Professional Documents
Culture Documents
CS-400 User Manual PDF
CS-400 User Manual PDF
Instruction:
Dear user, thanks for purchasing our CS-400 Auto-Chemistry Analyzer.
Please read the user manual carefully in order to operate the instrument correctly. Incorrect operation may affect
the precision and accuracy of the test results, or endanger personal safety.
Please keep the user manual safely for your any time reference.
Note:
● Instrument should be operated by medical inspection specialist, physician, nurse or lab assistant who are
specially trained.
● Instrument should be controlled by special software. Please install the software that is appointed by our
company. Installation of other software/hardware may interferer normal operation. Don’t operate other software
when instrument operating.
● Dust may accumulate on the surface of instrument after long time storage. Soft cloth or gauze can be used for
cleaning work, and a little detergent can be used if necessary. Please cut off the power supply before cleaning.
When instrument is not used, make sure shut the lid down.
● As to the use and storage method of the sample, reagent, Controls, Calibrator, please refer to the relevant
instructions.
● Sample, Controls, Calibrator and waster solution have the potential biochemical infectivity; the detergents are
corrosive that may hurt eyes, skin and mucosa. Operator should refer to the safety regulation for lab operation.
Protective measure should be taken to operator (Such as lab protective clothes and gloves).
● Avoid contact with eyes and skin, in case of skin contact, flush the area with water, rinse immediately with
plenty of water and seek medical advice.
● Operator should comply with the local regulation when draining and dealing with reagent, waste solution, waste
sample, consumable etc. Please dispose the waste solution and instrument consumable according to the regulation
of medical waste, infective waste and industrial waste.
I
User Manual of CS-400 Auto-Chemistry Analyzer
Warning:
● Instrument should be operated in a good ground condition, and an independent power supply is a must, the input
power should be conformed to instrument requirement.
● Don’t pull the electrical wire with wet hand, or there is a risk of electrical shock.
● Doesn’t stamp, twist, drag the wire and cable, or it may cause a fire.
● Please don’t open the back and side cover board before cutting the general power supply except Dirui special
service staff.
● If liquid occurs in instrument interior or there is an internal pipeline leakage, please immediately cut off the
general power supply, and contact Dirui customer service dept.
● Please don’t touch sample probe, reagent probe and stirring rod, etc. when instrument operating, don’t put your
hand into the opening part, or it may cause body injury or instrument damage.
● Cut off the power supply before replace light source lamp. Don’t touch the lamp before it is cool to avoid
burning.
● Periodic maintenance should be executed strictly according to the user manual. Or it may cause instrument
malfunction, and affect the accuracy and precision of test results.
● Make sure that the Auto-Chemistry Analyzer is operated according to the user manual, or the measuring result is
not a reliable one, and the damage on instrument may endanger human safety.
● Please don’t place combustible material around the instrument.
II
User Manual of CS-400 Auto-Chemistry Analyzer
Catalogue
Chapter 1 Brief Introduction ...........................................................................................................................1
1.1 Summary ......................................................................................................................................................1
1.2 Main technical index ...................................................................................................................................1
1.3 Composition of instrument .........................................................................................................................3
1.4 Configuration and function ........................................................................................................................8
1.5 Instrument Symbol ....................................................................................................................................21
Chapter 2 Function and Measuring Principle .............................................................................................22
2.1 Mechanism movement principle ..............................................................................................................22
2.2 Assay mode .................................................................................................................................................24
2.3 Check of measure value ............................................................................................................................57
2.4 ISE testing principle ..................................................................................................................................62
Chapter 3 Instrument Installation ................................................................................................................66
3.1 Installation requirement ...........................................................................................................................66
3.2 Open package .............................................................................................................................................67
3.3 Installation procedure ...............................................................................................................................68
Chapter 4 Accessory Device ...........................................................................................................................76
4.1 Sample disk barcode reader .....................................................................................................................76
4.2 Reagent barcode reader ............................................................................................................................77
4.3 Purified water equipment .........................................................................................................................78
Chapter 5 CS-400 Software Operation .........................................................................................................79
5.1 Software interface instruction ..................................................................................................................79
5.2 Software Operation ...................................................................................................................................83
5.3 Instrument standard specification ...........................................................................................................86
Chapter 6 Instrument Operation ..................................................................................................................88
6.1 Overview of operation ...............................................................................................................................88
6.2 Detailed operation .....................................................................................................................................89
Chapter 7 Calibration Information ............................................................................................................122
7.1 Colorimetric calibration..........................................................................................................................122
7.2 ISE calibration .........................................................................................................................................127
Chapter 8 Quality Control ...........................................................................................................................129
8.1 QC registration ........................................................................................................................................129
8.2 QC interval ...............................................................................................................................................133
8.3 Monthly quality control ..........................................................................................................................135
Chapter 9 System Setup ...............................................................................................................................137
9.1 Chemistry parameter ..............................................................................................................................137
9.2 Item combination .....................................................................................................................................145
9.3 Calculated item ........................................................................................................................................146
9.4 Cross contamination ................................................................................................................................147
9.5 Report sheet format .................................................................................................................................150
9.6 ISE Setup ..................................................................................................................................................153
9.7 Other setup ...............................................................................................................................................154
9.8 Manual item setup ...................................................................................................................................156
9.9 LIS communication setup .......................................................................................................................156
Chapter 10 System management .................................................................................................................158
10.1 User information ....................................................................................................................................158
10.2 Hospital information .............................................................................................................................159
10.3 Other information..................................................................................................................................160
10.4 Workload statistics ................................................................................................................................165
10.5 Database maintenance ...........................................................................................................................167
10.6 System log ...............................................................................................................................................168
Chapter 11 System Help ...............................................................................................................................169
11.1 System help application.........................................................................................................................169
Chapter 12 System Maintenance .................................................................................................................170
12.1 System maintenance preparation .........................................................................................................170
12.2 The Application of system maintenance menu....................................................................................171
12.3 Maintenance and checkup points and parts ........................................................................................180
III
User Manual of CS-400 Auto-Chemistry Analyzer
IV
User Manual of CS-400 Auto-Chemistry Analyzer
1.1 Summary
CS-400 Auto-Chemistry Analyzer is an instrument with discrete system, reagent open function, emergency
priority function as well as an external computer. The instrument is composed of humanized software operation
system, intelligent zed optical unit, complicated mechanism system, precision liquid path and accuracy electrical
system. The instrument could automatically realize sampling, reagent injection, anti-interference, mixture,
pre-temperature, reaction measurement, rinse, calculation, display and print function. The substitution of manual
operation for automatic operation could not only enhance the working efficient but also decrease the test error,
thus greatly enhance the accuracy and precision of test results.
CS-400 Auto-Chemistry Analyzer could carry out the immunology check and biochemical analyze of blood,
urine, ascites, cerebrospinal fluid and other body fluid. The instrument could also carry out clinic test, such as:
myocardium enzymogram, blood sugar, blood fat, liver function, renal function, immunoglobulin, etc.
Throughput: Constant speed, 400 tests/ hour (800 tests/ hour with ISE)
Liquid level sensor: Integration of sample probe and reagent probe with touch sensor and sample probe
block test function.
Sample disk: 115 samples position (50 routine samples, 34 Calibrators, 20 STAS samples, 8
Controls, 3 Detergents)
1
User Manual of CS-400 Auto-Chemistry Analyzer
Light source: 20W /12V Long life quartz halogen lamp (water cooling)
Reaction disk: 120 pcs of reusable rigid optical plastic reaction cuvette.
Calibration method: 1-point linearity, 2-point linearity, multi-point linearity, non-linearity method.
Barcode scanning: 3 internal barcode scanner ( scan the barcode on the routine sample, and reagent
on the R1, R2 disk )
Reagent volume test : Test and report the reagent remaining volume.
2
User Manual of CS-400 Auto-Chemistry Analyzer
① ②
③ ④
3
User Manual of CS-400 Auto-Chemistry Analyzer
① ② ③ ④ ⑤ ⑥ ⑦
① ISE pipetting syringe ② ISE diluting syringe ③ ISE inner standard solution syringe
4
User Manual of CS-400 Auto-Chemistry Analyzer
① ② ③ ④ ⑤ ⑥ ⑦ ⑧
① Power supply inlet ② RS-232 interface ③ Cooling fan ④ Diluted waste solution outlet
⑤ Purified water inlet ⑥ Vacuum tank waste solution outlet ⑦ Concentrated waste outlet
5
User Manual of CS-400 Auto-Chemistry Analyzer
① ② ③ ④ ⑤ ⑥
⑦ ⑧ ⑨ ⑩ ○
11 ○
12
⑦ Pilot lamp of sample disk rotation ⑧ cooling lip of inner track of sample disk ⑨ Sample disk
6
User Manual of CS-400 Auto-Chemistry Analyzer
① ② ③ ④
① Analytical unit switch (cooling power supply is excluded) ② Pilot lamp of cooling power supply (green)
③ General power supply switch ( breaker) ④ Pilot lamp of power supply ( red)
7
User Manual of CS-400 Auto-Chemistry Analyzer
CS-400 Auto-Chemistry Analyzer is composed by operating system and analytical system. The two parts is
connected by RS-232 serial wire.
Operating system is composed of host, 17 inch CRT display monitor, keyboard, mouse and printer.
CRT display monitor: Display all kinds of form, curve and test data of CS-400 software.
8
User Manual of CS-400 Auto-Chemistry Analyzer
! Warning:
△
z When instrument operating, be sure to close the cooling lid on the inner track of sample disk and screw the
knob.
z The sparkling of LED pilot lamp indicates that sample disk is turning or going to turn, do not change sample
or touch sample disk at this time, or it may cause body injury or instrument damage.
① ② ③ ④⑤ ⑥ ⑦⑧ ⑨
① Outer track ② Middle track ③ Inner track ④ Sample disk inner track pin
⑤ Handgrip of sample disk inner track ⑥ Sample disk guide pin ⑦ Locking block of inner/outer disk
Sample disk is composed of outer track, middle track, inner track and LED pilot lamp. The inner track has
cooling function, thus the Controls and Calibrator on inner track of disk can be restored in 5℃~15℃.
Place the containers (standard cup, micro cup, test tube) which contain Calibrator, sample, Controls on the
sample disk, and then the sample disk will send them to the sampling position in the sampling mechanism.
9
User Manual of CS-400 Auto-Chemistry Analyzer
(3)Movement
At Power on: Sample disk turns counterclockwise to move routine sample No.1 to the sampling position.
At analysis: At start of analysis, sample disk makes the same movement as “power on”. During analysis,
sample disk turns to the direction allowing a quicker access.
When mounting the sample disk, be sure to set the sample disk matching with the guide pin. Set the latch
on the inner track. Also, be sure to secure the cooling unit lid on the inner track, the outer track can be
demounted without removing the inner track.
Note:When mounting/demounting the routine sample disk (outer track), be sure to hold the hooks with both hands.
In order to change the sample in inner track, be sure that sample probe has stopped sampling or been under the
stand-by status and take out the disk cover first. Before operating, please check the disk position.
Single-click “ maintenance” key, select “ mechanism operation checkup”, input the check times,
single-click “ Execute ” button. Alarm is issued when abnormality exists.
! Warning:
△
● Please don’t touch the sampling mechanism when operating, or it may cause body injury or instrument
damage.
10
User Manual of CS-400 Auto-Chemistry Analyzer
① ② ③ ④
① Sample probe up and down mechanism ② Sample probe rotation arm
(1) Function
Assimilates a specified amount of sample from sample container and pipets it into reaction cuvette. The
sample probe possesses the liquid level detect function. Alarm is issued when touch occurs in descending
process, and alarm is issued also when sample probe is blocked.
(2)Specification
Sampling mechanism can assimilate 2.0-35.0 ul sample volume, set in 0.1ul stepping. Sample probe will
assimilate more than specified amount. Minimum sample volume requires more than 100 ul.
(3)Movement
At power on: The sample probe comes over above the reaction cup, and then returns above the sample
probe rinse bath.
At analysis: The sample probe circulates up and down motions as the following sequence:Sample
container, reaction cuvette, sample rinsing bath. Automatic sample probe rinsing is carried
out when sampling finished. At start of analysis, the sample probe makes the same movement
as at power on. In the rinsing bath, the inner and outer walls of the probe are rinsed. Sample
probe block test is carried out simultaneously.
Assimilate sample: sampling is taken when sample probe descent 1.7mm below liquid level.
Automatic rinsing of probe: After assimilating the sample, the sample probe will return above the rinsing
bath to wash the inner and outer walls of probe。When sampling is finished, the CS-alkaline detergent is
11
User Manual of CS-400 Auto-Chemistry Analyzer
(5)Operation check
Single-click the “Maintenance”key, select“mechanism operation checkup and input the check times. Click
“Execute “. If abnormality exists, instrument will issue alarm.
● Please don’t touch the lid of the reagent disk when running or it may cause body injury and instrument
damage.
① ② ③ ④ ⑤ ⑥ ⑦
① Reagent disk cover ② Locking knob of cover ③ Reagent bottle ( on the reagent rack)
④ Track pin of reagent disk ⑤ Handgrip of reagent disk. ⑥ Reagent disk guide pin
(1)Function
The reagent disk accommodates reagent bottles and carries the specific reagent, diluents and
CS-anti-bacterial phosphor-free detergent to the pipetting position in pipetting mechanism. Cooling system
can keep the reagent disk in a lower temperature to store reagent. There is a barcode reader on the wall of
reagent cooling unit, which can scan the barcode on reagent bottle.
(2)Specification
Reagent disk 1 (R1): R1, R4, diluents and detergent, total 45 bottles. Number 45 position is specially used
for CS-anti-bacterial phosphor-free detergent.
Reagent disk 2 (R2): R2, R3, diluents and detergent, total 45 bottles. Number 45 position is specially used
12
User Manual of CS-400 Auto-Chemistry Analyzer
Single-reagent usage: Single-reagent can be used as Reagent 1 or Reagent 2 that is putted into both two
reagent disk.
(3)Movement
At power on: Turning clockwise, Position 1 of R1 and R2 disks carry each bottle to the reagent pipetting
position.
At analysis: Initial operation is the same as at power on. Subsequently, the disks rotate in the direction
which allows a quicker access.
(4)Operation check
Single-click the “Maintenance”key, select“mechanism operation checkup and input the check times. Click
“Execute“. If abnormality exists, instrument will issue alarm.
(5)Mounting/ Dismounting
Fix the disk with 2 latches at the center of disk. To remove the disk, loose the latches. For mounting, be sure
to set the disk matching with the guide pin and fasten it with the latches. The reagent disk cover must be
attached except for replacement of the disk of reagent. During operation, the reagent probe may moves, so
avoid detaching the reagent disk cover at this time.
Note: If operator open the disk cover during stand-by or running state, alarm occurs. Reagent horizontal scanning
will be automatically carried out when cover the reagent disk lid at stand-by status.
! Warning:
△
● Be sure to close the lid of reagent disk when instrument operating, and fasten the knob.
13
User Manual of CS-400 Auto-Chemistry Analyzer
① ② ③ ④
(1)Function
Assimilate a specified amount of reagent from each reagent bottle and pipets it into a reaction cuvette. The
reagent probe acts as a sensor at the liquid surface level. The remaining amount of reagent in a bottle is
calculated from the descending distance of reagent probe and displayed on the “reagent info.’ menu.
(2)Specification
(3)Movement
At power on:Move toward the reaction cup side once and then returns above the probe rinsing bath.
At analysis: Reagent probe circulates the motion as the following sequence: reagent bottle--reaction
cuvette—reagent probe rinsing bath.
(4)Automatic rinsing
Assimilate CS-anti-bacterial phosphor-free detergent from the position 45 on the reagent disk for three
times and infuse it into reaction cuvette for three times and return to the probe rinsing bath to wash inner
and outer walls of the probe.
(5)Movement check
Single-click the “Maintenance”key, select“mechanism operation checkup and input the check times. Click
“Execute “. If abnormality exists, instrument will issue alarm.
14
User Manual of CS-400 Auto-Chemistry Analyzer
! Warning:
△
● Please don’t touch the reaction disks during operating, or it may cause mechanical failure.
① ② ③ ④ ⑤ ⑥
④ Fixing knob of reaction disk ⑤ Guide pin and hole ⑥ Reaction cup groupware handgrip
(1)Function
Fasten reaction cuvette by screw, and be sure to keep chemical reaction in constant temperature of 37℃.
Each reaction cuvette also serves as a cell for absorbance measurement.
(2)Specification
(3)Operation
At power on:Rotates and stops at the start position. At this time, the first reaction cuvette is located under
the first rinse nozzle.
At analysis: Initially operates the same as at power on, then circulates a cycle of half turn and one pitch
advance (covering 61 reaction cuvettes) followed by temporary stop. Each full turn takes about 18 seconds.
15
User Manual of CS-400 Auto-Chemistry Analyzer
(4)Rinsing
At position No. 45 of both reagent disks, set a bottle containing CS-anti-bacterial phosphor-free detergent
and open its cap. Carry out “ rinsing reaction cuvette” in the “maintenance” window, all the reaction
cuvettes will be rinsed. It is usually rinsed by CS-alkaline detergent in the front-left of instrument, so daily
maintenance for reaction cuvettes is unnecessary.
(5)Operation check
Single-click the “ Maintenance” key, select “ mechanism movement checkup”, and input the check times.
Click “ Execute“. If abnormal exist, instrument will issue alarm.
(6)Demounting
Reaction disk: Remove the rinse mechanism from above the reaction disk and completely loosen the fixing
knob at the center. The reaction disk can be lifted out. For mounting, align the guide pin of instrument with
the guide hole of the reaction disk and tighten the fixing screw.
Reaction cuvette: Remove the cuvette setscrew and pull up the reaction cuvette block by holding the
handgrip, and the reaction cuvette can be removed from the reaction disk.
Note:Keep reaction cuvette immersed in purified water. Besides, if the instrument will be not used more than 3
days, reaction cuvette should be removed and immersed in purified water.
! Warning:
△
● Keep the cleanness of purified water in incubation bath, or it may effect the test precision.
● When instrument startup or rinsing incubation bath, make sure there is enough CS-anti-bacterial
phosphor-free detergent at No.45 position.
(1)Function
(2)Operation
At power on: Automatic exchanges the constant temperature water once, the CS-anti-bacterial phosphor-free
detergent in position No.45 of both reagent disks is added in incubation bath.
At analysis: Incubation bath water is circulating. Instrument may automatically supply water when water
shortage comes in operation process.
Exchange water: In “maintenance” window, select “rinsing incubation bath” , and then the constant
temperature water may exchange, and then add 6ml CS anti-bacterial phosphor-free detergent in incubation
bath water.
Note: After running for 24 hours, instrument may require “incubation bath water exchange”, please carry out
“ Rinsing incubation bath”.
16
User Manual of CS-400 Auto-Chemistry Analyzer
! Warning:
△
● Please don’t touch stirring mechanism When operate, or it may cause body injury or instrument damage.
(1) Function
(2) Operation
At power on: Move to the side of reaction cuvette and then stops above the rinsing bath, move to the side
of reaction cuvette again, and then stops above the rinsing bath.
At analysis: The mechanism descends, rotates, rises and stops between two locations: reaction cuvette and
stirring rod rinsing bath.
Stirring is carried out after each addition of reagent 1(R1),reagent 2(R2), reagent3(R3), reagent 4(R4). R1
and R4 use Stirring rod mechanism 1. R2 and R3 use Stirring rod mechanism 2.
Automatic rinsing of stirring rod: when stirring rod descends into stirring rod rinsing bath, mechanism may
automatically rotate and washes the stirring rod with purified water.
Sampling finishing: Stirring rod is stirring in reaction cuvette in which detergent is added, thus rinse the
stirring rod.
Single-click the “maintenance” key, select “ mechanism operation check”, and input the check times. Click
“ Execute”. If abnormality exists, instrument will issue alarm.
! Warning:
△
● Please don’t touch the rinsing mechanism when operate, or it may cause body injury or instrument damage
● Avoid directly contact with body, or it may cause infection. Please adopt protective measure. In case of
skin contact, flush the area with water, rinse immediately with plenty of water and seek medical advice.
(1)Function
Eliminates the reaction solution, rinse the reaction cuvette, injects and eliminates purified water which
used for test cell blank
17
User Manual of CS-400 Auto-Chemistry Analyzer
F G F
C D A B A B B
E
Rotational
Rotational direction of 测4 4times
次杯空白(1
cell blank次停止,3 次通过)
measurement.
Reaction disk 1 stop, 3 pass.
Above figure shows that seven steps are needed when rinsing reaction cuvette. (Four times cell blank test is
added) , therefore, to finish rinsing one reaction cuvette, 11 steps are needed.
Step 1: Nozzle 1D assimilates reaction solution, then 1C discharges detergent to reaction cuvette.
Step 2: Nozzle 2B assimilate detergent from reaction cuvette, then 2A discharges purified water to reaction
cuvette.
Step 3: Nozzle 3B assimilates the purified water from reaction cuvette, then 3A discharges the purified
water to reaction cuvette.
Step 4: Nozzle 4B assimilates the purified water from reaction cuvette, and wipes the reaction cuvette at the
same time.
Step 6,7,8,and 9: Carry out the cell blank check measurement at the fifth nozzle. The reaction cuvette with
full purified water allows 4 times cell blank measurement. (1 time static measurement, 3 times
dynamic measurement when reaction cuvette passed by during reaction disk turning )
Step 10: Nozzle 6F assimilates the purified water, which has carried out the cell blank absorbency check.
Step 11: Nozzle 7 F assimilates the remaining water in reaction cuvette, and wipes the reaction cuvette at
the same time; Nozzle 7G only discharges the purified water in rinsing process before sampling,
rinse the nozzle tip.
(3)Operation
At analysis: Rinse the reaction cuvette and carry out the water blank test in the rotational direction of
arrangement of rinse nozzle in Figure.1-13.
(4)Operation check
Single-click the “maintenance” key, select “ mechanism operation check”, and input the check times. Click
“ Execute ”. If abnormality exist, instrument will issue alarm.
(5)Dismounting
The rinse mechanism can be shifted from the reaction disk by loosening the fixing screw at the head.
Cooling system is composed of reagent cooling system and sample cooling system, which could cool the
reagent, Controls and Calibrator respectively.
(2)Specifications
Temperature: 5 -- 15℃
! Warning:
△
● Even the analyzing system is power off, cooling system is still at working status. The cooling system
only stop working when main power supply is cut off.
● The usage and storage of Reagent should be performed strictly according to user manual.
Concave
Detector (340-750nm) diffraction
12 fixed wavelength grating
Light source
Reaction
lamp
cuvette
19
User Manual of CS-400 Auto-Chemistry Analyzer
(1) Function
When the reaction disk rotates, the absorbance of purified water or reaction solution is measured in each
reaction cuvette. As above figure shows.
(2)Specification:
20
User Manual of CS-400 Auto-Chemistry Analyzer
Symbol Meaning
AC symbol
Storage at
Batch code
Use by
Serial number
Measurement Control
Manufacture by
Grounding terminal
21
User Manual of CS-400 Auto-Chemistry Analyzer
CS-400 Auto-Chemistry Analyzer consists primarily of the sample disk, sampling mechanism, reagent
disk, reagent pipeting mechanism, reaction disk, reaction bath, rinsing mechanism and photometer.
