MINDRAY Digital Radiography System - DigiEye 760

Description
General requirement
ceiling suspension multi-purpose digital radiography system, which
can satisfy a full range of clinical application, high throughput with 1 detectors.

Main technical specifications and requirements


X-ray tube
anode type:Rotating anode
Target angle: 12°
Input Voltage: 40~ 150kV
Max Peak X-ray tube voltage: ≥150kV
Input current: Min. ≤400mA, Max. ≥1000mA
Nominal focus:0.6/1.2mm
Small focus output power: ≥22KW,big focus output power:≥54KW
Max anode heat capacity: ≥230KHU
Housing heat storage capacity: 1345 kHU
Max housing heat dissipation rate 272 HU/sec

Anode cooling rate: ≥1000HU/sec


Anode Frequency: 0-25kHz
Anode Rotating speed: 3200r/min (60Hz) 2700r/min (50Hz)

Inherent filtration: 1.3mmAl/75kV


Overheat capacity protection

Generator
Super High-frequency Generator
Frequency: 460KHz
output power: ≥50KW
Max.current: ≥640mA
Min. current: ≤10mA
Current steps: 10mA, 12.5mA, 16mA, 20mA, 25mA, 32mA, 40mA,
50mA, 64mA, 80mA, 100mA, 125mA, 160mA, 200mA, 250mA,
320mA, 400mA, 500mA, 640mA

KV range:40~ 150kV,step: 1 KV/step


Min. loading time: ≤1ms
Max. loading time: ≥8s
Current time product range: ≥0.1 mAs ~ 640 mAs
0.1mAs;0.125mAs;0.16mAs;0.2mAs;0.25mAs;0.32mAs;0
.4mAs;0.5mAs;0.64mAs;0.8mAs;1.0mAs;1.25mAs;1.6m As;2.0mAs;
2.5mAs;3.2mAs;4.0mAs;5.0mAs;6.4mAs;8. 0mAs;10mAs;12.5mAs;
16mAs;20mAs;25mAs;32mAs;4 0mAs;50mAs;64mAs;80mAs;
100mAs;125mAs;160mAs
;200mAs;250mAs;320mAs;400mAs;500mAs;640mAs

Exposure mode: Automatic Exposure Control(AEC) and Manual


Exposure Control both
Generator Size: ≤620mm×360mm×820mm(L*W*H) or less
380/400/415/440V~ 3ph 50/60Hz
Power: MAX:50VA Standby:3.5kVA

Detector
Number of detector: 1 unit in wall stand .
Detector type: Fixed digital flat panel detector

Material: Amorphous silicon (a-Si) semiconductor with cesium


lodide (CsI) scintillator
Cooling Method: Natural cooling
X-ray sensitive array: 3072*3072 pixels
Detector area: ≥43*44cm
Pixel size: 140μm
Spatial resolution: 3.6 Lp/mm
Dynamic range: ≥14bit
DQE(0.1LP/mm,RQA5): ≥70%
Imaging time of imaging system:3s
Preview time of imaging system:2.3s
Cycle time of imaging system: ≤6s
Data transmission method: Ethernet

rack system and wallstand


rack system: Single track ceiling suspension
Rotation range of Detector: ≥-20º ~+90º
Detector Vertical move range: ≥500mm ~ 1700mm
Vertical Rotation of Tube: ≥-110º ~+110º
Vertical move range of tube: 500mm ~ 1700m
Longitudinal motion of tube: 0~ 2000mm
SID: 70 ~200cm
Automatic/mannual collimator
Control mode:
1.Control panel on the Tube-cover;
2.Touchscreen on the Tube-cover;
3.Control panel on both side of Detector-cover;
4.Remote control pannel

Grid

Type: Fixed Grid, it can be removed without using any tool, 2 pacs.

Focus: 1.8m in wallstand for chest position, 1.0m under patient


table for laying position

anti-scatter grid: grid density: 40L/cm,grid ratio:8:1


3-field ionization chamber,with automatic exposure control

Chamber
Type: 3-field ionization chamber
inherent filtration: 0.4mmAl/100kV
Voltage range: ≥40kV~150kV
Loading time: 0.5ms~10s
Dose range: 0.174~174uGy

Patient table
Mobile patient table
move by wheels which can be locked
Aluminium equivalent of table-board≤1.0mmAl/100kV
table height≥70cm
table size≥200x65cm
maxmium Load capacity≥250kg

Warranty