Model 775 Warming Unit: Operator's Manual

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 12

Normothermia System

Model 775
Warming Unit
Operator’s Manual

1 3M™ Bair Hugger™ - Unidade de Aquecimento


Warming Unit Model 775 3M™ Bair Hugger™, Modelo 775
Operator’s Manual Manual do Usuário
2 areil de réchauffement = 3M™ Bair Hugger™
Bair Hugger™ 3M™, modèle 775 Μονάδα Θέρμανσης Μοντέλο 775
Manuel de l’utilisateur Εγχειρίδιο χειριστή
3 3M™ Bair Hugger™-Wärmeeinheit, H Model 775 aparatu do
Modell 775 terapii grzewczej 3M™ Bair Hugger™
Bedienungsanleitung Instrukcja obsługi
4 Unità riscaldante J 775-ös típusszámú
3M™ Bair Hugger™ modello 775 3M™ Bair Hugger™ melegítőegység
Manuale d’uso Felhasználói kézikönyv
5 Unidad de calentamiento F Ohřívací jednotka
modelo 775 Bair Hugger™ de 3M™ 3M™ Bair Hugger™, model 775
Manual del operador Návod k obsluze
6 3M™ Bair Hugger™-verwarmingsunit Q „3M™ Bair Hugger™“
model 775 775 modelio šildomasis įtaisas
Bedieningshandleiding Naudotojo vadovas
7 3M™ Bair Hugger™ värmeenhet, M Устройство конвекционного типа
modell 77521 Bair Hugger для обогрева пациентов с
Användarhandbok принадлежностями, модель 775
Руководство по эксплуатации
8 3M™ Bair Hugger™ varmeenhed,
model 775 d 3M™ Bair Hugger™
Betjeningsvejledning Isıtma Ünitesi Model 775
9 3M™ Bair Hugger™ Varmeenhet, Kullanım Kılavuzu
modell 775 I .˳#BJS)VHHFS˳㘗⸈庛㐼
Brukerhåndbok  乼⡲䩛ⱃ
0 3M™ Bair Hugger™ U 3M™ ‫ ﻣﻦ‬Bair Hugger™ ‫ ﻟﻮﺣﺪة اﻟﺘﺪﻓﺌﺔ‬775 ‫اﻟﻄﺮاز‬
-lämpöpuhallinmalli 775 ‫دﻟﻴﻞ اﻟﻤﺸﻐﻞ‬
Käyttöopas
Normothermia System

Model 775
Warming Unit
Operator’s Manual

1 English........................................... 1 - Português .................................123

2 Français ...................................... 13 = Ελληνικά ..................................135

3 Deutsch ..................................... 27 H Polski ......................................... 147

4 Italiano .......................................39 J Magyar......................................159

5 Español ....................................... 51 F Česky .........................................171

6 Nederlands ................................63 Q Lietuvių Kalba ......................... 183

7 Svenska ...................................... 75 M Русский....................................195

8 Dansk ......................................... 87 d Türkçe ..................................... 209

9 Norsk..........................................99 I傈劥铃........................................221

0 Suomi ..........................................111 U ..................................231

4394547 5
Description: 3M TM Bair Hugger™ Item Spec#: 34-8718-8811-0
Patient Adjustable Warming Unit
Manual 08.02.16 New Build kmh
Supersedes#:
08.04.16 Updated Final pages number. Jer
Base File Name: 08.25.16 Alts per PDF file kmh
Structure#:
PA: 34871888110.indb 08.30.16 Final release. DJ
GA: Cat/Product#:Model 775 09.24.16 Adjust Center Margins to 1 in. to accomodate spiral binding and Final Release kmh
Software: InDesign CC 2014 - BOOK
SGS Contact: Reference:
Email:
Requester: Frank Boeshart/Debra Staack
Printer:
Printer Location: Die # / Doc. Size: 8.5" x 11"
Supplier:

PROCESS PMS
BLACK 280

PDF Scaled to 65%


1 Table of Contents
Introduction ................................................................................................................................................................................... 3
Description Of The 3M™ Bair Hugger™ Temperature Management System ......................................................... 3
Indications.............................................................................................................................................................................. 3
Definition Of Symbols ......................................................................................................................................................... 3
Contraindications, Warnings, Cautions And Notices ........................................................................................................... 3
Explanation Of Signal Word Consequences ................................................................................................................... 3

CONTRAINDICATION: ...................................................................................................................................................... 4
WARNING: ............................................................................................................................................................................ 4
CAUTION: ............................................................................................................................................................................. 4
Notices ................................................................................................................................................................................... 5
Proper Use And Maintenance ............................................................................................................................................ 5
Read Before Servicing Equipment .................................................................................................................................... 5
Overview And Operation............................................................................................................................................................ 5
Warming Unit Power-On-Reset ......................................................................................................................................... 6
Overview Of Control Panel ................................................................................................................................................ 6
Mounting The Warming Unit To An IV Pole .....................................................................................................................7
Instructions For Use ......................................................................................................................................................................7
Viewing The Temperature Mode Timer............................................................................................................................7
What To Do In Case Of An Over-Temperature Condition ............................................................................................7
What To Do If A Fault Condition Occurs ..........................................................................................................................7
General Maintenance And Storage .......................................................................................................................................... 8
Cleaning Instructions........................................................................................................................................................... 8
Storage ................................................................................................................................................................................... 8
Technical Service And Order Placement ................................................................................................................................ 8
When You Call For Technical Support ............................................................................................................................. 8
In-Warranty Repair And Exchange ................................................................................................................................... 8
Specifications........................................................................................................................................................................ 8

GB / 34-8718-8811-0
1
1 3M™ Bair Hugger™ Warming Unit Model 775

Operator’s Manual
Introduction
Description of the 3M™ Bair Hugger™ Temperature Voltage, alternating current (AC)
Management System
The Bair Hugger temperature management system consists of a
Model 775 warming unit (with optional rolling stand, and other This system is subject to European WEEE
accessories) along with a 3M warming blanket, Bair Hugger warming Directive 2002/96/EC. This product contains
gown, or the 3M™ Ranger™ 241 blood/fluid warming set. electrical and electronic components and
The Bair Hugger warming unit is attached to the warming blanket or must not be disposed of using standard refuse
warming gown by means of a flexible hose. Warm air is generated collection. Please consult local directives for
in the warming unit and flows through the warming unit hose and disposal of electrical and electronic equipment.
into the blanket or gown. Depending on the model, the 3M blanket
or gown is placed either around, over, or underneath the patient.
Small perforations on the blanket or gown allow the warm air to be Protective earth ground
dispersed over the patient. For blood/fluid warming applications, the
Model 241 blood/fluid warming set is inserted in the warming unit
hose. When the warming unit is turned ON and a temperature setting
is selected, warm air flows over the Model 241 tubing and warmed No free hosing
blood/fluid exits from the distal end of the tubing. For additional
information on 3M blankets, gowns, the 241 warming set, or other
accessories visit us online at bairhugger.com. High fan setting
This manual includes operating instructions and warming unit
specifications for the Model 775 warming unit. You can use the
Model 775 warming unit in all clinical settings where the patient may Low fan setting
feel too warm or too cold; including the operating room to provide
patient temperature management. For information about using
3M blankets, gowns or the 241 blood/fluid warming set with the
Non-explosion proof
Model 775 warming unit, refer to the respective Instructions for Use.

Indications
The Bair Hugger temperature management system is intended Consult instructions for use
to prevent and treat hypothermia. In addition, the temperature
management system can be used to provide patient thermal comfort
when conditions exist that may cause patients to feel too warm or too
Follow instructions for use
cold. The temperature management system can be used with adult
and pediatric patients.
• The Bair Hugger temperature management system should Recycle to avoid environmental contamination.
only be used by trained medical professionals. This product contains recyclable parts. For
• Federal law (USA) restricts this device to sale by or on the information on recycling, please contact your
order of a licensed healthcare professional. nearest 3M Service Center for advice.