Operation of each mechanism is explained according to figure 2-1:
After starting, instrument carries out resetting first, rotates the R1 reagent disk and R2 reagent disk to
Position 1, then rotate them for 360 degrees, the R1 reagent probe, R2 reagent probe, sample probe, R1
stirring rod, R2 stirring rod are all stopped at the upper side of their rinsing bathes respectively.
Upon pressing the start key, the rinse mechanism starts rinsing from reaction cuvette No1. The reaction
disk rotates by 22 patches and stops temporarily, and then rotates by 37 patches and stops, sequently, 2 more
patches and then stops. This sequence is carried out again to cover one full turn plus 2 patches (18 seconds).
Cell blank is measured when the reaction cuvette passes through the photometric unit during rotation of the
reaction disk. The measured value of cell blank becomes the reference value for the subsequent absorbance
measurement. The liquid in the reaction cuvette is assimilated through the nozzle of rinse mechanism.
After rinsing the reaction cuvette for 3 minutes (the reaction disk rotates 10 circles), the sampling
mechanism begins to work when reaction disk rotates the 11th turns, and the sample probe moves above of the
sample cup and decends into the cup. Since the sample probe comprises a liquid level sensor, the probe stops
descending when its tip enters the sample. A set volume of sample is assimilated with the sample pipettor.
Next, the sample probe moves to the top of No.1 reaction cup and descends until its tip reaches the bottom of
the cuvette, where the sample is discharged. The sample probe further moves to the inside of probe rinsing
bath, where its inner and outer walls are washed with purified water.
On the other hand, the reagent pipetting mechanism assimilates reagent 1 (with R1 probe), while the
reaction disk rotates by 22 patches, stops temporarily and the rotates by 37 patches after the fore mentioned
sampling mechanism has discharged the sample into the reaction cuvette. When the reaction disk stops
temporarily after the above rotation, reagent 1 is discharged into the reaction cuvette. And when the reaction
disk stops temporarily after rotating by 2 patches, the R1 stirring mechanism mixes the sample and reagent.
The reagent pipetting mechanism rotates to the top of reagent bottle in the specified channel and descends
while assimilates and discharges reagent. A set volume of reagent is assimilated, then the reagent probe moves
to the top of reaction cuvette and discharges the reagent, followed by returning to the rinsing bath and
washing of its inner and outer walls. Add reagent R1 and start photometry. Measurement is made when each
reaction cuvette passes through the optical path during rotation of the reaction disk. The reaction disk
rotates15 turns plus 22 patches to the reagent 2 pipetting position, where reagent 2 is added with R2 probe.
Reagent 2 is stirred with the R2 stirring mechanism after the reaction disk rotates by 16 turn plus 2 patches
and stops temporarily, and then rotates 22 patches and stops temporarily.
After 26 turns plus 37 patches , reaction disk comes to the reagent 3 sampling position, and reagent
probe 2 assimilates reagent 3. R2 stirring rod begins to work after 2 more patches of reaction cuvette.
After 41 turns plus 37 patches , reaction disk comes to the reagent 4 sampling position, and reagent probe
1 assimilates reagent 4. R1 stirring rod begins to work after 2 more patches of reaction cuvette
After the lapse of about 15 minutes (60 circles), the first cup measurement is over, and the reaction
solution in No. 1 reaction cup is assimilated with the nozzle of cuvette rinsing mechanism, which will
discharge cleaning liquid and water into the reaction cup to rinse the cup. The instrument stops and goes to
22
User Manual of CS-400 Auto-Chemistry Analyzer
standby status when the last reaction cuvette has been washed, and goes to the the cell blank status. The
reagent probe and sample probe will be rinsed respectively by their own detergents added by themselves after
every test.
Reagent combination (R1, R2, R3, R4) of each item and reaction time (3~15 mins) is set in the “Analytical
parameter” form in “System setup”
光电检测位
63,
Photoelectric
detection
62,R1 stirring
Sop and wipe 62,R1 stirring
rod
position
搅拌探针
63,R4 stirring
position
Sam. probe
Sample probe
60 R1 Pipetting
R1, R4 Probe
61 R4 Pipetting
Reference position
number
23
User Manual of CS-400 Auto-Chemistry Analyzer
This instrument adopts the whole reaction monitoring system, which intermittently measures the absorbance of
reaction solution for a reaction time of 15 minutes.
The reaction disk rotates 1 turn plus 2 pitches in about 18 seconds and during this time the absorbance is
measured for all of 120 reaction cuvettes which go across the optical axis of the photometer. For each reaction
cuvette, measurement is made10 times in a reaction time of about 3 minutes. 13 times measurement is made
during 4 minutes (13 photo metric point.). 16 times measurement is made during 5 minutes (16 photo metric
point33 times measurement is made during 10 minutes (33 photo metric point.) .49 times measurement is made
during 15 minutes (49 photo metric point.)
CS-400 multi-wavelength photometer condenses the white light emitted from the light source lamp through the
lens, passes the condensed light through the reaction cuvette and separates the light with the concave diffraction
grating. The separated respective wavelength components are simultaneously received on the 12 fixed detectors
and amplified by 12 amplifiers, then logarithmically converted to obtain the absorbance or difference in
absorbance. In 2 wavelengths photometry, concentration is measured by the value of the difference of dominant
wavelength and complementary wavelength. This means that the photometer features a correcting effect for
lipemia, hemolysis and icterus of sample and has a compensating effect for fluctuation in source voltage, thus
realizing stable measurement.
The assay mode of Auto-Chemistry Analyzer is based on the Beer-Lambert law that the material selective
absorption light
The main principle is: When monochromatic light with specific wavelength passes through the cuvette with
sample, the monochromatic light absorbency and sample liquid concentration are varies directly as the distance
24
User Manual of CS-400 Auto-Chemistry Analyzer
I
A = lg(1/T)= lg( 0 )= ε b c
It
I0 - Incident intensity
It - Transmitted intensity
Solution layer thickness (b): Optical path, which is fixed by instrument. Molar absorption coefficient (ε) is the
correlation coefficient of the wavelength, solution and solution temperature. Linear relationship is displayed
between solution thickness and absorbency when in stable temperature and single wavelength(ε value is given
on the reagent bottle by factory)
If the sample liquid adequate distribution, interaction between liquid and incidence monochromatic light only
happens during absorbing process. No fluorescence, disperse and photochemical appear. No interaction between
substances in the solution while absorbing process. The absorbency possess conducts nature, and this condition
conforms to the Beer-Lambert law
25
User Manual of CS-400 Auto-Chemistry Analyzer
1-point L– 0– 0– 0 B1 + B 2 + B 3 AL + AL −1
Assay 3
1<L≤49 2
t :time
(minute)
2-point L– M– 0- 0 B1 + B 2 + B 3 ( AM + AM −1 ) − k ( AL + AL −1 )
between
Rate Assay 3
1<L<M≤49 2 photometry
point L、m
AM + AM −1 AL + AL −1
2-point L– M– 0– 0 B1 + B 2 + B 3 −
assay 2 2
1<L<M≤49 3
t
L– M – 0 - 0
B1 + B 2 + B 3
First half 1<L<M≤49 △A(M-L)
1-point & 3
L +2<M
Rate Assay
Second L– 0 – 0 – 0 B1 + B 2 + B 3 AL + AL −1
half 1<M<N≤L<P<Q≤49 3 2
M–N–P–Q
Rate A B1 + B 2 + B 3
1<M<N≤L<P<Q≤49 △(AQ- P)-k△(AN -M)
Assay 3
M+2<N,P+2<Q
Rate A
Assay with
Serum L– 0 – 0 – 0 B1 + B 2 + B 3 AL + AL −1
Index 3
Measurem
1<L≤M<N≤49 2
ent.
M–N–0–0 B1 + B 2 + B 3 ( AN + AN −1 ) − k ( AM + AM −1 )
First half
1<L≤M<N≤49 3 2
Rate B
Assay P-N ( different wavelength from the
L– M – 0 - 0
(mode 1 ) Second B1 + B 2 + B 3 Half) Two
3≤L<M<N<P≤49
half 3 △ A(P-N)–k △ A(M-L) ( same conditions
L +2<M
wavelength as the second half)
N–P–0–0
B1 + B 2 + B 3
First half 3≤L<M<N<P≤49 △A(M-L)
3
Rate B N+2<P
Assay
(mode 2 ) L– M – 0 – 0
Second B1 + B 2 + B 3
3≤L<M<N<P<Q<R≤49 △A(R-Q)– k△A(P-N)
half 3
L +2<M
Explanation of symbols:
26
User Manual of CS-400 Auto-Chemistry Analyzer
a
S + ∑ Rj
j =1
k= b
S + ∑ Ri
i =1
S : Sample volume
Note 1: The 5 th Photometric point won’t be Stirred after adding reagent 2. Stirring when the reaction disk
pauses after rotates one circle plus 2 pitches
Note 2: liquid in the reaction cuvette should be more than or equal to 150 ul, less than or equal to 450ul.
Cell blanks
(B1+B2+B3)/3 R2~R4
R1
A L + A L −1
S 2
SB B1 B2 B3
Time
Figure 2-2 1-point Assay
27
User Manual of CS-400 Auto-Chemistry Analyzer
AL + AL − 1
AX =
2
C X = {K × ( AX − B ) + C1 }× IFA + IFB
K: Factor
B: Absorbance of blank
Cell blanks
(B1+B2+B3)/3 R2~R4
A m + A m −1
R1
2
S A L + A L −1
2
SB B1 B2 B3
Time
Figure 2-3 2-point Assay
(a) Photometric point : 【L】-【M】-【0】-【0】 (1≤ L ≤ 49)
28
User Manual of CS-400 Auto-Chemistry Analyzer
The difference between the average of absorbance at measurement point m and m-1 and that at
measurement point l and l-1 is used.
AX=
( Am + Am − 1) − k ( AL + AL − 1)
2
a
S + ∑ Rj
j −1
k= b
S + ∑ Ri
i −1
C X = {K × ( AX − B ) + C1 }× IFA + IFB
K: Factor
B: Absorbance of blank
CRE, etc.
R2~R4
Cell blanks
User Manual of CS-400 Auto-Chemistry Analyzer
The difference between the average of absorbance at measurement points M and M-1 and that at
measurement points L and L-1, then divide the result by time.
( Am + Am − 1) ( AL + AL − 1)
−
AX= 2 2
t
t: Time (minute) between measurement points L and M
C X = {K × ( AX − B ) + C1 }× IFA + IFB
Ax: Average change in absorbance per minute between measurement points L and M
K: Factor
B: Absorbance of blank
30
User Manual of CS-400 Auto-Chemistry Analyzer
S
Absorbance
R1
R2~R4
Cell blanks
(B1+B2+B3)/3
AL Am
SB B1 B2 B3
Time
The change in absorbance per minute between measurement point L and M is obtained by the least
squares method
AX=△A(L-m)
C X = {K × ( AX − B ) + C1 }× IFA + IFB
31
User Manual of CS-400 Auto-Chemistry Analyzer
K: Factor
B: Absorbance of blank
1<m<n≤L<P<Q≤49
M+2<N,P+2<Q
For test A, the average of absorbance at measurement points L and L-1 is used, and used for the test B is
the difference obtained by subtracting the change in absorbance per minute between measurement points
32
User Manual of CS-400 Auto-Chemistry Analyzer
AL + AL − 1
AA=
2
AB=△Ap×q-k△Am×n
a
S + ∑ Rj
j −1
K= b
S + ∑ Ri
i −1
For each of tests A and B, calculation is made according to the following formula.
C X = {K × ( AX − B ) + C1 }× IFA + IFB
S: Sample volume
K: Factor
In the rate assay, the change in absorbance per minute is obtained by the least squares method
Absorbance
Cell blank
Time
1< L<M<N≤49
The average value of absorbance at photometric points L and L-1 is for test A, and test B, the difference
between the average absorbance at photometric points N and N-1 and that of photometric points Mind
M-1.
AL + AL − 1
AA =
2
( AN + AN −1 ) − k × ( AM + AM −1 )
AB =
2
a
S + ∑ Rj
j =1
k= b
S + ∑ Ri
i =1
C X = {K × ( AX − B ) + C1 }× IFA + IFB
S: Sample volume
K: Factor
35
User Manual of CS-400 Auto-Chemistry Analyzer
3≦L <m<n<p≦49。L+2<m、n+2<p
Used for test A is the change in absorbance per minute between measurement point l and m, which is
obtained by least squares method. Used for test B is the change in absorbance per minute between
measurement points n and p, which is obtained by the same method, but do not carry out blank
correction.
△AA=△A(M-L)
△AB=△A(P-N)
For each of tests A and B, calculation is made according to the following formula.
C X = {K × ( AX − B ) + C1 }× IFA + IFB
S: Sample volume
K: Factor
36
User Manual of CS-400 Auto-Chemistry Analyzer
3≦L<m<n<p≦49。L+2<m、n+2<p
Used for test A is the change in absorbance per minute between measurement points l and m, which is
obtained by least squares method. Used for test B is the difference obtained by subtracting the above
change for test A from the change in absorbance per minute between measurement points n and p, which
is obtained by the same method
△ AA =△A(L- M)
△ AB =△A(P- N)-k×△A(L- M)
a
S + ∑ Rj
j =1
k= b
S + ∑ Ri
i =1
For each of tests A and B, calculation is made according to the following formula.
C X = {K × ( AX − B ) + C1 }× IFA + IFB
S: Sample volume
K: Factor
3≦L<m<n<p≦49。L+2<m、n+2<R
Used for test A is the change in absorbance per minute between measurement points l and m, which is
obtained by the least squares method. Used for test B is the difference obtained by subtracting the change
in absorbance per minute between measurement points n and p from that between measurement points q
and r, which is obtained by the same method
△ AA =△A(L- M)
△ AB =△A(P- Q)-k×△A(P- N)
38
User Manual of CS-400 Auto-Chemistry Analyzer
a
S + ∑ Rj
j =1
k= b
S + ∑ Ri
i =1
For each of tests A and B, calculation is made according to the following formula.
C X = {K × ( AX − B ) + C1 }× IFA + IFB
S: Sample volume
K: Factor
39
User Manual of CS-400 Auto-Chemistry Analyzer
Absorbance
Concentration
K: Input value.
C X = {K × ( AX − B ) + C1 }× IFA + IFB
1-point assay, 2-point rate assay, 2-point assay, 3-point assay, 1-point&rate assay, rate A assay and rate B
assay
Blank (or standard 1) and standard sample (standard 2) are measured to prepare a linear working curve Figure
2-11 explains the linear method.
Absorbance
Concentration
Figure 2-11 2-point linearity
K: Calculated from measured values and input values of blank (standard 1) and standard sample (standard
2)
C 2 − C1
K=
A2 − B
C X = {K × ( AX − B ) + C1 }× IFA + IFB
1-point assay, 2-point rate assay, 2-point assay, 3-point assay, 1-point&rate assay, rate A assay and rate B
assay
Absorbance
Concentration
S1ABS (B):Linear primary regression intercept for absorbance or change in absorbance per minute of
blank (standard)
K: Inverse number of working curve slope in the result of linear primary regression.
X × Cr
S1ABS ( B) = A −
Y
Y
K = × 10 4 × 10 a
X
42
User Manual of CS-400 Auto-Chemistry Analyzer
( )( )
n
X : ∑ Cri − Cr × Ai − A
i =1
( )
n
Y : ∑ Cri − Cr
2
i =1
⎛ n ⎞
A : ⎜ ∑ Ai ⎟ / n
⎝ i =1 ⎠
⎛ n ⎞
Cr : ⎜ ∑ Cri ⎟ / n
⎝ i =1 ⎠
(c)Calculation of concentration
C X = {K × ( AX − B ) + C1 }× IFA + IFB
1-point assay, 2-point rate assay, 2-point assay, 3-point assay, 1-point&rate assay, rate A assay and rate B
assay
43
User Manual of CS-400 Auto-Chemistry Analyzer
B: the absorbance or approximate value measure of the absorbance change per minute when CX
approaches ∞.
K: blank (standard 1) absorbance or value calculate by the approximation formula of the absorbance
change per minute subtraction B
C X = (C + C1 ) × IFA + IFB
K
AX = B +
1 + aC )
1 ⎧ K − ( AX − B ) ⎫
C= ×⎨ ⎬
a ⎩ AX − B ⎭
44
User Manual of CS-400 Auto-Chemistry Analyzer
∑∑ (A )
N 2
, 2
IJ − A1
i =1 j =1
SD =
2N − 3
(N=3~6、j=1or 2)
(Aij-Ai’): Difference between approximate absorbance Ai’ and measured value Aij or A12. Each standard
sample is measured in duplicate so the number as measurement points Aij is 12 at maximum
Concentration
B: approximation for the absorbance or its change per minute when CX approaches ∞.
K: blank (standard 1) absorbance or value calculate by the approximation of the absorbance change per
minute subtraction B
45
User Manual of CS-400 Auto-Chemistry Analyzer
C X = (C + C1 ) × IFA + IFB
K
AX = B +
1 + aC b )
1 ⎧ K − ( AX − B ) ⎫
C = b× ×⎨ ⎬
a ⎩ AX − B ⎭
∑∑ (A )
N 2
, 2
IJ − A1
i =1 j =1
SD =
2N − 4
(N=4~6、j=1or 2)
(Aij-Ai’): Difference between approximate absorbance Ai’ and measured value Aij or A12. Each
standard sample is measured in duplicate so the number as measurement points Aij is 12 at maximum
46
User Manual of CS-400 Auto-Chemistry Analyzer
Absorbance
Concentration
B: approximation for the absorbance or its change per minute when CX approaches ∞.
AX − B
a+b·lnC+c·C-ln{ }=0
K − (A − BX )
C X = (C + C1 ) × IFA + IFB
K
1 + exp× (− a − b × ln C − c × C )
AX=B+
47
User Manual of CS-400 Auto-Chemistry Analyzer
∑∑ (A )
N 2
, 2
IJ − A1
i =1 j =1
SD =
2N − 4
(N=5~6、j=1 or 2)
(Aij-Ai’): Difference between approximate absorbance Ai’ and measured value Aij or A12. Each
standard sample is measured in duplicate so the number as measurement points Aij is 12 at maximum
Concentration
B: approximation formula for the absorbance or its change per minute of blank (standard 1)
C X = (C + C1 ) × IFA + IFB
∑∑ (A )
N 2
, 2
IJ − A1
i =1 j =1
SD =
2N − 5
(N=5~6、j=1 or 2)
(Aij-Ai’): Difference between approximate absorbance Ai’ and measured value Aij or A12. Each
standard sample is measured in duplicate so the number as measurement points Aij is 12 at maximum
49
User Manual of CS-400 Auto-Chemistry Analyzer
Absorbance
( ) + d (I) × (C
3
AX=a(I)+b(I) × (C X − C (1) + c(I)) × C X − C (1) - C(1) )
2
X
f × (C X − C ( I )) = a × ( I ) + b × ( I ) × (C X − C ( I )) + d × ( I ) × (C X − C ( I )) 2 + d ( I ) × (C X − C ( I )) 3 − AX
Calculate C according to the Newton approximation formula.
C X = (C + C1 ) × IFA + IFB
AX, A2~AN: Absorbance of sample and standard or its change per minute.
50
User Manual of CS-400 Auto-Chemistry Analyzer
∑∑ (A )
N 2
, 2
IJ − A1
i =1 j =1
SD =
2N − 4
(N=5~6、j=1 or 2)
(Aij-Ai’): Difference between approximate absorbance Ai’ and measured value Aij or A12. Each
standard sample is measured in duplicate so the number as measurement points Aij is 12 at maximum
Absorbance
Concentration
S1ABS is the two measure value (absorbance or it’s change) of STD (1)
C 2 − C1
K=
A2 − B
51
User Manual of CS-400 Auto-Chemistry Analyzer
C X = {K N × ( AX − AN ) + C N }× IFA + IFB
absorbance
A2 STD(2)
A3 A3’
STD(3)
STD(3)
AF AM’ Sample
Sample
A4 A4’
STD(4) STD(4)
B STD(1) B’ STD(1)
Isozymes method uses 2 reagent position. The total activity Cf is supposed to be calculated first with a
certain reagent in the isozyme P position. Calculate the activity Cm or Cn of isozymes M or N with a
reagent which can suppress N or M substance.
Generally speaking isozymes N inhibitor cannot suppress the activity of isozymes N completely, and the
activity of isozymes M is suppressed in a degree at the meantime.
The isozymes Method uses 2 channels to test the total activity and standard cost of isozymes M,N, K
value is tested by total activity channel, both two channels are used.
52
User Manual of CS-400 Auto-Chemistry Analyzer
M and N activity residual rate of M and N is calculated by inhibitor (ratio between absorbance of
standard 3 and standard 4 in two channel.).Calculate the total activity and isozymes M activity upon
above method. The activity of isozymes N is observed by calculation.
Reagent: Reagent for measurement of total activity, reagent for measurement of isozyme activity.
Standard sample: Standard sample F (containing both isozymes M and N), standard sample M
(containing isozyme M), standard sample N (containing isozyme N)
Make entry for each of the isozyme P and Q channels as shown in below Table:
Table 2-3
S1 to S4 are calibrator code numbers of calibrator 1 to calibrator 4 respectively. Enter the same calibrator
code number in both channels for each of calibrator 1,3,4. Set standard sample of isozyme M at position
Calibrator 3 and that of isozyme N at position Calibrator (4). It is unnecessary to enter the concentrations
of calibrator 3,4,2 for the isozyme Q channel. The channel number M of isozyme M is specified in the
isozyme P channel. But it need not be specified on the isozyme Q side. Unless M is entered, Isozyme
parameter error alarm occurs to disable analysis
(c) Working curve for total activity measurement channel (isozyme P) K factor is calculated by the
following formula:
C 2 − C1
K=
A2 − B
53
User Manual of CS-400 Auto-Chemistry Analyzer
CF={K·(AF-B)-C1}·IFA-IFB
C3={K·(A3-B)-C1}·IFA-IFB
C4={K·(A4-B)-C1}·IFA-IFB
CF 、C3 、C4: is sample, activity of standard 3, standard4; Ap 、A3 、A4: is absorbance or change of the
absorbance per minute of standard 3 , standard 4;IFA,IFB Instrument constants, representing slope and
intercept.
CM’={K·(AM’-B’)-C1}·IFA-IFB
C3’’={K·(A3’- B’)-C1}·IFA-IFB
C4’={K·(A4’- B’)-C1}·IFA-IFB
CM’: Isozyme M activity of sample;AM’: Isozyme absorbance of sample or its change per minute;C3’ 、
C4’:each inhibited activity of standard 3 and 4;A3’、A4’分 Each inhibited absorbance of standard 3 and 4
or its change per minute;B: Absorbance of blank or its change per minute IFA,IFB Instrument constants,
representing slope and intercept.
A3, -B,
α=
A3-B
A4,
-B,
β=
A4-B
CM’=α×CM+β×CN
54
User Manual of CS-400 Auto-Chemistry Analyzer
CM − α × CM
CM= CF-CN= CF -
β
β×CM=β×CF -CM’+α×CM
(α-β)×CM= CM-β×CF
C M '− β × CF
CM=
(α − β )
(h) Applicable assay
Note: Suggest operator use 6 calibrators in Logit-log5P, exponential, spline non-linear calibration.
A11、 A12:1st and 2nd absorbance values of standard (1) measured presently.
