Respiratory Medical Solutions

High-Flow Heated Respiratory Humidifiers

MAINTAIN
& SERVICE
MANUAL
HUMID-BHR/HUMID-BH/HUMID-BM

Prolinx GmbH
EC REP Brehmstr.56,40239 Düsseldorf Germany
Tel:0049-2131 4051968-0

Shenyang RMS Medical Tech Co.,Ltd


No. 21, Gaoke Road, Hunnan New District, Shenyang, China
Tel: 0086-024-31682686
Catalogue Quality Guarantee

Quality Guarantee......................................................................................... 1 RESPICARE HiFent TM high flow respiratory heated humidifier (vision) should be manufactured without
defects in technology and materials, such as
Preface ………………..................................................................................... 2
The warranty period shall be one year from the date of sale if the product is operated in accordance with the

1. Introduction of high flow respiratory heated humidifier................................... 2 product specifications.If the product doesn't work
Operate according to product specifications, the product will repair or replace defective materials or
1.1 Overview of Respicare HiFent TM high flow respiratory heated humidifier.. 2
components.This warranty does not include cause
Damage caused by accident, misuse, abuse, alteration and other defects not related to materials or
1.2 Service notification and technical support............................................... 2
processes.
3. Operating principles and instructions........................................................... 3 RESPICARE denies all responsibility for economic loss, profit loss, indirect loss or indirect damage
The loss may be due to the sale or use of the product.Some areas do not allow the exclusion or restriction
4. daily maintenance...................................................................................... 4
of incidental or
5. troubleshooting.......................................................................................... 5 Indirect damage, so the above limitation or exclusion may not apply to you.
This warranty supersedes all other express warranties.In addition, any implied warranty, including any
6. Repair and replacement.............................................................................. 6
merchantability or fitness

6.1 Defective heating plate.......................................................................... 6 All warranties of definite use are limited to one year.Some areas do not allow restrictions on the duration
of implied warranties, therefore
6.2 The display screen does not display........................................................ 6
The limitation may not apply to you.This guarantee gives you specific legal rights and you can have other
places as well
6.3 Start up but the fan does not work.......................................................... 6
Different rights.
7. Testing and calibration............................................................................... 7 The warranty period for repairs is 90 days and the replacement part is one year.
To exercise your rights under this guarantee, please contact your local authorized dealer or contact your
7.1 Re-aging and testing........................................................................... 7
company:
Welcome to our website at http://www.rms-care.com/