Definition of Symbols
The following symbols may appear on the product’s labeling or
exterior packaging. Attention; see instructions for use

An equipotentiality plug (grounded) conductor Keep dry


other than a protective earth conductor or a
neutral conductor, providing a direct connection
between electrical equipment and the potential
equalization busbar of the electrical installation. Temperature limit
Please consult IEC 60601-1 for requirements.

Fuse Contraindications, Warnings, Cautions and Notices


Explanation of Signal Word Consequences

Type BF equipment (patient applied) ! WARNING: Indicates a hazardous situation which, if not


avoided, could result in death or serious injury.

! CAUTION: Indicates a hazardous situation which, if not


avoided, could result in minor or moderate injury.
Date of manufacture
NOTICE: Indicates a situation which, if not avoided, could result in
property damage only.
Manufacturer

GB / 34-8718-8811-0
3
! CONTRAINDICATION: To reduce the risk of thermal injury: ! WARNING: To reduce the risk of injury due to patient falls: • Monitor cutaneous responses of patients who are incapable
of reacting, communicating and/or who cannot sense
Overview and Operation
• Do not apply heat to lower extremities during aortic • Do not use a warming blanket/gown to transfer or move
cross-clamping. Thermal injury may occur if heat is applied the patient. temperature a minimum of every 15 minutes or according to Carrying
institutional protocol. handle Alphanumeric
!
to ischemic limbs.
WARNING: To reduce the risks associated with hazardous display
!
• Adjust air temperature or discontinue therapy when the
WARNING: To reduce the risk of thermal injury: voltage and fire: therapeutic goal is reached, if elevated temperatures are
• The Bair Hugger Model 775 warming unit has been • Keep power cord visible and accessible at all times. The recorded or if there is an adverse cutaneous response in the
designed to operate safely ONLY with 3M disposable plug on the power cord serves as the disconnect device. warmed area.
warming products. Use with other products may cause • Do not place the warming unit on a soft uneven surface,
• Only connect to outlets marked “Hospital only,” “Hospital
thermal injury. To the full extent permitted by law, the such as a bed, or a visibly wet surface as the air intake may
Grade,” or a reliable grounded outlet.
manufacturer and/or importer declines all responsibility for become blocked and cause the warming unit to overheat,
thermal injury resulting from the warming unit being used in • Use only the power cord specified for this product and compromising the warming unit’s performance.
conjunction with non-3M products. certified for the country of use. • Perform all temperature testing of the warming unit with a
• Do not treat patients with the warming unit hose • Do not allow the power cord to get wet. 3M Model 22110 temperature test unit.
alone. Always attach the hose to a 3M warming
blanket/gown before providing warming therapy.
• Do not use the warming unit when it appears the warming
unit, power cord or any component is damaged. Replace
! CAUTION: To reduce the risk associated with
electromagnetic interference (EMI) due to portable and
Control Panel

• Do not allow the patient to lie on the warming unit hose. the warming unit. Contact 3M Technical Support. mobile radio frequency (RF) communications equipment: Figure A. Model 775 Warming Unit – front view
• Do not allow the warming hose to directly contact the • Do not disassemble the warming unit unless you are a • The 3M Bair Hugger temperature management system has
patient’s skin during warming therapy. qualified service technician. There are electrically live been tested to be resistant to both EMI and electrostatic
parts within the warming unit when it is connected to a discharge (ESD).
• Do not leave neonates, infants, children and other
power source. • Install and put into service the 3M Bair Hugger temperature IV pole clamp
vulnerable patient populations unattended during
• Connect each warming unit being tested to a separate management system according to the electromagnetic
warming therapy.
power source. compatibility (EMC) information provided in the Guidance
• Do not leave patients with poor perfusion unmonitored and Manufacturer’s Declaration.
during prolonged warming therapy.
! CAUTION: To reduce the risk of cross-contamination: • Should interference occur, move away from the portable or Hose
• Do not place the non-perforated side of the warming • Except for specific Bair Hugger warming blanket models, mobile RF communications equipment.
blanket/gown on the patient. Always place the perforated 3M warming blankets/gowns are not sterile. Each warming Notices
side (with the small holes) directly on top of the patient in blanket/gown is intended for single patient use ONLY.
contact with the patient’s skin. Placing a sheet between the warming blanket/gown and 1. The Bair Hugger warming unit meets medical electronic
the patient does not prevent contamination of the product. interference requirements. If radio frequency interference with
• Do not connect a torn or damaged warming blanket/gown other equipment should occur, connect the warming unit to a
to the warming unit. • Clean the warming unit and the outside of the warming unit different power source.
• For over-body warming blankets and gowns; do not place hose after each patient use. See “Cleaning Instructions” on
page 8. 2. To avoid warming unit damage:
patient securement device (i.e. safety strap or tape) over
the warming blanket/gown. • Follow applicable regulations when disposing of this • Use proper Electrostatic Discharge (ESD) procedures when
warming unit or any of its electrical components. performing maintenance.
• For underbody or side channel warming blankets; if a
securement device (i.e. safety strap, tape) is used, ensure • Do not modify this equipment without authorization from
• Do not attempt to clean the air filter as it may be
the warming channels are not occluded. the manufacturer. Power cord Hose connection
contaminated from use. Discard the filter in a manner
• Do not immerse the warming unit, warming unit parts connection
• Do not place the warming blanket/gown directly over a consistent with institutional protocol.
or accessories in any liquid or subject them to any
dispersive electrode pad. • Do not operate the Bair Hugger warming unit with the Figure B. Model 775 Warming Unit – rear view
sterilization process.
• Do not continue warming therapy if the red Over-temp hose detached from the 3M warming blanket/gown. The
• Do not use cleaning solutions with greater than 80%
indicator light illuminates and the alarm sounds. Unplug the Bair Hugger warming unit is intended to be used with Warming Unit Power-On-Reset
alcohol or solvents, including acetone and thinner, to clean
warming unit and contact a qualified service technician. the hose properly attached to its corresponding warming
the warming unit or hose. Solvents may damage the labels The Model 775 warming unit automatically performs the following
blanket/gown and in accordance with good practices for
• Do not continue 241 blood/fluid warming therapy if the and other plastic parts. power-on-reset sequence after you connect the warming unit to a
operating room sterile technique.
red Over-temp indicator light illuminates and the alarm properly grounded power source:
!
Proper Use and Maintenance
sounds. Immediately stop fluid flow, and discard the blood/ CAUTION: To reduce the risk of patient or caregiver injury: 3M assumes no responsibility for the reliability, performance, or • Performs all self-test functions.
fluid warming set. Unplug the warming unit, and contact a • If mounted on an IV pole, the distance from the bottom of safety of the warming unit if the following events occur:
qualified service technician. • Illuminates indicator lights and all pixels in the
the warming unit to the floor must be less than 44” (112 cm) alphanumeric display momentarily.
• Modifications or repairs are performed by
• For the Bair Hugger flex warming gown; ensure that the and the IV pole wheelbase diameter must be at least 28”
unqualified personnel. • Displays the text “MD 700’s” and the software revision
blood pressure cuff, ECG, IV or other lines or cords are (71 cm) to prevent tipping.
• The warming unit is used in a manner other than that level in the alphanumeric display.
not between the upper sleeve insert and the gown prior to
deploying the upper sleeve warming insert(s), as this could ! CAUTION: To reduce the risk of fire: described in the Operator’s or Service Manual.
• Sounds the alarm (three low-level clicks).
result in tearing of the insert during deployment. • 3M warming blankets and gowns are classified as Class I • The warming unit is installed in an environment
that does not meet the appropriate electrical and • Enters the Standby mode with the high fan setting selected
Normal Flammability as defined by the Consumer Product
• Do not perform the over-temperature detection system test by default.
Safety Commissions flammable fabric regulation, 16 CFR grounding requirements.
while the warming unit is being used for warming therapy.
1610. Follow standard safety protocols when using high • The warming unit not maintained in accordance with the Note: The low fan setting indicator light will not illuminate
! WARNING: To reduce the risk of patient injury or death due to intensity heat sources. procedures described in the Service Manual. during power up.
altered drug delivery:
• Do not use a warming blanket/gown over transdermal
! CAUTION: To reduce the risk of thermal injury, hyperthermia
or hypothermia:
Read Before Servicing Equipment
All repair, calibration and servicing of the warming unit require the
If the warming unit loses power for less than 1 second, the warming
unit software will return the warming unit to the operating mode
medication patches. you selected prior to the power loss. However, if the warming unit
• 3M recommends continuously monitoring core skill of a qualified, medical equipment service technician who is
loses power for longer than 1 second, the warming unit software will
! WARNING: To reduce the risk of injury due to interference
with ventilation:
temperature. In the absence of continuous monitoring,
monitor the temperature of patients who are incapable
familiar with good practice for medical device repair. If service does
not require the manufacturer’s attention, the Model 775 Service
reset when you restore power. The warming unit will then enter the
Standby mode with the high fan setting selected by default.
of reacting, communicating and/or who cannot sense Manual provides the technical information needed to service the
Do not allow the warming blanket/gown or head drape to warming unit. Perform all repairs and maintenance in accordance
temperature a minimum of every 15 minutes or according to
cover the patient’s head or airway when the patient is not with the instructions in the Service Manual. For additional service
institutional protocol.
mechanically ventilated. information please contact 3M technical support.

