1. SYSTEM OVERVIEW
1. 1. System Description
1.1.1. What is MIXEL?
Laser system MIXEL is composed of main body, handpiece, articulated arms, Foot switch, electrical accessories and
it is a carbon dioxide laser that uses fractional laser technology used as surgical tool for skin treatment requiring
cut or ablation.
● Material : CO2 LASER ● Laser Tube Type : All Metal sealed off RF CO2 Laser tube
● Spot Size : Max 20mm * 20mm ● Pulse Repetition Rate : 1Hz ~ 20Hz
g ) Protection
● Laser Output : Laser output does not occur when foot switch is pressed while it is not in ‘READY’ mode.
● Laser Over Current : Stops generating the laser if over-current flows more than 20% of rated voltage on laser diode.
● DSP Watchdog : Stops all the output if there are any abnormal operation in DSP or firmware.
● The protected circuit will start operating if over voltage current flows into circuit.
b. ARM
e. GOGGLES FOR
PRACTITIONER f. GOOGLES FOR PATIENT
a. MAIN BODY
⑨
②
⑩
③ ⑪
⑤
⑥
⑧
⑦
● Components and Description
NO Item DESCRIPTION (FUNCTION)
① Laser Output Arm Emits the energy generated from the body to the treatment area.
② Key Switch Power On / Off Key Switch
③ Emergency Button Stops the device immediately in an emergency situation.
④ LCD Control Panel This screen sets and displays the operation conditions.
⑤ Outlet for wind Product emits the heat through this outlet.
⑥ Interlock Switch It is a socket to connect Interlock switch.
⑦ Foot Switch Connector It is a socket to connect the foot switch.
⑧ Main Power Connector It is a socket to connect main power cord.
If you remove the sticker on the rear lock key and open it without
⑨ Rear Lock Key
permission from Hironic, you will be excluded from warranty. You can
⑩ Handle It is used when moving the product.
⑪ Handpiece Holder Hang the handpiece up on the holder
1.2.2.2. Handpiece
①
Symbol Definition
Dangers alert the user to information that is of the highest importance and vital to the
safety of the user and might cause severe damage to the device
[DANGER]
Warnings alert the user to information that is of importance and vital to the safety of
[WARNING] the user and might cause minor damage to the device
References alert the user to information that is vital and important to safely operate
IEC417/878-02-02 Class B
Dangerous voltage
IEC417/878-03-01
Power Off
IEC417/5008
Power On
IEC417/5007
Earth Grounding
IEC417/5021
Keep electrical waste separate from municipal waste and ask to Hrionic to handle.
This mark is accordance with EU Waste Electrical and Electronic Equipment
Directive (Directive 2002/96/EC &EN50419)
2. INSTALLATION
● Install the device at a maximum distance from other medical equipments generating high heat.
● The best safekeeping temperature of the device is 5~30℃ and the best usage temperature is 14~24℃.
● Do not connect the device to multiple power socket with other devices. It might cause fire or explosion due to
abnormal heat.
● Use the power outlet having ground terminal. It might cause electrical effect leading to malfunction.
[DANGER]
· In case of not following electrical requirements such as inappropriate usage or wrong voltage, the device
might be damaged or cause malfunction that could give fatal damages to human.
2.1.2. To relocate the device from the initial installation, please observe the following
● Disconnect cables prior to moving
● The handpiece should be firmly fixed to the holder to prevent falling down
[Warning]
· While moving, please do not give shock to the device
● Ensure that the power cord is connected to the socket properly. Do not handle the device with wet hands
because of the risk of electric shock.
● Key switch must be kept confidentially away from others to prevent using.
● Do not expose lamp light of the handpiece to reflective products [e.g. mirrors, jewelry and watches]. If the
lamp light is reflected and goes into the naked eyes, it might cause damage to the eyes.
● If the fuse is broken or the device do not power on because of overvoltage, exchange fuse on the lower rear
panel of the main body.
● If malfunction occurs during use, please contact customer service center of the local retailer immediately.
Only the authorized and trained personnel should check and fix the problem.
● Insert the key switch and turn the key clockwise to turn on.
● When turning off, go to Standby mode and turn the key switch to counterclockwise direction to turn off.
[DANGER]
·It might cause malfunction if the power is turned OFF at Ready condition.
·When restarting the system, take at least 5 minutes after turning off the power for the stability of air cooling.
[Reference]
· The screen does not display any messages when key switch is turning OFF.
· If the power does not turn on when the power cable is properly connected to the socket, please check if
the emergency stop switch is pushed in. If it is pressed in, rotate it to clockwise direction and ensure that
the switch is protruded in proper position.
① ⑩
⑥
②
⑤
⑦
③
⑧
④
⑮ 16 ⑪ ⑭
a) The main function of each button is as follows.
Class. Range Description
① POWER 1 ~ 60 mJ(1 step) Sets the laser energy
0.1 ~ 2.0 mm
② DENSITY Adjusts interval of dots.