01
Preface ●


Do not continue to operate if the source wire or plug is damaged, the liquid enters, the output gas
temperature is too high, abnormal noise, etc
Excessive use of heating circuit, nasal cannula, etc can lead to infection and other injuries.
● when the environment temperature and humidity conditions used in the regulations, the output of
1. Introduction of high flow respiratory heated humidifier
gas temperature is lower than 43 ℃.
● when the environment temperature is beyond the scope prescribed temperature 18 ℃ to 28 ℃, the
1.1 Overview of RESPICARE HiFent TM high flow respiratory heated humidifier system
temperature and humidity ‘value of the gas output will be affected .
TM
RESPICARE HiFent high flow respiratory heated humidifier system , As shown in figure,It is a microprocessor- ● The performance of the therapeutic device is greatly affected when used in excess of specified
controlled adjustable flow rate and humidified respiratory treatment auxiliary system.
For HiFent TM,the screen showing includes multi-function keys such as display working mode, monitoring oxygen environmental conditions (temperature, humidity, etc.) to damage, and may cause harm to the
concentration, flow rate and temperature setting, etc. Real - time interface display and system alarm user interface. human body.
HiFent TM is characterized by a centrifugal blower generating air flow as well as hardware and software platforms,
● The alternating current input should be within the range of + / - 10% of the rated voltage, beyond
patients can be provided with humidification by adjusting the flow rate, temperature
and oxygen concentration,Fixed flow of air or oxygen.The system will alarm automatically when it is running due to which the device may be damaged
improper use or failure
● when exposed to electrocautery, electrosurgical, defibrillation, X-ray, infrared radiation, transient
electromagnetic field,
1.2 Service notification and technical support
These include magnetic resonance imaging (MRI) and radio interference, which can interfere with
This service manual is prepared by Shenyang RMS Medical Tech Co., LTD., mainly for the use of authorized technical
the device.
personnel to serve RESPIRCARE HiFent TM high flow respiratory heated humidifier .
Shenyang RMS Medical Tech Co., LTD. Is committed to customer satisfaction and can work with any problem or ● If the therapeutic device is blocked, covered or heated, or the air inlet, output pipe and air outlet,
technical support which can cause equipment to be overheat or even damaged.
● The output oxygen concentration may be increased if the output pipe is blocked during the
Contact.Telephone: 0086-24 - 31682686
connection of oxygen.
● The bottom of the humidifying tank has a heat dissipation hole. Do not touch the bottom when
moving the device to prevent injury to human body
2. Warning and Attention
Keep the device from moving when humidification pot with water, and do not turn upside down or
● The therapeutic device cannot be used for life support. tilt to prevent water from flowing into the device, causing the damage of the therapeutic device,
● When output the gas by this device, producing positive airway pressure, which is necessary if airway pressure has there is a "do not tilt when there is water" on the top of the warning paste to remind.
side effects on patients.It must be considered. ● When you do not use the device, you must unplug the power.
● The therapeutic device should be set up and used with the guidance of a professional physician.
● Do not open the inside of the therapeutic device, it should be repaired by an authorized repairman.
● The doctor's advice cannot be substituted for the advice in the handbook.
● The therapeutic device should be kept at the end of its life and should not be throwed away.
● The device can only use the humidification pot , heating circuit, nasal cannula and others provided by Shenyang
RMS Medical co., LTD.The use of unauthorized accessories and consumables may affect the function of the device Note: these are general warnings and cautions.
or cause danger.
● Humidification pot, heating circuit, nasal cannula and other accessories provided by the device can only be used in
the device itself ,Otherwise, it may affect the function of other devices or cause danger 3. Operating principles and instructions
● In case of power failure or machine failure, remove the nasal cannula and turn off the oxygen supply.
● To ensure electromagnetic compatibility of this equipment, the equipment needs to be Begin to use
installed, debugged and used in accordance with random files. Connection of device
● The portable and mobile radio frequency communication equipment has the potential to affect the Connect the device according to the device connection method.
electromagnetic compatibility of the equipment, as such Please contact the company to solve the Set the parameters
situation. Please set the parameters under the guidance of the doctor. If you continue the previous setting, skip
● The device can only be connected to the cable mentioned in the random document, and the device this one.When on the top right corner of setting the interface lock appears,the setting parameters
can be connected with the specified accessories and cables Standby connection may result in cannot be changed.When the lock is opened, the setting parameters can

increased emission or reduced disturbance rejection of the device. be changed.In the system layout interface ,the setting parameters can be changed.

● This device shall not be used close to or superimposed with other equipment, if it must be used
close to or superimposed ,observe that the validation works in the configuration it is using.
● The therapeutic apparatus should be placed in a well-ventilated place on all sides and not on a soft
surface. Do not use inflammable and narcotic gases around the terminal.
If the heating circuit is damaged or bent, or the therapeutic device falls over, there is obvious
external damage and electricity
02 03
4. Cleaning and Maintenance Preheat
Long press the start/stop button for 3 seconds to start the therapeutic device
HUMID-BHR/HUMID-BH/HUMID-BM series High-Flow Heated Respiratory Humidifiers achieves Heat, the flow and dew point temperature will fluctuate and rise, and the value will remain unchanged after reaching the
humidification by water. If regular cleaning, disinfection and replacement of parts are not carried out, the preset value.Preheat the knot
instrument may be at risk of bacteria and patient infection. The words "READY" appear.
The respiratory humidification therapeutic instrument must be cleaned after use each time and before use Adjust oxygen
again.It must be cleaned and disinfected in case it is used for different patients. HUMID-BHR/HUMID-BM therapeutic device needs to manually adjust the flow of oxygen source to make the monitoring
value of oxygen concentration showing on the screen of the therapeutic device reached the target value.HUMID-BH the
Cleaning rapeutic
Daily cleaning: tubes, nasal cannulas, HUMID-BH humidification chamber device automatically adjusts oxygen to the set oxygen concentration.The oxygen concentration must be within the range
of the setting oxygen concentration alarm , otherwise the therapeutic device will alarm
After use, run the dry mode, remove and wash the nasal cannulas with clean water, and then connect the
Note: the FiO2 of patient may be less than the monitored value due to the dilution effect of the inhaled air.
tubes and dry the nasal cannulas; after drying, remove the HUMID-BH chamber for water rinse.
Using
Weekly cleaning: respiratory humidification therapeutic instrument, nasal cannulas, HUMID-BH
humidification chamber. Put the nasal cannula in place and start treatment.In the course of treatment, the machine will be alarm when the external
power is disconnected (press the mute button to turn off the sound), the user should disconnect the oxygen source,
Disconnect the instrument from the power, remove the nasal cannulas, HUMID-BH humidification chamber remove the nasal cannula and check the power source, after troubleshooting before continuing to use.
and wash the nasal cannulas with warm water added with neutral detergent. Transportation function
Warning: HUMID-BM humidification box, heating tubes are forbidden to be cleaned.
When standby, preheat or ready, press the mute button for 3 seconds to start the transportation function in the condition
Use a clean, lint-free disposable cloth damped with neutral detergent to wipe from both ends of the elbow, of low power consumption without heating the humidification pot ; After 20 minutes, it enters the preheating state
making sure that all dirties is removed. with heating the humidification pot .
Use a piece of alcohol cotton to wile the elbow and its vicinity, making sure that all dirties is removed. End
! Warnings:To prevent electric shock, the power of positive-pressure ventilating therapeutic instrument At the end of using ,first remove the nasal cannula, disconnect the oxygen source, long press the on/off button for 3
should be cut off before cleaning. seconds, then entering the drying mode, the humidifier will give a sound and screen reminder to turn off the oxygen, and
make sure to turn off the oxygen source.After that, press the on/off button gently to cancel the alert status.The drying
Do not place the positive-pressure ventilating therapeutic instrument in any of the liquid environment. mode is to dry the circuit by heating the circuit for 99 minutes. After the end, the device will be in standby mode.
At this time, external power needs to be turned off.If want to turn off the drying mode forcibly, pls press the on/off button
again for 3 seconds, but the moisture stored in the circuit may be affects to reuse.
Warning: after use, be sure to turn off the oxygen to avoid the accumulation of oxygen in the therapeutic device .
Operating mode : Warning: do not wear a nasal cannula during drying to avoid heat damage.

