Professional Documents
Culture Documents
MiSeq Field Service Guide (v9, Mar 2022) PDF
MiSeq Field Service Guide (v9, Mar 2022) PDF
FieldService Guide
E May 2015 Removed the Training Level from the table at the
beginning of every procedure per the Clinical
Transformation Initiative.
Corrected the reagent sipper tube FRU part number
to 15044014S.
Added two new procedures for Y-stage replacement
where isolator standoffs are used:
Remove the Y-Stage—Isolator Standoffs
Install the Y-Stage—Isolator Standoffs
Added the set of isolator standoffs (15072883) to the
MiSeq Toolkit.
Revised the section: Align and Qualify the Y-Stage.
Revised the section: Y-Stage Calibration.
Updated Z-stage FRU part number from 10524S to
15065545S.
Added more information to "Remove the Flow Cell
TEC".
Removed PhiX run as a qualification step for SBC
board replacement (change requested by Mike
Addison; confirmed by Product Support).
LED Replacement procedure:
Remove the LED: Revised the procedure to eliminate
removal of the LED cover. No need to remove the
cover to remove the LED from the instrument. Also
corrected which black connector to
disconnect/reconnect.
Install the LED: Removed and revised instructions to
exclude installing the cover. FRU comes with the
cover already installed.
Updated the FRU part number from 15022597S to
15069870S (E38897)
Updated Flow Cell TEC FRU part number from
15029741S to 15051681S
Removed the instructions for installing a V2 reagent
valve. Currently only v3 reagent valves used for
reagent valve replacements.
D January 2015 Revised the guide to facilitate its use for the MiSeq
FGx as follows:
• Changed all instances of MCS to the instrument
control software
• Added the MiSeq FGx user guide, safety and
compliance guide, and the site prep guide to the
MiSeq FGx reference document list under About
this Guide.
• Under About this Guide, updated the titles of the
references to the MiSeq install guide and checklist
to MiSeq/MiSeq FGx.
• Added MiSeq FGx Control Software to the
Software Required section.
• Added MiSeq FGx logins and passwords under
Logins and Passwords section.
• Under Rebooting the Computer and Shutting Down
the Computer, removed the MiSeq screen captures.
No screen captures required.
Revision History ii
Table of Contents ix
Chapter 2 Safety 9
Safety Information 10
Chapter 1
About this Guide 2
Tools Overview 3
About the Instrument Control Computer 5
About the Instrument Display Modes 6
Finding FRUs 7
Purpose
This guide describes how to service the MiSeq and MiSeq FGx instruments at customer
sites. It includes instructions for replacing field-replaceable units (FRUs).
Intended Audience
This manual is intended for qualified Field Service Engineers (FSEs) and Field Application
Scientists (FASs). Qualified FSEs and FASs have received the appropriate training, and are
authorized to carry out the service procedures described in this guide.
Reference Documents
The instrument-specific installation qualification and operational qualification document is
intended for use with this guide. Refer to this document for instrument specifications.
} MiSeq System Installation Qualification and Operational Qualification (document #
150632822)
Additional reference documents for the MiSeq include:
} MiSeq/MiSeq FGx Installation Guide (document # 15027239)
} MiSeq System User Guide (document # 15027617)
} MiSeq System Site Preparation Guide (document # 15027615)
} MiSeq System Safety and Compliance Guide (document # 15027616)
Wash bottle (empty PR2 bottle; 500 ml) Instrument accessory; customer-
supplied
Software Required
This software is used to perform the service procedures in this guide.
NOTE
In this guide, the term instrument control software refers to MiSeq Control Software and
MiSeq FGx Operating Software.
fas@forenseq.uas fasDefaultPwFGx
3 Enter the common name in the Search field, and select Search.
Figure 1 Search Using the FRU Search Criteria (Common Name)
Notice the common name is in the Notes field. A checkmark in the Active field
indicates that this is the current spare.
Figure 2 Notes and Active Fields
Safety
Chapter 2
Safety Information 10
Optical Procedures
Chapter 3
Camera 2 Replacement 12
Camera 1 Replacement 19
Compensator Repairs 24
Compensator Adjustment 32
Compensator Qualification 34
F4 and F8 Replacement 36
LED Replacement 39
LED Calibration 43
LED Qualification 44
M3 Mirror Inspection and Service 46
M3 Mirror Replacement 50
M3 Motor Replacement 56
NOTE
The FRU for both cameras is the same. To enable the FRU to operate as camera
2, you open the camera and change the position of one switch on the circuit
board.
Prerequisites
Make sure that you have full access to the back, front, and left side of the instrument.
Equipment Required
} ESD with ground strap
} Laboratory gloves
Workflow
1 Remove camera 2 from the instrument.
2 If replacing the camera, open the new camera, and set the switch position to camera 2.
3 Install the camera.
4 Qualify the camera.
5 Complete the post service call tasks.
Safety
No special safety precautions are required for this procedure.
Remove Camera 2
1 Close all applications, then 1) shutdown the PC, and 2) turn off the instrument.
2 Remove the right, left, and top skins.
3 Remove the appropriate covers from the optical bench.
7 Remove the tool, making sure you do not turn the knob.
2 Using a 2.5 mm Allen wrench, remove these 6 screws, and lift off the metal plate
(screws removed in this figure).
Figure 8 2.5 mm Screws
Install Camera 2
1 Place the camera against the mounting plate, and loosely install the mounting screw.
2 Attach the Y Offset tool to the camera.
Figure 11 Camera 1
Prerequisites
Make sure that you have full access to the back, front, and left side of the instrument.
Workflow
1 Remove the camera
2 Install the camera
3 Qualify the camera.
4 Complete the post service call tasks.
Safety
No special safety precautions are required for this procedure.
Remove Camera 1
1 Close all applications, then 1) shutdown the PC, and 2) turn off the instrument.
2 Remove the right, left, and top skins.
3 Remove the appropriate covers from the optical bench.
4 Using a 2.5 mm Allen wrench, remove the ground cable.
5 Unplug the cables on top of the camera.
Figure 12 Cables Connected to the Camera
A Power cable
B USB cable
C Control cable
A Power cable
B Control cable
6 Measure and record the relationship between the camera and the mounting plate in X
and Y (Figure 14).
7 Using an M3 Allen wrench, remove the bolt at the base of the camera.
Figure 14 Measure Camera to Mounting Plate in X and Y
Install Camera 1
1 Without fully tightening the mounting screw, attach the mount to the new camera.
Figure 15 Attach Mount to New Camera
Qualify Camera 1
If you replaced camera 2, skip these instructions and follow the steps under Adjust and
Qualify Camera 2 on page 17.
1 Turn on the instrument.
2 Launch MTS, initialize, and home the instrument.
3 Load a super flat beaded flow cell.
4 Focus the beads as follows:
a Go to the Motor tab, select tile 6, and click Move To Tile.
b Set the Z-Motor position to –0.17.
c Go to the Camera tab and:
— Select Camera C, red LED
— Click Continuous.
d On the Image tab, click Contrast, then Zoom Out.
Prerequisites
Make sure that you have full access to the front of the instrument.
IMPORTANT
Use part number 20032622 to replace the Compensator motor and 20032931 to replace the
Compensator assembly. Do not use previous part numbers. See below for the
manufacturing instrument serial number cut-ins for parts 20032622 and 20032931:
} MiSeqRUO ≥ M06488
} MiSeqFGx ≥ M50260
If … Then …
The bottom surface BoltPAC values are out Replace the compensator assembly.
of spec
Workflow
1 Remove the compensator.
2 Replace the appropriate part.
3 Install the compensator.
4 Qualify the compensator.
5 Complete the post-service call tasks.
Safety
No special safety precautions are required for this procedure.
1 Using a 2.5 mm Allen wrench, remove the arm from the compensator assembly.
2 Remove the 4 screws that attach the motor to the compensator mount.
3 Orient the replacement motor on the compensator mount as shown here.
a Position the pigtail to the right of the motor.
b Position the flat side of the motor shaft facing down.
c Align the mounting holes with the motor.
Figure 19 Orientation of Motor on Compensator Mount
2 Identify the positioning pin and wedge on the bottom of the compensator.
Figure 23 Positioning Guides
A Push the alignment pin up against the right front corner of the Z-stage
B Push the alignment wedge up against the back of the Z-stage
b Firmly push the compensator up against the Z-stage by applying pressure forward
and diagonally to the left.
Figure 25 Apply Pressure Forward and Diagonally to Left
5 Make sure that the compensator is firmly in place by trying to move it.
6 Make sure that there is at least 1 mm clearance between the heat shrink tube on the
compensator motor and the projection lens.
If there is not enough clearance, reseat the compensator and check again.
7 Connect the motor and sensor cables.
Figure 27 Motor and Sensor Cables Connected
A Make sure that there is 1 mm clearance between bottom of heat shrink tube and
projection lens
B Sensor cable
C Motor cable
Adjusting the compensator affects only the X value of tip/tilt for the bottom surface of the
flow cell. If the problem is common for both the top and bottom flow cell surfaces, you
typically need to adjust the optics plate.
Prerequisites
Make sure that you have full access to the front of the instrument.
Equipment Required
} Allen wrenches, 2.5 and 3.0 mm
Workflow
1 Adjust the compensator.
2 Qualify the compensator.
3 Complete the post-service call tasks.
Safety
No special safety precautions are required for this procedure.
5 Firmly butt the compensator up against the Z-stage by applying forward pressure and
diagonal pressure to the left.
6 Holding the compensator firmly against the Z-stage, tighten down the M4 screws.
Figure 31 Apply Pressure Forward and Diagonally to Left
7 Make sure that the compensator is firmly in place by trying to move it.
3 Run BoltTileTilt.exe on the top and bottom surfaces of a super flat flow cell (see Tip/Tilt
on page 420).
If … Then …
4 Run BoltPACnoD50.exe on the top and bottom surfaces of the flow cell; make
adjustments if necessary.
5 Find the best focus Z-position on the top surface of the flow cell.
Within spec 1. Record this value for input into the MiSeqOverride.cfg file.
2. Continue to the next step.
Prerequisites
Make sure that you have full access to the front and both sides of the of the instrument.
Workflow
} Remove the mirror.
} Install the mirror.
} Qualify the system.
} Complete the post service call tasks.
Safety
CAUTION
Wear safety glasses when removing and installing the spring to prevent eye injury.
Prerequisites
Make sure that you have full access to the front and left side of the instrument.
Equipment Required
} Allen wrenches, 3.0 mm
} Blackout cloth
Workflow
1 Remove the LED.
2 Install the LED.
3 Qualify the LED.
4 Complete the post-service call tasks.
Safety
No special safety precautions are required for this procedure.
2 Insert the 3 mm Allen wrench through the hole in the LED cover, and tighten the
mounting screw until the assembly is secure.
Figure 38 LED Mounting Screw
Materials Required
} MiSeq Single Point LED Calibration Parameter Generator (15063994)
Download this document from Box.
Overview
To calibrate the LEDs, you will:
} Measure the green and red LEDs in MTS.
} Generate a new calibration parameter for each LED using the workbook.
} Enter the new parameters into the MiSeqOverride.cfg file.
} Restart MTS.
Procedure
1 Open theMiSeq Single Point LED Calibration Parameter Generator(15063994), and click the
LED tab.
2 In MTS, click the LED tab.
3 Measure the green LED as follows:
a Disable the red LED.
b Set the green LED current to 2000 mA.
c Click Enable.
d Click Read.
e On the LED tab in the workbook, type the Photodiode [ADC] value in the orange
box underGREEN LED #2.
The new value is automatically calculated.
f Disable the green LED.
4 Measure the red LED as follows:
a Set the red LED current to 1500 mA.
b Click Enable.
c Click Read.
d On the LED tab in the workbook, type the Photodiode [ADC] value in the orange
box underRED LED #1.
The new value is automatically calculated.
5 Open the MiSeqOverride.cfg file.
6 Enter the new values for each LED.
7 Restart MTS.
8 If a problem occurs while restarting MTS, make sure that you entered the red and green
values correctly.
Prerequisites
Make sure that you have full access to the front and left side of the instrument.
Equipment Required
} Allen wrench, 2 mm
} Loctite 242 (PN 15020191)
} Screwdriver, flat-head
} Toothpick, wooden applicator stick, or other suitable applicator for Loctite
Workflow
1 Inspect the M3 cam screw for proper configuration.
2 Adjust or replace the M3 cam screw if necessary.
3 Inspect the Y-axis M3 set screw for proper configuration and Loctite.
4 Adjust and apply Loctite if necessary.
Safety
No special safety precautions are required for this procedure.
Can be turned The screw cannot The screw is stripped. No corrective action is
be tightened available. Contact Product Support for
enough to stop the instructions.
cam from turning
Can be turned The screw can be Tighten the screw until the cam does not
tightened enough turn.
to stop the cam Perform a tip/tilt adjustment on camera 1
from turning (Tip/Tilt Adjustment—Camera 1 on page 429).
If … Then …
Loctite is not present, or if you are unsure it Continue to the next step, and apply Loctite.
was applied
5 While counting the number of turns, back out the Y-axis screw until it is disengaged.
8 Rethread the screw to its approximate original position by counting the number of
turns recorded from step 5 above.
