Download as pdf or txt
Download as pdf or txt
You are on page 1of 254

User Manual of CS-400 Auto-Chemistry Analyzer

Instruction:
Dear user, thanks for purchasing our CS-400 Auto-Chemistry Analyzer.
Please read the user manual carefully in order to operate the instrument correctly. Incorrect operation may affect
the precision and accuracy of the test results, or endanger personal safety.
Please keep the user manual safely for your any time reference.

Note:
● Instrument should be operated by medical inspection specialist, physician, nurse or lab assistant who are
specially trained.
● Instrument should be controlled by special software. Please install the software that is appointed by our
company. Installation of other software/hardware may interferer normal operation. Don’t operate other software
when instrument operating.
● Dust may accumulate on the surface of instrument after long time storage. Soft cloth or gauze can be used for
cleaning work, and a little detergent can be used if necessary. Please cut off the power supply before cleaning.
When instrument is not used, make sure shut the lid down.
● As to the use and storage method of the sample, reagent, Controls, Calibrator, please refer to the relevant
instructions.
● Sample, Controls, Calibrator and waster solution have the potential biochemical infectivity; the detergents are
corrosive that may hurt eyes, skin and mucosa. Operator should refer to the safety regulation for lab operation.
Protective measure should be taken to operator (Such as lab protective clothes and gloves).
● Avoid contact with eyes and skin, in case of skin contact, flush the area with water, rinse immediately with
plenty of water and seek medical advice.
● Operator should comply with the local regulation when draining and dealing with reagent, waste solution, waste
sample, consumable etc. Please dispose the waste solution and instrument consumable according to the regulation
of medical waste, infective waste and industrial waste.

I
User Manual of CS-400 Auto-Chemistry Analyzer

Warning:
● Instrument should be operated in a good ground condition, and an independent power supply is a must, the input
power should be conformed to instrument requirement.
● Don’t pull the electrical wire with wet hand, or there is a risk of electrical shock.
● Doesn’t stamp, twist, drag the wire and cable, or it may cause a fire.
● Please don’t open the back and side cover board before cutting the general power supply except Dirui special
service staff.
● If liquid occurs in instrument interior or there is an internal pipeline leakage, please immediately cut off the
general power supply, and contact Dirui customer service dept.
● Please don’t touch sample probe, reagent probe and stirring rod, etc. when instrument operating, don’t put your
hand into the opening part, or it may cause body injury or instrument damage.
● Cut off the power supply before replace light source lamp. Don’t touch the lamp before it is cool to avoid
burning.
● Periodic maintenance should be executed strictly according to the user manual. Or it may cause instrument
malfunction, and affect the accuracy and precision of test results.
● Make sure that the Auto-Chemistry Analyzer is operated according to the user manual, or the measuring result is
not a reliable one, and the damage on instrument may endanger human safety.
● Please don’t place combustible material around the instrument.

II
User Manual of CS-400 Auto-Chemistry Analyzer

Catalogue
Chapter 1 Brief Introduction ...........................................................................................................................1 
1.1 Summary ......................................................................................................................................................1 
1.2 Main technical index ...................................................................................................................................1 
1.3 Composition of instrument .........................................................................................................................3 
1.4 Configuration and function ........................................................................................................................8 
1.5 Instrument Symbol ....................................................................................................................................21 
Chapter 2 Function and Measuring Principle .............................................................................................22 
2.1 Mechanism movement principle ..............................................................................................................22 
2.2 Assay mode .................................................................................................................................................24 
2.3 Check of measure value ............................................................................................................................57 
2.4 ISE testing principle ..................................................................................................................................62 
Chapter 3 Instrument Installation ................................................................................................................66 
3.1 Installation requirement ...........................................................................................................................66 
3.2 Open package .............................................................................................................................................67 
3.3 Installation procedure ...............................................................................................................................68 
Chapter 4 Accessory Device ...........................................................................................................................76 
4.1 Sample disk barcode reader .....................................................................................................................76 
4.2 Reagent barcode reader ............................................................................................................................77 
4.3 Purified water equipment .........................................................................................................................78 
Chapter 5 CS-400 Software Operation .........................................................................................................79 
5.1 Software interface instruction ..................................................................................................................79 
5.2 Software Operation ...................................................................................................................................83 
5.3 Instrument standard specification ...........................................................................................................86 
Chapter 6 Instrument Operation ..................................................................................................................88 
6.1 Overview of operation ...............................................................................................................................88 
6.2 Detailed operation .....................................................................................................................................89 
Chapter 7 Calibration Information ............................................................................................................122 
7.1 Colorimetric calibration..........................................................................................................................122 
7.2 ISE calibration .........................................................................................................................................127 
Chapter 8 Quality Control ...........................................................................................................................129 
8.1 QC registration ........................................................................................................................................129 
8.2 QC interval ...............................................................................................................................................133 
8.3 Monthly quality control ..........................................................................................................................135 
Chapter 9 System Setup ...............................................................................................................................137 
9.1 Chemistry parameter ..............................................................................................................................137 
9.2 Item combination .....................................................................................................................................145 
9.3 Calculated item ........................................................................................................................................146 
9.4 Cross contamination ................................................................................................................................147 
9.5 Report sheet format .................................................................................................................................150 
9.6 ISE Setup ..................................................................................................................................................153 
9.7 Other setup ...............................................................................................................................................154 
9.8 Manual item setup ...................................................................................................................................156 
9.9 LIS communication setup .......................................................................................................................156 
Chapter 10 System management .................................................................................................................158 
10.1 User information ....................................................................................................................................158 
10.2 Hospital information .............................................................................................................................159 
10.3 Other information..................................................................................................................................160 
10.4 Workload statistics ................................................................................................................................165 
10.5 Database maintenance ...........................................................................................................................167 
10.6 System log ...............................................................................................................................................168 
Chapter 11 System Help ...............................................................................................................................169 
11.1 System help application.........................................................................................................................169 
Chapter 12 System Maintenance .................................................................................................................170 
12.1 System maintenance preparation .........................................................................................................170 
12.2 The Application of system maintenance menu....................................................................................171 
12.3 Maintenance and checkup points and parts ........................................................................................180 

III
User Manual of CS-400 Auto-Chemistry Analyzer

12.4 Maintenance and check up method......................................................................................................183 


12.5 ISE device maintenance ........................................................................................................................206 
Chapter 13 Alarm Data Processing .............................................................................................................213 
13.1 Alarm information type ........................................................................................................................213 
13.2 Countermeasure to malfunction do not issue alarm...........................................................................213 
13.3 Instrument alarm list.............................................................................................................................214 
Chapter 14 Instrument Transportation and Storage .................................................................................244 
14.1 Transportation requirement .................................................................................................................244 
14.2 Storage requirement ..............................................................................................................................244 
14.3 Storage environment .............................................................................................................................244 
Addendum A Product Warranty .................................................................................................................245 
Addendum B Product Description ..............................................................................................................246 
Statement ..........................................................................................................................................................250 

IV
User Manual of CS-400 Auto-Chemistry Analyzer

Chapter 1 Brief Introduction

1.1 Summary

CS-400 Auto-Chemistry Analyzer is an instrument with discrete system, reagent open function, emergency
priority function as well as an external computer. The instrument is composed of humanized software operation
system, intelligent zed optical unit, complicated mechanism system, precision liquid path and accuracy electrical
system. The instrument could automatically realize sampling, reagent injection, anti-interference, mixture,
pre-temperature, reaction measurement, rinse, calculation, display and print function. The substitution of manual
operation for automatic operation could not only enhance the working efficient but also decrease the test error,
thus greatly enhance the accuracy and precision of test results.

CS-400 Auto-Chemistry Analyzer could carry out the immunology check and biochemical analyze of blood,
urine, ascites, cerebrospinal fluid and other body fluid. The instrument could also carry out clinic test, such as:
myocardium enzymogram, blood sugar, blood fat, liver function, renal function, immunoglobulin, etc.

1.2 Main technical index

Instrument structure: Discrete system

Throughput: Constant speed, 400 tests/ hour (800 tests/ hour with ISE)

Simultaneous analysis item No.: At most 88 colorimetric items,

3 ISE items (K, Na, Cl).

Sample volume: 2 to 35μl,(Stepping 0.1μl)

Reagent volume: 20 to 350μl,(Stepping 1μl)

Reaction solution volume: 150~450μl

Liquid level sensor: Integration of sample probe and reagent probe with touch sensor and sample probe
block test function.

Stirring: Independent stirring after reagent injecting.

Sample disk: 115 samples position (50 routine samples, 34 Calibrators, 20 STAS samples, 8
Controls, 3 Detergents)

Reagent disk: Dual disk

Disk 1: 45 positions for Reagent 1, Reagent 4, diluents and CS anti-bacterial


phosphor-free detergent.

Disk 2: 45 positions for Reagent 2, Reagent 3, CS anti-bacterial phosphor-free


detergent.

Photometer: Grating spectrophotometry system in a range of 340~750nm, wavelength: 340、


380、405、450、480、505、546、570、600、660、700、750nm

1
User Manual of CS-400 Auto-Chemistry Analyzer

Wave length accuracy: ±2nm

Light source: 20W /12V Long life quartz halogen lamp (water cooling)

Measurement range: 0 to 3.3Abs

Reaction disk: 120 pcs of reusable rigid optical plastic reaction cuvette.

Reaction cuvette optical diameter: 6mm

Reaction cuvette rinse: Automatic

Incubation bath temperature: 37℃±0.1℃

Reaction time: 15 minutes at most (optional for 3, 4,5,10 and 15 minutes)

Analysis method: Rate assay, end-point assay, 2-point assay.

Calibration method: 1-point linearity, 2-point linearity, multi-point linearity, non-linearity method.

Reagent bottle volume: 20ml, 70ml

Reagent cooling unit: All reagents keep at 5℃ - 15℃, semiconductor refrigeration.

Barcode scanning: 3 internal barcode scanner ( scan the barcode on the routine sample, and reagent
on the R1, R2 disk )

Reagent volume test : Test and report the reagent remaining volume.

Power supply: ~220/230V 50Hz

Ambient temperature: 15℃~32℃ Optimum temperature: 18℃ to 25℃

Relative humidity: 40% ~ 85%

Appearance dimension: 1060×790×1150mm ( length×width×height)

Fixing power: 2000VA

Weight: About 300Kg

2
User Manual of CS-400 Auto-Chemistry Analyzer

1.3 Composition of instrument

1.3.1 Appearance of instrument

1.3.1.1 Front of instrument

① ②

③ ④

① Model symbol ② Cover ③ Left front door ④ Right front door

Figure 1-1 Front of instrument

3
User Manual of CS-400 Auto-Chemistry Analyzer

1.3.1.2 Front of opening door

① ② ③ ④ ⑤ ⑥ ⑦

① ISE pipetting syringe ② ISE diluting syringe ③ ISE inner standard solution syringe

④ CS-Alkaline detergent port ⑤ Sample syringe ⑥ R2 syringe ⑦ R1 syringe

Figure 1-2 Front of opening door

4
User Manual of CS-400 Auto-Chemistry Analyzer

1.3.1.3 Rear of instrument

① ② ③ ④ ⑤ ⑥ ⑦ ⑧

① Power supply inlet ② RS-232 interface ③ Cooling fan ④ Diluted waste solution outlet

⑤ Purified water inlet ⑥ Vacuum tank waste solution outlet ⑦ Concentrated waste outlet

⑧ Port of concentrated waste level sensor

Figure 1-3 Rear of instrument

5
User Manual of CS-400 Auto-Chemistry Analyzer

1.3.1.4 Top of instrument

① ② ③ ④ ⑤ ⑥

⑦ ⑧ ⑨ ⑩ ○
11 ○
12

① Sample pipetting mechanism ② Cuvette rinsing mechanism ③ Reaction disk

④ R1 stirring mechanism ⑤ R1 reagent pipetting mechanism ⑥ R1 reagent disk

⑦ Pilot lamp of sample disk rotation ⑧ cooling lip of inner track of sample disk ⑨ Sample disk

⑩ R2 reagent pipetting mechanism ○


11 R2 stirring mechanism ○
12 R2 reagent disk

Figure 1-4 Top of instrument

6
User Manual of CS-400 Auto-Chemistry Analyzer

1.3.1.5 Right of instrument

① ② ③ ④

① Analytical unit switch (cooling power supply is excluded) ② Pilot lamp of cooling power supply (green)
③ General power supply switch ( breaker) ④ Pilot lamp of power supply ( red)

Figure 1-5 Right of instrument

7
User Manual of CS-400 Auto-Chemistry Analyzer

1.3.2 System configuration

Figure 1-6 System configuration

1.4 Configuration and function

CS-400 Auto-Chemistry Analyzer is composed by operating system and analytical system. The two parts is
connected by RS-232 serial wire.

1.4.1 Operating system

Operating system is composed of host, 17 inch CRT display monitor, keyboard, mouse and printer.

Host computer: Windows XP system

Special applied software and database.

Computer configuration: Basic frequency ≥2.8GHz.

Hard disk≥160G Memory ≥1G

With RS-232 serial interface, website interface and USB interface.

CRT display monitor: Display all kinds of form, curve and test data of CS-400 software.

Keyboard: Operation control and data input.

Mouse: Carry out software operation

Printer: Print out test data and chart.

8
User Manual of CS-400 Auto-Chemistry Analyzer

1.4.2 Analytical system


Analytical system is composed of sample disk, sample pipetting mechanism, reagent disk, reagent pipetting
mechanism, reaction disk, stirring mechanism, cooling system, rinsing mechanism, optical system etc.

1.4.2.1 Sample disk

! Warning:

z When instrument operating, be sure to close the cooling lid on the inner track of sample disk and screw the
knob.

z The sparkling of LED pilot lamp indicates that sample disk is turning or going to turn, do not change sample
or touch sample disk at this time, or it may cause body injury or instrument damage.

① ② ③ ④⑤ ⑥ ⑦⑧ ⑨

① Outer track ② Middle track ③ Inner track ④ Sample disk inner track pin

⑤ Handgrip of sample disk inner track ⑥ Sample disk guide pin ⑦ Locking block of inner/outer disk

⑧ Handgrip of sample disk outer track ⑨ Sample barcode reader

Figure 1-7 Sample disk

(1) Composition and function

Sample disk is composed of outer track, middle track, inner track and LED pilot lamp. The inner track has
cooling function, thus the Controls and Calibrator on inner track of disk can be restored in 5℃~15℃.

Place the containers (standard cup, micro cup, test tube) which contain Calibrator, sample, Controls on the
sample disk, and then the sample disk will send them to the sampling position in the sampling mechanism.

(2)Specifications ( Sample No.)

(Outer track): For routine samples (1-50)………………………………………….50cups

9
User Manual of CS-400 Auto-Chemistry Analyzer

(Middle track): For Calibrator (S1-S17)……………………………………………...17cups

For STAT sample (E51-E70)………………………………….…….. 20 cups

For detergent (W1-W3)………………………………………………..3 cups

(Inner track): For Calibrator (S18-S34)……………………………………………..17 cups

For Controls(C1-C8)………………………………………………..8 cups

(3)Movement

At Power on: Sample disk turns counterclockwise to move routine sample No.1 to the sampling position.

At analysis: At start of analysis, sample disk makes the same movement as “power on”. During analysis,
sample disk turns to the direction allowing a quicker access.

At resetting: Make the same movement as at “power on”.

(4) Mounting / dismounting

When mounting the sample disk, be sure to set the sample disk matching with the guide pin. Set the latch
on the inner track. Also, be sure to secure the cooling unit lid on the inner track, the outer track can be
demounted without removing the inner track.

Note:When mounting/demounting the routine sample disk (outer track), be sure to hold the hooks with both hands.
In order to change the sample in inner track, be sure that sample probe has stopped sampling or been under the
stand-by status and take out the disk cover first. Before operating, please check the disk position.

(5) Action check

Single-click “ maintenance” key, select “ mechanism operation checkup”, input the check times,
single-click “ Execute ” button. Alarm is issued when abnormality exists.

1.4.2.2 Sampling mechanism

! Warning:

● Please don’t touch the sampling mechanism when operating, or it may cause body injury or instrument
damage.

10
User Manual of CS-400 Auto-Chemistry Analyzer

① ② ③ ④
① Sample probe up and down mechanism ② Sample probe rotation arm

③ Sample probe ④ Rinsing bath of sample probe

Figure 1-8 Sampling mechanism

(1) Function

Assimilates a specified amount of sample from sample container and pipets it into reaction cuvette. The
sample probe possesses the liquid level detect function. Alarm is issued when touch occurs in descending
process, and alarm is issued also when sample probe is blocked.

(2)Specification

Sampling mechanism can assimilate 2.0-35.0 ul sample volume, set in 0.1ul stepping. Sample probe will
assimilate more than specified amount. Minimum sample volume requires more than 100 ul.

(3)Movement

At power on: The sample probe comes over above the reaction cup, and then returns above the sample
probe rinse bath.

At analysis: The sample probe circulates up and down motions as the following sequence:Sample
container, reaction cuvette, sample rinsing bath. Automatic sample probe rinsing is carried
out when sampling finished. At start of analysis, the sample probe makes the same movement
as at power on. In the rinsing bath, the inner and outer walls of the probe are rinsed. Sample
probe block test is carried out simultaneously.

At resetting: Makes the same movement as at power on.

Assimilate sample: sampling is taken when sample probe descent 1.7mm below liquid level.

(4) Automatic rinsing

Automatic rinsing of probe: After assimilating the sample, the sample probe will return above the rinsing
bath to wash the inner and outer walls of probe。When sampling is finished, the CS-alkaline detergent is

11
User Manual of CS-400 Auto-Chemistry Analyzer

assimilated from the position W1 of the sample disk.

(5)Operation check

Single-click the “Maintenance”key, select“mechanism operation checkup and input the check times. Click
“Execute “. If abnormality exists, instrument will issue alarm.

1.4.2.3 Reagent disk


! Warning:

● Please don’t touch the lid of the reagent disk when running or it may cause body injury and instrument
damage.

① ② ③ ④ ⑤ ⑥ ⑦

① Reagent disk cover ② Locking knob of cover ③ Reagent bottle ( on the reagent rack)

④ Track pin of reagent disk ⑤ Handgrip of reagent disk. ⑥ Reagent disk guide pin

⑦ Reagent disk open detector

Figure 1-9 Reagent disk

(1)Function

The reagent disk accommodates reagent bottles and carries the specific reagent, diluents and
CS-anti-bacterial phosphor-free detergent to the pipetting position in pipetting mechanism. Cooling system
can keep the reagent disk in a lower temperature to store reagent. There is a barcode reader on the wall of
reagent cooling unit, which can scan the barcode on reagent bottle.

(2)Specification

Reagent disk 1 (R1): R1, R4, diluents and detergent, total 45 bottles. Number 45 position is specially used
for CS-anti-bacterial phosphor-free detergent.

Reagent disk 2 (R2): R2, R3, diluents and detergent, total 45 bottles. Number 45 position is specially used

12
User Manual of CS-400 Auto-Chemistry Analyzer

for CS-anti-bacterial phosphor-free detergent.

Reagent bottle capacity: 70ml、20ml

Single-reagent usage: Single-reagent can be used as Reagent 1 or Reagent 2 that is putted into both two
reagent disk.

(3)Movement

At power on: Turning clockwise, Position 1 of R1 and R2 disks carry each bottle to the reagent pipetting
position.

At analysis: Initial operation is the same as at power on. Subsequently, the disks rotate in the direction
which allows a quicker access.

At resetting: Make sure same as at power on.

(4)Operation check

Single-click the “Maintenance”key, select“mechanism operation checkup and input the check times. Click
“Execute“. If abnormality exists, instrument will issue alarm.

(5)Mounting/ Dismounting

Fix the disk with 2 latches at the center of disk. To remove the disk, loose the latches. For mounting, be sure
to set the disk matching with the guide pin and fasten it with the latches. The reagent disk cover must be
attached except for replacement of the disk of reagent. During operation, the reagent probe may moves, so
avoid detaching the reagent disk cover at this time.

Note: If operator open the disk cover during stand-by or running state, alarm occurs. Reagent horizontal scanning
will be automatically carried out when cover the reagent disk lid at stand-by status.

1.4.2.4 Reagent pipetting mechanism

! Warning:

● Be sure to close the lid of reagent disk when instrument operating, and fasten the knob.

13
User Manual of CS-400 Auto-Chemistry Analyzer

① ② ③ ④

① Reagent probe up and down mechanism ② Rotating arm of reagent probe

③ Reagent probe ④ Rinsing bath of reagent probe

Figure 1-10 Reagent pipetting mechanism

(1)Function

Assimilate a specified amount of reagent from each reagent bottle and pipets it into a reaction cuvette. The
reagent probe acts as a sensor at the liquid surface level. The remaining amount of reagent in a bottle is
calculated from the descending distance of reagent probe and displayed on the “reagent info.’ menu.

(2)Specification

Reagent volume: 20 to 350ul set in 1 ul stepping.

(3)Movement

At power on:Move toward the reaction cup side once and then returns above the probe rinsing bath.

At analysis: Reagent probe circulates the motion as the following sequence: reagent bottle--reaction
cuvette—reagent probe rinsing bath.

At resetting: Make the same action as at power on.

(4)Automatic rinsing

Assimilate CS-anti-bacterial phosphor-free detergent from the position 45 on the reagent disk for three
times and infuse it into reaction cuvette for three times and return to the probe rinsing bath to wash inner
and outer walls of the probe.

(5)Movement check

Single-click the “Maintenance”key, select“mechanism operation checkup and input the check times. Click
“Execute “. If abnormality exists, instrument will issue alarm.

14
User Manual of CS-400 Auto-Chemistry Analyzer

1.4.2.5 Reaction disk

! Warning:

● Please don’t touch the reaction disks during operating, or it may cause mechanical failure.

① ② ③ ④ ⑤ ⑥

① Reaction cuvette rinsing mechanism ② Reaction disk ③ Fixing screw

④ Fixing knob of reaction disk ⑤ Guide pin and hole ⑥ Reaction cup groupware handgrip

Figure 1-11 Reaction disk

(1)Function

Fasten reaction cuvette by screw, and be sure to keep chemical reaction in constant temperature of 37℃.
Each reaction cuvette also serves as a cell for absorbance measurement.

(2)Specification

Reaction cuvettes No: 20 cuvettes/set * 6 sets ( 120 cuvettes in total)

Optical diameter: 6mm

Material of reaction cuvette: Optical plastic

(3)Operation

Always rotates counterclockwise

At power on:Rotates and stops at the start position. At this time, the first reaction cuvette is located under
the first rinse nozzle.

At analysis: Initially operates the same as at power on, then circulates a cycle of half turn and one pitch
advance (covering 61 reaction cuvettes) followed by temporary stop. Each full turn takes about 18 seconds.
15
User Manual of CS-400 Auto-Chemistry Analyzer

At resetting: Make the same action as at power on.

(4)Rinsing

At position No. 45 of both reagent disks, set a bottle containing CS-anti-bacterial phosphor-free detergent
and open its cap. Carry out “ rinsing reaction cuvette” in the “maintenance” window, all the reaction
cuvettes will be rinsed. It is usually rinsed by CS-alkaline detergent in the front-left of instrument, so daily
maintenance for reaction cuvettes is unnecessary.

(5)Operation check

Single-click the “ Maintenance” key, select “ mechanism movement checkup”, and input the check times.
Click “ Execute“. If abnormal exist, instrument will issue alarm.

(6)Demounting

Reaction disk: Remove the rinse mechanism from above the reaction disk and completely loosen the fixing
knob at the center. The reaction disk can be lifted out. For mounting, align the guide pin of instrument with
the guide hole of the reaction disk and tighten the fixing screw.

Reaction cuvette: Remove the cuvette setscrew and pull up the reaction cuvette block by holding the
handgrip, and the reaction cuvette can be removed from the reaction disk.

Note:Keep reaction cuvette immersed in purified water. Besides, if the instrument will be not used more than 3
days, reaction cuvette should be removed and immersed in purified water.

1.4.2.6 Incubation bath

! Warning:

● Keep the cleanness of purified water in incubation bath, or it may effect the test precision.

● When instrument startup or rinsing incubation bath, make sure there is enough CS-anti-bacterial
phosphor-free detergent at No.45 position.

(1)Function

Keep the reaction solution in the reaction cuvette at a constant temperature.

(2)Operation

At power on: Automatic exchanges the constant temperature water once, the CS-anti-bacterial phosphor-free
detergent in position No.45 of both reagent disks is added in incubation bath.

At analysis: Incubation bath water is circulating. Instrument may automatically supply water when water
shortage comes in operation process.

Exchange water: In “maintenance” window, select “rinsing incubation bath” , and then the constant
temperature water may exchange, and then add 6ml CS anti-bacterial phosphor-free detergent in incubation
bath water.

Note: After running for 24 hours, instrument may require “incubation bath water exchange”, please carry out
“ Rinsing incubation bath”.

16
User Manual of CS-400 Auto-Chemistry Analyzer

1.4.2.7 Stirring mechanism

! Warning:

● Please don’t touch stirring mechanism When operate, or it may cause body injury or instrument damage.

(1) Function

Stirring the reaction solution in each reaction cuvette.

(2) Operation

At power on: Move to the side of reaction cuvette and then stops above the rinsing bath, move to the side
of reaction cuvette again, and then stops above the rinsing bath.

At analysis: The mechanism descends, rotates, rises and stops between two locations: reaction cuvette and
stirring rod rinsing bath.

Stirring is carried out after each addition of reagent 1(R1),reagent 2(R2), reagent3(R3), reagent 4(R4). R1
and R4 use Stirring rod mechanism 1. R2 and R3 use Stirring rod mechanism 2.

(3) Automatic rinsing

Automatic rinsing of stirring rod: when stirring rod descends into stirring rod rinsing bath, mechanism may
automatically rotate and washes the stirring rod with purified water.

Sampling finishing: Stirring rod is stirring in reaction cuvette in which detergent is added, thus rinse the
stirring rod.

(4) Operation check

Single-click the “maintenance” key, select “ mechanism operation check”, and input the check times. Click
“ Execute”. If abnormality exists, instrument will issue alarm.

1.4.2.8 Reaction cuvette rinsing mechanism

! Warning:

● Please don’t touch the rinsing mechanism when operate, or it may cause body injury or instrument damage

● Avoid directly contact with body, or it may cause infection. Please adopt protective measure. In case of
skin contact, flush the area with water, rinse immediately with plenty of water and seek medical advice.

(1)Function

Eliminates the reaction solution, rinse the reaction cuvette, injects and eliminates purified water which
used for test cell blank

(2 ) Composition of rinsing nozzle

17
User Manual of CS-400 Auto-Chemistry Analyzer

F G F
C D A B A B B
E

Rotational
Rotational direction of 测4 4times
次杯空白(1
cell blank次停止,3 次通过)
measurement.
Reaction disk 1 stop, 3 pass.

Figure 12 Arrangement of Rinse Nozzles

Above figure shows that seven steps are needed when rinsing reaction cuvette. (Four times cell blank test is
added) , therefore, to finish rinsing one reaction cuvette, 11 steps are needed.

Step 1: Nozzle 1D assimilates reaction solution, then 1C discharges detergent to reaction cuvette.

Step 2: Nozzle 2B assimilate detergent from reaction cuvette, then 2A discharges purified water to reaction
cuvette.

Step 3: Nozzle 3B assimilates the purified water from reaction cuvette, then 3A discharges the purified
water to reaction cuvette.

Step 4: Nozzle 4B assimilates the purified water from reaction cuvette, and wipes the reaction cuvette at the
same time.

Step 5: Nozzle 5E discharges the purified water to reaction cuvette.

Step 6,7,8,and 9: Carry out the cell blank check measurement at the fifth nozzle. The reaction cuvette with
full purified water allows 4 times cell blank measurement. (1 time static measurement, 3 times
dynamic measurement when reaction cuvette passed by during reaction disk turning )

Step 10: Nozzle 6F assimilates the purified water, which has carried out the cell blank absorbency check.

Step 11: Nozzle 7 F assimilates the remaining water in reaction cuvette, and wipes the reaction cuvette at
the same time; Nozzle 7G only discharges the purified water in rinsing process before sampling,
rinse the nozzle tip.

Distribution of 11 rinsing nozzles:

A. For discharging purified water ………………………….2 nozzles

B. For assimilating rinse water…………………………….3 nozzles

C. For discharging detergent……..…………………………1 nozzle

D. For assimilating reaction solution ..……………………. 1 nozzle

E. For discharging purified water for cup blank……………1 nozzle

F. For aspirating purified water for cup blank……………….2 nozzles


18
User Manual of CS-400 Auto-Chemistry Analyzer

G. For discharging purified water ………………………..…1 nozzle

(3)Operation

At power on: Rises when already down

At analysis: Rinse the reaction cuvette and carry out the water blank test in the rotational direction of
arrangement of rinse nozzle in Figure.1-13.

(4)Operation check

Single-click the “maintenance” key, select “ mechanism operation check”, and input the check times. Click
“ Execute ”. If abnormality exist, instrument will issue alarm.

(5)Dismounting

The rinse mechanism can be shifted from the reaction disk by loosening the fixing screw at the head.

1.4.2.9 Reagent cooling system

(1) Composition and function:

Cooling system is composed of reagent cooling system and sample cooling system, which could cool the
reagent, Controls and Calibrator respectively.

(2)Specifications

Temperature: 5 -- 15℃

! Warning:

● Even the analyzing system is power off, cooling system is still at working status. The cooling system
only stop working when main power supply is cut off.

● The usage and storage of Reagent should be performed strictly according to user manual.

1.4.2.10 Optical system

Concave
Detector (340-750nm) diffraction
12 fixed wavelength grating

Light source
Reaction
lamp
cuvette

Figure 1-13 Photometer

19
User Manual of CS-400 Auto-Chemistry Analyzer

(1) Function

When the reaction disk rotates, the absorbance of purified water or reaction solution is measured in each
reaction cuvette. As above figure shows.

(2)Specification:

Carry out photometry with dual-wavelength or single-wavelength at wavelengths: 340nm、380nm、405nm、


450nm、480nm、505nm、546nm、570nm、600nm、660nm、700nm、750nm.

Wavelength accuracy: ±2nm

Measuring range: 0 -3.3 Abs

Spectral bandwidth: FHW 8 to 10nm

Detector: Silicon photodiode

Light source: 12V, 20W halogen lamp

20
User Manual of CS-400 Auto-Chemistry Analyzer

1.5 Instrument Symbol

Symbol Meaning

To perform as the instruction under the symbol,


emphasize the important information and special contents.

To perform as the instruction under the mark,


or it may cause biological infection

AC symbol

Only diagnostic use

Storage at

Batch code

Use by

Serial number

Measurement Control

Manufacture by

Grounding terminal

Table 1-2 Instrument symbol

21
User Manual of CS-400 Auto-Chemistry Analyzer

Chapter 2 Function and Measuring Principle


The function and measuring principle is composed of mechanism movement principle and analyzing assay.

2.1 Mechanism movement principle

CS-400 Auto-Chemistry Analyzer consists primarily of the sample disk, sampling mechanism, reagent
disk, reagent pipeting mechanism, reaction disk, reaction bath, rinsing mechanism and photometer.
Operation of each mechanism is explained according to figure 2-1:
After starting, instrument carries out resetting first, rotates the R1 reagent disk and R2 reagent disk to
Position 1, then rotate them for 360 degrees, the R1 reagent probe, R2 reagent probe, sample probe, R1
stirring rod, R2 stirring rod are all stopped at the upper side of their rinsing bathes respectively.
Upon pressing the start key, the rinse mechanism starts rinsing from reaction cuvette No1. The reaction
disk rotates by 22 patches and stops temporarily, and then rotates by 37 patches and stops, sequently, 2 more
patches and then stops. This sequence is carried out again to cover one full turn plus 2 patches (18 seconds).
Cell blank is measured when the reaction cuvette passes through the photometric unit during rotation of the
reaction disk. The measured value of cell blank becomes the reference value for the subsequent absorbance
measurement. The liquid in the reaction cuvette is assimilated through the nozzle of rinse mechanism.
After rinsing the reaction cuvette for 3 minutes (the reaction disk rotates 10 circles), the sampling
mechanism begins to work when reaction disk rotates the 11th turns, and the sample probe moves above of the
sample cup and decends into the cup. Since the sample probe comprises a liquid level sensor, the probe stops
descending when its tip enters the sample. A set volume of sample is assimilated with the sample pipettor.
Next, the sample probe moves to the top of No.1 reaction cup and descends until its tip reaches the bottom of
the cuvette, where the sample is discharged. The sample probe further moves to the inside of probe rinsing
bath, where its inner and outer walls are washed with purified water.
On the other hand, the reagent pipetting mechanism assimilates reagent 1 (with R1 probe), while the
reaction disk rotates by 22 patches, stops temporarily and the rotates by 37 patches after the fore mentioned
sampling mechanism has discharged the sample into the reaction cuvette. When the reaction disk stops
temporarily after the above rotation, reagent 1 is discharged into the reaction cuvette. And when the reaction
disk stops temporarily after rotating by 2 patches, the R1 stirring mechanism mixes the sample and reagent.
The reagent pipetting mechanism rotates to the top of reagent bottle in the specified channel and descends
while assimilates and discharges reagent. A set volume of reagent is assimilated, then the reagent probe moves
to the top of reaction cuvette and discharges the reagent, followed by returning to the rinsing bath and
washing of its inner and outer walls. Add reagent R1 and start photometry. Measurement is made when each
reaction cuvette passes through the optical path during rotation of the reaction disk. The reaction disk
rotates15 turns plus 22 patches to the reagent 2 pipetting position, where reagent 2 is added with R2 probe.
Reagent 2 is stirred with the R2 stirring mechanism after the reaction disk rotates by 16 turn plus 2 patches
and stops temporarily, and then rotates 22 patches and stops temporarily.
After 26 turns plus 37 patches , reaction disk comes to the reagent 3 sampling position, and reagent
probe 2 assimilates reagent 3. R2 stirring rod begins to work after 2 more patches of reaction cuvette.
After 41 turns plus 37 patches , reaction disk comes to the reagent 4 sampling position, and reagent probe
1 assimilates reagent 4. R1 stirring rod begins to work after 2 more patches of reaction cuvette
After the lapse of about 15 minutes (60 circles), the first cup measurement is over, and the reaction
solution in No. 1 reaction cup is assimilated with the nozzle of cuvette rinsing mechanism, which will
discharge cleaning liquid and water into the reaction cup to rinse the cup. The instrument stops and goes to
22
User Manual of CS-400 Auto-Chemistry Analyzer

standby status when the last reaction cuvette has been washed, and goes to the the cell blank status. The
reagent probe and sample probe will be rinsed respectively by their own detergents added by themselves after
every test.
Reagent combination (R1, R2, R3, R4) of each item and reaction time (3~15 mins) is set in the “Analytical
parameter” form in “System setup”

2.1.1 Operating Position

Operating principle of 120 reaction cuvette is showed as below:

Rinsing mechanism Reset point


Position No. of reaction cuvette.

