Monnal t75 Air Liquide Ventilator PDF

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 10

DATA SHEET

www.airliquidemedicalsystems.com
www.airliquide.com
[1] Monnal T75 / Air Liquide Medical Systems

CARACTERISTIQUES TECHNIQUES

APPLICATION AREA OPERATING CONDITIONS


___________________________________________ _________________________________________

Intensive care respirator designed for Operating


+10 to +40°C (+50 to +104°F)
Temperature:
Application: resuscitation, intensive care and the
postoperative room
Relative 0 to 90 % (at 40°C without
humidity: condensation)
Patient
Adult, Child and Infant.
categories: 700 (3075 mètres) to 1060 mbar
Atmospheric
pressure: Altitude effect compensation
Weight: 3 to 250 kg

STORAGE CONDITIONS
STANDARDS AND DIRECTIVES
_________________________________________
___________________________________________
Storage
CE 0459. The device complies with temperature: -20 to +70°C (-4 to +158°F)
the European requirements in the
CE Mark:
Directive 93/42/EEC concerning
Storage relative
medical devices.
humidity: 0 to 90 % (at 40°C without condensation)

Class IIb, according to Directive Storage


Medical classification:
93/42/CEE. atmospheric 500 (5550 meters) to 1060 mbar
pressure:
ISO 14971
Standards: CEI 60601-1
CEI 60601-2-12

Electromagnetic
compatibility Complying with CEI 60601-1-2
(2001)
(EMC):

France: end of 2008


First launch:
Other countries: please contact us

French, English, Italian, German,


Available Russian, Spanish, Lithuania, Polish,
languages: Czech,Portuguese, Chinese, Japanese
Other language: please contact us

PHYSICAL SPECIFICATIONS
___________________________________________

Size: H 35 x W 30 x D 40 cm

Ventilator with
H 135 x W 55 x D 60 cm
trolley:

16 kg - Ventilator
Weight:
31,8 kg - Ventilator + trolley

Noise level: 48 dB (A) at 1m


[2] Monnal T75 / Air Liquide Medical Systems

TECHNICAL SPECIFICATIONS

POWER SUPPLY PNEUMATIC SUPPLY


___________________________________________ _________________________________________

Oxygen pneumatic supply


Power supply: 100-240 VAC ± 5 %

Frequency: 50 – 60 Hz 2.8 – 6 bar / 280 – 600 kPa / 40 –


86 psi
High pressure:
Operating possible starting from 1.5
Electric power bar
250 VA
consumed:

Connection NF, NIST, DISS


Electric class: I
standards For other standards, please contact
available: us
Type: B

0 – 1.5 bar / 0 – 150 kPa / 0 – 21


Low pressure:
Connection EU, AFS, JP, US, AR, BR psi
standards Please contact us for more information
available: concerning other standards Air pneumatic supply

Turbine peak flow: 200 l/mn at Patm


Continuous
20 to 30 V DC
external supply:
Equipped with HEPA filter Monnal
Ambiant air inlet: Clean’In (High Efficiency Particulate
EXTERNAL BATTERY: Air Filter)
If no mains are available, switch to external battery

PATIENT SYSTEM GAS CONNECTOR


Type: Rechargeable, NiMH, 24V, 2x4500 mAh ________________________________________

Autonomy: 2,5 to 3 hours in standard ventilation


Inspiratory
breathing tube 22 mm male.
Charge time: 12 hours
connector:

-Interval between It is advised to perform one complete


two recharges: discharge cycle every 6 months
Expiratory
breathing tube 22 mm male.
INTERNAL BATTERY: connector:
If neither the mains nor the external source is available,
switch to internal battery
SCREEN
________________________________________
Type: Rechargeable, NiMH, 24V, 2x4500 mAh
Flat color touch screen
Type:
Autonomy: 2,5 to 3 hours in standard ventilation TFT-LCD Module

Charge time: 12 hours Technology: Resistive analogical technology

Size: 10.4 inches


-Intervalbetween It is advised to perform one complete
two recharges: discharge cycle every 6 months
Resolution: 640 x 480 pixels
[3] Monnal T75 / Air Liquide Medical Systems

TECHNICAL SPECIFICATIONS

MODES – INVASIVE VENTILATION


PENUMATIC SPECIFICATIONS ___________________________________________
___________________________________________
Volume Controlled Ventilation or Assist
VCV
volume controlled ventilation