9 Requalify the system by running the BoltPAC.exe script.
10 Review the data and make adjustments as necessary. Rerun the scripts until all of the
tests pass.
11 After all tests pass, attach CSV files for both surfaces to the work order.
Prerequisites
Make sure that you have full access to the front and left side of the instrument.
Equipment Required
} Loctite 242 (PN 15020191)
Make sure that the Loctite is not expired; check the expiration date on the bottle.
} Screwdrivers, flat-head and Phillips-head
} Toothpick, wooden applicator stick, or other suitable applicator for Loctite
} Wrenches, M3, 2, 2.5, 4 mm Allen
Workflow
1 Remove the M3 mirror.
2 Install the M3 mirror.
3 Align and qualify the M3 mirror, M3 motor, and Camera 1.
Safety
No special safety precautions are required for this procedure.
2 Close all applications, then 1) shutdown the PC, and 2) turn off the instrument.
3 Remove the right, left, and top skins.
4 Remove the appropriate covers from the optical bench.
5 Remove Camera 1 by following these instructions: Remove Camera 1 on page 20
A Tile tilt adjustment screw is turned so that it touches the cover plate
B Rear clamp flexure
7 Using a 4 mm Allen wrench, remove the 4 screws that attach the flexure to the chassis,
and remove the flexure.
8 Remove the sensor cable and sensor screw as follows:
a Remove the sensor cable from the groove in the imaging module plate (B).
b Using an M2 Allen wrench, remove the sensor screw (A).
Figure 49 Sensor Cable and Sensor Screw
A M3 sensor screw
B M3 sensor cable and groove on imaging module plate
• To remove the new style M3 mirror, remove the original screw and the new screw.
Figure 51 Location of new style M3 Mirror Screws
10 Slide the mirror up and off the motor shaft. Do not twist or turn the motor shaft. The
shaft has a flat edge to stop the mirror from slipping.
Figure 52 Bottom of M3 Mirror and Motor Shaft
6 Counting the number of turns, back out the Y-axis screw until it is disengaged.
8 Rethread the screw to its approximate original position by counting the number of
turns recorded in step 6.
9 Reinstall the M3 sensor screw.
10 Reinstall the rear clamp flexure.
11 Route the sensor cable in the groove on the imaging module bench.
Prerequisites
Make sure that you have full access to the back, front and left side of the instrument.
IMPORTANT
Use part number 20032622 to replace the M3 motor. Do not use previous part numbers. See
below for the manufacturing instrument serial number cut-ins for parts 20032622:
} MiSeqRUO ≥ M06488
} MiSeqFGx ≥ M50260
Equipment Required
} Allen wrenches, 1.5, 2.5 and 4.0 mm, and M2
Workflow
} Remove the defective motor
} Install the new motor
} Qualify the new motor, M3 mirror and Camera 1
} Complete the post-service call tasks
Safety
No special safety precautions are required for this procedure.
A Tile tilt adjustment screw is turned so that it touches the cover plate
B Rear clamp flexure
6 Using a 4 mm Allen wrench, remove the four screws that attach the flexure to the
chassis, and remove the flexure.
7 Remove the sensor cable and sensor screw as follows:
a Remove the sensor cable from the groove in the imaging module plate (B).
b Using an M2 Allen wrench, remove the sensor screw (A).
Figure 60 Sensor Cable and Sensor Screw
A M3 sensor screw
B M3 sensor cable and groove on imaging module plate
Electrical Procedures
Chapter 4
E-Box Components and Cables 62
E-Box Access or Removal 65
Combo Board or Main Board Replacement 71
E-Box Replacement 84
Main Board Replacement 94
Power Supply Replacement 95
SBC Board Replacement 99
New SBC Replacement (AIMB205) 112
Z-Stage Control Board Replacement 129
Flow Cell RFID Replacement 130
Chiller RFID Board Replacement 136
Reagent Cartridge RFID Replacement 140
Flashing the NVRAM 141
FPGA Upgrade 143
Drive Replacement and Configuration 145
A Combo Board
B Main Board (behind the Combo Board)
C Z-stage Control Board
D Fan
E Power supply
A SBC Board
B Speaker
C Hard drive
Main board, Combo board, Z- Proceed to Reposition or Remove the E-Box on page 68.
stage control board, or the power
supply
2 Label and remove the cables from the E-Box door only.
Be sure to label the camera cables.
3 Loosen the five thumb screws using a Phillips-head screwdriver.
Figure 70 Thumbscrews
Main board, Combo 1. Position the E-Box so that you can open and work inside of it.
board, Z-stage control Set the E-Box on the bench next to the instrument, or
board, or the power disconnect the ethernet and video cables on the back of the
supply box (Figure 73), and move it to another location.
2. Loosen the five thumb screws using a Phillips-head
screwdriver (Figure 70 above).
3. Swing the door open.
4. Proceed to one of the following:
• Remove the Combo or Main Board on page 72.
• Remove the Z-Stage Control Board on page 370.
• Remove the E-Box on page 85.
E-Box 1. Disconnect the ethernet and video cables at the back of the
E-Box (Figure 73).
2. Remove the E-Box.
IMPORTANT
Reflash the NVRAM when installing a used E-Box or Main Controller PCA. Parts taken from
other systems have the serial number of the original system written to the NVRAM of the
main board. The serial number of the original system must be reprogrammed using
Functional Test Software. New service parts do not have a serial number in the NVRAM, and
do not require reprogramming.
Prerequisites
Make sure that you have full access to the back, front and right side of the instrument.
Equipment Required
} Allen wrenches, long-handled, 3 mm and 2.5 mm
} ESD wristband with ground strap
} Lab tape
} Nut drivers, 3/16 inch and 1/4 inch
} Phillips head screwdriver
} Super flat beaded flow cell
Workflow
1 Remove the Combo board and Main board.
2 Attach the Combo board to the Main board.
3 Install the boards.
4 Qualify the boards.
Safety
No special safety considerations for this procedure.
6 Using a 3/16 inch nut driver, remove the nuts from each connector (18 nuts in total).
NOTE
The gap around the two LED connections maybe wide enough so that you do not have
to remove the nuts from these two connections. Be careful not to misplace any of the
nuts.
A LED connectors
8 Using an Allen wrench, remove the four socket head cap screws that secure the plastic
cover to the boards, then remove the plastic cover.
Figure 79 Plastic Cover Socket Head Cap Screws
10 Using a 1/4 inch nut driver, loosen the two captive nuts on the pin connector.
Figure 81 Captive Nuts on Pin Connector
A Combo Board
B Main Board
12 If you are replacing the Main Board, remove the Z-stage Control Board from the Main
Board now (see Remove the Z-Stage Control Board on page 370 for instructions).
You will install it on the new Main Board.
13 Set the defective board aside, making sure that you do not interchange the defective
board with the new board.
Defective boards can be discarded. We encourage you to discard in E-Waste recycling if
available.
3 Press down on the boards to push the pins into the connector (rocking the boards back
and forth can help).
Figure 84 Press Pins Into Connector
4 Using a 1/4 inch nut driver, tighten the two captive nuts on the pin connector so that
they are snug; do not over tighten (Figure 81 on page 76).
5 Reinstall the four standoffs using a 1/4 inch nut driver (Figure 80 on page 76). Be sure
to replace the Z-stage cable tie-down if one is present.
IMPORTANT
Do not over tighten the standoffs. Tighten to snug only. Over tightening can lead to
compression of the board layers which can impact functionality.
6 Reinstall the protective plastic sheet (Figure 79 on page 75).
7 If the new board has mounting hardware attached to the connectors on the back of the
board, remove that hardware now.
9 Using a 3/16 inch nut driver, reinstall the nuts onto each connector (Figure 77 on page
74).
4 Reattach the 11 cables to the boards, taking special care with the ribbon cable.
IMPORTANT
The 40-pin connector can be incorrectly inserted off-center. The guide ridge should be
completely within the slot on the board. If not properly attached, the instrument cannot
initialize and displays FPGA timeout errors.
5 Double check that all of the cables have been reattached, and that the connections are
secure.
IMPORTANT
Before closing the E-Box, make certain that all of the cables have been reattached to the
boards, and that the connections are secure.
6 Carefully close the E-Box and tighten the five captive screws.
7 Position the E-Box so that you can reconnect the cables to the faceplate, then reconnect
the cables.
d Repeat steps b and c using these Flow Rates: 1000 and 1500 .
e If each test passes, continue to the next step.
5 Run the volume test in the instrument control software, or use VCL in MTS to test
delivery for all reagent valve positions.
One or more of the optics requires 1. Align the optics as described in the
alignment MiSeq® System Installation Guide.
2. Rerun the scripts.
Prerequisites
Make sure that you have full access to the back, front and right side of the instrument.
IMPORTANT
Check the instrument serial number. If the serial number meets the criteria listed in Table 3,
and the label is missing from the instrument chassis, install a serial number label on the
chassis. A serial number label must be present on the chassis to maintain instrument
compliance. For instructions, see Serial Number Label Installation on page 90
Table 3 Affected Instruments
Instrument Serial Number
Equipment Required
} Allen wrenches, 3 mm and 2.5 mm
} Digital camera (can be your smart phone camera)
Required if you install a new serial number label
} ESD wristband with ground strap
} Lab tape
} Marking pen, black or blue ink, non-water soluble
Workflow
1 Remove the E-Box.
2 Install the E-Box.
3 If required, install a new serial number label on the instrument chassis.
4 Calibrate the flow sensors.
5 Qualify the flow sensors.
6 Qualify the E-Box.
7 Complete the post-service call tasks (Complete the Post-Service Call Tasks on page 447).
Safety
No special safety considerations for this procedure.
8 Open the E-Box, and remove the Z-stage Control Board from the Main Board (see
Remove the Z-Stage Control Board on page 370 for instructions).
You will install it in the new E-Box.
1 Open the E-Box, and attach the Z-stage Control Board to the Main Board (see Install the
Z-Stage Control Board on page 371 for instructions).
2 Position the E-Box on the instrument so that you can connect cables to the inner panel
(similar to the position shown in Figure 92 on page 88).
3 Attach the appropriate cables to the inner panel. (See E-Box Components and Cables on
page 62.)
When a spare E-Box is installed on these instruments, the new Instrument Compliance
label does not list the instrument serial number. As a result, the instrument is non-
compliant. To maintain compliance, a serial number label must be installed on the
instrument chassis. Blank labels (PN 15037891) are included with the spare E-Box.
1 Perform the instructions in this section if the instrument serial number is ≤M01353.
2 At the back of the instrument, locate the rear left end of the chassis.
3 Select from the following:
5 On the instrument, open the file MiSeqOverride.cfg located in C:\MiSeq Config, and
note the instrument Name. The Name is the instrument serial number.
Example: Name = M00806
6 Write S/N–<instrument serial number noted in step 5> legibly on one blank label with a
non-water soluble black or blue ink pen.
Example:
d Repeat steps b and c using these Flow Rates: 1000 and 1500 .
e If each test passes, continue to the next step.
5 Run the volume test in the instrument control software, or use VCL in MTS to test
delivery for all reagent valve positions.
Prerequisites
Make sure that you have full access to the back, front and right side of the instrument.
Equipment Required
} Cable ties
} Camera, digital
} Cutter, diagonal
} ESD wristband with ground strap
} Lab tape
} Marker, indelible
} Nut driver, 7/32 inch
} Screwdriver, Phillips-head
} Wrenches, Allen 2.5, 3, and 4 mm
Workflow
1 Remove the power supply.
2 Install the power supply.
3 Qualify the power supply.
Safety
No special safety considerations for this procedure.
6 Remove the fan (7/32 nut driver), and the power supply ground cable if present.
Figure 99 Fan Secured to E-Box
Figure 100 Tight Fit Between Power Supply and Main Controller U-Bracket
2 Secure the power supply to the E-Box with the four screws.
3 Position the fan in the E-Box, and reinstall the two nuts that secure it in place.
4 Connect the ground strap to the E-Box and power supply (TRUMP power supply
shown here).
Figure 101 Ground Strap Connected to E-Box and Power Supply
9 Bundle the cables on top of the E-Box, and secure them in the cable clamps.
10 Reinstall the six screws that attach the E-Box to the chassis.
11 If present, remove the imaging module-to-chassis grounding cable (no longer
necessary).
Figure 102 Remove Imaging Module-to-Chassis Grounding Cable
Prerequisites
} Make sure that you have full access to the back, front and right side of the instrument.
} New SBC boards include a new network adapter, which changes the MAC address.
The new address might require that the instrument be re-networked (DHCP server).
Alert the on-site IT department that the instrument has a new MAC address before you
replace the board.
Equipment Required
} Allen wrench, 2.5 and 3 mm
} Cable ties, approximately 6 small
} Cutter, diagonal
} ESD wristband with ground strap
} Marker, Sharpie
} Nut drivers, 7/32 inch
} Screwdrivers, Phillips-head and small flat-head
Workflow
1 Remove the SBC board.
2 Install the SBC board.
Safety
No special safety considerations for this procedure.
E ATX cable (power input) J P20 cable tied out of the way
3 Using a 2.5 mm Allen wrench, remove the four screws and washers that attach the
board to the E-Box (located at the four corners of the board).
