Professional Documents
Culture Documents
Making Controls With Patient Samples 20200516 v2.0
Making Controls With Patient Samples 20200516 v2.0
Making Controls With Patient Samples 20200516 v2.0
1. Negative serum collection: collect serum from those who are not infected with COVID-
19. Each sample should be clear, transparent, and free of hemolysis, lipolysis, or
jaundice. It is recommended to collect a sample volume of not less than 15mL for a
1-month quality control length. Mix the collected sera evenly. If you can not collect
totally 15mL in one day. the serum can be put into 2-8° C for max. 3 days to allow to
2. Positive serum collection: collect serum from those who are infected with COVID-19
(confirmation). Choose samples that are about 2 to 8 times of the cutoff value (for IgM
with the value of 2-8, for IgG with the value of 20-80). Each sample should be clear,
a sample volume of not less than 15mL for a 1-month quality control length. Mix the
for 30 min. If there are too few positive samples in the required concentration range,
you can use the negative samples to dilute the inactivated positive samples to the
required concentration. (Use negative serum to run 1:100, 1:50, 1:25, 1:10, 1:5, and 1:2
dilution so that you can determine the dilution factor.) It is recommend to confirm
the supernatant.
4. Aliquot and storage: after the preparation is completed, you can confirm on the
Mindray system whether the prepared quality control meets the requirements. After
confirming that it is correct, aliquot the negative control and the positive quality
control as 200uL / tube, respectively, and place it under -20 ℃ for long-term storage
(28 days).
5. Test: take a tube of quality control out from the fridge. Leave it at room temperature
for about 30 minutes to thaw thoroughly. Mix it and test it on the analyzer. Discard it
6. Quality control target value and range establishment: each day before the start of the
patient sample test and after the completion of the patient sample test, test one
negative control and one positive control in duplicate respectively. Run such tests for
5 days and you can obtain 20 results for each level of control. Calculate the mean and
The negative control has a target value range as the table below.
Negative Control
IgM 0~0.8
(When the negative control result is within the above range, it is considered acceptable.
If you would love to use the Levey-Jennings chart (mean±3SD), you can set the mean
as 4 U/mL for IgG and 0.4 for IgM and SD as 1.33 U/mL for IgG and 0.133 for IgM.
You can also set the mean and SD as per your needs and make sure that the range is
Positive control
7. The target values of the quality control materials are lot specific and target value
8. In case quality control fails for any reason, thaw a new aliquot and re-assess the
9. Controls should be run individually at least once every 24 hours when the test is in
10. Please follow the applicable government regulations and local guidelines for quality
control.