55
User Manual of CS-400 Auto-Chemistry Analyzer
a(I):I=1~N,N representing the number of standard solution and factor of the curve(refer to 5.3.3(8)
Logit-Log 5P method)
This calibration method needs be specified for each test on the calibration test selection screen of Routine
job Menu.
Note*:Only blank calibration can be carried out in case 1 is entered for the number of standard solutions (K
factor method)
Linear (2-point), Linear (multi-point), Linear (1-point: K factor), Isozyme P, Isozyme Q, Logit-log 3P,
Logit-log4P, Logit-log 5P, Exponential, Spline
56
User Manual of CS-400 Auto-Chemistry Analyzer
A11、 A12:1st and 2nd absorbance values of standard (1) measured presently.
Various checks are performed to enhance the reliability of measured results. Below are check description
In calibration, a warning –level alarm is issued if the measured absorbance of blank is not within the input
range of standard 1 absorbance. In this case, the result of measurement and alarm (S1ABS) are printed out. To
avoid check, enter-“-3.3~3.3”.
A warning level alarm is issued if the difference of the two times measured absorbance value is larger than the
set value. In calibration, each Calibrator (include reagent blank: Calibrator 1) is tested twice.
57
User Manual of CS-400 Auto-Chemistry Analyzer
If the difference of standard absorbance (average between 2 times measure) from Max. concentration standard
absorbance (sensitivity) exceeds the permissible absorbance sensitivity value (Sensitivity Limit), a warning-
level alarm is issued. In this case, alarm mark is printed out together with the result of measurement.
The working curve of the alarmed analytical item will be renewed, and the K value won’t be renewed. To
avoid the check, enter-“0”.
ASTD(N) − ASTD(1)
lower lim it ≤ ≤ upper lim it
C STD(N) − C STD(1)
If the fluctuation in factor K value between previous calibration and current calibration is 20% ,a warning
level alarm is issued. The working curve and K factor will be renewed and testing can be carried out. Make
sure check the reason of alarm.
K this − K last
× 100% ≤20%
(K this + K last ) / 2
(5) Drift rate check
In calibration, a warning –level alarm is issued if the difference between the calculated absorbance and tested
absorbance has exceeded the drift rate set value. The working curve and K factor will be renewed and testing
can be carried out. Make sure to check the reason of alarm. To avoid the check, enter“3.3”.
When 4 or more than 4 tested absorbance value is not accord with the set value of reaction limit absorbance,
alarm is issued as figure 2-21 shows:
58
User Manual of CS-400 Auto-Chemistry Analyzer
ABS
Time
Figure 2-21
Select “Linearity check” in “Alarm Info.”, and input the limit check value in corresponding textbox as figure
2-22 shows:
Figure 2-22
If not selected, even if input value in textbox, linearity check is not carried out.
ΔAf −ΔAb
ΔA ×100>Limit linearity value(%)
59
User Manual of CS-400 Auto-Chemistry Analyzer
Photometry point
Time
ΔAf ,−ΔAb,
ΔA, ×100>Limit linearity value(%)
Photometry point
Time
This instrument can check whether concentration is in the absorbance decreasing range (post zone). For prozone
check, the following 2 methods are available: antigen readdition method in 1-point assay with prozone check
and reaction rate method in 2-point assay. To Avoid check, input “-3.3 lower limit” in “checkup value” function
box in “analyze parameter” menu.
60
User Manual of CS-400 Auto-Chemistry Analyzer
Input method:
Analytical method: 【】
1≤Q1≤Q2≤49
Aq 2 + A( q 2 − 1) Aq1 + A( q1 − 1)
PC = −k×
2 2
K=total liquid volume when test q1/total liquid volume when test q2
Time
Figure 2-25 Antigen supplement method
Upper/lower limit: 【】
Analyze method: 【】
5≤q1≤q2≤49、5≤q3≤q4≤49
61
User Manual of CS-400 Auto-Chemistry Analyzer
( Aq 4 − Aq 3) /( q 4 − q 3)
PC = × 100
( Aq1 − Aq 2) /( q1 − q 2)
Time
Figure 2-26 Reaction speed ratio method
RT
E = E 0 + 2.303 × × l og(ai ) …………………………………………………………(1)
nF
ai = f × Ci ………………………………………………………………………………. (2)
T : Absolute temperature(t℃+273.15)(K)
ai :Ion(i)activity
f :Activity coefficient
Ci:Concentration
E ( H ) − E ( L)
SL = ……………………………(3)
C(H )
log
C ( L)
SL:Slope
E ( IS ) − E ( L )
C ( IS ) = C ( L) × 10 SL
………………………………(4)
E ( S ) − E ( IS )
C ( S ) = C ( IS ) × 10 SL
……………………………………….(5)
C(S):Sample concentration
C (VALUE ) = C (C ) − C ( X ) ……………………………..(6)
Item Specification
64
User Manual of CS-400 Auto-Chemistry Analyzer
Note: During operation, if ISE analysis is not taken more than 10 minutes, the internal standard liquid will be
pipetted once in order to activate electrode.
65
User Manual of CS-400 Auto-Chemistry Analyzer
Before installation, operator should check the space, power and environment requirement.
● Space between left (right) side of analyzer and the wall should ≥50cm
● Space between rear panel of analyzer and the wall should ≥50cm
● Make sure there is enough space for waste device and purified water equipment.
● Environment should with no dust, no mechanical vibration, no noise source and power interference
● Do not put the analyzer in the vicinity of brush motor, flicker fluorescent tube and other constant on-off
electrical equipment.
● Avoid direct sunlight, do not put the analyzer in front of heat source and wind source
● Power: 2000 VA
●A well grounded power supply socket is a must. (Socket at least with one 15 A and three 5A). Large electrical
appliance such as air condition, refrigerator, even cannot use the same electrical wire as analyzer. Instrument
is equipped with a three core electrical wire, red wire is live line, blue wire is zero line, and yellow green wire
is ground lead. As figure 3-1 shows:
66
User Manual of CS-400 Auto-Chemistry Analyzer
HOT
Figure 3-1
Note: The unqualified environment may cause test value inaccuracy, analyzer damage and it is also
harmful to human body.
3.2.1 Procedure
Check if there is a physical damage on the packing when analyzer arrived. If yes, contact Dirui company or
local distributor. If not, open the package according to below procedure.
● Make sure that arrow on the package is up, upright the package.
● Open the accessory box and mainframe box, check if parts in box are complete, if not, please contact Dirui
company or local distributor.
● Check the packing and appearance of the analyzer, if there is a damage, please contact Dirui company or local
distributor.
● Protect the display screen and sample probe of front panel from outside force while transportation.
● Avoid vibration while transportation, check and debug the analyzer after transportation.
67
User Manual of CS-400 Auto-Chemistry Analyzer
Install software only by professional staff of our company. User is not allowed to uninstall software unless
abnormality occurs. Uninstall the software according to the following procedure:
Put CS-400 applied software into CD-Rom, click “set up.exe ” file, installation program initialize as figure 3-2
shows:
Figure 3-2
Figure 3-3
68
User Manual of CS-400 Auto-Chemistry Analyzer
Click “next ” in figure 3-3, a user information form pops up as figure 3-4 shows:
Figure 3-4
Click “next” in figure 3-3, a software select window of install folder will pop up. The default destination folder
is “C:\Program Files\Changchun Dirui Industrial CO., LTD\CS-400 Auto-Chemistry Analyzer applied software”.
User could revise installation path by click “self-defining” button as figure figure 3-5 shows:
Figure 3-5
Click “next” button in figure 3-4, a software select window of install folder will pop up. The default destination
folder is “C:\Program Files\Changchun Dirui Industrial CO., LTD\CS-400 Auto-Chemistry Analyzer applied
software”. User could revise installation path by click “self-defining” button as figure 3-5shows:
69
User Manual of CS-400 Auto-Chemistry Analyzer
Figure 3-6
Click “install” button in figure 3-6, initialize the software as figure 3-7,3-8 shows:
Figure 3-7
70
User Manual of CS-400 Auto-Chemistry Analyzer
Figure 3-8
Note: please deal with the waste liquid under the local rules and laws.
b) Outlet pipeline diluted waste: connect one end of the concentrated waste pipeline taken with the analyzer
with the concentrated waste liquid outlet interface(4 in figure 1-3), and place the other end into the waste liquid
collector.
c) Connection of liquid level sensor of concentrated waste solution: connect one end of the concentrated waste
pipeline taken with the analyzer with the concentrated waste liquid outlet interface(7 in figure 1-3), and place
the other end into the waste liquid collector.
(2) Check if data line is connected well between printer and analyzer.
71
User Manual of CS-400 Auto-Chemistry Analyzer
(3) Check if there are papers in printer, printout after turn on the switch.
General power supply of the instrument, analytical system, computer host, display of computer and printer.
Then, click the icon on computer screen or click “start-up”, then find the software CS-400 in
“program” and click it, after that, enter “system logging” window as the figure3-9 and 3-10 show.
Figure 3-9
Figure 3-10
Input user name, password, click “login” to get into the main menu of software, as figure 3-11 shows: click
“back” cancel login.( Initial user name: 001, password: 001)
72
User Manual of CS-400 Auto-Chemistry Analyzer
Figure 3-11
After successfully login, the software show as offline state, browse menu, check alarm information, user logout
function can be used at this state. Click “on-line” button, user could then process all browse and testing
operation.
If the wrong user name and password is inputted, a hint form will pop up, as figure 3-12 show:
Figure 3-12
If 3 times failed consecutively the user name and password , the program will exit automatically.
a) Connection: in figure3-11, click “ ”, “connecting” will show, after success of access, the status bar
will display “sleep”. At this time, all operation and tests can be carried out.
After logging, the instrument partially connected or not connected the cable, click the “Connection”, the screen
will show as figure 3-13 shows.
73
User Manual of CS-400 Auto-Chemistry Analyzer
Figure 3-13
If this shows on the screen, connect after get the power supply connected.
b) Exit system: In the window as figure 3-11 shows, click “ ”, enter window figure 3-14 shows:
Figure 3-14
Only exit system at off-line state. If analyzer is on-line, click “offline” button to exit system, dormant could be
carried out when analyzer stand-by.
Note:
● In order to prevent data from being damaged or revised by other people, exiting software when doctor takes a
rest is strongly suggested. Periodically backup database in order to avoid data lose.
●. Input initial user name and password when first login, select “user information” in “management” menu, set
user name, password and access authority for next time login.
The analyzer will be in sleep status after 20 mins of power on(waiting for the stability of power and
temperature)
74
User Manual of CS-400 Auto-Chemistry Analyzer
Figure 3-15
75
User Manual of CS-400 Auto-Chemistry Analyzer
(b) Orifice of the cuvette should be regular. Deformation and extrusion is not allowed.
(b) Size: Blankness between start and finish line should within be 3mm when cutting barcode as figure 4-1
show:
Width
Code39 5~10
Code93 4~12
Code128 5~22
12of5 4~15
UPC-A 11
UPC-E 6
EAN-8 7
Figure 4-1
76
User Manual of CS-400 Auto-Chemistry Analyzer
In order to obtain correct barcode, 15mm-20mm between cuvette bottom and barcode lower line is required.
Make sure barcode is on the outer side of sample disk when place on the test tube rack., as figure 4-2 show:
Barcode
Figure 4-2
When the CODE39 is not capitalized, please add “+” before the corresponding capitalized letters of the ID code
displayed on the screen and the report sheet printed after scanning.
Note: the “ ”, ‘’
,()are not allowed for the barcode, or it cannot be identified.
Barcode reader will continually identify 1-50 sample on outer track of sample disk when processing barcode
reader checkup, and the scanned information will be showed in “ maintenance” menu.
(b) Size: width within: 12mm-15mm, length within 40mm. As figure 4-1 show.
77
User Manual of CS-400 Auto-Chemistry Analyzer
(c) Blankness between start and finish should within be 3mm when cutting barcode as figure 4-1 show.
Blankness between bottle bottom and barcode should be within 15mm-25mm, thus barcode could be read
correctly.
When checking reagent barcode reader in “system maintenance”, reader will identify barcode info. and
display the scanning result of R1, R2 reagent disks in the window “system maintenance”.
The read information could be showed in “reagent information” menu as “disc No.”, “position”, “reagent name”,
“reagent type”, “remaining reagent volume”, “remaining test No.”.
Position: Each reagent disk has 1-45 positions, 1-44 are used for place reagent, and 45 position is specially used
for place CS-anti-bacterial phosphor-free detergent.
Note: regular clearing of the reading window of sample disk and reagent disk barcode reader is required.
Instrument consumes 25L/Hour water at peak value. The purified water equipment should meet the following
requirement:
Note: To use/maintenance the purified water equipment, please refer to user manual, or consult the distributer or
manufacturer.
78
User Manual of CS-400 Auto-Chemistry Analyzer
(a) Status bar: Shows display status on the top of menu. As figure 5-1 shows:
Figure 5-1
Description:
: Represent display status: stand-by, testing, emergence stop, sampling stop, maintenance operation,
sleeping mode.
: Communication monitor mark. When communication is under normal status, the icon turns blue; when
communication abnormal, the icon turns black. Once the icon color turns from blue to black, that indicate
communication failure.
: Alarm icon. This icon occurs in status bar when alarm is issued. Click the icon, start alarm checkup, solve
the problem according to the remedy.
: Display temperature of circulating water in incubation bath, regular water temperature is within
37℃±0.1℃. Alarm issued when temperature above 45℃. Alarm also occurs during test status.
(b) Main function keypad: Select menu by single mouse click. Single click on corresponding function key, the
border will change color correspondingly. As figure 5-2 show:
Figure 5-2
(c) Working space: enter specific operating space, as figure 5-3 show. Single click corresponding function key
79
User Manual of CS-400 Auto-Chemistry Analyzer
Figure 5-3
(d) Hint bar: instruct user how to use software, hint the input range, input method, operation error, as figure 5-4
show:
Figure 5-4
(e) Shortcut key space: for convenience use, as figure 5-5 shows. Single mouse click on corresponding function
key or press F2-F9 on keypad.
Figure 5-5
80
User Manual of CS-400 Auto-Chemistry Analyzer
Figure 5-6
F7 User log-out
81
User Manual of CS-400 Auto-Chemistry Analyzer
Reagent Info Manual registry, remove reagent info, barcode scan, reagent level
System setup Report format Report Info, print sequence, print format setup
Figure 5-7
82
User Manual of CS-400 Auto-Chemistry Analyzer
Select function key of software by single click mouse button. Input value and character combine with keypad
(switch by shift + control, input method is depend on windows system).
Mode form: other menu cannot be opened until mode menu is closed. Click “exit” or “close” button to exit. As
figure 5-8 show:
Figure 5-8
Modeless form: other menu can be opened when modeless menu open, system will automatically close last menu
when other menu is opened. As figure 5-9 show:
83
User Manual of CS-400 Auto-Chemistry Analyzer
Figure 5-9
List box is used for displaying a information. List box is also used for finding and selecting needed information
from displayed information. As figure 5-10 shows:
Figure 5-10
Scroll bar is used for adjusting display range in list box. Scroll bar is divided into longitudinal scroll bar and
horizontal scroll bar. As figure 5-10 shows:
84
User Manual of CS-400 Auto-Chemistry Analyzer
Figure 5-11
Check box: more than two functions can be chosen at the same time, as figure 5-12 shows:
Figure 5-12
85
User Manual of CS-400 Auto-Chemistry Analyzer
Figure 5-16
86
User Manual of CS-400 Auto-Chemistry Analyzer
Note:according to different test condition,sometimes instrument processing capability is lower than 400 test/hour
Figure 5-17
87
User Manual of CS-400 Auto-Chemistry Analyzer
Table 6.1 shows the operation flow. For detailed operation, please refer to table 6.2.
Consult
Operation step Form / function key Operation
index
1. Check before operation Check before turning on power 6.2.1
2. Connect water unit and turn Open water faucet, turn on the power of
on CS-400 power purified water unit and CS-400 6.2.2
Software login System Login Input operator’s ID No. and password.
3. Check instrument status
1) Check alarm Alarm information Please refer to chapter 13 “alarm
maintenance”
Execute light quantity checkup, check if
test value is within regular range
2) Check light quantity of System maintenance Execute cell blank check, check if test 6.2.3
photometer value is within regular range
3) Check cell blank System maintenance Check if temperature of incubation bath
is within 37.0℃±0.1℃
4) Check temperature of Status bar
incubation bath.
4. Check analytical conditions
1) Item added System setup Add chemistry item
6.2.4
2) Chemistry parameter System setup Check chemistry parameter
3) Check K value Calibration info Check calibration curve and K factor
5. Reagent preparation (reagent
info)
1) Check reagent residual Reagent info Check reagent remaining volume and
volume remaining test times. Place the reagent at
the corresponding position on the 6.2.5
Reagent disk.
2) Preparation for Reagent info
photometry item and
preparation for ISE item
6. Setup of calibration and Calibration info Check item name of calibration analysis
6.2.6
control item Quality control Check item name of QC
7. Sample registration and test Sample registration Registry and check sample info, patient
6.2.7
info, and operator info.
8. Start analyze
1) Preparation of sample, Put routine sample, standard and control
standard solution, Controls, sample on sample disk
detergent start analyze Select test condition, and execute “start 6.2.8
analyze”.
2) Set starting condition and
start test
9. Testing Monitor the instrument when testing.
(1)System monitor System monitor
(2) Sampling stop/continue Sampling
stop/continue 6.2.9
(3) Emergence stop Emergence stop
(4) Sample super addition. Sample registry Carry out sample edit when testing and
click start analyze.
88
User Manual of CS-400 Auto-Chemistry Analyzer
Consult
Operation step Form / function key Operation
index
10. Check test result (result To search, amend, delete test result,
Result data 6.2.10
data) check sample reaction curve
In “recheck” form, click “start recheck”,
11. Recheck sample test Test result 6.2.11
and send the recheck instruction.
12 Completion of analysis
1) Recheck result Test result Check and printout the test result;
2) Database backup System management Periodical backup database, one week is
suggested.
3) System dormancy Instrument could automatic start at 6.2.12
specified time after setting
4) Turn off instrument Turn off instrument, computer, connect
5) Preparation of next water unit.
operation
(b) Check communication wire and power wire which connecting host computer, instrument and printer.
Make sure the wires are well connected .
(d) Check if there are water drops, contamination and bending of reagent probe, sample probe and stirring
mechanism.
(e) Check if the detergent is enough. Place CS anti-bacterial phosphor-free detergent at 45th position on R1,R2
reagent disk. CS alkaline detergent in detergent box should be enough, detergent at sample disk W1,W2,
W3 should be enough. For detailed information please refer to “12.1.3 detergent ”.
(f) Check if waste solution. bottle is empty. Ignore this operation if drainage device is connected with down
pipe.
(g) Check if there are air bubbles in syringes ( leakage and air bubble may cause incorrect data)
! Warning :
△
CS serial detergent is corrosive liquid. In case of skin contact, flush the area with water, rinse immediately with
plenty of water and seek medical advice.
(a) Turn on the power of purified water, and open the water faucet
(b) Turn on the power supply of CS-400.The main power switch lies in the right lower side of instrument. Turn
on the switch when there are reagents in reagent disk, therefore the cooling system may under normal
working condition. The power of analyzed part lies in the right upper side of instrument.
89
User Manual of CS-400 Auto-Chemistry Analyzer
(c) Login CS-400 software, make sure carry out online instruction before testing.
Single-click “ ” key in the shortcut area to check alarm. When alarm issues during operation, the
alarm code, level, description and time can be displayed on the screen. (As figure 6-1shows)
Figure 6-1
For detailed information, click the information, and the detailed description will be displayed in the textbox
(as figure 6-2 shows)
Figure 6-2
At the same time, the remedy will be displayed simultaneously in “remedy” textbox ( as figure 6-3 shows).
Figure 6-3
“ ”indicates buzzer on , the instrument will make a buzz when alarm issues.
90
User Manual of CS-400 Auto-Chemistry Analyzer
When alarm is issued, the alarm icon “ ” may occur. Click the icon to display the alarm information
form.
Click“ ” key , all the information in the “ alarm info.” form will be deleted. Select one of the
alarm information and click “ ” key, the selected information will be deleted.
ISE reagent remaining volume alarm: Setup the lowest test number of reference liquid, diluent , Internal
standard liquid of ISE test.
Sampling stop alarm: Select the function, and alarm will issue when sampling stops.
Linearity check: Input the limit value of Rate assay or linearity assay.
Figure 6-4
Single-click “ ” button, select “ light quantity check ” option, and click “execute” button.
Instrument will automatically carry out light quantity check. The light value of current time and last time can be
showed in result column. As figure 6-5 shows:
91
User Manual of CS-400 Auto-Chemistry Analyzer
Figure 6-5
As light source lamp aging gradually, the check value may increase daily. AD value of all wavelengths should
be less than 18000. If the AD value is more than 18000, please replace new light source according to chapter
“12.4.3”.
Single-click the “ ” key, select “cell blank check” option, and then click“ ” key.
Instrument will automatically carry out the cell blank check. The cell blank value can be displayed in result
column. Click key to print out the test result. (As figure 6-6 shows)
92
User Manual of CS-400 Auto-Chemistry Analyzer
Figure 6-6
Cell blank value of the first reaction cuvette should be less than 18000, the difference between cuvettes (the
difference between current one and the first one) should be within ±800. When test value out of this range,
please replace reaction cuvette according to chapter “12.2.4”.
Before a new reaction cuvette is replaced, immerse it into 2% CS anti-bacterial phosphor-free detergent for
more than 8 hours, and then flush it with purified water before install. Only carry out test when cell blank value
is qualified.
Observe the temperature in status bar on the top of screen, check if the temperature of incubation bath is within
37±0.1℃. A warning level alarm is issued when temperature over 37±0.5℃, but instrument will continue
analyze at this time. A warning level alarm is issued when temperature of incubation bath over 45℃ (computer
stand-by).
Note: After power on, it takes 20 minutes to ensure the constant (37±0.1) ℃ of incubation bath. And it also
takes several minutes to stabilize the light source lamp. Therefore, form information can be browsed, parameter
can be login, alarm info. can be check even computer is not in standby status. Sample testing can be only
carried out after computer stand-by.
Setup the “analyze item added, chemistry parameter, calibration parameter and control parameter” before
testing. As to the parameter setup, usage method and storage information, please refer to the user manual or
consult the relative manufacturer or distributor.
93
User Manual of CS-400 Auto-Chemistry Analyzer
Item should be added before parameter setup. Click the “ ” key, and its -click the
Item name and number can’t overlap, if the same item of different company is added, please distinguish them
with different letters or numbers. Test takes place in terms of the sequence (from small to big) of item No..
Set up/check the reagent chemistry parameter according to the reagent manual. Some parameters are necessary,
such as: test item name, wavelength, test light point, test time, sample volume, reagent volume, reagent position,
reaction type, reaction direction and decimal digits etc. and some parameters are significant to a reliable test
result, such as linearity range, absorbance limit, etc. Strongly suggest the operator input correct parameter
before test.
Click “ ” key, then click “ ” key. Please setup the chemistry parameter
according to the reagent manual. For detailed operation, please refer to chapter “9.1”
The function of profile item is to put the relative items together, such as perform the whole set of liver function
or the whole set of kidney function. Use one key to finish many items registration, which is convenience for a
rapid sample registration.
Click “ ” key, then click “ ” key. For detailed operation, please refer to chapter “9.2 ”.
The function of calculated test is to determine result on the basis of test A, B or more test in order to calculate
test result of a new item.