4 Remove the board from the instrument.
6 Connect the fan cable to the fan connector on the SBC board.
Figure 108 Fan Cable Connection
2 Record the MAC address, located on the label of the Ethernet ports.
Provide the customer’s IT department with the new MAC address when you are
finished with this service procedure. It might be required for network authentication.
Figure 112 MAC Address Location
4 Make sure that all of the cable receptacles are properly seated in the port bracket.
NOTE
You might have to slightly bend the port bracket to seat the board.
5 Secure the board to the standoffs by reinstalling the four screws with washers using a
3 mm Allen wrench.
NOTE
Apply horizontal pressure against the other grounding taps to align the holes in the
standoffs with the holes in the E-Box door.
Be sure to add a washer between each screw and the board.
E ATX cable (power input) J P20 cable tied out of the way
i Audio cable—keyed
Figure 116 Audio Cable - Keyed
6 Log on as sbsuser.
7 Once Windows and the instrument control software have started, go to Adjust the Clock,
Date, and Time (Windows 7) on page 440 and:
a Adjust the time zone, date, and time.
The SBC board starts up with US PDT settings.
b Synchronize the instrument to the Amazon S3 servers.
8 For new SBCs (AIMB205) complete the following steps:
a Open Control Panel.
b Select Network and Sharing Center.
c Select Change adapter settings.
d Right-click on Ethernet and select Properties.
e Select Internet Protocol Version 4 and select Properties.
f Select Obtain an IP address automatically and Obtain DNS server address
automatically.
NOTE
If this action is not performed, then the Ethernet ports may be switched.
g Select OK.
h Open Control Panel.
i Select Network and Sharing Center.
n Select OK.
9 Verify the network settings; adjust them if necessary (the new MAC address might be
required by the local IT department for their DHCP server).
10 Reboot the system to make sure that all of the settings are OK.
This section provides instructions for the replacement of the current (AIMB-272) SBC board
with the new (AIMB-205) SBC board (Windows 10). The new SBC boards require a new
panel to be installed on the E-Box enclosure. Cable routing to the SBC, both inside and
outside the E-Box enclosure, is different with the new board.
Prerequisites
} Compatible with Windows 10 (v3.1.C02 image) only.
} Make sure that you have full access to the back, front and right side of the instrument.
} New SBC boards include a new network adapter, which changes the MAC address.
The new address might require that the instrument be re-networked (DHCP server).
Alert the on-site IT department that the instrument has a new MAC address before you
replace the board.
Equipment Required
} Allen wrench, 2.5 and 3 mm
} Cable ties, approximately 6 small
} Cutter, diagonal
} ESD wristband with ground strap
} Nut drivers, 7/32 inch
} Screwdrivers, Phillips-head and small flat-head
Workflow
1 Remove the SBC board.
2 Install the SBC board.
3 Configure the SBC board.
5 Remove board bezel from the sheet metal door by pushing the bezel from the outside
towards the inside of the E-Box.
6 Locate the USB to 10 pin HDR cable that goes from the SBC to the two USB ports (next
to the disk drive) on the back of the instrument.
7 Cut and remove cable ties to access the standoffs connecting the cable to the sheet
metal.
8 Remove the standoffs and the USB cable.
9 Remove the 8 screws holding the IO panel with the camera USB ports in place.
Figure 120
10 Remove both the IO panel and the cable for the camera USB ports from the E-Box door.
11 Remove the standoffs holding the cable to the IO panel and set the standoffs aside for
later use.
Figure 121
12 Identify the new IO panel with USB 3.0 label and the new dual USB to dual USB cable.
13 Using the standoffs that were holding the camera USB cable to the original IO panel,
attach the new dual USB to dual USB cable to the new IO panel (20040380) with the
USB 3.0 label.
Figure 122
14 Identify the new SBC (AIMB-205). The board bezel should be included with the SBC.
15 Identify the new SBC board and locate the board bezel for the new SBC.
16 Remove slugs from bezel.
17 Apply and align single board bezel onto E -Box door. Listen for click. Attach with 8
2.5mm screws.
18 Verify single board bezel is facing away from the inside of the E-Box.
19 Verify the orientation of the bezel.
20 Ensure the tabs at bezel are bent on the ethernet connector.
Figure 124
21 Mount SBC onto E-Box. Align the SBC onto 4 standoffs on sheet metal door.
NOTE
New SBC is only compatible with Windows 10 and not Windows 7.
3 Attach the free end of the new dual usb to dual usb cable to the cutouts in the rear of
the ebox where the old USB cable was removed.
4 Use the standoffs from the cable that was removed from the same location to secure the
new dual USB cable to the sheet metal.
NOTE
Remove the 24-way power connector and rotate the USB connector to access the lower stand
off.
Figure 134
2 Route and install the other end of ATX Main cable labeled SBC onto the SBC connector.
5 Install prepared 4-port serial SBC Main cable onto SBC PCA as shown.
NOTE
There are two headers next to each other that the ribbon cable can connect to, the cable must
be installed on the left header (as pictured).
7 IMPORTANT: Use cable ties to bundle and organize cables and place cables in cable
channel before closing the E-Box.
NOTE
Make sure cables stay clear of the CPU fan when the E-Box is closed. (No fan guard included
at this time.)
8 Identify the two USB extension cables.
9 Close the E-Box and attach the USB extension cables from the SBC USB ports to the IO
panel, as shown.
10 Connect the cameras and LCD display to the external USB ports on the SBC in the
locations shown.
Figure 143
Prerequisites
Make sure that you have full access to the front and right side of the instrument.
Equipment Required
} Allen wrenches, 2.0 and 2.5 mm
} PR2 bottle filled with DI water
} Screwdriver, Phillips-head
} Wash tray filled with DI water
Workflow
1 Remove the flow cell RFID.
2 Install the flow cell RFID.
3 Qualify the flow cell RFID.
4 Complete the post-service call tasks.
Safety
No special safety precautions are required for this procedure.
3 Using a 2.5 mm Allen wrench, remove the four inner screws that attach the TEC to the
Y-stage.
NOTE
The TEC drops when the screws are removed.
Prerequisites
Make sure that you have full access to the front and right side of the instrument.
Equipment Required
} Allen wrench, 2.5 mm
} Screwdriver, Phillips-head
Workflow
1 Remove the PR2 RFID.
2 Install the PR2 RFID.
3 Qualify the PR2 RFID.
4 Complete the post-service call tasks.
Safety
No special safety precautions are required for this procedure.
6 Using a 2.5 mm Allen wrench, remove the four screws that attach the board to the
instrument.
Figure 154 PR2 RFID Board Screws
11 When complete:
a Close the FTS software.
b Open the MiSeqOverride.cfg file, and remove the 2 lines added in step 4 ([Cloud]
and PerformCloudTest = false).
c Save and close the MiSeqOverride.cfg file.
12 Launch MCS, and make sure that the new serial number is correct.
13 Remove FTS from the MiSeq as follows:
a Run the FunctionalComponentTestInstaller.msi installer again
b Select Remove when prompted in the wizard.
Figure 158 Remove FTS from the Instrument
14 Remove the FTS zipped and unzipped folders from the desktop, making sure to empty
them from the recycle bin.
SPARE,MiSeq,RUO,SSD,1TB
Contains RUO SSD only. Order for MiSeqRUO with existing
2-bay drive carrier.
SPARE,MiSeq,FGx,SSD,1TB
Contains FGx SSD only. Order for a MiSeqFGx with existing
2-bay drive carrier. If not in SFDC, contact ILMN-Customer
Solutions Parts Inquiry (ILMN-CustomerSolutionsPartsInquiry@
illumina.com)
Prerequisites
NOTE
Check the DCP Technical Attributes for the specific MiSeq instrument. If the hard drive type
is Enterprise, a carrier upgrade is required.
MiSeq FGx:
• If a spare Enterprise hard drive is available,
replace just the hard drive.
• SPARE, KIT, ENTERPRISE HD+1Bay Miseq
FGx (15069836S)
• If a spare Enterprise hard drive is not available,
upgrade the instrument to a 2-bay drive carrier
with SSD.
• FGx: UPGRADE KIT,MiSeq,FGx,HDD to SSD
2-bay hard drive carrier Replace the drive (standard or solid state) with the
appropriate SSD:
• RUO: SPARE,MiSeq,RUO,SSD,1TB
• FGx: SPARE,MiSeq,FGx,SSD,1TB
Make sure that you have full access to the front and back of the instrument.
Workflow
1 Copy data from the existing hard drive.
2 Do one of these:
• Retrofit the instrument with a 2-bay drive carrier and SSD.
• Replace a currently installed SSD.
• Replace a currently installed Enterprise HD with an Enterprise HD (MiSeq FGx
only)
3 Configure the drive.
4 Qualify the drive.
5 Update the Installed Product page in SFDC.
Safety
No special safety considerations for this procedure.
5 To change the date and time select Change date and time, then set as appropriate.
Select OK to save changes.
Figure 170 Date and Time Settings
CAUTION Speak with the on-site IT professionals. Advise them that the
Make sure that the network setting for Windows Updates must be turned off.
network setting for If turned on, Windows Updates override the instrument and
Windows Updates is forces it to reboot. If a run is in progress, the run fails.
turned off.
Task How To
If Local Run Manager is 1. Launch Chromium from the desktop, and log in to Local
installed on the Dx SSD, Run Manager (User Name=admin; Password=password).
create an 2. In the Local Run Manager dashboard navigation bar,
administrative select System | User Management.
customer account.
3. Select Create User in the User Management page.
4. In the Create New User field, enter first and last names,
user name, and password information. The password is
temporary, and is reset at initial login.
5. Toggle to Admin in the Role field by clicking User, then
select Create User and Continue.
6. Log out of Local Run Manager, and close the Chromium
browser.
7. Provide the customer the newly created user name and
temporary password and have them test the account log
in. Direct them to the Local Run Manager Software Reference
Guide for MiSeqDx (1000000011880).
Make sure that the Open folder and make sure that the CloudAPIKey.xml file is
CloudAPIKey.xml file present:
is present and ensure Dx: C:\ProgramData\Illumina\MiSeq
the serial number in RUO: C:\ProgramData\Illumina\MiSeqControlSoftware
this file matches the
serial number of the Note: Program Data may not show--select Hidden Items
instrument. from View tab of file explorer
Synchronize the See Synchronize the Instrument with an Internet Time Server
instrument with an on page 463.
internet time server
Fluidics Procedures
Chapter 5
Chiller Disassembly 160
Chiller Assembly Replacement 167
Chiller Elevator Motor Replacement 176
Chiller TEC (Peltier Heater) Replacement 182
Chiller Thermistor or Reagent Cartridge RFID Assembly Replacement 188
Fluidics Line Replacement 194
Sipper Elevator Assembly Replacement 201
Sipper Piercing Probe Assembly Replacement 210
Reagent Sipper Tube Replacement 216
Sipper Lead Screw Lubrication 220
Flow Cell Holder Replacement 224
Flow Cell Inlet (Common) Line Replacement 229
Flow Cell Manifold Replacement 233
Flow Cell TEC Replacement 241
Flow Cell TEC Thermal Calibration 250
Flow Cell TEC Qualification 255
Flow Sensor Assembly Replacement 256
Flow Sensor Calibration 259
Flow Sensor Assembly Qualification 260
Reagent Valve Replacement 261
Reagent Valve Power/Comm Cable Replacement 275
Sipper Console Replacement 278
PR2 Sipper Tube Replacement 290
Syringe Replacement 299
Syringe Pump Replacement 304
Estimated Time See estimated time for the specific part being replaced.
Prerequisites
You must have full access to the front, back and right side of the instrument.
Equipment Required
} Allen wrenches, 1.5 mm, 2.5 mm (6 inch and 9 inch), and 3 mm
} Screwdriver, Phillips head (< 5 inches)
Workflow
} Disassemble the chiller as described in this section.
} Replace and qualify the defective part as described under:
• Chiller Elevator Motor Replacement on page 176
• Chiller TEC (Peltier Heater) Replacement on page 182
• Chiller Thermistor or Reagent Cartridge RFID Assembly Replacement on page 188
• Fluidics Line Replacement on page 194
• Sipper Elevator Assembly Replacement on page 201
• Sipper Piercing Probe Assembly Replacement on page 210
• Reagent Sipper Tube Replacement on page 216
Safety
No special safety considerations for this procedure.
8 If a ground strap is present, use a 3 mm Allen wrench to disconnect it from the E-Box.
Figure 184 Ground Strap
9 Remove the two screws that attach the back of the chiller to the base plate using a
3 mm Allen wrench.
10 Loosen the 2 captive screws that secure the front of the chiller to the instrument:
a Slide the instrument forward so that front of it overhangs the lab bench by
approximately 5 inches (allowing access to the two screws underneath the chiller.
b Loosen or remove the 2 screws under the front of the instrument (directly below the
reagent compartment).
NOTE
Depending upon the instrument serial number, these screws can be captive or Phillips-head.
Figure 186 Captive Screws on the Underside of the Front of the Instrument
11 Open the chiller door, and remove the two Phillips head screws (use a screwdriver that
is < 5 in.).