Mode
Flow : 2-25L/min; 5. Troubleshooting
Low flow mode
Temp : 34℃;
(HUMID-BH/HUMID-BM)
FiO2: 21-100% Alarms Description Actions
10-80L/min(HUMID-BH/HUMID-BM);
After startup, in case of faults of internal
Temp : 31℃(10-60L/min); parts such as the motor, a fault alarm will Shutdown for restart. If the alarm persists, please
High flow mode Flow 34℃(10-80L/minHUMID-BH/HUMID-BM) Internal Fault be displayed and meanwhile a sound contact your supplier or RMS Medical.
37℃(10-60L/min) alarm will be generated in 5s.
FiO2: 21-100%
After startup, if a heated respiratory
Alarm Settings: tube is not detected by the Confirm that the respiratory tube is not
Alarm Parameter setting Default Respiratory instrument, a respiratory tube damaged and is inserted correctly. If the
Tube Alarm alarm will be displayed and alarm persists, replace the respiratory tube.
Alarm of lower FiO2 Off / 21-25% meanwhile a sound alarm will
be generated in 5s.
Off / 30-50%(HUMID-BHR) 50%
Alarm of higher FiO2
Off / 30-100%(HUMID-BH/HUMID-BM) 90%
If abnormal leak is detected by the Check whether the water pot is installed
Ambient temperature
Off /On Off
therapeutic instrument, a leak alarm properly;
alarm sound Leak Alarm will be displayed and meanwhile a Check whether the heating tube is
sound alarm will be generated in 5s. damaged and is inserted correctly.
Alarm of Block Off /On On
When selected, the alarm setting will
Alarm restore restore the preset value If a blocking is detected by the
therapeutic instrument, a blocking Check whether the water pot, heating tube
system layout Blocking Alarm alarm will be displayed and or nasal cannulas are blocked;
meanwhile a sound alarm will be Check whether the inlet cover and cotton
The clinical setting such as mode and alarm filter are blocked.
Clinical Settings cannot be changed when the condition is off generated in 10s.
or on.
When the oxygen concentration is
Modify the display language in Chinese detected to be below the limit value,
Language
and English a low oxygen concentration alarm Check whether the humidifier's oxygen inlet is
Low Oxygen will be displayed and meanwhile a connected to the oxygen source correctly;
Concentration Alarm sound alarm will be generated in 20s. Adjust the flow of oxygen source as needed.
The range of low FiO2 is 21-25%
(with a step of 1%).The original setting
is 21%.