4 3M™Bair Hugger™ Model 775 Warming Unit – Operator’s Manual GB / 34-8718-8811-0


5
! CONTRAINDICATION: To reduce the risk of thermal injury: ! WARNING: To reduce the risk of injury due to patient falls: • Monitor cutaneous responses of patients who are incapable
of reacting, communicating and/or who cannot sense
Overview and Operation
• Do not apply heat to lower extremities during aortic • Do not use a warming blanket/gown to transfer or move
cross-clamping. Thermal injury may occur if heat is applied the patient. temperature a minimum of every 15 minutes or according to Carrying
institutional protocol. handle Alphanumeric
!
to ischemic limbs.
WARNING: To reduce the risks associated with hazardous display
!
• Adjust air temperature or discontinue therapy when the
WARNING: To reduce the risk of thermal injury: voltage and fire: therapeutic goal is reached, if elevated temperatures are
• The Bair Hugger Model 775 warming unit has been • Keep power cord visible and accessible at all times. The recorded or if there is an adverse cutaneous response in the
designed to operate safely ONLY with 3M disposable plug on the power cord serves as the disconnect device. warmed area.
warming products. Use with other products may cause • Do not place the warming unit on a soft uneven surface,
• Only connect to outlets marked “Hospital only,” “Hospital
thermal injury. To the full extent permitted by law, the such as a bed, or a visibly wet surface as the air intake may
Grade,” or a reliable grounded outlet.
manufacturer and/or importer declines all responsibility for become blocked and cause the warming unit to overheat,
thermal injury resulting from the warming unit being used in • Use only the power cord specified for this product and compromising the warming unit’s performance.
conjunction with non-3M products. certified for the country of use. • Perform all temperature testing of the warming unit with a
• Do not treat patients with the warming unit hose • Do not allow the power cord to get wet. 3M Model 22110 temperature test unit.
alone. Always attach the hose to a 3M warming
blanket/gown before providing warming therapy.
• Do not use the warming unit when it appears the warming
unit, power cord or any component is damaged. Replace
! CAUTION: To reduce the risk associated with
electromagnetic interference (EMI) due to portable and
Control Panel

• Do not allow the patient to lie on the warming unit hose. the warming unit. Contact 3M Technical Support. mobile radio frequency (RF) communications equipment: Figure A. Model 775 Warming Unit – front view
• Do not allow the warming hose to directly contact the • Do not disassemble the warming unit unless you are a • The 3M Bair Hugger temperature management system has
patient’s skin during warming therapy. qualified service technician. There are electrically live been tested to be resistant to both EMI and electrostatic
parts within the warming unit when it is connected to a discharge (ESD).
• Do not leave neonates, infants, children and other
power source. • Install and put into service the 3M Bair Hugger temperature IV pole clamp
vulnerable patient populations unattended during
• Connect each warming unit being tested to a separate management system according to the electromagnetic
warming therapy.
power source. compatibility (EMC) information provided in the Guidance
• Do not leave patients with poor perfusion unmonitored and Manufacturer’s Declaration.
during prolonged warming therapy.
! CAUTION: To reduce the risk of cross-contamination: • Should interference occur, move away from the portable or Hose
• Do not place the non-perforated side of the warming • Except for specific Bair Hugger warming blanket models, mobile RF communications equipment.
blanket/gown on the patient. Always place the perforated 3M warming blankets/gowns are not sterile. Each warming Notices
side (with the small holes) directly on top of the patient in blanket/gown is intended for single patient use ONLY.
contact with the patient’s skin. Placing a sheet between the warming blanket/gown and 1. The Bair Hugger warming unit meets medical electronic
the patient does not prevent contamination of the product. interference requirements. If radio frequency interference with
• Do not connect a torn or damaged warming blanket/gown other equipment should occur, connect the warming unit to a
to the warming unit. • Clean the warming unit and the outside of the warming unit different power source.
• For over-body warming blankets and gowns; do not place hose after each patient use. See “Cleaning Instructions” on
page 8. 2. To avoid warming unit damage:
patient securement device (i.e. safety strap or tape) over
the warming blanket/gown. • Follow applicable regulations when disposing of this • Use proper Electrostatic Discharge (ESD) procedures when
warming unit or any of its electrical components. performing maintenance.
• For underbody or side channel warming blankets; if a
securement device (i.e. safety strap, tape) is used, ensure • Do not modify this equipment without authorization from
• Do not attempt to clean the air filter as it may be
the warming channels are not occluded. the manufacturer. Power cord Hose connection
contaminated from use. Discard the filter in a manner
• Do not immerse the warming unit, warming unit parts connection
• Do not place the warming blanket/gown directly over a consistent with institutional protocol.
or accessories in any liquid or subject them to any
dispersive electrode pad. • Do not operate the Bair Hugger warming unit with the Figure B. Model 775 Warming Unit – rear view
sterilization process.
• Do not continue warming therapy if the red Over-temp hose detached from the 3M warming blanket/gown. The
• Do not use cleaning solutions with greater than 80%
indicator light illuminates and the alarm sounds. Unplug the Bair Hugger warming unit is intended to be used with Warming Unit Power-On-Reset
alcohol or solvents, including acetone and thinner, to clean
warming unit and contact a qualified service technician. the hose properly attached to its corresponding warming
the warming unit or hose. Solvents may damage the labels The Model 775 warming unit automatically performs the following
blanket/gown and in accordance with good practices for
• Do not continue 241 blood/fluid warming therapy if the and other plastic parts. power-on-reset sequence after you connect the warming unit to a
operating room sterile technique.
red Over-temp indicator light illuminates and the alarm properly grounded power source:
!
Proper Use and Maintenance
sounds. Immediately stop fluid flow, and discard the blood/ CAUTION: To reduce the risk of patient or caregiver injury: 3M assumes no responsibility for the reliability, performance, or • Performs all self-test functions.
fluid warming set. Unplug the warming unit, and contact a • If mounted on an IV pole, the distance from the bottom of safety of the warming unit if the following events occur:
qualified service technician. • Illuminates indicator lights and all pixels in the
the warming unit to the floor must be less than 44” (112 cm) alphanumeric display momentarily.
• Modifications or repairs are performed by
• For the Bair Hugger flex warming gown; ensure that the and the IV pole wheelbase diameter must be at least 28”
unqualified personnel. • Displays the text “MD 700’s” and the software revision
blood pressure cuff, ECG, IV or other lines or cords are (71 cm) to prevent tipping.
• The warming unit is used in a manner other than that level in the alphanumeric display.
not between the upper sleeve insert and the gown prior to
deploying the upper sleeve warming insert(s), as this could ! CAUTION: To reduce the risk of fire: described in the Operator’s or Service Manual.
• Sounds the alarm (three low-level clicks).
result in tearing of the insert during deployment. • 3M warming blankets and gowns are classified as Class I • The warming unit is installed in an environment
that does not meet the appropriate electrical and • Enters the Standby mode with the high fan setting selected
Normal Flammability as defined by the Consumer Product
• Do not perform the over-temperature detection system test by default.
Safety Commissions flammable fabric regulation, 16 CFR grounding requirements.
while the warming unit is being used for warming therapy.
1610. Follow standard safety protocols when using high • The warming unit not maintained in accordance with the Note: The low fan setting indicator light will not illuminate
! WARNING: To reduce the risk of patient injury or death due to intensity heat sources. procedures described in the Service Manual. during power up.
altered drug delivery:
• Do not use a warming blanket/gown over transdermal
! CAUTION: To reduce the risk of thermal injury, hyperthermia
or hypothermia:
Read Before Servicing Equipment
All repair, calibration and servicing of the warming unit require the
If the warming unit loses power for less than 1 second, the warming
unit software will return the warming unit to the operating mode
medication patches. you selected prior to the power loss. However, if the warming unit
• 3M recommends continuously monitoring core skill of a qualified, medical equipment service technician who is
loses power for longer than 1 second, the warming unit software will
! WARNING: To reduce the risk of injury due to interference
with ventilation:
temperature. In the absence of continuous monitoring,
monitor the temperature of patients who are incapable
familiar with good practice for medical device repair. If service does
not require the manufacturer’s attention, the Model 775 Service
reset when you restore power. The warming unit will then enter the
Standby mode with the high fan setting selected by default.
of reacting, communicating and/or who cannot sense Manual provides the technical information needed to service the
Do not allow the warming blanket/gown or head drape to warming unit. Perform all repairs and maintenance in accordance
temperature a minimum of every 15 minutes or according to
cover the patient’s head or airway when the patient is not with the instructions in the Service Manual. For additional service
institutional protocol.
mechanically ventilated. information please contact 3M technical support.