13 Slide the chiller towards the back of the instrument away from the monitor.
14 Remove the top of the chiller as follows:
a Loosen the four screws that attach the top to the bottom of the chiller using a
9 inch 2.5 mm Allen wrench.
b Lift off the top of the chiller.
Figure 190 Removing Chiller Top from Chiller Bottom
Prerequisites
Make sure that you have full access to the back, front and sides of the instrument.
Equipment Required
} Allen wrench, 3 mm
} Flow cell, wash
} Fluidics wrench
} Screwdriver, Phillips head
} Torque wrench and fittings
} Wash bottle filled with DI water
} Wash tray filled with DI water
Workflow
1 Remove the chiller assembly.
2 Install the chiller assembly.
3 Qualify the chiller assembly.
Safety
No special safety considerations for this procedure.
5 If a ground strap is present, disconnect it from the E-Box using a 3 mm Allen wrench.
Figure 196 Ground Strap
7 At the front of the chiller, loosen or remove the 2 screws that attach it to the chassis:
a Slide the instrument forward so that it overhangs the bench by approximately
5 inches.
b Loosen or remove the 2 screws under the front of the instrument (directly below the
reagent compartment).
NOTE
Depending upon the instrument serial number, these screws can be captive or Phillips head.
3 Review the data and make adjustments as necessary. Rerun the scripts until all of the
tests pass.
4 Record the chiller temperature in the work order.
5 Attach CSV files for both surfaces to the work order.
6 Attach a screen shot of the Volume Test to the work order.
7 Complete the post-service call tasks (Complete the Post-Service Call Tasks on page 447).
Prerequisites
Make sure that you have full access to the back, front and sides of the instrument.
Equipment Required
} Allen wrench, 1.5 mm
} Allen wrench, 2.5 mm (if replacing the motor only)
} Phillips head screwdriver (if replacing the motor assembly)
} PR2 bottle with RFID tag
} Reagent cartridge, full with sealed tubes, PN 15028689
Workflow
1 Remove the defective motor assembly.
2 Install the new motor assembly.
3 Reassemble the chiller.
4 Qualify the new motor.
Safety
No special safety considerations for this procedure.
The motor only 1. Loosen all four screws on the motor coupling using a
1.5 mm Allen wrench (A).
2. Remove the four screws that secure the motor to the
bracket using a 2.5 mm Allen wrench (B).
3. Lift off the motor, leaving the coupling in place.
The motor assembly 1. Loosen the lower two screws on the motor coupling
using a 1.5 mm Allen wrench (D).
2. Remove the four screws and washers that secure the
motor assembly to the top of the chiller using a
Phillips head screwdriver (C).
3. Lift off the motor assembly.
The motor only Reinstall the four screws that secure the motor to the
bracket using a 2.5 mm Allen wrench (B in Figure 203 on
page 178).
The motor assembly Reinstall the four screws and washers that secure the
motor assembly to the top of the chiller using a Phillips
head screwdriver (C in Figure 203 on page 178).
Prerequisites
Make sure that you have full access to the back, front and sides of the instrument.
Equipment Required
} Allen wrench, 3 mm
Workflow
1 Disassemble the chiller, and remove the defective chiller TEC.
2 Install the new TEC.
3 Reassemble the chiller.
4 Qualify the new TEC.
Safety
No special safety considerations for this procedure.
3 Install the ferrite on the RFID/thermistor cable in the position shown here.
Figure 209 Ferrite Position
A Chiller thermistor
B Reagent cartridge RFID assembly
Prerequisites
Make sure that you have full access to the back, front and sides of the instrument.
Equipment Required
} Phillips-head screwdriver
Workflow
1 Disassemble the chiller and remove the defective thermistor or RFID assembly.
2 Install the new thermistor or RFID assembly.
3 Reassemble the chiller.
4 Qualify the new thermistor or RFID assembly.
Safety
No special safety considerations for this procedure.
3 Bend the plastic tab to pop out one side of the metal plate.
Figure 217 Bend Plastic Tab to Release One End of Metal Plate
5 If replacing the RFID assembly, use a small screwdriver to open the two clips and
remove the board from the tray.
Figure 219 Clips Holding PC Board in Place
Prerequisites
Make sure that you have full access to the back, front and sides of the instrument.
Equipment Required
} Allen wrench, 2.5 mm
} Cable ties, two
} Fluidics wrench
} Torque wrench and fittings
Workflow
1 Disassemble the chiller.
2 Remove the fluidics line.
3 Install the fluidics line.
4 Reassemble the chiller.
5 Qualify the new fluidics line.
Easily accessible on the reagent Remove defective line using the fluidics wrench,
valve leaving the valve in place.
Not easily accessible, particularly the Remove the reagent valve and bracket assembly
lines on the left side of the reagent from the instrument as described under Remove
valve the Reagent Valve on page 262.
Remove defective line using the fluidics wrench.
2 Cut the cable tie, and remove the spiral wrap from around the lines.
3 Remove the two screws that secure the gasket clamp to the chiller using a 2.5 mm
Allen wrench.
A Maximum height
8 Reinstall the reagent valve if you had to remove it while detaching the defective line.
9 Reassemble the reagent chiller by reversing the instructions under Disassemble the
Chiller on page 160.
7 Review the data and make adjustments as necessary. Rerun the scripts until all of the
tests pass.
8 Record the chiller temperature in the work order.
9 Attach CSV files for both surfaces to the work order.
Prerequisites
Make sure that you have full access to the back, front and sides of the instrument.
Equipment Required
} Allen wrench, 2.5 mm
} Cable ties, two
} Fluidics fitting wrench
} Reagent cartridge, full with sealed tubes, PN 15028689
} Screwdriver, Phillips head
} Torque wrench and fittings
Workflow
1 Disassemble the chiller.
2 Remove the defective sipper elevator assembly.
3 Install the new sipper elevator assembly.
4 Reassemble the chiller.
5 Qualify the new sipper elevator assembly.
Safety
No special safety considerations for this procedure.
5 Remove the 6 screws that attach the sipper elevator to the bottom of the chiller using a
Phillips head screwdriver.
NOTE
This photograph was taken in manufacturing. At a customer site, the gasket
would be present.
Figure 235 Top Down View of Empty Chiller—Sipper Elevator Screw Locations
b Insert the lines into the grooves on one piece of the black gasket.
Inserting the lines in numerical order helps keep the lines neat and makes height
adjustment easier later on in the procedure.
NOTE
The translucent guide shown here is used in manufacturing and is not available
to the field.
[
d Route the tubing through the pass thru.
e Tap the ends of the lines on the palm of your hand several times to dislodge any
bits of foam or other particulates that might be adhering to them.
f Secure the clamp to the chiller enclosure (Figure 234 on page 204).
5 With the elevator in its highest position, pull on each tube and adjust the bow so that
all of the lines are within the height range (Figure 239).
The bow of each tube should be neat and tight, but not so tight as to kink the line.
6 Check the tubing while the elevator moves as follows:
a In MTS on the Pump tab, move the elevator up and down (toggle between Sipper:
Move Up and Sipper: Move Down).
b Watch how the tubing bends to make sure that none of the lines kink in any
position.
7 Bundle the lines in the spiral wrap, then cable tie the lines (Figure 240).
A Maximum height
6 Review the data and make adjustments as necessary. Rerun the scripts until all of the
tests pass.
7 Record the chiller temperature in the work order.
8 Attach CSV files for both surfaces to the work order.
9 Attach a screen shot of Volume Test to the work order.
10 Complete the post-service call tasks (Complete the Post-Service Call Tasks on page 447).
Prerequisites
Make sure that you have full access to the back, front and sides of the instrument.
Equipment Required
} Allen wrench, 2.5 mm
} Cable ties, two
} Fluidics fitting wrench
} Screwdriver, Phillips head
} Torque wrench and fittings
Workflow
1 Disassemble the chiller.
2 Remove the sipper elevator assembly.
3 Remove the defective piercing probe assembly.
4 Install the new piercing probe assembly.
5 Reinstall the sipper elevator assembly.
6 Reassemble the chiller.
7 Qualify the new piercing probe assembly.
Safety
No special safety considerations for this procedure.
d Lift the fluidics lines up through the pass thru on the side of the chiller.
e Gently push the black molded gasket out of the clamp.
f Open and remove the gasket (the gasket is two pieces).
3 Unscrew the guide pin (black rod) from the defective assembly.
Install the alignment pin (PN 15020819) onto the new assembly.
Figure 246 Guide Pin and Guide Plate
A Guide pin
B Guide plate
c Insert the lines into the grooves on one piece of the black gasket.
Inserting the lines in numerical order helps keep the lines neat and makes height
adjustment easier later on in the procedure.
NOTE
The translucent guide shown here is used in manufacturing and is not available
to the field.
[
e Route the tubing through the pass thru.
f Tap the ends of the lines on the palm of your hand several times to dislodge any
bits of foam or other particulates that might be adhering to them.
g Secure the clamp to the chiller enclosure (Figure 244 on page 211).
4 With the elevator in its highest position, pull on each tube and adjust the bow so that
all of the lines are within the height range shown in Figure 250.
The bow of each tube should be neat and tight, but not so tight as to kink the line.
5 Check the tubing while the elevator moves as follows:
a In MTS on the Pump tab, move the elevator up and down (toggle between Sipper:
Move Up and Sipper: Move Down).
b Watch how the tubing bends to make sure that none of the lines kink in any
position.
Figure 250 Proper Fluidics Line Height
A Maximum height
Make sure that you have full access to the back, front and sides of the instrument.
Equipment Required
No additional equipment required.
Workflow
1 Disassemble the chiller, and remove the sipper tube.
2 Install the sipper tube.
3 Reassemble the chiller.
4 Qualify the sipper tube.
Safety
No special safety considerations for this procedure.
6 Review the data and make adjustments as necessary. Rerun the scripts until all of the
tests pass.
7 Record the chiller temperature in the work order.
8 Attach CSV files for both surfaces to the work order.
9 Attach a screen shot of Volume Test to the work order.
10 Complete the post-service call tasks (Complete the Post-Service Call Tasks on page 447).
Prerequisites
Make sure that you have full access to the back, front and sides of the instrument.
Equipment Required
} Allen wrenches, 1.5 mm, 2.5 mm (6 inch and 9 inch), and 3 mm
} Lubricant, PTFE, part # 15018061S
} Screwdriver, Phillips head (< 5 inches)
} Wash bottle
} Wash tray
Workflow
1 Remove the top from the chiller.
2 Apply lubricant to the top and bottom of the lead screw.
3 Reinstall the top of the chiller.
4 Qualify the sipper assembly.
Safety
No special safety considerations for this procedure.
3 Apply 0.1—0.3 ml of lubricant along the lead screw starting from the end of the
cartridge (apply < 15 mm of lubricant along the lead screw).
Figure 257 Lubricate Top of Lead Screw
Willing to allow further troubleshooting Replace the chiller elevator motor assembly.
If the issue is not resolved, then replace the
sipper elevator assembly.
Not willing to allow further troubleshooting Replace the sipper elevator assembly.
Prerequisites
Make sure that you have full access to the front and left side of the of the instrument.
Equipment Required
} Alignment tools, manifold and Y-stage
} Allen wrenches, 2.5 mm and 3 mm
} Flow cell, open or wash
} Flow cell, super flat, beaded
} Fluidics wrench
} OTT
} PR2 bottle filled with DI water
} Screwdriver, Phillips-head
} Torque wrench and fittings
} Wash tray filled with DI water
Workflow
1 Remove the flow cell holder assembly.
2 Install the new flow cell holder assembly.
3 Qualify the new flow cell holder assembly.
4 Complete the post-service call tasks.
3 Remove and set aside the flow cell clamp button and spring.
You reinstall these parts later.
4 Supporting the flow cell TEC, use a 2.5 mm Allen wrench to loosen the 4 inner screws
that attach the TEC to the Y-stage.
NOTE
Set the TEC down gently on the chassis. Do not let it drop when the screws are
loosened.
The flow cell inlet line is also referred to as the common line.
Figure 265 Flow Cell Inlet Line FRU (current PN might differ)
Prerequisites
Make sure that you have full access to the front of the of the instrument.
Equipment Required
} Flow cell, open or wash
} Fluidics wrench
} PR2 bottle filled with DI water
} Screwdriver, Phillips-head
} Torque wrench and fittings
} Wash tray filled with DI water
Workflow
1 Remove the flow cell inlet line.
2 Install the flow cell inlet line.
3 Qualify the flow cell inlet line.
Safety
No special safety precautions are required for this procedure.
4 Disconnect the inlet (green) line from the manifold and from the reagent valve head.
5 Cut any cable ties that secure the line to the instrument.
6 Remove the line from the instrument, noting where the line is fed through the flow cell
clamp.
6 Open and close the flow cell clamp 3 to 4 times, making sure that the lines are not
pinched and that the clamp opens fully and freely.
CAUTION
Be sure that the fluidics lines are not pinched when the flow cell clamp is closed.
A pinched line can cut off reagent flow.
If the flow cell inlet line was Then run these tests …
2 Complete the post-service call tasks (Complete the Post-Service Call Tasks on page 447).
Prerequisites
Make sure that you have full access to the front of the instrument.