Click “ ” in main function key field, then click “ ” key,. For detailed operation,
please refer to chapter “9.3 ”
The function of cross contamination is to decrease or avoid cross contamination among different items. It
includes the cross contamination of reagent probe, reaction disk and sample probe.
Note: Due to the different reagent formula, the analytical result of other items can be effect. The contamination
degree is different according to varied reagents. For detailed information, please consult the manufacture or
distributor. Strongly suggest the operator setup the reagent which cross contamination may occur separately. Or
through the “Cross contamination” setup to decrease the cross contamination among tests.
94
User Manual of CS-400 Auto-Chemistry Analyzer
Click “ ” key, then click “ ” key. For detailed operation, please refer to
chapter “ 9.4 ”
If 1 point linearity method is used for calibration, Click “ ” key, then click “ calibration result “ key,
as figure 6-7 shows.
Figure 6-7
(1)Activate ISE
If the instrument has equipped with ISE module, please activate ISE in “system setup” form first, then carry out
If sign “√” in “ washing after test”, sample probe will assimilate the CS-ISE detergent (locate in W2 position of
95
User Manual of CS-400 Auto-Chemistry Analyzer
sample disk), then rinse the diluent bath and ISE pipeline. This operation won’t take place if it is not selected.
Select item name and sample type in “ISE setup” menu. In calibration parameter work area, select Calibrator
position from the pull-down menu, input the concentration of Calibrator. There are three kinds of Calibrators:
low concentration, high concentration and blood sample. 15ul is a fixed sampling volume of Calibrator, there is
no need to set again. Input the reference range, linearity range, decimal digits, QC interval (default value:0.
input interval time if needed), slope and intercept, and then click“ save parameter” key, as figure 6-8 shows:
Figure 6-8
Note : The sample volume of ISEis fixed, 15ul, and no set value is needed.
(a) Reagent confect, use and storage should be performed strictly according to reagent manual. Avoid air
bubbles occur in reagent. Due to the surface active agent of reagent may cause bubbles. And when liquid level
sensor touches the bubbles, it may misjudge it as reagent, thus cause the inaccuracy of sampling volume and
affect the test result.
96
User Manual of CS-400 Auto-Chemistry Analyzer
If the added reagent is coming from different plant or different lots of the same plant, it may cause the change of
the reagent component, thus effect the test result.
(c) Turn the reagent knob counter-clockwise, open the lid, and put reagent bottle at the relevant position on
reagent disk. Table 6-10 shows the reagent type:
Reagent 1
Reagent 4 Reagent disk1
Diluent
Reagent 2
Reagent disk 2
Reagent 3
! Warning:
△
Execute resetting process before barcode reading. Sample probe, reagent probe and stirring rod mechanism
will start up later. So please leave the instrument after closing the cover of reagent disk. Don’t extend your
hand into instrument in order to avoid body injury.
(a) Single-click “ ” key , and then click “reagent information “ key, input reagent information in
“manual registration” form as figure 6-9 shows.
Figure 6-9
Select 1( reagent disk 1), 2( reagent disk 2) in the pull-down menu. Input reagent position from 1 to 44.
Position 45 of both reagent disks is for CS-anti-bacterial phosphor-free detergent. Select reagent name, reagent
97
User Manual of CS-400 Auto-Chemistry Analyzer
type, bottle specification in the pull-down menu, then click “ register” key.
(c) Reagent scanner cannot recognize unclear barcode, to solve this, click “ barcode’ key to registry by manual,
input valid barcode in “barcode” input box, click “registration” to register reagent information.
Note: Reagent information cannot be only registered or removed under running status.
Reagent can be automatically registered through “barcode scan” when barcode information are clear and
complete.
Single-click “barcode setup” key, input the barcode number and item name, and then click “ add” key, as figure
6-10 shows.
Figure 6-10
Select the information to be deleted, then click “ delete” to eliminate the information. Single-click “ close” key
to exit the form.
Click “ ” key in “reagent info.” menu, instrument will automatically scan reagent on both
reagent disk, relative reagent information (such as reagent position, name, type, shelf time, model, batch
number and bottle standard) will be displayed in “ reagent info.” list. When reagent remaining test number is 0,
gray color will appear; when expire, red color will appear. As figure 6-11 shows.
98
User Manual of CS-400 Auto-Chemistry Analyzer
Figure 6-11
(4)Reagent Horizontal
Click “ ” key in “ ” form, instrument will check remaining reagent volume and
remaining test number, and the test information can be showed in “ reagent info.” list.
Once reagent probe sucked reagent, the remaining reagent volume and remaining test number are tested and
updated.
Select the information to be deleted by mouse, click “ ” key, a delete form will pop up, as figure 6-12
shows.
Figure 6-12
99
User Manual of CS-400 Auto-Chemistry Analyzer
Note:
(a) If the reagent disk cover is open during measuring, a warning level alarm will be issued. Do not open the
cover when testing in order to avoid dangerous and instrument damage.
(b) After reagent information registry, carry out reagent level check before test in order to check the reagent
remaining volume and remaining test number.
Open the top cover, place internal standard liquid (IS), diluent ( DIL) , reference electrode (REF) in ISE
position at the left side of unit system.
Input the reagent volume of internal standard solution (IS), diluent (DIL) , reference electrode (REF) in “ISE
setup” Form. Click save button to same parameter. Remaining volume will automatically reduce after
instrument use ISE reagent.
(3) Carry out ISE ( all) rinse in “maintenance” menu . If one reagent is changed, do the relevant rinse.
Table 6-3
Table 6-4
100
User Manual of CS-400 Auto-Chemistry Analyzer
(a) Click“ ” key, select calibration method in “ calibration ” pull down menu in
Calibration Standard
Object Applied assay mode Application example
type solution volume
Blank solution renew reagent All calibration methods K coefficient method, when
Blank
blank value standard solution test is
calibration
omitted
Span One point except Renew k value Two point linear and Recheck standard solution
calibration of blank solution. multi-points linear point 1
Blank solution Renew reagent Two point linear and Linear 2 points method
and span points blank value and multi-points linear finish standard curve.
2 points
K value Multi-point method,omitted
calibration
the quantity of calibration
solution.
All concentration Renew standard Multi-point linear, Multipoint standard curve,
registered curve by all points isozyme Q, isozyme method ,
Full points
isozyme P , electrolytical standard
calibration
and nonlinear working curve
curve calibration
Table 6-5
(b) Select calibration item in “ test item” work area. Sign “√” in the function block by single-clicking mouse,
show as: “ ”, double-click to cancel it, and “ ” is showed.
(c) Click “calibration register ” functional key , the registered item name and item type will be displayed in the
browse area. In order to check the concentration and position of calibration item, single-click the desired item
name, the calibration parameter will be displayed in the right side. As figure 6-13 shows:
Figure 6-13
101
User Manual of CS-400 Auto-Chemistry Analyzer
(d) Delete the calibration item; Select the desired item, click “ delete” key , the item will be deleted.
(e) In order to execute ISE calibration, single-click “ ” button, select “execute ISE”.
Note: The test item should in “execute” status, and select “calibration before analyze” in “start analyze” form.
Click “ ” key, register the name, position, batch No. information according to chapter 9, and set up its
target value and standard deviation.
Note: QC registration of colorimetry item is the same as ISE item, select it in “QC test” pull down menu.
Click “ ” key, and then check “ sample info.” form. As figure 6-14 show.
Click “ “ to exit.
Figure 6-14
102
User Manual of CS-400 Auto-Chemistry Analyzer
(a) Input sample number in “ sample number” functional block, or click “ previous / next” key to select the
sample number.
(b) In order to register routine sample , select the sample disk number from 1 to 9 in “ disk number” menu.
Input sample position No. from 1 to 50 in “ position number” functional block. If sample disk number is
changed, instrument will issue stop level alarm, but analysis do not stop. Single-click “continue” key, sample
probe begin to sampling from new sample disk.
In order to register emergent sample, single-click “emergence” key, input number 51-70 in “ position”
functional block.
(c) Set up sample type, sample volume, cup type and check date etc. information.
(e) Click the item name to be tested by mouse in item information work area. “ ” indicate select. Items can
be also inputted the by profile function.
(f) After editing , click “ Register sample” key . The registered information will be displayed at the right side.
Sample number: Input sample number to functional block, the number is uniqueness, one sample only has
one number in a day.
Barcode number: barcode number is stick on the outside of sample tube, when scanning, the scanned
value will be displayed in “ barcode number” functional block. If barcode number scanning failure, input the
effective number in “ barcode number” functional block.
Sample type: Select sample type in pull down menu. This function is in “ other info.” of the “ system
administration” form. For detailed operation please refer to chapter “8.3”.
Repeated test: The test times to same record. The default time is 1 , the max. is 100
Dilute: indicate if sample need to be diluted. Select diluent position in chemistry parameter. Input sample
volume, diluent volume, sample volume after diluting. Dilution function must be selected in sample
information.
Previous sample: when register or delete sample, sample number decease 1 by single-click mouse button.
Next sample: when register or delete sample, sample number increase 1 by single-click mouse button.
When various samples are registered in same test item, batch sample registration can be used.
Single-click “ ” key in “sample register ” menu. enter “ batch sample registration” menu as
figure 6-20 shows.
103
User Manual of CS-400 Auto-Chemistry Analyzer
Figure 6-20
(a) Input the start sample number in the first functional block and the last sample number in the second
functional block of “ Sample No. range”. The last sample number should be large than the first number.
(b) Select the start disk number in “ disk No.” menu, input the beginning position No. (1 to 50) in “position” .
Sample disk No. and sample position No. will automatically increase from little to large when batch
registration.
(c) Select the relevant info. in “ sample type” “ sample volume “ menu.
(d) Single-click item name by mouse , select test item , item compounding function can be used also.
(e) After edit, click ” register” key. If one of batch registration samples is the same as single registration
sample, instrument will remind user that registration failed, display as figure 6-16.
Figure 6-16
Click “ OK” key to exit. Edit the right batch samples again.
104
User Manual of CS-400 Auto-Chemistry Analyzer
Figure 6-22
(a) Input sample No. in “ sample No.” functional block, click “ Previous” “ Next” key to choose sample No.
(b) In functional block, choose or input patient name, age, gender, case history No., patient type, delivery dept. ,
delivery doctor, bed No., Check doctor, audit doctor, delivery date, clinic diagnoses and remark etc.
Patient name must be input when edit patient info.. Or instrument will refuse registration.
(c) Single-click “ patient registration” key, patient name will be automatically displayed in the browse area.
(d) Single-click “ sample info.” key enter the menu. Click “ close” key to exit the menu.
Age: Select age from pull-down menu: year, month, day, and time. And then input numbers.
Patient type: Select “clinic, medical insurance, hospital, physical exam” in pull-down menu. This
function can be setup in “ other info.” of “ ” menu. For detailed operation, please refer to
chapter “ 10.3” .
Dept.: Select the type of sample delivery dept. in pull-down menu. This function can be setup in “ hospital
Doctor: Select the doctor name. This function can be setup in “ hospital info.” of “ ”menu.
105
User Manual of CS-400 Auto-Chemistry Analyzer
Auditor: Select the doctor name. For detailed operation refer to chapter “ 10.1”.
Single-click the desired item in browse area, or click “ previous” “ next ” key to choose the right sample, or
directly input the sample number to be modified in “ sample No.” functional bock, then click “OK” key to edit
new information, click “ sample registration” key to enter “overlay confirmation” menu. As figure 6-18 shows.
Figure 6-18
In order to delete sample, click “ ” key in “ sample info.” menu, as figure 6-19 shows:
Figure 6-19
In the left functional block, input the start sample No. to be deleted. Input the end sample No. in the right one.
For deleting one piece of information, input the same sample No. in the start and end functional block . The
start sample No. should be less than/ equal to the end No..
Note 1: Registered sample (Not yet test) can be modified or deleted during stand-by and operating .
Note 2: Operator can register sample info. firstly, then register patient info.
Place routine sample, Calibrator, Controls and detergent. in the relevant position on sample disk.
Test colorimetric item, place CS- alkaline detergent in the position of sample W1 disk; test ISE item, please
place CS- ISE detergent in the position of sample W2disk; avoid crossed contamination, place CS- acidic
detergent in the set position.
According to the Calibrator position, which is set in “ chemistry parameter”, place the relevant Calibrator in
middle or inner track of sample disk.
Place the relevant QC sample in inner track in terms of the QC position set in “ QC registration” menu.
According to the routine sample registration position of “ sample register” menu. , place routine sample in
positions 1 to 50 of outer track, and place the emergency in the positions 51~70 of middle track.
Note: QC sample and Calibrator should be tested in standard cup and micro cup.
After sample registration, setup “start analysis” condition before test. Click“ ” key to start analysis.
As figure 6-20 shows.
Figure 6-20
At the right side of start menu, operator can setup according to the requirement. For calibration test, click the
After test order sent successfully, “ start conditions” form will automatically close. Or click “ ”
key to exit.
Note : Instrument began to test according to “ calibration – QC – sample” sequence. If there are ISE items, test
according to “ ISE – colorimetric” sequence.
107
User Manual of CS-400 Auto-Chemistry Analyzer
6.2.9 Testing
(1)System monitor
During sample testing, the status of sample disk, analysis disk and reagent disk can be real time monitored.
“ ”, and then click the reagent position in reagent disk chart, all the relevant
information will come forth , as figure 6-21shows:
Figure 6-21
Reagent remaining volume: Single-click the corresponding position of reagent in the reagent disk chart,
the reagent remaining volume will be showed directly by chart and percent mode in menu.
Remaining volume test: In terms of the remaining volume and the set volume in chemistry parameter,
instrument will automatically calculate the remaining test quantity.
Remaining volume: Display the reagent remaining volume in the current position in ml unit.
Reagent normal: Test volume conforms to the test requirement. Show as green color.
Reagent shortage: Reagent volume or reagent remaining test times is less than setup value in “System
setup” form. This is called reagent shortage. Show as violet color.
Reagent not in use: Reagents has registered but it is not used for test. Show as white color.
(b) Enter the“ ” menu, single-click “ ”key, and then click the relevant position in
reaction disk chart. All information of reaction cuvette will be displayed as figure 6-22 show:
Figure 6-22
Test status: Display the current reaction cuvette status. In the monitor chart, different color indicates
different status.
Vacancy : The reaction cuvette is not used for current test. Show as white color.
Reagent 1, 4 : Reagent in the reaction cuvette is infused with reagent 1(R1) and reagent 4 ( R4) . Show
as blue color.
Reagent 2, 3 : Reagent in the reaction cuvette is infused with reagent 2(R2) and reagent 3 ( R3) . Show
as pink color.
Completion : Result has been calculated from the tested sample of the reaction cuvette. Show as green
color.
Dirty cuvette: The cell blank value has exceeded the normal range. Show as red color.
Sample No.: The No. of sample that is tested in the reaction cuvette.
Test No.: Instrument will automatically produce the serial number according to the test sequence.
109
User Manual of CS-400 Auto-Chemistry Analyzer
Item name: The name of analyzed item, which is tested in the reaction cuvette.
(c) Enter the“ ” menu, click “ ” key, select the disk No. from “ sample disk” menu and
the sample No. from “ sample No.” menu, or click the monitor chart of sample disk by mouse, the relevant
information will be displayed as figure 6-23 shows:
Figure 6-23
Test status: Display the current sample status. In sample disk chart, different color indicates different status.
Waiting for test: Sample in this position is registered, but not sampling. Show as green color.
Sampling: Sample probe is assimilating in this position and adding them to reaction cuvette. Show as
yellow color.
Analysis: Sampling Complete, and sample is being analyzed. Waiting for the result. Show as pink color.
Sampling stop can be execute only when testing, click the “continue” key to continue sampling.
If disk number changes, instrument sampling may temporarily stop, click “ continue” key to go on sampling.
110
User Manual of CS-400 Auto-Chemistry Analyzer
Click “ ” key when testing, instrument may stop the current action. Emergence stop is not allowed
when scanning sample barcode.
When testing, other sample could be edited in “ ” form. Click “ ” key to send
“sample add” order.
● Sample with lipemia, homolysis and icterus could affect the test result.
● Make sure that sample is not cloudy and with no clot, or sample probe could be jammed and effect test
result.
● Substance in sample, such as medicine, anticoagulant , preservative , may disturb test result.
● Avoid long time contact with air, or sample will volatilize, thus affects test result.
● System maintenance that not conform to user manual could cause contamination and instrument damage,
thus effects test result.
● Revise or add test result is not recommended by our company. And we are not responsible for this
operation.
Click “ ” key. Operator could check, delete, modify, audit, report preview, report printout, manual
recheck, history review of test result.
For checking daily result, select “the same day results” in “ ”. All daily sample
information can be displayed in the “ result data” menu as figure 6-24 shows.
111
User Manual of CS-400 Auto-Chemistry Analyzer
Figure 6-24
Sample information is at the left side of menu. Sample test result is at the right side of menu.
In order to check sample reaction curve, click “ ” key, select the desired sample number and
test item, Select the wavelength type, that is dominant wavelength and secondary wavelength. The absorbency
of each light point is connected by line. As figure 6-25:
Figure 6-25
112
User Manual of CS-400 Auto-Chemistry Analyzer
In order to check the absorbency value of test light point, select the desired point in pull down menu. The
absorbency value can be displayed.
(b) Review
In order to audit single sample, single-click the record “ “ key. In order to check batch
sample, single-click “ ” key, input the start and end sample number in “ sample number
range” functional block, and then click “ “ key. As figure 6-26 show:
Figure 6-26
Note 1: The start sample No. should be less than or equal to the end sample No..
113
User Manual of CS-400 Auto-Chemistry Analyzer
Figure 6-27
In order to print single sample report, choose the record, then click “ ”key.
In order to print batch sample report, single-click “ ” key. Input the start and end sample No.
in “ sample No. range” functional block, and then click “ “ key . For example, print sample
result from 1 to 14, as figure 6-28 show:
Figure 6-28
Single-click “print the audited report”, this procedure only apply for checked sample report, not for the
unchecked report. If don’t choose this function, all report will be printed out.
114
User Manual of CS-400 Auto-Chemistry Analyzer
Note: When printing batch report, the number of start sample must be less than the number of end sample.
Double- click the sample record to be modified, enter the menu. Input the new result in the “ check result”
functional block, and then click “ save” key. As figure 6-29 shows.
Figure 6-29
Note: addition of result and modification are not allowed if no result happens to some samples.
Single- click the sample record to be deleted. Then click ” key , as figure 6-30 shows .
Figure 6-30
All samples except those are being tested without result can be added with both analytical item and manual
item.
Click “Super add Result” key in “test result” to enter addition result form, select , as
figure6-31 shows.
115
User Manual of CS-400 Auto-Chemistry Analyzer
Figure 6-31
Addition of analytical items: this method is applicable to the sample registered already which is being tested.
● Select analytical item sample No. in “sample No.” pull-down functional block.
● Select short form of analytical item sample No. in “analytical item” pull-down functional block.
● Select disk No. in “disk No.” pull-down functional block. Input the place of addition sample in “position”
input block.
● Display other information of the sample in the input block “sample type”, “sample volume”, test date”.
● Click “start” key in “start condition” form to carry out the addition test.
Add manual item: this method is applicable to the items not tested in the instrument but needed to add result in
“test result”.
Click “ ” key in “addition result” form to start manual item addition, as figure 6-32 show.
116
User Manual of CS-400 Auto-Chemistry Analyzer
Figure 6-32
● select additional manual item sample No. in “sample” pull-down functional block..
● select additional manual item short form in “item short form” functional block.
● display other information of the sample in the input block “item full form”, “unit”, “reference range”.
117
User Manual of CS-400 Auto-Chemistry Analyzer
Figure 6-33
Select the inquiry record as the start date: ” to the end date
Input the patient ’s whole name or last name in “ ” functional block, click
118
User Manual of CS-400 Auto-Chemistry Analyzer
“ ”key. The qualified record will be displayed in the functional block. Contents of the
pull-down menu is from “ hospital info.” from.
Select the delivery doctor in the pull-down menu of “ ” , input either the doctor’s
whole name or the last name , then click “ ” functional key , The qualified record will be
displayed in the functional block. Contents of the pull-down menu is from “ hospital Info.” form
(1)Automatic registration.
After sample test finished, instrument will automatically add sample info into reset form. The reset information
is defaulted set the same as first test. Operator can alter the data. Relationship between sample type and
sample volume is as figure 6-34 shows:
Figure 6-34
119
User Manual of CS-400 Auto-Chemistry Analyzer
(2)Manual registration
Click “ ” key in the result data menu to add this item to recheck menu. Set up the sample
Figure 6-35
Single-click “ ” key, instrument will carry out recheck test. Click the “ ” key
to exit.
After finishing measurement, confirm, audit or print the recheck sample in “ ” menu.
(2)Database backup
120
User Manual of CS-400 Auto-Chemistry Analyzer
(3)System sleep
Sleep indicates that instrument is at half-stop status, only cooling system keeps working. Instrument will
automatically start up in specified time.
Set up the time to awaken the instrument in “ system setup “ menu. Click “ sleep “ key. Instrument will shut off
all power supply besides cooling system. Status bar remind of “ instrument sleep”. Sleep mode can be set only
in the stand-by state.
If awakened time is not coming, click “ awaken” key to relieve sleep mode, instrument will carry out the same
operations as power on.
Exit the software program of CS-400, turn off the power supply according to the following sequence: printer
power supply, host computer power supply, display power supply, analysis system of control power supply,
general power supply (general power cannot be cut when reagents are in cooling unit).
Check if the reagent lid is closed well or not. Take away the sample cup or test tube with Calibrator, Controls,
diluent and sample. Drain the waste solution. in waste barrel; Check if reagent probe, sample probe and stirring
rod are contaminated or bended. Check if the surface of analyze unit /operation unit is dirty.
121
User Manual of CS-400 Auto-Chemistry Analyzer
Single-click the key to carry out the registration of calibration information and the check of
calibration result.
Figure 7-1
After registering, carry out calibration setup in “ ” form. Select “calibration before test” function.
Registration in “ calibration” form is not enough, without register in “start analysis” form, instrument will only
accept the calibration application, and calibration test won’t be carried out. Calibration rest cannot be renewed
either.
(a) Select the proper calibration type in “calibration type” menu. Calibration type is as table 7-2 shows:
122
User Manual of CS-400 Auto-Chemistry Analyzer
Span One point except Renew k value Two points linear and Recheck standard solution
calibration blank solution. multi-points linear point 1
Calibration
Logit- Log3P
Logit- Log4P
2 points Blank solution Renew reagent Linear two points and Linear 2 points method
calibration and span points blank value and multi-points linear finish standard curve.
K value calibration Multi-point method,omitted
Logit- Log3P the quantity of calibration
Logit- Log4P solution.
Full points All registered Renew standard Linear multi-point Multipoint standard curve
calibration calibrator curve by all Calibration isozyme method
points isozyme Q
isozyme P
nonlinear working
curve calibration
Table 7-1
Select the item name to be calibrated in “ registered calibration item” form. As figure 7-2 shows:
Figure 7-2
(a) Single-click “register calibration item” key, the selected items will automatically be displayed in registration
list. New registration item will carry out calibration. If not, even select the “calibration before test” function,
instrument will not carry out calibration. The concentration and position of Calibrator can be showed in the
form.
(b) In order to delete calibration item , select it and then click “delete” key.
