Sample Information: Flutamida 250 MG Lote: 19284

You might also like

Download as docx, pdf, or txt
Download as docx, pdf, or txt
You are on page 1of 1

SAMPLE INFORMATION

Sample Name: MR 100% Acquired By: Analista_1


Sample T ype: Unknown Sample Set Name: Secuencia UPLC
Vial: Acq. Method Set: Fluorouracilo Met Set UPLC
Injection #: 5, 3, 2, 1, 4 Processing Method: Met Pro 9
Injection Volume: 2.00 ul Channel Name: W2996 266nm-1.2nm
Run Time: 1.7 Minutes Proc. Chnl. Descr.: W2996 266nm-1.2nm

Date Acquired:
15/7/2019 10:42:43 ETC, 15/7/2019 10:46:15 ETC, 15/7/2019 10:49:47 ETC, 15/7/2019 10:53:19 ETC, 15/7/2019 10:56:51 ETC

Date Processed: 17/7/2019 8:43:21 ETC, 17/7/2019 8:44:06 ETC, 17/7/2019 8:44:51 ETC, 17/7/2019 8:45:36 ETC, 17/7/2019 8:46:21 ETC

Flutamida 250 mg
Lote: 19284

Peak Summary with Statistics


Name: Flutamida
Sample Retention Amount
Vial Inj Name Area Height
Name Time (min) (%)
1 MR 100% 1: A,2 5 Flutamida 5.241 4588358 593552 99.56
2 MR 100% 1: A,3 Flutamida
3 5.240 4588330 593562 99.56
3 MR 100% 1: A,4 Flutamida
2 5.241 4588288 593560 99.55
4 MR 100% 1: A,5 Flutamida
1 5.241 4588355 593546 99.56
5 MR 100% 1: A,6 Flutamida
4 5.237 4588340 593574 99.56
Mean 5.240 99.56
Std. Dev. 0.001
% RSD 0.03

Reported by User: Analista_1 Project Name: Análisis de Ptoductos/Flutamida


Report Method: M MP Date Printed:
Report Method ID: 2051 17/07/19
Page: 1 of 1 8:47:04 America/Guayaquil

You might also like