Equipment Required
} Allen wrench, 1.5 and 2.5 mm
} Fluidics fitting wrench
} Manifold alignment tool
} Screwdriver, Phillips-head
} Torque wrench and fittings
} Wash tray or PR2 bottle filled with DI water
Workflow
} Remove the defective flow cell manifold
} Install the new manifold
} Qualify the new manifold
} Complete the post-service call tasks
Safety
No special safety precautions are required for this procedure.
3 Using the 3 alignment pins on the TEC, position the manifold alignment tool on the
TEC.
Figure 277 Manifold Alignment Tool on TEC
5 Using a cross pattern, tighten the 4 screws that secure the TEC (2.5 mm Allen wrench).
Figure 279 Tighten Screws in Cross Pattern
2 Complete the post-service call tasks (Complete the Post-Service Call Tasks on page 447).
Prerequisites
Make sure that you have full access to the front and the left side of the instrument.
Equipment Required
} OTT
} Beaded flow cell
} Manifold alignment tool
} Y-stage alignment tool
} 2.5 mm and 3 mm Allen wrenches
} Phillips head screwdriver
Workflow
1 Remove the flow cell TEC.
2 Install the flow cell TEC.
3 Calibrate the flow cell TEC.
4 Qualify the flow cell TEC.
5 Complete the postservice call tasks.
Safety
No special safety precautions are required for this procedure.
5 Cut the RFID cable tie (if present), and disconnect the cable from the back of the RFID
board.
6 Disconnect the 15-pin and 25-pin connectors from the back of the TEC, and remove it
from the instrument.
Figure 285 RFID Cable and Cables Attached to Back of TEC
A RFID Cable
2 Using a 2.5 mm Allen wrench, loosen the 4 screws that secure the TEC.
Figure 287 Four Inner Screws of Flow Cell Holder
4 Lower the manifold until it snaps in place onto the 2 pins on the alignment tool.
Figure 289 Manifold Snaps in Place onto the Alignment Tool
A Carefully insert flat head screwdriver in these slots to move the TEC
The flow cell Peltier used in this instrument is referred to as a thermoelectric cooler (TEC).
Recalibrate the flow cell TEC as described in this procedure whenever:
} The TEC is replaced.
} The thermal ramp test fails.
IMPORTANT
For the best accuracy:
• Make sure that the skins are on the instrument.
• Close the flow cell compartment door so that no outside air can flow around
it and affect the calibration.
Prerequisites
Make sure that you have full access to the front and left side of the instrument.
Workflow
1 Prepare the instrument.
2 Attach a thermocouple to a flow cell and load onto the instrument.
3 Calibrate the flow cell TEC.
Safety
No special safety precautions are required for this procedure.
c When the graph has stabilized, record the Fluke meter reading.
d Deselect Enable.
4 Determine the high measured temperature as follows:
a Enter 96 in the Target field, and click Set.
b Select Enable.
c When the graph has stabilized, record the Fluke meter reading.
d Deselect Enable.
5 Calibrate the flow cell TEC as follows:
a In the Calibration pane, make sure that 16 and 96 are the values displayed in the
Low and High Target fields.
b Enter the Fluke meter readings in the Low and High Measured fields.
6 Do the following:
a Repeat step 3. Make sure that the value is 16° C ± 2° C.
b Repeat step 4. Make sure that the value is 96° C ± 2° C.
c If the either of the values is not within spec, the TEC was not calibrated. Repeat
this procedure.
7 Enter these values in the MiSeqOverride.cfg file:
a The high measured temperature from the Fluke meter that you entered into MTS to
calibrate the flow cell TEC (step 4).
b The low measured temperature from the Fluke meter that you entered into MTS to
calibrate the flow cell TEC (step 3).
c Save and close the file.
Figure 298 Measured High and Low Temperatures in the MiSeqOverride.cfg File
Prerequisites
Make sure that you have full access to the front and both sides of the of the instrument.
Equipment Required
} Allen wrench, 2.5 mm
} Cable ties, approximately 5
} Diagonal cutter
} Flow cell, open beaded or wash
} Fluidics wrench
} Screwdriver, Phillips-head
} Sharpie marker
} Wrench, 9/32
Workflow
1 Remove the reagent valve.
2 Remove the flow sensors.
3 Install the flow sensors.
4 Install the reagent valve.
5 Qualify the reagent valve.
6 Complete the post-service call tasks.
Safety
No special safety precautions are required for this procedure.
2 Using a cable tie, bundle the flow sensor and power/comm cables approximately 6
inches from the sensors.
d Repeat steps b and c using these Flow Rates: 1000 and 1500 .
e If each test passes, continue to the next step.
5 Run the volume test in the instrument control software, or use VCL in MTS to test
delivery for all reagent valve positions.
Prerequisites
Make sure that you have full access to the front and both sides of the of the instrument.
Equipment Required
} Allen wrench, 2.5 mm
} Cable ties, approximately 5
} Diagonal cutter
} Flow cell, open beaded or wash
} Fluidics wrench
} Screwdriver, Phillips-head
} Sharpie marker
} Torque wrench and fittings
} Wrench, 9/32
Workflow
1 Remove the reagent valve.
2 Install the reagent valve.
3 Qualify the reagent valve.
4 Complete the post-service call tasks.
Safety
No special safety precautions are required for this procedure.
5 Lift and slide out the valve assembly with the power and flow sensor cables.
6 If this is a v2 valve, grasp the rubber bushing with fluidics lines and remove it from
the bracket.
Figure 307 Rubber Bushing
3 Add one extra cable tie around the cables to hold them more securely against the
bracket.
Figure 310 Extra Cable Tie Installed
5 Grasp the black ring, and unscrew the valve head from the valve body.
Figure 312 Unscrew Valve Head from the Body
6 Transfer the plug from position 24 of the old valve to position 24 on the new valve.
8 Set the valve head down away from the installation area to provide more work space.
Figure 314 Move Valve Head Out of the Way
12 Position and secure the cables inside the instrument along the back and right sides of
the metal bracket (use cable ties to secure the cables).
CAUTION
It is critical to properly position and secure these cables. If not secured in place, these
cables will pop up during instrument operation, and will interfere with Y-stage
movement.
If you remove any cable ties, be sure to replace them using the same size cable ties.
13 Install the valve into the instrument as follows (Figure 319 on page 273):
a Slide the valve into position with the waste line routed over the valve head.
b Using a 9/32 wrench, secure the valve bracket to the chassis with two nuts and two
lock washers.
15 Manually rotate the lead screw to move the flow cell stage out approximately one inch
(provides slack to connect the inlet line).
16 As shown in Figure 319:
a Using the marks made earlier, insert the same section of waste line into the bubble
sensors.
b Secure the waste line to the bracket with two cable ties.
IMPORTANT
Use small cable ties. Larger ties can hit the reagent valve head or get caught on the
valve cables when the Y-stage moves.
c Using a torque wrench:
— Connect the inlet line to the center of the valve head between positions 1 and 3.
— Connect line A1 to position 3 on the valve head.
d Secure the waste and inlet lines to the flow cell RFID cable with a cable tie (Figure
320).
NOTE
Refer to these documents on SFDC for more troubleshooting information:
• Reagent Valve Error in the instrument control software or MTS
• MiSeq System Fluidics Troubleshooting
Prerequisites
Make sure that you have full access to the front and both sides of the of the instrument.
Equipment Required
} Allen wrench, 2.5 mm
} Cable ties, approximately 5
} Diagonal cutter
} Flow cell, open beaded or wash
} Fluidics wrench
} Screwdriver, Phillips-head
} Sharpie marker
} Wrench, 9/32
Workflow
1 Remove the reagent valve.
2 Remove the power/comm cable.
3 Install the power/comm cable and reagent valve.
4 Qualify the reagent valve.
5 Complete the post-service call tasks.
Safety
No special safety precautions are required for this procedure.
2 Bundle the flow sensor and power/comm cables approximately 6 inches from the
sensors with a cable tie.
3 Install the valve and cables back into the instrument as described under Install a
Version 3 Reagent Valve on page 266.
Prerequisites
Make sure that you have full access to the front, left, and right side of the instrument.
Equipment Required
} Cable ties
} Diagonal cutters
} Fluidics wrench
} Level
} Nut driver, 9/32
} PR2 bottle with RFID chip
} PR2 bottle filled with DI water
} Screwdrivers, flat-head and Phillips-head
} Torque wrench and fittings
} Wrenches, 2.5 mm ballpoint hex, 4.0 mm hex
} Wash tray filled with DI water
Workflow
1 Remove the sipper console.
2 Install the sipper console.
3 Qualify the sipper console.
4 Complete the post-service call tasks.
Safety
No special safety precautions are required for this procedure.
4 Disconnect all of the cables from the back of the console except for those included in
the chiller/PR2 cable assembly, PN 15029259, as follows (cable assembly shown in
Figure 327 on page 281):
a Cut the cable ties securing the flow sensor and reagent power/comm cables along
the left side of the console.
b Disconnect the flow sensor and reagent valve power/comm connectors from the
back of the console.
d Carefully disconnect and move the Z-stage cables out of the way.
e Disconnect the chiller motor cable and the reagent door cable.
f Using a flat-head screwdriver, disconnect the chiller ribbon cable (top) from the
sipper in/out cable (bottom) by pushing down and disconnecting the sipper in/out
cable first. Then remove the chiller cable from the cutout.
Figure 325 Disconnect Sipper In/Out Cable from Chiller Ribbon Cable
b On the left side of the console, remove the 3 screws that secure it to the imaging
module.
Figure 330 Screws Securing Imaging Module to Sipper Console
Figure 331 Chiller/PR2 Sensor Cable Assembly Disconnected from E-Box (Out and Bypass lines
on syringe pump should be disconnected)
11 Reconnect the Z-stage cables, and secure them with cable ties along the top back of the
sipper console (Figure 333).
Notice how the Z-stage cables are routed across the top of the console, and then looped
back and down through the cutout between the reagent door (C) and waste bottle cables
(D).
12 Reinstall and level the reagent compartment door.
Prerequisites
Make sure that you have full access to the back, front and sides of the instrument.
Equipment Required
} Allen wrench, 2.5 mm
} Cable ties, two
} Cutter, diagonal
} Fluidics fitting wrench
} PR2 bottle filled with DI water
} Screwdriver, Phillips-head
} Torque wrench and fittings
} Wash tray filled with DI water
Workflow
1 Remove the PR2 sipper tube assembly.
2 Install the PR2 sipper tube assembly.
3 Qualify the PR2 sipper tube assembly.
Safety
No special safety considerations for this procedure.
Figure 336 PR2 Cover Plate Screws and Cable Tie Locations
4 Cut the two PR2 lines, and remove the sipper tube.
Figure 338 Cutting the PR2 Sipper Lines
6 Detach the bypass line from the pump (labeled BP on old style pump; B on new style
pump).
Figure 340 Fluidics Line Labeling
8 Cut enough off the old PR2 lines so that they are not visible in the PR2 compartment.
9 Reinstall the rubber grommet on the cutout (Figure 341 on page 294).
10 Cut off the old PR2 lines close to the cutout and secure them with a cable tie.
11 Reinstall the PR2 sipper cover plate.
12 Secure the fluidics lines and cables to the cover plate with cable ties (Figure 336 on
page 291).
13 Raise and lower the sipper tube handle several times to make sure that:
• Sipper movement is smooth.
• All tubing and cable is secure, and does not interfere with sipper handle
movement.
Prerequisites
Make sure that you have full access to the right side of the instrument.
Equipment Required
} Flow cell, open or wash
} PR2 bottle filled with DI water
} Wash tray filled with DI water
Workflow
1 Remove the syringe.
2 Install the syringe.
3 Qualify the syringe.
4 Complete the post-service call tasks.
Safety
No special safety precautions are required for this procedure.
Old Style Syringe and Pump New Style Syringe and Pump
A Barrel
B Plunger
C Pump arm and lock screw
Old style (Figure 348) Use your fingertip to rotate the worm screw counterclockwise and
lower the lock screw. To avoid thread damage, do not push on the
lock screw.
New style (Figure 349) Using your fingertip, push down on the pump arm to lower the
lock screw. To avoid thread damage, do not push on the lock
screw.
A Worm screw
Old Style Syringe and Pump New Style Syringe and Pump
Pump only.
Pump only.
Prerequisites
Make sure that you have full access to the right side of the instrument.
Equipment Required
} Allen wrench, 2.5 mm
} Flow cell, open or wash
} Nut driver, 9/32
} PR2 bottle filled with DI water
} Screwdriver, flat-head
} Torque wrench
} Wash tray filled with DI water
Workflow
1 Remove the pump.
2 Install the pump.
3 Qualify the pump.
4 Complete the post-service call tasks.
Safety
No special safety precautions are required for this procedure.
Old Style Syringe Pump Assembly New Style Syringe Pump Assembly
Power Cable on Old Style Pump Power Cable on New Style Pump
8 Using a 2.5 mm Allen wrench, remove the screws that attach the pump to the bracket.
If … Then …
New style bracket Attach the bracket to the old or new style pump.
Old style bracket Attach the bracket to the old or new style pump, ensuring that it is
positioned as shown in Figure 358.
IMPORTANT: Later you will seat the pump as far back into the
instrument as possible. Positioning the old style bracket as shown is
important to prevent the plunger lock screw from rubbing against
the skin and becoming loose. When sufficiently loosened, pump
operation ceases.
