx. 10 minutes.

Info_02B

www.anton-paar.com Page 1