4 3M™Bair Hugger™ Model 775 Warming Unit – Operator’s Manual GB / 34-8718-8811-0


5
Overview of Control Panel TEMP IN RANGE INDICATOR LIGHT INSTRUCTIONS FOR USE 5. Press the Standby button when warming therapy is complete.
The Temp in Range indicator light illuminates when the temperature at The instructions below describe how to operate the Model Disconnect the warming unit hose from the warming blanket/
Alphanumeric Display
Temperature the blanket or gown end of the hose is ±1.5°C of the selected setting; 775 warming unit. For information about using 3M blankets, gowns, gown and discard the blanket/gown per hospital policy.
Modes this indicator light does not illuminate in the ambient mode. or the 241 blood/fluid warming set with the Model 775 warming unit, 6. Unplug the warming unit to disconnect it from the
refer to the respective Instructions for Use. power source.
TEMPERATURE
High Fan FAULT INDICATOR LIGHT
HIGH
Setting When a system fault occurs, the amber Fault indicator light flashes 1. If the Model 775 warming unit is not mounted on an IV pole or Note: At any time, the fan speed setting can be switched between
MODEL 775 43 C AIR FLOW
and an audible alarm sounds. Please refer to What to Do in Case of a Bair Hugger rolling stand, place the warming unit on a flat, hard, high fan setting ( ) and low fan setting ( ) by pressing the
Low Fan Fault Condition on page 7 for additional information. corresponding button. However, these buttons will not place the
MEDIUM
Setting dry surface, such as a table, before beginning warming therapy.
38 C warming unit into Standby.

LOW
UNDER-TEMP FAULT
If the warming unit senses an under-temperature condition, the
! Caution: Do not place the warming unit on a soft uneven
surface, such as a bed, or a visibly wet surface as the air
Viewing the Temperature Mode Timer
32 C The temperature mode timer records the amount of time the warming
amber Fault indicator light flashes and an audible alarm sounds. intake may become blocked and cause the warming unit to
Please refer to What to Do in Case of a Fault Condition on page 7 for unit has been operating in the selected temperature mode. The timer
overheat, compromising the warming unit’s performance.
System
TEMP IN RANGE
additional information. resets itself every time you select a different temperature mode.
AMBIENT FAULT
2. Insert the end of the Bair Hugger warming unit hose in the To view the temperature mode timer, press the button and hold for
OVER-TEMP INDICATOR LIGHT warming blanket or gown hose port. Use a twisting motion to 3 seconds for the current temperature mode. The alphanumeric
OVER TEMP
If the warming unit senses an over-temperature condition, the red ensure a snug fit. A visual marker is located around the display will show the temperature mode time, then go back to
STANDBY Over-temp indicator light flashes and an alarm sounds. Please refer to mid-section of the hose end to guide the depth of hose showing the temperature mode setting.
What to Do in Case of an Over-Temperature Condition on page 7  for insertion. (Figures E). Support the hose to ensure
additional information. secure attachment. What to Do in Case of an Over-Temperature Condition

!
If an over-temperature condition occurs, the red Over-temp indicator
Mounting the Warming Unit to an IV Pole Warning: Do not treat patients with the Bair Hugger light flashes and an audible alarm sounds and the warming unit
Figure C. Model 775 Warming Unit Control Panel
To prevent tipping, clamp the Model 775 warming unit to an IV pole warming unit hose alone. Always attach the hose to a automatically turns OFF the heater, blower, and operating mode
at a height that provides stability. Clamp the warming unit no higher Bair Hugger warming blanket/gown before providing indicator lights. A fault code appears in the alphanumeric display.
ALPHANUMERIC DISPLAY than 44” (112 cm) from the floor on an IV pole with a minimum 28” warming therapy.
Note: Pressing any button will silence the alarm.
The alphanumeric display shows the temperature at the end of the (71 cm) diameter wheelbase.
Hose
!
Bair Hugger warming unit hose in degrees Celsius. IF AN OVER-TEMPERATURE CONDITION OCCURS:
!
Caution: If mounted on an IV pole, the distance from the
TEMPERATURE MODES bottom of the warming unit to the floor must be less than 44” Warning: Do not continue warming therapy if the red
• Press the 32˚C (Low), 38˚C (Med), or 43˚C (High) button to (112 cm) and the IV pole wheelbase diameter must be at least Over-temp indicator light illuminates and the alarm
select the desired temperature. 28” (71 cm) to prevent tipping. sounds. Unplug the warming unit and contact a qualified
service technician.
• Press the Ambient button to supply room temperature air. METHOD
When you select a temperature mode, the following events occur: To mount the Model 775 warming unit to an IV pole, simply ! Warning: Do not continue 241 blood/fluid warming therapy
if the red Over-temp indicator light illuminates and the alarm
turn the clamp handle clockwise to tighten the clamp on
• Corresponding temperature indicator and fan setting the IV pole and counterclockwise to release it. Ensure sounds. Immediately stop fluid flow, and discard the blood/fluid
lights illuminate. Insert
the IV pole bumper rests against the pole. See Figure D. marker Hose port warming set. Unplug the warming unit, and contact a qualified
• Blower operates at selected fan speed. service technician.
Clamp 1. Discontinue all warming therapy. If you are using the 241 blood/
• Temperature at the end of the hose appears in the Handle Figure E.
alphanumeric display. fluid warming set, immediately stop fluid flow and discard the
blood/fluid warming set.
• Heater activates except in ambient mode. 3. Connect the warming unit to a properly grounded power
• Temperature mode timer and hour meter activate. source. The warming unit will be in Standby mode and 2. Press any button on the control panel to silence the alarm.
the Standby indicator light will illuminate. The high fan
• Temp in range indicator light illuminates when the 3. Unplug the warming unit.
setting ( ) will be pre-selected by default with its indicator
temperature at the blanket end of the hose is within +/- light illuminated. The low fan setting ( ) may be pre-selected 4. Replace warming unit. Do not attempt to restart or reset the
1.5°C of the selected setting; this indicator light does not prior to selecting the desired temperature mode. warming unit.
illuminate in the ambient mode.
4. Press the appropriate button (i.e., 32˚C, 38˚C, 43˚C, or 5. Contact a qualified service technician.
AIRFLOW MODES Ambient) to select the desired temperature. This will turn What to Do If a Fault Condition occurs
The Model 775 warming unit has two preset fan speed settings: a the blower and heater ON. When the warming unit reaches
The Model 775 warming unit software recognizes several
default or high fan setting ( ) and a reduced or low fan setting ( ). the selected temperature, the Temp in Range indicator light
non-hazardous conditions and reports those conditions as faults.
The Model 775 warming unit will reset/default to the high fan setting IV Pole will illuminate; this indicator light does not illuminate in the
When a system fault occurs, the software stores the fault code
each time the warming unit is powered up or Standby is selected. 44 in. (112 cm) Bumper Ambient mode.
in memory, the amber Fault indicator light flashes, and an alarm
The low fan setting ( ) may be pre-selected while in Standby prior to
selecting the desired temperature.
maximum
! Caution: Patient Monitoring Recommendations: sounds. The warming unit automatically turns the heater, blower,
and operating mode indicator lights OFF. The control panel will not
• 3M recommends continuously monitoring core
STANDBY respond to commands.
temperature. In the absence of continuous monitoring,
Upon power up, the warming unit defaults to Standby and the high monitor the temperature of patients who are incapable Note: Pressing any button will silence the alarm.
fan setting. When the warming unit is in 32˚C (Low), 38˚C (Med), or of reacting, communicating and/or who cannot sense
43˚C (High) or Ambient mode, press the Standby button to place the IF A FAULT CONDITION OCCURS:
temperature a minimum of every 15 minutes or according to
warming unit in Standby mode. When you select the Standby mode, institutional protocol. 1. Unplug the warming unit and wait for 30-60 seconds.
the following events occur:
• Monitor cutaneous responses of patients who are incapable 2. Reconnect the warming unit to a grounded power source. The
• Standby indicator light illuminates. of reacting, communicating and/or who cannot sense warming unit will perform the normal power-on-reset sequence
28 in.
• Blower and heater are turned OFF. (71 cm) minimum temperature a minimum of every 15 minutes or according to and then enter the Standby mode.
institutional protocol.
• Alphanumeric display deactivates. 3. Reselect the temperature setting.
Figure D. Model 775 warming unit on an IV Pole • Adjust air temperature or discontinue therapy when the
• Alarm and fault detection functions remain active. 4. If the warming unit does not return to normal operation, contact
therapeutic goal is reached, if elevated temperatures are
• Operating timer pauses. a qualified service technician.
recorded or if there is an adverse cutaneous response in the
• Airflow mode resets/defaults to the high fan setting. warmed area.