Single-click “ ” key, check calibration result, such as: Reagent blank , K factor,
123
User Manual of CS-400 Auto-Chemistry Analyzer
approximate function of constant A.B.C from multi-points calibration curve etc. As figure 7-3 shows:
Figure 7-3
In order to modify the calibration result, delete the old result, input the new one, and then single-click
“ ” key.
(a) Single-click “ ” key in “ calibration result “ menu. and then select the desired item name to
be checked. The item of calibration type, S1ABS, K, A, B, C will be displayed in the form.
(b) The reaction curve of the calibration item is showed as figure 7-4. The abscissa represents the concentration,
the ordinate represents the absorbency. Absorbency range can be changed by revise “ ”. Click
“ ” key to exit.
124
User Manual of CS-400 Auto-Chemistry Analyzer
Figure 7-4
(3)Calibration tracing
Instrument will automatically store the absorbency of Calibrator. The tracing graph can check the stability of
absorbency variety; therefore the calibration accuracy can be checked.
(a) Single-click “ ” key, select the item name, the number of Calibrator. then click “ renew”
key. The 50 times absorbency value will be displayed in the graph. The abscissa represents the calibration
times, the ordinate represents the absorbency value. Absorbency range can be altered by revise
125
User Manual of CS-400 Auto-Chemistry Analyzer
Figure 7-5
(4)Reaction process
Check the absorbency variety of each item in different time point through reaction monitor form. Check the
reaction status and check if the test value of absorbency is stable or not through reaction curve graph.
(a) Single-click “ ” key in “calibration ” menu, select the item name and the Calibrator number.
Because of each Calibrator is tested twice, select the test times, main wavelength, sub wavelength etc., then the
reaction curve graph of Calibrator will be displayed. The abscissa represents the photometric point, the ordinate
represent the absorbency as figure 7-6 shows:
Figure 7-7
(b) In order to check the detailed absorbency value of one photometric point, select the desired point in
“ photometric point” pull-down list. The absorbency value will be displayed.
126
User Manual of CS-400 Auto-Chemistry Analyzer
(c) Change the absorbency range by revising “ ”. Single-click “ ” key to print out
Main wavelength- Sub wavelength: display reaction curve of main wavelength subtracting sub wavelength.
Note: after adding new item, calibration of the new should be implemented first, and in correct calibration
result may affect accuracy of result.
Click “print result” to preview the result of colorimetric item calibration. In order to print the result, click
“print” key upper side of screen.
Figure 7-7
127
User Manual of CS-400 Auto-Chemistry Analyzer
If carry out the calibration for ISE items, sign “√” in front of the item.
Instrument will automatically display the calibration result when carries out ISE test calibration. Alarm is issued
when abnormality exists.
In ISE calibration, Calibrator has been measured for three times. In calculation, the mean value of the second
and the third times result is used.
Check the slope value, the concentration of internal standard liquid, compensation value in “ calibration result “.
After finishing ISE calibration, instrument will automatically calculate the compensation value. If the
compensation value is the difference between the input value and the test value of Calibrator 3. In order to
revise the compensation value, delete the origin value, and input the number, then click “ ” key.
Click “print result” to preview the result of ISE item calibration. In order to print the result, click “print” key
upper side of screen.
128
User Manual of CS-400 Auto-Chemistry Analyzer
Random error: The difference between test result and the mean value of the same target tested many times in a
repetitious conditions is called random error.
System error: The difference between true value and the mean value of the same target tested many times in a
repetitious conditions is called system error.
Accuracy: The integration of system error and random error in the test result, indicate the consistent degree
between test result and true value.
Precision: The consistent degree among many test result of one target in a specified conditions, indicate the
degree of random error magnitude among the test results.
L-J ( levey Jennings) QC chart: QC chart is a kind of graph with quality control limit. QC limit is controllable
analysis method to known specimen (QC sample) carry out repetitious test to get the mean value ( X ) and
8.1 QC registration
Single-click the “ ” key, then click the “ ” key. At most eight QC samples can be used
simultaneously to carry out quality control. As figure 8-1 shows.
129
User Manual of CS-400 Auto-Chemistry Analyzer
Figure 8-1
Figure 8-2
Operator could select the needed QC regulation, click “ ” to save the setup. After setup, the
“QC interval” and “Monthly QC” may process QC analyze according to the regulation.
According to Westgard multi-rule judgment base, carry out the incontrollable analysis to the test result, as figure
130
User Manual of CS-400 Auto-Chemistry Analyzer
8-2 shows:
QC Data
NO
12S Under control
YES NO
NO NO NO NO
13S 22S R4S 41S 10 X
Out of control
Figure 8-3
1 2S : One QC result exceeds mean value ±2 SD, which is judged as warning regulation, and initializes other
regulation to check QC data.
1 3S : One QC result exceeds mean value ±3 SD, which is judged as lose control, this regulation is sensitive to
random error.
2 2S : Two consecutive QC result simultaneously exceed mean value +2 SD or -2 SD, which is judged as lose
control, this regulation is sensitive to system error.
R 4S : One control result exceeds mean value +2 SD, another exceeds -2 SD, which is judged as lose control,
and this regulation is sensitive to random error.
4 1S: Four consecutive QC results exceed simultaneously mean value +1 SD or -1 SD, which is judged as lose
control , this regulation is sensitive to system error.
10 : Ten consecutive QC results all are in the same side of mean value (higher or lower than mean value, no
X
requirement to the degree of deflection), which is judged as lose control, this regulation is sensitive to
system error.
131
User Manual of CS-400 Auto-Chemistry Analyzer
Figure 8-4
In order to delete, select the QC name in pull down menu, and click “ ”key, click
“ ”key to exit.
(c) Input lots No. of Controls in the “ QC lot number” functional field.
(d) Select type of Controls, such as blood, urine, in “sample type” pull down menu.
(f) Click “add” key when above parameters are correctly inputted. All inputted parameter is saved in the left
work area.
Note: after register the QC item, make sure check which item may carry out QC test, click “execute” key in
front of QC item.
In order to carry out QC test according to QC interval, input QC interval in “chemistry parameter”; In order to
carry out one time QC test before sample test, click “ ” key in “start analysis” form, and
then click “ ”.
132
User Manual of CS-400 Auto-Chemistry Analyzer
color changes after being clicked. Click “ ” key to finish this operation. All information of
this item will be deleted.
8.2 QC interval
The QC interval is set in the “chemistry parameter” menu, and instrument will automatically carry out QC test
according to the set interval sample number. After analysis finish, check the QC result in the “ ”
menu, and a QC result chart is showed as well. In the chart, the abscissa represent test times, the ordinate
represent concentration.
(a) Select QC item and lot No. in the “ ” pull-down menu, QC result will be displayed in the QC
chart, as figure 8-5 shows:
Figure 8-5
133
User Manual of CS-400 Auto-Chemistry Analyzer
(b) Single-click “ ” key to analyze QC data according to Westgard Multi-rule Judgment base.
(c) Single-click “ ” to check and modify the QC result, in order to revise the result , input the
new data , click “ result revise” key as figure 8-6 shows
Figure 8-6
In order to add QC result, single-click “Add” key after input desired QC item result in the “Result” functional
block.
In order to modify QC result, single- click the desired QC result in the left side working area, input
“modification result” into the “Result” functional block.
In order to delete QC result, single-click the desired QC result in the left side of working area, the click
“Delete” key.
After QC test finish, instrument will automatically calculated real test QC target value (mean value), standard
deviation, coefficient of variation, range (alteration range) etc. data.
- ∑ Xi
i =1
Target value( X ):
N
134
User Manual of CS-400 Auto-Chemistry Analyzer
∑ ( Xi − MV )
2
i =1
Standard deviation(SD):
N −1
SD
Coefficient Variation(CV%): ×100%
mv
Range: X max − Xmin
N: test times, Xi : test result.
Select “ QC before test” in the “start analyze” form when test. After finish analysis, check the QC result in the
“ ” form. This abscissa of QC chart represent test date, ordinate indicate concentration.
(a) Select QC item and lot No. and month in the “ ” pull-down menu, QC result will be displayed in
the QC chart as figure 8-7 shows:
Figure 8-7
(b) Single-click “ ” key to analyze QC data according to Westgard Multi-rule Judgment base.
135
User Manual of CS-400 Auto-Chemistry Analyzer
(c) Single-click “ ” to check and modify the QC result, in order to revise the result , input the
new data , and click “ result modify” key.
After Qc test finish, instrument will automatically calculated real test QC target value (mean value), standard
deviation, coefficient of variation, range (alteration range) etc, data.
136
User Manual of CS-400 Auto-Chemistry Analyzer
Figure 9-1
Note: After the parameter of each menu is edited, operator should click “ ” to save the data.
Click “ ” key at the lower right side of form, add or delete the test items as figure 9-2 shows:
137
User Manual of CS-400 Auto-Chemistry Analyzer
In order to delete item, move the scroll bar at the right side of list box to look for desired item, click it by mouse,
the color of chosen item will change, appear in front of chosen item, click “ delete “key, The item will be
deleted.
Click “ ”key in “ ” menu. Operator can edit or revise the analysis parameter
of colorimetric items. As figure 9-3 shows:
138
User Manual of CS-400 Auto-Chemistry Analyzer
Figure 9-3
Item name: Select the item’s abbreviation from the pull-down menu. All items may display automatically in
the list box.
Decimal digits: Operator can select the decimal digits of test result and printout result in the pull-down list
box.
Items full name: Input the full name of testing item, such as ALT, whose full name is Alanine
Aminotransferase.
Quality control interval: Input the quantity of interval sample. Input the integral number of 10 times, the least
interval quantity is 10, the most is 1000.
Test method: Select one method in the pull-down menu which is conform to reagent requirement, 1 point
assay, 2 point assay, rate A assay, rate B assay, 1 point rate assay, 2 point rate assay, 3 point assay. For
detailed introduction, please refer to 2.2.1
Item unit: Choose the chemistry item in the pull-down menu, For the add and delete of item unit, please
refer to chapter “ 8.3 ”
Test time: Test time can be selected from the “test result” pull-down menu.
Photometry point: Instrument will record one time absorbency every 18 seconds. Please input proper
photometry point according to reagent instruction. Effective photometry point should be inputted within 2 to
49 (0 represent no input). Tested absorbency value of each test light points can be searched from reaction
curve.
Main wavelength: There are 12 optional wavelengths in pull down menu, select them according to reagent
manual. Main wavelengths : 340 nm、380 nm、405 nm、450 nm、480 nm、505 nm、546 nm、570 nm、
600 nm、660 nm、700 nm and 750nm.
139
User Manual of CS-400 Auto-Chemistry Analyzer
Sub wavelength: When adopting double wavelength assay to test or analyze sample, select the wavelength
from 12 wavelength in pull down menu. Secondary wavelength include: 340 nm、380 nm、405 nm、450
nm、480 nm、505 nm、546 nm、570 nm、600 nm、660 nm、700 nm and 750nm . The difference of the
absorbance value between main wavelength and secondary wavelength is used for calculated result. when
select single wavelength test, select “0” of secondary wavelength.
Instrument factor (Y=a X + b): Carry out relation calibration. The test result will be higher or lower than
expected result or result from other instrument. In order to make result in accordance with expected result or
result from other analyzer, add the calibration relation in result calculation.
Relation equation:
Y=a X + b
When test result of the analyzer is the same as expected value, or when results of any two analyzer is
accordant, let a=1, b=o.
When results from two analyzers are different, analyzer can get an accordant result by calibration of slope
value and intercept value. Slope value is a positive number, which is less than 8 digits. Intercept value is a
real number, which is less than 8 digit.
“sample volume” work area: Sample volume includes normal volume, increased value and decreased value.
Left side of sample volume work area is for serum sample, right side is an optional area. Operator can select
sample type from pull down menu, as figure 9-5 shows:
Normal volume
In “ normal volume” work area operator can specify sample normal volume. This work area are divided
three functional key field: sample volume, diluted sample volume, diluent volume.
[ Normal/ sample volume] : The sample volume (2ul to 35ul) is absorbed from sample container( sample
cup or tube) . If predilute is not required, the total volume of sample and reagent should be greater than or
140
User Manual of CS-400 Auto-Chemistry Analyzer
equal to 150ul, smaller than or equal to 450ul. If diluent is required, the total volume of sample and
detergent should be greater than or equal to 105ul, the total volume of diluted sample and reagent should
be greater than 150ul, less than or equal to 450ul.
[ Normal/ diluted sample volume ] : If predilute is required, the parameter is used to set up diluted sample
volume which sucked from dilute cup and infuse to reaction cuvette that is used for analyze reagent. The
input value is within 2ul to 35ul. Input 0 to avoid predilute.
[Normal/diluent volume]: If predilute is required, the parameter is used for set up diluent volume for
dilute sample. The input value is within 2ul to 350ul. Input 0 to avoid predilute.
“Decreased volume” work area is used for specify sample volume when sample concentration exceed the
upper limit of reagent linearity range. (lower than normal volume). This area is divided into 3 functional
key field: sample volume, diluted sample volume, diluent volume. After test, instrument will automatically
change decreased test result into normal volume and display on result information.
[Decrease/ sample volume]: Select a sample volume (from 2ul to 35ul, and less than normal volume),
which is sucked from sample container (sample cup or tube). If predilute is not required, the total volume of
sample and detergent should be greater than or equal to 150ul, smaller than or equal to 450ul. If dilute is
required, the total volume of sample and detergent should be greater than or equal to 150ul, the total
volume of diluted sample and reagent should be larger than 150ul, less than or equal to 450ul.
[Decrease/diluted sample volume] : If predilute is required, the parameter is used for set up diluted sample
volume which is sucked from dilute cup and infuse to reaction cuvette that is used for analyze reagent. The
input value is within 2ul to 35ul, Input 0 to avoid predilute.
[ Decrease/diluent volume] : If predilute is required, the parameter is used for set up diluent volume for
dilute sample . The input value is within 2ul to 350ul, Input 0 to avoid predilute.
“Increased volume” work area is used to specify sample volume when sample concentration exceed the
lower limit of reagent linearity range. (more than normal volume). This area is divided into 3 functional key
field: sample volume, diluted sample volume, diluent volume.
[Increase/ sample volume] : Select the sample volume (2ul to 35ul,more than normal volume) ,which is
sucked from sample container ( sample cup or tube). If predilute is not required, the total volume of sample
and reagent should be greater than or equal to 150ul, smaller than or equal to 450ul. If predilute is required,
the total volume of sample and detergent should be larger than or equal to 105ul, the total volume of diluted
sample and reagent should be larger than 150 ul, smaller than or equal than 450ul. Input 0 to avoid predilute
[Increase/diluted sample volume] : If predilute is required, the parameter is used for set up diluted sample
volume which sucked from dilute cup and infuse to reaction cuvette that is used for analyze, The input
value is within 2ul to 35ul. Input 0 to avoid predilute.
[Increase/diluent volume] : If predilute is required, the parameter is used to set up diluent volume for
dilute sample . The input value is within 2ul to 350ul, Input 0 to avoid predilute.
“Reagent “ work area: Contents of reagent include reagent volume and reagent position. Reagent 1(R1) ,
141
User Manual of CS-400 Auto-Chemistry Analyzer
reagent 2 (R2), reagent 3 (R3), reagent 4 (R4) are reagent type . R1 and R4 are in reagent disk 1, R1 probe
absorb reagent. R2 and R3 are in reagent disk 2, R2 probe absorb reagent. As figure 9-5 shows:
Figure 9-5
Reagent volume: Reagent unit: ul. Reagent probe can exactly suck 20ul to 350ul solution. “0” represent no
reagent is added.
Position: Display reagent position in reagent disk. Register this parameter in “ reagent info.”
Second half item of two test analyze: In order to carry out two test analyze, select the name of second half
item in “second half item of two test analyze” pull down menu.
Prozone check: input the range value of prozone check, setup the upper limit and lower limit.
Absorbance limit: Input the range value of absorbance, setup the positive reaction and negative reaction.
142
User Manual of CS-400 Auto-Chemistry Analyzer
Figure 9-6
(2) In terms of reagent instruction, Check calibration type, calibration point, span point, and weight
coefficient etc. parameter.
(4) Input the parameter of calibration check. For detailed operation, please refer to chapter “ 2.3.1 ”
(5) For automatic calibration, input the overtime of automatic calibration according to relevant calibration
type. If automatic calibration time comes when stand-by, instrument will automatically carry out
calibration before test next time. In same item, if automatic calibration conflict with manual calibration,
carry out manual calibration only. Click “0” in time position if automatic calibration to avoid automatic
calibration.
Calibration point: Input the quantity of Calibrator in functional field, (among 1-6).
Span point: Input the relevant span point in functional field (among 2-6).
Click “ ” key in the “ ” form to setup the reference value range and
linearity range. Click “save ” button to save the parameter.
143
User Manual of CS-400 Auto-Chemistry Analyzer
Figure 9-7
Special reference range: If patient age, and patient gender are different, the reference value range are
different too. Click “ ” key. For example, the special reference range of UA:
0.1~0.34 for male aged 0~15, female 0.12~0.33 aged the same; 0.21~0.43 for male aged 14~50, 0.15~0.36
for female aged the same; 0.28~0.50 for male aged more than 50, 0.21~0.43 for female aged the same. As
figure 9-8 shows:
Figure 9-8
Default reference range: select default reference value range “ ”when the reference value
144
User Manual of CS-400 Auto-Chemistry Analyzer
range are the same despite the difference patient age and gender.
For example, the default reference value range of urine amylase is 0~640U/L , as figure 9-9 shows.
Figure 9-9
Linearity range: Input the upper limit and lower limit of reagent linearity in functional field. Alarm is issued
when test result exceed the range.
Note: The chemistry parameter in the manual is taken as an example, not real test parameter. Operator should set
parameter according to reagent manual .
Click key “ system setup” form, set up item combination, as figure 9-10 shows:
Figure 9-10
(1) Input the number of combination item in “ number” functional field, the number can’t be repeated, or it cannot
be saved.
(2) Input the name of item combination in “ item combination name” functional field, Character and number are
all allowed, but can’t be repeated. Or it cannot be saved.
145
User Manual of CS-400 Auto-Chemistry Analyzer
(3) Select item of item combination, click check box in front of item name, indicate that is selected.
Click check box again, and indicate that it is cancel.
(4) Click “ ” key , Combination item number, name will be displayed in the right side of functional
field. Click the number or name, all items will be automatically showed in functional field.
(5) Select the number or name of desired combination item, click “ ” key to delete the
combination item.
Test Calculation is on the base of two or more item test results, use special calculation method to get a new item,
such as A/G..
Figure 9-11
(3) Select the unit of new calculated item in “unit” pull down menu. Select the decimal digits of new calculated
item in “ decimal digits” pull down menu.
146
User Manual of CS-400 Auto-Chemistry Analyzer
(4) Select the reference value range of new calculated item in “ reference value range” pull down menu.
(5)Edit the calculation formula the edited information will be showed in “ calculation formula” functional field,
click “add” key to finish edition work.
(6) In order to delete calculation item, click the desired item, and then click “delete ”key.
Set up item name, unit, decimal digit, reference value range according to the above procedure, select 【ALB】
in “test name” pull down menu, select【/】 、【(】in “sign” pull down menu, select 【TP】 in “item name” pull
down menu, select 【-】 in “sign” pull down menu. The edit of formula can be showed in “calculation formula”.
If there are some number in formula, select any number among 0、1、2、3、4、5、6、7、8、9 in “number” pull
down menu. After input finish, click key, the formula will be showed at the right side of
functional field.
Figure 9-12
Click key beside the calculation formula functional field to delete the wrong content.
Cross contamination obviation is a function to avoid cross contamination among analyzing items. The degree of
cross contamination is different due to different reagent ingredient. For avoiding cross contamination among
reagents, we strongly suggest separate the items with cross contamination and the items without cross
contamination. If all item cannot be separated, automatic washing function can be added before tested item in
order decrease the cross contamination in maximum extent. But the cross contamination function may decrease
the test velocity.
Cross contamination include: reagent probe, reaction cuvette, sample probe. Detergent is located at position 45
in reagent disk 1 and reagent disk 2.
147
User Manual of CS-400 Auto-Chemistry Analyzer
(1) Select reagent probe R1 or R2 in reagent work field. indicate R1 has been selected.
Input the detergent volume (ul).
(2) Select the reagent type in “item” pull down menu in “from reagent” work area.
(3) Select reagent type in “item” pull down menu in “to reagent ” work area
(4) Click “ ” key, the set information will be displayed in left side functional field.
148
User Manual of CS-400 Auto-Chemistry Analyzer
Figure 9-14
(1) Select item that needs to be set in “item name” pull down menu.
(3) Input R2 detergent volume (ul) in “R2 detergent volume” work area.
(4) Click “ ” key, the set information will be displayed in left side of functional field.
(5) Click “ ” key, information will be deleted. And click “ close” key to exit the current menu.
Click “ ” key to set “sample cross contamination obviation”, as figure 9-15 shows:
149
User Manual of CS-400 Auto-Chemistry Analyzer
Figure 9-15
(1) Select the items to be set in “ item name ” pull -down menu.
(3) Click “ add” key, the set information will be displayed in the left side of functional field.
(4) Click “delete” key , information will be deleted. Click “close” key to exit the current menu.
Click key to set printout info. and format. As figure 9-16 shows:
150
User Manual of CS-400 Auto-Chemistry Analyzer
Figure 9-16
(2) If endnotes of report is used, please sign “√” in “ report endnotes” functional field. Input the first and second
line of report content. Endnotes may not be inputted if no need.
(3) Please sign “√” in “ automatic add calculated item” functional field to add the item name into print list.
small to large. If select 0, instrument will print according to the item number . Click “ ” to save
the setup parameter.
151
User Manual of CS-400 Auto-Chemistry Analyzer
(a) Click “ ” to select report template in pull down menu, preview the printout report
template at the right side of form, adjust the preview scale in “ ” functional field.
Figure 9-17
Figure 9-18
(d) Click “ ” key to input new template name, As figure 9-18 show.
(a) Click “ ” key to set printout report option . As figure 9-19 show.
Figure 9-19
(b) Select the contents of report and sign “√” in functional field. Select “print all item”, and all reminding, item
code, check result , item name, unit , reference value will be printed.
Click “ ”key in “report format” work area. And select a report format.
Single-click “ ” key in “ system setup” menu to setup ISE parameter. As figure 9-20 shows.
153
User Manual of CS-400 Auto-Chemistry Analyzer
Figure 9-20
(1) Select the items name and sample type, decimal digits, item unit, QC interval, etc.
(2) Input the concentration of Calibrator in “calibration parameter” work area and select the relevant calibration
position. Calibrator 1 is ISE low concentration Calibrator. Calibrator 2 is ISE high concentration Calibrator.
Calibrator 3 is compensation liquid.
(3) Input the remaining volume of Reference Liquid, Internal Standard Liquid and Diluent in reagent remaining
volume work area in “ml”, click “save ” key. Instrument will automatically subtract the wastage and display the
remaining volume in work area.
Single-click “ ” key to set the sample barcode, reagent barcode, sample alarm, reagent alarm, ISE,
time wakening etc. function. Single-click “ ” key to save the parameter. Click
Figure 9-21
Check barcode device, if barcode device is connected well, and use would like to execute barcode device
checkup of sample disk, R1 reagent disk, R2 reagent disk, select “√”in functional block, if barcode device is
not connected, use should not select “√”,or it alarm may occurs. Scanning barcode before sample analyze, in
order to checkup sample barcode scanning function, select “√”in functional block.