Method of
Flow and pressure Pressure Controlled Ventilation or Assist
triggering: PCV
pressure controlled ventilation

Pressure Regulated Volume Controlled


PRVC
Max. Operating ventilation
90 cmH2O
pressure:
Synchronized Intermittent Mandatory
SIMV
Ventilation
Bias flow: 3 L/min (flow by)
Pressure Synchronized Intermittent
PSIMV
Mandatory Ventilation

Max. Spontaneous ventilation with Pressure


Inspiratory 180 L/min (regarding supply pressures) PSV Support and PEEP
peak flow:
CPAP Continuous Positive Airway Pressure

Spontaneous ventilation with Pressure


PS-Pro
Support, PEEP and variable frequencies
INSPIRATORY CHANNEL
___________________________________________ Duo-Levels Alternation between two levels of CPAP

Apnea ventilation (VT, RR and T apnea


Safety
can be adjusted, constant flow rate,
Max. 4 cm H2O at a flow rate of 1L/s ventilation
Pressure drop: Ti.Ttot 33%)
(without filter)

MODES – NON INVASIVE VENTILATION


___________________________________________
Gaz supplier Turbine and proportional solenoid valve
system: managed by microprocessor Non-invasive spontaneous ventilation
PSV-VNI
with Pressure Support and PEEP

CPAP Continuous Positive Airway Pressure


Inspiratory flow
0 to 3,3 L/s
range:
Duo-Levels Alternation between two levels of CPAP

OXYGENOTHERAPY
___________________________________________
EXPIRATORY CHANNEL
___________________________________________
Flow Adult: 2 to 80 L/min

Child: 2 to 60 L/min
Pressure drop: Max. 3 cm H2O at a flow rate of 1L/s
Infant: 2 to 60 L/min

FiO2 21% to 100%


Expiratory flow
0 to 3,3 L/s
range: This function requires specific nasal cannula and use a
heating humidifier.
[4] Monnal T75 / Air Liquide Medical Systems

TECHNICAL SPECIFICATIONS

PARAMETER SETTINGS PARAMETER SETTINGS (NEXT)


___________________________________________ ___________________________________________

PARAMETERS: SETTINGS RANGE: PARAMETERS: SETTINGS RANGE:


______________________ ________________________ ______________________ ________________________

Inspiratory plateau (Tplat): 0 to 60 % of TI


Inspiratory tidal volume
20 to 2000 mL
(VT): Adult: 0 to 60 % of TI

Child: 0 to 40 % of TI
Adult: 100 to 2000 mL

Infant: 0 to 40 % of TI
Child: 50 to 500 mL

Infant: 20 to 75 mL Inspiratory ratio (TI/Ttot): 3 to 80 %

Respiratory Rate (RR): 4 to 120 Bpm Adult: 3 to 80 %

Adult: 4 to 80 Bpm Child: 5 to 60 %

Child: 5 to 120 Bpm


Infant: 5 to 60 %

Infant: 10 to 120 Bpm


Inspiratory time (I/E): 1/0,5 (=2) to 1/19

SIMV respiratory rate


4 to 120 Bpm Adult: 1/0,5 to 1/19
(SIMV RR):

Child: 1/0,5 to 1/19


Adult: 4 to 80 Bpm

Infant: 1/0,7 to 1/19


Child: 5 to 120 Bpm

Infant: 10 to 120 Bpm Inspiratory time (Ti): 0,2 to 10 s

Minimum frequency: 1 to 100 Bpm Adult: 0,2 to 10 s

Child: 0,2 to 8 s
Adult: 4 to 80 Bpm

Infant: 0,2 to 3 s
Child: 5 to 100 Bpm

Infant: 10 to 100 Bpm


Max. inspiratory time in
spontaneous modes (TI 0,2 to 5 s
Apnea time: Adjustable VCV mode max.):

Adult: 15 to 60 s
Adult: 0,3 to 5 s

Child: 4 to 60 s
Child: 0,3 to 5 s

Infant: 2 to 30 s
Infant: 0,2 to 2,5 s
[5] Monnal T75 / Air Liquide Medical Systems

