Download as pdf or txt
Download as pdf or txt
You are on page 1of 5

Ultrasound EQC - South Asia Region

Customer Information
Customer Name

GRANT MEDICAL FOUNDATION

Service Record #: System ID Customer Equipment ID:

1-9306776438 083037211847720 NA

Service Manual Document # Model Name: Software Version:

DOC2070873 Revision: 9 VENUE GO R2

Model Description: EQC Frequency:

VENUE GO 12 Months

>
Serial Number :

VGB001160

Form Status: Complete Incomplete

Start Date Country Modality

2023-02-16 INDIA-IN UL
>

No. of Probes:

3
>

UL EQC

ULTRASOUND EQC

Preliminary Checks

PM Task Procedure Results Comments

Ask/listen PSM/BUM Pass


>

Verify Power up PSM/BUM Pass 230V


>

/
Back up network & presets PSM/BUM Pass

>
Review Error Logs(Optional) PSM/BUM Pass

>
Back up software (if PSM/BUM Pass

>
applicable)

Perform diagnostics (optional) PSM/BUM Pass

>
Free archive space%( PSM/BUM Pass

>
discuss with customer )
Check free archive space. If
disk is greater than 75% full
advice customer to backup
and remove data.

-In the contact GE Section, PSM/BUM Pass

>
Clear the Queue and the
Machine Queue
-Configure and / or Verify
Insite Agent Configuration,
Check and Validate the InSite
Configuration (if applicable)
Confirm CRM # field matches
the SID (if applicable) in
Siebel
-Configure Default User for
Insite RFS / Auto-RFS
-Do Not Send an INSITE
Checkout RFS (USCAN)

Clear Spooler ( discuss with PSM/BUM Pass


>

customer )

Cleaning

PM Task Procedure Results Comments

-Clean Console PSM/BUM Pass


>

-Clean Control Panel


-Clean Monitor
-Clean Touch Screen
-Clean Probe Holder
-Clean Filters
/
Inspect cleaning agents and PSM/BUM Pass

>
ensure approved agents
being used (discuss with
customer ) look for yellowing
cables / console for misuse of
saniclothhb/af3 on parts of
system not intended.
Probe cleaning Site
(https://www.gehealthcare.com/products/ultrasound/ultrasound-
transducers)
System cleaning Site
(https://cleaning.gehealthcare.com/)

Physical Inspection

PM Task Procedure Results Comments

-Check external I/O PSM/BUM Pass

>
-Check wheels and brakes
-Check control panel
movement and operation
-If applicable, check monitor
movement
-Check cables and
connectors
-Check shielding and covers
-Check console
-Check that system date /
Time are correct

Inspect battery function (if PSM/BUM Pass


>

applicable)

Inspect probe holder function PSM/BUM Pass


>

Functional Checks

PM Task Procedure Results Comments

Check monitor operation. PSM/BUM Pass


>

Check touch panel operation. PSM/BUM Pass


>

/
Check keyboard operation. PSM/BUM Pass

>
-Check B-Mode Operation PSM/BUM Pass

>
-Check M-mode operation
-Check CF-mode operation
-Check 3D-mode operation (If
Applicable)
-Check 4D-mode operation (If
Applicable)
-Check Doppler modes
operation
-Check peripherals operation

Probe Identification Inspection Results Comments

Type Visual Pass

>
C1-5 RS Functional

Model Care Assmnt


5384874

Serial

853460WX4

Probe Identification Inspection Results Comments

Type Visual Pass


>

12L RS Functional

Model Care Assmnt


5141337

Serial

838201WX2

Probe Identification Inspection Results Comments

Type Visual Pass


>

3SC RS Functional

Model Care Assmnt


47237516

Serial

961965WX1
/
Comments, tools, and attachments
Comments Character(s) 2000

Tools used
Description Serial Number Bar Code Cal Due Date

MULTIMETER 44470318 2023-06-28

Attachments
Choose Files No file chosen
Attached Documents :

GE Representative
Number of Signers

1
>

Name: SSO #: SR #

NUPUR SINGH 212364483 1-9306776438

Date complete Form complete

2023-02-16

Customer Signature

Yes Not Required

Document Number: BBLNP1254 Revision: 1