3 Connect the power cable to the back of the pump (Figure 356 on page 308).
4 Slide the assembly back into the instrument, and secure the bracket to the chassis.
CAUTION
Seat the pump as far back into the instrument as possible. If the pump is too close to the
skin, the plunger lock screw might rub against the skin and become loose. When
sufficiently loosened, pump operation ceases.
5 Make sure that the other end of the power cable is securely connected to the E-Box.
Figure 359 Syringe Pump Power Cable Connection on E-Box
A Overflow port
Mechanical Procedures
Chapter 6
Air Filter Replacement and Maintenance 314
Y-Stage Replacement 316
Y-Stage Calibration 359
Noisy Y-Stage Troubleshooting and Lubrication (Ibex only) 365
Z-Stage Replacement 368
Prerequisites
Make sure that you have full access to the left side of the instrument.
Equipment Required
No special equipment required.
Workflow
} Remove the air filter
} Inspect the air filter
} Clean or replace the air filter
} Complete the post-service call tasks
Safety
No special safety precautions are required for this procedure.
Prerequisites
Make sure that you have full access to the back, front, and left side of the instrument.
Workflow
1 Visually inspect the lead screw length.
2 Remove the Y-stage.
3 Install Y-stage.
4 Qualify the Y-stage (includes setting the Y-stage hard stop to 14 tiles).
5 Complete the post-service call tasks.
Safety
Wear a properly grounded electrostatic discharge (ESD) wristband while performing this
procedure.
Detach the Flow Cell TEC and the Input and Output Lines
1 Remove the flow cell TEC shroud as follows:
a Loosen the thumb screw (Phillips-head screwdriver).
b Push down on the tray.
c Pull out the tray.
Figure 366 Flow Cell TEC Cover
7 Using a Phillips-head screwdriver, remove the fluidics line cover plate behind the
clamp.
Make sure that the lines are free and not crimped.
Figure 372 Fluidics Line Cover Plate Behind Clamp
9 Supporting the TEC with one hand, remove the 4 screws that attach the flow cell TEC
to the instrument using a 2.5 mm Allen wrench.
IMPORTANT
The TEC will fall when the screws are removed. Make sure that you support the TEC,
and set it down gently when the screws are removed.
10 Cut the RFID cable tie (if present), and disconnect the cable from the back of the RFID
board.
Figure 375 RFID Cable Disconnected
11 Disconnect the 15-pin and 25-pin connectors from the back of the TEC, and remove it
from the instrument.
2 Using a 2 mm Allen wrench, remove the 3 screws on the bottom of the RFID board.
Figure 378 Flow Cell RFID Board
3 Cut the cable tie that bundles these cables: Y-stage power, Motor M3/C2, LED Fan.
Figure 382 Cable Tie
4 Using the M6 T-handle Allen wrench, lock the imaging module so that you can
disconnect the cables in the next step.
Figure 383 Imaging Module Locked
2 Using an M3 wrench, loosen the 2 front adjustment screws on the imaging module.
Figure 395 Loosen the Adjustment Screws
4 Slide each pin outward and away from the stage; do not remove the pins.
Figure 397 Slide Pin Outwards
c Slide your arm under Y-stage and imaging module for support and:
— Remove the standoffs.
— Slowly lower the assembly onto the chassis, making sure that no cables are
caught. Avoid bumping the flow cell compartment mount (Figure 391 on page
329).
— Check again to make sure that no cables are caught.
12 Reinstall the flow cell TEC, RFID card, and antenna as follows:
a Screw the RFID standoffs onto the bottom of the Y-stage.
b Position the RFID board and antenna on the Y-stage.
c Secure the RFID board to the standoffs (the connector faces the rear of the
instrument; 2 mm Allen wrench).
f Slide the input and output lines up through the hole in the clamp assembly.
Figure 414 Fluidics Line Hole
g Reattach the input and output lines to the flow cell manifold.
h Reinstall the springs.
i Position the lines in the grooves and resintall the rear fluidics line cover.
Make sure that the lines are free and not crimped.
j Reinstall the flow cell TEC as follows:
4 Disconnect the input and output lines from the manifold using the fluidics wrench.
5 Remove the input and output line cover plate using a Phillips-head screwdriver
(behind the flow cell clamp; see also Figure 266 on page 230).
Figure 421 Input and Waste Line Cover Plate
2 Disconnect the optical sensor/encoder cable and Y-motor power cable at the junction
box.
The optical sensor/encoder is the large connector and cable; the Y-motor power cable is
the located directly beneath it.
Note how the tubes are routed through the instrument so that you can reroute them
during the Y-stage installation. Then, tuck the cables under the stage to keep them out
of the way.
3 Loosen the lock screws for the Y-stage adjustment knob using a 3 mm Allen wrench.
Figure 427 Y-Stage Adjustment Lock Screws
c Lift up the imaging module, then engage the lockdown rod using an M6 Allen
wrench.
d Lower the imaging module so that the hanger on each side rests on the lockdown
rod.
Figure 431 Imaging Module Lifted and Resting on Lockdown Rod
7 While supporting the Y-stage, slide both the left and right sliding pins until they are
flush with the inner edge of the imaging module.
Figure 433 Remove Sliding Pin
8 Gently slide the stage out through the front of the instrument, being careful not to catch
and pull any tubing or cables.
NOTE
It helps to use a tool to line up the sliding pin holes before reinstalling the screws.
See Figure 402 on page 336 for an example of what you can do.
5 Lower the imaging module into place as follows:
a Lift up the imaging module.
b Disengage the lockdown rod (M6 Allen wrench).
c Lower the imaging module into place on the isolator collars.
6 Lock the 3 isolator collars (M4 driver).
If you had to detach the reagent valve bracket, reattach it now.
NOTE
Do not secure the LED cable harness in the bottom clamp.
If the Average
ΔZ value Then …
calculated is …
The purpose of this procedure is to calibrate the Y-stage tile 1 and hard stop positions, and
update those values in the MiSeqOverride.cfg file.
Items Required
} 15059843_A_MiSeq2Y-stageCalibrationWorksheet.xlsm
Procedure
These instructions assume that you are at tile 1 on the OTT.
1 Open 15059843_A_MiSeq2Y-stageCalibrationWorksheet.xlsm, and select the Y Stage
tab.
2 Open the MiSeqOverride.cfg file and:
a Change these parameters:
— YStageHomeToEdgeOfSlide = 54.
— actualedgeofslide = 57.4.
Figure 442 System Measurement Parameters
b To locate tile 1, jog the Y-stage in decreasing increments starting with 0.5 mm until
you see the pattern in Figure 444.
c Fine tune the Y-position by 0.01 mm increments until the upper edge of the top-
most stripe is near 2600 pixels (Y=2600 pixels, ± 200 pixels).
d Enter the MTS Y-pixel value into line 24 of the worksheet.
e Enter the new Y Motor Position value from MTS into line 21 of the worksheet.
Figure 444 Y-Motor Position Value
8 Enter the newly calculated values from the worksheet (lines 26, 27, and 28) for these
parameters into the MiSeqOverride.cfg file:
• YStageHomeToEdgeOfSlide
• actualedgeofslide
• SoftLimitPositiveMm
9 If present in the MiSeqOverride.cfg file, delete the parameter
retractpositionforchemistrymm.
10 Save and close the MiSeqOverride.cfg file.
12 Using a 2.5 mm Allen wrench as a shim, adjust the gap between the hard stop screw
and the contact point of the Y-stage to 0.1 inches as follows:
NOTE
On instruments with the Parker stage, the hard stop is a block. On instruments with Ibex
stages, the hard stop screw is on the left or right. The gap needs to be set to 0.1 inches.
You can use a 2.5 mm Allen wrench as a shim.
A Hard stop block—loosen both screws and move the block to adjust the gap
B 0.1 inch gap
C Y-stage
a Loosen the hard stop lock screw (loosen both screws on Parker stages).
b Insert a shim (flat sides of a 2.5 mm Allen wrench) between the stage and the
hard stop adjuster.
c Move the hard stop screw in or out using a long-handled Allen wrench.
You can insert the wrench through the back or front of the instrument
depending upon the orientation of the hard stop screw.
d Tighten the lock screw such that the gap is maintained, and the shim can be
removed.
e Remove the shim (the 2.5 mm Allen wrench).
The same as the pattern shown here Continue to the next step.
Not close to the pattern shown here An error was made. Repeat the procedure
Y-Stage Calibration on page 359.
6 To verify, recalculate the values for tile 1 and the hard stop as follows:
a Enter the Y Motor Position value (Motor tab in MTS) into line 45 of the
worksheet.
Figure 449 Verify Calibration
If there is … Then …
CAUTION
There must be a gap between the hard stop screw and the Y-stage. If there is no gap,
damage will occur to the Y-stage.
9 Once the hard stop is good, return to the next step under Align and Qualify the Y-Stage
on page 357.
IMPORTANT
This procedure is for the Ibex Y-stage only. Do not lubricate Parker Y-stage lead screws.
Lubricating the lead screw on Parker Y-stages can cause the stage to stop moving.
Prerequisites
} Make sure that you have full access to the back, front and sides of the instrument.
} Look up the SoftLimitPositiveMm value in the MiSeqOverride.cfg file.
Equipment Required
} Lubricant, Krytox, part # 210981S
} Flow cell, super flat
Workflow
1 Apply lubricant to the Ibex Y-stage lead screw.
2 Qualify the Ibex Y-stage.
Safety
No special safety considerations for this procedure.
An Ibex Continue to the next step, and lubricate the lead screw.
6 Move the stage to the opposite end by entering the SoftLimitPositiveMm value
from the MiSeqOverride.cfg file into the Y motor position field.
7 Apply 0.1–0.3 ml of Krytox lubricant in < 15 mm length at the other end of the lead
screw.
Figure 451 Apply Lubricant to Opposite End of Ibex Lead Screw
8 Distribute the lubricant by moving the Y-stage between 0 and the SoftLimitPositiveMm
position 10 times.
9 If the Y-stage is still noisy, contact SIS.
10 Reinstall the instrument skins.
Prerequisites
Make sure that you have access to the front, back, and left side of the instrument.
Equipment Required
} Allen wrench, 3 mm
} Cable ties, 10
} Cable cutter
} ESD with ground strap
} MiSeq Piezo ground cable
Order from SFDC: MiSeq,Ground cable,P-603K010 Piezo Z-Stage, 15056887
} Phillips-head screwdriver
Workflow
1 Remove the Z-stage control board.
2 Install Z-stage control board.
3 Remove the Z-stage.
4 Install the Z-stage.
5 Qualify the Z-stage and control board.
6 Calibrate the Y-stage at tile 1.
7 Complete the post service call tasks.
Safety
No special safety precautions are required for this procedure.
1 Close all applications, then 1) shutdown the PC, and 2) turn off the instrument.
2 Put on an ESD and ground the strap.
3 Drain the energy from the Z-stage Piezo as follows:
a If connected, disconnect the Z-Stage Piezo Voltage and SGS cables from the E-Box.
Figure 452 Piezo and SGS Cables
b Connect the MiSeq Piezo ground cable connector (15056887) to the Z-stage Piezo
Voltage cable (Figure 453) for 30 seconds.
Figure 453 Z-stage 3-Pin Piezo Voltage Cable
c Disconnect the ground cable from the Z-stage Piezo Voltage cable.
d Reconnect the Z-Stage Piezo Voltage and SGS cables to the E-Box.
A Piezo
B SGS
5 Carefully reinstall the four connectors onto the board (Figure 454 on page 370).
Initializes and homes Proceed to the next step, and reinstall the compensator.
Does not initialize and home 1. Drain the energy from the Z-stage Piezo as described
under Drain the Energy from the Z-Stage on page 369 using
the MiSeq Piezo ground cable (156887; order from
SFDC).
2. Repeat this step until the instrument initializes and
homes.
4 Reinstall the compensator (Install the Compensator on page 28; make sure that the
clearance is adequate).
5 Turn on the instrument, and make sure that it initializes and homes.
6 Adjust the rough Z-motor position (Z-Motor Rough Position Adjustment on page 431).
7 Calibrate the Y-stage tile 1 and hard stop locations (Y-Stage Calibration on page 359).
8 Reinstall the optical covers and instrument skins.
Qualify the Z-Stage
Follow these steps to qualify the Z-stage. The Z-stage is qualified when all the tests pass,
and the best focus Z-position is within spec.
1 Run the Z-stage test using the instrument control software or MTS.
Adjustment Procedures
Chapter 7
Imaging Module Components 380
Camera 2 Center Adjustment 382
Camera XY Offset Adjustment 386
Chromatism Adjustment 391
Compensator Adjustment 394
Determine the Best Focus Z–Position 396
Focus an OTT 398
Illumination Footprint Check 402
Illuminated Area Adjustment 406
Lane Center Adjustment 412
Lane Rotation Adjustment 416
Check the LED Power 419
Tip/Tilt 420
Z-Motor Rough Position Adjustment 431
3 M3 Focus 13 F4
(Sensor 1 X/Y tilt)
4 Red LED 14 M4
5 Green LED 15 M1
21 Compensator
If the … Then …
Top-bottom value and/or the left-right Adjust camera 2 (see Adjust Camera 2
value is > 0.20 μm Center on page 383).