TECHNICAL SPECIFICATIONS

PARAMETERS SETTINGS: SPECIAL FEATURES


PARAMETER SETTINGS (NEXT) ________________________________________________
___________________________________________

PARAMETERS: SETTING RANGE: NON INVASIVE VENTILATION:


______________________ ________________________

Positive End of Expiration


0 to 15 cmH2O
Pressure (PEEP):
Positive End of Expiration
0 to 50 cmH2O Pressure Support (PS): 2 to 25 cmH2O
Pressure (PEEP):
Low VMe alarm: OFF position available
Pressure Support (PS): 2 to 40 cmH2O

Insufflation Pressure (PI): 2 to 99 cmH20


PRCV MODE:
Max. airways pressure: 90 cmH20

Inspiratory slope: 20 to 200 cmH2O / s OFF to maximum of each


Target Tidal volume:
patient category
Flow rate shape: Constant, decelerated
Insufflation pressure (PI): AUTO, 2 to 99 cmH2O
O2 concentration (FiO2): 21 to 100 %
Max. Insufflation pressure (PI
2 to 99 cmH2O
max):
1 to 10 L/min
Inspiratory flow trigger: OFF position available in VCV, PCV
and PRCV modes
PS-PRO MODE:

Always activated and


Inspiratory pressure trigger: automatically set in function of Support frequency
Adult: 4 to 99 Bpm
the flow trigger setting (RRsupport)

Child: 5 to 120 Bpm

0 (OFF) to 90 % of inspiratory Infant: 5 to 120 Bpm


Expiratory trigger:
peak flow
Pressure Support (PS) : AUTO, 2 to 40 cmH2O

Adult: 5 to 90% Max. Insufflation pressure (PI


2 to 99 cmH2O
max):
Child: 0 to 90%
OFF to maximum of each
Inspiratory plateau: 0 to 15 s Target Tidal volume :
patient category
Expiratory plateau: 0 to 15 s O2 : ASPIRATION
INTELLIGENTE
Sigh frequency: 1 sigh every 9 to 200 cycles
Durée de pré-oxygénation 30 à 900 s
Sigh Tidal volume: 1 to 2 per VT
Durée de post-oxygénation 30 à 300 s
Sigh insufflation pressure: 1 to 2 per PI
Cible FiO2 21% à 100%
Infant: 0 to 90%

TC: TUBE COMPENSATION


O2: INTELLIGENT SUCTION:

Applicable on: Pressure modes only


Pre-oxygénation time: 30 to 900 s
Compensation level: 0 to 100%
Post-oxygénation time: 30 to 300 s
Tube type: Endotracheal or tracheostomy
FiO2 target 21% to 100% Tube diameter: 2.5 to 11 mm
[6] Monnal T75 / Air Liquide Medical Systems

TECHNICAL SPECIFICATIONS

MONITORING LOOPS AND WAVEFORMS PRESENTATION


___________________________________________ ___________________________________________

Pressure curve
Real-time curves:
MEASURED MEASURE TREND Flow curve
PARAMETERS: RANGE: VALUE*:
______________________ ____________ ___________ Loop curves: Pressure / Volume
Volume / Flow

Expired Minute Volume (VMe) 0 to 99 L/min Yes Flow / Pressure

Expired Tidal Volume(Vte) 0 to 5000 mL Yes CONFIGURATION


___________________________________________
Insufflated Tidal Volume
0 to 5000 mL Yes
(VTi)
Selection of Enabled, Disabled (except VCV always
Respiratory Rate (RR) 0 to 120 Bpm Yes ventilation mode: enabled)

Peak Airways pressure (Ppeak) 0 to 120 cmH2O Yes Alarms: Save alarms current settings

Save current ventilation settings in the


Settings:
Positive End of Expiration ventilation mode used
0 to 99 cmH2O Yes
Pressure (PEEP):
Alarm volume
25% to 100% of custom alarm
Plateau pressure (Pplat) ** 0 to 99 cmH2O Yes setting:

FiO2 15 to 100% Yes Min. frequency: Enabled, Disabled (only in PS and NIV-PS)

Mean pressure 0 to 99 cmH2O Yes Setting in VCV


Ti, Flow, TI/Ttot or I/E
mode:
Ti / Ttot 0 to 99% Yes
Setting in PCV /
Ti, TI/Ttot or I/E
RR / Vte 0 to 5000 pm/L Yes PRCV mode:

