GEHC Site Planning Technical Guide CT Quick Reference Guide - XLS - Rev5

You might also like

Download as xls, pdf, or txt
Download as xls, pdf, or txt
You are on page 1of 11

CT QUICK REFERENCE - 15.Aug.

11

BRIGHTSPEED
SELECT
EDGE, EXCEL, ELITE
&
ELITE ENANCED
gantry/table flr.
anchored? YES

gantry cable access REAR

conduit sizes -
OC to Gantry 3 1/2"
-OC to PDU - 2 1/2"
PDU to Gantry 2 1/2" and 3 1/2"

39" to gantry, OC, AW,


injector electronics and
PDU & UPS varian & laserlight
interference limits controls
gauss field OC = 10G
proximity limit Gantry = 1G

OC - Gtry = 60'
standard cable OC - PDU = 60'
lengths PDU - Gtry = 20'

OC - Gtry = 75'
alternate cable OC - PDU = 75'
lengths PDU - Gtry = 55'

voltage/amps/MDC 480V/90A
cat NO. E4502AB
B7999ZA or E4502KY
UPS
recommended ceiling
height 9'-0"
absolute min. ceiling
height 8'-0"
absolute min. room
size
(LIMITED ACCESS) 10'-10" X 19'-5"

typical room size 12'-10" X 19'-3"


CT QUICK REFERENCE - 15.Aug.11

BRIGHTSPEED ELITE,
EDGE & EXCEL
gantry/table flr.
anchored? YES

gantry cable access REAR

conduit sizes -
OC to Gantry 3 1/2"
-OC to PDU - 2 1/2"
PDU to Gantry 2 1/2" and 3 1/2"

39" to gantry, OC, AW,


injector electronics and
PDU & UPS varian & laserlight
interference limits controls
gauss field OC = 10G
proximity limit Gantry = 1G

OC - Gtry = 60'
standard cable OC - PDU = 60'
lengths PDU - Gtry = 20'

OC - Gtry = 75'
alternate cable OC - PDU = 75'
lengths PDU - Gtry = 55'

voltage/amps/MDC 480V/90A
cat NO. E4502AB
B7999ZA or E4502KY
UPS
recommended ceiling
height 9'-0"
absolute min. ceiling
height 8'-0"
absolute min. room
size
(LIMITED ACCESS) 11'-0" X 20'-0"

typical room size 12'-10" X 19'-3"


CT QUICK REFERENCE - 15.Aug.11
BRIGHTSPEED
ELITE ENHANCED
gantry/table flr.
anchored? YES

gantry cable access REAR

conduit sizes -
OC to Gantry 3 1/2"
-OC to PDU - 2 1/2"
PDU to Gantry 2 1/2" and 3 1/2"

39" to gantry, OC, AW,


injector electronics and
PDU & UPS varian & laserlight
interference limits controls
gauss field OC = 10G
proximity limit Gantry = 1G

OC - Gtry = 60'
standard cable OC - PDU = 60'
lengths PDU - Gtry = 20'

OC - Gtry = 75'
alternate cable OC - PDU = 75'
lengths PDU - Gtry = 55'

voltage/amps/MDC 480V/90A
cat NO. E4502AB
B7999ZA or E4502KY
UPS
recommended ceiling
height 9'-0"
absolute min. ceiling
height 8'-0"
absolute min. room
size
(LIMITED ACCESS) 12'-10" X 19'-4"

typical room size 13'-0" X 20'-8"


CT QUICK REFERENCE - 15.Aug.11

DISCOVERY CT750
HD LIGHTSPEED VCT
gantry/table flr.
anchored? YES

gantry cable access REAR

conduit sizes -
OC to Gantry 3 1/2"
-OC to PDU - 2 1/2"
PDU to Gantry 2 1/2" and 3 1/2"

PDU & UPS


interference limits 39" to gantry and OC
gauss field OC = 10G
proximity limit Gantry = 1G

OC - Gtry = 60'
standard cable OC - PDU = 60'
lengths PDU - Gtry = 20'

