1133010I Rev. 02

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 2

HDL-CHOLESTEROL

REF 1133010
2 x 40 mL HDL-CHOLESTEROL
CONTENTS
DIFFERENTIAL PRECIPITATION
Enzymatic colorimetric test
R1. Reagent 2 x 40 mL
CAL. Standard 1 x 3 mL ENDPOINT
For in vitro diagnostic use only

PRINCIPLE INTERFERENCES

This technique1 uses a separation method based on the selective  Lipemia (triglycerides 10 g/L) does not interfere.
precipitation of apoliprotein B-containing lipoproteins (VLDL, LDL  Bilirubin (10 mg/dL), hemoglobin (5 g/L), may affect the results.
and (a)Lpa) by phosphotungstic acid/MgCl2, sedimentation of the
precipitant by centrifugation, and subsequent enzymatic analysis of  Other drugs and substances may interfere3.
high density lipoproteins (HDL) as residual cholesterol remaining in
the clear supernatant.
MATERIALS REQUIRED

REAGENT COMPOSITION I. Precipitation

R1 Precipitating reagent. Phosphotungstic acid 0.63 mmol/L,  Dilutor and pipettes.


magnesium chloride 25 mmol/L. Stabilizers.  Centrifuge tubes (13 x 100 m/m).
 Vortex mixer.
CAL Cholesterol standard. Cholesterol 50 mg/dL 1.3 mmol/L).  Desktop centrifuge.
Organic matrix based primary standard. Concentration value
is traceable to Standard Reference Material 1951a.
II . Colorimetry
R2 Cholesterol MR. Optative. Ref: 1118005, 1118010,
1118015.
 Kit for measuring Total Cholesterol.
 Constant temperature incubator set at 37ºC.
STORAGE AND STABILITY  Photometer or colorimeter capable of measuring absorbance at
 500  10 nm.
Store at 2-8ºC.
All the kit compounds are stable until the expiry date stated on the PROCEDURE
label.
Discard If appear signs of deterioration:
- Presence of particles and turbidity. I. Precipitation
- Blank absorbance (A) at 500 nm > 0.100 in 1cm cuvette.
1. Bring reagents and samples to room temperature.
2. Pipette into labelled centrifuge tubes:
REAGENT PREPARATION

The Reagents and Standard are ready-to-use. Store the vials tightly
Sample 1
closed, protected from light and prevented contaminations during Sample or Standard 0.2 mL Ratio =
the use. Reagent 2

Precipitating reagent 0.4 mL Dil. factor = 3


SAMPLES

Serum .EDTA or heparinized plasma, free of hemolysis. Obtained 3. Vortex and allow to stand for 10 minutes at room temperature.
from the patient after an overnight fast. Remove from cells within 3 4. Centrifuge for 10 minutes at 4000 r.p.m., or 2 minutes at 12000
hours of venipuncture. Samples may be kept at 4-8ºC for 2 weeks, r.p.m.
and at –20ºC for 3 months with no alteration of HDL cholesterol. 5. Separate off the clear supernatant within 2 hours.
The supernate containing the HDL fraction is conveniently
prepared on the day of sample collection and may be analysed In case of turbid supernatants caused by elevated triglycerides
after 2 weeks at 4-8ºC or 3 months at –20ºC in a non-selfdefrosting (350 g/dL) the sample should be diluted 1:2 with saline and steps
freezer.2 2,3,4 and 5 repeated. Multiply the result of the colorimetry by 2.

QUALITY SYSTEM CERTIFIED LINEAR CHEMICALS, S.L.U. Joaquim Costa 18 2ª planta. 08390 Montgat (Barcelona) SPAIN
ISO 9001 ISO 13485 Telf. (+34) 934 694 990; E-mail: info@linear.es ; website: www.linear.es NIF-VAT:B60485687
CLINICAL SIGNIFICANCE
II. Colorimetry
1. Bring the Cholesterol MR Monoreagent and the cholesterol Low HDL-cholesterol is a strong independent predictor of coronary
standard (50 mg/dL) of the kit to room temperature heart disease. In ATP III4, low HDL cholesterol is defined
2. Pipette into labelled tubes: categorically as a level  40 mg/dL (1.04 mmol/L), a change from
the level of  35 mg/dL in ATPII (1993).
Low HDL cholesterol is used as a risk factor to estimate 10-year
Sample Standard risk for coronary heart disease, having several causes: elevated
TUBES Blank
Supernat Supernat triglycerides, overweigh and obesity, physical inactivity, and type 2
diabetes. Other causes are, cigarrete smoking, very high
Monoreagent 1.0 mL 1.0 mL 1.0 mL carbohydrate intakes ( 60% of calories), and certain drugs as
anabolic steroids and progestional agents.
Supernate  50 L 
Standard   50 L
ANALYTICAL PERFORMANCE