MVe Spont 0 to 99 L/min Yes


EVENT HISTORY
RR Spont 0 to 120 Bpm Yes
___________________________________________
Insp Peak Flow ** 0 to 200 L/min Yes
Alarms history: 200 last triggered alarms recorded
Exp Peak Flow 0 to 200 L/min Yes
Technical messages with error codes
Leak Flow (in VNI) 0 to 150 L/min Yes

Spontaneous ratio -- to 100 Yes COMMUNICATION / INTERFACE


___________________________________________
Leak ratio -- to 100 Yes

P0.1 ** 0 to 20 Yes Serial port: RS 232 port (X 2)


OTP: Open Taema Protocol
0 to 500
Rstat ** Yes Video output: Screen report to external monitor.
cmH2O/(L/s)
BOW MEDICAL Communication with DIANE anaesthesia
Interface: sheet
0 to 150
Cstat ** Yes
mL/ cmH2O DATACAPTOR: Capsuleteck

CLINISOFT Communication with the GE data collection


Auto-PEEP ** 0 to 99 cmH2O Yes
interface: software

* Stored trend values for up to 80 hours


** Only in Invasive Ventilation
[7] Monnal T75 / Air Liquide Medical Systems

TECHNICAL SPECIFICATIONS

ALARMS SETTINGS ALARMS SYSTEMS


___________________________________________ ___________________________________________

Ppeak high 10 to 100 cmH2O Visual alarms 3 priority levels with specific colours
classified by
Pplat high OFF, 1 to 50 cmH2O priority: 1 information level

RR low Adult: 1 to 70 Bpm


Sound alarms
Child: 1 to 110 Bpm classified by 3 priority levels with specific melodies
priority:
Infant: 1 to 110 Bpm
Technical alarms: Categorized by severity
RR high Adult: 11 to 80 Bpm
Supply alarms: O2 supply failure
Child: 11 to 120 Bpm
Switch over to external battery
Infant: 11 to 120 Bpm Switch over to internal battery
Vti low Adult: OFF to 2900 mL Internal battery power low
Internal battery exhausted
Child: OFF to 1900 mL
Internal battery absent
Infant: OFF to 500 mL
Alarm indicating total loss of electrical
Vti high Adult: 100 to 3000 mL power
Child: 100 to 2000 mL Automatic
Direct access key
thresholds:
Infant: 100 to 800 mL
Sound inhibition: 2 minutes
Vte low Adult: OFF to 2000 mL
Apnea ventilation: Adjustable VCV mode
Child: OFF to 2000 mL
Non-adjustable
Infant: OFF to 500 mL Disconnection
patient alarms:

Vte high Adult: 10 to 3000 mL PEEP > PEEP set-point + 5 cmH2O

Child: 10 to 3000 mL Active patient. Increase inspiratory flow


rate
Infant: 10 to 800 mL

VMe low 0,5 to 39 L/mn

VMe high Adult: 1,5 to 40 L/mn


CLEANING AND STERILIZATION
Child: 1,5 to 40 L/mn ___________________________________________
Infant: 1 to 40 L/mn
Autoclavable: Reusable expiratory valve
FiO2 low 18 to 95 % Patient circuit
FiO2 high 23 to 100 % Reusable spirometry sensor
Single use: Single use expiratory valve
Patient circuit
ALARMS SETTINGS: SPECIAL FEATURES OF NIV Single use spirometry sensor
___________________________________________
Turbine inlet: « Monnal Clean’in » filter (quaterly
inspection)
VMe low OFF to 39 L/mn

VMe high Adult: 1,5 to 60 L/min


MAINTENANCE
Child: 1,1 to 60 L/min
___________________________________________
Infant: 1 to 40 L/min
Periodic
VTi Low and High not measured Annual, performance check
maintenance:
[8] Monnal T75 / Air Liquide Medical Systems

CONSUMABLES & ACCESSORIES

LIST OF CONSUMABLES LIST OF ACCESSORIES


___________________________________________ ___________________________________________
Oxygen tank hook (maximum height: 80 cm) KY665000
Autoclavable expiratory flow sensor KY632200
External battery (plus 3 hours battery life) YR109900
Autoclavable expiratory valve Monnal EVA (per Autoclavable patient circuit plug (3 parts: plug,
KY694500 KY669800
unit) screw, and clip) included in the stand

