GEHC Site Planning Technical Guide CT Quick Reference Guide - XLS - Rev5
GEHC Site Planning Technical Guide CT Quick Reference Guide - XLS - Rev5
GEHC Site Planning Technical Guide CT Quick Reference Guide - XLS - Rev5
11
BRIGHTSPEED
SELECT
EDGE, EXCEL, ELITE
&
ELITE ENANCED
gantry/table flr.
anchored? YES
conduit sizes -
OC to Gantry 3 1/2"
-OC to PDU - 2 1/2"
PDU to Gantry 2 1/2" and 3 1/2"
OC - Gtry = 60'
standard cable OC - PDU = 60'
lengths PDU - Gtry = 20'
OC - Gtry = 75'
alternate cable OC - PDU = 75'
lengths PDU - Gtry = 55'
voltage/amps/MDC 480V/90A
cat NO. E4502AB
B7999ZA or E4502KY
UPS
recommended ceiling
height 9'-0"
absolute min. ceiling
height 8'-0"
absolute min. room
size
(LIMITED ACCESS) 10'-10" X 19'-5"
BRIGHTSPEED ELITE,
EDGE & EXCEL
gantry/table flr.
anchored? YES
conduit sizes -
OC to Gantry 3 1/2"
-OC to PDU - 2 1/2"
PDU to Gantry 2 1/2" and 3 1/2"
OC - Gtry = 60'
standard cable OC - PDU = 60'
lengths PDU - Gtry = 20'
OC - Gtry = 75'
alternate cable OC - PDU = 75'
lengths PDU - Gtry = 55'
voltage/amps/MDC 480V/90A
cat NO. E4502AB
B7999ZA or E4502KY
UPS
recommended ceiling
height 9'-0"
absolute min. ceiling
height 8'-0"
absolute min. room
size
(LIMITED ACCESS) 11'-0" X 20'-0"
conduit sizes -
OC to Gantry 3 1/2"
-OC to PDU - 2 1/2"
PDU to Gantry 2 1/2" and 3 1/2"
OC - Gtry = 60'
standard cable OC - PDU = 60'
lengths PDU - Gtry = 20'
OC - Gtry = 75'
alternate cable OC - PDU = 75'
lengths PDU - Gtry = 55'
voltage/amps/MDC 480V/90A
cat NO. E4502AB
B7999ZA or E4502KY
UPS
recommended ceiling
height 9'-0"
absolute min. ceiling
height 8'-0"
absolute min. room
size
(LIMITED ACCESS) 12'-10" X 19'-4"
DISCOVERY CT750
HD LIGHTSPEED VCT
gantry/table flr.
anchored? YES
conduit sizes -
OC to Gantry 3 1/2"
-OC to PDU - 2 1/2"
PDU to Gantry 2 1/2" and 3 1/2"
OC - Gtry = 60'
standard cable OC - PDU = 60'
lengths PDU - Gtry = 20'
OC - Gtry = 75'
alternate cable OC - PDU = 75'
lengths PDU - Gtry = 55'
voltage/amps/MDC
cat NO. 480V/125A
11'-8" X 19'-2"
absolute min. room (GT1700)
size 11'-8" X 21'-2"
(LIMITED ACCESS) (GT2000)
13'-6" X 20'-0"
(GT1700)
13'-6" X 22'-0"
typical room size (GT2000)
DISCOVERY CT590
RT, OPTIMA CT580
LIGHTSPEED
RT16/XTRA
gantry/table flr.
anchored? YES
conduit sizes -
OC to Gantry 3 1/2"
-OC to PDU - 2 1/2"
PDU to Gantry 2 1/2" and 3 1/2"
OC - Gtry = 60'
standard cable OC - PDU = 60'
lengths PDU - Gtry = 20'
OC - Gtry = 75'
alternate cable OC - PDU = 75'
lengths PDU - Gtry = 55'
voltage/amps/MDC 480V/125A
cat NO. E4502AE
B7999ZA or E4502KY
UPS
recommended ceiling
height 9'-0"
absolute min. ceiling
height 8'-0"
12'-2" X 20'-0"
absolute min. room (GT1700)
size 12'-2" X 22'-0"
(LIMITED ACCESS) (GT2000)
conduit sizes -
OC to Gantry 3 1/2"
-OC to PDU - 2 1/2"
PDU to Gantry 2 1/2" and 3 1/2"
OC - Gtry = 60'
standard cable OC - PDU = 60'
lengths PDU - Gtry = 20'
OC - Gtry = 75'
alternate cable OC - PDU = 75'
lengths PDU - Gtry = 55'
voltage/amps/MDC 480V/125A
cat NO. E4502AE
B7999ZA or E4502KY
UPS
recommended ceiling
height 9'-0"
absolute min. ceiling
height 8'-0"
LIGHTSPEED 5x
SERIES LIGHTSPEED
16 & 16PRO
gantry/table flr.
anchored? YES
conduit sizes -
OC to Gantry 3 1/2"
-OC to PDU - 2 1/2"
PDU to Gantry 2 1/2" and 3 1/2"
OC - Gtry = 60'
standard cable OC - PDU = 60'
lengths PDU - Gtry = 20'
OC - Gtry = 75'
alternate cable OC - PDU = 75'
lengths PDU - Gtry = 55'
480V/90A
E4502AB (LS 16),
voltage/amps/MDC 480V/125A
cat NO. E4502AE (16PRO)
B7999ZA or E4502KY
UPS
recommended ceiling
height 9'-0"
absolute min. ceiling
height 8'-0"
LIGHTSPEED
SERIES Qxi, PLUS,
ULTRA
gantry/table flr.
anchored? YES
conduit sizes -
OC to Gantry 3 1/2"
-OC to PDU - 2 1/2"
PDU to Gantry 2 1/2" and 3 1/2"
OC - Gtry = 60'
standard cable OC - PDU = 60'
lengths PDU - Gtry =20'
OC - Gtry = 75'
alternate cable OC - PDU = 75'
lengths PDU - Gtry = 55'
voltage/amps/MDC 480V/90A
cat NO. E4502AB
B7999ZA or
UPS E4502KY
recommended ceiling
height 9'-0"
absolute min. ceiling
height 8'-0"
absolute min. room
size 13'-6" X 19'-4"
YMS HISPEED
SERIES Xi, Xi PLUS,
Xi PRO, Nxi, Nxi
PLUS, Nxi PRO
gantry/table flr.
anchored? YES
conduit sizes
-OC to Gantry
-OC to PDU 2"
-PDU to 2"
Gantry (2) 2 1/2"
OC - Gantry = 59'
standard cable OC - PDU = 59'
lengths PDU - Gantry = 59'
alternate cable
lengths NO
voltage/amps/MDC 480V/90A
cat NO. E4502AD
E4504AL (req's
UPS E4502AD)
recommended
ceiling height 8'-6"
absolute min.
ceiling height 7'-6"