光电检测位
63,
Photoelectric
detection

62,R1 stirring
Sop and wipe 62,R1 stirring
rod
position
搅拌探针
63,R4 stirring
position
Sam. probe
Sample probe

60 R1 Pipetting
R1, R4 Probe
61 R4 Pipetting

Reference position
number

Reagent probe 2,3 Stirring probe


Reagent probe Stirring

Figure 2-1 Each mechanism operation position

2.1.2 Analytical flow

The analytical flow of CS-400 Auto-Chemistry Analyzer is show in figure 2-2:

23
User Manual of CS-400 Auto-Chemistry Analyzer

Table 2-1 Analytical flow

2.1.3 Photometric Features

This instrument adopts the whole reaction monitoring system, which intermittently measures the absorbance of
reaction solution for a reaction time of 15 minutes.

The reaction disk rotates 1 turn plus 2 pitches in about 18 seconds and during this time the absorbance is
measured for all of 120 reaction cuvettes which go across the optical axis of the photometer. For each reaction
cuvette, measurement is made10 times in a reaction time of about 3 minutes. 13 times measurement is made
during 4 minutes (13 photo metric point.). 16 times measurement is made during 5 minutes (16 photo metric
point33 times measurement is made during 10 minutes (33 photo metric point.) .49 times measurement is made
during 15 minutes (49 photo metric point.)

CS-400 multi-wavelength photometer condenses the white light emitted from the light source lamp through the
lens, passes the condensed light through the reaction cuvette and separates the light with the concave diffraction
grating. The separated respective wavelength components are simultaneously received on the 12 fixed detectors
and amplified by 12 amplifiers, then logarithmically converted to obtain the absorbance or difference in
absorbance. In 2 wavelengths photometry, concentration is measured by the value of the difference of dominant
wavelength and complementary wavelength. This means that the photometer features a correcting effect for
lipemia, hemolysis and icterus of sample and has a compensating effect for fluctuation in source voltage, thus
realizing stable measurement.

2.2 Assay mode

The assay mode of Auto-Chemistry Analyzer is based on the Beer-Lambert law that the material selective
absorption light

The main principle is: When monochromatic light with specific wavelength passes through the cuvette with
sample, the monochromatic light absorbency and sample liquid concentration are varies directly as the distance

24
User Manual of CS-400 Auto-Chemistry Analyzer

which is passed through sample liquid by light:

I
A = lg(1/T)= lg( 0 )= ε b c
It

A - Absorbency of the light when passing through liquid

T - Transmitted intensity and incident intensity ratio: transmittance It/I0;

I0 - Incident intensity

It - Transmitted intensity

ε - Molar absorption coefficient of solution(ml×mmol 1×cm 1);


- -

c - Mol concentration of the solution(mmol/ml);

b - Solution layer thickness(cm);

Solution layer thickness (b): Optical path, which is fixed by instrument. Molar absorption coefficient (ε) is the
correlation coefficient of the wavelength, solution and solution temperature. Linear relationship is displayed
between solution thickness and absorbency when in stable temperature and single wavelength(ε value is given
on the reagent bottle by factory)

If the sample liquid adequate distribution, interaction between liquid and incidence monochromatic light only
happens during absorbing process. No fluorescence, disperse and photochemical appear. No interaction between
substances in the solution while absorbing process. The absorbency possess conducts nature, and this condition
conforms to the Beer-Lambert law

2.2.1 Assay mode variety


As to how to set the assay parameter and standard liquid parameter, please refer to user manual. Assay mode is
showed as table 2-2:

Method Item Photometry point Cell blank Formula Note

25
User Manual of CS-400 Auto-Chemistry Analyzer

1-point L– 0– 0– 0 B1 + B 2 + B 3 AL + AL −1
Assay 3
1<L≤49 2
t :time
(minute)
2-point L– M– 0- 0 B1 + B 2 + B 3 ( AM + AM −1 ) − k ( AL + AL −1 )
between
Rate Assay 3
1<L<M≤49 2 photometry
point L、m

AM + AM −1 AL + AL −1
2-point L– M– 0– 0 B1 + B 2 + B 3 −
assay 2 2
1<L<M≤49 3
t
L– M – 0 - 0
B1 + B 2 + B 3
First half 1<L<M≤49 △A(M-L)
1-point & 3
L +2<M
Rate Assay
Second L– 0 – 0 – 0 B1 + B 2 + B 3 AL + AL −1
half 1<M<N≤L<P<Q≤49 3 2
M–N–P–Q
Rate A B1 + B 2 + B 3
1<M<N≤L<P<Q≤49 △(AQ- P)-k△(AN -M)
Assay 3
M+2<N,P+2<Q

Rate A
Assay with
Serum L– 0 – 0 – 0 B1 + B 2 + B 3 AL + AL −1
Index 3
Measurem
1<L≤M<N≤49 2
ent.

M–N–0–0 B1 + B 2 + B 3 ( AN + AN −1 ) − k ( AM + AM −1 )
First half
1<L≤M<N≤49 3 2
Rate B
Assay P-N ( different wavelength from the
L– M – 0 - 0
(mode 1 ) Second B1 + B 2 + B 3 Half) Two
3≤L<M<N<P≤49
half 3 △ A(P-N)–k △ A(M-L) ( same conditions
L +2<M
wavelength as the second half)
N–P–0–0
B1 + B 2 + B 3
First half 3≤L<M<N<P≤49 △A(M-L)
3
Rate B N+2<P
Assay
(mode 2 ) L– M – 0 – 0
Second B1 + B 2 + B 3
3≤L<M<N<P<Q<R≤49 △A(R-Q)– k△A(P-N)
half 3
L +2<M

Table 2-2 Assay mode table

Explanation of symbols:

L,m,n,p,q,r : Photometric points

Rn : Volume of nth reagent ,n=1 to 4

26
User Manual of CS-400 Auto-Chemistry Analyzer

SB : Stopped cell blank

B1,B2,B3 : Passed cell blanks

Ax : Absorbance at photometric point x

△A(m-L) : Change in absorbance per minute between photometric points L and M

k : Liquid volume correction factor

a
S + ∑ Rj
j =1
k= b
S + ∑ Ri
i =1

S : Sample volume

Rj 、Ri a: No. of reagents with correction (at Al measurement)

B: No. of reagents without correction (at Am measurement)

Note 1: The 5 th Photometric point won’t be Stirred after adding reagent 2. Stirring when the reaction disk
pauses after rotates one circle plus 2 pitches

Note 2: liquid in the reaction cuvette should be more than or equal to 150 ul, less than or equal to 450ul.

Note 3: Do input 0 if the photometric point is not used.

(1) 1-point Assay


Endpoint assay in which absorbance is measured at a designated photometric point (specific time point when
reaction reach balance) after addition of reagents. Figure 2-2 explains the 1-point assay.
Absorbance

Cell blanks
(B1+B2+B3)/3 R2~R4

R1
A L + A L −1
S 2

SB B1 B2 B3

Time
Figure 2-2 1-point Assay

(a) Photometric point : 【L】-【0】-【0】-【0】 (1< L ≤ 49)

(b) Calculation of absorbance

27
User Manual of CS-400 Auto-Chemistry Analyzer

The average of absorbance at measurement points Land L-1 is used.

AL + AL − 1
AX =
2

(c) Calculation of concentration

C X = {K × ( AX − B ) + C1 }× IFA + IFB

SB: Stopped cup blank

R1~R3: Passed cup blank

R1~R4: Reagent adding position

Cx: Concentration of standby sample

C1: Concentration of standard 1 solution(reagent blank)

K: Factor

B: Absorbance of blank

IFA and IFB: Instrument constants, representing slope and intercept

(d) Analytical items

TP, ALB, etc.

(2) 2-point Assay


Endpoint assay in which measurement is made twice at different points to obtain the difference in absorbance.
One point is measured as the action initial, the other point is measured when the action reach endpoint or
balance. The difference between the absorbance of two photometric points is used for calculation sample
concentration. Figure 2-3 explains the 2-point assay:
Absorbance

Cell blanks
(B1+B2+B3)/3 R2~R4
A m + A m −1
R1
2
S A L + A L −1
2

SB B1 B2 B3

Time
Figure 2-3 2-point Assay
(a) Photometric point : 【L】-【M】-【0】-【0】 (1≤ L ≤ 49)

28
User Manual of CS-400 Auto-Chemistry Analyzer

(b) Calculation of absorbance

The difference between the average of absorbance at measurement point m and m-1 and that at
measurement point l and l-1 is used.

AX=
( Am + Am − 1) − k ( AL + AL − 1)
2

a
S + ∑ Rj
j −1
k= b
S + ∑ Ri
i −1

a: No. of reagents at Al measurement

b: No. of reagents at Am measurement

(c) Calculation of concentration

C X = {K × ( AX − B ) + C1 }× IFA + IFB

SB: Stopped cup blank

R1~R3: Passed cup blank

R1~R4: Reagent adding position

Ax: the deference between the photometric point M and L

Cx concentration of standby sample

C1: Concentration of standard 1 solution(reagent blank)

K: Factor

B: Absorbance of blank

IFA and IFB: Instrument constants, representing slope and intercept

(d) Analytical items

CRE, etc.

(3) 2-point Rate Assay


Measurement is made twice at different measurement points (The two point are neither measured initial nor
endpoint) to determine the change in absorbance per minute in order to calculate sample concentration. For
check of reaction limit level, refer to Figure 2-4:
Absorbance

Reaction limit level


Am-1 Am
29

R2~R4
Cell blanks
User Manual of CS-400 Auto-Chemistry Analyzer

Figure 2-4 2-point Rate Assay

(a) Photometric point : 【L】-【M】-【0】-【0】 (1< L <M≤ 49)

(b) Calculation of absorbance

The difference between the average of absorbance at measurement points M and M-1 and that at
measurement points L and L-1, then divide the result by time.

( Am + Am − 1) ( AL + AL − 1)

AX= 2 2
t
t: Time (minute) between measurement points L and M

(c) Calculation of concentration

C X = {K × ( AX − B ) + C1 }× IFA + IFB

SB: Stopped cup blank

B1~B3: Passed cup blank

R1~R4: Reagent adding position

Ax: Average change in absorbance per minute between measurement points L and M

Cx: Concentration of standby sample

C1: Concentration of standard solution 1(reagent blank)

K: Factor

B: Absorbance of blank

IFA and IFB: Instrument constants, representing slope and intercept.

(d) Analytical items

30
User Manual of CS-400 Auto-Chemistry Analyzer

BUN, CRE etc.

(4) Rate A Assay


Ordinary Rate Assay. The concentration or activity level is obtained from the change in absorbance between the
specified measurement points. Figure 2-5 explains the Rate A Assay.

S
Absorbance

R1
R2~R4

Cell blanks
(B1+B2+B3)/3

AL Am

Reaction limit level

SB B1 B2 B3

Time

Figure 2-5 Rate A Assay

(a) Photometric point : 【L】-【M】-【0】-【0】 (1<L <M≤ 49) L+2<m)

(b) Calculation of absorbance

The change in absorbance per minute between measurement point L and M is obtained by the least
squares method

AX=△A(L-m)

(c) Calculation of concentration

C X = {K × ( AX − B ) + C1 }× IFA + IFB

SB: Stopped cup blank

B1~B3: Passed cup blank

R1~R4: Reagent adding position

△A(L-m): Change in absorbance per minute between measurement point L and M

Cx: Concentration of standby sample

C1: Concentration of standard solution 1(reagent blank)

31
User Manual of CS-400 Auto-Chemistry Analyzer

K: Factor

B: Absorbance of blank

IFA and IFB: Instrument constants, representing slope and intercept.

(d) Analytical items

AST, ALT, etc.

(5) 1-point & Rate Assay


Two tests are analyzed by the endpoint assay and rate assay in a single channel. Test A is analyzed by the
endpoint assay in the first half of the specified reaction time and test B is analyzed by the rate assay in the
second half. Figure 2-6 explains the 1-point & rate assay.

Figure 2-6 1-point& Rate Assay

(a)Input of “analytical parameter”

Correct input of test A and test B is required respectively

(Text A): Photometric point : 【L】-【0】-【0】-【0】

Dual tests assay: designate B

(Text B) :Photometric point: 【M】-【N】-【P】-【Q】

1<m<n≤L<P<Q≤49

M+2<N,P+2<Q

(b) Calculation of absorbance

For test A, the average of absorbance at measurement points L and L-1 is used, and used for the test B is
the difference obtained by subtracting the change in absorbance per minute between measurement points

32
User Manual of CS-400 Auto-Chemistry Analyzer

L and M from that between measurement points N and P

AL + AL − 1
AA=
2

AB=△Ap×q-k△Am×n

a
S + ∑ Rj
j −1
K= b
S + ∑ Ri
i −1

a:△A(N-M): No. of reagents.

b:△A(Q-P) : No. of reagents.

(c) Calculation of concentration.

For each of tests A and B, calculation is made according to the following formula.

C X = {K × ( AX − B ) + C1 }× IFA + IFB

SB: Stopped cup blank

B1~B3: Passed cup blank

Rn: Reagent adding position

AA and AB : Each calculated absorbance of tests A and B

k: Liquid volume correction factor

S: Sample volume

Rj、Ri: Volume of each reagent

Cx: concentration of standby sample

C1: Concentration of standard solution 1(reagent blank)

K: Factor

AX: Calculated absorbance

B: absorbance of standard solution 1(reagent blank)

IFA and IFB: Instrument constants, representing slope and intercept.

In the rate assay, the change in absorbance per minute is obtained by the least squares method

(6) 3-point dual assay


Endpoint assay and endpoint assay without sample blank are carried out simultaneously in the same cup. The
first half reaction time is used for test A, and the left half is used for test B. Figure 2-7 explains the rate A assay
33
User Manual of CS-400 Auto-Chemistry Analyzer

with serum index measurement.

Absorbance
Cell blank

Time

Figure 2-7 3-point dual assay

(a)Input of analytical Parameters

Correct input of test A and test B is required respectively

(Test A)Photometric point : 【L】-【0】-【0】-【0】

Dual tests assay: designate B

(Test B) Photometric point: 【M】-【N】-【0】-【0】

1< L<M<N≤49

(b) Calculation of absorbance

The average value of absorbance at photometric points L and L-1 is for test A, and test B, the difference
between the average absorbance at photometric points N and N-1 and that of photometric points Mind
M-1.

AL + AL − 1
AA =
2
( AN + AN −1 ) − k × ( AM + AM −1 )
AB =
2
a
S + ∑ Rj
j =1
k= b
S + ∑ Ri
i =1

a:△ AM : No. of reagent when tested


b:△ AN : No. of reagent when tested
(c) Calculation of concentration
34
User Manual of CS-400 Auto-Chemistry Analyzer

C X = {K × ( AX − B ) + C1 }× IFA + IFB

SB: Stopped cup blank

B1~B3: Passed cup blank

Rn: Reagent adding position

AA and AB : Each calculated absorbance of tests A and B

k: Liquid volume correction factor

S: Sample volume

Rj、Ri: Volume of each reagent

Cx: concentration of standby sample

C1: Concentration of standard solution 1(reagent blank)

K: Factor

AX: Calculated absorbance

B: absorbance of standard solution 1(reagent blank)

IFA and IFB: Instrument constants, representing slope and intercept.

(7) Rate B Assay (mode 1)


Two tests are analyzed by the rate assay in a single channel. In the first half of reaction time, test A is measured,
and test B is measured in the second half. In the rate B assay, correction is possible with sample blank and for
an endogenous reaction. However, the method of correction of such reaction varies with measuring wavelength.
So this assay is categorized into modes 1 and 2 for easier explanation. Mode 1 is subdivided depending on
whether or not measuring wavelength is the same between test A and B. Figure 2-8 explains the rate B assay
(mode 1).

35
User Manual of CS-400 Auto-Chemistry Analyzer

Figure 2-8 Rate B Assay (mode 1)

When wavelength differs between tests A and B

(a) Input of analytical parameters

Respective entry is required for each of tests A and B.

(Test A) photometric point: 【L】-【M】-【0】-【0】

(Test B ) photometric point: 【N】-【P】-【0】-【0】

3≦L <m<n<p≦49。L+2<m、n+2<p

(b) Calculation of absorbance

Used for test A is the change in absorbance per minute between measurement point l and m, which is
obtained by least squares method. Used for test B is the change in absorbance per minute between
measurement points n and p, which is obtained by the same method, but do not carry out blank
correction.

△AA=△A(M-L)

△AB=△A(P-N)

(c) Calculation of concentration.

For each of tests A and B, calculation is made according to the following formula.

C X = {K × ( AX − B ) + C1 }× IFA + IFB

SB: Stopped cup blank

B1~B3: Passed cup blank

AA and AB : Each calculated absorbance of tests A and B

k: Liquid volume correction factor

S: Sample volume

Rj、Ri: Volume of each reagent

Cx: concentration of standby sample

C1: Concentration of standard solution 1(reagent blank)

K: Factor

AX: Calculated absorbance

B: absorbance of standard solution 1(reagent blank)

IFA and IFB: Instrument constants, representing slope and intercept.

36
User Manual of CS-400 Auto-Chemistry Analyzer

When wavelength is the same between tests A and B

(a) Input of analytical parameters

Respective entry is required for each of tests A and B.

(Test A) photometric point: 【L】-【M】-【0】-【0】

(Test B ) photometric point: 【N】-【P】-【0】-【0】

3≦L<m<n<p≦49。L+2<m、n+2<p

(b) Calculation of absorbance

Used for test A is the change in absorbance per minute between measurement points l and m, which is
obtained by least squares method. Used for test B is the difference obtained by subtracting the above
change for test A from the change in absorbance per minute between measurement points n and p, which
is obtained by the same method

△ AA =△A(L- M)

△ AB =△A(P- N)-k×△A(L- M)

a
S + ∑ Rj
j =1
k= b
S + ∑ Ri
i =1

a: △A(M - L)No. of reagents with correction

b: △A(P- Q)No. of reagents without correction

(c) Calculation of concentration

For each of tests A and B, calculation is made according to the following formula.

C X = {K × ( AX − B ) + C1 }× IFA + IFB

SB: Stopped cup blank

B1~B3: Passed cup blank

AA and AB : Each calculated absorbance of tests A and B

k: Liquid volume correction factor

S: Sample volume

Rj、Ri: Volume of each reagent

Cx: concentration of standby sample

C1: Concentration of standard solution 1(reagent blank)


37
User Manual of CS-400 Auto-Chemistry Analyzer

K: Factor

AX: Calculated absorbance

B: absorbance of standard solution 1(reagent blank)

IFA and IFB: Instrument constants, representing slope and intercept.

(8) Rate B Assay (mode 2)


Two tests are analyzed by the rate assay in a single channel. In the first half of reaction time, test A is measured,
and test B is measured in the second half. Correction of endogenous reaction is carried out by using the ratio of
change in absorbance within a time period after measurement of the first test in the first half of reaction time.
Figure 2-9 explains the rate B assay (mode 2).

Figure 2-9 Rate B Assay (mode 2)

(a) Input of analytical parameters

Respective entry is required for each of tests A and B.

(Test A) photometric point: 【L】-【M】-【0】-【0】

Dual tests assay: designate B

(Test B ) photometric point: 【N】-【P】-【Q】-【R】

3≦L<m<n<p≦49。L+2<m、n+2<R

(b) Calculation of absorbance

Used for test A is the change in absorbance per minute between measurement points l and m, which is
obtained by the least squares method. Used for test B is the difference obtained by subtracting the change
in absorbance per minute between measurement points n and p from that between measurement points q
and r, which is obtained by the same method

△ AA =△A(L- M)

△ AB =△A(P- Q)-k×△A(P- N)

38
User Manual of CS-400 Auto-Chemistry Analyzer

a
S + ∑ Rj
j =1
k= b
S + ∑ Ri
i =1

a: △A(P- N)No. of reagents

b: △A(R- Q)No. of reagents

(c) Calculation of concentration

For each of tests A and B, calculation is made according to the following formula.

C X = {K × ( AX − B ) + C1 }× IFA + IFB

SB: Stopped cup blank

B1~B3: Passed cup blank

AA and AB : Each calculated absorbance of tests A and B

k: Liquid volume correction factor

S: Sample volume

Rj、Ri: Volume of each reagent

Cx: Concentration of standby sample

C1: Concentration of standard solution 1(reagent blank)

K: Factor

AX: Calculated absorbance

B: absorbance of standard solution 1(reagent blank)

IFA and IFB: Instrument constants, representing slope and intercept.

2.2.2 Calibration Method


(1) Linearity Method (1-point linearity)
The absorbance and input K value of blank (or standard 1) is measured to prepare a working curve. Figure 2-10
explains the linear method.

39
User Manual of CS-400 Auto-Chemistry Analyzer

Absorbance

Concentration

Figure 2-10 1-point linearity

(a) Calibration Parameter input

Calibration type: 【1-point linear】

Calibration point: 【1】 (number of standard sample )

Span point: 【0】

(b) Input K factor in “calibration result” menu

Input K factor in the “calibration result ”

(c) Calculation of parameters for working curve

S1ABS (B): Change in absorbance per minute of blank (standard 1)

K: Input value.

C1: Concentration of standard 1(reagent blank ), input value.

(d) Calculation of concentration

C X = {K × ( AX − B ) + C1 }× IFA + IFB

Cx: Concentration of standby sample

AX: Calculated absorbance or change of absorbance per minute.

IFA and IFB: Instrument constants, representing slope and intercept.

(e) Applicable assay

1-point assay, 2-point rate assay, 2-point assay, 3-point assay, 1-point&rate assay, rate A assay and rate B
assay

(2) Linearity Method ( 2-point linearity )


40
User Manual of CS-400 Auto-Chemistry Analyzer

Blank (or standard 1) and standard sample (standard 2) are measured to prepare a linear working curve Figure
2-11 explains the linear method.

Absorbance

Concentration
Figure 2-11 2-point linearity

(a) Calibration Parameter input

Calibration type: 【2-point linearity】

Calibration point: 【2】 (number of standard sample )

Span point: 【2~6】

(b) Calculation of parameters for working curve

S1ABS (B): Absorbance or change in absorbance per minute of blank (standard 1)

K: Calculated from measured values and input values of blank (standard 1) and standard sample (standard
2)

C1: Concentration of standard 1(reagent blank)

C2: Concentration of standard 2

A2: Absorbance or change in absorbance per minute of standard 2.

C 2 − C1
K=
A2 − B

(c) Calculation of concentration

C X = {K × ( AX − B ) + C1 }× IFA + IFB

Cx: Concentration of standby sample

AX: Absorbance or change in absorbance per minute

IFA and IFB: Instrument constants, representing slope and intercept.


41
User Manual of CS-400 Auto-Chemistry Analyzer

(d) Applicable assay

1-point assay, 2-point rate assay, 2-point assay, 3-point assay, 1-point&rate assay, rate A assay and rate B
assay

(3) Linearity Method (Multi-point linearity)


Blank (or standard 1) and standard samples (standard 2 and standards 6) are measured and linear working curve.
Figure 2-12 explains the linear method.

Absorbance

Concentration

Figure 2-12 Multi-point linearity

(a) Calibration Parameter input

Calibration type:【 multi-point linearity】

Calibration point: 【3-6 】(number of standard sample)

Span point: 【3-6】

(b)Calculation of parameters for working curve

S1ABS (B):Linear primary regression intercept for absorbance or change in absorbance per minute of
blank (standard)

K: Inverse number of working curve slope in the result of linear primary regression.

S1ABS and K values can be calculated by the formulas below:

X × Cr
S1ABS ( B) = A −
Y

Y
K = × 10 4 × 10 a
X

42
User Manual of CS-400 Auto-Chemistry Analyzer

( )( )
n
X : ∑ Cri − Cr × Ai − A
i =1

( )
n
Y : ∑ Cri − Cr
2

i =1

⎛ n ⎞
A : ⎜ ∑ Ai ⎟ / n
⎝ i =1 ⎠

⎛ n ⎞
Cr : ⎜ ∑ Cri ⎟ / n
⎝ i =1 ⎠

A1,A2: Each measured absorbance in duplicate measurement of standard(1)

n: No. of standards (N) ×2

Cri: Concentration of standard (i)

(c)Calculation of concentration

C X = {K × ( AX − B ) + C1 }× IFA + IFB

Cx: Concentration of standby sample

AX: Absorbance of sample or its change per minute.

IFA and IFB: Instrument constants, representing slope and intercept.

(d) Applicable assay

1-point assay, 2-point rate assay, 2-point assay, 3-point assay, 1-point&rate assay, rate A assay and rate B
assay

(4) Logit-log 3P (Non-linearity Method)


This is applied to a working curve in which the absorbance converges as the concentration increase. Figure 2-13
explains the non-linearity method.

43
User Manual of CS-400 Auto-Chemistry Analyzer

Figure 2-13 Logit-log3P

(a) Calibration Parameter input

Calibration type: 【Logit-log3P】

Calibration point: 【3-6】(number of standard sample)

Span point: 【0】span calibration invalid

(b) Calculation of parameters for working curve

B: the absorbance or approximate value measure of the absorbance change per minute when CX
approaches ∞.

K: blank (standard 1) absorbance or value calculate by the approximation formula of the absorbance
change per minute subtraction B

a: Constants in approximation formula. Automatically calculated.

B,K,a are displayed on the Calibration List screen.

(c) Calculation of concentration.

C X = (C + C1 ) × IFA + IFB

K
AX = B +
1 + aC )

1 ⎧ K − ( AX − B ) ⎫
C= ×⎨ ⎬
a ⎩ AX − B ⎭

Cx: Concentration of standby sample

C1: Blank concentration.

AX: Absorbance of sample or its change per minute.

K: Constants in approximation formula. The more Cx approaches ∞, AX approaches B

When K<0, AX≤B+K or K>0,When AX≥B+K,C=C1

IFA,IFB Instrument constants, representing slope and intercept.

44
User Manual of CS-400 Auto-Chemistry Analyzer

(d) Calculation of SD value

∑∑ (A )
N 2
, 2
IJ − A1
i =1 j =1
SD =
2N − 3

(N=3~6、j=1or 2)

(Aij-Ai’): Difference between approximate absorbance Ai’ and measured value Aij or A12. Each standard
sample is measured in duplicate so the number as measurement points Aij is 12 at maximum

(e) Applicable assay

1-point assay, 2-point rate assay, 2-point assay, Rate A assay.

(5) Logit-log4P (Non-linearity method 2)


It is applied to a working curve in which the absorbance converges as the concentration increases. Figure 2-14
explains the non-linearity method.
Absorbance

Concentration

Figure 2-14 Logit-log4P

(a) “Calibration Parameter” input

Calibration type: 【Logit-log4P】

Calibration point: 【4-6】 (number of standard sample)

Span point: 【0】

(b) Calculation of parameters for working curve

B: approximation for the absorbance or its change per minute when CX approaches ∞.

K: blank (standard 1) absorbance or value calculate by the approximation of the absorbance change per
minute subtraction B

45
User Manual of CS-400 Auto-Chemistry Analyzer

a ,b : Constants in approximation formula. Automatically calculated.

B,K,a are displayed on the Calibration List screen.

(c) Calculation of concentration.

C X = (C + C1 ) × IFA + IFB

K
AX = B +
1 + aC b )

1 ⎧ K − ( AX − B ) ⎫
C = b× ×⎨ ⎬
a ⎩ AX − B ⎭

Cx: Concentration of standby sample

C1: Blank concentration.

AX: Absorbance of sample or its change per minute.

K: Constants in approximation formula. The more Cx approaches ∞, AX approaches B

When K<0, AX≤B+K or when K>0,AX≥B+K C1=0

IFA,IFB Instrument constants, representing slope and intercept.

(d) Calculation of SD value

∑∑ (A )
N 2
, 2
IJ − A1
i =1 j =1
SD =
2N − 4

(N=4~6、j=1or 2)

(Aij-Ai’): Difference between approximate absorbance Ai’ and measured value Aij or A12. Each
standard sample is measured in duplicate so the number as measurement points Aij is 12 at maximum

(e) Applicable assay

1-point assay, 2-point rate assay, 2-point assay, Rate A assay.

(6) Logit-log5P (Non-linear method 3)


There is no distinct difference between the working curves prepared by the non-linear method 2 and 3. However,
in some cases, the non-linear method 3 allows more accurate approximation because this method has one more
calculation because this method has one more calculation parameter than the non-linear method 2. Figure 2-15
explain the non-linear method 3.

46
User Manual of CS-400 Auto-Chemistry Analyzer

Absorbance

Concentration

Figure 2-15 Logit-log5P

(a) “Calibration Parameter” input

Calibration type : 【Logit-log5P】

Calibration point: 【5-6】( number of standard sample )

Span point: 【0】Span point calibration invalid.

(b) Calculation of parameters for working curve

B: approximation for the absorbance or its change per minute when CX approaches ∞.

K,a,b,c: Constants in approximation formula. Automatically calculated.

B,K,a ,b,c :are displayed as S1ABS,K,A,B,C on the Calibration List screen.

(c) Calculation of concentration.

AX − B
a+b·lnC+c·C-ln{ }=0
K − (A − BX )

Calculate C according to the Newton approximation formula.

C X = (C + C1 ) × IFA + IFB

K
1 + exp× (− a − b × ln C − c × C )
AX=B+

Cx: Concentration of standby sample

C1: Blank concentration.

AX: Absorbance of sample or its change per minute.

K: Constants in approximation formula. The more Cx approaches ∞, AX approaches B

47
User Manual of CS-400 Auto-Chemistry Analyzer

When K<0, AX≤B or when K>0,AX≥B,C=0

IFA,IFB Instrument constants, representing slope and intercept.

(d) Calculation of SD value

∑∑ (A )
N 2
, 2
IJ − A1
i =1 j =1
SD =
2N − 4

(N=5~6、j=1 or 2)

(Aij-Ai’): Difference between approximate absorbance Ai’ and measured value Aij or A12. Each
standard sample is measured in duplicate so the number as measurement points Aij is 12 at maximum

(e) Applicable assay

1-point assay, 2-point rate assay, 2-point assay, Rate A assay.

(7) Exponential function method (Non-linear method)


Unlike non-linear methods 1,2 and 3.Exponetial function method prepares a working curve in which the
absorbance disperses as the concentration increases. Figure 2-16 explains the exponential function method.
Absorbance

Concentration

Figure 2-16 Exponential function method

(a) Calibration Parameter input

Calibration type: 【exponential function】

Calibration point: 【5-6】(number of standard sample)

Span point: 【0】Span point calibration invalid.

(b) Calculation of parameters for working curve

B: approximation formula for the absorbance or its change per minute of blank (standard 1)

K,a,b,c: Constants in approximation formula. Automatically calculated.


48
User Manual of CS-400 Auto-Chemistry Analyzer

B,K,a ,b,c :are displayed as S1ABS,K,A,B,C on the Calibration List screen.

(c) Calculation of concentration.

AX=B+K × exp{a × (ln C ) + b ×(ln C ) + c × (ln C )


2 3
}
⎛ A − B⎞
a × (ln C ) + b × (ln C ) + c × (ln C ) -ln ⎜ X
2 3
⎟=0
⎝ K ⎠

Calculate C according to the Newton approximation formula.

C X = (C + C1 ) × IFA + IFB

Cx: Concentration of standby sample

C1, C2~CN: Blank and standard concentration.

AX: Absorbance of sample or its change per minute.

When K>0, AX≤B or when K<0,AX≥B,C=0

IFA,IFB Instrument constants, representing slope and intercept.

(d) Calculation of SD value

∑∑ (A )
N 2
, 2
IJ − A1
i =1 j =1
SD =
2N − 5

(N=5~6、j=1 or 2)

(Aij-Ai’): Difference between approximate absorbance Ai’ and measured value Aij or A12. Each
standard sample is measured in duplicate so the number as measurement points Aij is 12 at maximum

(e) Applicable assay

1-point Assay, 2-point Rate assay, 2-point Assay, Rate A assay.

(8) Spline function method (Non-linear method)


In this method, line is connected in each section so as to form a curve as a whole. Since each section is
smoothed including the error in measured value, more accurate approximation is possible than the polygonal
line approximation. Figure 2-17 explains this method.

49
User Manual of CS-400 Auto-Chemistry Analyzer

Absorbance

Figure 2-17 Spline function method

(a) “Calibration Parameter” input

Calibration type: 【Spline function】

Calibration point: 【5-6】(number of standard sample)

Span point: 【0】Span point calibration invalid.

(b) Calculation of parameters for working curve

A(1),b(1),c(1),d(1): Constants in approximation formula, l=1~N

In the “calibration” menu, S1ABS show as a (1) (intercept of absorbance axis ).

(c) Calculation of concentration.

( ) + d (I) × (C
3
AX=a(I)+b(I) × (C X − C (1) + c(I)) × C X − C (1) - C(1) )
2
X

f × (C X − C ( I )) = a × ( I ) + b × ( I ) × (C X − C ( I )) + d × ( I ) × (C X − C ( I )) 2 + d ( I ) × (C X − C ( I )) 3 − AX
Calculate C according to the Newton approximation formula.

C X = (C + C1 ) × IFA + IFB

Cx: Concentration of standby sample

C1~CN: Blank and standard concentration.

AX, A2~AN: Absorbance of sample and standard or its change per minute.

IFA, IFB Instrument constants, representing slope and intercept.