6 3M™Bair Hugger™ Model 775 Warming Unit – Operator’s Manual GB / 34-8718-8811-0


7
Overview of Control Panel TEMP IN RANGE INDICATOR LIGHT INSTRUCTIONS FOR USE 5. Press the Standby button when warming therapy is complete.
The Temp in Range indicator light illuminates when the temperature at The instructions below describe how to operate the Model Disconnect the warming unit hose from the warming blanket/
Alphanumeric Display
Temperature the blanket or gown end of the hose is ±1.5°C of the selected setting; 775 warming unit. For information about using 3M blankets, gowns, gown and discard the blanket/gown per hospital policy.
Modes this indicator light does not illuminate in the ambient mode. or the 241 blood/fluid warming set with the Model 775 warming unit, 6. Unplug the warming unit to disconnect it from the
refer to the respective Instructions for Use. power source.
TEMPERATURE
High Fan FAULT INDICATOR LIGHT
HIGH
Setting When a system fault occurs, the amber Fault indicator light flashes 1. If the Model 775 warming unit is not mounted on an IV pole or Note: At any time, the fan speed setting can be switched between
MODEL 775 43 C AIR FLOW
and an audible alarm sounds. Please refer to What to Do in Case of a Bair Hugger rolling stand, place the warming unit on a flat, hard, high fan setting ( ) and low fan setting ( ) by pressing the
Low Fan Fault Condition on page 7 for additional information. corresponding button. However, these buttons will not place the
MEDIUM
Setting dry surface, such as a table, before beginning warming therapy.
38 C warming unit into Standby.

LOW
UNDER-TEMP FAULT
If the warming unit senses an under-temperature condition, the
! Caution: Do not place the warming unit on a soft uneven
surface, such as a bed, or a visibly wet surface as the air
Viewing the Temperature Mode Timer
32 C The temperature mode timer records the amount of time the warming
amber Fault indicator light flashes and an audible alarm sounds. intake may become blocked and cause the warming unit to
Please refer to What to Do in Case of a Fault Condition on page 7 for unit has been operating in the selected temperature mode. The timer
overheat, compromising the warming unit’s performance.
System
TEMP IN RANGE
additional information. resets itself every time you select a different temperature mode.
AMBIENT FAULT
2. Insert the end of the Bair Hugger warming unit hose in the To view the temperature mode timer, press the button and hold for
OVER-TEMP INDICATOR LIGHT warming blanket or gown hose port. Use a twisting motion to 3 seconds for the current temperature mode. The alphanumeric
OVER TEMP
If the warming unit senses an over-temperature condition, the red ensure a snug fit. A visual marker is located around the display will show the temperature mode time, then go back to
STANDBY Over-temp indicator light flashes and an alarm sounds. Please refer to mid-section of the hose end to guide the depth of hose showing the temperature mode setting.
What to Do in Case of an Over-Temperature Condition on page 7  for insertion. (Figures E). Support the hose to ensure
additional information. secure attachment. What to Do in Case of an Over-Temperature Condition