(0.1 step)
Single – When foot S/W is pushed, the laser shoots once.
Single ~ 2.0 sec
③ REPEAT 0.5 ~ 2.0 – Continuously shoots the laser energy with regular interval with
(0.5 step)
pre-set value when pushing the foot switch.
④ OVER LAP 1 ~ 10 (1 step) Sets the number of laser emission per one each dot.
2*2 ~ 20*20 mm
⑤ SCAN FIELD Sets the size of laser emitting area
(1 step)
⑦ SHAPE SQUARE, CIRCLE Selects the shape of laser emitting area ( Square or Circle )
b) CO2 Mode
100~1,000us
④ ON TIME In Pulse laser mode, set the emitting time for one-time shooting.
(20 step)
1~10ms(1step)
10~100ms(10step)
⑤ OFF TIME In Pulse laser mode, set the standby time for the next shooting.
100~1,000ms
(100step)
3.2. Operation
For detailed operation method, please refer to the appendix Protocol.
3.3. Notice
3.3.1. Safety checks before use
● Update patient’s treatment record information prior to treatment.
● Based on a clinical treatment guideline and patient’s condition, set a proper parameter.
● Check the patient skin condition for the laser treatment. Hairs after shaving, any black substances, lotion,
perfumes, sun blocks, and local anesthetics should not be left.
● The device should always be kept in ‘Stand by’ mode if it is not operated.
● Confirm the existence of contaminants on the emitting area and remove it by using gauze and alcohol.
● Not allowed to use acne medicine and bleaching or retinol containing products from 1 day before treatment.
● Control the entrance during operation, as laser is not seen through eyes and might give severe effect.
● As laser pointer emits very strong energy, do not emit when touching this area.
● Do not look at the laser pointer when laser emits in Ready mode.
● Please contact Hironic or local agency for safe delivery when you move the device to far distance.
Problem Troubleshooting
· Turn off the power by pressing the emergency stop switch or turning off
the key switch
· Disconnect the power cord from the socket
When burning smell comes
· Contact a local agency or Hironic C/S center
out from the device
[Warning] If you use the device continuously, this might cause
damage to module inside that leads to high risk of fire and other
safety threats
· Turn off the power by pressing the emergency stop switch or turning off
the key switch
· Disconnect the power cord from the socket
When explosion sound comes
· Contact a local agency or Hironic C/S center
out from the device
[Warning] If you use the device continuously, this might cause
damage to module inside that leads to high risk of fire and other
safety threats.
screen is broken
[Warning] If you use the device continuously when the glass of LCD
touch screen is broken, emitting energy will be irregular causing
damage to skin.
Problem Troubleshooting
· When power does not turn OFF when manipulating Key switch during
product operation, push the emergency stop switch immediately to turn
OFF the power.
When power key switch does ∙ After disconnecting the power cord from the plug, rotate the key switch
not work left to right for 2-3 times and check the operating condition.
∙ Plug in the power cord and turn ON the power to check the operation.
∙ If the same situation occurs after repeating 2-3 times, please contact a
local agency or Hironic C/S center.
∙ In case of power turning off when handpiece emits a shot during use
Symptoms with lack of
∙ In case of touch screen operation error when handpiece emits a shot
power supply when using the
during use
product
∙ In case of often occurrence in fuse disconnection
6. GUARANTEE
6.1. Outline
The following explains about the gratuitous guarantee when operating device properly and limited guarantee
matters during the warranty period.
[DANGER]
· For the gratuitous guarantee, any change or repair should be accomplished by an authorized
service person by Hironic. If the device is used by unintentional purpose or users do not
observe this manual, then users will be excluded from gratuitous guarantee.
· Hironic has the responsibility and rights to judge products’ troubleshooting and those causes.
● When the handpiece connection parts got damaged due to hard pulling or twisting
● When the user opens, dismantles or changes the device without approval from Hironic. And when the user
use the device for unintentional purposes
● Consumables which maintain the device
● When the device is broken because of the user’s misuse or fault
- In case the handpiece is broken because it is dropped on the floor
- In case the handpiece is broken by excessive force at both ends of the hose where connected to the body
● When anyone who is non-certified service personnel decompose or repair the device arbitrarily
● When the seal sticker at rear panel is removed without the approval from Hironic.
PRODUCT OVERVIEW
Letter of Guarantee
The seller does not have any responsibility for guaranteeing marketability or condition of the device. This letter of
Hospital Name
Contact Person
Address
Office Telephone
Cell Phone
Installation Date ●
Guarantee ●
Expiration Date
Seller Name /
Company
Seller Telephone
Others
Important!
For the guarantee, please send above content to the following address by email or fax
PRODUCT OVERVIEW
MEMO
Tel : +82-31-698-4900
Fax: +82-31-777-9713
Homepage : www.hironic.com
E-mail : service@hironic.com