(d) Calculation of SD value

50
User Manual of CS-400 Auto-Chemistry Analyzer

∑∑ (A )
N 2
, 2
IJ − A1
i =1 j =1
SD =
2N − 4

(N=5~6、j=1 or 2)

(Aij-Ai’): Difference between approximate absorbance Ai’ and measured value Aij or A12. Each
standard sample is measured in duplicate so the number as measurement points Aij is 12 at maximum

(e) Applicable assay

1-Point Assay, 2-Point Rate Assay, 2-Point Assay, Rate A Assay.

(9) Polygon method (Non-linear method)


The range between standard samples 1 to 6 is subject to approximation in consideration of measured values
across them and line is connected in each section so as to form a curve as a whole. Since each section is
smoothed including the error in measured value, more accurate approximation is possible than the polygonal
line approximation. Figure 2-18 explains the polygonal line method

Absorbance

Concentration

Figure 2-18 Polygon method

(a) “Calibration Parameter” input

Calibration type: 【polygon】

Calibration point: 【3-6】(number of standard sample)

Span point: 【0】Span point calibration invalid.

(b) Calculation of parameters for working curve

S1ABS is the two measure value (absorbance or it’s change) of STD (1)

C 2 − C1
K=
A2 − B

B: Absorbance or it’s change of standard (1)

51
User Manual of CS-400 Auto-Chemistry Analyzer

A2: Absorbance or it’s change of standard (2)

C1: standard (1) concentration (input value)

C2: standard (2) concentration (input value)

(c) Calculation of concentration.

C X = {K N × ( AX − AN ) + C N }× IFA + IFB

(e) Applicable assay

1-Point Assay, 2-Point Rate Assay, 2-Point Assay, Rate A assay.

(10) Isozyme Method


In a sample in which 2 different isozymes coexist, a reagent containing inhibitor may fail to completely
suppress the activity of either isozyme alone. In this case, isozyme activity is determined from total activity and
activity residual rate. Each working curve for total activity and isozyme activity are prepared by using 2
channels. If the activity of a specific isozyme of the coexistent two can be suppressed completely by using
monoclonal antibody, etc. this calibration method is unnecessary. As figure 2-19, 2-20 shows:
absorbance

absorbance

A2 STD(2)

A3 A3’
STD(3)
STD(3)
AF AM’ Sample
Sample
A4 A4’
STD(4) STD(4)

B STD(1) B’ STD(1)

C1 CF C2 concentration C1 CM’ concentration

Figure 2-19 Isozyme Method Figure2-20 Isozyme Method

(a) calibration principle

Isozymes method uses 2 reagent position. The total activity Cf is supposed to be calculated first with a
certain reagent in the isozyme P position. Calculate the activity Cm or Cn of isozymes M or N with a
reagent which can suppress N or M substance.

The isozymes M inhibitor testing isozymes M.

Generally speaking isozymes N inhibitor cannot suppress the activity of isozymes N completely, and the
activity of isozymes M is suppressed in a degree at the meantime.

The isozymes Method uses 2 channels to test the total activity and standard cost of isozymes M,N, K
value is tested by total activity channel, both two channels are used.

52
User Manual of CS-400 Auto-Chemistry Analyzer

M and N activity residual rate of M and N is calculated by inhibitor (ratio between absorbance of
standard 3 and standard 4 in two channel.).Calculate the total activity and isozymes M activity upon
above method. The activity of isozymes N is observed by calculation.

(b) Input of parameters:

Reagent and standard samples.

Reagent: Reagent for measurement of total activity, reagent for measurement of isozyme activity.

Standard sample: Standard sample F (containing both isozymes M and N), standard sample M
(containing isozyme M), standard sample N (containing isozyme N)

Reagent position: isozymes M and N are placed in different positions

(d) Entry on Chemistry parameters screen.

Make entry for each of the isozyme P and Q channels as shown in below Table:

Con. and Pos.of F Activity Con.and Pos. of M isozyme


Calibrator
(Isozyme P) (Isozyme Q) (Isozyme Q)

(concentration) (position) (concentration) (position)

Calibrator(1) Blank concentration …………………….. [S1] Blank concentration……………..[S1]

Calibrator(2) Concentration value of calibrator F………[S2] 0…………………………………….0

Calibrator(3) 0 …………………[S3] (Isozyme M Calibrator) 0………..[S3](Isozyme M Calibrator)

Calibrator(4) 0………………….[S4] (Isozyme N Calibrator) 0………. [S4](Isozyme N Calibrator)

Table 2-3

S1 to S4 are calibrator code numbers of calibrator 1 to calibrator 4 respectively. Enter the same calibrator
code number in both channels for each of calibrator 1,3,4. Set standard sample of isozyme M at position
Calibrator 3 and that of isozyme N at position Calibrator (4). It is unnecessary to enter the concentrations
of calibrator 3,4,2 for the isozyme Q channel. The channel number M of isozyme M is specified in the
isozyme P channel. But it need not be specified on the isozyme Q side. Unless M is entered, Isozyme
parameter error alarm occurs to disable analysis

(c) Working curve for total activity measurement channel (isozyme P) K factor is calculated by the
following formula:

C 2 − C1
K=
A2 − B

B: Absorbance of blank (standard 1) or its change per minute

A2: Absorbance of standard sample F (standard 2) or its change per minute

53
User Manual of CS-400 Auto-Chemistry Analyzer

C1: Activity of blank (standard 1 )

C2:Activity of standard sample F(standard 2 )

(d) Calculation of activity for total activity measurement channel (isozyme P)

CF={K·(AF-B)-C1}·IFA-IFB

C3={K·(A3-B)-C1}·IFA-IFB

C4={K·(A4-B)-C1}·IFA-IFB

CF 、C3 、C4: is sample, activity of standard 3, standard4; Ap 、A3 、A4: is absorbance or change of the
absorbance per minute of standard 3 , standard 4;IFA,IFB Instrument constants, representing slope and
intercept.

(e) Calculation of activity for isozyme measurement channel (Q)

CM’={K·(AM’-B’)-C1}·IFA-IFB

C3’’={K·(A3’- B’)-C1}·IFA-IFB

C4’={K·(A4’- B’)-C1}·IFA-IFB

CM’: Isozyme M activity of sample;AM’: Isozyme absorbance of sample or its change per minute;C3’ 、
C4’:each inhibited activity of standard 3 and 4;A3’、A4’分 Each inhibited absorbance of standard 3 and 4
or its change per minute;B: Absorbance of blank or its change per minute IFA,IFB Instrument constants,
representing slope and intercept.

(f) Calculation of activity residual rate

{K × (A 3 '-B) + C1}× IFA + IFB


α=
{K × (A 3 - B) + C1}× IFA + IFB

{K × (A 4 '-B) + C1}× IFA + IFB


β=
{K × (A 4 - B) + C1}× IFA + IFB
when C1=0, IFA=1,IFB=0

A3, -B,
α=
A3-B

A4,
-B,
β=
A4-B

(g) Calculation of isozyme M activity Cm

CM’=α×CM+β×CN

54
User Manual of CS-400 Auto-Chemistry Analyzer

CM − α × CM
CM= CF-CN= CF -
β

β×CM=β×CF -CM’+α×CM

(α-β)×CM= CM-β×CF

C M '− β × CF
CM=
(α − β )
(h) Applicable assay

1-Point Assay, 2-Point Rate Assay, 2-Point Assay, Rate A Assay.

Note: Suggest operator use 6 calibrators in Logit-log5P, exponential, spline non-linear calibration.

2.2.3 Calibration points


According to the number of calibration points, calibration can be effected in different ways. Blank calibration:
Only reagent blank (standard 1 ) is calibrated. Span calibration: Only one standard solution other than the
reagent blank is calibrated. 2-point calibration: Reagent blank and a single standard solution are calibrated.
Full-point calibration: All standard solutions specified on the Chemistry Parameters screen are calibrated. These
are selectively usable so as to meet your analytical purpose. Each calibration is explained below.

(1) Blank Calibration


Only reagent blank (standard 1) is calibrated. The previously measured working curve is corrected for a change
in absorbance or absorbance change rate (through calculation of S1ABS). Table 2-4 lists the calculation method
for each calibration type. (a) S1ABS calculation

Calibration type SIABS calculation

2 point linearity (A11+ A12)/2


Multi-point linearity {(AU+ A12)/2-(AU’+ A12’)/2}+SIABS’
Isozyme P (A11+ A12)/2
Isozyme Q (A11+ A12)/2
Logit-Log 3P {(A11+ A12)/2-(A11’+ A12’)/2}+SIABS’
Logit-Log 4P {(A11+ A12)/2-(A11’+ A12’)/2}+SIABS’
Logit-Log 5P (A11+ A12)/2
Exponential function (A11+ A12)/2
Spline function {(A11+ A12)/2- SIABS’}+a(I)
Polygon method (A11+ A12)/2

Table 2-4 S1ABS Calculation

A11、 A12:1st and 2nd absorbance values of standard (1) measured presently.

55
User Manual of CS-400 Auto-Chemistry Analyzer

A11’, A12’:1 st and 2 nd absorbance values of standard (1) measured previously.

SIABS’:Previous SIABS value

a(I):I=1~N,N representing the number of standard solution and factor of the curve(refer to 5.3.3(8)
Logit-Log 5P method)

This calibration method needs be specified for each test on the calibration test selection screen of Routine
job Menu.

Note*:Only blank calibration can be carried out in case 1 is entered for the number of standard solutions (K
factor method)

(b)Applicable calibration types

Linear (2-point), Linear (multi-point), Linear (1-point: K factor), Isozyme P, Isozyme Q, Logit-log 3P,
Logit-log4P, Logit-log 5P, Exponential, Spline

(2) Span Calibration


Only one standard solution other than reagent blank is calibrated. The standard solution which corresponds to
the span point entered on the Chemistry Parameters screen is measured and the previously measured working
curve is corrected for its slope. Table 2-5 lists the calculation method for each calibration type. This calibration
method need be specified for each test on the Calibration Test Selection screen.

(a) K factor and S1ABS calculation

Calibration type K factor calculation S1ABS

Two-point calibration (C2-C1)/(A2-S1ABS) Previous value


Full-point calibration (C2-C1)/(AN-S1ABS) Previous value

Table 2-5 S1ABS, K value calculation

C2:Concentration of standard (2)

C1:Concentration of standard (1)

CN:Concentration of standard N (N represents span point)

A2:Average of measured absorbance values of standard (2)

AN:Average of measured absorbance values of standard (N)

(b) Applicable assay

Linear (2-point), Linear (multi-point),

(3) 2-point calibration


Reagent blank and a single standard solution are calibrated. The standard solution and reagent blank, which
correspond to the span points entered on the Chemistry Parameters screen, are measured and the previously
measured working curve is corrected for S1ABS and slope. Table 2-6 explains the calculation method.

(a) K factor and S1ABS calculation

56
User Manual of CS-400 Auto-Chemistry Analyzer

Calibration type S1ABS Calculation K factor calculation

Two-point linearity (A11+ A12)/2 (C2-C1)/(A2-S1ABS)


Full-point linearity (AU+ A12)/2 (CN-C1)/(AN-S1ABS)

Table 2-6 S1ABS, K value calculation

A11、 A12:1st and 2nd absorbance values of standard (1) measured presently.

C2:Concentration of standard (2)

C1:Concentration of standard (1)

CN:Concentration of standard N (N represents span point)

A2:Average of measured absorbance values of standard (2)

AN:Average of measured absorbance values of standard (N)

A1: Average of measured absorbance values of standard(1)

(b) Applicable assay

Linear (2-point), Linear (multi-point),

(4) Full-point Calibration


All standard solutions (including reagent blank) specified on the “Chemistry Parameters” screen are calibrated.
After this calibration, all working curve parameters S1ABS,K,a,b and c displayed on the Calibration List screen
are updated.

(a) Calculation formula

The calculation varies from the calibration methods.

(b)Applicable calibration types

Linearity (multi-point),Isozyme P, Isozyme Q, Logit-log 3P, Logit-log 4P, Exponential, Spline

2.3 Check of measure value

Various checks are performed to enhance the reliability of measured results. Below are check description

2.3.1 Calibration check


(1) Blank level check

In calibration, a warning –level alarm is issued if the measured absorbance of blank is not within the input
range of standard 1 absorbance. In this case, the result of measurement and alarm (S1ABS) are printed out. To
avoid check, enter-“-3.3~3.3”.

(2) Discrete check

A warning level alarm is issued if the difference of the two times measured absorbance value is larger than the
set value. In calibration, each Calibrator (include reagent blank: Calibrator 1) is tested twice.

57
User Manual of CS-400 Auto-Chemistry Analyzer

To avoid the check, enter“3.3”.

Discrete check is performed by the below formula:

≤Absorbance Discrete Check(ABS)

(3) Sensitivity check

If the difference of standard absorbance (average between 2 times measure) from Max. concentration standard
absorbance (sensitivity) exceeds the permissible absorbance sensitivity value (Sensitivity Limit), a warning-
level alarm is issued. In this case, alarm mark is printed out together with the result of measurement.

The working curve of the alarmed analytical item will be renewed, and the K value won’t be renewed. To
avoid the check, enter-“0”.

Check the permissible sensitivity by the below formula:

ASTD(N) − ASTD(1)
lower lim it ≤ ≤ upper lim it
C STD(N) − C STD(1)

(4) K factor check

If the fluctuation in factor K value between previous calibration and current calibration is 20% ,a warning
level alarm is issued. The working curve and K factor will be renewed and testing can be carried out. Make
sure check the reason of alarm.

Check the K factor by the below formula:

K this − K last
× 100% ≤20%
(K this + K last ) / 2
(5) Drift rate check

In calibration, a warning –level alarm is issued if the difference between the calculated absorbance and tested
absorbance has exceeded the drift rate set value. The working curve and K factor will be renewed and testing
can be carried out. Make sure to check the reason of alarm. To avoid the check, enter“3.3”.

2.3.2 Absorbance of reaction limited check


As to the Rate assay, correct data won’t be obtained when concentration or activity exceeds the quantitative
span. Thus, set the upper limit value and lower limit value of the absorbance, print the alarm sign. Input the
calibration value on the screen. To avoid the check, enter-0 (decrease) or 3.3 (increase).

When 4 or more than 4 tested absorbance value is not accord with the set value of reaction limit absorbance,
alarm is issued as figure 2-21 shows:

58
User Manual of CS-400 Auto-Chemistry Analyzer

ABS

Reaction limit level

Time

Input photometry range

Figure 2-21

2.3.3 Linearity Abnormal Check


In the rate assay, relation between absorbance change and time should be Linearity. Thus, check on the
linearity is a necessity.

Select “Linearity check” in “Alarm Info.”, and input the limit check value in corresponding textbox as figure
2-22 shows:

Figure 2-22

If not selected, even if input value in textbox, linearity check is not carried out.

(1) When number of measurement points (N) more than 9 (N>9)


Linearity is checked by dividing the difference in absorbance change between the first and last 6
measurement points by the average absorbance change for all. If the value thus obtained is beyond the limit
linearity value, alarm is printed out together with the result of measurement as figure 2-23 shows:

ΔAf −ΔAb
ΔA ×100>Limit linearity value(%)

59
User Manual of CS-400 Auto-Chemistry Analyzer

Photometry point

Time

Figure 2-23 Linearity check N≥9

(2) When number of measurement points (N) between 4 and 8 (4≤N≤8)


Linearity is checked by dividing the difference in absorbance change between the first and last 6 measurement
points by the average absorbance change for all. If the value thus obtained is beyond the limit linearity value,
alarm is issued as figure 2-24 shows:

ΔAf ,−ΔAb,
ΔA, ×100>Limit linearity value(%)

Photometry point

Time

Figure 2-24 Linearity check(4≤N≤8)


2.3.4 Prozone check
In immunoreaction, working curve descends if antigen concentration is abnormally high beyond the suitable
range (prozone area). This is called prozone or zone phenomenon.

This instrument can check whether concentration is in the absorbance decreasing range (post zone). For prozone
check, the following 2 methods are available: antigen readdition method in 1-point assay with prozone check
and reaction rate method in 2-point assay. To Avoid check, input “-3.3 lower limit” in “checkup value” function
box in “analyze parameter” menu.

(1) Antigen supplement method:


Take 1-point essay for example, measure reagent 1, and take it’s value as reference value. Replace reagent with
serum diluents, which contain antigen, add 20ul. Compare the prozone limit value with absorbance difference

60
User Manual of CS-400 Auto-Chemistry Analyzer

(before add reagent 2 and after add reagent 2).

Input method:

Prozone check value (PC value): 【 】

Upper /lower limit: 【】

Analytical method: 【】

Photometric point: 【Q1】 【Q2】 【0】 【0】

1≤Q1≤Q2≤49

Aq 2 + A( q 2 − 1) Aq1 + A( q1 − 1)
PC = −k×
2 2

K=total liquid volume when test q1/total liquid volume when test q2

When 1≤Q1≤Q2≤4 or 5≤Q1≤Q2≤16, 17≤Q1≤Q2≤31, 32≤Q1≤Q2≤49 K=1.

AQ1、AQ2:absorbance of photometric point:Q1,Q2


Absorbance

Time
Figure 2-25 Antigen supplement method

(2) Reaction speed ratio method:


Take 2-point assay for example, the ratio between average reaction speed and reaction speed with serum
(prozone check value: PC value), then compare the PC value with prozone limit value.

Prozone checkup value(PC value) 【 】

Upper/lower limit: 【】

Analyze method: 【】

photometric point: 【q1】 【q2】 【q3】 【q4】

5≤q1≤q2≤49、5≤q3≤q4≤49

61
User Manual of CS-400 Auto-Chemistry Analyzer

( Aq 4 − Aq 3) /( q 4 − q 3)
PC = × 100
( Aq1 − Aq 2) /( q1 − q 2)

photometric point: Aq1、Aq2、Aq3、Aq4:absorbance: q1、q2、q3、q4

photometric point q1、q2、q3、q4 test time point:q1、q2、q3、q4

Do not carry out prozone check when following condition occur:

① test point: q1=q2=q3=q4。

② test for No.1 standard liquid (reagent blank)

Figure 2-26 explains antigen supplement method:


Absorbance

Time
Figure 2-26 Reaction speed ratio method

2.4 ISE testing principle

2.4.1 Operation principle


Once analysis starts, SIP injection pump aspirates 65UL reference liquid and infuses it into the reference
electrode liquid line---sample probe aspirates15uL sample and infuse it into diluting cup---DIL injection pump
infuses 450uL diluent---SIP injection pump aspirates the diluted sample and infuses it into electrode liquid line,
and test its potential, and the rest of diluted sample will be assimilated by vacuum nozzle---IS injection pump
infuses 600uL internal standard liquid into the diluting cup for rinsing it, and the internal standard liquid will be
sucked by vacuum nozzle sequently---IS injection pump infuses 450uL internal standard liquid into the diluting
cup---SIP injection pump aspirates 65uL reference liquid and infuses it into the reference electrode liquid line
first, and then SIP injection pump aspirates it and infuses it into the electrode liquid line, and test its
potential---the vacuum nozzle assimilates the redundant internal standard liquid in order to empty the diluting
cup for next turn test.

2.4.2 Electric potential generation principle


The electric potential is calculate by “Nernst” formula.
62
User Manual of CS-400 Auto-Chemistry Analyzer

RT
E = E 0 + 2.303 × × l og(ai ) …………………………………………………………(1)
nF

ai = f × Ci ………………………………………………………………………………. (2)

E 0 :Standard electrode potential of tested system


R:Gas constant(8.314510 J*mol-1*K-1)

T : Absolute temperature(t℃+273.15)(K)

F : Faraday constant(9.6485309×104C* mol-1)

ai :Ion(i)activity
f :Activity coefficient

Ci:Concentration

n :electric charge of given icon(Cation is positive, Anion is negative)

2.4.3 Test method


The following procedure explains the preparation of working curve, the test of internal standard liquid
concentration, calculation of concentration, and result revision.

2.4.3.1 Working curve preparation


Test low concentration slope liquid (S1) and high concentration slope liquid (S2), set the slope (sensitivity) of
K,NA, Cl electrode.

E ( H ) − E ( L)
SL = ……………………………(3)
C(H )
log
C ( L)

SL:Slope

E(H): Electrode potential of high concentration slope liquid

E(L):Electrode potential of low concentration slope liquid

C(H):Concentration value of high concentration slope liquid (input value 0)

C(L): Concentration value of low concentration slope liquid (input value 0)

2.4.3.2 Internal standard liquid concentration test


Test the concentration of internal standard liquid when working curve prepared.

E ( IS ) − E ( L )
C ( IS ) = C ( L) × 10 SL
………………………………(4)

C(IS):Concentration of internal standard liquid

E(IS): Electrode potential of internal liquid.


63
User Manual of CS-400 Auto-Chemistry Analyzer

2.4.3.3 Concentration calculation.


Make internal standard liquid concentration as reference value, calculate basic sample, emergency sample, QC
product concentration. The internal standard liquid is tested according to different sample.

E ( S ) − E ( IS )
C ( S ) = C ( IS ) × 10 SL
……………………………………….(5)

C(S):Sample concentration

E(S): Sample electrode

2.4.3.4 Result revision


Test Calibrator (S3) after calibration, and the calculate the concentration, make the difference between test
concentration and input value as compensation value, and add/ subtract the sample concentration of standby
sample.

C (VALUE ) = C (C ) − C ( X ) ……………………………..(6)

C(VALUE): Compensate value

C(C): Input value of serum Calibrator concentration

C(X):Measure value of serum Calibrator concentration

C ' ( S ) = IF {C ( S ) + C (VALUE )} …………………(7)

C’(S):Sample concentration after revision

IF: Instrument constant(1.0)

2.4.3.5 Standard specification of ISE

Item Specification

Sample syringe 15ul


Diluent liquid volume 450ul
Capacity 80 sample/hour(only for electrolyte test)
Na 20~200mmol/L(only for serum test)
10~400mmol/L(for urine test)
K+ 1.0~15.0mmol/L(only for serum test)
Range
1~200mmol/L(for urine test)
Cl- 20~200mmol/L(only for serum test)
10~400mmol/L(for urine test)
Internal standard liquid 1050ul/sample (only for electrolyte continual test)
Reagent consumption
Diluent 450 ul/sample
volume
Reference electrode liquid 130 ul/sample

Table 2-7 ISE standard specification

64
User Manual of CS-400 Auto-Chemistry Analyzer

Note: During operation, if ISE analysis is not taken more than 10 minutes, the internal standard liquid will be
pipetted once in order to activate electrode.

65
User Manual of CS-400 Auto-Chemistry Analyzer

Chapter 3 Instrument Installation


Analyzer has passed all tests before transportation. Packing it gently in order to avoid damage while
transportation. To make sure normal operation, install or initialize the CS-400 only by authorized staffs from Dirui
Company.

3.1 Installation requirement

Before installation, operator should check the space, power and environment requirement.

3.1.1 Space Requirement


To make sure the space of maintenance, please follow the instruction as below:

● Space between left (right) side of analyzer and the wall should ≥50cm

● Space between rear panel of analyzer and the wall should ≥50cm

● Space in front of analyzer should≥100cm

● Make sure there is enough space for waste device and purified water equipment.

3.1.2 Environment requirement


Operate or store the analyzer according to the following requirement:

● Working environment: 15℃~32℃

● Relative humidity: 40% ~85%

● Atmospheric pressure: 76kPa~106kPa

● Environment should with no dust, no mechanical vibration, no noise source and power interference

● Do not put the analyzer in the vicinity of brush motor, flicker fluorescent tube and other constant on-off
electrical equipment.

● Avoid direct sunlight, do not put the analyzer in front of heat source and wind source

3.1.3 Power requirement:


● Power supply: ~2 20 V, 50 Hz

● Power: 2000 VA

● Circuit breaker: 250V, 20A

●A well grounded power supply socket is a must. (Socket at least with one 15 A and three 5A). Large electrical
appliance such as air condition, refrigerator, even cannot use the same electrical wire as analyzer. Instrument
is equipped with a three core electrical wire, red wire is live line, blue wire is zero line, and yellow green wire
is ground lead. As figure 3-1 shows:

66
User Manual of CS-400 Auto-Chemistry Analyzer

HOT

Figure 3-1

Note: The unqualified environment may cause test value inaccuracy, analyzer damage and it is also
harmful to human body.

3.2 Open package

3.2.1 Procedure
Check if there is a physical damage on the packing when analyzer arrived. If yes, contact Dirui company or
local distributor. If not, open the package according to below procedure.

● Make sure that arrow on the package is up, upright the package.

● Open the accessory box and mainframe box, check if parts in box are complete, if not, please contact Dirui
company or local distributor.

● Check the packing and appearance of the analyzer, if there is a damage, please contact Dirui company or local
distributor.

3.2.2 Transportation method


● Do not move analyzer until “discharge liquid ”operation finished. For detailed operation please refer to
12.4.8.

● Only push the analyzer in short and smooth distance

● Protect the display screen and sample probe of front panel from outside force while transportation.

● Make sure analyzer stands upright, no slope, no side lay.

● Avoid vibration while transportation, check and debug the analyzer after transportation.

67
User Manual of CS-400 Auto-Chemistry Analyzer

3.3 Installation procedure

3.3.1 Software installation


Install hardware only by professional staff of company. Do not disassemble the analyzer except normally
system maintenance.

Install software only by professional staff of our company. User is not allowed to uninstall software unless
abnormality occurs. Uninstall the software according to the following procedure:

Put CS-400 applied software into CD-Rom, click “set up.exe ” file, installation program initialize as figure 3-2
shows:

Figure 3-2

Figure 3-3
68
User Manual of CS-400 Auto-Chemistry Analyzer

Click “next ” in figure 3-3, a user information form pops up as figure 3-4 shows:

Figure 3-4

Click “next” in figure 3-3, a software select window of install folder will pop up. The default destination folder
is “C:\Program Files\Changchun Dirui Industrial CO., LTD\CS-400 Auto-Chemistry Analyzer applied software”.
User could revise installation path by click “self-defining” button as figure figure 3-5 shows:

Figure 3-5

Click “next” button in figure 3-4, a software select window of install folder will pop up. The default destination
folder is “C:\Program Files\Changchun Dirui Industrial CO., LTD\CS-400 Auto-Chemistry Analyzer applied
software”. User could revise installation path by click “self-defining” button as figure 3-5shows:

69
User Manual of CS-400 Auto-Chemistry Analyzer

Figure 3-6

Click “install” button in figure 3-6, initialize the software as figure 3-7,3-8 shows:

Figure 3-7

70
User Manual of CS-400 Auto-Chemistry Analyzer

Figure 3-8

Click “finish” button in figure 3-8 to complete the installation process.

3.3.2 Connection of peripheral equipment


3.3.2.1 Connection of pure water pipeline
Connect the outlet interface of pure water supplier and the pure water inlet interface of the analyzer.

3.3.2.2 Connection of waste liquid outlet pipeline


a) Outlet pipeline of concentrated waste: connect one end of the concentrated waste pipeline taken with the
analyzer with the concentrated waste liquid outlet interface(6 in figure 1-3), and place the other end into the
waste liquid collector.

Note: please deal with the waste liquid under the local rules and laws.

b) Outlet pipeline diluted waste: connect one end of the concentrated waste pipeline taken with the analyzer
with the concentrated waste liquid outlet interface(4 in figure 1-3), and place the other end into the waste liquid
collector.

c) Connection of liquid level sensor of concentrated waste solution: connect one end of the concentrated waste
pipeline taken with the analyzer with the concentrated waste liquid outlet interface(7 in figure 1-3), and place
the other end into the waste liquid collector.

3.3.2.3 Connection of computer


Connect one end of the communication cable taken with the analyzer with the interface of “RS-232” in the
analyzer, and place the other end into the concentrated waste liquid collector.

3.3.2.4 Printer installation


Make sure do the following checkup before print:

(1) Check if install the driver of printer.

(2) Check if data line is connected well between printer and analyzer.
71
User Manual of CS-400 Auto-Chemistry Analyzer

(3) Check if there are papers in printer, printout after turn on the switch.

3.3.3 System login


After installation, following power supply need to be got through:

General power supply of the instrument, analytical system, computer host, display of computer and printer.

Then, click the icon on computer screen or click “start-up”, then find the software CS-400 in
“program” and click it, after that, enter “system logging” window as the figure3-9 and 3-10 show.

Figure 3-9

Figure 3-10

Input user name, password, click “login” to get into the main menu of software, as figure 3-11 shows: click
“back” cancel login.( Initial user name: 001, password: 001)

72
User Manual of CS-400 Auto-Chemistry Analyzer

Figure 3-11

After successfully login, the software show as offline state, browse menu, check alarm information, user logout
function can be used at this state. Click “on-line” button, user could then process all browse and testing
operation.

If the wrong user name and password is inputted, a hint form will pop up, as figure 3-12 show:

Figure 3-12

If 3 times failed consecutively the user name and password , the program will exit automatically.

The function keys in the figure3-11 are described as following:

a) Connection: in figure3-11, click “ ”, “connecting” will show, after success of access, the status bar
will display “sleep”. At this time, all operation and tests can be carried out.

After logging, the instrument partially connected or not connected the cable, click the “Connection”, the screen
will show as figure 3-13 shows.

73
User Manual of CS-400 Auto-Chemistry Analyzer

Figure 3-13

If this shows on the screen, connect after get the power supply connected.

b) Exit system: In the window as figure 3-11 shows, click “ ”, enter window figure 3-14 shows:

Figure 3-14

Click “ok” in figure 3-14 to exit software.

Only exit system at off-line state. If analyzer is on-line, click “offline” button to exit system, dormant could be
carried out when analyzer stand-by.

Note:

● In order to prevent data from being damaged or revised by other people, exiting software when doctor takes a
rest is strongly suggested. Periodically backup database in order to avoid data lose.

●. Input initial user name and password when first login, select “user information” in “management” menu, set
user name, password and access authority for next time login.

The analyzer will be in sleep status after 20 mins of power on(waiting for the stability of power and
temperature)

3.3.4 Uninstallation of software


If the uninstallation of applied software of full-automatical chemistry analyzer of CS-400 is needed, please enter
“addition or cancel program” in setting board, click “cancel” button, window “addition or cancel program” will
pop up. Then window like figure 3-15 will show:

74
User Manual of CS-400 Auto-Chemistry Analyzer

Figure 3-15

In the figure 3-15, click “Yes” to complete the uninstallation.

75
User Manual of CS-400 Auto-Chemistry Analyzer

Chapter 4 Accessory Device

4.1 Sample disk barcode reader

4.1.1 Scan range of barcode reader


Barcode reader on sample disk is used to identify barcode of 1-50 sample positioned on the outer track of
sample disk.

4.1.2 Container requirement


(a) Specification: cuvette: Φ10mm×75mm, Φ10mm×100mm, Φ13mm×75mm, Φ13mm×100mm(±1 mm)

Standard cup: Φ14mm×37mm(±1 mm)

(b) Orifice of the cuvette should be regular. Deformation and extrusion is not allowed.

4.1.3 Barcode requirement


(a) Type: Code bar、code 128、code 39、code 93、12of5、UPC/EAN

(b) Size: Blankness between start and finish line should within be 3mm when cutting barcode as figure 4-1
show:

Width

Start blank Length End blank

Figure 4-1 Blank and length

(c) Digit of different barcode types

Sample barcode type Digit

Code39 5~10
Code93 4~12
Code128 5~22
12of5 4~15
UPC-A 11
UPC-E 6
EAN-8 7

Figure 4-1
76
User Manual of CS-400 Auto-Chemistry Analyzer

4.1.4 Stick requirement


(a) No cockle, no contamination one the label, no line deformity when stick barcode , otherwise the barcode
reader cannot correctly read the barcode.

(b) Stick barcode on correct place:

In order to obtain correct barcode, 15mm-20mm between cuvette bottom and barcode lower line is required.
Make sure barcode is on the outer side of sample disk when place on the test tube rack., as figure 4-2 show:

Barcode

Figure 4-2

When the CODE39 is not capitalized, please add “+” before the corresponding capitalized letters of the ID code
displayed on the screen and the report sheet printed after scanning.

Note: the “ ”, ‘’
,()are not allowed for the barcode, or it cannot be identified.

4.1.5 Barcode reader checkup


When test startup, sample disk will stop turning on barcode reader position, and then barcode reader will read
barcode. If barcode is not identified correctly, the barcode reader will repeat scanning three times. Sample
supplement cannot be taken when scanning, it can be taken only after scanning. If “scan sample barcode” is set,
sample probe will stop sampling operation; sample disk will turn to barcode reader position, start scanning.
Sample disk turns to sampling position when scanning finish. Scanning information will be showed in “sample
register” and “test result” menu.

Barcode reader will continually identify 1-50 sample on outer track of sample disk when processing barcode
reader checkup, and the scanned information will be showed in “ maintenance” menu.

“??” means no effective barcode exist.

4.2 Reagent barcode reader

4.2.1 Scanning range


Barcode reader on reagent disk is used for identifying barcode on reagent disk 1 and reagent disk 2.

4.2.2Reagent bottle requirement


Specification: 70ml, 20ml.

4.2.3 Operation requirement


(a) Type: Code 128(17 digit)

(b) Size: width within: 12mm-15mm, length within 40mm. As figure 4-1 show.

77
User Manual of CS-400 Auto-Chemistry Analyzer

(c) Blankness between start and finish should within be 3mm when cutting barcode as figure 4-1 show.

4.2.4 Stick requirement


(a) Stick the barcode with no cockle, make sure there is no deformity on barcode line. Contamination is not
allowed on label, or barcode cannot be read correctly.

(b) Stick the barcode on the correct place.

Blankness between bottle bottom and barcode should be within 15mm-25mm, thus barcode could be read
correctly.

4.2.5 Barcode reader checkup


Carry out “barcode scanning” in “reagent info.” menu, reagent barcode reader will continually scan on the two
reagent disks twice. If barcode cannot be identified while scanning, barcode reader will repeat scanning three
times

When checking reagent barcode reader in “system maintenance”, reader will identify barcode info. and
display the scanning result of R1, R2 reagent disks in the window “system maintenance”.

“??” means no effective barcode exist.

4.2.6 Reagent barcode rule


Reagent barcode information can only be read by barcode reader, the information will coupling with chemistry
parameter which stored in instrument, and this process is called reagent registry information. Reagent
information registration could check reagent position on reagent disk.

The read information could be showed in “reagent information” menu as “disc No.”, “position”, “reagent name”,
“reagent type”, “remaining reagent volume”, “remaining test No.”.