3. Mix and let the tubes stand for 10 minutes at room temperature - Detection limit: 1.2 mg/dL
or 5 minutes at 37ºC.
4. Read the absorbance (A) of the supernatant and the standard - Linearity: Up to 200 mg/dL
at 500 nm against the reagent blank. - Precision:
mg/dL Within-run Between-run
The color is stable for at least 30 minutes protected from light. Mean 42.1 45.8 54.6 42.1 45.8 54.6
SD 0.23 0.23 0.2 0.27 0.28 0.31

CALCULATIONS CV% 0.54 0.5 0.34 0.64 0.61 0.52


N 10 10 10 10 10 10

A Supernatant - Sensitivity: 2,5 mA/mg/dL HDL


x C Standard = mg/dL HDL-Cholesterol - Correlation: This assay (y) was compared with a similar commercia
A Standard method (x). The results were:
N = 25 r = 0.995 y = 0.985 + 2.6
If results are to be expressed as SI units apply:
mg/dL x 0.0259 = mmol/L The analytical performances have been generated using on automatic
instrument. Results may vary depending on the instrument.

REFERENCE VALUES4
NOTES

Clinical values of HDL-Cholesterol used to classify risk groups. 1. This method may be used with different instruments. Any
application to an instrument should be validated to demonstrate
Cholesterol from lipoproteins of high density RISK that results meets the performance characteristics of the
method. It is recommended to validate periodically the
 55 mg/dL (  1.42 mmol/L) Low instrument. Contact to the distributor for any question on the
Men 35-55 mg/dL (0.90-1.42 mmol/L) Moderate application method.
 40 mg/dL ( 1.04 mmol/L) High 2. Clinical diagnosis should not be made on findings of a single test
result, but should integrate both clinical and laboratory data.
 65 mg/dL (  1.68 mmol/L) Low
Women 45-65 mg/dL (1.16-1.68 mmol/L) Moderate
REFERENCES
 45 mg/dL (  1.16 mmol/L) High

1. Burstein, M., Scholnick, H.R. and Morfin, R. Scand. J. Clin.


QUALITY CONTROL Lab. Invest. 40 : 560 (1980).
2. Finley, P.R., Shifman, R.B., Williams, R.S. and Lichti, D.I. Clin.
The use of a standard to calculate results allows to obtain an Chem. 24 : 931 (1971).
accuracy independent of the system or instrument used. 3. Tietz. N.W. Clinical Guide to Laboratory Tests, 3rd Edition.
To ensure adequate quality control (QC) each run should include a W.B. Saunders Co. Philadelphia, PA. (1995).
set of controls (normal and abnormal) with assayed values handled 4. SPECIAL REPORT. Executive Summary of the Third Report
as unknowns. of the National Cholesterol Education Program (NCEP) Expert
If the values are found outside of the defined range, check the Panel on Detection, Evaluation, and Treatment of High Blood
instrument, reagents and procedure. Cholesterol in Adults (Adult Treatment Panel III). JAMA. 285 :
Each laboratory should establish its own Quality Control scheme 2486 (2001).
and corrective actions if controls do not meet the acceptable Young DS. Effects of drugs on clinical laboratory tests, 5th ed.
tolerances. AACC Press, 2000.

B1133-2/0901
R1.ing

QUALITY SYSTEM CERTIFIED LINEAR CHEMICALS, S.L.U. Joaquim Costa 18 2ª planta. 08390 Montgat (Barcelona) SPAIN
ISO 9001 ISO 13485 Telf. (+34) 934 694 990; E-mail: info@linear.es ; website: www.linear.es NIF-VAT:B60485687

You might also like