Ind560x – Programming

Supervisor protocol: SPIDER


Supervisor comunication parameters:9600,N,8,1 Dialog

To access Setup menu’ press:

Setup soft key:

Setup
Terminal
Nazione
Inglese
User
Set following password to admin user: “ADMIN” (admin user must be lowercase,
password ADMIN must be uppercase)
Comunication
Connection
COM1. In case you are using ACM200 setup COM4 or COM5 (com4 and com5
must be enabled before configure)
Com4 or 5 with ACM200, must be configured as STANDARD

Edit Connection

new connection configure COMx as


Assigment: SICS

use following edit soft key to modify the connection.


SERIAL
COM1 or COM4 or COM5
9600,NONE,8 – RS232
SCALE
Filter
Average Condition
Fine Filling
Zero
AZM & Display
Power UP (Restart)

Note2:
In case scale start with following error message: “switch non protetto” from Setup menu set up “NO
APPROVE” instead of “APPROVE”
Ind560x – Calibration

To access Setup menu’ press:

Setup soft key:

If scale ask for a user to enter in setup menu,


Type “admin” LOWERCASE as user
And Password: “INK” or “ADMIN” UPPERCASE.

SCALE
CALIBRATION
(softkey) “SRVC MODE”

RETURN PRESS “NO”


RESET PRESS “NO”
NATION PRESS “NO”
SCALE PARAMETERS PRESS “NO”
LINEARITY PRESS “NO”
CALIBRATION PRESS “OK”
CAL INTERNAL PRESS “OK”
SET PRELOAD PRESS “OK” (SET THE FINAL PRELOAD ON PLATFORM)
------ CALIBRATION ------
(WAIT THE END….)
CAL EXTERNAL PRESS “OK”
SET PRELOAD PRESS “OK” (SET THE FINAL PRELOAD ON PLATFORM)
SET FULLCAP (LOAD CALIBRATION WEIGHT ON THE SCALE) THEN  PRESS “OK”
(set calibration weight)
CA xxxx KG (BLINKING)  PRESS”OK”
------ CALIBRATION ------
(WAIT THE END….)
UNLOAD  (UNLOAD SCALE PLATFORM) THEN PRESS”OK”
ADAPTION PRESS “NO”
SAVE PARAMETERS PRESS “OK”
(reinizializing IDNET….. Wait…)

Then press two times left display soft key “ARROW” to go out from programming menu.