!
If an over-temperature condition occurs, the red Over-temp indicator
Mounting the Warming Unit to an IV Pole Warning: Do not treat patients with the Bair Hugger light flashes and an audible alarm sounds and the warming unit
Figure C. Model 775 Warming Unit Control Panel
To prevent tipping, clamp the Model 775 warming unit to an IV pole warming unit hose alone. Always attach the hose to a automatically turns OFF the heater, blower, and operating mode
at a height that provides stability. Clamp the warming unit no higher Bair Hugger warming blanket/gown before providing indicator lights. A fault code appears in the alphanumeric display.
ALPHANUMERIC DISPLAY than 44” (112 cm) from the floor on an IV pole with a minimum 28” warming therapy.
Note: Pressing any button will silence the alarm.
The alphanumeric display shows the temperature at the end of the (71 cm) diameter wheelbase.
Hose
!
Bair Hugger warming unit hose in degrees Celsius. IF AN OVER-TEMPERATURE CONDITION OCCURS:
!
Caution: If mounted on an IV pole, the distance from the
TEMPERATURE MODES bottom of the warming unit to the floor must be less than 44” Warning: Do not continue warming therapy if the red
• Press the 32˚C (Low), 38˚C (Med), or 43˚C (High) button to (112 cm) and the IV pole wheelbase diameter must be at least Over-temp indicator light illuminates and the alarm
select the desired temperature. 28” (71 cm) to prevent tipping. sounds. Unplug the warming unit and contact a qualified
service technician.
• Press the Ambient button to supply room temperature air. METHOD
When you select a temperature mode, the following events occur: To mount the Model 775 warming unit to an IV pole, simply ! Warning: Do not continue 241 blood/fluid warming therapy
if the red Over-temp indicator light illuminates and the alarm
turn the clamp handle clockwise to tighten the clamp on
• Corresponding temperature indicator and fan setting the IV pole and counterclockwise to release it. Ensure sounds. Immediately stop fluid flow, and discard the blood/fluid
lights illuminate. Insert
the IV pole bumper rests against the pole. See Figure D. marker Hose port warming set. Unplug the warming unit, and contact a qualified
• Blower operates at selected fan speed. service technician.
Clamp 1. Discontinue all warming therapy. If you are using the 241 blood/
• Temperature at the end of the hose appears in the Handle Figure E.
alphanumeric display. fluid warming set, immediately stop fluid flow and discard the
blood/fluid warming set.
• Heater activates except in ambient mode. 3. Connect the warming unit to a properly grounded power
• Temperature mode timer and hour meter activate. source. The warming unit will be in Standby mode and 2. Press any button on the control panel to silence the alarm.
the Standby indicator light will illuminate. The high fan
• Temp in range indicator light illuminates when the 3. Unplug the warming unit.
setting ( ) will be pre-selected by default with its indicator
temperature at the blanket end of the hose is within +/- light illuminated. The low fan setting ( ) may be pre-selected 4. Replace warming unit. Do not attempt to restart or reset the
1.5°C of the selected setting; this indicator light does not prior to selecting the desired temperature mode. warming unit.
illuminate in the ambient mode.
4. Press the appropriate button (i.e., 32˚C, 38˚C, 43˚C, or 5. Contact a qualified service technician.
AIRFLOW MODES Ambient) to select the desired temperature. This will turn What to Do If a Fault Condition occurs
The Model 775 warming unit has two preset fan speed settings: a the blower and heater ON. When the warming unit reaches
The Model 775 warming unit software recognizes several
default or high fan setting ( ) and a reduced or low fan setting ( ). the selected temperature, the Temp in Range indicator light
non-hazardous conditions and reports those conditions as faults.
The Model 775 warming unit will reset/default to the high fan setting IV Pole will illuminate; this indicator light does not illuminate in the
When a system fault occurs, the software stores the fault code
each time the warming unit is powered up or Standby is selected. 44 in. (112 cm) Bumper Ambient mode.
in memory, the amber Fault indicator light flashes, and an alarm
The low fan setting ( ) may be pre-selected while in Standby prior to
selecting the desired temperature.
maximum
! Caution: Patient Monitoring Recommendations: sounds. The warming unit automatically turns the heater, blower,
and operating mode indicator lights OFF. The control panel will not
• 3M recommends continuously monitoring core
STANDBY respond to commands.
temperature. In the absence of continuous monitoring,
Upon power up, the warming unit defaults to Standby and the high monitor the temperature of patients who are incapable Note: Pressing any button will silence the alarm.
fan setting. When the warming unit is in 32˚C (Low), 38˚C (Med), or of reacting, communicating and/or who cannot sense
43˚C (High) or Ambient mode, press the Standby button to place the IF A FAULT CONDITION OCCURS:
temperature a minimum of every 15 minutes or according to
warming unit in Standby mode. When you select the Standby mode, institutional protocol. 1. Unplug the warming unit and wait for 30-60 seconds.
the following events occur:
• Monitor cutaneous responses of patients who are incapable 2. Reconnect the warming unit to a grounded power source. The
• Standby indicator light illuminates. of reacting, communicating and/or who cannot sense warming unit will perform the normal power-on-reset sequence
28 in.
• Blower and heater are turned OFF. (71 cm) minimum temperature a minimum of every 15 minutes or according to and then enter the Standby mode.
institutional protocol.
• Alphanumeric display deactivates. 3. Reselect the temperature setting.
Figure D. Model 775 warming unit on an IV Pole • Adjust air temperature or discontinue therapy when the
• Alarm and fault detection functions remain active. 4. If the warming unit does not return to normal operation, contact
therapeutic goal is reached, if elevated temperatures are
• Operating timer pauses. a qualified service technician.
recorded or if there is an adverse cutaneous response in the
• Airflow mode resets/defaults to the high fan setting. warmed area.

6 3M™Bair Hugger™ Model 775 Warming Unit – Operator’s Manual GB / 34-8718-8811-0


7
General Maintenance and Storage Specifications Safety System POWER CORD
Cleaning Instructions Physical Characteristics THERMOSTAT 15 ft., SJT, 3 cond., 13 A
Independent electronic circuit; thermal cutoff shuts the heater OFF 15 ft., SJT, 3 cond., 15 A
1. Disconnect the warming unit from the power source DIMENSIONS 4.6 m, HAR, 3 cond., 10 A
to ensure hose end air remains below 56°C (53°C ± 3°C typical);
before cleaning. 13" high x 13" deep x 14" wide (33 cm high x 33 cm deep x
back-up over-temperature detection at warming unit hose inlet. DEVICE RATINGS
36 cm wide)
2. Cleaning should be performed in accordance with hospital 110-120 VAC, 50/60 Hz, 11.7 A, or
ALARM SYSTEM
practices for cleaning OR equipment. After every use; wipe WEIGHT 220-240 VAC, 50/60 Hz, 7.2 A, or
the warming unit, the outside of the warming unit hose, and Over-temperature (≤56°C, 53°C ± 3°C typical): red Over-Temp
16 lb (7.3kg) 100 VAC, 50/60 Hz, 15 A
any other surfaces that may have been touched. Use a damp, indicator light flashes, alarm sounds, heater and blower shut
soft cloth and a hospital approved mild detergent, germicidal RELATIVE NOISE LEVEL down, operating indicator lights turn OFF, and control panel FUSES
disposable wipes, disinfecting towelettes, or antimicrobial 53 dBA (High fan setting) becomes unresponsive.
TYPE
spray. 48 dBA (Low fan setting) FAULT Fast acting ceramic fuses, 250 VAC
3. Let air dry or dry with a separate clean soft cloth. Amber Fault indicator light flashes, alarm sounds.
HOSE AMP RATING
Notice: Flexible, compatible with the Ranger 241 blood/fluid warming set. OVERCURRENT PROTECTION 12 A (110 - 120 VAC)
• Do not use cleaning solutions with greater than 80% Length: 80" (203 cm) Dual input fused lines.
8 A (220 - 240 VAC)
alcohol or solvents, including acetone and thinner, to clean
the warming unit or hose. Solvents may damage the labels UNIT INTAKE FILTER Electrical Characteristics 15 A (100 VAC)
and other plastic parts. MERV 14* HEATING ELEMENT
OPERATING SPEED
1400 W Resistive
• Do not immerse the Bair Hugger warming unit, warming * See Frequently Asked Questions on Filtration (p. 243) F (Fast Acting)
unit parts, or accessories in any liquid or subject them to LEAKAGE CURRENT
any sterilization process. RECOMMENDED FILTER CHANGE BREAKING CAPACITY
Meets UL 60601-1 and IEC 60601-1 requirements.
Every 12 months or 500 hours of use. 15 A, 12 A: 750 A @ 250 Vac
STORAGE BLOWER MOTOR
Store all components in a cool, dry place when not in use. MOUNTING 15 A, 12 A: 10,000 A @ 125 Vac
Operating speed: 4,700 rpm (high fan setting)
Can be clamped to an IV pole, placed on a hard surface, or mounted 8 A: 200 A @ 250 Vac
to the rolling stand accessory. 4,100 rpm (low fan setting)
Technical Service and Order Placement 8 A: 10,000 A @ 125 Vac
Temperature Characteristics Airflow: up to 48 cfm or 23 L/s
USA CERTIFICATIONS
+1-800-733-7775 RECOMMENDED OPERATING POWER CONSUMPTION IEC 60601-1; IEC/EN 60601-1-2; UL 60601-1; CAN/CSA-C22.2,
ENVIRONMENT TEMPERATURE Peak: 1550 W
OUTSIDE USA No.601.1; EN 55011; EN 80601-2-35
15°C-25°C
Contact your local service provider or distributor. Average: 800 W
HUMIDITY
When You Call for Technical Support
Max relative humidity 80% (up to 31°C) decreasing linearly to 50%
We will need to know the serial number of your Bair Hugger relative humidity at 40°C.
warming unit. The serial number label is located on the back of the
warming unit. ALTITUDE
Max 2,000m
In-Warranty Repair and Exchange
USA TEMPERATURE CONTROL
Call 3M customer service if your Model 775 warming unit requires Electronically controlled
factory service. A customer service representative will give you a HEAT GENERATED
Return Authorization (RA) number. Please use this RA number on High fan setting: 1600 BTU/hr (average), 470 W (average)
all correspondence concerning your warming unit. Your customer
service representative will also send a shipping carton to you at no Low fan setting: 1330 BTU/hr (average), 390 W (average)
charge, if needed. Call your local supplier or sales representative
OPERATING TEMPERATURES
to inquire about borrowing a warming unit while we service your
warming unit. For more detailed instructions on returning warming Average temperatures at the end of the hose:
units for service, please refer to the Model 775 Service Manual. HIGH:43° ± 1.5°C (109.4° ± 2.7°F)

OUTSIDE OF THE USA MED:38° ± 1.5°C (100.4° ± 2.7°F)


Contact your local distributor concerning in-warranty repair LOW:32° ± 1.5°C (89.6° ± 2.7°F)
and exchange.
TIME TO REACH OPERATING TEMPERATURE
2 - 5 minutes (dependent on blanket model)
Time required for the contact surface temperature to heat up from
23 ± 2°C to 37°C (73 ± 2°F to 99°F).