Reagent name: Chemical name of analyze item.

Position: Each reagent disk has 1-45 positions, 1-44 are used for place reagent, and 45 position is specially used
for place CS-anti-bacterial phosphor-free detergent.

Note: regular clearing of the reading window of sample disk and reagent disk barcode reader is required.

4.3 Purified water equipment

Instrument consumes 25L/Hour water at peak value. The purified water equipment should meet the following
requirement:

① water should be obtained from tap water pipe

② water conductivity should within 1uS/cm

③ water supply volume should reach 50L/h or more

④ The hydraulic pressure should within 49-343 Kpa

Note: To use/maintenance the purified water equipment, please refer to user manual, or consult the distributer or
manufacturer.

78
User Manual of CS-400 Auto-Chemistry Analyzer

Chapter 5 CS-400 Software Operation

5.1 Software interface instruction

5.1.1 Main interface composition


The main menu of software is composed of status bar, main function keypad, workspace, software information.

(a) Status bar: Shows display status on the top of menu. As figure 5-1 shows:

Figure 5-1

Description:

: Represent display status: stand-by, testing, emergence stop, sampling stop, maintenance operation,
sleeping mode.

: Communication monitor mark. When communication is under normal status, the icon turns blue; when
communication abnormal, the icon turns black. Once the icon color turns from blue to black, that indicate
communication failure.

: Alarm icon. This icon occurs in status bar when alarm is issued. Click the icon, start alarm checkup, solve
the problem according to the remedy.

: Display temperature of circulating water in incubation bath, regular water temperature is within
37℃±0.1℃. Alarm issued when temperature above 45℃. Alarm also occurs during test status.

: Display ID information of current user. To setup, change or remove the


information, click “user information” in “management” menu.

: Display computer system time (year-month-day week hour-minute-second).


Click “time and date” button to change time and date.

(b) Main function keypad: Select menu by single mouse click. Single click on corresponding function key, the
border will change color correspondingly. As figure 5-2 show:

Figure 5-2

(c) Working space: enter specific operating space, as figure 5-3 show. Single click corresponding function key
79
User Manual of CS-400 Auto-Chemistry Analyzer

enter working space menu, the border color change correspondingly.

Figure 5-3

(d) Hint bar: instruct user how to use software, hint the input range, input method, operation error, as figure 5-4
show:

Figure 5-4

(e) Shortcut key space: for convenience use, as figure 5-5 shows. Single mouse click on corresponding function
key or press F2-F9 on keypad.

Figure 5-5

(f) Software information space:

Click “ ” key to display edition information of software. Edition No.:


1.0. As figure 5-6 show. This form is a mode form, click “ok” key to close.

80
User Manual of CS-400 Auto-Chemistry Analyzer

Figure 5-6

5.1.2 Keypad function


(a) Num Lock

This button is used for checking if number keypad is open.

(b) Caps Lock

This button is used for switch letter case.

(c) Function key

F1 Software help shortcut

F2 Start condition shortcut

F3 Sampling stop shortcut

F4 Emergence stop shortcut

F5 System monitor shortcut

F6 Alarm information shortcut

F7 User log-out

F8 Exit system shortcut.

81
User Manual of CS-400 Auto-Chemistry Analyzer

5.1.3 Software function frame

Sample test Edit sample, doctor, patient test information

Delete result, result query, review, preview, audit, print,


Test result batch print, batch print, reaction curve

Reagent Info Manual registry, remove reagent info, barcode scan, reagent level

Calibration registry Calibration registry type and item


Calibration
Information
Calibration result Calibration data and reaction curve.

QC registry QC registry and parameter setup

QC control QC interval Analyze QC test data

Chart monthly Analyze QC data in one month


CS-400

Parameter Analyze, calibration, range parameter


Automatic

Item combination Edit item combination information

Calculate item Edit calculation item information

Cross infect Cross contamination avoiding


Biochemistry

System setup Report format Report Info, print sequence, print format setup

ISE setup ISE parameter setup


Analyzer

System setup Barcode, ISE, time awaken setup.

Manual item setup Add manual item.

LIS setup ISE communication setup.

User Info. Operator ID, name, password, access authority.

Hospital info Test delivery department and doctor.

Sample type, patient type, clinic


System Other Info. diagnose, report remark、item unit.
Management
Workload statistic

Database backup and restore

System log Login, maintenance, operation, alarm log


System
Maintenance Periodical maintenance and checkup

System help Provide help to user

Figure 5-7
82
User Manual of CS-400 Auto-Chemistry Analyzer

5.2 Software Operation

Select function key of software by single click mouse button. Input value and character combine with keypad
(switch by shift + control, input method is depend on windows system).

5.2.1 Cursor move


Cursor moves as single mouse click on target input space or target item.
5.2.2 Function key select
Select function key by single mouse click.

5.2.3 Open Form


In order to open form, click function key corresponding to form. Form is divided into mode form and modeless
form.

Mode form: other menu cannot be opened until mode menu is closed. Click “exit” or “close” button to exit. As
figure 5-8 show:

Figure 5-8

Modeless form: other menu can be opened when modeless menu open, system will automatically close last menu
when other menu is opened. As figure 5-9 show:

83
User Manual of CS-400 Auto-Chemistry Analyzer

Figure 5-9

5.2.4 List box and scroll bar


(a) List box

List box is used for displaying a information. List box is also used for finding and selecting needed information
from displayed information. As figure 5-10 shows:

Figure 5-10

(b) Scroll bar

Scroll bar is used for adjusting display range in list box. Scroll bar is divided into longitudinal scroll bar and
horizontal scroll bar. As figure 5-10 shows:

84
User Manual of CS-400 Auto-Chemistry Analyzer

5.2.5 Pull-down menu


Click “ ” on the right side of menu to open or close pull-down menu. More information can be showed in
pull-down menu. Once select , the selected item can be showed on the top column, and pull-down menu
disappears at the same time.

5.2.6 Button box and Check box


Button box: only one function can be chosen between two functions, as figure 5-11 shows:

Figure 5-11

Check box: more than two functions can be chosen at the same time, as figure 5-12 shows:

Figure 5-12

85
User Manual of CS-400 Auto-Chemistry Analyzer

5.3 Instrument standard specification

Performance index Standard specification


Plat image grating spectra photometry system,photometry by
Wavelength
12 wavelength. wavelength:340 、380 、405 、450 、480 、
range
505 、546 、570 、600 、660 、700 、750nm
Wavelength
±2nm
Basic precision
feature Reaction temperature 37℃±0.1℃
Reagent open,88 colorimetry test,3 ISE item can be taken at
Test item
the same time.
Measure method End point essay、Rate essay、turbidimetry
Measure speed Constant speed 400 test /hour, 800 tests /hour with ISE
Sample disk、 115 position(basic sample :50,standard product: 34,emergence
Sample position sample 20,QC product: 8, detergent:3
R1, R2 cooling reagent disk,each reagent disk conclude 45
Reagent disk、
reagent position ( 45 position is special for placing
Sample Reagent position
CS-anti-bacterial phosphor-free detergent)
system Sample barcode identify system 3 build-in barcode reader (basic sample, reagent 1, reagent 2)
Sample type Serum, blood, urine, cerebrospinal fluid, ascetic fluid
Sample volume 2~35μl
Sample liquid level sensor Integrate with sample probe.
Reagent volume 20~350μl
Reagent Reagent bottle volume 20mL 、70mL
system Reagent restore temperature Temperature within 5~15℃,adopt semi conducting cooling
Reagent liquid level sensor Integrate with reagent probe
Reaction cuvette type Discrete system
Reaction cuvette
6mm
optical diameter
Reaction position 6sets(20),total :120
Reaction time No more than 15 minute(3、4、5、10, 14minute can be setup.)
Analyze Reaction liquid volume 150~450μl
system Light source 20W/12Vlong life quartz halogen lamp
Absorbance range 0~3.3Abs
Quality control QC interval, Monthly QC
Automatic rinsing Rinsing cuvette automatically.
Stirring rod mechanism Singleness stirring after reagent adding
Interface TCP/IP interface,standard RS-232 and USB2.0 interface
Data
Printer stylus printer,support user-defined mode
system
Connect with LIS/HIS system Connect with LIS/HIS system
Integrated Weight About 330Kg
Equipment size 1060 mm×790 mm×1150mm(length×width×heighth)
system Power consumption(VA) 2000VA

Figure 5-16

86
User Manual of CS-400 Auto-Chemistry Analyzer

Note:according to different test condition,sometimes instrument processing capability is lower than 400 test/hour

Test condition Processing capability lower degree(estimated)

Test after sample diluent 133tests/hour(test under pre-diluent condition)

At least 200test/hour(reaction cuvette、sample probe)


Cross contamination avoiding function
200~400test/hour(reagent probe)

R1-R4 test At least 200 test/hour

Figure 5-17

87
User Manual of CS-400 Auto-Chemistry Analyzer

Chapter 6 Instrument Operation

6.1 Overview of operation

Table 6.1 shows the operation flow. For detailed operation, please refer to table 6.2.

Consult
Operation step Form / function key Operation
index
1. Check before operation Check before turning on power 6.2.1
2. Connect water unit and turn Open water faucet, turn on the power of
on CS-400 power purified water unit and CS-400 6.2.2
Software login System Login Input operator’s ID No. and password.
3. Check instrument status
1) Check alarm Alarm information Please refer to chapter 13 “alarm
maintenance”
Execute light quantity checkup, check if
test value is within regular range
2) Check light quantity of System maintenance Execute cell blank check, check if test 6.2.3
photometer value is within regular range
3) Check cell blank System maintenance Check if temperature of incubation bath
is within 37.0℃±0.1℃
4) Check temperature of Status bar
incubation bath.
4. Check analytical conditions
1) Item added System setup Add chemistry item
6.2.4
2) Chemistry parameter System setup Check chemistry parameter
3) Check K value Calibration info Check calibration curve and K factor
5. Reagent preparation (reagent
info)
1) Check reagent residual Reagent info Check reagent remaining volume and
volume remaining test times. Place the reagent at
the corresponding position on the 6.2.5
Reagent disk.
2) Preparation for Reagent info
photometry item and
preparation for ISE item
6. Setup of calibration and Calibration info Check item name of calibration analysis
6.2.6
control item Quality control Check item name of QC
7. Sample registration and test Sample registration Registry and check sample info, patient
6.2.7
info, and operator info.
8. Start analyze
1) Preparation of sample, Put routine sample, standard and control
standard solution, Controls, sample on sample disk
detergent start analyze Select test condition, and execute “start 6.2.8
analyze”.
2) Set starting condition and
start test
9. Testing Monitor the instrument when testing.
(1)System monitor System monitor
(2) Sampling stop/continue Sampling
stop/continue 6.2.9
(3) Emergence stop Emergence stop
(4) Sample super addition. Sample registry Carry out sample edit when testing and
click start analyze.

88
User Manual of CS-400 Auto-Chemistry Analyzer

Consult
Operation step Form / function key Operation
index
10. Check test result (result To search, amend, delete test result,
Result data 6.2.10
data) check sample reaction curve
In “recheck” form, click “start recheck”,
11. Recheck sample test Test result 6.2.11
and send the recheck instruction.
12 Completion of analysis
1) Recheck result Test result Check and printout the test result;
2) Database backup System management Periodical backup database, one week is
suggested.
3) System dormancy Instrument could automatic start at 6.2.12
specified time after setting
4) Turn off instrument Turn off instrument, computer, connect
5) Preparation of next water unit.
operation

Table 6-1 Overview of the operation process

6.2 Detailed operation

6.2.1 Check before measurement

(a) Check power supply and voltage.

(b) Check communication wire and power wire which connecting host computer, instrument and printer.
Make sure the wires are well connected .

(c) Check if print paper is enough, add paper if necessary.

(d) Check if there are water drops, contamination and bending of reagent probe, sample probe and stirring
mechanism.

(e) Check if the detergent is enough. Place CS anti-bacterial phosphor-free detergent at 45th position on R1,R2
reagent disk. CS alkaline detergent in detergent box should be enough, detergent at sample disk W1,W2,
W3 should be enough. For detailed information please refer to “12.1.3 detergent ”.

(f) Check if waste solution. bottle is empty. Ignore this operation if drainage device is connected with down
pipe.

(g) Check if there are air bubbles in syringes ( leakage and air bubble may cause incorrect data)

! Warning :

CS serial detergent is corrosive liquid. In case of skin contact, flush the area with water, rinse immediately with
plenty of water and seek medical advice.

6.2.2 Power on and software login

(a) Turn on the power of purified water, and open the water faucet

(b) Turn on the power supply of CS-400.The main power switch lies in the right lower side of instrument. Turn
on the switch when there are reagents in reagent disk, therefore the cooling system may under normal
working condition. The power of analyzed part lies in the right upper side of instrument.
89
User Manual of CS-400 Auto-Chemistry Analyzer

(c) Login CS-400 software, make sure carry out online instruction before testing.

6.2.3 Check instrument status

6.2.3.1 Check alarm

(a)Letter display of alarm information

Single-click “ ” key in the shortcut area to check alarm. When alarm issues during operation, the
alarm code, level, description and time can be displayed on the screen. (As figure 6-1shows)

Figure 6-1

For detailed information, click the information, and the detailed description will be displayed in the textbox
(as figure 6-2 shows)

Figure 6-2

At the same time, the remedy will be displayed simultaneously in “remedy” textbox ( as figure 6-3 shows).

Figure 6-3

(b) Buzzer Hint:

“ ”indicates buzzer on , the instrument will make a buzz when alarm issues.

90
User Manual of CS-400 Auto-Chemistry Analyzer

“ ” indicates buzzer off . Even if the alarm is issued, there is no buzz.

(c)Alarm icon hint

When alarm is issued, the alarm icon “ ” may occur. Click the icon to display the alarm information
form.

(d)Delete alarm information

Click“ ” key , all the information in the “ alarm info.” form will be deleted. Select one of the

alarm information and click “ ” key, the selected information will be deleted.

(e)Alarm information setup. (As figure 6-4 shows )

— ISE reagent remaining volume alarm: Setup the lowest test number of reference liquid, diluent , Internal
standard liquid of ISE test.

— Sampling stop alarm: Select the function, and alarm will issue when sampling stops.

— Linearity check: Input the limit value of Rate assay or linearity assay.

Figure 6-4

6.2.3.2 Check light quantity

Single-click “ ” button, select “ light quantity check ” option, and click “execute” button.
Instrument will automatically carry out light quantity check. The light value of current time and last time can be
showed in result column. As figure 6-5 shows:

91
User Manual of CS-400 Auto-Chemistry Analyzer

Figure 6-5

Click “ ” button to print out the test result.

As light source lamp aging gradually, the check value may increase daily. AD value of all wavelengths should
be less than 18000. If the AD value is more than 18000, please replace new light source according to chapter
“12.4.3”.

6.2.3.3 Check cell blank

Single-click the “ ” key, select “cell blank check” option, and then click“ ” key.
Instrument will automatically carry out the cell blank check. The cell blank value can be displayed in result

column. Click key to print out the test result. (As figure 6-6 shows)

92
User Manual of CS-400 Auto-Chemistry Analyzer

Figure 6-6

Cell blank value of the first reaction cuvette should be less than 18000, the difference between cuvettes (the
difference between current one and the first one) should be within ±800. When test value out of this range,
please replace reaction cuvette according to chapter “12.2.4”.

Before a new reaction cuvette is replaced, immerse it into 2% CS anti-bacterial phosphor-free detergent for
more than 8 hours, and then flush it with purified water before install. Only carry out test when cell blank value
is qualified.

6.2.3.4 Check the temperature of incubation bath

Observe the temperature in status bar on the top of screen, check if the temperature of incubation bath is within
37±0.1℃. A warning level alarm is issued when temperature over 37±0.5℃, but instrument will continue
analyze at this time. A warning level alarm is issued when temperature of incubation bath over 45℃ (computer
stand-by).

Note: After power on, it takes 20 minutes to ensure the constant (37±0.1) ℃ of incubation bath. And it also
takes several minutes to stabilize the light source lamp. Therefore, form information can be browsed, parameter
can be login, alarm info. can be check even computer is not in standby status. Sample testing can be only
carried out after computer stand-by.

6.2.4 Check analyze condition

Setup the “analyze item added, chemistry parameter, calibration parameter and control parameter” before
testing. As to the parameter setup, usage method and storage information, please refer to the user manual or
consult the relative manufacturer or distributor.

6.2.4.1 Check analysis conditions of colorimetric item

(1)Add / Remove item

93
User Manual of CS-400 Auto-Chemistry Analyzer

Item should be added before parameter setup. Click the “ ” key, and its -click the

“ ” in “ ” form. For detailed operation please refer to chapter “ 9.1 ”.

Item name and number can’t overlap, if the same item of different company is added, please distinguish them
with different letters or numbers. Test takes place in terms of the sequence (from small to big) of item No..

(2)Check chemistry parameter

Set up/check the reagent chemistry parameter according to the reagent manual. Some parameters are necessary,
such as: test item name, wavelength, test light point, test time, sample volume, reagent volume, reagent position,
reaction type, reaction direction and decimal digits etc. and some parameters are significant to a reliable test
result, such as linearity range, absorbance limit, etc. Strongly suggest the operator input correct parameter
before test.

Click “ ” key, then click “ ” key. Please setup the chemistry parameter
according to the reagent manual. For detailed operation, please refer to chapter “9.1”

(3)Profile item setup

The function of profile item is to put the relative items together, such as perform the whole set of liver function
or the whole set of kidney function. Use one key to finish many items registration, which is convenience for a
rapid sample registration.

Click “ ” key, then click “ ” key. For detailed operation, please refer to chapter “9.2 ”.

(4)Calculated test setup

The function of calculated test is to determine result on the basis of test A, B or more test in order to calculate
test result of a new item.

Click “ ” in main function key field, then click “ ” key,. For detailed operation,
please refer to chapter “9.3 ”

(5)Cross contamination setup

The function of cross contamination is to decrease or avoid cross contamination among different items. It
includes the cross contamination of reagent probe, reaction disk and sample probe.

Note: Due to the different reagent formula, the analytical result of other items can be effect. The contamination
degree is different according to varied reagents. For detailed information, please consult the manufacture or
distributor. Strongly suggest the operator setup the reagent which cross contamination may occur separately. Or
through the “Cross contamination” setup to decrease the cross contamination among tests.

94
User Manual of CS-400 Auto-Chemistry Analyzer

Click “ ” key, then click “ ” key. For detailed operation, please refer to
chapter “ 9.4 ”

(6)Check calibration K factor

If 1 point linearity method is used for calibration, Click “ ” key, then click “ calibration result “ key,
as figure 6-7 shows.

Figure 6-7

Input K factor in column, and click “ ” key .

6.2.4.2 Check analysis conditions of ISE item

(1)Activate ISE

If the instrument has equipped with ISE module, please activate ISE in “system setup” form first, then carry out

ISE test. “ ” means ISE is activated. And “ ” means


ISE is not selected, all ISE function can’t be used.

If sign “√” in “ washing after test”, sample probe will assimilate the CS-ISE detergent (locate in W2 position of

95
User Manual of CS-400 Auto-Chemistry Analyzer

sample disk), then rinse the diluent bath and ISE pipeline. This operation won’t take place if it is not selected.

Click “save” key after setup.

(2)ISE parameter setup

Select item name and sample type in “ISE setup” menu. In calibration parameter work area, select Calibrator
position from the pull-down menu, input the concentration of Calibrator. There are three kinds of Calibrators:
low concentration, high concentration and blood sample. 15ul is a fixed sampling volume of Calibrator, there is
no need to set again. Input the reference range, linearity range, decimal digits, QC interval (default value:0.
input interval time if needed), slope and intercept, and then click“ save parameter” key, as figure 6-8 shows:

Figure 6-8

Note : The sample volume of ISEis fixed, 15ul, and no set value is needed.

6.2.5 Reagent Preparation (Reagent information)

6.2.5.1 Colorimetric item Preparation

(1) Reagent usage

(a) Reagent confect, use and storage should be performed strictly according to reagent manual. Avoid air
bubbles occur in reagent. Due to the surface active agent of reagent may cause bubbles. And when liquid level
sensor touches the bubbles, it may misjudge it as reagent, thus cause the inaccuracy of sampling volume and
affect the test result.

(b) Don’t replenish reagent.

96
User Manual of CS-400 Auto-Chemistry Analyzer

If the added reagent is coming from different plant or different lots of the same plant, it may cause the change of
the reagent component, thus effect the test result.

(c) Turn the reagent knob counter-clockwise, open the lid, and put reagent bottle at the relevant position on
reagent disk. Table 6-10 shows the reagent type:

Reagent type Reagent disk

Reagent 1
Reagent 4 Reagent disk1
Diluent
Reagent 2
Reagent disk 2
Reagent 3

Table 6-2 Reagent type and Reagent disk

! Warning:

Execute resetting process before barcode reading. Sample probe, reagent probe and stirring rod mechanism
will start up later. So please leave the instrument after closing the cover of reagent disk. Don’t extend your
hand into instrument in order to avoid body injury.

(2)Reagent manual registration

(a) Single-click “ ” key , and then click “reagent information “ key, input reagent information in
“manual registration” form as figure 6-9 shows.

Figure 6-9

(b) Select reagent disk number and reagent position .

Select 1( reagent disk 1), 2( reagent disk 2) in the pull-down menu. Input reagent position from 1 to 44.
Position 45 of both reagent disks is for CS-anti-bacterial phosphor-free detergent. Select reagent name, reagent
97
User Manual of CS-400 Auto-Chemistry Analyzer

type, bottle specification in the pull-down menu, then click “ register” key.

(c) Reagent scanner cannot recognize unclear barcode, to solve this, click “ barcode’ key to registry by manual,
input valid barcode in “barcode” input box, click “registration” to register reagent information.

Note: Reagent information cannot be only registered or removed under running status.

(3)Barcode scanning ( automatic registration)

Reagent can be automatically registered through “barcode scan” when barcode information are clear and
complete.

Single-click “barcode setup” key, input the barcode number and item name, and then click “ add” key, as figure
6-10 shows.

Figure 6-10

Select the information to be deleted, then click “ delete” to eliminate the information. Single-click “ close” key
to exit the form.

Click “ ” key in “reagent info.” menu, instrument will automatically scan reagent on both
reagent disk, relative reagent information (such as reagent position, name, type, shelf time, model, batch
number and bottle standard) will be displayed in “ reagent info.” list. When reagent remaining test number is 0,
gray color will appear; when expire, red color will appear. As figure 6-11 shows.

98
User Manual of CS-400 Auto-Chemistry Analyzer

Figure 6-11

(4)Reagent Horizontal

Click “ ” key in “ ” form, instrument will check remaining reagent volume and
remaining test number, and the test information can be showed in “ reagent info.” list.

Once reagent probe sucked reagent, the remaining reagent volume and remaining test number are tested and
updated.

(5)Delete reagent information.

Select the information to be deleted by mouse, click “ ” key, a delete form will pop up, as figure 6-12
shows.

Figure 6-12

99
User Manual of CS-400 Auto-Chemistry Analyzer

Click “ OK ” key, the reagent information will be deleted.

Note:

(a) If the reagent disk cover is open during measuring, a warning level alarm will be issued. Do not open the
cover when testing in order to avoid dangerous and instrument damage.

(b) After reagent information registry, carry out reagent level check before test in order to check the reagent
remaining volume and remaining test number.

6.2.5.2 ISE item preparation


(1) Reagent placement

Open the top cover, place internal standard liquid (IS), diluent ( DIL) , reference electrode (REF) in ISE
position at the left side of unit system.

(2) Input remaining reagent volume.

Input the reagent volume of internal standard solution (IS), diluent (DIL) , reference electrode (REF) in “ISE
setup” Form. Click save button to same parameter. Remaining volume will automatically reduce after
instrument use ISE reagent.

(3) Carry out ISE ( all) rinse in “maintenance” menu . If one reagent is changed, do the relevant rinse.

(4) The reagent wastage (ml) is showed in table 6-3, 6-4:

Operation Change reagent(ISE prime)


Reagent Power on Start up Adjust
(ml/test) IS DIL REF IS+DIL ALL
Internal
2.4 2.4 1.05 1.05 22.7 10.7 10.7 22.7 22.7
standard
Diluent 1.2 1.2 0.45 0.15 0.2 11.0 0.2 11.0 11.0
Reference
0.9 0.9 0.13 0.065 0.8 0.8 12.8 0.8 12.8
electrolyte

Table 6-3

Reagent ISE check


Internal
2.4+0.45*N
Standard
Diluent 1.2

Reference electrolyte 0.9+0.65*N

Table 6-4

Note: “N”is the time of checking.

6.2.6 Calibration item and QC items registration

(1)Calibration item registration.

100
User Manual of CS-400 Auto-Chemistry Analyzer

(a) Click“ ” key, select calibration method in “ calibration ” pull down menu in

“ ” form, and select test name and test item .

The four kinds of calibration methods is showed in Table 6-5:

Calibration Standard
Object Applied assay mode Application example
type solution volume
Blank solution renew reagent All calibration methods K coefficient method, when
Blank
blank value standard solution test is
calibration
omitted
Span One point except Renew k value Two point linear and Recheck standard solution
calibration of blank solution. multi-points linear point 1
Blank solution Renew reagent Two point linear and Linear 2 points method
and span points blank value and multi-points linear finish standard curve.
2 points
K value Multi-point method,omitted
calibration
the quantity of calibration
solution.
All concentration Renew standard Multi-point linear, Multipoint standard curve,
registered curve by all points isozyme Q, isozyme method ,
Full points
isozyme P , electrolytical standard
calibration
and nonlinear working curve
curve calibration

Table 6-5

(b) Select calibration item in “ test item” work area. Sign “√” in the function block by single-clicking mouse,
show as: “ ”, double-click to cancel it, and “ ” is showed.

(c) Click “calibration register ” functional key , the registered item name and item type will be displayed in the
browse area. In order to check the concentration and position of calibration item, single-click the desired item
name, the calibration parameter will be displayed in the right side. As figure 6-13 shows:

Figure 6-13

101
User Manual of CS-400 Auto-Chemistry Analyzer

(d) Delete the calibration item; Select the desired item, click “ delete” key , the item will be deleted.

(e) In order to execute ISE calibration, single-click “ ” button, select “execute ISE”.

(f) Click “ ” key to exit the “ calibration info.” menu.

Note: The test item should in “execute” status, and select “calibration before analyze” in “start analyze” form.

(2)QC item registration

Click “ ” key, register the name, position, batch No. information according to chapter 9, and set up its
target value and standard deviation.

Note: QC registration of colorimetry item is the same as ISE item, select it in “QC test” pull down menu.

6.2.7 Sample registration

Click “ ” key, and then check “ sample info.” form. As figure 6-14 show.

Click “ “ to exit.

Figure 6-14

102
User Manual of CS-400 Auto-Chemistry Analyzer

(1)Single sample registration

(a) Input sample number in “ sample number” functional block, or click “ previous / next” key to select the
sample number.

(b) In order to register routine sample , select the sample disk number from 1 to 9 in “ disk number” menu.
Input sample position No. from 1 to 50 in “ position number” functional block. If sample disk number is
changed, instrument will issue stop level alarm, but analysis do not stop. Single-click “continue” key, sample
probe begin to sampling from new sample disk.

In order to register emergent sample, single-click “emergence” key, input number 51-70 in “ position”
functional block.

(c) Set up sample type, sample volume, cup type and check date etc. information.

(d) For sample diluting, click “ dilute ” key.

(e) Click the item name to be tested by mouse in item information work area. “ ” indicate select. Items can
be also inputted the by profile function.

Setup and deletion of compounding item, please refer to chapter “ 7.2 ”

(f) After editing , click “ Register sample” key . The registered information will be displayed at the right side.

— Sample number: Input sample number to functional block, the number is uniqueness, one sample only has
one number in a day.

— Barcode number: barcode number is stick on the outside of sample tube, when scanning, the scanned
value will be displayed in “ barcode number” functional block. If barcode number scanning failure, input the
effective number in “ barcode number” functional block.

— Sample type: Select sample type in pull down menu. This function is in “ other info.” of the “ system
administration” form. For detailed operation please refer to chapter “8.3”.

— Check date: Display the current date.

— Repeated test: The test times to same record. The default time is 1 , the max. is 100

— Dilute: indicate if sample need to be diluted. Select diluent position in chemistry parameter. Input sample
volume, diluent volume, sample volume after diluting. Dilution function must be selected in sample
information.

— Previous sample: when register or delete sample, sample number decease 1 by single-click mouse button.

— Next sample: when register or delete sample, sample number increase 1 by single-click mouse button.

(2)Registration of batch routine sample

When various samples are registered in same test item, batch sample registration can be used.

Single-click “ ” key in “sample register ” menu. enter “ batch sample registration” menu as
figure 6-20 shows.

103
User Manual of CS-400 Auto-Chemistry Analyzer

Figure 6-20

(a) Input the start sample number in the first functional block and the last sample number in the second
functional block of “ Sample No. range”. The last sample number should be large than the first number.

(b) Select the start disk number in “ disk No.” menu, input the beginning position No. (1 to 50) in “position” .
Sample disk No. and sample position No. will automatically increase from little to large when batch
registration.

(c) Select the relevant info. in “ sample type” “ sample volume “ menu.

(d) Single-click item name by mouse , select test item , item compounding function can be used also.

(e) After edit, click ” register” key. If one of batch registration samples is the same as single registration
sample, instrument will remind user that registration failed, display as figure 6-16.

Figure 6-16

Click “ OK” key to exit. Edit the right batch samples again.

(f) Single-click “ close” key to exit the batch registration menu.

( 3 ) Edit the patient info

Single-click “ ” key in “ sample register” menu. As figure 6-22 shows:

104
User Manual of CS-400 Auto-Chemistry Analyzer

Figure 6-22

(a) Input sample No. in “ sample No.” functional block, click “ Previous” “ Next” key to choose sample No.

(b) In functional block, choose or input patient name, age, gender, case history No., patient type, delivery dept. ,
delivery doctor, bed No., Check doctor, audit doctor, delivery date, clinic diagnoses and remark etc.

Patient name must be input when edit patient info.. Or instrument will refuse registration.

(c) Single-click “ patient registration” key, patient name will be automatically displayed in the browse area.

(d) Single-click “ sample info.” key enter the menu. Click “ close” key to exit the menu.

— Age: Select age from pull-down menu: year, month, day, and time. And then input numbers.

— Patient type: Select “clinic, medical insurance, hospital, physical exam” in pull-down menu. This

function can be setup in “ other info.” of “ ” menu. For detailed operation, please refer to
chapter “ 10.3” .

— Dept.: Select the type of sample delivery dept. in pull-down menu. This function can be setup in “ hospital

info.” of “ ” menu. For detailed operation, please refer to chapter “ 10.2”.

— Doctor: Select the doctor name. This function can be setup in “ hospital info.” of “ ”menu.

105
User Manual of CS-400 Auto-Chemistry Analyzer

For detailed operation, please refer to chapter “ 10.2”.

— Auditor: Select the doctor name. For detailed operation refer to chapter “ 10.1”.

(4)Modification and deletion of sample information

Single-click the desired item in browse area, or click “ previous” “ next ” key to choose the right sample, or
directly input the sample number to be modified in “ sample No.” functional bock, then click “OK” key to edit
new information, click “ sample registration” key to enter “overlay confirmation” menu. As figure 6-18 shows.

Figure 6-18

In order to delete sample, click “ ” key in “ sample info.” menu, as figure 6-19 shows:

Figure 6-19

In the left functional block, input the start sample No. to be deleted. Input the end sample No. in the right one.
For deleting one piece of information, input the same sample No. in the start and end functional block . The
start sample No. should be less than/ equal to the end No..

Note 1: Registered sample (Not yet test) can be modified or deleted during stand-by and operating .

Registered Sample (Already test) cannot be modified or deleted during operating.

Note 2: Operator can register sample info. firstly, then register patient info.

6.2.8 Test preparation


(1)Prepare routine sample, Calibrator, Controls and detergent.

Place routine sample, Calibrator, Controls and detergent. in the relevant position on sample disk.

(a) Detergent placement .


106
User Manual of CS-400 Auto-Chemistry Analyzer

Test colorimetric item, place CS- alkaline detergent in the position of sample W1 disk; test ISE item, please
place CS- ISE detergent in the position of sample W2disk; avoid crossed contamination, place CS- acidic
detergent in the set position.

(b) Calibrator placement

According to the Calibrator position, which is set in “ chemistry parameter”, place the relevant Calibrator in
middle or inner track of sample disk.

(c) Placement of controls

Place the relevant QC sample in inner track in terms of the QC position set in “ QC registration” menu.

(d) Sample placement

According to the routine sample registration position of “ sample register” menu. , place routine sample in
positions 1 to 50 of outer track, and place the emergency in the positions 51~70 of middle track.

Note: QC sample and Calibrator should be tested in standard cup and micro cup.

(2)Start condition setup .

After sample registration, setup “start analysis” condition before test. Click“ ” key to start analysis.
As figure 6-20 shows.

Figure 6-20

At the right side of start menu, operator can setup according to the requirement. For calibration test, click the

“ ” key. For quality control test, click the “ ”key. After

completion of all selection, click “ ” key. Instrument will begin to work.

After test order sent successfully, “ start conditions” form will automatically close. Or click “ ”
key to exit.

Note : Instrument began to test according to “ calibration – QC – sample” sequence. If there are ISE items, test
according to “ ISE – colorimetric” sequence.

107
User Manual of CS-400 Auto-Chemistry Analyzer

6.2.9 Testing
(1)System monitor

During sample testing, the status of sample disk, analysis disk and reagent disk can be real time monitored.

(a) Enter “ ” menu, click “ ” key, select reagent disk in

“ ”, and then click the reagent position in reagent disk chart, all the relevant
information will come forth , as figure 6-21shows:

Figure 6-21

— Reagent remaining volume: Single-click the corresponding position of reagent in the reagent disk chart,
the reagent remaining volume will be showed directly by chart and percent mode in menu.

— Remaining volume test: In terms of the remaining volume and the set volume in chemistry parameter,
instrument will automatically calculate the remaining test quantity.

— Remaining volume: Display the reagent remaining volume in the current position in ml unit.

— Reagent status: Different color indicates different reagent status.

Reagent normal: Test volume conforms to the test requirement. Show as green color.