OC - Gtry = 75'
alternate cable OC - PDU = 75'
lengths PDU - Gtry = 55'

voltage/amps/MDC
cat NO. 480V/125A

B7864PZ, P5064PS &


E4502F
((OLD) B7864PP or
UPS E4502KT)
recommended ceiling
height 9'-0"
absolute min. ceiling
height 7'-6"

11'-8" X 19'-2"
absolute min. room (GT1700)
size 11'-8" X 21'-2"
(LIMITED ACCESS) (GT2000)

13'-6" X 20'-0"
(GT1700)
13'-6" X 22'-0"
typical room size (GT2000)

Delivery: min. 5ft. corr. = 78"


opening, incl. 6 ft. corr. = 66"
Dollies w/o side 7 ft. corr. = 56"
rails 8 ft. corr = 48"
CT QUICK REFERENCE - 15.Aug.11

DISCOVERY CT590
RT, OPTIMA CT580
LIGHTSPEED
RT16/XTRA
gantry/table flr.
anchored? YES

gantry cable access REAR

conduit sizes -
OC to Gantry 3 1/2"
-OC to PDU - 2 1/2"
PDU to Gantry 2 1/2" and 3 1/2"

PDU & UPS


interference limits 39" to gantry and OC
gauss field OC = 10G
proximity limit Gantry = 1G

OC - Gtry = 60'
standard cable OC - PDU = 60'
lengths PDU - Gtry = 20'

OC - Gtry = 75'
alternate cable OC - PDU = 75'
lengths PDU - Gtry = 55'

voltage/amps/MDC 480V/125A
cat NO. E4502AE

B7999ZA or E4502KY
UPS
recommended ceiling
height 9'-0"
absolute min. ceiling
height 8'-0"

12'-2" X 20'-0"
absolute min. room (GT1700)
size 12'-2" X 22'-0"
(LIMITED ACCESS) (GT2000)

14'-0" X 20'-0" (GT1700)


14'-0" X 24'-
typical room size 0" (GT2000)

Delivery: min. 5ft. corr. = 78"


opening, incl. 6 ft. corr. = 66"
Dollies w/o side 7 ft. corr. = 56"
rails 8 ft. corr = 48"
CT QUICK REFERENCE - 15.Aug.11
LIGHTSPEED RT
gantry/table flr.
anchored? YES

gantry cable access REAR

conduit sizes -
OC to Gantry 3 1/2"
-OC to PDU - 2 1/2"
PDU to Gantry 2 1/2" and 3 1/2"

PDU & UPS


interference limits 39" to gantry and OC
gauss field OC = 10G
proximity limit Gantry = 1G

OC - Gtry = 60'
standard cable OC - PDU = 60'
lengths PDU - Gtry = 20'

OC - Gtry = 75'
alternate cable OC - PDU = 75'
lengths PDU - Gtry = 55'

voltage/amps/MDC 480V/125A
cat NO. E4502AE

B7999ZA or E4502KY
UPS
recommended ceiling
height 9'-0"
absolute min. ceiling
height 8'-0"

absolute min. room


size
(LIMITED ACCESS) 12'-2" X 19'-3"

typical room size 14'-0" X 19'-3"

Delivery: min. 5ft. corr. = 78"


opening, incl. 6 ft. corr. = 66"
Dollies w/o side 7 ft. corr. = 56"
rails 8 ft. corr = 48"
CT QUICK REFERENCE - 15.Aug.11

LIGHTSPEED 5x
SERIES LIGHTSPEED
16 & 16PRO
gantry/table flr.
anchored? YES

gantry cable access REAR

conduit sizes -
OC to Gantry 3 1/2"
-OC to PDU - 2 1/2"
PDU to Gantry 2 1/2" and 3 1/2"

PDU & UPS


interference limits 39" to gantry and OC
gauss field OC = 10G
proximity limit Gantry = 1G

OC - Gtry = 60'
standard cable OC - PDU = 60'
lengths PDU - Gtry = 20'

