Temp.

at 50 % RH
Uncertainty Procidure : -
UNCERTAINITY MEASUREMENT
(The values having ( ) right to them are variables
Range 50.00
TYPE - A LC UUC 0.01
DegC Uncertainity of std. 0.30
1st Reading 50.24 50.00 -0.24
2nd Reading 50.24 50.00 -0.24
3rd Reading 50.25 50.00 -0.25
4th Reading 50.25 50.00 -0.25
5th Reading 50.24 50.00 -0.24
6th Reading 50.23 49.99 -0.24
7th Reading 50.23 49.99 -0.24
8th Reading 50.23 49.99 -0.24
9th Reading 50.22 49.99 -0.23
10th Reading 50.22 49.99 -0.23

Average -0.24
-0.24
Standard Deviation 0.006666667 DegC

Total readings (n) 10

Under Root (n) 3.16227766

Uncertainity Ua( Under Test ) 0.00211 DegC UUC

TYPE - B

Uncertainity due to master 0.30 DegC Master Indicator


Normal Dis. K=2 2.000
U1= 0.150 DegC

Uncertainity due to Stability 0.01 DegC Stability of Chamber


Chamber 1.732
Rect. Dis. K=1.73205
U2= 0.01 DegC

Uncertainity due to Uniformity 0.0 DegC Uniformity of Chamber


Chamber 1.732
Rect. Dis. K=1.73205
U3= 0.00 DegC

Uncertainity due to 0.06 DegC Drift of master


Drift of master 1.732
Rect. Dis. K=1.73205
U4= 0.03 DegC

Uncertainty due to L.C. of UUC 0.01 DegC Least Count


Rect. Dis.K=1.73205 1.732
U5= 0.00 DegC

Combin Uncertainity 0.154 DegC


Degrees of freedom(v) 256916854.9
Coverage Factor 2
Expanded Uncertainity 0.31 DegC
Final Uncertinity (+/-) 0.31 DegC