0005 Err D+
0005 Err D+
Detection Always
Control section only
Detecting Communication error has occurred with monitoring section (CPU2).
condition ・Transmission retry error has occurred (3 times at intervals of 3 sec.).
Operation Lamp Sound BP stop HP stop Clamp close Bypass UF stop Heater OFF Hyd. line stop
Red blinks Beep ○ ○ Venous line ○ ○ ○ ○
Reset Turn OFF and ON power supply.
condition
Note Stops all operations in present process.
Monitor PCB is disconnected during treatment Measures: Turn off the machine and reconnect the PCB.
Harness between control and monitor PCBs is disconneMeasures: Turn off the machine and reconnect the harness.
See Maintenance Manual 11.5(wiring diagram)
Harness between control and monitor PCBs is damagedMeasures: Replace the harness.
See Maintenance Manual 11.5(wiring diagram)
Power to monitor PCB is not normal. Analyzing: Check the connection of harness between power
supply unit and mother PCB.
See Maintenance Manual 11.5(wiring diagram)
Measures: Reconnect the harness.