Equipment Required
} Indelible marker or other type of marker
} Vertical adjustment tool
Procedure
TIP
Using an indelible marker or other type of marker, draw a line to use as a reference before
making an adjustment.
1 Remove the instrument skins.
2 Adjust camera 2 in the X direction as follows:
a Loosen the X-axis lockdown screw (Figure 467 on page 384).
b Move the camera left or right towards the highest left-right value.
Example Figure 468 on page 384:
Because 1.91 is the highest left-right value; you would move the camera left.
c Tighten the X-axis lockdown screw.
d Run BoltCameraCenteringD50.exe and review the image.
e Continue adjusting the camera position and rerunning
BoltCameraCenteringD50.exe until camera center is within spec.
3 Adjust camera 2 vertically as follows:
a Install the vertical adjustment tool shown in Figure 467.
b Loosen the Y-axis lockdown screw.
c Turn the micrometer on the adjustment tool so that the camera moves in the
direction of the highest top-bottom value.
NOTE
If turning the micrometer counterclockwise, move up the camera flush against the
tool before tightening the Y-axis lockdown screw.
Example Figure 468:
Because 1.97 is the highest top-bottom value; you would move the camera down.
d Tighten the Y-axis lockdown screw.
e Run BoltCameraCenteringD50.exe and review the image.
f Continue adjusting the camera position and rerunning
BoltCameraCenteringD50.exe until camera center is within spec.
NOTE
In this particular image, the camera is already centered. The image is being used to
illustrate the direction in which you would move the camera when center is out of spec.
Equipment Required
} Cam adjustment tool (for camera X-offset)
} Camera vertical adjustment tool
To the left 1. Place the cam adjustment tool in 1 of the holes to the right
of the camera (Figure 474 on page 390).
2. Turn the tool clockwise to push the camera to the left.
3. Cover the imaging module with a blackout cloth.
4. View the image, and make further adjustments until the
target is on the cursor.
Be sure that the appropriate camera is turned off.
5. Tighten the lockdown screw.
To the right 1. Place the cam adjustment tool in 1 of the holes to the right
of the camera.
2. Turn the tool counterclockwise 1/4 to 1/2 turn to create a
space, leaving the tool in place.
3. Move the camera to the right up against the tool.
4. Cover the imaging module with a blackout cloth.
5. View your image and make further adjustments until the
target is on the cursor.
Be sure that the appropriate camera is turned off.
6. Tighten the lockdown screw.
Equipment Required
} L1 adjustment tool
} Phillips screwdriver
Run BoltChromatism.exe
1 Load a super flat beaded flow cell.
2 Using MTS, move to any tile.
3 Generate 2 through-focus image sets (top and bottom of flow cell).
You can use any tile to extend the life of the flow cell. See Generate Through Focus Image
Sets (Stacks) on page 502.
4 Run BoltChromatism.exe for each image set.
See Running Scripts in MTS on page 500 for instructions.
5 View the results files.
A in Figure 475 indicates that the instrument has 15 microns of chromatism.
6 If chromatism is out of spec, adjust L1 as recommended in the last line of the results
file (B in Figure 475).
IMPORTANT
BoltPAC.exe chromatism results include only the information in the row labeled A in
Figure 475. The direction (negative or positive sign) of the recommended correction is
the opposite of the correct direction. BoltChromastism.exe results include the
information in rows A and B. The direction of the adjustment in the row labeled B is
correct.
A BoltChromatism.exe result
B Adjustment recommended to improve chromatism
Adjust L1
1 Remove the instrument skins.
2 Install the L1 adjustment tool as follows:
a Make sure that the micrometer is installed as shown in Figure 476 so that you can
use the horizontal line on the barrel as a reference when turning the knob. Loosen
the nut to reposition the barrel.
b Position the tool in groove on base plate.
c Loosely tighten the adjustment tool lockdown screw.
d Tighten the micrometer to the point where it touches the lens and the ball spring is
slightly compressed.
e Lock down the tool.
Figure 476 L1 Adjustment Tool Installed
5 Tighten the lens lockdown screw, and back out the micrometer.
6 Reinstall the optical covers (no need to tighten them down).
7 Cover the imaging module with the blackout cloth.
8 Generate new through-focus image sets.
9 Run BoltChromatism.exe and assess the results.
10 Select from the following:
A Compensator adjuster
B M3 screw
C M4 screws
6 Holding the compensator firmly against the Z-stage, tighten down the M4 screws.
7 Make sure that the compensator is firmly in place by trying to move it.
8 Using MTS or the instrument control software, check tip/tilt on the bottom surface of a
super flat beaded flow cell.
b Click Zoom All, then click Contrast to view the image, and look for a small black
crosshair.
— If the image is not usable, try the following:
— Adjust the exposure time (for example, 400, 600).
— If the G channel is not usable, try other channels and exposure times until
the image quality is acceptable. You can also try using Camera 1.
— If no black crosshair is visible, try moving the upper and lower contrast bars.
NOTE
The Contrast button in the Image pane works as follows:
1. It measures the foreground and background based on the Z-motor position.
2. It adjusts the contrast to what it calculates as best focus.
If the Z-motor position changes by a large value, or changes several times, click the
Contrast and the Zoom All buttons again to “reset” the software for the new position.
Also click these buttons if you adjusted the position and no image is displayed.
11 Select the Motor tab, and focus the image as follows:
a Zoom in on the cross.
b Move the Z-motor up or down to bring the cross into good focus.
Increments of 0.002–0.0005 typically work well.
e Move the right slider all the way down, and make sure that beads appear across
the bottom of the image (no blank strip across the bottom of the image).
9 Select from the following:
Within spec The illumination footprint is OK, record the result on the
IQ/OQ.
Out of spec Adjust the illuminated area until it is within spec (See
Illuminated Area Adjustment on page 406).
Notice the X value on the right side of the image is 3212, which is < 3620.
Figure 493 Illuminated Area Out of Spec in X
3 If you are unable to bring the illumination area into spec by adjusting L8:
a Measure the LED power as described under Check the LED Power on page 419.
b If the power is too low, replace the LED assembly.
4 Attach CSV files for both surfaces to the work order.
5 Complete the post-service call tasks (Complete the Post-Service Call Tasks on page 447).
Out of spec for lane center 1. Adjust lane center by following the instructions
under Lane Center Adjustment on page 413.
2. Record the lane center values on the IQ/OQ.
3. Proceed to the next step.
Out of spec for lane 1. Adjust lane rotation by following the instructions under
rotation Lane Rotation Adjustment on page 416.
2. Record the lane rotation value on the IQ/OQ.
3. Change the exposure value back to 200 (Camera tab).
Equipment Required
} Blackout cloth
} OTT
3 Cover the imaging module with the blackout cloth, making sure that the cloth does not
cover the rear fan.
Equipment Required
} OTT
} Beaded flow cell
} Y-stage alignment tool
} 2.5 mm and 3 mm Allen wrenches
} Phillips head screwdriver
6 Using a flat head screwdriver, carefully move the TEC so that the 3 TEC alignment
pins are seated up against the alignment tool.
A Carefully insert the flat head screwdriver in these slots to move the TEC
Check Tip/Tilt
Use the instrument control software or the instrument control software to check tip/tilt (also
referred to as tile tilt). Always use a super flat beaded flow.
If using … Then …
MTS Navigate to the results folder that you created earlier for each
through-focus image set, double-click the image for Camera 1,
and assess the results.
If tip/tilt is … Then …
out of spec and a manual 1. Review the Brenner Metrics images to make sure that
adjustment are required the through-focus image set was generated
successfully (Review the Brenner Metrics Images on
page 423).
2. If tip/tilt must be adjusted, follow the instructions
listed under:
• Tip/Tilt Adjustment—Camera 2 on page 425 or
• Tip/Tilt Adjustment—Camera 1 on page 429
Figure 516 illustrates failed through-focus image set generation. A graph with this
appearance can be caused by not optimizing the Z-motor position before generating the
through-focus image set. Repeat the procedure listed under Generate Through Focus Image
Sets (Stacks) on page 502 to optimize the Z-motor position and generate a new set of
through-focus images. Then run BoltTileTilt.exe again.
Figure 516 Failed Through Focus Image Set Generation Due to Instrument Vibration
Tip/Tilt Adjustment—Camera 2
When manually adjusting tip/tilt for camera 2, you essentially change the focal plane of the
sample as you make your adjustments. Loosening and tightening the lockdown screws for
the optical stage causes the Y-stage to shift.
If … Then …
the results indicate that tip/tilt is out of continue to Adjust Tip/Tilt on page 426.
spec
Adjust Tip/Tilt
1 Remove the instrument skins.
2 Make sure that the spring loaded screw and the micrometer are touching the optical
stage.
Figure 518 Micrometer and Spring Loaded Screw Just Touching the Optical Stage
A Back actuator
B Front actuator
6 Tighten all of the lock screws to barely snug (do not fully tighten).
a Gently tighten down each screw the same amount until barely snug.
b Gently tighten down each screw a little more by the same amount.
If tip/tilt … Then …
Is out of spec for the top surface of the flow Return to step 3 and repeat this procedure.
cell
Passes for the top surface of the flow cell, but 1. Adjust the compensator (Compensator
is out of spec for the bottom surface Adjustment on page 394).
2. Repeat this procedure for the bottom
surface only.
A Screw actuator
B Cam actuator
C Locking screw
3 Load a super flat beaded flow cell and move to the center of the flow cell.
Center = actualedgeofslide + 14 mm
4 Cover the imaging module with a blackout cloth.
5 Run the test again on both surfaces of the flow cell (Check Tip/Tilt on page 420).
6 Repeat steps 2 through 5 until tip/tilt is within spec.
7 If the cam actuator was rotated, lock it down now.
8 Run the tip/tilt test again for both surfaces of the flow cell, and make sure that tip/tilt is
still within spec.
9 If any settings were modified (eg, camera exposure, LED current), restore the default
values now.
10 Reinstall the instrument skins.
b Open the MiSeqOverride.cfg file and find the value for actualedgeofslide.
A Adjustment screw
B Straight and angled lockdown screws
C Insert Allen wrench through this opening to loosen the angled lockdown screw
Beads that are in You are on the bottom surface of the flow cell. Confirm by
relatively good focus 1. Click the up arrow one time to move back to the top
surface (correct result: you still see focused bead clusters.).
2. Click the up arrow 1 more time. Correct result: No bead
clusters are visible.
The Z-motor position is OK.
If the beaded flow cell is focused on the wrong (bottom) surface of the flow cell, a Brenner
graph similar to Figure 530is generated.
Figure 530 Brenner Graph Indicating Z-Stage Focused on Wrong (Bottom) Surface
Appendix A
Specifications
Instrument Specifications 438
NOTE
The following specs are different for the top and bottom surface of a lane:
• Camera 1–Camera2 Z Offset
• Tip/Tilt
Max D50 < 2.90 pixels < 2.90 pixels < 2.90 pixels < 2.90 pixels
Common Procedures
Appendix B
Adjust the Clock, Date, and Time (Windows 7) 440
Cleaning the Imaging Module Bench 442
Cleaning Optical Components 443
Complete the Post-Service Call Tasks 447
Flow Cell Cleaning, Loading, and Imaging 448
Removing and Reinstalling the Instrument Skins 453
Rebooting the Computer 461
Shutting Down the Computer 462
Synchronize the Instrument with an Internet Time Server 463
Turning the Instrument On and Off 464
1 Select the time/date displayed in the lower right corner of the screen.
2 Select Change date and time settings.
Figure 531 Open the Clock
3 Select Change time zone to open the time zone settings window.
4 Open the drop-down menu and select the appropriate time zone.
5 Deselect this checkbox: Automatically adjust clock for Daylight Saving Time.
6 Select OK.
Figure 532 Automatic Adjustment for Daylight Saving Time Off
NOTE
If the date and time are not correct, the connection to BaseSpace Broker fails.
Items Required
} Air puffer
} Flashlight (LED preferred)
} Hemostat tool (optional)
} HiSeq Optics Cleaning Kit available in SFDC. This kit includes:
• Swabs, microfiber
• Purosol Optical Cleaner
• Cleaning Wipes
} Latex gloves, powder-free
If the contamination is dust or something similar, use the air puffer to blow off particles
before using a chemical for cleaning.
} For the objective lens, look at it at a 45° angle and at a 90° angle as shown here.
Figure 535 Inspecting the Objective Lens
g For an objective lens, use an optical quality swab or an optical quality Q-tip. Use a
circular motion to clean the top surface of the objective lens.
3 Lightly rinse the flow cell with labaoratory-grade water until both the glass and plastic
cartridge are thoroughly rinsed of excess salts.
Excess salts can affect flow cell seating on the instrument. If salt drys in the imaging
area, imaging can also be affected.
Figure 542 Rinse the Flow Cell
5 Clean the flow cell glass with an alcohol wipe, making sure that the glass is free of
streaks, fingerprints, and lint or tissue fibers.
NOTE
Do not use the alcohol wipe on the flow cell port gasket. Make sure that the
cartridge is securely closed before use.