Reagent shortage: Reagent volume or reagent remaining test times is less than setup value in “System
setup” form. This is called reagent shortage. Show as violet color.

Reagent absence: Reagent volume is 0 ml. Show as red color.


108
User Manual of CS-400 Auto-Chemistry Analyzer

Reagent not in use: Reagents has registered but it is not used for test. Show as white color.

(b) Enter the“ ” menu, single-click “ ”key, and then click the relevant position in
reaction disk chart. All information of reaction cuvette will be displayed as figure 6-22 show:

Figure 6-22

— Test status: Display the current reaction cuvette status. In the monitor chart, different color indicates
different status.

Vacancy : The reaction cuvette is not used for current test. Show as white color.

Sampling: The reaction cuvette is sampling. Show as yellow color.

Reagent 1, 4 : Reagent in the reaction cuvette is infused with reagent 1(R1) and reagent 4 ( R4) . Show
as blue color.

Reagent 2, 3 : Reagent in the reaction cuvette is infused with reagent 2(R2) and reagent 3 ( R3) . Show
as pink color.

Completion : Result has been calculated from the tested sample of the reaction cuvette. Show as green
color.

Dirty cuvette: The cell blank value has exceeded the normal range. Show as red color.

— Sample No.: The No. of sample that is tested in the reaction cuvette.

— Test No.: Instrument will automatically produce the serial number according to the test sequence.

109
User Manual of CS-400 Auto-Chemistry Analyzer

— Item name: The name of analyzed item, which is tested in the reaction cuvette.

(c) Enter the“ ” menu, click “ ” key, select the disk No. from “ sample disk” menu and
the sample No. from “ sample No.” menu, or click the monitor chart of sample disk by mouse, the relevant
information will be displayed as figure 6-23 shows:

Figure 6-23

— Test status: Display the current sample status. In sample disk chart, different color indicates different status.

Vacancy: Sample in this position doesn’t register. Show as white color.

Waiting for test: Sample in this position is registered, but not sampling. Show as green color.

Sampling: Sample probe is assimilating in this position and adding them to reaction cuvette. Show as
yellow color.

Analysis: Sampling Complete, and sample is being analyzed. Waiting for the result. Show as pink color.

Completion: Get the sample result.

(2) Sampling stop or sampling continue

Sampling stop can be execute only when testing, click the “continue” key to continue sampling.

If disk number changes, instrument sampling may temporarily stop, click “ continue” key to go on sampling.

(3) Emergence stop

110
User Manual of CS-400 Auto-Chemistry Analyzer

Click “ ” key when testing, instrument may stop the current action. Emergence stop is not allowed
when scanning sample barcode.

(4) Sample addition

When testing, other sample could be edited in “ ” form. Click “ ” key to send
“sample add” order.

6.2.10 Test result checkup (Result data)


! Warning :

● Sample with lipemia, homolysis and icterus could affect the test result.

● Make sure that sample is not cloudy and with no clot, or sample probe could be jammed and effect test
result.

● Substance in sample, such as medicine, anticoagulant , preservative , may disturb test result.

● Avoid long time contact with air, or sample will volatilize, thus affects test result.

● Incorrect parameter setup could effects test result.

● System maintenance that not conform to user manual could cause contamination and instrument damage,
thus effects test result.

● Revise or add test result is not recommended by our company. And we are not responsible for this
operation.

Click “ ” key. Operator could check, delete, modify, audit, report preview, report printout, manual
recheck, history review of test result.

(1) Daily result

For checking daily result, select “the same day results” in “ ”. All daily sample
information can be displayed in the “ result data” menu as figure 6-24 shows.

111
User Manual of CS-400 Auto-Chemistry Analyzer

Figure 6-24

Sample information is at the left side of menu. Sample test result is at the right side of menu.

(a) Reaction curve.

In order to check sample reaction curve, click “ ” key, select the desired sample number and
test item, Select the wavelength type, that is dominant wavelength and secondary wavelength. The absorbency
of each light point is connected by line. As figure 6-25:

Figure 6-25

112
User Manual of CS-400 Auto-Chemistry Analyzer

In order to check the absorbency value of test light point, select the desired point in pull down menu. The
absorbency value can be displayed.

(b) Review

In order to audit single sample, single-click the record “ “ key. In order to check batch

sample, single-click “ ” key, input the start and end sample number in “ sample number

range” functional block, and then click “ “ key. As figure 6-26 show:

Figure 6-26

Single-click “ close” key to exit.

Note 1: The start sample No. should be less than or equal to the end sample No..

Note 2: Sample without result or name is not allowed to audit.

Note 3: Sample review cannot be modified and deleted.

(c) Report preview and print

Single-click “ ” to preview the record to be printed, as figure 6-27shows. Click “ cancel


“ key to exit.

113
User Manual of CS-400 Auto-Chemistry Analyzer

Figure 6-27

In order to print single sample report, choose the record, then click “ ”key.

In order to print batch sample report, single-click “ ” key. Input the start and end sample No.

in “ sample No. range” functional block, and then click “ “ key . For example, print sample
result from 1 to 14, as figure 6-28 show:

Figure 6-28

Single-click “print the audited report”, this procedure only apply for checked sample report, not for the
unchecked report. If don’t choose this function, all report will be printed out.

Single-click “ ” key to exit.

114
User Manual of CS-400 Auto-Chemistry Analyzer

Note: When printing batch report, the number of start sample must be less than the number of end sample.

(d) Modify and delete the result

Double- click the sample record to be modified, enter the menu. Input the new result in the “ check result”
functional block, and then click “ save” key. As figure 6-29 shows.

Figure 6-29

Click “ close” key to exit the current menu.

Note: addition of result and modification are not allowed if no result happens to some samples.

Single- click the sample record to be deleted. Then click ” key , as figure 6-30 shows .

Figure 6-30

Click “ ok” key to finish this procedure.

(e) Addition result.

All samples except those are being tested without result can be added with both analytical item and manual
item.

Click “Super add Result” key in “test result” to enter addition result form, select , as
figure6-31 shows.

115
User Manual of CS-400 Auto-Chemistry Analyzer

Figure 6-31

Addition of analytical items: this method is applicable to the sample registered already which is being tested.

● Select analytical item sample No. in “sample No.” pull-down functional block.

● Select short form of analytical item sample No. in “analytical item” pull-down functional block.

● Select disk No. in “disk No.” pull-down functional block. Input the place of addition sample in “position”
input block.

● Display other information of the sample in the input block “sample type”, “sample volume”, test date”.

● Click “addition” key

● Click “start” key in “start condition” form to carry out the addition test.

Add manual item: this method is applicable to the items not tested in the instrument but needed to add result in
“test result”.

Click “ ” key in “addition result” form to start manual item addition, as figure 6-32 show.

116
User Manual of CS-400 Auto-Chemistry Analyzer

Figure 6-32

● select additional manual item sample No. in “sample” pull-down functional block..

● select additional manual item short form in “item short form” functional block.

● input manual item result in “test result” input block.

● display other information of the sample in the input block “item full form”, “unit”, “reference range”.

● click “addition” key.

(2) Check results within three days

Single-click “ ” radio button.

(3) Query all the result

Single-click “ ” key to check all data. As figure 6-33 shows:

117
User Manual of CS-400 Auto-Chemistry Analyzer

Figure 6-33

(a) Query by date

Select the inquiry record as the start date: ” to the end date

“ ”. Single-click “ ” key, the qualified record will be displayed in the


functional block. In order to check one day record, input the same date in block.

(b) Query by patient name

Input the patient ’s whole name or last name in “ ” functional block, click

“ ” key. The qualified record will be displayed in the functional block.

(c) Query by sample number

Input the sample number in “ ” functional block, click “ ”key.


The qualified record will be displayed in the functional block.

(d) Query by case number

Input the case number in “ ” functional block, click “ ” key. The


qualified record will be displayed in the functional block.

(e) Query by bed number

118
User Manual of CS-400 Auto-Chemistry Analyzer

Input bed number in “ ” functional block, then click “ ”key. The


qualified record will be displayed in the functional block.

(f) Query by barcode

Input the sample barcode in “ ” functional block click

“ ”key. The qualified record will be displayed in the functional block.

(g) Query by delivery dept.

Select the delivery dept. in the pull-down menu of “ ” , click

“ ”key. The qualified record will be displayed in the functional block. Contents of the
pull-down menu is from “ hospital info.” from.

(h) Query by delivery doctor

Select the delivery doctor in the pull-down menu of “ ” , input either the doctor’s

whole name or the last name , then click “ ” functional key , The qualified record will be
displayed in the functional block. Contents of the pull-down menu is from “ hospital Info.” form

Click“ ” key to exit, and then continue other operation.

6.2.11 Sample recheck


There are two register methods: automatically registration and manual registration. Sample number is not
change when recheck. Reset the sample volume in “chemistry parameter” form.

(1)Automatic registration.

After sample test finished, instrument will automatically add sample info into reset form. The reset information
is defaulted set the same as first test. Operator can alter the data. Relationship between sample type and
sample volume is as figure 6-34 shows:

Conditions for automatic registration Sample type

The lower limit of technical limit over Increment

The upper limit of technical limit over Decrement

The limit of reaction absorbency over (substrate exhaust) Decrement

Prozone check value over (Antigen overmuch) Decrement

Figure 6-34

119
User Manual of CS-400 Auto-Chemistry Analyzer

Single-click “ ” key, instrument will carry out recheck test.

(2)Manual registration

Click “ ” key in the result data menu to add this item to recheck menu. Set up the sample

position and sample volume, click “ ” key, as figure 6-35 shows:

Figure 6-35

Single-click “ ” key, instrument will carry out recheck test. Click the “ ” key
to exit.

6.2.12 Analyze complete


(1)Recheck test result

After finishing measurement, confirm, audit or print the recheck sample in “ ” menu.

(2)Database backup

120
User Manual of CS-400 Auto-Chemistry Analyzer

Backup database In “ ” menu to avoid data lost.

(3)System sleep

Sleep indicates that instrument is at half-stop status, only cooling system keeps working. Instrument will
automatically start up in specified time.

Set up the time to awaken the instrument in “ system setup “ menu. Click “ sleep “ key. Instrument will shut off
all power supply besides cooling system. Status bar remind of “ instrument sleep”. Sleep mode can be set only
in the stand-by state.

If awakened time is not coming, click “ awaken” key to relieve sleep mode, instrument will carry out the same
operations as power on.

(4)Turn off instrument

Exit the software program of CS-400, turn off the power supply according to the following sequence: printer
power supply, host computer power supply, display power supply, analysis system of control power supply,
general power supply (general power cannot be cut when reagents are in cooling unit).

(5)Preparation for next measuring

Check if the reagent lid is closed well or not. Take away the sample cup or test tube with Calibrator, Controls,
diluent and sample. Drain the waste solution. in waste barrel; Check if reagent probe, sample probe and stirring
rod are contaminated or bended. Check if the surface of analyze unit /operation unit is dirty.

121
User Manual of CS-400 Auto-Chemistry Analyzer

Chapter 7 Calibration Information

Single-click the key to carry out the registration of calibration information and the check of
calibration result.

7.1 Colorimetric calibration

7.1.1 Calibration registration for colorimetric items

Single-click “ ” key in calibration information menu, then click the

“ ” key, as the figure 7-1 shows:

Figure 7-1

After registering, carry out calibration setup in “ ” form. Select “calibration before test” function.
Registration in “ calibration” form is not enough, without register in “start analysis” form, instrument will only
accept the calibration application, and calibration test won’t be carried out. Calibration rest cannot be renewed
either.

(a) Select the proper calibration type in “calibration type” menu. Calibration type is as table 7-2 shows:
122
User Manual of CS-400 Auto-Chemistry Analyzer

Calibration Standard Calibration


Calibration method Application example
Type solution volume result
Blank Blank solution renew reagent All points calibration K coefficient method,
calibration blank value method when standard solution
test is omitted.

Span One point except Renew k value Two points linear and Recheck standard solution
calibration blank solution. multi-points linear point 1
Calibration
Logit- Log3P
Logit- Log4P

2 points Blank solution Renew reagent Linear two points and Linear 2 points method
calibration and span points blank value and multi-points linear finish standard curve.
K value calibration Multi-point method,omitted
Logit- Log3P the quantity of calibration
Logit- Log4P solution.

Full points All registered Renew standard Linear multi-point Multipoint standard curve
calibration calibrator curve by all Calibration isozyme method
points isozyme Q
isozyme P
nonlinear working
curve calibration

Table 7-1

For detailed information please refer to chapter “2.2.2, 2.2.3”

(1) Select desired calibration item

Select the item name to be calibrated in “ registered calibration item” form. As figure 7-2 shows:

Figure 7-2

(a) Single-click “register calibration item” key, the selected items will automatically be displayed in registration
list. New registration item will carry out calibration. If not, even select the “calibration before test” function,
instrument will not carry out calibration. The concentration and position of Calibrator can be showed in the
form.

(b) In order to delete calibration item , select it and then click “delete” key.

7.1.2 Calibration result of colorimetric item


(1) Calibration result

Single-click “ ” key, check calibration result, such as: Reagent blank , K factor,

123
User Manual of CS-400 Auto-Chemistry Analyzer

approximate function of constant A.B.C from multi-points calibration curve etc. As figure 7-3 shows:

Figure 7-3
In order to modify the calibration result, delete the old result, input the new one, and then single-click

“ ” key.

(2) Calibration curve

(a) Single-click “ ” key in “ calibration result “ menu. and then select the desired item name to
be checked. The item of calibration type, S1ABS, K, A, B, C will be displayed in the form.

(b) The reaction curve of the calibration item is showed as figure 7-4. The abscissa represents the concentration,

the ordinate represents the absorbency. Absorbency range can be changed by revise “ ”. Click

“ ” key to exit.

124
User Manual of CS-400 Auto-Chemistry Analyzer

Figure 7-4

(3)Calibration tracing

Instrument will automatically store the absorbency of Calibrator. The tracing graph can check the stability of
absorbency variety; therefore the calibration accuracy can be checked.

(a) Single-click “ ” key, select the item name, the number of Calibrator. then click “ renew”
key. The 50 times absorbency value will be displayed in the graph. The abscissa represents the calibration
times, the ordinate represents the absorbency value. Absorbency range can be altered by revise

“ ”. as figure 7-6 show.

(b) In order to print the calibration trace graph, single-click “ ” key.

(c) Single-click “ ” key to exit.

125
User Manual of CS-400 Auto-Chemistry Analyzer

Figure 7-5

(4)Reaction process

Check the absorbency variety of each item in different time point through reaction monitor form. Check the
reaction status and check if the test value of absorbency is stable or not through reaction curve graph.

(a) Single-click “ ” key in “calibration ” menu, select the item name and the Calibrator number.
Because of each Calibrator is tested twice, select the test times, main wavelength, sub wavelength etc., then the
reaction curve graph of Calibrator will be displayed. The abscissa represents the photometric point, the ordinate
represent the absorbency as figure 7-6 shows:

Figure 7-7

(b) In order to check the detailed absorbency value of one photometric point, select the desired point in
“ photometric point” pull-down list. The absorbency value will be displayed.

126
User Manual of CS-400 Auto-Chemistry Analyzer

(c) Change the absorbency range by revising “ ”. Single-click “ ” key to print out

reaction curve, Single-click “ ” key to exit.

— Main wavelength: display the reaction curve of main wavelength.

— Sub wavelength: display the reaction curve of sub wavelength

— Main wavelength- Sub wavelength: display reaction curve of main wavelength subtracting sub wavelength.

Note: after adding new item, calibration of the new should be implemented first, and in correct calibration
result may affect accuracy of result.

(5) print result

Click “print result” to preview the result of colorimetric item calibration. In order to print the result, click
“print” key upper side of screen.

7.2 ISE calibration

Single-click “ ” key in “ ” menu to check the registration and result of ISE


calibration items, as figure 7-7 shows:

Figure 7-7

127
User Manual of CS-400 Auto-Chemistry Analyzer

(1) Calibration registration

If carry out the calibration for ISE items, sign “√” in front of the item.

(2) Calibration result

Instrument will automatically display the calibration result when carries out ISE test calibration. Alarm is issued
when abnormality exists.

In ISE calibration, Calibrator has been measured for three times. In calculation, the mean value of the second
and the third times result is used.

Check the slope value, the concentration of internal standard liquid, compensation value in “ calibration result “.
After finishing ISE calibration, instrument will automatically calculate the compensation value. If the
compensation value is the difference between the input value and the test value of Calibrator 3. In order to

revise the compensation value, delete the origin value, and input the number, then click “ ” key.

(3) print result

Click “print result” to preview the result of ISE item calibration. In order to print the result, click “print” key
upper side of screen.

128
User Manual of CS-400 Auto-Chemistry Analyzer

Chapter 8 Quality Control


The goal of quality control in lab is to guarantee test result reliability for each sample. The reliability include two
aspects, one is precision: the test result is in good repetition, daily test result changes little, the main purpose is to
eliminate or reduce the influence caused by random error; the other one is high accuracy: that the test result is
correct, close to the truth, and eliminate or reduce the influence caused by system error.

Random error: The difference between test result and the mean value of the same target tested many times in a
repetitious conditions is called random error.

System error: The difference between true value and the mean value of the same target tested many times in a
repetitious conditions is called system error.

Accuracy: The integration of system error and random error in the test result, indicate the consistent degree
between test result and true value.

Precision: The consistent degree among many test result of one target in a specified conditions, indicate the
degree of random error magnitude among the test results.

L-J ( levey Jennings) QC chart: QC chart is a kind of graph with quality control limit. QC limit is controllable

analysis method to known specimen (QC sample) carry out repetitious test to get the mean value ( X ) and

standard deviation (SD). X ± 2 SD is warning limit. X ± 3SD is out of control limit.

8.1 QC registration

Single-click the “ ” key, then click the “ ” key. At most eight QC samples can be used
simultaneously to carry out quality control. As figure 8-1 shows.

129
User Manual of CS-400 Auto-Chemistry Analyzer

Figure 8-1

8.1.1 QC regulation setup

(a) Click “ ” key, the Westgard QC regulation is showed in figure 8-2:

Figure 8-2

Operator could select the needed QC regulation, click “ ” to save the setup. After setup, the
“QC interval” and “Monthly QC” may process QC analyze according to the regulation.

According to Westgard multi-rule judgment base, carry out the incontrollable analysis to the test result, as figure

130
User Manual of CS-400 Auto-Chemistry Analyzer

8-2 shows:

QC Data

NO
12S Under control

YES NO

NO NO NO NO
13S 22S R4S 41S 10 X

YES YES YES YES YES

Out of control

Figure 8-3

Judgment benchmark instruction:

1 2S : One QC result exceeds mean value ±2 SD, which is judged as warning regulation, and initializes other
regulation to check QC data.

1 3S : One QC result exceeds mean value ±3 SD, which is judged as lose control, this regulation is sensitive to
random error.

2 2S : Two consecutive QC result simultaneously exceed mean value +2 SD or -2 SD, which is judged as lose
control, this regulation is sensitive to system error.

R 4S : One control result exceeds mean value +2 SD, another exceeds -2 SD, which is judged as lose control,
and this regulation is sensitive to random error.

4 1S: Four consecutive QC results exceed simultaneously mean value +1 SD or -1 SD, which is judged as lose
control , this regulation is sensitive to system error.

10 : Ten consecutive QC results all are in the same side of mean value (higher or lower than mean value, no
X

requirement to the degree of deflection), which is judged as lose control, this regulation is sensitive to
system error.

8.1.2 QC name setup

Click “ ”button, and click “ ”, as figure 8-4 shows:

131
User Manual of CS-400 Auto-Chemistry Analyzer

Figure 8-4

Input Controls name in “QC name”, and then click “ ”key.

In order to delete, select the QC name in pull down menu, and click “ ”key, click

“ ”key to exit.

8.1.3 QC item registration


(a) Select the position of QC sample from C1-C8 in the “position” pull down menu.

(b) Select the item name in the “ QC name” pull-down menu.

(c) Input lots No. of Controls in the “ QC lot number” functional field.

(d) Select type of Controls, such as blood, urine, in “sample type” pull down menu.

(e) Input target value and standard deviation.

(f) Click “add” key when above parameters are correctly inputted. All inputted parameter is saved in the left
work area.

Note: after register the QC item, make sure check which item may carry out QC test, click “execute” key in
front of QC item.

In order to carry out QC test according to QC interval, input QC interval in “chemistry parameter”; In order to

carry out one time QC test before sample test, click “ ” key in “start analysis” form, and

then click “ ”.

132
User Manual of CS-400 Auto-Chemistry Analyzer

8.1.4 QC parameter modification


If registered parameter need modify, single-click the desired items in the left side of filed, the color will change.
Saved QC parameter will display in the right side of field. Directly input the parameter on the item, then

click“ ” key to finish this procedure.

8.1.5 Delete QC item


Single-click the item to be deleted in the left side of field to eliminate the registered

color changes after being clicked. Click “ ” key to finish this operation. All information of
this item will be deleted.

8.2 QC interval

The QC interval is set in the “chemistry parameter” menu, and instrument will automatically carry out QC test

according to the set interval sample number. After analysis finish, check the QC result in the “ ”
menu, and a QC result chart is showed as well. In the chart, the abscissa represent test times, the ordinate
represent concentration.

(a) Select QC item and lot No. in the “ ” pull-down menu, QC result will be displayed in the QC
chart, as figure 8-5 shows:

Figure 8-5
133
User Manual of CS-400 Auto-Chemistry Analyzer

(b) Single-click “ ” key to analyze QC data according to Westgard Multi-rule Judgment base.

(c) Single-click “ ” to check and modify the QC result, in order to revise the result , input the
new data , click “ result revise” key as figure 8-6 shows

Figure 8-6

In order to add QC result, single-click “Add” key after input desired QC item result in the “Result” functional
block.

In order to modify QC result, single- click the desired QC result in the left side working area, input
“modification result” into the “Result” functional block.

In order to delete QC result, single-click the desired QC result in the left side of working area, the click
“Delete” key.

Single-click “Close” to exit “QC result”.

(d) Single-click “ ” key to check the whole reaction process of QC test.

(e) Single-click “ “ key to print out the QC chart.

After QC test finish, instrument will automatically calculated real test QC target value (mean value), standard
deviation, coefficient of variation, range (alteration range) etc. data.

- ∑ Xi
i =1
Target value( X ):
N

134
User Manual of CS-400 Auto-Chemistry Analyzer

∑ ( Xi − MV )
2

i =1
Standard deviation(SD):
N −1

SD
Coefficient Variation(CV%): ×100%
mv
Range: X max − Xmin
N: test times, Xi : test result.

8.3 Monthly quality control

Select “ QC before test” in the “start analyze” form when test. After finish analysis, check the QC result in the

“ ” form. This abscissa of QC chart represent test date, ordinate indicate concentration.

(a) Select QC item and lot No. and month in the “ ” pull-down menu, QC result will be displayed in
the QC chart as figure 8-7 shows:

Figure 8-7

(b) Single-click “ ” key to analyze QC data according to Westgard Multi-rule Judgment base.

135
User Manual of CS-400 Auto-Chemistry Analyzer

(c) Single-click “ ” to check and modify the QC result, in order to revise the result , input the
new data , and click “ result modify” key.

(d) Single-click “ ” key to check the whole reaction process of QC test.

(e) Single-click “ “ key to print out the QC chart.

After Qc test finish, instrument will automatically calculated real test QC target value (mean value), standard
deviation, coefficient of variation, range (alteration range) etc, data.

136
User Manual of CS-400 Auto-Chemistry Analyzer

Chapter 9 System Setup


“ System setup “ menu includes: chemistry parameter, profile item, calculated item, cross contamination, report
format, ISE setup, system setup. As the figure 9-1 shows:

Figure 9-1

Click “ ” key to exit “ system setup “ menu.

9.1 Chemistry parameter

Click “ ” functional key in main functional field “ ”. Chemistry parameter


has three sub-menu: analyze parameter, calibration parameter, range parameter.

Note: After the parameter of each menu is edited, operator should click “ ” to save the data.

9.1.1 Add/delete item


Before edit chemistry parameter, add chemistry item first.

Click “ ” key at the lower right side of form, add or delete the test items as figure 9-2 shows:

137
User Manual of CS-400 Auto-Chemistry Analyzer

Figure 9-2 Add and delete items


After enter “add item” menu, input the “item number” “item name”, and click “add” key to finish this operation.

In order to delete item, move the scroll bar at the right side of list box to look for desired item, click it by mouse,
the color of chosen item will change, appear in front of chosen item, click “ delete “key, The item will be
deleted.

Carry out other operation after click “close” key.

9.1.2 Analysis parameter

Click “ ”key in “ ” menu. Operator can edit or revise the analysis parameter
of colorimetric items. As figure 9-3 shows:

138
User Manual of CS-400 Auto-Chemistry Analyzer

Figure 9-3

— Item name: Select the item’s abbreviation from the pull-down menu. All items may display automatically in
the list box.

— Decimal digits: Operator can select the decimal digits of test result and printout result in the pull-down list
box.

— Items full name: Input the full name of testing item, such as ALT, whose full name is Alanine
Aminotransferase.

— Quality control interval: Input the quantity of interval sample. Input the integral number of 10 times, the least
interval quantity is 10, the most is 1000.

— Test method: Select one method in the pull-down menu which is conform to reagent requirement, 1 point
assay, 2 point assay, rate A assay, rate B assay, 1 point rate assay, 2 point rate assay, 3 point assay. For
detailed introduction, please refer to 2.2.1

— Item unit: Choose the chemistry item in the pull-down menu, For the add and delete of item unit, please
refer to chapter “ 8.3 ”

— Test time: Test time can be selected from the “test result” pull-down menu.

— Photometry point: Instrument will record one time absorbency every 18 seconds. Please input proper
photometry point according to reagent instruction. Effective photometry point should be inputted within 2 to
49 (0 represent no input). Tested absorbency value of each test light points can be searched from reaction
curve.

— Main wavelength: There are 12 optional wavelengths in pull down menu, select them according to reagent
manual. Main wavelengths : 340 nm、380 nm、405 nm、450 nm、480 nm、505 nm、546 nm、570 nm、
600 nm、660 nm、700 nm and 750nm.
139
User Manual of CS-400 Auto-Chemistry Analyzer

— Sub wavelength: When adopting double wavelength assay to test or analyze sample, select the wavelength
from 12 wavelength in pull down menu. Secondary wavelength include: 340 nm、380 nm、405 nm、450
nm、480 nm、505 nm、546 nm、570 nm、600 nm、660 nm、700 nm and 750nm . The difference of the
absorbance value between main wavelength and secondary wavelength is used for calculated result. when
select single wavelength test, select “0” of secondary wavelength.

— Instrument factor (Y=a X + b): Carry out relation calibration. The test result will be higher or lower than
expected result or result from other instrument. In order to make result in accordance with expected result or
result from other analyzer, add the calibration relation in result calculation.

Relation equation:

Y=a X + b

Y : Result after calibration

X : Real result from analyzer

a : Slope value (multiplication calibration factor)

b : Intercept value (compensation calibration factor)

When test result of the analyzer is the same as expected value, or when results of any two analyzer is
accordant, let a=1, b=o.

When results from two analyzers are different, analyzer can get an accordant result by calibration of slope
value and intercept value. Slope value is a positive number, which is less than 8 digits. Intercept value is a
real number, which is less than 8 digit.

— “sample volume” work area: Sample volume includes normal volume, increased value and decreased value.
Left side of sample volume work area is for serum sample, right side is an optional area. Operator can select
sample type from pull down menu, as figure 9-5 shows:

Figure 9-4 sample volume

Normal volume

In “ normal volume” work area operator can specify sample normal volume. This work area are divided
three functional key field: sample volume, diluted sample volume, diluent volume.

[ Normal/ sample volume] : The sample volume (2ul to 35ul) is absorbed from sample container( sample
cup or tube) . If predilute is not required, the total volume of sample and reagent should be greater than or

140
User Manual of CS-400 Auto-Chemistry Analyzer

equal to 150ul, smaller than or equal to 450ul. If diluent is required, the total volume of sample and
detergent should be greater than or equal to 105ul, the total volume of diluted sample and reagent should
be greater than 150ul, less than or equal to 450ul.

[ Normal/ diluted sample volume ] : If predilute is required, the parameter is used to set up diluted sample
volume which sucked from dilute cup and infuse to reaction cuvette that is used for analyze reagent. The
input value is within 2ul to 35ul. Input 0 to avoid predilute.

[Normal/diluent volume]: If predilute is required, the parameter is used for set up diluent volume for
dilute sample. The input value is within 2ul to 350ul. Input 0 to avoid predilute.

Decreased volume ( sample volume decrease)

“Decreased volume” work area is used for specify sample volume when sample concentration exceed the
upper limit of reagent linearity range. (lower than normal volume). This area is divided into 3 functional
key field: sample volume, diluted sample volume, diluent volume. After test, instrument will automatically
change decreased test result into normal volume and display on result information.

[Decrease/ sample volume]: Select a sample volume (from 2ul to 35ul, and less than normal volume),
which is sucked from sample container (sample cup or tube). If predilute is not required, the total volume of
sample and detergent should be greater than or equal to 150ul, smaller than or equal to 450ul. If dilute is
required, the total volume of sample and detergent should be greater than or equal to 150ul, the total
volume of diluted sample and reagent should be larger than 150ul, less than or equal to 450ul.

[Decrease/diluted sample volume] : If predilute is required, the parameter is used for set up diluted sample
volume which is sucked from dilute cup and infuse to reaction cuvette that is used for analyze reagent. The
input value is within 2ul to 35ul, Input 0 to avoid predilute.

[ Decrease/diluent volume] : If predilute is required, the parameter is used for set up diluent volume for
dilute sample . The input value is within 2ul to 350ul, Input 0 to avoid predilute.

Increased volume( sample volume increase)

“Increased volume” work area is used to specify sample volume when sample concentration exceed the
lower limit of reagent linearity range. (more than normal volume). This area is divided into 3 functional key
field: sample volume, diluted sample volume, diluent volume.

[Increase/ sample volume] : Select the sample volume (2ul to 35ul,more than normal volume) ,which is
sucked from sample container ( sample cup or tube). If predilute is not required, the total volume of sample
and reagent should be greater than or equal to 150ul, smaller than or equal to 450ul. If predilute is required,
the total volume of sample and detergent should be larger than or equal to 105ul, the total volume of diluted
sample and reagent should be larger than 150 ul, smaller than or equal than 450ul. Input 0 to avoid predilute

[Increase/diluted sample volume] : If predilute is required, the parameter is used for set up diluted sample
volume which sucked from dilute cup and infuse to reaction cuvette that is used for analyze, The input
value is within 2ul to 35ul. Input 0 to avoid predilute.

[Increase/diluent volume] : If predilute is required, the parameter is used to set up diluent volume for
dilute sample . The input value is within 2ul to 350ul, Input 0 to avoid predilute.

— “Reagent “ work area: Contents of reagent include reagent volume and reagent position. Reagent 1(R1) ,

141
User Manual of CS-400 Auto-Chemistry Analyzer

reagent 2 (R2), reagent 3 (R3), reagent 4 (R4) are reagent type . R1 and R4 are in reagent disk 1, R1 probe
absorb reagent. R2 and R3 are in reagent disk 2, R2 probe absorb reagent. As figure 9-5 shows:

Figure 9-5

Reagent volume: Reagent unit: ul. Reagent probe can exactly suck 20ul to 350ul solution. “0” represent no
reagent is added.

Position: Display reagent position in reagent disk. Register this parameter in “ reagent info.”

— Second half item of two test analyze: In order to carry out two test analyze, select the name of second half
item in “second half item of two test analyze” pull down menu.

— Prozone check: input the range value of prozone check, setup the upper limit and lower limit.

— Absorbance limit: Input the range value of absorbance, setup the positive reaction and negative reaction.

9.1.3 Calibration parameter

Click “ ”button in “ ” form, as figure 9-6 shows:

142
User Manual of CS-400 Auto-Chemistry Analyzer

Figure 9-6

(1) Select item name in “select test ”pull-down menu.

(2) In terms of reagent instruction, Check calibration type, calibration point, span point, and weight
coefficient etc. parameter.

(3) Input the concentration and position of Calibrator.

(4) Input the parameter of calibration check. For detailed operation, please refer to chapter “ 2.3.1 ”

(5) For automatic calibration, input the overtime of automatic calibration according to relevant calibration
type. If automatic calibration time comes when stand-by, instrument will automatically carry out
calibration before test next time. In same item, if automatic calibration conflict with manual calibration,
carry out manual calibration only. Click “0” in time position if automatic calibration to avoid automatic
calibration.

(6) After check the parameter, click ” save ” key to save.

— Calibration point: Input the quantity of Calibrator in functional field, (among 1-6).

— Span point: Input the relevant span point in functional field (among 2-6).

9.1.4 Range parameter

Click “ ” key in the “ ” form to setup the reference value range and
linearity range. Click “save ” button to save the parameter.

143
User Manual of CS-400 Auto-Chemistry Analyzer

Figure 9-7

— Special reference range: If patient age, and patient gender are different, the reference value range are

different too. Click “ ” key. For example, the special reference range of UA:
0.1~0.34 for male aged 0~15, female 0.12~0.33 aged the same; 0.21~0.43 for male aged 14~50, 0.15~0.36
for female aged the same; 0.28~0.50 for male aged more than 50, 0.21~0.43 for female aged the same. As
figure 9-8 shows:

Figure 9-8

— Default reference range: select default reference value range “ ”when the reference value
144
User Manual of CS-400 Auto-Chemistry Analyzer

range are the same despite the difference patient age and gender.

For example, the default reference value range of urine amylase is 0~640U/L , as figure 9-9 shows.

Figure 9-9
— Linearity range: Input the upper limit and lower limit of reagent linearity in functional field. Alarm is issued
when test result exceed the range.

Note: The chemistry parameter in the manual is taken as an example, not real test parameter. Operator should set
parameter according to reagent manual .

9.2 Item combination

Click key “ system setup” form, set up item combination, as figure 9-10 shows:

Figure 9-10

(1) Input the number of combination item in “ number” functional field, the number can’t be repeated, or it cannot
be saved.

(2) Input the name of item combination in “ item combination name” functional field, Character and number are
all allowed, but can’t be repeated. Or it cannot be saved.