Scan reagent open area barcode: Reagent close area must execute reagent barcode scanning, in order to scan
reagent open area barcode, select “√”in functional block.
Execute ISE: Sign “√” in ISE setup if any test required ISE test.
If sign “√” in functional block, the instrument will automatically be awakened in specified time. Input the
desired date in awaking date list, and the wakening time in functional block ( such as : ** hour ** minute)
Reagent remaining times: input reagent remaining test times. When instrument detect the remaining test times is
less than set value, an alarm will be issued.
Reagent remaining volume: input reagent remaining volume. When instrument detect the remaining volume is
less than set value, an alarm will be issued.
Reference Liquid remaining volume: Input the Reference Liquid volume for alarm in “ml”
Internal Standard Liquid remaining volume: Input the Internal Standard Liquid volume for alarm in “ml”.
Diluent remaining volume: Input the Diluent volume for alarm in “ml”.
“Manual item ” means the test is not processed in the instrument, but to add the result by handwork in “test
result” block.
Single-click “manual item setup” key to setup the item short form, item name, item unit, reference range.
Single-click “add” to save parameter after setup as figure 9-22 shows.
Figure 9-22
(a) Input the manual item short form in the “item short form” functional block.
(b) Input the manual item name in the “item name” functional block.
(d) Input the manual item reference value range in the “reference range” functional block.
(e) In order to delete the set information, single-click the set information in work area, single-click “delete” key.
For LIS communication setup function, please refer to LIS function user manual
156
User Manual of CS-400 Auto-Chemistry Analyzer
Figure 9-23
157
User Manual of CS-400 Auto-Chemistry Analyzer
The user with administrator permission can add, delete or modify the user information.
Single-click the “ ” key in main functional filed, then click “ ” key to set the operator ID,
name, password, reconfirm password ( two times inputted password should be the same) , mnemonics, etc. As
figure 10-1 shows:
Figure 10-1
Single-click “ ” key, the relevant info. will be displayed in the form. Click “ ”
key to eliminate the user information.
Administrator: User with administrator permission can set, delete, check , browse and test all functions
Inquiry: some function windows are available to user, setup and test unavailable.
Operation: User with operator permission can set, delete, check, browse and test all functions except for user
information.
158
User Manual of CS-400 Auto-Chemistry Analyzer
Single-click the “ ” key in “ ” menu, to set the delivery dept. delivery doctor, sample
type, patient type and item unit, as figure 10-2 shows :
Figure 10-2
These items will be displayed automatically after setup in “patient info.” form corresponding pull-down block.
Dept. No.: Input the dept. No. The information will be displayed in pull-down menu of “ sample register ”
menu.
Delivery dept.: Input the name of delivery dept., the information will be displayed in the relevant pull-down
list of “ sample register” menu.
Mnemonics: Help the user input the information quickly. For example: the mnemonics of department
numbers can be set as KSBH, then the contents of patient type option can be replaced by inputting KSBH,
click “enter” key, the department number will be automatically inputted.
159
User Manual of CS-400 Auto-Chemistry Analyzer
Single-click “ ” key in functional field to set “Doctor No.” “ Doctor Name” etc. information. As figure
10-3 shows:
Figure 10-3
Click “ ” key, all the information will add to delivery doctor menu. Select the info. bar to be
deleted, then click “ delete” key , the info. bar will be eliminated
Single-click “ ” key in “ ” form, then set the serial number, sample type,
mnemonics etc information . As figure 10-4 shows:
160
User Manual of CS-400 Auto-Chemistry Analyzer
Figure 10-4
Select the info. bar to be deleted, then click “ ” key , the info. bar will be eliminated.
Single-click “ ” key in work area to set serial number, patient type, mnemonics etc. as the figure
10-5 shows:
161
User Manual of CS-400 Auto-Chemistry Analyzer
Figure 10-5
Select the info. bar to be deleted, then click “ ” key , the info. bar will be eliminated.
Single-click “ ” key in work area to set the clinic diagnosis information. As the figure 10-6
shows:
162
User Manual of CS-400 Auto-Chemistry Analyzer
Figure 10-6
Select the info. bar to be deleted, then click“ ” key , the info. bar will be eliminated
Single-click “ ” key in working area to set the report remark information, as the figure 10-7 shows:
163
User Manual of CS-400 Auto-Chemistry Analyzer
Figure 10-7
Input No. remark, mnemonics in functional block, click “add” to register remark.
Select the info. bar to be deleted, then click “ ” key , the info. bar will be eliminated.
164
User Manual of CS-400 Auto-Chemistry Analyzer
Figure 10-8
Click “ ” key, the inputted information can be showed in test unit form. Select the info. bar to be
Workload statistics function is used for checking the workload of delivery department, deliver doctor and check
doctor. According to the desired time slice select the statistic contents, click “ ” key to
complete statistic, result is showed as statistic chart, as figure 10-9 shows:
165
User Manual of CS-400 Auto-Chemistry Analyzer
Figure 10-9
Click “print” key to preview and print the statistics chart, as figure 10-10 shows:
Figure 10-10
166
User Manual of CS-400 Auto-Chemistry Analyzer
Figure 10-11
Database backup: Before database backup, user could select the save path, or the database will be saved to the
default software installation folder. The file name of database is the current date plus the current time, the
postfix is *.back. Periodically backup database can avoid the data lost. When the path of backup is selected,
click “ backup” key. The form will display the status of database backup and provide backup finish hint.
Database recover: When the software can’t be used, the database backup file can recover the former data.
Select the save path of backup file, then select the backup file according to the date and time, click “ recover”
key. The form will display the information of recovered database. If the path and file of database backup are
not selected, a hint information will pop up.
167
User Manual of CS-400 Auto-Chemistry Analyzer
Figure 10-12
System log realizes these function to check system operation, including user login, operation log, maintenance
log and alarm log. Select one log type in “type” work area, select the start date and finish date in “date” work
area, click “ “ key, all the relevant log information will be listed.
168
User Manual of CS-400 Auto-Chemistry Analyzer
Any questions arise when operate instrument, the user could click “ ” to seek help. As figure 11-1 shows:
Figure 11-1
(1) Single-click the “catalogue” key, check the desired menu in the list. Single-click the “ back” key to return
to the last menu. If instruction contents are not displayed completely, pull the scroll bar to browse the rest.
169
User Manual of CS-400 Auto-Chemistry Analyzer
To ensure the accuracy and precision of the CS-400, the user should operate strictly according to the CS-400
User Manual, and a regular maintenance is also a necessity. This is the only way to make sure a long useful life
and a reliable result, which is provided by instrument.
Please prepare the following items before carry out system maintenance
Stainless steel ware (with diameter 03mm and 0.5mm)…..(for cleaning sample probe and reagents probe)
Buckets (two)…………………………………(for drainage the cold water and discharge the reagent waste)
12.1.3 Detergent
The detergent is used for cleaning all parts of instrument. All kinds of detergents could be purchased from Dirui
company. Other brand detergents may cause the uncleanness of cuvette, reagent probe, sample probe, stirring
rod, pipe line, and finally result a cross contamination. Our company is not responsible for the inaccuracy,
which is conducted by the other kind of detergent.
may grow in the incubation bath. Owing to there is no conductance, the liquid sensor can not detect liquid level
normally. The reagent probe may automatically suck detergent on 45 position for reagent probe cleaning. Wipe
all parts of instrument or immersion cuvette with 2% CS anti-bacterial phosphate-free detergent.
CS-alkaline detergent: The CS-alkaline detergent in the detergent bottle which located in the front of
instrument is used for cleaning cuvette; Detergent on W1 position in sample disk is used for cleaning sample
probe.
CS-acidic detergent: The CS-acidic detergent on W3 position is used for the cleaning which could avoid
sample probe cross contamination.
CS-ISE detergent: The CS-ISE detergent on W2 position is used for cleaning for sample probe, dilution bath,
ISE pipe line after ISE test.
CS-solenoid valve cleaning liquid: Clean the concentrated waste liquid pipeline.
Click “ ”button of the main function menu to start the instrument maintenance. Select maintenance
information in the list by mouse, or remove “↑”“↓”key on keyboard, click “ ” button, start
maintenance. Instrument will carry out reset operation first among all maintenance operation.
Some maintenance item allow stop in the midway, click “ ” to finish the maintenance operation.
For those not allow stop in the midway, take other operation after maintenance finish. In order to exit the system
12.2.1 Reset
Select “Instrument Reset” in “maintenance item list” work area, then click “Execute”. The instrument will
automatically return to the initial position. There will be an “alarm” hint if mistakes happen. Emergency stop is
not allowed while resetting. Take other operation when computer stand-by. Strongly suggest user execute reset
operation after emergency stop or after adjustment of reagent probe, sample probe and stirring rod.
time. (As figure 12-1 shows). Check the menu to choose printout the result or not. The absorbency value should
be ≤ 18000. Click “end maintenance” to complete the light quantity checkup operation.
Normally, carry out the light quantity checkup once a month. Carry out the light quantity checkup after replace
bulb, then proceed the test after the absorbency value qualified.
Figure 12-1
The cell blank check value will be showed on the system maintenance menu (as figure 12-2 shows). Click
“print” button to print out the cell blank check value. Click “end maintenance” key to complete the cell blank
check operation.
Normally, carry out the cell blank check once a week is suggested. Carry out cell blank check after replace the
cuvette. Then proceed the test after the cell blank check value qualified. Do not proceed sample test if the cell
blank check value is abnormal, it may influence the accuracy of the test result.
172
User Manual of CS-400 Auto-Chemistry Analyzer
Figure 12-2
“340-750”tandem: display the cell blank value of 120 cuvettes corresponding to the different wavelength of
340、380、405、450、480、505、546、570, 600、660、700、750(nm).
“1”(No.1 cuvette ) rank: display the cell blank value of 12 type different wavelength of No.1 cuvette. A
<18000 cell blank value is considered as a qualified one.
“2”(No.2 cuvette) rank: display the difference between two cuvette: the difference cell blank value between
No.1 cuvette and No.2 cuvette, a ±800 difference value is considered as a qualified one.
Carry out Air exhaustion function when replace syringe or replace the connection pipeline of syringe.
Carry out this function when detergent bottle with CS-alkaline-detergent discharged, and there is air in the
173
User Manual of CS-400 Auto-Chemistry Analyzer
connection pipeline.
Rinsing cuvette once a week in order to avoid the dirt in cuvette influence the test result. Carry out Rinsing
cuvette function when cell blank value abnormal, if cell blank value is still not qualified after Rinsing, replace
the cuvette.
We suggest user execute Rinsing ISE once a day after analyze finish. ISE calibration must be execute before
next test.
Note : If ISE device is not connected, or connected but not set, all ISE maintenance cannot be used, the hint bar
may hint: please check if ISE device can be used.
We suggest user execute Rinsing ISE + Reaction cuvette once a week. If rinsing all is executed, rinsing ISE +
Reaction cuvette should not be carried out again.
Carry out incubation bath water replace function when constant temperature water is contaminated. Incubation
bath will automatically replace water when instrument start up. If the instrument is continuously used for more
than 24 hour, alarm issued to hint to carry out the incubation bath water replace function.
For detail introduction of sample probe vertical checkup, please refer to “12.4.1(4)”.
174
User Manual of CS-400 Auto-Chemistry Analyzer
Figure 12-3
For detail introduction of sample probe horizontal checkup, please refer to “12.4.1(4)”.
This operation is taken when carry out sample probe position adjustment or sample probe position checkup.
For detail introduction of reagent probe vertical checkup, please refer to “12.4.1(4)”.
175
User Manual of CS-400 Auto-Chemistry Analyzer
For detail introduction of reagent probe horizontal checkup, please refer to “12.4.1(4)”.
This operation is taken when carry out reagent probe position adjustment or reagent probe position checkup.
This function is carry out when adjust the stirring mechanism position (at the side of cuvette or on the top of
rinsing bath).
Figure 12-4
176
User Manual of CS-400 Auto-Chemistry Analyzer
result work area, click “end maintenance ”key to end the operation.
Execute ISE checkup when exchange electrode or after issue ISE alarm.
Note: The difference of two test value of same electrode should be less than 0.2.
177
User Manual of CS-400 Auto-Chemistry Analyzer
Figure12-6
Single-click “ ” in main keypad and select “ISE pipeline exhaust”, and then execute; the instrument
will carry out pipeline exhaust automatically after implement this function when bubbles exist in ISE pipeline
and ISE injection pump.
(a) Replace the CS-phosphor-free anti-bacterial cleaning liquid at the positions 45th of R1 reagent and R2
reagent disks with CS-solenoid valve cleaning liquid.
(b) Single-click “ ” in main keypad and select “Automatically rinse concentrated waste liquid
pipeline”, and then single-click “Execute”; the instrument will carry out pipeline rinsing automatically as figure
12-5 shows:
178
User Manual of CS-400 Auto-Chemistry Analyzer
Figure 12-5
(a) Clamp the concentrated waste liquid outlet at the right lower side of instrument back cover board with a
hemostatic clamp (about 50mm away from the back cover board).
(b) Unplug the nozzle 1 of rinsing mechanism first as figure 12-6 shows:
Nozzle 1
Figure 12-6
(c) Take a 70ml reagent bottle (containing CS- solenoid valve cleaning liquid and make sure the distance
179
User Manual of CS-400 Auto-Chemistry Analyzer
between liquid level and bottle lip is approximately 20mm), and insert nozzle 1 into the bottle, and then
single-click “ ” key, then select “Manually rinse concentrated waste liquid pipeline”, afterwards, the
instrument will execute rinsing automatically after click “Execute” as figure 12-7 shows:
(d) After about 5~6 minutes,remove the clamp on the outlet of concentrated liquid, and waste liquid will be
discharged automatically after rinsing.
(e) Replace the CS-reagent solenoid valve cleaning liquid in the 70ml bottle with purified water, and repeat the
operation carried out above.
(○: denotes periodic cleaning and check up and ●: denotes periodic replacement part.)
180
User Manual of CS-400 Auto-Chemistry Analyzer
Frequency
Quantity Yearly
No Part Refer to
Per use Quantity Every 3 Every 6
Daily Required Weekly Monthly Yearly
month month
1 Sample cup ● ——
CS-anti-bacterial
12 phosphor free ● 12.1.3
CS-alkaline detergent
20 Waste discharge ○
21 Detergent bottle ○ 12.4.2
(a) Statistics in above table is a maximum statistics, if cell blank value is still qualified after 1 month,
181
User Manual of CS-400 Auto-Chemistry Analyzer
continue to use it, if cell blank value is abnormal after Rinsing, replace it.
(b) Replace the lamp as soon as the photometer check value (340nm wavelength) exceeds 18,000 hours, or
the lamp use time more than 750 hours
(c) Instrument could use laser, ink mist, stylus printer, select accessory according to different printer.
(d) A cell blank alarm may occur if cell blank is not executed every week
(e) If the purified water has exceeded 1us/cm, consult the water supplier
Quantity to
item Model name Description
Stock
1 Halogen lamp (light source lamp) 12V 20W 2
2 Reaction cuvette set(20pcs*6set) 72set
3 3603 ethylene tube 1/8 *1/4 inch 5m
4 3603 ethylene tube 1/16 *1/8 inch 5m
5 Teflon FEP rigid tube 1.5mm * 2.5mm 5m
6 Teflon FEP rigid tube 0.03 *1/16 inch 3m
7 Silica gel tube 8mm*14mm 10m
Proper
8 Ribbon cassette For printer
amount
Proper
9 Printing paper For printer
amount
For supply water
10 Water supply filter 1
connection
11 Sample probe For sample 1
12 Reagent probe For sampling 2
13 Stirring rod For stirring 2
14 Nozzle 1,2 of rinsing mechanism For cleaning 2
R1 reagent probe
15 R1 reagent probe syringe 1
sampling
R2 reagent probe
16 R2 reagent probe syringe 1
sampling
Sample probe
17 Sample probe syringe 1
sampling
Table 12-2
182
User Manual of CS-400 Auto-Chemistry Analyzer
! Warning:
△
◆ Do not spill water, reagent or detergent over the instrument or mechanical /electrical parts in order to
avoid the damage.
◆ Do not touch the suction mechanism, sampling mechanism, stirring mechanism, reaction cuvette rinsing
mechanism during operation, or there will be a risk of infection or injury.
◆ Protective measures should be taken to the operator, such as with protective gloves, protective glasses
and work uniform during operation. Otherwise, there maybe an infection when touch the contaminated
areas and contaminated liquid. Corrosive liquids may cause a skin injury. If the contaminated liquid or
corrosive liquids accidentally touched the body, please rinse with water immediately, and seek medical
advise.
(a) Remove the sample disc cover, place 1ml CS-alkaline detergent (as figure 12-7 shows) on W1 position of
sample disc. Placed 1ml CS-acidity detergent on W3 position to avoid cross-contamination. Detergent
quantity should be taken according to the frequency of rinsing.
Figure 12-7
(b) Set 70ml CS-anti bacterial phosphor-free detergent on 45 position in the reagent disc(R1,R2).
When sample and reagent probe finish sampling respectively, they may automatically assimilate CS-anti
bacterial phosphor-free detergent or CS alkaline detergent to process the rinsing.
Note: Place the detergent on a designated position When doing cuvette, sample probe avoiding
cross-contamination rinsing.
183
User Manual of CS-400 Auto-Chemistry Analyzer
Figure 12-8
(2) Cleaning outside of probe tip
(b) Remove the sample or reagent disk cover, and move the probe arm to the top of disk by hand.
Figure 12-9
(c) Using gauze moistened with alcohol solution, clean the outside of probe.
Figure 12-10
184
User Manual of CS-400 Auto-Chemistry Analyzer
(d) Turn on the POWER switch of analyzing unit. Each probe will then return to its reset (home) position
automatically.
Note: As alcohol is flammable, Pay attention to it and do not place large amount alcohol in the vicinity of the
instrument.
(b) Pinch the jaw of probe arm and remove the cover, and loosen the connector as figure 12-11 shows:
jaw
Figure 12-11
Tube
Figure 12-12
(d) Penetrate through the needle from the top with stainless steel wire to clean up, as Figure 12-13 shows:
185
User Manual of CS-400 Auto-Chemistry Analyzer
Figure 12-13
(e) Let injector assimilate 10 ml pure water after cleaning the probe with stainless steel, connect both
interfaces of probe and injector, push injector plunger until the 10 ml water has been discharged completely as
figure 12-14 shows.
Figure 12-14
Note: When cleaning the clogged sample probe, use a stainless steel wire having outside diameter of 0.3 mm.
When cleaning the clogged reagent probe, use a stainless steel wire having outside diameter of 0.5 mm.
(c) Click “sample probe horizontal checkup”, “sample probe vertical checkup ”, “reagent sample horizontal
checkup” and “reagent sample vertical checkup” in the “maintenance item”.
Note: During adjusting sample probe, make sure to implement the horizontal checkup firstly, then implement
the vertical checkup. The tip of sample probe should be at the center of the reaction cuvette.
(d)During sample probe horizontal checkup, the probe stops above the reaction cuvette. At this step, adjust the
reagent probe so that its tip will be aligned with the center of reaction cuvette.
186
User Manual of CS-400 Auto-Chemistry Analyzer
Reaction
cuvette
Probe tip
Figure 12-15
Please contact maintenance man if the probe tip is not aligned with the center of reaction cuvette.
(e) During reagent probe horizontal checkup, the probe stops above the reaction cuvette. At this step adjust
the reagent probe so that its tip will be aligned with the center of the reaction cuvette.
(f) To execute “stirring mechanism horizontal checkup” , when the stirring rod stops above the reaction
cuvette, at this step, check muddler probe tip is aligned with the center of reaction cuvette.
From overlooking angle to view the sample probe and reagent probe and stirring rod, the relevant position of
the reaction cuvette is as following:
Reagent probe 1
stirring rod
Reaction
cuvette
Reagent probe 2
Figure 12-16
Please contact maintenance man if the probe tip and stirring rod probe tip are not aligned with the center of
reaction cuvette.
Sample probe → above the reaction cuvette → Rinsing bath (pause)→ outer track 1 position on sample
disk(descend) → above the reaction cuvette → Rinsing bath (pause) → Middle track W1 position on sample
disk (descend) → Above the reaction cuvette → Rinsing bath (pause) → Inner track C8 position on sample
disk (descend) → Above the reaction cuvette → Rinsing bath (pause) → ISE diluent bath →Above the
187
User Manual of CS-400 Auto-Chemistry Analyzer
Note: “ISE diluent bath →Above the reaction cuvette” action only be taken where ISE function is provided.
Home position of the sample probe is reset point.
Reagent probe (R1,R2) → Above the reaction cuvette → Rinsing bath → Reagent bottle → Rinsing bath
(pause) →Above the reaction cuvette → repeated the whole process.
*Sample probe R1 and R2 have the same movement according to the sample probe movement.
Stirring rod → Above the reaction cuvette → Rinsing bath →Above the reaction cuvette → repeated the
whole process.
R1,R2 have the same movement according to the Stirring rod movement
Select “sample probe vertical check”, in “maintenance” key, and then click “Execute ”key, thus, the sample
probe descends until the bottom of sample cup is detected.
(h) During reagent probe vertical checkup, place an empty and dry reagent bottle at position 1 on reagent disk
1 and reagent disk 2, single-click “next”, the reagent probe will descend from top to the bottom to memorize
the descent distance as the reference value of the remaining reagent.
(a) If the Rinsing bath is contaminated, use a tube brush cleaning it with 2% CS-anti-bacterial phosphor-free
detergent as figure 12-17 shows:
Rinsing bath
Figure 12-17
188
User Manual of CS-400 Auto-Chemistry Analyzer
Figure 12-18
After cleaning, contamination can be eliminated and bacterial can be restrained. Cleaning can be taken every
month. If the instrument is contaminated while the operation, please cleaning it in time.
(b) Select “cell blank” function in the “Maintenance” menu, click “Execute ”, the instrument will carry out
the cell blank check automatically.
(c) The cell blank value of the first reaction cuvette should ≤18000, and the difference between 2-120
cuvette should within the range of ±800.
The cell blank value is the absorbency of each reaction cuvette before adjust the photometer, data value
represent the absorbency of each reaction cuvette from the second reaction cuvette, data value represent the
margin between each reaction coveter ( from the second reaction cuvette) and the first reaction cuvette.
(d) If the cell blank value is not within a rage of ±800, the relevant reaction cuvette is contaminated. Clean the
reaction cuvette.
Note: The cell blank value can be displayed or printed(previous value will be replaced when the second test
finished)
If the cell blank value is not within a rage of ±800, the relevant reaction cuvette is contaminated. Clean the
reaction cuvette. If the usage of the reaction cuvette has exceed the time limit for replacement, please place a
new one.
189
User Manual of CS-400 Auto-Chemistry Analyzer
(a) Place the CS-anti-bacterial phosphor free detergent (70ml/bottle)on position 45 in the R1,R2 reagent disk
(b) Click “Rinsing reaction cuvette ”button in “maintenance item list” work area, then click “Execute ”key
(c) After cleaning reaction cuvette, carry out the cell blank check again. If the cell blank value exceeds the
±800, please place a new one.