5 Gently press down on the flow cell clamp to close it over the flow cell.
As the flow cell clamp closes, alignment pins position the flow cell. An audible click
indicates that tthe flow cell clamp is secure.
Figure 547 Close the Flow Cell Clamp
Equipment Required
} Allen wrench, 4 mm
} Screwdrivers, Phillips-head, and flat-head
Skin Removal
Remove the skins (also referred to as panels) from the instrument in the order indicated
here.
Figure 549 MiSeq Skin Removal
1 Using a 4 mm Allen wrench, remove the captive screw, centered in bottom of skin
(panel).
2 Pull the panel out and up.
Two clips are on the bottom, inner side of the panel. These clips are the same on the
left and right panels.
1 Using a 4 mm Allen wrench, remove the captive screw, centered in bottom of panel.
2 Pull the panel out and up.
Two clips are on the bottom, inner side of the panel.
Figure 554 Proper Position to Rest the Top Cover on the Floor
Figure 556 Flow Cell Compartment Skin—Shoulder Screws and Slide Lock Holes
Skin Reinstallation
Reinstall the skins in the reverse order in which they were removed.
NOTE
The skins are designed to reseat easily. Do not try to force any skins back on. Forcing the
skins can create torque on internal components, and can affect optical alignment. If the skins
do not seat easily, especially the top skin, check the cabling in the affected area. Cables can
shift or be moved during repair or alignment.
TIP
Moving the flow cell stage to tile 12 makes it easier to reinstall the top skin and the flow cell
compartment skin.
} Flow cell compartment skin—Remember to:
• Remove the interlock defeat cable and reattach the cable on the flow cell
compartment skin to the connector on the chassis.
• Open the reagent chiller door when reinstalling this skin.
TIP
Curl one hand under the instrument, and guide both shoulder lock screws into the slide
lock holes (Figure 556 on page 458).
} Top skin—Fit it onto the back of the instrument first:
• Slide the slot on the back of the cover over the captive screw on the back of the
instrument.
• Position the bottom of the cover in the track, and slide it to the left so that the
cutout fits around the cables.
CAUTION
Be careful not to pinch any cables when installing the top skin.
Follow these steps to synchronize the instrument to an internet time server by enabling the
internet time feature.
1 From the Start menu, open the Control Panel, then select Clock, Language, and
Region.
2 Select the Internet Time tab.
3 Select Change settings.
4 Do the following:
a Select Synchronize with an Internet time server.
b In the Server field, select time.windows.com.
c Select Update now.
d Select OK.
5 On the Date and Time dialog box, select Change time zone.
6 Select Automatically adjust clock for Daylight Saving Time.
Figure 561 Automatically Adjust for Daylight Saving Time
7 Select OK.
8 Restart the BaseSpace Broker service.
Appendix C
Data Required For Troubleshooting 466
4 Navigate to the D:\Illumina Maintenance Logs folder, and copy these log files:
• Green LED Log.csv
• Red LED Log.csv
• Temperature Log Chiller.csv
• Temperature Log Flowcell #1.csv
Figure 565 Illumina Maintenance Log Files
5 Collect the MiSeq Reporter (MSR) output files and save in the desired location as
follows:
a Navigate to D:\Illumina\MiSeqAnalysis folder, and copy these files:
— MiSeq Reporter Error.txt
— MiSeq Reporter Log.txt
b Navigate to D:\Illumina\MiSeqAnalysis\<run folder> folder, and copy
these files:
— AnalysisError.txt
— AnalysisLog.txt
— Samplesheet.csv
Appendix D
Backflushing Fluidics Lines 470
Materials Required
} Flow cell, used or open
} Tissue such as Kimwipes
} Wash tray
} Wash bottle filled with DI water
9 Make sure that DI water is emitted onto the tissue from the correct sipper tube.
Refer to the positions marked on the wash tray to identify which reagent valve position
corresponds to each sipper tube.
SBC does not power on: Verify the main connector is latched securely to the SBC
board.
Verify the 12 V power cable is latched securely to the SBC
board.
Verify the ATX/AT Mode Jumper is set to AT Mode.
MiSeq Control Software will Verify COM port settings in config files.
not load: Verify the ribbon cable is connected to the correct header on
the SBC board.
Service Tools
Appendix E
Service Tools 476
Toolkits
The following toolkits are used to install, refresh, and service MiSeq instruments:
} MiSeq Alignment Toolkit (part # 15026398S)
} MiSeq Generic Tool Box (part # 15030145S)
} Field Service Repair Kit (part # 15028821S)
Interlock 15029220 1
override cable
Chromatism 15023706 1
adjustment tool
Manifold 15032247 1
alignment tool
Y-stage 15029808 1
alignment tool
Level 15042044 1
• MiSeq 15026129 1
Beaded Flow
Cell, open
• MiSeq 15026717 1
Beaded Flow
Cell, sealed
Syringe 1
(SPARE, SYRINGE, XC PUMP,
500 μl)
NOTE: Old style syringe here.
New style syringe in Syringe
Replacement on page 299.
Syringe Pump 1
(SPARE, SYRINGE PUMP, Tecan
XC, CERAMIC 3-WAY
DISTRIBUTION)
NOTE: Old style pump here. New
style pump in Syringe Pump
Replacement on page 304
Chiller Thermistor 1
Appendix F
Returning an Instrument to Illumina 482
Prerequisites
For instructions on the following prerequisites, see Ordering the Crate Kit on page 482.
} Place a service order
} Order the crate kit
Items Required
} Allen wrenches, M6, and 5/64 inch T-handle
} Box for instrument accessories
} Crate kit
} Shipping request form
Gather the instrument accessories and place them in a box. Ship the accessory box in the
crate with the instrument. Accessories include:
} Instrument safety and compliance guide
} Instrument user guide
} Wash tray (1 or 2 depending upon when the instrument was shipped)
} PR2 bottle
} Waste bottle with cap
} M6 T-handle hex wrench
} 5/64 inch T-handle hex wrench
5 Retract the Y-stage all the way into the instrument by manually turning the shaft.
Figure 568 Retract Y-Stage Into Instrument
4 Place the sides of the crate in position and lock them in place.
Figure 576 Position and Lock Sides of Crate
Appendix G
Cable Interconnect Diagram for the MiSeq 490
Subsystem Block Diagram for the MiSeq 491
Optical System for the MiSeq 492
Fluidics System for the MiSeq 493
Appendix H
About MTS 496
Running Scripts in MTS 500
Generate Through Focus Image Sets (Stacks) 502
About the BoltPAC.exe Script 507
Running Bolt Scripts Offline 510
Using MTS
MTS is used for some instrument installation activities, and for service procedures. This
section includes a brief description of MTS, and includes instructions on how to run the
MATLAB scripts. You can use MTS to run the Bolt scripts used to check instrument
alignment (automated in the first time setup screens). Most of these tests can also be run
from the instrument control software (Welcome screen | Manage Instrument).
To begin using MTS:
1 Double-click the MTS icon on the desktop.
Each view has a tab. However, all the tabs are not visible across the top of the left pane. If
a tab is not displayed, you can access it as follows:
1 Open the quick access view drop-down menu and selecting the tab.
2 Click the Views button and selecting the tab.
You can also hide tabs by deselecting them in the Views drop-down. If you hide a tab, it is
not displayed in the tab drop-down list.
Figure 579 Accessing Views (Tabs) in MTS
Select a channel Click the channel button so that the color of the button is
solid (A channel). A channel is off when the top of the
button fades to white (G channel).
4 Double-click the script to open a window with a command prompt (Figure 582).
5 Go to the D drive and open the folder with the top surface through-focus image stack.
Find Best Focus on the Top or Bottom Surface of the Flow Cell
NOTE
When generating an image stack for the bottom surface of the flow cell, it is good practice to
focus the beads on the top surface first. Then move the compensator in, and focus the beads
on the bottom surface of the flow cell.
1 On the Motor tab, set the Mirror Position to Normal for Imaging.
2 Type –0.17 in the Z Motor Position field, and hit Enter.
3 Open the LED tab, change the Green LED Default Snap Curr to 2000, and click Set.
The default snap current of 1500 for the red LED is OK.
Figure 585 LED Tab
6 Using the wheel on the mouse and the sliders, zoom in on an area of beads.
See Image Tab in the Right Pane on page 498 for additional information.
Figure 587 Example of Well Focused Beads
7 On the Motor tab, bring the beads into sharp focus by moving the Z Motor up or down.
Increments of 0.002–0.0005 typically work well.
8 Confirm that the value for best focus is within spec.
9 Turn off the camera.
• Z-Center field: paste the Z-Motor Position value copied from the Camera tab
• Select the Save Images checkbox.
• Deselect Compute Best Focus.
• Output Directory field: browse to the D drive and create a folder for the image
stack (eg, D:\Bottom\20July2012_OutputFolder)
Installation Prerequisites
1 Download MCRInstaller_x64_v14.exe, and install it on the local PC.
Matlab compiler runtime and vcredist_x64 are installed.
NOTE
This installer only installs on a 64-bit PC. If it is run on a 32-bit PC, an error
reading "The installation package is not supported by this processor type." is
displayed. Although there is a 32-bit MCRInstaller, the Bolt scripts themselves
are compiled for a 64-bit system.
The Matlab compiler required is v14, not v9 which is used to run Mockingbird
on HiSeq and HiScan instruments. The two versions are not cross-compatible.
2 Download the Bolt scripts and unzip them on the local PC.
Verification
Appendix I
Verification Run Requirements 512
Prepare the Reagent Cartridge 513
Prepare the PhiX Library 515
Perform a Post-Run Wash 517
Pipettes and pipette tips (P10 or P20, P200, As required Customer supplied
and P1000)
4 Allow the reagent cartridge to thaw in the room temperature water bath for at
approximately 60 minutes or until thawed.
5 Remove the cartridge from the water bath and gently tap it on the bench to dislodge
water from the base of the cartridge.
6 If not using the reagent cartridge immediately, place on ice or in 4° C refrigerator.
7 Invert the reagent cartridge ten times to mix the thawed reagents.
8 Visually inspect the USM-2 and IMT-1 reagents to make sure that they are fully mixed
and free of precipitates.
9 Gently tap the cartridge on the bench to remove air bubbles in the reagents.
WARNING
A component in this set of reagents contains formamide, an aliphatic amide that is a
probable reproductive toxin. Personal injury can occur through inhalation, ingestion,
skin contact, and eye contact.
Dispose of containers and any unused contents in accordance with the governmental
safety standards for your region.
The SDSs for this kit are available at www.illumina.com/sds.
Table 22 Reagent Cartridge Map
Position Reagent Position Reagent
Items Required
Table 23 Items Required for Post-Run Wash
Items Quantity Part Number
Appendix J
Using a Torque Driver on Fluidic Fittings 520
Troubleshooting SBC Board Upgrade 522
Overview
When replacing any fluidics-related part (for example, a reagent valve, front/rear manifold,
any fluidics line), use the torque driver to help achieve the most consistent and highest-
quality fluidics connections possible.
Refer to this section to torque the fluidics fittings to the required spec correctly.
Requirements
} Torque driver, part # 15050210S
} 6-40 Fitting Adapter (small green fitting), part # 15045811S
} ¼-28 and 10-32 Fitting Adapter and (larger brown/tan/blue fitting), part #15036791S
Procedure
1 Identify the type of fitting and thread to be connected and find the recommended torque
spec from the following table.
• Fitting Type
— Flare: Clear plastic flat disc at the end of tubing; fitting consists of nut and thin
metal washer ring
— Swage: Opaque yellowish plastic ferrule at the end of tubing; fitting consists of
nut and metal ring (larger than metal washer on flare fittings) that compress
onto ferrule
• Thread Type
— 6-40: Smaller green nut; often used on the 24-position reagent valves and
manifolds.
— 1/4-28: Larger tan/brown nut; often used on the 10-position reagent valves and
syringe pumps.
Table 24 Torque Recommendations
Fitting Thread Recommended
Image
Type Type Torque Spec
2 Choose the appropriate fitting adapter and place it on the end of the torque driver.
• For 6-40 fittings, use adapter part # 15045811S
• For ¼-28 and 10-32 fittings, use adapter part # 15036791S
3 Set the torque driver to the recommended spec from the previous table (adjustment
procedure instructions are included in box that the torque driver is shipped in):
a Pull down on the red knob to engage the adjusting mechanism.
b Rotate the handle until the required torque is selected on the calibrated scale.
— Numbers along the vertical shaft indicate the 0.5–4.0 in-lb settings.
— Numbers around the circular base (0–0.9) indicate the decimal place of the
torque setting.
NOTE
When setting the torque, the vendor recommends always starting below the desired setting
and adjusting the value upward. So if adjusting to a lower setting from a higher setting,
rotate the handle counterclockwise past the desired setting, then increase torque clockwise to
the required value.
4 Put the adapter on the fluidic fitting and turn the torque driver clockwise. When the
driver slips, the fitting is torqued correctly.
SBC does not power on: Verify the main connector is latched securely to the SBC
board.
Verify the 12 V power cable is latched securely to the SBC
board.
Verify the ATX/AT Mode Jumper is set to AT Mode.
MiSeq Control Software will Verify COM port settings in config files.
not load: Verify the ribbon cable is connected to the correct header on
the SBC board.
Indonesia 0078036510048