Group of 5 single-use expiratory valves Monnal Fast-locking patient circuit arm, no clamps
KY694800 (30+43+43+27 cm) Mounted on the plate of the KB024200
EVA
stand (for reference KB027500)
Group of 20 single-use expiratory valves Monnal
KY694900 Fast-locking patient circuit arm, with clamps
EVA (10+33+33+27 cm) To be mounted on a rail KB019200
(adapted for reference KB022600)
Oxygen cell (electrochemical O2 sensor) YR049700
Oxygen filtrabloc KB002800 Patient circuit arm with three locking joints, no
KB005200
claims (8+43+43+55 cm)
Monnal Clean’In filter (HEPA filter, per-unit) KB030100
Mains power cord, 3 m long, EU standard KY120500
Mains power cord, 3 m long, AFS standard KY647300
Group of 20 single-use adult-patient circuits, 1.6
Side hook alone with pincers (max 3 kg) KY669600
meter without any water-trap, but with a smooth
inside surface KG020200 Cable/hose winder (rear-fastening) KY669700
Can be connected to a pneumatic nebulizer O2 supply hose (NF plugs) 1.5 m long BF030600
O2 supply hose (NF plugs) 3 m long BF030200 BF030200
Group of 10 single-use adult-patient circuits, 1.6 O2 supply hose (NF plugs) 5 m long BF030000
KG019300
meter with inspiratory/expiratory water trap Monnal T75 cover (dimensions: 80x60x46 cm) KB028500
Clamp for 3-positions arm mounting rail (for
KY252800
Group of 12 single-use child-patient circuits, 1.6 KB005200)
meter with inspiratory/expiratory water trap KG019400 Rear-fastening stainless steel basket
Can be connected to a pneumatic nebulizer KY664700
(dimensions: H11xL30xP17 cm)
Stand alone (3 parts: platform, column, and 4-
KY659700
wheel base, 2 wheels having latches)
Group of 15 single-use adult-patient circuits, 1,6
meter with a water-trap on the inspiratory circuit Infusion pole (106 cm, can be lowered) KB029600
and a corrugated surface expiratory circuit VD315100
Base with 3 additional outlets (EU standard with
Cannot be connected to a pneumatic nebulizer YR110000
fastening)
Side tablet with pincers and hook (max 3 kg,
KY669500
dimensions 32x23 cm)
Group of 10 single-use child-patient circuits 1.5
meter With a water-trap on the inspiratory circuit Wall tablet (47.5 x 45 x 12 cm) KY690500
and a corrugated surface expiratory surface VD317600
Cannot be connected to a pneumatic nebulizer PNEUMATIC NEBULIZERS
___________________________________________
Group of 5 single-use expiratory flow sensors KY664500
Group of 20 single-use pneumatic nebulizer KB025200
Group of 20 single-use expiratory flow sensors KY664600
Full kit autoclavable pneumatic nebulizer KB007200
Group of 6 air intake filters KY650300
Group of 50 bacteria filters (inspiratory output) KV103300 ULTRASONIC NEBULIZERS
Membrane Monnal EVA x 5 KY665300 ___________________________________________

Air intake filter mount KY652700 International autoclavable Aeroneb nebulizer


KB025500
(English)
Autoclavable Aeroneb nebulizer, Switzerland KB029100
HUMIDIFIERS Autoclavable Aeroneb nebulizer, France KB028600
___________________________________________
Autoclavable Aeroneb nebulizer, Italy KB028700

Heating humidifier MR850 230V EU, FR / ES / EN VD324500 Autoclavable Aeroneb nebulizer, Latin America KB029000
Autoclavable Aeroneb nebulizer, Portugal KB028900
Heating humidifier MR850 115V US, PT / ES / EN VD324600
Autoclavable Aeroneb nebulizer, Spain KB028800
Heating humidifier MR850 230V EU, IT / ES / PT VD324700
Single-use Aeroneb nebulizer, FR/DE/NL/IT KB029300
Autoclavable humidifier adult accessories kit VD324900 Single-use Aeroneb nebulizer, International KB029200
Pediatric humidifier accessories kit VD324800 Single-use Aeroneb nebulizer ES/IT/PT/EL/TR KB029400
Air Liquide Medical Systems reserves the right to modify the design and the specifications contained herein without prior notice.