STORAGE/TRANSPORT TEMPERATURE
-20°C to 45°C (-4°F to 113°F)
Store all components in a cool, dry place when not in use.

8 3M™Bair Hugger™ Model 775 Warming Unit – Operator’s Manual GB / 34-8718-8811-0


9
General Maintenance and Storage Specifications Safety System POWER CORD
Cleaning Instructions Physical Characteristics THERMOSTAT 15 ft., SJT, 3 cond., 13 A
Independent electronic circuit; thermal cutoff shuts the heater OFF 15 ft., SJT, 3 cond., 15 A
1. Disconnect the warming unit from the power source DIMENSIONS 4.6 m, HAR, 3 cond., 10 A
to ensure hose end air remains below 56°C (53°C ± 3°C typical);
before cleaning. 13" high x 13" deep x 14" wide (33 cm high x 33 cm deep x
back-up over-temperature detection at warming unit hose inlet. DEVICE RATINGS
36 cm wide)
2. Cleaning should be performed in accordance with hospital 110-120 VAC, 50/60 Hz, 11.7 A, or
ALARM SYSTEM
practices for cleaning OR equipment. After every use; wipe WEIGHT 220-240 VAC, 50/60 Hz, 7.2 A, or
the warming unit, the outside of the warming unit hose, and Over-temperature (≤56°C, 53°C ± 3°C typical): red Over-Temp
16 lb (7.3kg) 100 VAC, 50/60 Hz, 15 A
any other surfaces that may have been touched. Use a damp, indicator light flashes, alarm sounds, heater and blower shut
soft cloth and a hospital approved mild detergent, germicidal RELATIVE NOISE LEVEL down, operating indicator lights turn OFF, and control panel FUSES
disposable wipes, disinfecting towelettes, or antimicrobial 53 dBA (High fan setting) becomes unresponsive.
TYPE
spray. 48 dBA (Low fan setting) FAULT Fast acting ceramic fuses, 250 VAC
3. Let air dry or dry with a separate clean soft cloth. Amber Fault indicator light flashes, alarm sounds.
HOSE AMP RATING
Notice: Flexible, compatible with the Ranger 241 blood/fluid warming set. OVERCURRENT PROTECTION 12 A (110 - 120 VAC)
• Do not use cleaning solutions with greater than 80% Length: 80" (203 cm) Dual input fused lines.
8 A (220 - 240 VAC)
alcohol or solvents, including acetone and thinner, to clean
the warming unit or hose. Solvents may damage the labels UNIT INTAKE FILTER Electrical Characteristics 15 A (100 VAC)
and other plastic parts. MERV 14* HEATING ELEMENT
OPERATING SPEED
1400 W Resistive
• Do not immerse the Bair Hugger warming unit, warming * See Frequently Asked Questions on Filtration (p. 243) F (Fast Acting)
unit parts, or accessories in any liquid or subject them to LEAKAGE CURRENT
any sterilization process. RECOMMENDED FILTER CHANGE BREAKING CAPACITY
Meets UL 60601-1 and IEC 60601-1 requirements.
Every 12 months or 500 hours of use. 15 A, 12 A: 750 A @ 250 Vac
STORAGE BLOWER MOTOR
Store all components in a cool, dry place when not in use. MOUNTING 15 A, 12 A: 10,000 A @ 125 Vac
Operating speed: 4,700 rpm (high fan setting)
Can be clamped to an IV pole, placed on a hard surface, or mounted 8 A: 200 A @ 250 Vac
to the rolling stand accessory. 4,100 rpm (low fan setting)
Technical Service and Order Placement 8 A: 10,000 A @ 125 Vac
Temperature Characteristics Airflow: up to 48 cfm or 23 L/s
USA CERTIFICATIONS
+1-800-733-7775 RECOMMENDED OPERATING POWER CONSUMPTION IEC 60601-1; IEC/EN 60601-1-2; UL 60601-1; CAN/CSA-C22.2,
ENVIRONMENT TEMPERATURE Peak: 1550 W
OUTSIDE USA No.601.1; EN 55011; EN 80601-2-35
15°C-25°C
Contact your local service provider or distributor. Average: 800 W
HUMIDITY
When You Call for Technical Support
Max relative humidity 80% (up to 31°C) decreasing linearly to 50%
We will need to know the serial number of your Bair Hugger relative humidity at 40°C.
warming unit. The serial number label is located on the back of the
warming unit. ALTITUDE
Max 2,000m
In-Warranty Repair and Exchange
USA TEMPERATURE CONTROL
Call 3M customer service if your Model 775 warming unit requires Electronically controlled
factory service. A customer service representative will give you a HEAT GENERATED
Return Authorization (RA) number. Please use this RA number on High fan setting: 1600 BTU/hr (average), 470 W (average)
all correspondence concerning your warming unit. Your customer
service representative will also send a shipping carton to you at no Low fan setting: 1330 BTU/hr (average), 390 W (average)
charge, if needed. Call your local supplier or sales representative
OPERATING TEMPERATURES
to inquire about borrowing a warming unit while we service your
warming unit. For more detailed instructions on returning warming Average temperatures at the end of the hose:
units for service, please refer to the Model 775 Service Manual. HIGH:43° ± 1.5°C (109.4° ± 2.7°F)

OUTSIDE OF THE USA MED:38° ± 1.5°C (100.4° ± 2.7°F)


Contact your local distributor concerning in-warranty repair LOW:32° ± 1.5°C (89.6° ± 2.7°F)
and exchange.
TIME TO REACH OPERATING TEMPERATURE
2 - 5 minutes (dependent on blanket model)
Time required for the contact surface temperature to heat up from
23 ± 2°C to 37°C (73 ± 2°F to 99°F).

STORAGE/TRANSPORT TEMPERATURE
-20°C to 45°C (-4°F to 113°F)
Store all components in a cool, dry place when not in use.