145
User Manual of CS-400 Auto-Chemistry Analyzer

(3) Select item of item combination, click check box in front of item name, indicate that is selected.
Click check box again, and indicate that it is cancel.

(4) Click “ ” key , Combination item number, name will be displayed in the right side of functional
field. Click the number or name, all items will be automatically showed in functional field.

(5) Select the number or name of desired combination item, click “ ” key to delete the
combination item.

9.3 Calculated item

Test Calculation is on the base of two or more item test results, use special calculation method to get a new item,
such as A/G..

Click key in “ ” form. As figure 9-11 shows:

Figure 9-11

(1) Input test name in “ test item” Functional field.

(2) Input item name in “ test full name” functional field.

(3) Select the unit of new calculated item in “unit” pull down menu. Select the decimal digits of new calculated
item in “ decimal digits” pull down menu.
146
User Manual of CS-400 Auto-Chemistry Analyzer

(4) Select the reference value range of new calculated item in “ reference value range” pull down menu.

(5)Edit the calculation formula the edited information will be showed in “ calculation formula” functional field,
click “add” key to finish edition work.

(6) In order to delete calculation item, click the desired item, and then click “delete ”key.

Edition method of calculation formula ( for instance: A/G )

Set up item name, unit, decimal digit, reference value range according to the above procedure, select 【ALB】
in “test name” pull down menu, select【/】 、【(】in “sign” pull down menu, select 【TP】 in “item name” pull
down menu, select 【-】 in “sign” pull down menu. The edit of formula can be showed in “calculation formula”.
If there are some number in formula, select any number among 0、1、2、3、4、5、6、7、8、9 in “number” pull

down menu. After input finish, click key, the formula will be showed at the right side of
functional field.

Figure 9-12

Click key beside the calculation formula functional field to delete the wrong content.

9.4 Cross contamination

Cross contamination obviation is a function to avoid cross contamination among analyzing items. The degree of
cross contamination is different due to different reagent ingredient. For avoiding cross contamination among
reagents, we strongly suggest separate the items with cross contamination and the items without cross
contamination. If all item cannot be separated, automatic washing function can be added before tested item in
order decrease the cross contamination in maximum extent. But the cross contamination function may decrease
the test velocity.

Cross contamination include: reagent probe, reaction cuvette, sample probe. Detergent is located at position 45
in reagent disk 1 and reagent disk 2.

Click key in “ “ menu.

9.4.1 Reagent probe cross contamination

Click “ ” key in “ “ menu to setup avoid cross contamination of reagent


probe. As figure 9-13 shows:

147
User Manual of CS-400 Auto-Chemistry Analyzer

Figure 9-13 Reagent probe cross contamination

(1) Select reagent probe R1 or R2 in reagent work field. indicate R1 has been selected.
Input the detergent volume (ul).

(2) Select the reagent type in “item” pull down menu in “from reagent” work area.

(3) Select reagent type in “item” pull down menu in “to reagent ” work area

(4) Click “ ” key, the set information will be displayed in left side functional field.

(5) Click “ ” key, relevant information will be deleted.

9.4.2 Reaction cuvette cross contamination

Click “ ” key to set avoiding cross contamination. As figure 9-14 shows:

148
User Manual of CS-400 Auto-Chemistry Analyzer

Figure 9-14

(1) Select item that needs to be set in “item name” pull down menu.

(2) Input R1 detergent volume (ul) in “ R1 detergent Volume” work area.

(3) Input R2 detergent volume (ul) in “R2 detergent volume” work area.

(4) Click “ ” key, the set information will be displayed in left side of functional field.

(5) Click “ ” key, information will be deleted. And click “ close” key to exit the current menu.

9.4.3 Sample probe cross contamination

Click “ ” key to set “sample cross contamination obviation”, as figure 9-15 shows:

149
User Manual of CS-400 Auto-Chemistry Analyzer

Figure 9-15

(1) Select the items to be set in “ item name ” pull -down menu.

(2) Select detergent position from W1,W2,W3.

(3) Click “ add” key, the set information will be displayed in the left side of functional field.

(4) Click “delete” key , information will be deleted. Click “close” key to exit the current menu.

9.5 Report sheet format

Click key to set printout info. and format. As figure 9-16 shows:

150
User Manual of CS-400 Auto-Chemistry Analyzer

Figure 9-16

9.5.1 Basic information setup


(1) Input the first name and second name of the organization in “basic report information” work area. If there is
no second name, do not input it.

(2) If endnotes of report is used, please sign “√” in “ report endnotes” functional field. Input the first and second
line of report content. Endnotes may not be inputted if no need.

(3) Please sign “√” in “ automatic add calculated item” functional field to add the item name into print list.

(4) Click “ ” key to save basic information of report.

9.5.2 Print sequence setup


Input the print sequence of report item in “ print order” work area. Print them in the sequence of 0,1,2,3,from

small to large. If select 0, instrument will print according to the item number . Click “ ” to save
the setup parameter.

9.5.3 Report printout format setup

Click “ ” key to set report template and print option.

9.5.3.1 Report template setup

151
User Manual of CS-400 Auto-Chemistry Analyzer

(a) Click “ ” to select report template in pull down menu, preview the printout report

template at the right side of form, adjust the preview scale in “ ” functional field.

As figure 9-17 shows.

Figure 9-17

Figure 9-18

(b) Click “ ” key to edit template in template menu.

(c) Select one template and click “ ” key to delete template..

(d) Click “ ” key to input new template name, As figure 9-18 show.

(e) Click “ ” key to exit “ report template setup “ menu.


152
User Manual of CS-400 Auto-Chemistry Analyzer

9.5.3.2 Print option setup

(a) Click “ ” key to set printout report option . As figure 9-19 show.

Figure 9-19

(b) Select the contents of report and sign “√” in functional field. Select “print all item”, and all reminding, item
code, check result , item name, unit , reference value will be printed.

(c) Select and click “ ” to save select information.

(d) Click “ ” key to exit “ print option setup” menu.

9.5.3.3 default format setup

Click “ ”key in “report format” work area. And select a report format.

9.6 ISE Setup

Single-click “ ” key in “ system setup” menu to setup ISE parameter. As figure 9-20 shows.

153
User Manual of CS-400 Auto-Chemistry Analyzer

Figure 9-20

(1) Select the items name and sample type, decimal digits, item unit, QC interval, etc.

(2) Input the concentration of Calibrator in “calibration parameter” work area and select the relevant calibration
position. Calibrator 1 is ISE low concentration Calibrator. Calibrator 2 is ISE high concentration Calibrator.
Calibrator 3 is compensation liquid.

(3) Input the remaining volume of Reference Liquid, Internal Standard Liquid and Diluent in reagent remaining
volume work area in “ml”, click “save ” key. Instrument will automatically subtract the wastage and display the
remaining volume in work area.

(4) Set reference value range in parameter setup work area.

(5) Single-click “ ” key to save the set data.

(6) Single-click “ ” key to exit the menu.

9.7 Other setup

Single-click “ ” key to set the sample barcode, reagent barcode, sample alarm, reagent alarm, ISE,

time wakening etc. function. Single-click “ ” key to save the parameter. Click

“ ” key to exit. As figure 9-21 shows:


154
User Manual of CS-400 Auto-Chemistry Analyzer

Figure 9-21

(a) Barcode setup

Check barcode device, if barcode device is connected well, and use would like to execute barcode device
checkup of sample disk, R1 reagent disk, R2 reagent disk, select “√”in functional block, if barcode device is
not connected, use should not select “√”,or it alarm may occurs. Scanning barcode before sample analyze, in
order to checkup sample barcode scanning function, select “√”in functional block.

Scan reagent open area barcode: Reagent close area must execute reagent barcode scanning, in order to scan
reagent open area barcode, select “√”in functional block.

(b) ISE setup

Execute ISE: Sign “√” in ISE setup if any test required ISE test.

(c) Timing wakening:

If sign “√” in functional block, the instrument will automatically be awakened in specified time. Input the
desired date in awaking date list, and the wakening time in functional block ( such as : ** hour ** minute)

(d) Reagent alarm

Reagent remaining times: input reagent remaining test times. When instrument detect the remaining test times is
less than set value, an alarm will be issued.

Reagent remaining volume: input reagent remaining volume. When instrument detect the remaining volume is
less than set value, an alarm will be issued.

(e) ISE reagent alarm


155
User Manual of CS-400 Auto-Chemistry Analyzer

Reference Liquid remaining volume: Input the Reference Liquid volume for alarm in “ml”

Internal Standard Liquid remaining volume: Input the Internal Standard Liquid volume for alarm in “ml”.

Diluent remaining volume: Input the Diluent volume for alarm in “ml”.

9.8 Manual item setup

“Manual item ” means the test is not processed in the instrument, but to add the result by handwork in “test
result” block.

Single-click “manual item setup” key to setup the item short form, item name, item unit, reference range.
Single-click “add” to save parameter after setup as figure 9-22 shows.

Figure 9-22

Single-click “close” to exit.

(a) Input the manual item short form in the “item short form” functional block.

(b) Input the manual item name in the “item name” functional block.

(c) Select the manual unit in the “unit” pull-down block.

(d) Input the manual item reference value range in the “reference range” functional block.

(e) In order to delete the set information, single-click the set information in work area, single-click “delete” key.

9.9 LIS communication setup

For LIS communication setup function, please refer to LIS function user manual

156
User Manual of CS-400 Auto-Chemistry Analyzer

Figure 9-23

157
User Manual of CS-400 Auto-Chemistry Analyzer

Chapter 10 System management

10.1 User information

The user with administrator permission can add, delete or modify the user information.

Single-click the “ ” key in main functional filed, then click “ ” key to set the operator ID,
name, password, reconfirm password ( two times inputted password should be the same) , mnemonics, etc. As
figure 10-1 shows:

Figure 10-1

Single-click “ ” key, the relevant info. will be displayed in the form. Click “ ”
key to eliminate the user information.

Click “ ” key to exit .

— Administrator: User with administrator permission can set, delete, check , browse and test all functions

— Inquiry: some function windows are available to user, setup and test unavailable.

— Operation: User with operator permission can set, delete, check, browse and test all functions except for user
information.
158
User Manual of CS-400 Auto-Chemistry Analyzer

10.2 Hospital information

Single-click the “ ” key in “ ” menu, to set the delivery dept. delivery doctor, sample
type, patient type and item unit, as figure 10-2 shows :

Figure 10-2

Single-click “ ” key to exit the current menu.

These items will be displayed automatically after setup in “patient info.” form corresponding pull-down block.

10.2.1 Delivery dept

Single-click the “ ” key to set department number, name and mnemonics

— Dept. No.: Input the dept. No. The information will be displayed in pull-down menu of “ sample register ”
menu.

— Delivery dept.: Input the name of delivery dept., the information will be displayed in the relevant pull-down
list of “ sample register” menu.

— Mnemonics: Help the user input the information quickly. For example: the mnemonics of department
numbers can be set as KSBH, then the contents of patient type option can be replaced by inputting KSBH,
click “enter” key, the department number will be automatically inputted.
159
User Manual of CS-400 Auto-Chemistry Analyzer

10.2.2 Delivery doctor

Single-click “ ” key in functional field to set “Doctor No.” “ Doctor Name” etc. information. As figure
10-3 shows:

Figure 10-3

Click “ ” key, all the information will add to delivery doctor menu. Select the info. bar to be
deleted, then click “ delete” key , the info. bar will be eliminated

10.3 Other information

10.3.1 Sample type

Single-click “ ” key in “ ” form, then set the serial number, sample type,
mnemonics etc information . As figure 10-4 shows:

160
User Manual of CS-400 Auto-Chemistry Analyzer

Figure 10-4

Select the info. bar to be deleted, then click “ ” key , the info. bar will be eliminated.

10.3.2 Patient type

Single-click “ ” key in work area to set serial number, patient type, mnemonics etc. as the figure
10-5 shows:

161
User Manual of CS-400 Auto-Chemistry Analyzer

Figure 10-5

Select the info. bar to be deleted, then click “ ” key , the info. bar will be eliminated.

10.3.3 Clinic diagnosis

Single-click “ ” key in work area to set the clinic diagnosis information. As the figure 10-6
shows:

162
User Manual of CS-400 Auto-Chemistry Analyzer

Figure 10-6

Select the info. bar to be deleted, then click“ ” key , the info. bar will be eliminated

10.3.4 Report remark

Single-click “ ” key in working area to set the report remark information, as the figure 10-7 shows:

163
User Manual of CS-400 Auto-Chemistry Analyzer

Figure 10-7

Input No. remark, mnemonics in functional block, click “add” to register remark.

Select the info. bar to be deleted, then click “ ” key , the info. bar will be eliminated.

10.3.5 Test unit

Click “ ”key to set the test unit, as figure 10-8 shows:

164
User Manual of CS-400 Auto-Chemistry Analyzer

Figure 10-8

Click “ ” key, the inputted information can be showed in test unit form. Select the info. bar to be

deleted, then click “ ” key , the info. bar will be eliminated

10.4 Workload statistics

Workload statistics function is used for checking the workload of delivery department, deliver doctor and check

doctor. According to the desired time slice select the statistic contents, click “ ” key to
complete statistic, result is showed as statistic chart, as figure 10-9 shows:

165
User Manual of CS-400 Auto-Chemistry Analyzer

Figure 10-9

Click “print” key to preview and print the statistics chart, as figure 10-10 shows:

Figure 10-10

166
User Manual of CS-400 Auto-Chemistry Analyzer

10.5 Database maintenance

In order to prevent data lost, database should be backup periodically.

Single-click “ ” to backup and recover the data. As the figure 10-11shows:

Note: Please implement database backup and recovery in offline status.

Figure 10-11

Database backup: Before database backup, user could select the save path, or the database will be saved to the
default software installation folder. The file name of database is the current date plus the current time, the
postfix is *.back. Periodically backup database can avoid the data lost. When the path of backup is selected,
click “ backup” key. The form will display the status of database backup and provide backup finish hint.

— Database recover: When the software can’t be used, the database backup file can recover the former data.
Select the save path of backup file, then select the backup file according to the date and time, click “ recover”
key. The form will display the information of recovered database. If the path and file of database backup are
not selected, a hint information will pop up.

167
User Manual of CS-400 Auto-Chemistry Analyzer

10.6 System log

Click ““ ” key in “ ” menu. As the figure 10-12 shows:

Figure 10-12

System log realizes these function to check system operation, including user login, operation log, maintenance
log and alarm log. Select one log type in “type” work area, select the start date and finish date in “date” work

area, click “ “ key, all the relevant log information will be listed.

168
User Manual of CS-400 Auto-Chemistry Analyzer

Chapter 11 System Help

Any questions arise when operate instrument, the user could click “ ” to seek help. As figure 11-1 shows:

Figure 11-1

11.1 System help application

(1) Single-click the “catalogue” key, check the desired menu in the list. Single-click the “ back” key to return
to the last menu. If instruction contents are not displayed completely, pull the scroll bar to browse the rest.

(2) Single- click the “ close “ key to finish this application.

169
User Manual of CS-400 Auto-Chemistry Analyzer

Chapter 12 System Maintenance

12.1 System maintenance preparation

To ensure the accuracy and precision of the CS-400, the user should operate strictly according to the CS-400
User Manual, and a regular maintenance is also a necessity. This is the only way to make sure a long useful life
and a reliable result, which is provided by instrument.

Please prepare the following items before carry out system maintenance

12.1.1 Instrument and Tools


(1) Accessories (prepared with CS-400)

Cross screwdriver……………..(for demounting instrument cover board)

Stainless steel ware (with diameter 03mm and 0.5mm)…..(for cleaning sample probe and reagents probe)

Fixing block…………………………………..(for adjusting the height of the stirring rod)

Cleaning probe tool…………………...….(for cleaning probe when blocked)

(2) To be prepared by user

Clean gauze…………………………………… (cleaning parts )

Swab………………………………………….(for cleaning sample probe and reagents probe)

Vacuum Cleaner………………………………(for cleaning cooling fan and radiator filter)

Buckets (two)…………………………………(for drainage the cold water and discharge the reagent waste)

Test tube brush………………………………….(for cleaning the cleaning tank)

Graduate or breaker (3L or 5L)…………………(for water supply to tank)

12.1.2 Pure water


For routine run, maintenance and checkup, please use purified water with conductivity 1 us/cm max. A regular
maintenance of the purified water equipment is also a necessity. Please operate according to the manual of the
purified water equipment or contact the supplier of the purified water equipment.

12.1.3 Detergent
The detergent is used for cleaning all parts of instrument. All kinds of detergents could be purchased from Dirui
company. Other brand detergents may cause the uncleanness of cuvette, reagent probe, sample probe, stirring
rod, pipe line, and finally result a cross contamination. Our company is not responsible for the inaccuracy,
which is conducted by the other kind of detergent.

There are five kinds detergents for CS serial:

CS-anti-bacterial phosphor-free detergent: Place CS anti-bacterial phosphor-free detergent on the 45


position on reagent disk 1, reagent disk2. Put 6ml CS anti-bacterial phosphor-free detergent in incubation bath
when exchange water. If the detergent is not added in, air bubble may attach on the cuvette, and the bacterial
170
User Manual of CS-400 Auto-Chemistry Analyzer

may grow in the incubation bath. Owing to there is no conductance, the liquid sensor can not detect liquid level
normally. The reagent probe may automatically suck detergent on 45 position for reagent probe cleaning. Wipe
all parts of instrument or immersion cuvette with 2% CS anti-bacterial phosphate-free detergent.

CS-alkaline detergent: The CS-alkaline detergent in the detergent bottle which located in the front of
instrument is used for cleaning cuvette; Detergent on W1 position in sample disk is used for cleaning sample
probe.

CS-acidic detergent: The CS-acidic detergent on W3 position is used for the cleaning which could avoid
sample probe cross contamination.

CS-ISE detergent: The CS-ISE detergent on W2 position is used for cleaning for sample probe, dilution bath,
ISE pipe line after ISE test.

CS-solenoid valve cleaning liquid: Clean the concentrated waste liquid pipeline.

12.2 The Application of system maintenance menu

Click “ ”button of the main function menu to start the instrument maintenance. Select maintenance

information in the list by mouse, or remove “↑”“↓”key on keyboard, click “ ” button, start
maintenance. Instrument will carry out reset operation first among all maintenance operation.

Some maintenance item allow stop in the midway, click “ ” to finish the maintenance operation.
For those not allow stop in the midway, take other operation after maintenance finish. In order to exit the system

maintenance menu, click “ ” button.

If there is an abnormity, an “alarm” hint will be showed on the menu.

12.2.1 Reset
Select “Instrument Reset” in “maintenance item list” work area, then click “Execute”. The instrument will
automatically return to the initial position. There will be an “alarm” hint if mistakes happen. Emergency stop is
not allowed while resetting. Take other operation when computer stand-by. Strongly suggest user execute reset
operation after emergency stop or after adjustment of reagent probe, sample probe and stirring rod.

12.2.2 Cleaning water tank


Select “water tank” in “maintenance item list” work area, and then click “Execute”. The instrument will
automatically cleaning the water tank immediately. Emergency stop is not allowed during operation. Take other
operation when computer stand-by. The water quality will be contaminated if bacterial grow in the incubation
bath.

12.2.3 Light quantity checkup


Select “light quantity checkup” in “Maintenance item list”, and then click “Execute”. The instrument will carry
out the “light quantity” checkup. The previous “absorbency value” and current one could be showed at the same
171
User Manual of CS-400 Auto-Chemistry Analyzer

time. (As figure 12-1 shows). Check the menu to choose printout the result or not. The absorbency value should
be ≤ 18000. Click “end maintenance” to complete the light quantity checkup operation.

Normally, carry out the light quantity checkup once a month. Carry out the light quantity checkup after replace
bulb, then proceed the test after the absorbency value qualified.

Figure 12-1

12.2.4 Cell blank check


Select “cell blank test” in “Maintenance item list” work area, then click “Execute”. The instrument will carry
out cell blank check for 120 cuvettes.

The cell blank check value will be showed on the system maintenance menu (as figure 12-2 shows). Click
“print” button to print out the cell blank check value. Click “end maintenance” key to complete the cell blank
check operation.

Normally, carry out the cell blank check once a week is suggested. Carry out cell blank check after replace the
cuvette. Then proceed the test after the cell blank check value qualified. Do not proceed sample test if the cell
blank check value is abnormal, it may influence the accuracy of the test result.

172
User Manual of CS-400 Auto-Chemistry Analyzer

Figure 12-2

“Cuvette number” tandem: display the number of 1-120 cuvettes

“340-750”tandem: display the cell blank value of 120 cuvettes corresponding to the different wavelength of
340、380、405、450、480、505、546、570, 600、660、700、750(nm).

“1”(No.1 cuvette ) rank: display the cell blank value of 12 type different wavelength of No.1 cuvette. A
<18000 cell blank value is considered as a qualified one.

“2”(No.2 cuvette) rank: display the difference between two cuvette: the difference cell blank value between
No.1 cuvette and No.2 cuvette, a ±800 difference value is considered as a qualified one.

3-120 cuvette rank: display the difference between 3-120 cuvette.

12.2.5 Air exhaustion of syringe


Select “syringe exhaust” in “Maintenance” work area. then click “Execute”. The plunger of syringe move up
and down in order to exhaust the air of syringe. Emergency stop is not allowed during operation. Take other
operation when computer stand-by.

Carry out Air exhaustion function when replace syringe or replace the connection pipeline of syringe.

12.2.6 Rinsing /air exhaust detergent pipeline


Select “Rinsing / air exhaustion of detergent pipeline” function, click “Execute”. The instrument will
automatically exhaust the air in detergent pipeline. Emergency stop is not allowed during operation. Take other
operation when computer stand-by.

Carry out this function when detergent bottle with CS-alkaline-detergent discharged, and there is air in the
173
User Manual of CS-400 Auto-Chemistry Analyzer

connection pipeline.

12.2.7 Rinsing reaction cuvette


Select “Rinsing reaction cuvette” function, click “Execute”, instrument will automatically Rinsing 120 cuvettes.
Click “end maintenance” key to end this operation.

Rinsing cuvette once a week in order to avoid the dirt in cuvette influence the test result. Carry out Rinsing
cuvette function when cell blank value abnormal, if cell blank value is still not qualified after Rinsing, replace
the cuvette.

12.2.8 Rinsing ISE


Select “Rinsing ISE” key, and click “execute” after ISE device is well connected. Instrument will automatically
rinse ISE system (Dilute bath and ISE pipeline),click “end maintenance” key to end this operation.

We suggest user execute Rinsing ISE once a day after analyze finish. ISE calibration must be execute before
next test.

Note : If ISE device is not connected, or connected but not set, all ISE maintenance cannot be used, the hint bar
may hint: please check if ISE device can be used.

12.2.9 Rinsing ISE+ Reaction cuvette


If ISE device is well connected, select “Rinsing ISE+ Reaction cuvette” key, click “execute” key, instrument
will rinsing reaction cuvette and ISE system (Dilute bath and ISE pipeline) together. Click “end maintenance”
key to end this operation.

We suggest user execute Rinsing ISE + Reaction cuvette once a week. If rinsing all is executed, rinsing ISE +
Reaction cuvette should not be carried out again.

12.2.10 Rinsing incubation bath


Click “Rinsing Incubation bath” function in “Maintenance item list”. And then click “Execute” .The instrument
will automatically carry out the whole process, expel the water from the incubation bath and infuse new purified
water. Meanwhile, reagent probe 1 and reagent probe 2 will suck the CS-anti-bacterial phosphor free detergent
at 45 position of reagent 1 disk and reagent 2 disk. Each probe suck 6 times, each time 500ul. 6 ml
CS-anti-bacterial phosphor free detergent is added in the incubation bath. Stop is not allowed in this operation.

Carry out incubation bath water replace function when constant temperature water is contaminated. Incubation
bath will automatically replace water when instrument start up. If the instrument is continuously used for more
than 24 hour, alarm issued to hint to carry out the incubation bath water replace function.

12.2.11 Sample probe vertical checkup


Select “sample probe vertical checkup” function, click “Execute”, instrument will carry out single-step vertical
operation of sample probe lift mechanism. Click “next” to proceed next operation, click “end maintenance” to
stop the maintenance operation. Stop is not allowed during maintenance operation, take other operation when
computer stand-by.

For detail introduction of sample probe vertical checkup, please refer to “12.4.1(4)”.

174
User Manual of CS-400 Auto-Chemistry Analyzer

Figure 12-3

12.2.12 Sample probe horizontal checkup


Select “sample probe horizontal checkup” function, click “Execute”, instrument will carry out single-step
horizontal operation of sample probe lift mechanism. Click “next” to proceed next operation, click “end
maintenance” to stop the maintenance operation. Stop is not allowed during maintenance operation, take other
operation when computer stand-by.

For detail introduction of sample probe horizontal checkup, please refer to “12.4.1(4)”.

This operation is taken when carry out sample probe position adjustment or sample probe position checkup.

12.2.13 Reagent probe vertical checkup


Select “reagent probe vertical checkup” function, click “Execute”, instrument will carry out single-step vertical
operation of reagent probe lift mechanism. Click “next” to proceed next operation, click “end maintenance” to
stop the maintenance operation. Instrument will automatically memorize the altitude of the test, and this would
view as reference value to calculate the remaining volume of reagent. Stop is not allowed during maintenance
operation, take other operation when computer stand-by.

For detail introduction of reagent probe vertical checkup, please refer to “12.4.1(4)”.

12.2.14 Reagent probe horizontal checkup


Select “reagent probe horizontal checkup” function, click “Execute”, instrument will carry out single-step
horizontal operation of reagent probe lift mechanism. Click “next” to proceed next operation, click “end
maintenance” to stop the maintenance operation. Stop is not allowed during maintenance operation, take other
operation when computer stand-by.

175
User Manual of CS-400 Auto-Chemistry Analyzer

For detail introduction of reagent probe horizontal checkup, please refer to “12.4.1(4)”.

This operation is taken when carry out reagent probe position adjustment or reagent probe position checkup.

12.2.15 Stirring mechanism horizontal checkup


Select “stirring mechanism horizontal checkup” function, click “Execute”, instrument will carry out single-step
operation checkup of stirring rod lift mechanism.

This function is carry out when adjust the stirring mechanism position (at the side of cuvette or on the top of
rinsing bath).

12.2.16 Mechanism operation checkup


Select “mechanism operation checkup” function, input check times, click “Execute”. Instrument will
automatically carry out mechanism operation checkup.

Alarm issued to hint to carry out the mechanism operation checkup

12.2.17 Barcode reader checkup


Click “bar code reader checkup” function, list all kinds barcode in system maintenance: reagent disk barcode
checkup, sample disk barcode checkup, select one of them and click “Execute” key. Click “end maintenance”
button to complete scan. Figure 12-4 explain the reagent disk barcode scan and sample disk barcode scan:

Figure 12-4

12.2.18 ISE checkup


Select “ISE checkup” function, input check times, and click “Execute”. The check value can be displayed in the

176
User Manual of CS-400 Auto-Chemistry Analyzer

result work area, click “end maintenance ”key to end the operation.

Execute ISE checkup when exchange electrode or after issue ISE alarm.

Note: The difference of two test value of same electrode should be less than 0.2.

figure 12-5 ISE check

12.2.19 ISE rinsing reagent pipeline


Select “ISE Rinsing reagent pipeline ” function in “Maintenance item list” menu, and click “Execute” key.
Click “end maintenance” key to end this operation.

We suggest user execute SIE rinsing reagent pipeline once a month.

177
User Manual of CS-400 Auto-Chemistry Analyzer

Figure12-6

12.2.20 ISE pipeline exhaust

Single-click “ ” in main keypad and select “ISE pipeline exhaust”, and then execute; the instrument
will carry out pipeline exhaust automatically after implement this function when bubbles exist in ISE pipeline
and ISE injection pump.

12.2.21 Automatically rinse the pipeline of concentrated liquid


The centrifuged serum may contain fibrin to make concentrated waste liquid pipeline blocked if the serum
sample are not concreted completely when testing, or blocked by bacteria that may be growing it. After test
1000 samples, upper machine software automatically prompts: "Please rinse concentrated waste liquid pipeline",
and execute rinsing as following steps:

(a) Replace the CS-phosphor-free anti-bacterial cleaning liquid at the positions 45th of R1 reagent and R2
reagent disks with CS-solenoid valve cleaning liquid.

(b) Single-click “ ” in main keypad and select “Automatically rinse concentrated waste liquid
pipeline”, and then single-click “Execute”; the instrument will carry out pipeline rinsing automatically as figure
12-5 shows:

178
User Manual of CS-400 Auto-Chemistry Analyzer

Figure 12-5

12.2.22 Manually rinse concentrated waste liquid pipeline


When concentrated waste liquid pipeline is dirty, execute manual rinsing of the pipeline as following steps:

(a) Clamp the concentrated waste liquid outlet at the right lower side of instrument back cover board with a
hemostatic clamp (about 50mm away from the back cover board).

(b) Unplug the nozzle 1 of rinsing mechanism first as figure 12-6 shows:

Nozzle 1

Figure 12-6

(c) Take a 70ml reagent bottle (containing CS- solenoid valve cleaning liquid and make sure the distance

179
User Manual of CS-400 Auto-Chemistry Analyzer

between liquid level and bottle lip is approximately 20mm), and insert nozzle 1 into the bottle, and then

single-click “ ” key, then select “Manually rinse concentrated waste liquid pipeline”, afterwards, the
instrument will execute rinsing automatically after click “Execute” as figure 12-7 shows:

Figure 12-7 Rinse concentrated waste liquid pipeline

(d) After about 5~6 minutes,remove the clamp on the outlet of concentrated liquid, and waste liquid will be
discharged automatically after rinsing.

(e) Replace the CS-reagent solenoid valve cleaning liquid in the 70ml bottle with purified water, and repeat the
operation carried out above.

12.3 Maintenance and checkup points and parts

12.3.1 Periodic cleaning ,checkup and parts replacement


Table 12-1 gives periodic cleaning and replacement parts (based on use of 5 hours daily).

(○: denotes periodic cleaning and check up and ●: denotes periodic replacement part.)

180
User Manual of CS-400 Auto-Chemistry Analyzer

Frequency
Quantity Yearly
No Part Refer to
Per use Quantity Every 3 Every 6
Daily Required Weekly Monthly Yearly
month month
1 Sample cup ● ——

2 Sample probe ○ 12.4.1

3 Reagent probe ○ 12.4.1


Rinsing bath of
reagent probe
Rinsing bath of
4
sample probe
○ 12.4.1
Rinsing bath of
stirring rod
Reaction cuvette sets
5(a)
(20pcs*6set)
6set 72set ○ ● 12.4.2

Drain filter of the


6
incubation bath
○ 12.4.2

7(b) Halogen lamp 1 2 ● 12.4.3


Rinsing mechanism
8
nozzle
○ 12.4.4

9 Stirring rod ○ 12.4.5

Sample probe syringe


10
Reagent probe syringe
● 12.4.11

11 Water supply filter ○ 12.4.6

CS-anti-bacterial
12 phosphor free ● 12.1.3
CS-alkaline detergent

13 Vacuum tank ○ 12.4.7

14 Cooling water bath ○ 12.4.8

Reagent cooling unit


15
Sample cooling unit
○ 12.4.9

16 Cooling fan ○ 12.4.10


specifica
17(c) Printer ribbon cassette ● tion
18(d) Cell blank check ○ 12.2.4

Purified water specifica


19(e)
equipment
○ ● tion

20 Waste discharge ○
21 Detergent bottle ○ 12.4.2

Table 12-1 Periodic Cleaning, Checkup and Parts Replacement List


Note:

(a) Statistics in above table is a maximum statistics, if cell blank value is still qualified after 1 month,

181
User Manual of CS-400 Auto-Chemistry Analyzer

continue to use it, if cell blank value is abnormal after Rinsing, replace it.

(b) Replace the lamp as soon as the photometer check value (340nm wavelength) exceeds 18,000 hours, or
the lamp use time more than 750 hours

(c) Instrument could use laser, ink mist, stylus printer, select accessory according to different printer.

(d) A cell blank alarm may occur if cell blank is not executed every week

(e) If the purified water has exceeded 1us/cm, consult the water supplier

12.3.2 Periodical replacement parts list


Make it a rule to stock as many spare parts as necessary for operation.

Quantity to
item Model name Description
Stock
1 Halogen lamp (light source lamp) 12V 20W 2
2 Reaction cuvette set(20pcs*6set) 72set
3 3603 ethylene tube 1/8 *1/4 inch 5m
4 3603 ethylene tube 1/16 *1/8 inch 5m
5 Teflon FEP rigid tube 1.5mm * 2.5mm 5m
6 Teflon FEP rigid tube 0.03 *1/16 inch 3m
7 Silica gel tube 8mm*14mm 10m
Proper
8 Ribbon cassette For printer
amount
Proper
9 Printing paper For printer
amount
For supply water
10 Water supply filter 1
connection
11 Sample probe For sample 1
12 Reagent probe For sampling 2
13 Stirring rod For stirring 2
14 Nozzle 1,2 of rinsing mechanism For cleaning 2
R1 reagent probe
15 R1 reagent probe syringe 1
sampling
R2 reagent probe
16 R2 reagent probe syringe 1
sampling
Sample probe
17 Sample probe syringe 1
sampling

Table 12-2

182
User Manual of CS-400 Auto-Chemistry Analyzer

12.4 Maintenance and check up method

! Warning:

◆ Do not spill water, reagent or detergent over the instrument or mechanical /electrical parts in order to
avoid the damage.

◆ Do not touch the suction mechanism, sampling mechanism, stirring mechanism, reaction cuvette rinsing
mechanism during operation, or there will be a risk of infection or injury.

◆ Protective measures should be taken to the operator, such as with protective gloves, protective glasses
and work uniform during operation. Otherwise, there maybe an infection when touch the contaminated
areas and contaminated liquid. Corrosive liquids may cause a skin injury. If the contaminated liquid or
corrosive liquids accidentally touched the body, please rinse with water immediately, and seek medical
advise.

12.4.1 Sample probe and reagent probe


If the probe inside or outside is contaminated, the serum, reagent, water, etc. might easily adhere, thereby
degrading the sampling accuracy and precision or clogging the interior. Wash or clean the probe from time to
time.