OC - Gtry = 75'
alternate cable OC - PDU = 75'
lengths PDU - Gtry = 55'

480V/90A
E4502AB (LS 16),
voltage/amps/MDC 480V/125A
cat NO. E4502AE (16PRO)

B7999ZA or E4502KY
UPS
recommended ceiling
height 9'-0"
absolute min. ceiling
height 8'-0"

absolute min. room


size 13'-6" X 20'-0"

typical room size 13'-6" X 20'-0"

Delivery: min. 5ft. corr. = 78"


opening, incl. 6 ft. corr. = 66"
Dollies w/o side 7 ft. corr. = 56"
rails 8 ft. corr = 48"
CT QUICK REFERENCE - 15.Aug.11

LIGHTSPEED
SERIES Qxi, PLUS,
ULTRA
gantry/table flr.
anchored? YES

gantry cable access REAR

conduit sizes -
OC to Gantry 3 1/2"
-OC to PDU - 2 1/2"
PDU to Gantry 2 1/2" and 3 1/2"

PDU & UPS


interference limits 39" to gantry and OC
gauss field OC = 10G
proximity limit Gantry = 1G

OC - Gtry = 60'
standard cable OC - PDU = 60'
lengths PDU - Gtry =20'

OC - Gtry = 75'
alternate cable OC - PDU = 75'
lengths PDU - Gtry = 55'
voltage/amps/MDC 480V/90A
cat NO. E4502AB
B7999ZA or
UPS E4502KY
recommended ceiling
height 9'-0"
absolute min. ceiling
height 8'-0"
absolute min. room
size 13'-6" X 19'-4"

Delivery: min. 5ft. corr. = 78"


opening, incl. 6 ft. corr. = 66"
Dollies w/o side 7 ft. corr. = 56"
rails 8 ft. corr = 48"
CT QUICK REFERENCE -15.Aug.11
HISPEED Qxi
gantry/table flr.
anchored? YES

gantry cable access REAR

conduit sizes - 3 1/2"


OC to Gantry 2 1/2"
-OC to PDU 2
-PDU to Gantry 1/2" and 3 1/2"

PDU & UPS 39" to gantry and


interference limits OC
gauss field OC = 1G
proximity limit Gantry = 1G
OC - Gtry = 75'
OC - PDU =
standard cable 75' PDU -
lengths Gtry = 55'
alternate cable
lengths NO
voltage/amps/MDC 480V/90A
cat NO. E4502AD
B7999ZA
UPS E4502KY
recommended
ceiling height 9'-0"
absolute min. ceiling
height 7'-6"
absolute min. room
size 13'-0" X 19'-4"

Delivery: min. 5ft. corr. = 78"


opening, incl. 6 ft. corr. = 66"
Dollies w/o side 7 ft. corr. = 56"
rails 8 ft. corr = 48"
CT QUICK REFERENCE - 15.Aug.11

YMS HISPEED
SERIES Xi, Xi PLUS,
Xi PRO, Nxi, Nxi
PLUS, Nxi PRO
gantry/table flr.
anchored? YES

gantry cable access REAR

conduit sizes
-OC to Gantry
-OC to PDU 2"
-PDU to 2"
Gantry (2) 2 1/2"

PDU & UPS 20" to gantry


interference limits 39" to OC

gauss field OC = 10G


proximity limit Gantry = 1G

OC - Gantry = 59'
standard cable OC - PDU = 59'
lengths PDU - Gantry = 59'
alternate cable
lengths NO
voltage/amps/MDC 480V/90A
cat NO. E4502AD

E4504AL (req's
UPS E4502AD)

recommended
ceiling height 8'-6"

absolute min.
ceiling height 7'-6"

absolute min. room


size 11'-6" X 16'-6"

Delivery: min. 5ft. corr. = 61"


opening, incl.
6 ft. corr. = 52"
Dollies w/o side
rails
Delivery: min.
opening, incl.
Dollies w/o side
rails 7 ft. corr. = 46"
8 ft. corr = 43"