04 05
Alarms Description Actions
6.2 The display works improperly

When the oxygen concentration is detected


to exceed the limit value, a high oxygen
concentration alarm will be displayed and
High Oxygen meanwhile a sound alarm will be generated
Concentration Alarm in 20s. The range of high FiO2 is 30-100 %
Adjust the flow of oxygen source as needed.
(HUMID-BH/BM), or 30-50%(HUMID-BHR).
The original setting is 90%/50%.
(with a step of 1%).

When the instrument fails to reach the Check whether the water pot, heating tube or The LCD screen is not Check if the shielded
Not Reaching pre-set flow in 10min (±1 min), a alarm nasal cannulas are blocked; on after electrification wire fixed properly
will be displayed for not reaching the Check whether the inlet cover and cotton filter
Target Flow Alarm target flow and meanwhile a sound are blocked. Remove 5 screws on the front and back shell of the device, remove the top cover of the high-flow respiratory
alarm will be generated. Check whether the pre-set flow is too high.
humidifier and remove it
When water in water pot runs out, a Get the shielded wires and reconnect the wires of the display. Observe whether the circuit board wiring
alarm will be displayed for checking Check whether water in the pot boils away, harness is virtual connection with the screen wiring, if not; go ahead into the next step.
Check-water Alarm water and meanwhile a sound alarm which may cause damage to the pot;
will be generated in 20 min Make sure that water in the pot will not run out.
(flow>20L/min) or 40min (flow<20L/min).
When the instrument fails to reach
the pre-set temperature in 30min When the instrument's flow is too high while
Not Reaching Target (±3 min), a alarm will be displayed
for not reaching the target temperature the ambient temperature is too low, lower the
Temperature Alarm target flow or target temperature.
and meanwhile a sound alarm will
be generated.
Replace the LCD screen Fix the wires properly LCD display panel
When the ambient temperature is beyond the
When starting the instrument, if the range of 10-30℃, do not use the instrument Connect the lead wire such as the screen wire, turn on the power, and confirm whether the LCD screen has
ambient temperature is beyond the since it may be unable to reach to pre-set
Check-working- range of 10-30℃, a alarm will be backlight, and if the whole machine works well after start.
temperature.
condition Alarm displayed for checking the working A sudden change of working conditions may If the power supply is connected normally, there is backlight, and the whole machine is working normally,
conditions and meanwhile a sound cause an alarm. Start the instrument at ambient the LCD screen is not qualified, and please replace the LCD display and re-assemble on the main machine.
alarm will be generated in 60s (±6s). temperature and keep it run for 30 min, then If the display is still abnormal, replace the LCD panel, the product can work normally.
restart the instrument.
6.3 The blower works abnormally after starting
In the started state, once the external Check to confirm that the power supply is
power supply is disconnected, a sound connected well;
Power Failure Alarm alarm will be generated in 10s and Check whether the networks power supply
persists more than 120s. is normal.

6. Repair and Replacement


Wires of fan Fan and fan seal The power supply board The main board
6.1 Defective heating plate rubber rings

Heating element Remove the whole machine screw, remove the top shell and check whether the vertical socket is damaged.
sensor wire If it is damaged, replace the socket with the side out socket seat, and then cover the shell to confirm whether
it can work normally, such as abnormal further analysis, replace the power panel on the side. Power plate
12V often causes the fan to not work.
Heating Heating plate
element wire sensor wire If the main board or power plate unqualified, replace the power plate and main board, and assemble properly
to confirm whether the fan is working correctly.
If it works abnormally, the fan is not good, please replace with a new fan.

7. Testing and calibration


7.1 Re-Aging and testing

Test the resistance of Fix the screw


the heating plate and sensor with of heating plate
multimeter

Remove the 5pcs screws fixed on the covers both front and back. Push the clip of shielded wire, remove the
shielded wire, and then remove the top cover. Test the wire of heating plate by multimeter, the resistance is The finished products aging Record aging data
325Ω±10%. The resistances of two sensors on heating plate are 1K and 10k, the values differ from
temperature, and detailed value refers to specification. If resistance shows as open circuit or short circuit, After conforming the product works normally, transfer them into the aging room for aging. The aging time
please replace the heating plate.Remove the screws on the heating plate by screwdriver. Take the plate requires above 24 hours. In the aging environment, test the temperature and humidity value at three
down. As shown on the picture, replace with new heating plate and fix the wires properly. humidity 31, 34, 37 degrees separately to see if they met the technical and quality requirements.
The finished product is then inspected and finished the repair..
06 07