Note: In order to avoid uncleanness cleaning after long time use, immerse the reaction cuvette in 2% CS
anti-bacterial phosphor-free detergent for more than 8 hours every week. Wash the immersed reaction cuvette
with water, and then wash the reaction cuvette with purified water, and then mount the cuvette on the reaction
disk, carry out cell blank check, test after cell blank check value is qualified.
If the blank cell value is not qualified after cleaning or if the cuvette is used for over two month, replace it
with a new one.
Note: A new reaction cuvette should be immersed in 2% CS-anti-bacterial phosphor detergent for 8 hours, clean
the surface of the cuvette by purified water, and then Mount the cuvette on the reaction disk, carry out testing.
The abnormal cell blank value will influence the accuracy and repetition of the testing result.
(b) Remove the setscrew of rinsing mechanism when wear protective gloves.
Figure 12-19
190
User Manual of CS-400 Auto-Chemistry Analyzer
Figure 12-20
Figure 12-21
(e) Mount new reaction cuvette on the reaction disk. 6 sets reaction cuvette should be placed at the same time
in counter direction
(g) Select “cell blank test” in “maintenance” menu. Make sure carry out the cell blank check after replace
reaction cup. Testing can be carried out after the cell blank value is qualified.
Note 1: A once-employed reaction cuvette might be contaminated heavily if allowed to dry. Immerse it in
purified water to store it. If the instrument will be left unemployed for 3 days or more, remove the reaction
cuvettes from the reaction disk and keep them immersed in purified water.
Note 2: Never use any organic solvent (benzene, alcohol) for washing the cuvettes.
(4) Cleaning Incubation Bath and the drain filter of the Incubation Bath.
A clog on drain filter or Incubation contamination will cause the inaccuracy of the testing data. Thus, Clean the
incubation bath periodically. (once a month)
(a) Select “Rinsing incubation bath ” function in the “maintenance item list ”work area, after closing the light
191
User Manual of CS-400 Auto-Chemistry Analyzer
source lamp,the constant temperature water will be drained from the Incubation bath. Turn off the power
switch.
(b) Loosen the rinsing mechanism retaining screw, and remove the rinsing nozzle head, as figure 12-22
shows:
Figure 12-22
(c) Take out the 6 set reaction cuvette into purified water, Loose the retaining screw of the reaction disk, take
out the reaction disk. Place the reaction cuvette in a location free from dust.
Note: If take out the reaction disk and cuvette at the same time, water droplet attached on the outside of reaction
cuvette will drop into the instrument so that cause the instrument malfunction. Therefore take over the reaction
cuvette firstly and then take over the reaction disk secondly.
(d) Using washed gauze moisturized with water, clean the reaction bath and photometric window as figure
12-23 shows. Be careful not to flaw them or attached scrape.
Photometry
window
Figure 12-23
(e) Take out the drain filter of the incubation bath as figure 12-24 shows, cleaning by water, and then return it
192
User Manual of CS-400 Auto-Chemistry Analyzer
in place.
Filter
Figure 12-24
(f) Click “next” key in the “maintenance” form, infuse pure water into the incubation bath, turn light and
circulation syringe power on.
(g) Mount the reaction disk and reaction cuvette after Incubation bath rinsing.
(h) Return the rinsing mechanism nozzle head in place and secure it.
(i) Select “cell blank check ” function in “maintenance item list” work area. Test can be carried out after the
cell blank check value is qualified.
Take out the sensor out of the incubation bath, wipe the outside of sensor with gauze moisturized with 2%
detergent. In order to prevent water from being contaminated by sensor probe, rinsing once a month is advised.
As the CS-alkaline detergent in the detergent bottle is added timely and by this way maintain the day-to-day use.
After a period of time, there will be dust or white substance separate out Thus, clean it up monthly.
(a) Take over the detergent bottle from the instrument, loose the detergent bottle cap which is located on the
upper sprue. Insert the accessory tool from upper sprue to bottom outfall, then tighten the screw to prevent
leakage when detach the interface of the bottle bottom. Clean the bottle when there is only a few detergent
existed.
193
User Manual of CS-400 Auto-Chemistry Analyzer
Bottle lid
Figure 12-25
(b) Detach the bottom interface after rag prepared, Take care that the pipe port position must be lower than
bottle.
Pipe joint
Figure 12-26
(c) Take out the detergent bottle, clean the outside and inside of the bottle, then wash it by water, At last wash
it by purified water. Detach the tool and wash away the attached detergent.
(d) Dry the inside water droplet, wipe the outside water droplet, return the detergent bottle in place, add
CS-alkaline detergent.
(e) Select “maintenance item list” work area, implement " pipeline for pouring detergent exhaust " function.
① Select “Light quantity check” function in “maintenance” form. The photometer light quantity is checked
and the results will be displayed by AD value or printed out.
194
User Manual of CS-400 Auto-Chemistry Analyzer
Method 1:
Figure 12-27
Note: Do not touch the glass surface of the new lamp. Otherwise the lamp characteristics may change. If the
glass surface is found stained with fingerprints, etc, wipe them off with gauze wetted with alcohol.
(b) Select “Rinsing incubation bath ” function, click “Execute” key, the constant water in incubation bath is
automatically discharged.
(d) Loosen the retaining button of the reaction disk, take out the reaction disk. Place the reaction disk in a
position free from dust.
Note :reaction disk should be placed in dry and clean area, prevent the droplet on reaction cuvette dropping
on the inside of the instrument.
(e) Loosen two retaining terminals of lamp lead wire and disconnect the lamp lead wires.
Retaining
terminal
Figure 12-28
195
User Manual of CS-400 Auto-Chemistry Analyzer
(f) Loosen the retaining screw of the retaining light seat,take out the lamp as figure 12-29 show.
Retaining
screw
Figure 12-29
(g)In the reverse order of step 6-8, mount the new light source lamp. Do not distort the rubber tube used for
cooling light room. Make sure that the lead wire of the lamp not in loose status.
(h) Mount the reaction disk and cuvette and the Rinsing mechanism, turn the power on, execute “next” in
“maintenance”, infuse pure water into reaction cuvette. Execute “light volume check” function. While light
volume value is qualified, testing can be proceeded.
Method 2:
(d)Take over a set reaction cuvette instead of Rinsing mechanism. Keep the reaction cuvette clean. Turning
the reaction disk by holding the retaining button of the reaction disk, make the reaction disk, which is
detached from reaction cuvette turning underneath the Rinsing mechanism. Loosen the retaining button and
take over the reaction disk.
(a) Loosen the rinsing mechanism retaining screw by turning it counterclockwise, and remove the rinsing
nozzle head.
(b) Wipe the outside of the nozzle with gauze wetted in 2% anti-bacterial phosphor-free detergent.
196
User Manual of CS-400 Auto-Chemistry Analyzer
Nozzle
Wipe
Figure 12-30
(c) Insert a 0.5mm stainless steel wire into the nozzle from its bottom end for cleaning as figure 12-31 shows.
Figure 12-31
Note: If the nozzle chip is contaminated heavily or worn excessively, replace it with a new one.
(d) Take out wiping block lightly, rinse it with pure water after rinsing it with 2% CS-anti-bacterial
phosphor-free detergent.
(e) Return the wiping block (keep lower horizontal level of wiping block the same level with the rinsing
probe) as figure 12-32:
197
User Manual of CS-400 Auto-Chemistry Analyzer
Rinsing nozzle
Press
into
Nozzle
Reaction
cuvette
Figure 12-32
(g) Carry out the Mechanism Check 10 times, wiping block can’t touch the reaction cuvette and the rinsing
water is filled up to the upper limit level in the reaction cuvette and it does not overflow from the incubation
bath.
A contaminated stirring rod would cause a cross contamination so that influence the accuracy and precision of
the testing result. Clean the Stirring Rod periodically.
Wipe the Stirring rod using gauze moistened with 2% anti-bacterial phosphor-free detergent, then wash away
the detergent on the surface of the Stirring rod by gauze moistened with purified water.
Figure 12-33
198
User Manual of CS-400 Auto-Chemistry Analyzer
(b) Loosen the two setscrews one round as figure 12-34 shows.
Figure 12-34
(c) wipe front of the new stirring rod with gauze moisturized with 2% CS-anti-bacterial phosphor-free
detergent.
(d) Insert the new stirring rod until its end touches the bottom of axis motor. Then, secure it by M2 as figure
12-35 shows.
Stirring rod
Motor
Figure 12-35
(e) Place stirring rod adjust block on the rack of reaction cuvette, then move the stirring rod above the adjust
block as figure 12-36 shows.
199
User Manual of CS-400 Auto-Chemistry Analyzer
Stirring mechanism
Reaction cuvette
71 position
Height adjust screw
Adjust block
Figure 12-36
(f) loosen M2 screw, adjust the stirring rod position, its tip and upper side of adjust block are supposed to be
attached as figure 12-37 shows: tighten M2 screw.
Stirring rod
Figure 12-37
(g) Select “stirring mechanism horizontal check” in “maintenance” form, single-click “Execute”, “next” to
confirm stirring rod position weather correct. Please contact with after service.
(a) Turn off the purified water supply unit to stop the water supply
(c) Make the gauze as a cushion of the filter cap, Loosen the filter cap. Prepare a water container to receive
the water flow.
200
User Manual of CS-400 Auto-Chemistry Analyzer
Figure 12-38
(d) Pull out the filter, rinse the filter with water. Then reassemble it in the reverse order of the above.
(b) Loosen the two fixing screws of rubber hose on the back of the analyzer, and remove the hose as figure
12-39 shows.
(c) Unplug the cork and collect the outflow waste liquid with barrel or kind of containers.
Figure 12-39
(d) Plug the cork of rubber hose after empty the vacuum tank, and fix the hose on the cover with the two
screws.
201
User Manual of CS-400 Auto-Chemistry Analyzer
If water in the cooling water tank is used many times, it will be dirty and will be in bad water circulation, and
the volume will be reduced due to evaporation To prevent this, exchanging water in the cooling water bath once
a year is a must.
(b)open the left front door of the instrument, remove the cover of front left lower side of the instrument.
Figure 12-41
(c) Remove the rubber plug from the cooling water bath, and drain water using the hose pump (prepare a
water reservoir or bucket).
Figure 12-42
202
User Manual of CS-400 Auto-Chemistry Analyzer
(d) Fill purified water into the cooling water bath until purified water spills out of the high level of the hose.
(e) Switch on the main power for several minutes. Replenish purified water again into the low water level of
hose until the purified water spills again from the high level of the hose (approximate 3L water).
In case of transportation, discharge the purified water in the cooling water tank first.
Constant
temperature
Water tank
Robber plug
for hose
Figure 12-43
(d) Fill in the rubber plug after discharge the water, and mount the cover of instrument.
(a) Remove the reagent disk and clean the inside of the reagent cooling unit with gauze
203
User Manual of CS-400 Auto-Chemistry Analyzer
Figure 12-44
(b) Then, with gauze, clean the read-out window of the reagent barcode reader.
Barcode reader
window
Figure 12-45
(c) Remove the sample disk, and clean the inside of the sample disk tray with gauze.
Figure 12-46
204
User Manual of CS-400 Auto-Chemistry Analyzer
(a) Hold the handles of cooling fan at the both sides and remove the cover directly as figure 12-47 shows:
Figure 12-47
205
User Manual of CS-400 Auto-Chemistry Analyzer
LED Display
Figure 12-49
The LED display of cooling machine will circulated display the circuit value of semi-conductor cooling chip
1-4, water tank temperature of cooling system, and instrument interior temperature.
When the temperature of water tank of cooling system < 5~15℃, or the C1-C4 current value is less than 5-7 A,
alarm will be issued, please contact after service department.
Table 12-3 gives periodic cleaning and replacement parts (based on use of 5 hours daily).
(○:denotes periodic cleaning and check up and ●: denotes periodic replacement part.)
Frequency
Every Every Every Refer
No Item
Daily Required Weekly Monthly 2 3 6 to
Month Month Month
1 Sample syringe(SIP、IS、DIL) ● 12.4.11
2 Vacuum tank ○ 12.4.7
3 Sample pipeline rinsing ○ 12.5.2
4 Waste pipeline rinsing ○ 12.5.7
5 Reagent pipeline rinsing ○ 12.5.3
6 Na electrode ● 12.5.4
7 K electrode ● 12.5.4
8 Cl electrode ● 12.5.4
9 Indicator electrode ● 12.5.4
10 SIP tube(suction tube) ● 12.5.5
11 Cleaning of the waste ○ 12.5.6
Table12-3
206
User Manual of CS-400 Auto-Chemistry Analyzer
The pipe line could be contaminated by bacteria, fat and protein when testing electrolyte. Thus daily rinsing is a
necessity.
(b) Set “rinsing after test” in “Maintenance” menu as effective before the last test. If rinsing after test,
calibration should be taken when test ISE next time.
The ISE reagent pipeline could be contaminated after long time use. Rinsing it monthly according to the
following procedure.
(b) Infuse the internal standard liquid, diluent bottle with above diluted CS-ISE, carry out ISE (IS+DIL)
infuse in “maintenance” menu.
(c) After infusion finished, wash out the CS-ISE detergent which attached on the tube with water, wipe out the
water droplet with gauze, and put it back..
(2)ISE check up
Carry out 30 times ISE checkup in “maintenance” menu, and the result is showed.
Do not open ISE cover under normal situation, or temperature control error may cause test result inaccuracy.
When slope value is abnormal, alarm will be issued, as table 12-4 shows:
207
User Manual of CS-400 Auto-Chemistry Analyzer
Slope value
Alarm
Na K Cl
ISE prepare
above 68.1mV abov68.1mV below -68.1mV
abnormal
ISE prepare
37mV~32mV 37mV~32mV -32mV~-25mV
abnormal
ISE slope value
below 32mV below 32mV above -25mV
abnormal
Table 12-4
When ISE prepare alarm issued, analysis of current day could be proceed as normal, replace a new electrode
next day. When ISE slope value error occur, replace a new electrode immediately.
If alarm issued even the slope value within normal rage, that indicates the electrode response error. This is
caused by pipeline contamination. Rinse the pipeline to solve this problem.
If the calibration value is normal in previous day, but slope value change rapidly now, do not exclude other
reasons except electrode error. Check if there is a leakage or block in the pipeline.
(a) Open ISE cover at the left side of analyzer, turning-in the left side handgrip as figure 12-50 shows:
Hand shank
Figure 12-50
Remove the electrode by hand or by forceps after loosen the electrode as figure 12-51 shows:
208
User Manual of CS-400 Auto-Chemistry Analyzer
Figure 12-51
(b) Pull out of the electrode lead, connect the colored lead after replace the new electrode. As figure 12-52
shows:
Figure 12-52
Note 1: In order to prevent remaining of conductive component while replacing electrode, wipe out the liquid
in the vicinity of the electrode.
Note 2: In order to ensure gas tightness of electrode tube, a tube with O-ring is a must. Check if there is an
O-ring after replace new electrode.
209
User Manual of CS-400 Auto-Chemistry Analyzer
Debugging the electrode according to following sequence after replace a new electrode.
(b) Carry out 10 times ISE checkup in “maintenance” menu ten minutes later. The result of the ISE checkup
will be showed in maintenance form.
(c) Carry out ISE (full point)calibration, check if the slope value in within reference range.
When the slope value of Na, K, Cl are all unstable or lower, please replace a new one.
(a) Open left ISE cover of analyzing unit pull-in the handgrip, remove any electrode of Na, K, Cl after loosen
the electrode.
(c) Remove the fixing block of reference electrode, as figure 12-53 shows:
Fixed Reference
block electrode
Figure 12-53
Note1: Push down the wrench on the side of Na, K, Cl electrode side, fix the fixing block, the electrode could
be pull out easily by doing so.
Note3: Owing to the conductivity of the dropping liquid, wipe out the liquid by gauze wetted with purified
water.
210
User Manual of CS-400 Auto-Chemistry Analyzer
(b) Carry out 10 times ISE checkup in system maintenance ten minutes later. The result of the ISE checkup
will be showed in maintenance form. The value difference of the same electrode should within ±2Mv.
(c) Carry out ISE(full point)calibration, check if the slope value is within reference range.
(b) Pull out of the tube and replace a new one. Make sure the tube is not slack while connecting. As figure
12-54 shows:
(c) Wipe out the dropping liquid with gauze wetted with purified water.
SIP tube
Figure 12-54
(a) Wash the crystal attached on the ISE waste interface into waste container by purified water, and then wash
the waste container by purified water. As figure 12-55 shows:
211
User Manual of CS-400 Auto-Chemistry Analyzer
Waste
interface
Waste
container
Detergent bottle
Figure 12-55
(b) Wipe out the liquid by gauze wetted with purified water, Thus there will be no conductive component in
the waste interface and waste container.
Note: Touching the waste tube may cause noise jamming while testing.
(a) Place 1 ml CS-alkaline detergent at W1 position on sample disk, Place 1 ml CS-ISE detergent at W2
position on the sample disk.
(b) Place the CS-anti-bacterial phosphor-free detergent at 45 position on reagent disk 1 and reagent disk 2.
Note: Owing to the pipeline of electrolyte is cleaned at the same time, the electrode state may change, thus
carry out calibration before ISE test.
212
User Manual of CS-400 Auto-Chemistry Analyzer
Alarm level includes: Warning level, sample stop level, stop level, as table 13-1 shows:
Note appear both in data alarm and instrument malfunction. Buzzer calling, but the operation
Warning
proceed as usual, stop or proceed is judged by operator.
Alarm for instrument malfunction. Buzzer calling. Stop adding new sample. Sample on testing
Sampling stop
will proceed analyze.
Stop Alarm for malfunction. Buzzer calling. Instrument stop immediately. Test result is invalid.
Table 13-1
Table 13-2
213
User Manual of CS-400 Auto-Chemistry Analyzer
Table 13-3
214
User Manual of CS-400 Auto-Chemistry Analyzer
The absorbance to be used for 1.Check if the reagent is confected or positioned in the right way. 2.Check if the incubation
0-1 Warning Absorbance over calculation after cell blank bath is contaminated. 3.Check if there is the impurity interfuse. 4.Check if the reaction
correction exceeds 3.3ABS cuvette is blurred. 5.Check if there is an obstacle on the optical path of photometer.
215
User Manual of CS-400 Auto-Chemistry Analyzer
0-20 Warning Barcode repeat Barcode repeat on R1 disk Check the barcode stick information. 2.Scan barcode again after check.
Reagent R2 and R3
Reagent R2 and R3 is found on R1
0-21 Warning is found on R1 Check the barcode stick information. 2.Scan barcode again after check.
disk.
disk.
216
User Manual of CS-400 Auto-Chemistry Analyzer
Reagent R1 and R4
R1 and R4 detergent is found on R1
0-26 Warning is found on R2 Check the barcode stick information. 2.Scan barcode again after check.
or R4
disk
Communication abnormal. No
Communication 1.Check if serial port line is well connected. 2.Restart the instrument and software and
0-254 Stop response when instruction send to
abnormal then carry out on-line instruction.
host computer.
217
User Manual of CS-400 Auto-Chemistry Analyzer
218
User Manual of CS-400 Auto-Chemistry Analyzer
R1 Stirrer
The R1 stirring mechanism has 1.Call up the mechanism check screen and carry out the check program "mechanism check ".
1-9 Stop mechanism
ascend /descend error 2.Various malfunctions or trouble can not be restored. Contact the service department.
abnormal
R1 Stirrer
The R1 stirring mechanism has sway 1.Call up the mechanism check screen and carry out the check program "mechanism check ".
1-10 Stop mechanism
error 2.Various malfunctions or trouble can not be restored. Contact the service department.
abnormal
219
User Manual of CS-400 Auto-Chemistry Analyzer
R2 Stirrer
The R2 stirring mechanism has 1.Call up the mechanism check screen and carry out the check program "mechanism check ".
2-9 Stop mechanism
ascend /descend error 2.Various malfunctions or trouble can not be restored. Contact the service department.
abnormal
R2 Stirrer
The R2 stirring mechanism has sway 1.Call up the mechanism check screen and carry out the check program "mechanism check ".
2-10 Stop mechanism
error 2.Various malfunctions or trouble can not be restored. Contact the service department.
abnormal
1.This trouble is liable to occur after the reaction disk is washed. In this case, thoroughly
Reaction disk The reaction disk cannot detect wipe water droplets off the bottom of the reaction disk. 2.Check if water droplets adhere
4-1 Stop
abnormal the stop position. to the light coupler and code wheel located below the reaction disk. 3.if trouble can not
be restored, contact the service department.
1.This trouble is liable to occur after the reaction disk is washed. In this case, thoroughly
Reaction disk The reaction disk does not stop at wipe water droplets off the bottom of the reaction disk. 2.Check if water droplets adhere
4-2 Stop
abnormal the specified position to the light coupler and code wheel located below the reaction disk. 3.if trouble can not
be restored, contact the service department.
220
User Manual of CS-400 Auto-Chemistry Analyzer
1.This trouble is liable to occur after the reaction disk is washed. In this case, thoroughly
Reaction disk In resetting, the reaction disk wipe water droplets off the bottom of the reaction disk. 2.Check if water droplets adhere
4-3 Stop
abnormal cannot detect the home position to the light coupler and code wheel located below the reaction disk. 3.if trouble can not
be restored , contact the service department.
In resetting, the first reaction 1.This trouble is liable to occur after the reaction disk is washed. In this case, thoroughly
Reaction disk cuvette on the reaction disk does wipe water droplets off the bottom of the reaction disk. 2.Check if water droplets adhere
4-4 Stop
abnormal not stop at the specified to the light coupler and code wheel located below the reaction disk. 3.if trouble can not
position. be restored , contact the service department.
CS-alkaline
CS-alkaline detergent place in
4-6 Warning detergent Please add CS-alkaline detergent in detergent bottle.
front of instrument is shortage.
shortage
221
User Manual of CS-400 Auto-Chemistry Analyzer
1.Call up the mechanism check screen and carry out the check program "mechanism check".
The sample probe goes down 2.Various malfunctions or trouble can not be restored. Contact the service department.
Sample probe
5-4 Stop abnormally in descending motion 3.Check if check the altitude between probe and cup bottom is set properly. 4.Call up the
abnormal
(at the reaction cuvette side) system maintenance menu, carry out "sample probe horizontal checkup". 5.Check if the sample
cup is placed deflective.
Sample probe Sample probe descend abnormally 1.Call up the mechanism check screen and carry out the check program "mechanism check".
5-10 S. Stop
abnormal when discharge ISE test sample 2.Various malfunctions or trouble can not be restored. Contact the service department.
222
User Manual of CS-400 Auto-Chemistry Analyzer
Sample probe In rotation of reagent-1 probe, it 1.Call up the mechanism check screen and carry out the check program "mechanism check".
5-12 S. Stop
abnormal is not set at the upper dead point. 2.Various malfunctions or trouble can not be restored. Contact the service department.
Sample probe Sample probe cannot reach sample 1.Check if there is obstacle on sample disk. 2.Various malfunctions or trouble can not be
5-14 S. Stop
abnormal liquid when descend. restored. Contact the service department.
Sample probe Sample probe mechanism ascend/ 1.Call up the mechanism check screen and carry out the check program "mechanism check".
5-15 Stop
abnormal descend abnormal 2.Various malfunctions or trouble can not be restored. Contact the service department.
Sample probe Sample probe mechanism sway 1.Call up the mechanism check screen and carry out the check program "mechanism check".
5-16 Stop
abnormal abnormal 2.Various malfunctions or trouble can not be restored. Contact the service department.