(1) Daily washing (automatic washing)

(a) Remove the sample disc cover, place 1ml CS-alkaline detergent (as figure 12-7 shows) on W1 position of
sample disc. Placed 1ml CS-acidity detergent on W3 position to avoid cross-contamination. Detergent
quantity should be taken according to the frequency of rinsing.

Figure 12-7

(b) Set 70ml CS-anti bacterial phosphor-free detergent on 45 position in the reagent disc(R1,R2).

When sample and reagent probe finish sampling respectively, they may automatically assimilate CS-anti
bacterial phosphor-free detergent or CS alkaline detergent to process the rinsing.

Note: Place the detergent on a designated position When doing cuvette, sample probe avoiding
cross-contamination rinsing.

183
User Manual of CS-400 Auto-Chemistry Analyzer

Figure 12-8
(2) Cleaning outside of probe tip

(a) Turn off the POWER switch of analyzer.

(b) Remove the sample or reagent disk cover, and move the probe arm to the top of disk by hand.

Figure 12-9

(c) Using gauze moistened with alcohol solution, clean the outside of probe.

Figure 12-10
184
User Manual of CS-400 Auto-Chemistry Analyzer

(d) Turn on the POWER switch of analyzing unit. Each probe will then return to its reset (home) position
automatically.

Note: As alcohol is flammable, Pay attention to it and do not place large amount alcohol in the vicinity of the
instrument.

(3) Cleaning clogged probe

(a) Turn off the POWER switch of analyzing unit

(b) Pinch the jaw of probe arm and remove the cover, and loosen the connector as figure 12-11 shows:

jaw

Figure 12-11

Loosen the probe retaining nut as figure 12-12 shows.

Tube

Figure 12-12

(c) Remove the probe.

(d) Penetrate through the needle from the top with stainless steel wire to clean up, as Figure 12-13 shows:

185
User Manual of CS-400 Auto-Chemistry Analyzer

Figure 12-13

(e) Let injector assimilate 10 ml pure water after cleaning the probe with stainless steel, connect both
interfaces of probe and injector, push injector plunger until the 10 ml water has been discharged completely as
figure 12-14 shows.

Figure 12-14

Note: When cleaning the clogged sample probe, use a stainless steel wire having outside diameter of 0.3 mm.

When cleaning the clogged reagent probe, use a stainless steel wire having outside diameter of 0.5 mm.

(4) Adjusting probe position

(a)Turn on the POWER switch of analyzer

(b)Open the system maintenance menu

(c) Click “sample probe horizontal checkup”, “sample probe vertical checkup ”, “reagent sample horizontal
checkup” and “reagent sample vertical checkup” in the “maintenance item”.

Note: During adjusting sample probe, make sure to implement the horizontal checkup firstly, then implement
the vertical checkup. The tip of sample probe should be at the center of the reaction cuvette.

(d)During sample probe horizontal checkup, the probe stops above the reaction cuvette. At this step, adjust the
reagent probe so that its tip will be aligned with the center of reaction cuvette.

186
User Manual of CS-400 Auto-Chemistry Analyzer

Reaction
cuvette

Probe tip

Figure 12-15

Please contact maintenance man if the probe tip is not aligned with the center of reaction cuvette.

(e) During reagent probe horizontal checkup, the probe stops above the reaction cuvette. At this step adjust
the reagent probe so that its tip will be aligned with the center of the reaction cuvette.

(f) To execute “stirring mechanism horizontal checkup” , when the stirring rod stops above the reaction
cuvette, at this step, check muddler probe tip is aligned with the center of reaction cuvette.

From overlooking angle to view the sample probe and reagent probe and stirring rod, the relevant position of
the reaction cuvette is as following:

Reagent probe 1

stirring rod

Reaction
cuvette

Reagent probe 2

Figure 12-16

Click “end maintenance” key to end operation

Please contact maintenance man if the probe tip and stirring rod probe tip are not aligned with the center of
reaction cuvette.

Process of sample probe up-down mechanism horizontal check:

The movement of the sample probe:

Sample probe → above the reaction cuvette → Rinsing bath (pause)→ outer track 1 position on sample
disk(descend) → above the reaction cuvette → Rinsing bath (pause) → Middle track W1 position on sample
disk (descend) → Above the reaction cuvette → Rinsing bath (pause) → Inner track C8 position on sample
disk (descend) → Above the reaction cuvette → Rinsing bath (pause) → ISE diluent bath →Above the

187
User Manual of CS-400 Auto-Chemistry Analyzer

reaction cuvette → repeated the whole process.

Note: “ISE diluent bath →Above the reaction cuvette” action only be taken where ISE function is provided.
Home position of the sample probe is reset point.

Process of reagent probe up-down mechanism horizontal check:

The movement of the Reagent probe:

Reagent probe (R1,R2) → Above the reaction cuvette → Rinsing bath → Reagent bottle → Rinsing bath
(pause) →Above the reaction cuvette → repeated the whole process.

*Sample probe R1 and R2 have the same movement according to the sample probe movement.

The movement of the Stirring rod

Stirring rod → Above the reaction cuvette → Rinsing bath →Above the reaction cuvette → repeated the
whole process.

R1,R2 have the same movement according to the Stirring rod movement

(g) Sample probe vertical check.

Select “sample probe vertical check”, in “maintenance” key, and then click “Execute ”key, thus, the sample
probe descends until the bottom of sample cup is detected.

(h) During reagent probe vertical checkup, place an empty and dry reagent bottle at position 1 on reagent disk
1 and reagent disk 2, single-click “next”, the reagent probe will descend from top to the bottom to memorize
the descent distance as the reference value of the remaining reagent.

(5) Cleaning rinsing bath

(a) If the Rinsing bath is contaminated, use a tube brush cleaning it with 2% CS-anti-bacterial phosphor-free
detergent as figure 12-17 shows:

Rinsing bath

Figure 12-17

(b) Infuse 10ml 2% CS-anti-bacterial phosphor-free detergent into Rinsing bath.

188
User Manual of CS-400 Auto-Chemistry Analyzer

Figure 12-18

(c) Then, infuse 100ml water into Rinsing bath to wash.

After cleaning, contamination can be eliminated and bacterial can be restrained. Cleaning can be taken every
month. If the instrument is contaminated while the operation, please cleaning it in time.

12.4.2 Reaction cuvette


A contaminated reaction cuvette or incubation bath would cause faulty data. Beside, reaction cuvette get
aging after a long period usage. Periodically rinsing reaction cuvette, check the cell blank value of reaction
cuvette, if cell blank value is abnormal, replace the reaction cuvette .

(1) The confirmation of the contaminated Reaction cuvette

(a) Turn on the POWER switch of analyzer.

(b) Select “cell blank” function in the “Maintenance” menu, click “Execute ”, the instrument will carry out
the cell blank check automatically.

(c) The cell blank value of the first reaction cuvette should ≤18000, and the difference between 2-120
cuvette should within the range of ±800.

The cell blank value is the absorbency of each reaction cuvette before adjust the photometer, data value
represent the absorbency of each reaction cuvette from the second reaction cuvette, data value represent the
margin between each reaction coveter ( from the second reaction cuvette) and the first reaction cuvette.

(d) If the cell blank value is not within a rage of ±800, the relevant reaction cuvette is contaminated. Clean the
reaction cuvette.

Note: The cell blank value can be displayed or printed(previous value will be replaced when the second test
finished)

(2) Reaction cuvette cleaning

If the cell blank value is not within a rage of ±800, the relevant reaction cuvette is contaminated. Clean the
reaction cuvette. If the usage of the reaction cuvette has exceed the time limit for replacement, please place a
new one.
189
User Manual of CS-400 Auto-Chemistry Analyzer

(a) Place the CS-anti-bacterial phosphor free detergent (70ml/bottle)on position 45 in the R1,R2 reagent disk

(b) Click “Rinsing reaction cuvette ”button in “maintenance item list” work area, then click “Execute ”key

(c) After cleaning reaction cuvette, carry out the cell blank check again. If the cell blank value exceeds the
±800, please place a new one.

Note: In order to avoid uncleanness cleaning after long time use, immerse the reaction cuvette in 2% CS
anti-bacterial phosphor-free detergent for more than 8 hours every week. Wash the immersed reaction cuvette
with water, and then wash the reaction cuvette with purified water, and then mount the cuvette on the reaction
disk, carry out cell blank check, test after cell blank check value is qualified.

(3) Replace reaction cuvette.

If the blank cell value is not qualified after cleaning or if the cuvette is used for over two month, replace it
with a new one.

Note: A new reaction cuvette should be immersed in 2% CS-anti-bacterial phosphor detergent for 8 hours, clean
the surface of the cuvette by purified water, and then Mount the cuvette on the reaction disk, carry out testing.
The abnormal cell blank value will influence the accuracy and repetition of the testing result.

(a) Turn off the power switch.

(b) Remove the setscrew of rinsing mechanism when wear protective gloves.

Figure 12-19

(c) Remove the setscrew of reaction cuvette as figure 12-20 shows.

190
User Manual of CS-400 Auto-Chemistry Analyzer

Figure 12-20

(d)Take out six set reaction cuvette as figure 12-21 shows:

Figure 12-21

(e) Mount new reaction cuvette on the reaction disk. 6 sets reaction cuvette should be placed at the same time
in counter direction

(f) Turn on the power switch.

(g) Select “cell blank test” in “maintenance” menu. Make sure carry out the cell blank check after replace
reaction cup. Testing can be carried out after the cell blank value is qualified.

Note 1: A once-employed reaction cuvette might be contaminated heavily if allowed to dry. Immerse it in
purified water to store it. If the instrument will be left unemployed for 3 days or more, remove the reaction
cuvettes from the reaction disk and keep them immersed in purified water.

Note 2: Never use any organic solvent (benzene, alcohol) for washing the cuvettes.

(4) Cleaning Incubation Bath and the drain filter of the Incubation Bath.

A clog on drain filter or Incubation contamination will cause the inaccuracy of the testing data. Thus, Clean the
incubation bath periodically. (once a month)

(a) Select “Rinsing incubation bath ” function in the “maintenance item list ”work area, after closing the light
191
User Manual of CS-400 Auto-Chemistry Analyzer

source lamp,the constant temperature water will be drained from the Incubation bath. Turn off the power
switch.

(b) Loosen the rinsing mechanism retaining screw, and remove the rinsing nozzle head, as figure 12-22
shows:

Figure 12-22

(c) Take out the 6 set reaction cuvette into purified water, Loose the retaining screw of the reaction disk, take
out the reaction disk. Place the reaction cuvette in a location free from dust.

Note: If take out the reaction disk and cuvette at the same time, water droplet attached on the outside of reaction
cuvette will drop into the instrument so that cause the instrument malfunction. Therefore take over the reaction
cuvette firstly and then take over the reaction disk secondly.

(d) Using washed gauze moisturized with water, clean the reaction bath and photometric window as figure
12-23 shows. Be careful not to flaw them or attached scrape.

Photometry
window

Figure 12-23

(e) Take out the drain filter of the incubation bath as figure 12-24 shows, cleaning by water, and then return it

192
User Manual of CS-400 Auto-Chemistry Analyzer

in place.

Filter

Figure 12-24

(f) Click “next” key in the “maintenance” form, infuse pure water into the incubation bath, turn light and
circulation syringe power on.

(g) Mount the reaction disk and reaction cuvette after Incubation bath rinsing.

(h) Return the rinsing mechanism nozzle head in place and secure it.

(i) Select “cell blank check ” function in “maintenance item list” work area. Test can be carried out after the
cell blank check value is qualified.

(5) Liquid level sensor of the incubation bath

Take out the sensor out of the incubation bath, wipe the outside of sensor with gauze moisturized with 2%
detergent. In order to prevent water from being contaminated by sensor probe, rinsing once a month is advised.

(6) Cleaning detergent bottle

As the CS-alkaline detergent in the detergent bottle is added timely and by this way maintain the day-to-day use.
After a period of time, there will be dust or white substance separate out Thus, clean it up monthly.

(a) Take over the detergent bottle from the instrument, loose the detergent bottle cap which is located on the
upper sprue. Insert the accessory tool from upper sprue to bottom outfall, then tighten the screw to prevent
leakage when detach the interface of the bottle bottom. Clean the bottle when there is only a few detergent
existed.

193
User Manual of CS-400 Auto-Chemistry Analyzer

Bottle lid

Figure 12-25

(b) Detach the bottom interface after rag prepared, Take care that the pipe port position must be lower than
bottle.

Pipe joint

Figure 12-26

(c) Take out the detergent bottle, clean the outside and inside of the bottle, then wash it by water, At last wash
it by purified water. Detach the tool and wash away the attached detergent.

(d) Dry the inside water droplet, wipe the outside water droplet, return the detergent bottle in place, add
CS-alkaline detergent.

(e) Select “maintenance item list” work area, implement " pipeline for pouring detergent exhaust " function.

12.4.3 Light source lamp


If the light source lamp deteriorates, the quantity of light would be out of the specified photometric range or
noise would increase so that an accurate measurement would be impossible. If the check value exceeds, replace
the lamp.

(1) Light quantity check

① Select “Light quantity check” function in “maintenance” form. The photometer light quantity is checked
and the results will be displayed by AD value or printed out.

194
User Manual of CS-400 Auto-Chemistry Analyzer

② Usually the value is a maximum at 340nm.

(2) Replace the light source lamp

Method 1:

(a) Prepare a new lamp.

Figure 12-27
Note: Do not touch the glass surface of the new lamp. Otherwise the lamp characteristics may change. If the
glass surface is found stained with fingerprints, etc, wipe them off with gauze wetted with alcohol.

(b) Select “Rinsing incubation bath ” function, click “Execute” key, the constant water in incubation bath is
automatically discharged.

(c) Wait a few minute, cooling light room ( 30 minute ).

(d) Loosen the retaining button of the reaction disk, take out the reaction disk. Place the reaction disk in a
position free from dust.

Note :reaction disk should be placed in dry and clean area, prevent the droplet on reaction cuvette dropping
on the inside of the instrument.

(e) Loosen two retaining terminals of lamp lead wire and disconnect the lamp lead wires.

Retaining
terminal

Figure 12-28

195
User Manual of CS-400 Auto-Chemistry Analyzer

(f) Loosen the retaining screw of the retaining light seat,take out the lamp as figure 12-29 show.

Retaining
screw

Figure 12-29

(g)In the reverse order of step 6-8, mount the new light source lamp. Do not distort the rubber tube used for
cooling light room. Make sure that the lead wire of the lamp not in loose status.

(h) Mount the reaction disk and cuvette and the Rinsing mechanism, turn the power on, execute “next” in
“maintenance”, infuse pure water into reaction cuvette. Execute “light volume check” function. While light
volume value is qualified, testing can be proceeded.

Method 2:

(a)~(c) are the same as above mentioned.

(d)Take over a set reaction cuvette instead of Rinsing mechanism. Keep the reaction cuvette clean. Turning
the reaction disk by holding the retaining button of the reaction disk, make the reaction disk, which is
detached from reaction cuvette turning underneath the Rinsing mechanism. Loosen the retaining button and
take over the reaction disk.

Following steps are the same as Method 1.

12.4.4 Cleaning the rinsing nozzle


If the rinsing nozzle clogs, the reaction cuvette may not be cleaned adequately so that cause a data error or other
instrument malfunction. Beside, rinsing water may overflow on the reaction disk to make it impossible to
provide correct data.

(a) Loosen the rinsing mechanism retaining screw by turning it counterclockwise, and remove the rinsing
nozzle head.

(b) Wipe the outside of the nozzle with gauze wetted in 2% anti-bacterial phosphor-free detergent.

196
User Manual of CS-400 Auto-Chemistry Analyzer

Nozzle
Wipe

Figure 12-30

(c) Insert a 0.5mm stainless steel wire into the nozzle from its bottom end for cleaning as figure 12-31 shows.

Figure 12-31

Note: If the nozzle chip is contaminated heavily or worn excessively, replace it with a new one.

(d) Take out wiping block lightly, rinse it with pure water after rinsing it with 2% CS-anti-bacterial
phosphor-free detergent.

(e) Return the wiping block (keep lower horizontal level of wiping block the same level with the rinsing
probe) as figure 12-32:

197
User Manual of CS-400 Auto-Chemistry Analyzer

Rinsing nozzle

Press
into

Nozzle

Reaction
cuvette
Figure 12-32

(f) Mount the rinsing mechanism in home position.

(g) Carry out the Mechanism Check 10 times, wiping block can’t touch the reaction cuvette and the rinsing
water is filled up to the upper limit level in the reaction cuvette and it does not overflow from the incubation
bath.

12.4.5 Stirring rod


(1) Cleaning the Stirring Rod

A contaminated stirring rod would cause a cross contamination so that influence the accuracy and precision of
the testing result. Clean the Stirring Rod periodically.

(a) Turn off the power switch.

Wipe the Stirring rod using gauze moistened with 2% anti-bacterial phosphor-free detergent, then wash away
the detergent on the surface of the Stirring rod by gauze moistened with purified water.

Caution: Do not bend the Stirring Rod.

Figure 12-33

198
User Manual of CS-400 Auto-Chemistry Analyzer

(2)Replacing the stirring rod

(a) Turn off the power switch.

(b) Loosen the two setscrews one round as figure 12-34 shows.

Figure 12-34

(c) wipe front of the new stirring rod with gauze moisturized with 2% CS-anti-bacterial phosphor-free
detergent.

(d) Insert the new stirring rod until its end touches the bottom of axis motor. Then, secure it by M2 as figure
12-35 shows.

Stirring rod
Motor

M2 Screw Stirring rod

Figure 12-35

(e) Place stirring rod adjust block on the rack of reaction cuvette, then move the stirring rod above the adjust
block as figure 12-36 shows.

199
User Manual of CS-400 Auto-Chemistry Analyzer

Stirring mechanism

Reaction cuvette
71 position
Height adjust screw

Place the adjust block here

Adjust block

Figure 12-36

(f) loosen M2 screw, adjust the stirring rod position, its tip and upper side of adjust block are supposed to be
attached as figure 12-37 shows: tighten M2 screw.

Stirring rod

Adjust block top


Reaction cuvette tray

Figure 12-37

(g) Select “stirring mechanism horizontal check” in “maintenance” form, single-click “Execute”, “next” to
confirm stirring rod position weather correct. Please contact with after service.

(h) Execute 10 times “mechanism movement check” to confirm no abnormality exists.

12.4.6 Supply water filter


The main function of the supply water filter is to prevent the rubber grain and ion exchange resin get into the
inner line of instrument. If these filters are clogged, water supplied to the incubation bath and water used for
probe/ stirring rod rinsing are decreased significantly so that cause a data error. To prevent this, clean them
every month.

(a) Turn off the purified water supply unit to stop the water supply

(b) Turn off the power switch of analyzing unit.

(c) Make the gauze as a cushion of the filter cap, Loosen the filter cap. Prepare a water container to receive
the water flow.

200
User Manual of CS-400 Auto-Chemistry Analyzer

Figure 12-38
(d) Pull out the filter, rinse the filter with water. Then reassemble it in the reverse order of the above.

12.4.7 The vacuum tank


Vacuum tank must be emptied if the waste liquid spills into a vacuum tank. Otherwise, abnormality of other
parts may occur (Contact the company after-sales service staffs when instrument abnormality occurs).

(a) Switch off the instrument main power.

(b) Loosen the two fixing screws of rubber hose on the back of the analyzer, and remove the hose as figure
12-39 shows.

(c) Unplug the cork and collect the outflow waste liquid with barrel or kind of containers.

Two fixing screws of rubber hose

Figure 12-39
(d) Plug the cork of rubber hose after empty the vacuum tank, and fix the hose on the cover with the two
screws.

201
User Manual of CS-400 Auto-Chemistry Analyzer

12.4.8 Cooling water tank


Cooling water tank is located in the front left of instrument, and constant temperature water tank is located in
the lower left of the instrument.

(1) Adding purified water in water tank

If water in the cooling water tank is used many times, it will be dirty and will be in bad water circulation, and
the volume will be reduced due to evaporation To prevent this, exchanging water in the cooling water bath once
a year is a must.

(a)Turn off the power switch of analyzer

(b)open the left front door of the instrument, remove the cover of front left lower side of the instrument.

Figure 12-41

(c) Remove the rubber plug from the cooling water bath, and drain water using the hose pump (prepare a
water reservoir or bucket).

High level tube

Low level tube

Figure 12-42

202
User Manual of CS-400 Auto-Chemistry Analyzer

(d) Fill purified water into the cooling water bath until purified water spills out of the high level of the hose.

(e) Switch on the main power for several minutes. Replenish purified water again into the low water level of
hose until the purified water spills again from the high level of the hose (approximate 3L water).

(f)recover the rubber plug.

(g)Return the cover of the analyzer in place.

(2) Discharge the purified water in constant temperature water tank.

In case of transportation, discharge the purified water in the cooling water tank first.

(a) Turn off the main power supply of instrument.

(b) Demount the front left cover of instrument.

(c)Pull of the rubber plug as figure 12-43 shows.

Constant
temperature
Water tank

Robber plug
for hose

Figure 12-43

(d) Fill in the rubber plug after discharge the water, and mount the cover of instrument.

12.4.9 Reagent cooling unit and Sample disk tray


The reagent cooling unit and the sample disk tray will be contaminated with sample or dust. Clean them at least
once a month.

(a) Remove the reagent disk and clean the inside of the reagent cooling unit with gauze

203
User Manual of CS-400 Auto-Chemistry Analyzer

Reagent cooling unit

Figure 12-44

(b) Then, with gauze, clean the read-out window of the reagent barcode reader.

Barcode reader
window

Figure 12-45

(c) Remove the sample disk, and clean the inside of the sample disk tray with gauze.

Figure 12-46
204
User Manual of CS-400 Auto-Chemistry Analyzer

12.4.10 Cooling fan and dustproof cover


Long-term use of the instrument can make the cooling fan and dustproof cover surface accumulated dust, so
every 6 months cleaning should be carried out once.

(1) Cooling fan cleaning

(a) Switch off instrument main power.

(b) Clean the dust on fan with vacuum cleaner.

(2) Dustproof cover cleaning

(a) Hold the handles of cooling fan at the both sides and remove the cover directly as figure 12-47 shows:

Handles of dustproof cover

Figure 12-47

(b) Clean the dust on the cover with vacuum cleaner.


(c) Rinse the cover with clean water.
(d) Remount the cover after wipe it with cloth.
12.4.11 Sample syringe
The Syringe includes R1 reagent probe syringe, R2 reagent probe syringe, sample probe syringe. If connected
with ISE, it may include SIP Syringe, DIL syringe, IS syringe too. The span of the syringe can be 1 million times,
and replacement is must in 15 months if use correctly. Contact the after service department in order to change a
new one.

12.4.12 Cooling machine


The cooling machine is at the rear of instrument, as figure 12-49 shows:

205
User Manual of CS-400 Auto-Chemistry Analyzer

LED Display

Figure 12-49

The LED display of cooling machine will circulated display the circuit value of semi-conductor cooling chip
1-4, water tank temperature of cooling system, and instrument interior temperature.

When the temperature of water tank of cooling system < 5~15℃, or the C1-C4 current value is less than 5-7 A,
alarm will be issued, please contact after service department.

12.5 ISE device maintenance

12.5.1 Periodical cleaning, checkup and parts replacement


Periodic Cleaning, Checkup and Parts Replacement list

Table 12-3 gives periodic cleaning and replacement parts (based on use of 5 hours daily).

(○:denotes periodic cleaning and check up and ●: denotes periodic replacement part.)

Frequency
Every Every Every Refer
No Item
Daily Required Weekly Monthly 2 3 6 to
Month Month Month
1 Sample syringe(SIP、IS、DIL) ● 12.4.11
2 Vacuum tank ○ 12.4.7
3 Sample pipeline rinsing ○ 12.5.2
4 Waste pipeline rinsing ○ 12.5.7
5 Reagent pipeline rinsing ○ 12.5.3
6 Na electrode ● 12.5.4
7 K electrode ● 12.5.4
8 Cl electrode ● 12.5.4
9 Indicator electrode ● 12.5.4
10 SIP tube(suction tube) ● 12.5.5
11 Cleaning of the waste ○ 12.5.6

Table12-3

206
User Manual of CS-400 Auto-Chemistry Analyzer

12.5.2 Daily rinsing


(automatically rinsing)

The pipe line could be contaminated by bacteria, fat and protein when testing electrolyte. Thus daily rinsing is a
necessity.

(a) Input 1ml CS-ISE detergent at W2 position in sample disk.

(b) Set “rinsing after test” in “Maintenance” menu as effective before the last test. If rinsing after test,
calibration should be taken when test ISE next time.

12.5.3 ISE reagent pipeline rinsing


(ISE infuse)

The ISE reagent pipeline could be contaminated after long time use. Rinsing it monthly according to the
following procedure.

(1) Carry out the infusion.

(a) Dilute the CS-ISE detergent 20 times with purified water.

(b) Infuse the internal standard liquid, diluent bottle with above diluted CS-ISE, carry out ISE (IS+DIL)
infuse in “maintenance” menu.

(c) After infusion finished, wash out the CS-ISE detergent which attached on the tube with water, wipe out the
water droplet with gauze, and put it back..

(d) Carry out twice ISE(IS+DIL) infusion in “system maintenance” menu.

(2)ISE check up

Carry out 30 times ISE checkup in “maintenance” menu, and the result is showed.

Note : print value of the same electrode should within ±0.2

12.5.4 Electrode replacement


Note:

Do not open ISE cover under normal situation, or temperature control error may cause test result inaccuracy.

Replace the electrode and check the instrument within 1 hour.

12.5.4.1 Na, K, Cl electrode replacement


After long time usage of ion selective electrode, the electric potential becomes weak, Thus replace a new
electrode is in need.

(1) Electrode replacement time

When slope value is abnormal, alarm will be issued, as table 12-4 shows:

207
User Manual of CS-400 Auto-Chemistry Analyzer

Slope value
Alarm
Na K Cl

ISE prepare
above 68.1mV abov68.1mV below -68.1mV
abnormal

68.1mV~37mV 68.1mV~37mV -68.1mV~-32mV Normal

ISE prepare
37mV~32mV 37mV~32mV -32mV~-25mV
abnormal
ISE slope value
below 32mV below 32mV above -25mV
abnormal

Table 12-4

When ISE prepare alarm issued, analysis of current day could be proceed as normal, replace a new electrode
next day. When ISE slope value error occur, replace a new electrode immediately.

If alarm issued even the slope value within normal rage, that indicates the electrode response error. This is
caused by pipeline contamination. Rinse the pipeline to solve this problem.

If the calibration value is normal in previous day, but slope value change rapidly now, do not exclude other
reasons except electrode error. Check if there is a leakage or block in the pipeline.

(2) Replacement method of electrode

(a) Open ISE cover at the left side of analyzer, turning-in the left side handgrip as figure 12-50 shows:

Hand shank

Figure 12-50

Remove the electrode by hand or by forceps after loosen the electrode as figure 12-51 shows:

208
User Manual of CS-400 Auto-Chemistry Analyzer

Figure 12-51

(b) Pull out of the electrode lead, connect the colored lead after replace the new electrode. As figure 12-52
shows:

Reference Na+ K+ Cl-

Reference Na+ K+ Cl-

Figure 12-52

(c)Range in Cl (green),K(red), Na(yellow), sequence, turning-in the handgrip

Note 1: In order to prevent remaining of conductive component while replacing electrode, wipe out the liquid
in the vicinity of the electrode.

Note 2: In order to ensure gas tightness of electrode tube, a tube with O-ring is a must. Check if there is an
O-ring after replace new electrode.

(3) Electrode debugging

209
User Manual of CS-400 Auto-Chemistry Analyzer

Debugging the electrode according to following sequence after replace a new electrode.

(a) Carry out ISE (IS,REF) infusion twice in “maintenance” menu.

(b) Carry out 10 times ISE checkup in “maintenance” menu ten minutes later. The result of the ISE checkup
will be showed in maintenance form.

(c) Carry out ISE (full point)calibration, check if the slope value in within reference range.

12.5.4.2 Reference electrode replacement


(1) Replacement time

When the slope value of Na, K, Cl are all unstable or lower, please replace a new one.

(2) Replacement method

(a) Open left ISE cover of analyzing unit pull-in the handgrip, remove any electrode of Na, K, Cl after loosen
the electrode.

(b) Pull out of the electrode lead.

(c) Remove the fixing block of reference electrode, as figure 12-53 shows:

Fixed Reference
block electrode

Figure 12-53

Note1: Push down the wrench on the side of Na, K, Cl electrode side, fix the fixing block, the electrode could
be pull out easily by doing so.

Note2: In order to easily replacement, nip the reference electrode by forceps.

Note3: Owing to the conductivity of the dropping liquid, wipe out the liquid by gauze wetted with purified
water.

(d) Replace new electrode, connect the electrode lead.

210
User Manual of CS-400 Auto-Chemistry Analyzer

(3) Reference electrode debugging

(a) Carry out ISE( REF) infusion twice in “maintenance” menu.

(b) Carry out 10 times ISE checkup in system maintenance ten minutes later. The result of the ISE checkup
will be showed in maintenance form. The value difference of the same electrode should within ±2Mv.

(c) Carry out ISE(full point)calibration, check if the slope value is within reference range.

12.5.5 SIP tube replacement


(a) Open the ISE cover of instrument analyze unit.

(b) Pull out of the tube and replace a new one. Make sure the tube is not slack while connecting. As figure
12-54 shows:

(c) Wipe out the dropping liquid with gauze wetted with purified water.

(d) Close the ISE cover.

SIP tube

Figure 12-54

12.5.6 ISE waste cleaning


The crystal substance attached on the ISE waste interface may cause inaccuracy of the test result. Cleaning it
once a week.

(a) Wash the crystal attached on the ISE waste interface into waste container by purified water, and then wash
the waste container by purified water. As figure 12-55 shows:

211
User Manual of CS-400 Auto-Chemistry Analyzer

REF Tube Screw

Waste
interface

Waste
container

Detergent bottle

Figure 12-55

(b) Wipe out the liquid by gauze wetted with purified water, Thus there will be no conductive component in
the waste interface and waste container.

Note: Touching the waste tube may cause noise jamming while testing.

12.5.7 Waste pipeline cleaning


The sample probe, waste hose, electrode, ISE channel should be cleaned once a week.

(a) Place 1 ml CS-alkaline detergent at W1 position on sample disk, Place 1 ml CS-ISE detergent at W2
position on the sample disk.

(b) Place the CS-anti-bacterial phosphor-free detergent at 45 position on reagent disk 1 and reagent disk 2.

(c) Carry out “Rinsing reaction cuvette + ISE ” in maintenance menu.

Note: Owing to the pipeline of electrolyte is cleaned at the same time, the electrode state may change, thus
carry out calibration before ISE test.

212
User Manual of CS-400 Auto-Chemistry Analyzer

Chapter 13 Alarm Data Processing

13.1 Alarm information type

Alarm level includes: Warning level, sample stop level, stop level, as table 13-1 shows:

Alarm level Instrument action

Data alarm Alarm for testing result. Operation proceeds as normal.

Note appear both in data alarm and instrument malfunction. Buzzer calling, but the operation
Warning
proceed as usual, stop or proceed is judged by operator.
Alarm for instrument malfunction. Buzzer calling. Stop adding new sample. Sample on testing
Sampling stop
will proceed analyze.

Stop Alarm for malfunction. Buzzer calling. Instrument stop immediately. Test result is invalid.

Table 13-1

13.2 Countermeasure to malfunction do not issue alarm

13.2.1 Data malfunction which do not issue alarm


Some instrument malfunction is neither displayed in the data note nor issued the buzzer. These instrument
malfunction could be tested through recheck, observe instrument state and test control sample.

Dada malfunction and solution as figure 13-2 shows.

Malfunction Description Countermeasure


1.Maintenance is not carried out periodically. 1.Maintenance according to the user manual.
2.Reagent deteriorate, insoluble matter exist 2.Replace a new reagent, use or preserve it correctly.
3.Purified water unqualified. 3.Water conductivity should within 1uS/cm.
4.CS-anti-bacterial phosphor-free detergent, 4.Add detergent, carry out rinsing reaction cuvette in
CS-alkaline detergent shortage. “maintenance”.
Repeatability
5.Cristal occurs in reagent cooling unit. (low value 5.Use detergent produced by Dirui Company.
error
repetitive error.) 6.Separate place the reagent which may cause cross
6.Cross contamination exist between analysis item. contamination, or use cross contamination obviation
7.Sample disqualification, fibrin exists in sample. function.
7.Separate the disqualified sample, or collect sample
again.
1.Standard liquid concentrate failure. 1.Immediately use the standard liquid as long as it’s
2.Reagent concoct fault. put into the sample cup.
Accuracy error
3.Analyze condition set error. 2.Replace a new reagent.
3.Correctly setup the parameter.

Table 13-2

213
User Manual of CS-400 Auto-Chemistry Analyzer

13.2.2 Instrument malfunction which do not issue alarm

Malfunction Description Countermeasure

1.Sample probe contaminated.


Sample probe with water 1.Clean it through maintenance check.
2.Sampling system (pipeline, syringe)
droplet 2.Carry out maintenance check.
leakage.
1.CS-anti-bacterial phosphor-free
1.Fill it.
detergent for cleaning reaction cuvette
Water drop from washing 2.Check interface
has run down
mechanism 3.Clean it through maintenance check. For
2.Rinsing mechanism pipeline leakage
replacement purpose, contact service department.
3.Nozzle and pipeline block.
1.Outlet the air in pump.
1.Water supply pump with air.
Instant temperature 2.Clean it through maintenance check.
2.Incubation bath filter block.
Water do not outflow 3.Connect with power, do not use the same circuit
3.Purified unit do not electrify.
with instrument.

Water cannot outflow Clean it through maintenance check . For replacement


Nozzle, pipe line block.
from Rinsing nozzle purpose, contact the service department.

Water cannot outflow


from Rinsing bath (sample Rinsing outlet port. For replacement purpose contact
Outlet port, pipeline block.
probe, reagent nozzle, the service department.
stirring rod)
Constant temperature
water(from rinsing Clean it through maintenance check . For replacement
Nozzle, pipeline block.
mechanism to reaction purpose, contact the service department.
cuvette) overflow.
1.Incubation water level lower or
1.Carry out water exchange in incubation bath
Incubation bath with incubation bath is contaminated.
2.Carry out water exchange after purified water unit
bubble. 2.Incubation water exchange carry out
electrified
before Purified water unit power on.

