Art-8-3 PPB Dissolved Oxygen Measurement
Art-8-3 PPB Dissolved Oxygen Measurement
➔ OH- ions are produced and Clare hydroxide (KOH) as the most frequently used
consumed in the overall reaction which electrolyte. The cathode must be a noble metal
alters the chemistry of the electrolyte. such as silver or gold in order for the cathode
This causes a zero shift, and over time, potential to reduce molecular oxygen when the
the electrolyte will need to be replaced. cell circuit is closed. The anode should be a
base metal (iron, lead, cadmium, copper, zinc,
➔ The need for an external power source
or silver) with good stability and without the
to be applied to the electrode means
tendency for passivation. Galvanic technology
that as soon as the probe is
overcomes many of the disadvantages affiliated
disconnected, the power supply is cut
with polarographic sensors, for example:
off. Upon reconnection, the user must
wait for the probe to be polarized. ➔ No polarization time required. The probe
Polarization can take up to 1 hour and is ready to measure since it is 'self
any measurement taken before this time polarized' as long as electrolyte is
will typically result in a false high present.
reading.
➔ Provides an absolute zero measurement
as galvanic current is zero when zero
Galvanic Measuring Cell
oxygen is present.
By using carefully selected electrodes in
contact with an appropriate electrolyte, a ➔ Non-depleting electrolyte. The reaction
spontaneous chemical reaction occurs in the consumer and produces OH- at the same
presence of oxygen. In this reaction, the rate which as means no drift or zero
cathode reduces the oxygen into hydroxide, shift.
releasing four electrons for each oxygen ➔ Overall reaction produces metallic
molecule. hydroxide which flakes off the anode
leaving exposed anode for continuous
Voltmeter
reaction. Therefore, requires reduced
Silver Cathode
anode polishing.
O2+2H2O+4e-->4OH-
Teflon Membrane
The model 869 offers ppb D.O. measurement
for portable use mounted on a stainless steel
Sacrificial
Cathode
stand. The 869 offers two sensor and flow cell
Lead Anode
configurations; a CPVC option and a stainless
KOHFill Solution
Sensor Body
steel option for ultra low level accuracy.
Sacrificial Cathode
Insulator
SUMMARY
There are a variety of dissolved oxygen sensor
technologies, each with its own advantages and
disadvantages. One sensor technology will not
be suitable in all applications. However, in
comparing polarographic and galvanic
technologies, IC Controls saw compelling
evidence to substantiate the superiority of a
galvanic sensor and has opted to utilize only
galvanic technology for its ppb line of dissolved
oxygen measurement.
Common applications :
➔ deaerator evaluation
➔ breweries
➔ semiconductor manufacturing