Sample disk The sample disk cannot detect the 1.Call up the mechanism check screen and carry out the check program "mechanism check".
6-5 S. Stop
abnormal stop position on the inner track. 2.Various malfunctions or trouble can not be restored. Contact the service department.
223
User Manual of CS-400 Auto-Chemistry Analyzer
Sample disk In resetting, the sample disk 1.Call up the mechanism check screen and carry out the check program "mechanism check".
6-7 Stop
abnormal cannot detect the home position. 2.Various malfunctions or trouble can not be restored. Contact the service department.
Sample disk In resetting, the sample disk does 1.Call up the mechanism check screen and carry out the check program "mechanism check".
6-8 Stop
abnormal not leave the home position 2.Various malfunctions or trouble can not be restored. Contact the service department.
sample barcode Sample barcode reader cannot be 1.Check if barcode device is well connected. 2.Trouble cannot be restored, Contact the service
6-10 Warning
device abnormal found. department.
Sample syringe The sample syringe is not at the 1.Call up the mechanism check screen and carry out the check program "mechanism check".
7-1 S. Stop
abnormal upper dead point. 2.Various malfunctions or trouble can not be restored. Contact the service department.
Sample syringe The sample syringe does not leave 1.Call up the mechanism check screen and carry out the check program "mechanism check".
7-2 S. Stop
abnormal the upper dead point. 2.Various malfunctions or trouble can not be restored. Contact the service department.
1.Check if there is still reagent in the bottle. 2.Check if the reagent bottle is open. 3.Check
R1 probe The R1 probe goes down abnormally
8-2 Warning if the reagent disk cover is put in its proper place. 4.Check if the reagent-1 probe is bent.
abnormal in descending motion
5.Check if reagent bottle is lean placed.
224
User Manual of CS-400 Auto-Chemistry Analyzer
R1 probe In rotation of R1 probe, it is not 1.Call up the mechanism check screen and carry out the check program "mechanism check".
8-7 Stop
abnormal set at the upper dead point. 2.Various malfunctions or trouble can not be restored. Contact the service department.
R1 probe R1 probe cannot reach reagent 1.Check if there is obstacle on R1 reagent disk. 2.trouble cannot be restored, contact the
8-9 Stop
abnormal liquid level when descend. service department.
R1 probe R1 probe mechanism ascend/ 1.Call up the mechanism check screen and carry out the check program "mechanism check".
8-10 Stop
abnormal descend abnormal 2.Various malfunctions or trouble can not be restored. Contact the service department.
R1 probe 1.Call up the mechanism check screen and carry out the check program "mechanism check".
8-11 Stop R1 probe mechanism sway abnormal
abnormal 2.Various malfunctions or trouble can not be restored. Contact the service department.
1.Check if there is still reagent in the bottle. 2.Check if the reagent bottle is open. 3.Check
R2 probe The R2 probe goes down abnormally
9-2 Warning if the reagent disk cover is put in its proper place. 4.Check if the reagent-1 probe is bent.
abnormal in descending motion
5.Check if reagent bottle is lean placed.
225
User Manual of CS-400 Auto-Chemistry Analyzer
R2 probe In rotation of R2 probe, it is not 1.Call up the mechanism check screen and carry out the check program "mechanism check".
9-7 Stop
abnormal set at the upper dead point. 2.Various malfunctions or trouble can not be restored. Contact the service department.
R2 probe R2 probe cannot reach reagent 1.Check if there is obstacle on R2 reagent disk. 2.Trouble cannot be restored, contact the
9-9 Stop
abnormal liquid level when descend. service department.
The R1 disk cannot detect the stop 1.Call up the mechanism check screen and carry out the check program "mechanism check".
10-1 Stop R1 disk abnormal
position 2.Various malfunctions or trouble can not be restored. Contact the service department.
226
User Manual of CS-400 Auto-Chemistry Analyzer
The R1 disk does not stop at the 1.Call up the mechanism check screen and carry out the check program "mechanism check".
10-2 Stop R1 disk abnormal
specified position 2.Various malfunctions or trouble can not be restored. Contact the service department.
The R1 disk cannot detect the home 1.Call up the mechanism check screen and carry out the check program "mechanism check".
10-3 Stop R1 disk abnormal
position. 2.Various malfunctions or trouble can not be restored. Contact the service department.
The R2 disk does not stop at the 1.Call up the mechanism check screen and carry out the check program "mechanism check".
11-2 Stop R2 disk abnormal
specified position 2.Various malfunctions or trouble can not be restored. Contact the service department.
The R2 disk cannot detect the home 1.Call up the mechanism check screen and carry out the check program "mechanism check".
11-3 Stop R2 disk abnormal
position. 2.Various malfunctions or trouble can not be restored. Contact the service department.
227
User Manual of CS-400 Auto-Chemistry Analyzer
SIP sipper does not reach the 1.Call up the system maintenance menu,first carry out the program "reset",then go on
Electrolyte SIP
16-1 Warning lower dead point (when reset and "mechanism check" 2.Various malfunctions or trouble can not be restored. Contact the service
Sipper abnormal
operate) department.
1.Call up the system maintenance menu,first carry out the program "reset",then go on
Electrolyte SIP SIP sipper does not leave the
16-2 Warning "mechanism check" 2.Various malfunctions or trouble can not be restored. Contact the service
Sipper abnormal lower dead point
department.
Electrolyte 1.Call up the system maintenance menu,first carry out the program "reset",then go on
The vacuum nozzle does not reach
17-1 Warning vacuum Sipper "mechanism check" 2.Various malfunctions or trouble can not be restored. Contact the service
the lower dead point
abnormal department.
Electrolyte 1.Call up the system maintenance menu,first carry out the program "reset",then go on
The vacuum nozzle does not leave
17-2 Warning vacuum Sipper "mechanism check" 2.Various malfunctions or trouble can not be restored. Contact the service
the lower dead point.
abnormal department.
Electrolyte
SIP syringe does not at the upper
18-1 Warning sample Syringe Contact the service department.
dead point.
abnormal
Electrolyte
SIP syringe does not leave the
18-2 Warning sample Syringe Contact the service department.
upper dead point.
abnormal
Electrolyte
The diluent syringe is not at the
18-3 Warning sample Syringe Contact the service department.
upper dead point.
abnormal
Electrolyte
The diluent syringe is not leave
18-4 Warning sample Syringe Contact the service department.
the upper dead point.
abnormal
228
User Manual of CS-400 Auto-Chemistry Analyzer
Incubation bath
The water temperature in the 1.Check if cooling fan is operate normally. 2.check if colling fan dustproof cover is cloged
20-1 Warning water temp
incubation bath is over 45C by dirt 3. Trouble can not be restored, contact the service department.
abnormal
1.Clean liquid level sensor probe of incubation bath. 2.Check if there are water in purified
Incubator water The water level in the incubation
21-1 Warning water machine and pipeline 3. Please restart the instrument. 4.If the trouble cann't be
level error bath is too low
restored, contact the service department.
Water level too The water level in the water tank The distilled water tank is not supplied with water, Check the water pressure, water cock,
23-1 Stop
low is too low and other water supply systems.
229
User Manual of CS-400 Auto-Chemistry Analyzer
Automatic Timing calibration is not carried 1.Call up the system maintenance screen, first carry out the program "reset", then go on
51-1 Warning calibration out when reagent residual volume "mechanism check". 2.Various malfunctions or trouble can not be restored. Contact the service
disable cannot complete 10 tests. department.
230
User Manual of CS-400 Auto-Chemistry Analyzer
231
User Manual of CS-400 Auto-Chemistry Analyzer
232
User Manual of CS-400 Auto-Chemistry Analyzer
233
User Manual of CS-400 Auto-Chemistry Analyzer
Time sync Time sync instruction 1.Call up the system maintenance menu and carry out the check program "reset". 2.Various
143-1 Stop
failure transmitting failure. malfunctions or trouble can not be restored . Contact the service department.
1.Check the supplying pipe whether air exists in.check the height of water tank whether it
Add water Water tank liquid path system is more than 1.5 meters (distance from ground)if it will be used.2.Check if water supplier,
143-2 Warning
overtime error, add water overtime. pipeline and filter is normal. 3.Various malfunctions or trouble can not be restored. Contact
the service department.
AD board reset Error occurs when AD board reset, 1.Call up the system maintenance menu and carry out the check program "reset". 2.Various
143-3 Warning
failure AD board reset failure. malfunctions or trouble can not be restored. Contact the service department.
Error occurs when reaction disk 1.Check if there is waterdrop on the code wheel. 2.Call up the system maintenance menu and
Reaction disk
143-4 Warning resetting and reaction reset carry out the check program "reset". 3.Various malfunctions or trouble can not be restored.
reset failure
failure. Contact the service department.
234
User Manual of CS-400 Auto-Chemistry Analyzer
Water discharge Incubation bath discharge 1.Check if incubation filter and diluent waste pipeline is clog. 2.Various malfunctions or
143-8 Warning
failure failure. trouble can not be restored. Contact the service department.
R1 liquid level R1 probe cannot detect the liquid 1.Call up the system maintenance menu and carry out the check program "reset" 2.Various
143-10 Warning
detect failure level when adding detergent. malfunctions or trouble can not be restored. Contact the service department.
R2 liquid level R2 probe cannot detect the liquid 1.Call up the system maintenance menu and carry out the check program "reset". 2.Various
143-11 Warning
detect failure level when adding detergent. malfunctions or trouble can not be restored. Contact the service department.
Incubation bath 1.Check if liquid level sensor of incubation bath is clean and if there are water in incubation
Incubation bath liquid level
143-12 Warning liquid level bath 2.Restart the instrument. 3.Various malfunctions or trouble can not be restored.
detect failure.
detect failure. Contact the service department.
Incubation bath 1.Check if liquid level sensor of incubation bath is clean and if there are water in incubation
Add water overtime in incubation
143-13 Warning liquid path bath 2.Restart the instrument. 3.Various malfunctions or trouble can not be restored.
bath.
error Contact the service department.
Barcode scanning 1.Call up the system maintenance menu and carry out the check program "reset". 2.Various
143-14 Warning Barcode scanning overtime.
overtime. malfunctions or trouble can not be restored. Contact the service department.
235
User Manual of CS-400 Auto-Chemistry Analyzer
1.Call up the system maintenance menu and carry out the check program "reset". 2.Various
143-15 Warning Degas overtime Degas overtime
malfunctions or trouble can not be restored. Contact the service department.
Reaction
1.Call up the system maintenance menu and carry out the check program "reset". 2.Various
143-16 Stop initialize Reaction initialize failure
malfunctions or trouble can not be restored. Contact the service department.
failure
Reaction disk 1.Call up the system maintenance menu and carry out the check program "reset". 2.Various
143-17 Stop Reaction disk stop failure
stop failure malfunctions or trouble can not be restored. Contact the service department.
Sample probe
143-18 S .Stop Sample probe is block To remove the block please according to 12.4.1 in "user manual".
block
Add sample
143-19 S .Stop Last add sample failure After testing upon added sample, call up system maintenance menu, carry out "reset".
failure
143-20 S .Stop Add R1 failure Last add R1 failure. After testing upon added sample, call up system maintenance menu, carry out "reset"
143-21 S .Stop Add R2 failure Last add R2 failure After testing upon added sample, call up system maintenance menu, carry out "reset"
143-22 S .Stop Add R3 failure Last add R3 failure After testing upon added sample, call up system maintenance menu, carry out "reset"
R1 stirring
143-24 S .Stop Last R1 stirrer failure After testing upon added sample, call up system maintenance menu, carry out "reset"
failure
R2 stirring
143-25 S .Stop Last R2 stirrer failure After testing upon added sample, call up system maintenance menu, carry out "reset"
failure
R3 stirring
143-26 S .Stop Last R3 stirrer failure After testing upon added sample, call up system maintenance menu, carry out "reset"
failure
236
User Manual of CS-400 Auto-Chemistry Analyzer
R4 stirring
143-27 S .Stop Last R4 stirrer failure After testing upon added sample, call up system maintenance menu, carry out "reset"
failure
Waste bottle
143-28 Warning The waste bottle is full Empty the waste bottle.
full
Reagent
horizontal Reagent horizontal scanning 1.Call up the system maintenance menu and carry out the check program "reset ". 2.Various
143-30 Warning
scanning overtime. malfunctions or trouble can not be restored. Contact the service department.
overtime.
ISE check
1.Call up the system maintenance menu and carry out the check program "reset ". 2.Various
143-34 Warning operation ISE check operation overtime.
malfunctions or trouble can not be restored. Contact the service department.
overtime
ISE pipeline 1.Call up the system maintenance menu and carry out the check program "reset ". 2.Various
143-35 Warning ISE pipeline rinsing overtime.
rinsing overtime malfunctions or trouble can not be restored. Contact the service department.
237
User Manual of CS-400 Auto-Chemistry Analyzer
ISE testing Error occur when ISE testing, ISE 1.Call up the system maintenance menu and carry out the check program "reset ". 2.Various
143-36 Warning
error testing stop. malfunctions or trouble can not be restored. Contact the service department.
R1 disk cover
143-38 Warning R1 disk cover is open.
open
R2 disk cover
143-40 Warning R2 disk cover is open.
open.
Instrument Error occurs between instrument 1.Pwer off and restart the instrument. 2.Various malfunctions or trouble cannot be restored.
143-44 Stop
module error modules. Contact the service department.
238
User Manual of CS-400 Auto-Chemistry Analyzer
Continue
Continually 5 contaminated cup Carry out cell blank test to judge cup states. If cell blank value is abnormal, please change
143-45 Stop contaminated cup
occur. reaction cuvette cup, if cell blank value is normal please contact maintenance department.
occur
In resetting, if trouble cannot remove, or other malfunctions exist, please contact
143-46 Stop AD data error AD data error
maintenance department.
ISE test internal standard liquid In resetting, if trouble cannot remove, or other malfunctions exist, please contact
143-47 Warning Test ISE error
failure maintenance staff.
Read edition 1.Call up the system maintenance menu and carry out the check program "reset". 2.Various
143-48 Warning Read edition number overtime
number overtime malfunctions or trouble can not be restored. Contact the service department.
Cooling system
144-2 Warning Cooling circuit abnormal Please contact the maintenance department.
abnormal
Cooling system
144-3 Warning Cooling chip abnormal Please contact the maintenance department.
abnormal
Cooling system
144-4 Warning Cooling liquid level abnormal Please add water in cooling system water take according to 12.4.8 in the "user manual".
abnormal
Cooling system
144-5 Warning Cooling status abnormal Please contact the maintenance department.
abnormal
239
User Manual of CS-400 Auto-Chemistry Analyzer
240
User Manual of CS-400 Auto-Chemistry Analyzer
ISE Internal
Internal standard liquid shortage 1. Add ISE internal standard liquid. 2. Call up system maintenance screen, carry out ISE
38-1 Warning standard liquid
(Less than setup volume). internal standard liquid pipeline washing. 3. Carry out ISE calibration. .
shortage
R1 reagent
10-5 Stop number R1 reagent number invalidity. Please contact the service department.
invalidity
R2 reagent
11-5 Stop number R2 reagent number invalidity. Please contact the service department.
invalidity
241
User Manual of CS-400 Auto-Chemistry Analyzer
The R2 disk cannot detect the stop 1.Call up the mechanism check screen and carry out the check program "mechanism check".
11-1 Stop R2 disk abnormal
position 2.Various malfunctions or trouble can not be restored. Contact the service department.
Cooling fan
144-8 Warning Cooling fan running abnormal. Please contact maintenance staff.
malfunction
242
User Manual of CS-400 Auto-Chemistry Analyzer
Concentrated
waste liquid Concentrated waste liquid Call up the system maintenance menu, carry out the program "Auto Rinsing Concentrated waste
0-14 Warning
pipeline need to pipeline need to Rins liquid pipeline"
Rins
××× — ×××
Small sort
Big sort
243
User Manual of CS-400 Auto-Chemistry Analyzer
Waterproof and moisture proof is required while transportation. Do not extrude or vibrate the instrument.
Handle it gently while carrying and loading.
Instrument should be stored in clean room with no chemical, no corrosive gas. Height above sea level within
2000 meter, temperature within -10℃~40℃, relative humidity within: 40% ~85% atmospheric
pressure within: 76kPa~106kPa.
244
User Manual of CS-400 Auto-Chemistry Analyzer
Thank you for purchasing CS-400 Auto-Chemistry Analyzer of our company. We can offer you the following
service:
2. One year warranty from the day purchased. If instrument malfunction is caused by design defect,
manufacture defect, our company will repair them without payment.
Address:95,Yunhe Street, New & High Tech. Development Zone, Changchun, China
Fax:0431-85100405
Zip code:130012
E-mail:dirui@dirui.com.cn
Website: http://www.dirui.com.cn
245
User Manual of CS-400 Auto-Chemistry Analyzer
Belong to biochemical analyze system in clinic test analyze instrument (6840), type II in management type.
Following table explains the accessory reagent (often used biochemical reagent parameter table).
246
User Manual of CS-400 Auto-Chemistry Analyzer
Measurable Reagent
wavelength (nm) volume Blank
Photo Calibr
Item Assay Sample Span Reaction Deviation Discrete Sensitivity horizo
Time Main Sub metric Calibration method ation
name mode volume point direction rate check check check ntal
point point
wavel wavel R1 R2 check
ength ength
0.8-2
ALT Rate A 10 340 405 21-31 15 240 60 2-point linearity 2 2 Negative 3.3 0.05 0
.5
0.8-2
AST Rate A 10 340 405 21-31 15 240 60 2-point linearity 2 2 Negative 3.3 0.05 0
.5
ALP Rate A 10 405 505 21-31 4 200 50 2-point linearity 2 2 Positive 3.3 0.05 0 0-1.0
GGT Rate A 10 405 505 21-31 25 200 50 2-point linearity 2 2 Positive 3.3 0.05 0 0-1.2
2point
TBA rate 10 405 700 22-28 4 270 90 2-point linearity 2 2 Positive 3.3 0.05 0 0-1.2
assay
1point
TBIL 10 546 660 31 20 0 20 2-point linearity 2 2 Positive 3.3 0.05 0 0-0.5
assay
1 point
DBIL 10 546 600 31 20 0 20 2-point linearity 2 2 Positive 3.3 0.05 0 0-0.5
assay
1 point
TP 10 546 700 31 5 250 0 2-point linearity 2 2 Positive 3.3 0.05 0 0-0.5
assay
1 point
ALB 10 600 700 12 2 300 0 2-point linearity 2 2 Positive 3.3 0.05 0 0-0.5
assay
LAP Rate A 10 405 505 21-31 15 240 60 2-point linearity 2 2 Positive 3.3 0.05 0 0-1.2
SCH
Rate A 10 405 505 20-26 3 240 60 2-point linearity 2 2 Positive 3.3 0.05 0 0-3.3
E
ICDH Rate A 10 340 405 21-31 15 240 60 2-point linearity 2 2 Positive 3.3 0.05 0 0-0.8
247
User Manual of CS-400 Auto-Chemistry Analyzer
Reagent
Measurable(nm)
Photom volume Calibr Deviation Discrete Sensit Blank
Item Assay Sample Calibration Span Reaction
Time Main- Sub-w etric ation Rate rate ivity horizontal
name mode volume method point direction
wavel avelen point R1 R2 point check check check check
ength gth
Rate A 2-point
GLDH assay
10 340 405 21-31 15 240 60 2 2 Negative 3.3 0.05 0 0.8-2.5
linearity
Rate A 2-point
AMY assay
10 405 505 21-26 5 160 40 2 2 Positive 3.3 0.05 0 0-1.2
linearity
2-point 2-point
BUN assay
10 340 405 20-26 3 240 60 2 2 Negative 3.3 0.05 0 0.8-2.2
linearity
2-point 2-point
CRE assay
10 546 700 16-31 7.5 240 60 2 2 Positive 3.3 0.05 0 0-0.8
linearity
2-point 2-point
CRE assay
10 505 660 20-26 20 150 150 2 2 Positive 3.3 0.05 0 0-1.2
linearity
2-point 2-point
GLU assay
22 505 660 16-110 2 240 60 2 2 Positive 3.3 0.05 0 0-0.8
linearity
GLU 2-point 2-point
10 340 405 16-31 2 240 60 2 2 Positive 3.3 0.05 0 0-1.0
(HK) assay linearity
2-point 2-point
FMN assay
10 546 700 16-31 15 300 0 2 2 Positive 3.3 0.05 0 0-0.8
linearity
2-point 2-point
TC assay
10 505 660 16-31 3 240 60 2 2 Positive 3.3 0.05 0 0-0.5
linearity
2-point 2-point
TG assay
10 546 700 16-31 3 240 60 2 2 Positive 3.3 0.05 0 0-0.5
linearity
2-point 2-point
HDL-C assay
10 546 660 16-31 3 225 75 linearity 2 2 Positive 3.3 0.05 0 0-0.8
2-point 2-point
LDL-C assay
10 546 660 16-31 4 300 100 2 2 Positive 3.3 0.05 0 0-0.8
linearity
Rate A 2-point
CK-MB assay
10 340 405 25-36 10 200 50 2 2 Positive 3.3 0.05 0 0-0.8
linearity
Rate A 2-point
CK assay
10 340 405 21-31 5 200 50 2 2 Positive 3.3 0.05 0 0-1.2
linearity
248
User Manual of CS-400 Auto-Chemistry Analyzer
Reagent
Wavelength(nm)
Photom volume Calibr Deviati Sensit Blank
Item Assay Sample Calibration Span Reaction Discrete
Time Main-w Sub-w etric ation on rate ivity horizontal
name mode volume method point direction check
avelen avelen point R1 R2 point check check check
gth gth
Rate A 2-point
HBDH assay
10 340 405 21-31 6 240 60 2 2 Negative 3.3 0.05 0 0.8-2.5
linearity
Rate A 2-point
LDH assay
10 340 405 21-31 5 240 60 2 2 Positive 3.3 0.05 0 0-0.8
linearity
2-point 2-point
UA assay
10 546 700 16-31 4 200 50 2 2 Positive 3.3 0.05 0 0-1.5
linearity
1-point 2-point
P assay
10 340 405 10 4 200 0 2 2 Positive 3.3 0.05 0 0-0.8
linearity
2-point 2-point
Ca-CPC assay
10 570 660 16-31 5 150 150 2 2 Positive 3.3 0.05 0 0-0.8
linearity
1-point 2-point
Cl assay
10 505 660 31 3 300 0 2 2 Positive 3.3 0.05 0 0-0.5
linearity
1-point 2-point
Ca-ARS assay
10 660 750 31 3 300 0 2 2 Positive 3.3 0.05 0 0-0.8
linearity
1-point 2-point
Mg assay
10 546 750 10 3 300 0 2 2 Positive 3.3 0.05 0 0-1.2
linearity
Note: For specific parameter, please refer to reagent instruction. In order to add items, please set according to the parameters of reagent instruction. To calculate sample
result, calibration by calibration serum with the practical factor is advised if the items are tested by rate A assay.
249
User Manual of CS-400 Auto-Chemistry Analyzer
Statement
Dirui Co. LTD. has the final power of interpretation.
Dirui Co. LTD. is responsible for the security, reliability and capability of CS-240 B auto-Chemistry analyzer
under following circumstance.
1) Installation, adjustment, improvement and repair are proceeded by Dirui company professionals.
2) Relevant electric equipments are qualified according to state norms.
3) User Manual should be obeyed when operating instrument.
250