8 3M™Bair Hugger™ Model 775 Warming Unit – Operator’s Manual GB / 34-8718-8811-0


9
CLASSIFICATION Guidance and manufacturer’s declaration – electromagnetic immunity
Guidance and manufacturer’s declaration – electromagnetic emissions The Model 775 warming unit is intended for use in the electromagnetic environment specified below. The customer or the user of the Model
The Model 775 warming unit is intended for use in the electromagnetic environment specified below. The customer or the user of the Model 775 warming should assure that it is used in such an environment.
775 warming unit should assure that it is used in such an environment. Immunity Test IEC 60601 test level Compliance Electromagnetic environment - guidance
Emissions test Compliance Electromagnetic environment - guidance level
RF emissions Group 1 The Model 775 warming unit uses RF energy only for its internal function. Therefore, Conducted RF 3 Vrms 3 Vrms Portable and mobile RF communications equipment should be used no
its RF emissions are very low and are not likely to cause any interference in nearby IEC 61000-4-6 150 kHz to 80 MHz closer to any part of the Model 775 warming unit, including cables, than
CISPR 11 3 V/m
electronic equipment. the recommended separation distance calculated from the equation
Radiated RF 3 V/m
applicable to the frequency of the transmitter.
RF emissions Class B The Model 775 warming unit is suitable for use in all establishments, including domestic IEC 61000-4-3 80 MHz to 2,5 GHz
establishments and those directly connected to the public low-voltage power supply Recommended separation distance
CISPR 11
network that supplies buildings used for domestic purposes. d = 1.2
Harmonic emissions Class A
d = 1.2  80 MHz to 800 MHz
IEC 61000-3-2
d = 2.3  800 MHz to 2,5 GHz
Voltage fluctuations/ Complies
where P is the maximum output power rating of the transmitter in watts
Flicker emissions
(W) according to the transmitter manufacturer and d is the recommended
IEC 61000-3-3 separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site surveya, should be less than the compliance level in
Guidance and manufacturer’s declaration – electromagnetic immunity each frequency range.b
The Model 775 warming unit is intended for use in the electromagnetic environment specified below. The customer or the user of the Model Interference may occur in the vicinity of equipment marked with the
775 warming unit should assure that it is used in such an environment. following symbol:
Immunity Test IEC 60601 test level Compliance level Electromagnetic environment - guidance
Electrostatic ±6 kV contact ±6 kV contact Floors should be wood, concrete or ceramic tile. If floors are
discharge (ESD) covered with synthetic material, the relative humidity should NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
±8 kV air ±8 kV air
be at least 30%.
IEC 61000-4-2 NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
Electrical ±2 kV for power ±2 kV for power Mains power quality should be that of a typical commercial or structures, objects and people.
fast transient/burst supply lines supply lines hospital environment. a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio,
IEC 61000-4-4 ±1 kV for input/output ±1 kV for input/output AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment
lines lines due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which
the Model 775 warming unit is used exceeds the applicable RF compliance level above, the Model 775 warming unit should be observed to
Surge ±1 kV line to line ±1 kV line to line Mains power quality should be that of a typical commercial or verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the
hospital environment. Model 775 warming unit.
IEC 61000-4-5 ±2 kV line(s) to earth ±2 kV line(s) to earth
b
Voltage dips, short <5% UT <5% UT Mains power quality should be that of a typical commercial or Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
interruptions and (>95% dip in UT) (>95% dip in UT) hospital environment. If the user of the Model 775 warming
voltage variations for 0,5 cycle for 0,5 cycle unit requires continued operation during power mains
on power supply interruptions, it is recommended that the Model 775 warming Recommended separation distances between portable and mobile RF communications equipment and the Model 775 warming unit
40% UT 40% UT
input lines unit be powered from an uninterruptible power supply or
(60% dip in UT) (60% dip in UT) The Model 775 warming unit is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The
a battery.
IEC 61000-4-11 for 5 cycles for 5 cycles customer or the user of the Model 775 warming unit can help prevent electromagnetic interference by maintaining a minimum distance
70% UT 70% UT between portable and mobile RF communications equipment (transmitters) and the Model 775 warming unit as recommended below,
(30% dip in UT) (30% dip in UT) according to the maximum output power of the communications equipment.
for 25 cycles for 25 cycles Rated maximum output power of Separation distance according to frequency of transmitter m
<5% UT <5% UT transmitter W
150 kHz to 80 MHz 80 MHz to 800MHz 800 MHz to 2,5 GHz
(>95% dip in UT) (>95% dip in UT)
d = 1.2 d = 1.2 d = 2.3
for 5 sec for 5 sec
0.01 0.12 0.12 0.23
Power frequency 3 A/m 3 A/m Power frequency magnetic fields should be at levels
(50/60 Hz) characteristic of a typical location in a typical commercial or 0.1 0.38 0.38 0.73
magnetic field hospital environment. 1 1.2 1.2 2.3
IEC 61000-4-8 10 3.8 3.8 7.3
NOTE UT is the a.c. mains voltage prior to application of the test level. 100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated
using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts
(W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.

10 3M™Bair Hugger™ Model 775 Warming Unit – Operator’s Manual GB / 34-8718-8811-0


11
CLASSIFICATION Guidance and manufacturer’s declaration – electromagnetic immunity
Guidance and manufacturer’s declaration – electromagnetic emissions The Model 775 warming unit is intended for use in the electromagnetic environment specified below. The customer or the user of the Model
The Model 775 warming unit is intended for use in the electromagnetic environment specified below. The customer or the user of the Model 775 warming should assure that it is used in such an environment.
775 warming unit should assure that it is used in such an environment. Immunity Test IEC 60601 test level Compliance Electromagnetic environment - guidance
Emissions test Compliance Electromagnetic environment - guidance level
RF emissions Group 1 The Model 775 warming unit uses RF energy only for its internal function. Therefore, Conducted RF 3 Vrms 3 Vrms Portable and mobile RF communications equipment should be used no
its RF emissions are very low and are not likely to cause any interference in nearby IEC 61000-4-6 150 kHz to 80 MHz closer to any part of the Model 775 warming unit, including cables, than
CISPR 11 3 V/m
electronic equipment. the recommended separation distance calculated from the equation
Radiated RF 3 V/m
applicable to the frequency of the transmitter.
RF emissions Class B The Model 775 warming unit is suitable for use in all establishments, including domestic IEC 61000-4-3 80 MHz to 2,5 GHz
establishments and those directly connected to the public low-voltage power supply Recommended separation distance
CISPR 11
network that supplies buildings used for domestic purposes. d = 1.2
Harmonic emissions Class A
d = 1.2  80 MHz to 800 MHz
IEC 61000-3-2
d = 2.3  800 MHz to 2,5 GHz
Voltage fluctuations/ Complies
where P is the maximum output power rating of the transmitter in watts
Flicker emissions
(W) according to the transmitter manufacturer and d is the recommended
IEC 61000-3-3 separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site surveya, should be less than the compliance level in
Guidance and manufacturer’s declaration – electromagnetic immunity each frequency range.b
The Model 775 warming unit is intended for use in the electromagnetic environment specified below. The customer or the user of the Model Interference may occur in the vicinity of equipment marked with the
775 warming unit should assure that it is used in such an environment. following symbol:
Immunity Test IEC 60601 test level Compliance level Electromagnetic environment - guidance
Electrostatic ±6 kV contact ±6 kV contact Floors should be wood, concrete or ceramic tile. If floors are
discharge (ESD) covered with synthetic material, the relative humidity should NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
±8 kV air ±8 kV air
be at least 30%.
IEC 61000-4-2 NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
Electrical ±2 kV for power ±2 kV for power Mains power quality should be that of a typical commercial or structures, objects and people.
fast transient/burst supply lines supply lines hospital environment. a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio,
IEC 61000-4-4 ±1 kV for input/output ±1 kV for input/output AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment
lines lines due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which
the Model 775 warming unit is used exceeds the applicable RF compliance level above, the Model 775 warming unit should be observed to
Surge ±1 kV line to line ±1 kV line to line Mains power quality should be that of a typical commercial or verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the
hospital environment. Model 775 warming unit.
IEC 61000-4-5 ±2 kV line(s) to earth ±2 kV line(s) to earth
b
Voltage dips, short <5% UT <5% UT Mains power quality should be that of a typical commercial or Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
interruptions and (>95% dip in UT) (>95% dip in UT) hospital environment. If the user of the Model 775 warming
voltage variations for 0,5 cycle for 0,5 cycle unit requires continued operation during power mains
on power supply interruptions, it is recommended that the Model 775 warming Recommended separation distances between portable and mobile RF communications equipment and the Model 775 warming unit
40% UT 40% UT
input lines unit be powered from an uninterruptible power supply or
(60% dip in UT) (60% dip in UT) The Model 775 warming unit is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The
a battery.
IEC 61000-4-11 for 5 cycles for 5 cycles customer or the user of the Model 775 warming unit can help prevent electromagnetic interference by maintaining a minimum distance
70% UT 70% UT between portable and mobile RF communications equipment (transmitters) and the Model 775 warming unit as recommended below,
(30% dip in UT) (30% dip in UT) according to the maximum output power of the communications equipment.
for 25 cycles for 25 cycles Rated maximum output power of Separation distance according to frequency of transmitter m
<5% UT <5% UT transmitter W
150 kHz to 80 MHz 80 MHz to 800MHz 800 MHz to 2,5 GHz
(>95% dip in UT) (>95% dip in UT)
d = 1.2 d = 1.2 d = 2.3
for 5 sec for 5 sec
0.01 0.12 0.12 0.23
Power frequency 3 A/m 3 A/m Power frequency magnetic fields should be at levels
(50/60 Hz) characteristic of a typical location in a typical commercial or 0.1 0.38 0.38 0.73
magnetic field hospital environment. 1 1.2 1.2 2.3
IEC 61000-4-8 10 3.8 3.8 7.3
NOTE UT is the a.c. mains voltage prior to application of the test level. 100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated
using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts
(W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.