1.Radiator filter with dust block. 1.Clean it through maintenance check.


Reagent cooling error
2.Cooling unit malfunction. 2.Contact service department.

Sample syringe with


Interface installation error Re-install after leakage confirmed.
leakage

1.Re-install after air enter in confirmed


Sample syringe with 1.Interface installation error 2.Carry out again. If small air bubble exist, tap the
bubble 2.Sample syringe degas insufficient syringe while reagent or detergent flow. Eliminate it
with shake.

Table 13-3

13.3 Instrument alarm list

214
User Manual of CS-400 Auto-Chemistry Analyzer

Code Level Name Describe Countermeasure

The absorbance to be used for 1.Check if the reagent is confected or positioned in the right way. 2.Check if the incubation
0-1 Warning Absorbance over calculation after cell blank bath is contaminated. 3.Check if there is the impurity interfuse. 4.Check if the reaction
correction exceeds 3.3ABS cuvette is blurred. 5.Check if there is an obstacle on the optical path of photometer.

The mean value of two tests with


Reagent blank
standard solution 1 is larger than
0-2 Warning horizontal Check if reagent is still valid.
the setup value of cell blank
checkup over.
horizontal check.

The difference between two times


1.Check if the value of deviation permissible absorbance is correct. 2.Check if sample syringe
Divergence test standard solution absorbance
0-3 Warning is leakage and if the incubation bath is contaminated. 3.Check if there is a cross
checkup over is larger than deviation
contamination.
permissible absorbance.

The difference between calculate


K factor check 1.Check if K factor is proper. 2.Check if the standard solution is deteriorate. 3.Check if
0-4 Warning K factor and previous one is more
value over reagent is deteriorate.
than plus or minus 20%

The measured value is out of the


Technical limit
0-5 Warning technical limit range of test 1.Dilution the sample and rerun the test. 2.Check setup value.
over
item.

Calculation in the specified


Linearity photometric range, and there is a 1.Check if there is impurity in the sample. 2.Check if there is impurity and bubble in the
0-6 Warning
abnormal different exceeding the linearity reagent. 3.Dilution the sample and rerun the test. 4.Check Stirring mechanism.
limit value.

Sensitivity is checked for


linear(2-6 points), non-linear or
Isozyme-P calibration. This error
1.Call up main screen, check if sensitivity limit value is properly setup. 2.Check if reagent
Sensitivity is indicated if the difference in
0-7 Warning is deteriorate. 3.Check the input concentration value of standard solution is properly setup.
abnormal mean absorbance between standard
4.Check if standard solution deteriorate.
solution 1 and standard solution
(N) is smaller than the
sensitivity limit (input value).

215
User Manual of CS-400 Auto-Chemistry Analyzer

Code Level Name Describe Countermeasure

In the Antigen addiction assay and


Prozone check reaction assay, the prozone check 1.Dilution the sample and rerun the test. 2.Check if the reagent is confected or positioned
0-8 Warning
value over value (PC value),exceeds the in the right way. 3.Check if the specified upper/ lower limit is properly set up.
specified upper/lower limit.

Absorbency In Rate A assay and Rate B assay,


1.Test after dilute the sample. 2.Check if the reagent is confected or positioned in the
0-9 Warning reaction limit The check value of substrate is
right way. 3.Check if the check value is properly setup.
over larger than the setup limit value.

1.The denominator becomes zero in


calculation. 2.An over flow
occurs in logarithmic or 1.Check if there is a logical error in calculation formulas. 2.Check the concentration setup
Calculation
0-10 Warning exponential calculation. 3,the of standard liquid when multi-point calibration. 3.Check the absorbance value of photometry
disable
absorbance of test item is less point of the test.
than the absorbance of standard
liquid 1.

In multi-point calibration, the


Calculation absorbance of standard liquid do 1.Check if calibration liquid is positioned in the right way. 2.Check if absorbance value
0-11 Warning
disable not increase when concentration of different concentrationis increase. 3.Please rerun the calibration.
increase.

0-20 Warning Barcode repeat Barcode repeat on R1 disk Check the barcode stick information. 2.Scan barcode again after check.

Reagent R2 and R3
Reagent R2 and R3 is found on R1
0-21 Warning is found on R1 Check the barcode stick information. 2.Scan barcode again after check.
disk.
disk.

Detergent placed Detergent should be placed at 45


error at 45 position on R1 disk.or barcode
0-22 Warning Check the barcode stick information. 2.Scan barcode again after check.
position on R1 stick error at 45 position on R1
disk. disk.

1.when reagent residual volume is


R1 disk barcode less than setup reagent residual
0-23 Warning Check the barcode stick information. 2.Scan barcode again after check.
blank out. volume, the barcode is blank out.
2.barcode is used after blank out.

216
User Manual of CS-400 Auto-Chemistry Analyzer

Code Level Name Describe Countermeasure

1.Barcode checkup failure.


2.Barcode reagent name do not
R1 disk barcode 1.Check barcode stick information. 2.Scan barcode again after check. 3.Corresponding item
0-24 Warning occur in "barcode item setup". 3,
invalid is added in barcode item setup.
Barcode reagent type checkup
error.
R2 disk barcode
0-25 Warning R2 disk barcode repeat. Check the barcode stick information. 2.Scan barcode again after check.
repeat

Reagent R1 and R4
R1 and R4 detergent is found on R1
0-26 Warning is found on R2 Check the barcode stick information. 2.Scan barcode again after check.
or R4
disk

Detergent placed Detergent should be placed at 45


error at 45 position on R2 disk.Barcode
0-27 Warning Check the barcode stick information. 2.Scan barcode again after check.
position on R2 stick error at 45 position on R2
disk. disk.

1.when reagent residual volume is


R2 disk barcode less than setup reagent residual
0-28 Warning Check the barcode stick information. 2.Scan barcode again after check.
blank out. volume, the barcode is blank out.
2.barcode is used after blank out.

1.Barcode checkup failure.


2.Barcode reagent name do not
R2 disk barcode 1.Check barcode stick information. 2.Scan barcode again after check. 3.Corresponding item
0-29 Warning occur in "barcode item setup". 3,
invalid. is added in barcode item setup.
Barcode reagent type checkup
error.

Communication abnormal. No
Communication 1.Check if serial port line is well connected. 2.Restart the instrument and software and
0-254 Stop response when instruction send to
abnormal then carry out on-line instruction.
host computer.

Abnormal occur when communicate


Communication 1.Check if serial port line is well connected. 2.restart the instrument and software and
0-255 Stop with host computer and
abnormal then carry out on-line instruction.
communication halt.

217
User Manual of CS-400 Auto-Chemistry Analyzer

Code Level Name Describe Countermeasure

The R1 stirring mechanism does not


R1 Stirrer
reach the upper dead point in 1.Call up the mechanism check screen and carry out the check program "mechanism check ".
1-1 Stop mechanism
ascending motion,(at the rinsing 2.Various malfunctions or trouble can not be restored. Contact the service department.
abnormal
bath side)

The R1 stirring mechanism does not


R1 Stirrer
reach the upper dead point in 1.Call up the mechanism check screen and carry out the check program "mechanism check ".
1-2 Stop mechanism
ascending motion( at the reaction 2.Various malfunctions or trouble can not be restored. Contact the service department.
abnormal
cuvette side)

R1 Stirrer The R1 stirring mechanism does not


1.Call up the mechanism check screen and carry out the check program "mechanism check ".
1-3 Stop mechanism leave the upper dead point in
2.Various malfunctions or trouble can not be restored. Contact the service department.
abnormal descending motion.

R1 Stirrer The R1 stirring mechanism does not


1.Call up the mechanism check screen and carry out the check program "mechanism check ".
1-4 Stop mechanism reach the rinsing bath position in
2.Various malfunctions or trouble can not be restored. Contact the service department.
abnormal movement to the rinsing bath side.

The R1 stirring mechanism does not


R1 Stirrer
reach the reaction cuvette 1.Call up the mechanism check screen and carry out the check program "mechanism check ".
1-5 Stop mechanism
position in movement to the 2.Various malfunctions or trouble can not be restored. Contact the service department.
abnormal
reaction cuvette side

R1 Stirrer In resetting, the R1 stirring


1.Call up the mechanism check screen and carry out the check program "mechanism check ".
1-6 Stop mechanism mechanism does not reach the
2.Various malfunctions or trouble can not be restored. Contact the service department.
abnormal rinsing bath position

R1 Stirrer In resetting, the R1 stirring


1.Call up the mechanism check screen and carry out the check program "mechanism check ".
1-7 Stop mechanism mechanism does not leave the
2.Various malfunctions or trouble can not be restored. Contact the service department.
abnormal rinsing bath position

R1 Stirrer In rotation of R1 stirring


1.Call up the mechanism check screen and carry out the check program "mechanism check ".
1-8 Stop mechanism mechanism, it is not set at the
2.Various malfunctions or trouble can not be restored. Contact the service department.
abnormal upper dead point.

218
User Manual of CS-400 Auto-Chemistry Analyzer

Code Level Name Describe Countermeasure

R1 Stirrer
The R1 stirring mechanism has 1.Call up the mechanism check screen and carry out the check program "mechanism check ".
1-9 Stop mechanism
ascend /descend error 2.Various malfunctions or trouble can not be restored. Contact the service department.
abnormal

R1 Stirrer
The R1 stirring mechanism has sway 1.Call up the mechanism check screen and carry out the check program "mechanism check ".
1-10 Stop mechanism
error 2.Various malfunctions or trouble can not be restored. Contact the service department.
abnormal

The R2 stirring mechanism does not


R2 Stirrer
reach the upper dead point in 1.Call up the mechanism check screen and carry out the check program "mechanism check ".
2-1 Stop mechanism
ascending motion (at the rinsing 2.Various malfunctions or trouble can not be restored. Contact the service department.
abnormal
bath side)

The R2 stirring mechanism does not


R2 Stirrer
reach the upper dead point in 1.Call up the mechanism check screen and carry out the check program "mechanism check ".
2-2 Stop mechanism
ascending motion(at reaction 2.Various malfunctions or trouble can not be restored. Contact the service department.
abnormal
cuvette side)

R2 Stirrer The R2 stirring mechanism does not


1.Call up the mechanism check screen and carry out the check program "mechanism check ".
2-3 Stop mechanism leave the upper dead point in
2.Various malfunctions or trouble can not be restored. Contact the service department.
abnormal descending motion.

R2 Stirrer The R2 stirring mechanism does not


1.Call up the mechanism check screen and carry out the check program "mechanism check ".
2-4 Stop mechanism reach the rinsing bath position in
2.Various malfunctions or trouble can not be restored. Contact the service department.
abnormal movement to the rinsing bath side

The R2 stirring mechanism does not


R2 Stirrer
reach the reaction cuvette 1.Call up the mechanism check screen and carry out the check program "mechanism check ".
2-5 Stop mechanism
position in movement to the 2.Various malfunctions or trouble can not be restored. Contact the service department.
abnormal
reaction cuvette side

R2 Stirrer In resetting, the R2 stirring


1.Call up the mechanism check screen and carry out the check program "mechanism check ".
2-6 Stop mechanism mechanism does not reach the
2.Various malfunctions or trouble can not be restored. Contact the service department.
abnormal rinsing bath position

219
User Manual of CS-400 Auto-Chemistry Analyzer

Code Level Name Describe Countermeasure

R2 Stirrer In resetting, the R2 stirring


1.Call up the mechanism check screen and carry out the check program "mechanism check ".
2-7 Stop mechanism mechanism does not leave the
2.Various malfunctions or trouble can not be restored. Contact the service department.
abnormal rinsing bath position

R2 Stirrer In rotation of R2 stirring


1.Call up the mechanism check screen and carry out the check program "mechanism check ".
2-8 Stop mechanism mechanism, it is not set at the
2.Various malfunctions or trouble can not be restored. Contact the service department.
abnormal upper dead point.

R2 Stirrer
The R2 stirring mechanism has 1.Call up the mechanism check screen and carry out the check program "mechanism check ".
2-9 Stop mechanism
ascend /descend error 2.Various malfunctions or trouble can not be restored. Contact the service department.
abnormal

R2 Stirrer
The R2 stirring mechanism has sway 1.Call up the mechanism check screen and carry out the check program "mechanism check ".
2-10 Stop mechanism
error 2.Various malfunctions or trouble can not be restored. Contact the service department.
abnormal

Rinsing The rinse mechanism does not reach


1.Call up the mechanism check screen and carry out the check program "mechanism check ".
3-1 Stop mechanism the upper dead point in ascending
2.Various malfunctions or trouble can not be restored. Contact the service department.
abnormal motion

Rinsing The rinse mechanism dose not leave


1.Call up the mechanism check screen and carry out the check program "mechanism check ".
3-2 Stop mechanism the upper dead point in the
2.Various malfunctions or trouble can not be restored. Contact the service department.
abnormal descending motion.

1.This trouble is liable to occur after the reaction disk is washed. In this case, thoroughly
Reaction disk The reaction disk cannot detect wipe water droplets off the bottom of the reaction disk. 2.Check if water droplets adhere
4-1 Stop
abnormal the stop position. to the light coupler and code wheel located below the reaction disk. 3.if trouble can not
be restored, contact the service department.

1.This trouble is liable to occur after the reaction disk is washed. In this case, thoroughly
Reaction disk The reaction disk does not stop at wipe water droplets off the bottom of the reaction disk. 2.Check if water droplets adhere
4-2 Stop
abnormal the specified position to the light coupler and code wheel located below the reaction disk. 3.if trouble can not
be restored, contact the service department.

220
User Manual of CS-400 Auto-Chemistry Analyzer

Code Level Name Describe Countermeasure

1.This trouble is liable to occur after the reaction disk is washed. In this case, thoroughly
Reaction disk In resetting, the reaction disk wipe water droplets off the bottom of the reaction disk. 2.Check if water droplets adhere
4-3 Stop
abnormal cannot detect the home position to the light coupler and code wheel located below the reaction disk. 3.if trouble can not
be restored , contact the service department.

In resetting, the first reaction 1.This trouble is liable to occur after the reaction disk is washed. In this case, thoroughly
Reaction disk cuvette on the reaction disk does wipe water droplets off the bottom of the reaction disk. 2.Check if water droplets adhere
4-4 Stop
abnormal not stop at the specified to the light coupler and code wheel located below the reaction disk. 3.if trouble can not
position. be restored , contact the service department.

In rotation of the reaction disk


any of the sample probe, reagent
probe, stirring mechanism and
Reaction disk 1.Call up the mechanism check screen and carry out the check program "mechanism check".
4-5 Stop rinse mechanism is not at the
abnormal 2.Various malfunctions or trouble can not be restored. Contact the service department.
upper dead point (at the reaction
cuvette side) Warning: any other
alarm may occur simultaneously

CS-alkaline
CS-alkaline detergent place in
4-6 Warning detergent Please add CS-alkaline detergent in detergent bottle.
front of instrument is shortage.
shortage

The sample probe does not reach


Sample probe the upper dead point in descending 1.Call up the mechanism check screen and carry out the check program "mechanism check".
5-1 S .Stop
abnormal motion.(at other than the 2.Various malfunctions or trouble can not be restored. Contact the service department.
reaction cuvette side)

The sample probe does not reach


Sample probe the upper dead point in descending 1.Call up the mechanism check screen and carry out the check program "mechanism check".
5-2 Stop
abnormal motion.(at the reaction cuvette 2.Various malfunctions or trouble can not be restored. Contact the service department.
side)

The sample probe goes down


Sample probe 1.Check if there is sample in sample cup. 2.check if sample cup place in the correct position.
5-3 Warning abnormally in descending motion
abnormal 3.Check if sample cup is bend.
(at the sample cup side)

221
User Manual of CS-400 Auto-Chemistry Analyzer

Code Level Name Describe Countermeasure

1.Call up the mechanism check screen and carry out the check program "mechanism check".
The sample probe goes down 2.Various malfunctions or trouble can not be restored. Contact the service department.
Sample probe
5-4 Stop abnormally in descending motion 3.Check if check the altitude between probe and cup bottom is set properly. 4.Call up the
abnormal
(at the reaction cuvette side) system maintenance menu, carry out "sample probe horizontal checkup". 5.Check if the sample
cup is placed deflective.

The sample probe does not leave


Sample probe 1.Call up the mechanism check screen and carry out the check program "mechanism check".
5-5 S. Stop the upper dead point in descending
abnormal 2.Various malfunctions or trouble can not be restored. Contact the service department.
motion.(at the sample cup side)

The sample probe does not leave


Sample probe the upper dead point in descending 1.Call up the mechanism check screen and carry out the check program "mechanism check".
5-6 Stop
abnormal motion.(at the reaction cuvette 2.Various malfunctions or trouble can not be restored. Contact the service department.
side)

Sample probe Sample probe stay in liquid level


5-7 S. Stop Please contact service department when this error occur repetitively.
abnormal detect effective status.

The sample probe cannot detect the


Sample probe 1.Call up the mechanism check screen and carry out the check program "mechanism check".
5-8 S. Stop cell position in rotation to the
abnormal 2.Various malfunctions or trouble can not be restored. Contact the service department.
reaction cuvette side.

The sample probe does not leave


the reaction cuvette position in
Sample probe 1.Call up the mechanism check screen and carry out the check program "mechanism check".
5-9 S. Stop rotation from the reaction
abnormal 2.Various malfunctions or trouble can not be restored. Contact the service department.
cuvette side to any other
position.

Sample probe Sample probe descend abnormally 1.Call up the mechanism check screen and carry out the check program "mechanism check".
5-10 S. Stop
abnormal when discharge ISE test sample 2.Various malfunctions or trouble can not be restored. Contact the service department.

Abnormal touch occur (at the


Sample probe 1.Call up the mechanism check screen and carry out the check program "mechanism check".
5-11 Warning sample cup side) when sample probe
abnormal 2.Various malfunctions or trouble can not be restored. Contact the service department.
descend.

222
User Manual of CS-400 Auto-Chemistry Analyzer

Code Level Name Describe Countermeasure

Sample probe In rotation of reagent-1 probe, it 1.Call up the mechanism check screen and carry out the check program "mechanism check".
5-12 S. Stop
abnormal is not set at the upper dead point. 2.Various malfunctions or trouble can not be restored. Contact the service department.

Sample probe Sample probe cannot find sample


5-13 Warning Check if there is sample cup on sample disk.
abnormal cup

Sample probe Sample probe cannot reach sample 1.Check if there is obstacle on sample disk. 2.Various malfunctions or trouble can not be
5-14 S. Stop
abnormal liquid when descend. restored. Contact the service department.

Sample probe Sample probe mechanism ascend/ 1.Call up the mechanism check screen and carry out the check program "mechanism check".
5-15 Stop
abnormal descend abnormal 2.Various malfunctions or trouble can not be restored. Contact the service department.

Sample probe Sample probe mechanism sway 1.Call up the mechanism check screen and carry out the check program "mechanism check".
5-16 Stop
abnormal abnormal 2.Various malfunctions or trouble can not be restored. Contact the service department.

The sample disk does not detect


Sample disk 1.Call up the mechanism check screen and carry out the check program "mechanism check".
6-1 S. Stop the specified position on the
abnormal 2.Various malfunctions or trouble can not be restored. Contact the service department.
outer track

The sample disk does not stop in


Sample disk 1.Call up the mechanism check screen and carry out the check program "mechanism check".
6-2 S. Stop the specified position on the
abnormal 2.Various malfunctions or trouble can not be restored. Contact the service department.
outer track

The sample disk cannot detect the


Sample disk 1.Call up the mechanism check screen and carry out the check program "mechanism check".
6-3 S. Stop stop position on the intermediate
abnormal 2.Various malfunctions or trouble can not be restored. Contact the service department.
track.

The sample disk cannot stop in the


Sample disk 1.Call up the mechanism check screen and carry out the check program "mechanism check".
6-4 S. Stop specific position on the
abnormal 2.Various malfunctions or trouble can not be restored. Contact the service department.
intermediate track

Sample disk The sample disk cannot detect the 1.Call up the mechanism check screen and carry out the check program "mechanism check".
6-5 S. Stop
abnormal stop position on the inner track. 2.Various malfunctions or trouble can not be restored. Contact the service department.

223
User Manual of CS-400 Auto-Chemistry Analyzer

Code Level Name Describe Countermeasure

The sample disk cannot stop in the


Sample disk 1.Call up the mechanism check screen and carry out the check program "mechanism check".
6-6 S. Stop specific position on the inner
abnormal 2.Various malfunctions or trouble can not be restored. Contact the service department.
track

Sample disk In resetting, the sample disk 1.Call up the mechanism check screen and carry out the check program "mechanism check".
6-7 Stop
abnormal cannot detect the home position. 2.Various malfunctions or trouble can not be restored. Contact the service department.

Sample disk In resetting, the sample disk does 1.Call up the mechanism check screen and carry out the check program "mechanism check".
6-8 Stop
abnormal not leave the home position 2.Various malfunctions or trouble can not be restored. Contact the service department.

In resetting the sample disk does


Sample disk 1.Call up the mechanism check screen and carry out the check program "mechanism check".
6-9 Stop not stop at sample 1 position on
abnormal 2.Various malfunctions or trouble can not be restored. Contact the service department.
the outer track.

sample barcode Sample barcode reader cannot be 1.Check if barcode device is well connected. 2.Trouble cannot be restored, Contact the service
6-10 Warning
device abnormal found. department.

Sample syringe The sample syringe is not at the 1.Call up the mechanism check screen and carry out the check program "mechanism check".
7-1 S. Stop
abnormal upper dead point. 2.Various malfunctions or trouble can not be restored. Contact the service department.

Sample syringe The sample syringe does not leave 1.Call up the mechanism check screen and carry out the check program "mechanism check".
7-2 S. Stop
abnormal the upper dead point. 2.Various malfunctions or trouble can not be restored. Contact the service department.

The R1 probe does not reach the


R1 probe 1.Call up the mechanism check screen and carry out the check program "mechanism check".
8-1 Stop upper dead point in ascending
abnormal 2.Various malfunctions or trouble can not be restored. Contact the service department.
motion

1.Check if there is still reagent in the bottle. 2.Check if the reagent bottle is open. 3.Check
R1 probe The R1 probe goes down abnormally
8-2 Warning if the reagent disk cover is put in its proper place. 4.Check if the reagent-1 probe is bent.
abnormal in descending motion
5.Check if reagent bottle is lean placed.

The R1 probe does not leave the


R1 probe 1.Call up the mechanism check screen and carry out the check program "mechanism check".
8-3 Stop upper dead point in descending
abnormal 2.Various malfunctions or trouble can not be restored. Contact the service department.
motion

224
User Manual of CS-400 Auto-Chemistry Analyzer

Code Level Name Describe Countermeasure

R1 probe R1 probe stay in liquid level


8-4 Warning Please contact service department when this error occur repetitively.
abnormal detect effective status.

The R1 probe cannot detect the


R1 probe 1.Call up the mechanism check screen and carry out the check program "mechanism check".
8-5 Stop cell position in rotation to the
abnormal 2.Various malfunctions or trouble can not be restored. Contact the service department.
reaction cuvette side.

The R1 probe cannot leave the cell


R1 probe 1.Call up the mechanism check screen and carry out the check program "mechanism check".
8-6 Stop position in rotation to the
abnormal 2.Various malfunctions or trouble can not be restored. Contact the service department.
reaction cuvette side.

R1 probe In rotation of R1 probe, it is not 1.Call up the mechanism check screen and carry out the check program "mechanism check".
8-7 Stop
abnormal set at the upper dead point. 2.Various malfunctions or trouble can not be restored. Contact the service department.

R1 probe R1 probe cannot find reagent


8-8 Warning 1.Check if there is reagent bottle on R1 reagent disk.
abnormal bottle

R1 probe R1 probe cannot reach reagent 1.Check if there is obstacle on R1 reagent disk. 2.trouble cannot be restored, contact the
8-9 Stop
abnormal liquid level when descend. service department.

R1 probe R1 probe mechanism ascend/ 1.Call up the mechanism check screen and carry out the check program "mechanism check".
8-10 Stop
abnormal descend abnormal 2.Various malfunctions or trouble can not be restored. Contact the service department.

R1 probe 1.Call up the mechanism check screen and carry out the check program "mechanism check".
8-11 Stop R1 probe mechanism sway abnormal
abnormal 2.Various malfunctions or trouble can not be restored. Contact the service department.

The R2 probe does not reach the


R2 probe 1.Call up the system maintenance menu and carry out the check program "reset". 2.Various
9-1 Stop upper dead point in ascending
abnormal malfunctions or trouble can not be restored. Contact the service department.
motion

1.Check if there is still reagent in the bottle. 2.Check if the reagent bottle is open. 3.Check
R2 probe The R2 probe goes down abnormally
9-2 Warning if the reagent disk cover is put in its proper place. 4.Check if the reagent-1 probe is bent.
abnormal in descending motion
5.Check if reagent bottle is lean placed.

225
User Manual of CS-400 Auto-Chemistry Analyzer

Code Level Name Describe Countermeasure

The R2 probe does not leave the


R2 probe 1.Call up the mechanism check screen and carry out the check program "mechanism check".
9-3 Stop upper dead point in descending
abnormal 2.Various malfunctions or trouble can not be restored. Contact the service department.
motion

R2 probe R2 probe stay in liquid level


9-4 Warning Contact the service department when this error occurs repetitively.
abnormal detect effective status.

The R2 probe cannot detect the


R2 probe 1.Call up the mechanism check screen and carry out the check program "mechanism check".
9-5 Stop cell position in rotation to the
abnormal 2.Various malfunctions or trouble can not be restored. Contact the service department.
reaction cuvette side.

The R2 probe cannot leave the cell


R2 probe 1.Call up the mechanism check screen and carry out the check program "mechanism check".
9-6 Stop position in rotation to the
abnormal 2.Various malfunctions or trouble can not be restored. Contact the service department.
reaction cuvette side.

R2 probe In rotation of R2 probe, it is not 1.Call up the mechanism check screen and carry out the check program "mechanism check".
9-7 Stop
abnormal set at the upper dead point. 2.Various malfunctions or trouble can not be restored. Contact the service department.

R2 probe R2 probe cannot find reagent


9-8 Warning Check if there is reagent bottle on R2 reagent disk.
abnormal bottle.

R2 probe R2 probe cannot reach reagent 1.Check if there is obstacle on R2 reagent disk. 2.Trouble cannot be restored, contact the
9-9 Stop
abnormal liquid level when descend. service department.

R2 probe R2 probe mechanism ascend/descend


9-10 Stop Various malfunctions or trouble can not be restored. Contact the service department.
abnormal abnormal

R2 probe R2 probe mechanism ascend/descend


9-11 Stop Various malfunctions or trouble can not be restored. Contact the service department.
abnormal abnormal

The R1 disk cannot detect the stop 1.Call up the mechanism check screen and carry out the check program "mechanism check".
10-1 Stop R1 disk abnormal
position 2.Various malfunctions or trouble can not be restored. Contact the service department.

226
User Manual of CS-400 Auto-Chemistry Analyzer

Code Level Name Describe Countermeasure

The R1 disk does not stop at the 1.Call up the mechanism check screen and carry out the check program "mechanism check".
10-2 Stop R1 disk abnormal
specified position 2.Various malfunctions or trouble can not be restored. Contact the service department.

The R1 disk cannot detect the home 1.Call up the mechanism check screen and carry out the check program "mechanism check".
10-3 Stop R1 disk abnormal
position. 2.Various malfunctions or trouble can not be restored. Contact the service department.

Bar code device


Barcode reader of R1 disk cannot 1.Check if barcode device is well connected. 2.Various malfunctions or trouble can not be
10-4 Warning of R1 reagent
be find restored. Contact the service department.
disk abnormal

The R2 disk does not stop at the 1.Call up the mechanism check screen and carry out the check program "mechanism check".
11-2 Stop R2 disk abnormal
specified position 2.Various malfunctions or trouble can not be restored. Contact the service department.

The R2 disk cannot detect the home 1.Call up the mechanism check screen and carry out the check program "mechanism check".
11-3 Stop R2 disk abnormal
position. 2.Various malfunctions or trouble can not be restored. Contact the service department.

Bar code device


Barcode reader of R2 disk cannot 1.Check if barcode device is well connected. 2.Various malfunctions or trouble can not be
11-4 Warning of R2 reagent
be find restored . Contact the service department.
disk abnormal

R1 Syringe The R1 syringe is not at the upper


14-1 Stop When this error occurs repetitively, please contact service department.
abnormal dead point.

R1 Syringe The R1 syringe does not leave from


14-2 Stop When this error occurs repetitively, please contact service department.
abnormal the upper dead point

R2 Syringe The R2 syringe is not at the upper


15-1 Stop When this error occurs repetitively, please contact service department.
abnormal dead point.

R2 Syringe The R2 syringe does not leave from


15-2 Stop When this error occurs repetitively, please contact service department.
abnormal the upper dead point

227
User Manual of CS-400 Auto-Chemistry Analyzer

Code Level Name Describe Countermeasure

SIP sipper does not reach the 1.Call up the system maintenance menu,first carry out the program "reset",then go on
Electrolyte SIP
16-1 Warning lower dead point (when reset and "mechanism check" 2.Various malfunctions or trouble can not be restored. Contact the service
Sipper abnormal
operate) department.

1.Call up the system maintenance menu,first carry out the program "reset",then go on
Electrolyte SIP SIP sipper does not leave the
16-2 Warning "mechanism check" 2.Various malfunctions or trouble can not be restored. Contact the service
Sipper abnormal lower dead point
department.

Electrolyte 1.Call up the system maintenance menu,first carry out the program "reset",then go on
The vacuum nozzle does not reach
17-1 Warning vacuum Sipper "mechanism check" 2.Various malfunctions or trouble can not be restored. Contact the service
the lower dead point
abnormal department.

Electrolyte 1.Call up the system maintenance menu,first carry out the program "reset",then go on
The vacuum nozzle does not leave
17-2 Warning vacuum Sipper "mechanism check" 2.Various malfunctions or trouble can not be restored. Contact the service
the lower dead point.
abnormal department.

Electrolyte
SIP syringe does not at the upper
18-1 Warning sample Syringe Contact the service department.
dead point.
abnormal

Electrolyte
SIP syringe does not leave the
18-2 Warning sample Syringe Contact the service department.
upper dead point.
abnormal

Electrolyte
The diluent syringe is not at the
18-3 Warning sample Syringe Contact the service department.
upper dead point.
abnormal

Electrolyte
The diluent syringe is not leave
18-4 Warning sample Syringe Contact the service department.
the upper dead point.
abnormal

Electrolyte The internal standard solution


18-5 Warning sample Syringe syringe is not at the upper dead Contact the service department.
abnormal point.

228
User Manual of CS-400 Auto-Chemistry Analyzer

Code Level Name Describe Countermeasure

Electrolyte The internal standard solution


18-6 Warning sample Syringe syringe is not leave the upper Contact the service department.
abnormal dead point.

The ISE function stops due to an


1.Call up the system maintenance menu,first carry out the program "reset",then go on
alarm.Warning:Displayed if
19-1 Warning ISE stop "mechanism check" 2.Various malfunctions or trouble can not be restored. Contact the service
restart is given in sample stop
department.
status.

Incubation bath
The water temperature in the 1.Check if cooling fan is operate normally. 2.check if colling fan dustproof cover is cloged
20-1 Warning water temp
incubation bath is over 45C by dirt 3. Trouble can not be restored, contact the service department.
abnormal

The water temperature in the


Incubation bath 1.Check if the room temperature is within a range of 15-32C. 2.Check if cooling fan is operated
incubation bath is over the range
20-2 Warning water temp normally. 3.Check if cooling fan dustproof cover is cloged by dirt. 4.Check if the incubation
of 37 plus or minus 0.5C.(check in
abnormal bath water is circulation. 5.Trouble can not be restored, contact the service department.
operating)

1.Clean liquid level sensor probe of incubation bath. 2.Check if there are water in purified
Incubator water The water level in the incubation
21-1 Warning water machine and pipeline 3. Please restart the instrument. 4.If the trouble cann't be
level error bath is too low
restored, contact the service department.

A period of 24 hours has passed


Incubation bath
22-1 Warning since exchanging water in the Carry out "Incubator water exchange" or restart the instrument.
water exchange
incubation bath.

Water level too The water level in the water tank The distilled water tank is not supplied with water, Check the water pressure, water cock,
23-1 Stop
low is too low and other water supply systems.

Waste water in The vacuum tank contains some


28-1 Warning Empty the water in vacuum tank according to the 《User manual".
Vacuum Tank waste water

Reagent residual volume in R1 or


R4 reagent bottle on R1 disk is
Reagent 1.Check if reagent residual volume on the reagent information screen is lower than its setup
31-1 Warning less than the setup value of
shortage(R1) value. 2.Check if the setup value of reagent residual volume is reasonable..
reagent residual volume in system
setup interface.

229
User Manual of CS-400 Auto-Chemistry Analyzer

Code Level Name Describe Countermeasure

Reagent volume in R2 or R3 reagent


Reagent bottle on R2 disk is less than the 1.Check if reagent residual volume on the reagent information screen is lower than its setup
32-1 Warning
shortage(R2) setup value of reagent residual value. 2.Check if the setup value of reagent residual volume is reasonable..
volume in system setup interface.

The Detergent volume on No.45


Detergent
33-1 Warning position of R1 disk is less than Call for the reagent information screen, check if detergent is enough.
low(R1)
5ml.

The Detergent volume on No.45


Detergent
34-1 Warning position of R2 disk is less than Call for the reagent information screen, check if detergent is enough.
low(R2)
5ml.

1.Reagent name setup error,


reagent residual test time cannot
Reagent
be checked. 2.reagent type setup 1.Check barcode stick information, scan the barcode again after check. 2.register the reagent
35-1 Warning information
error, reagent residual test time information manually.
setup error(R1)
cannot be checked. 3.the item is
not set in analyze parameter.

1.Reagent name setup error,


reagent residual test time cannot
Reagent
be checked. 2.reagent type setup 1.Check barcode stick information, scan the barcode again after check. 2.register the reagent
36-1 Warning information
error, reagent residual test time information manually.
setup error(R2)
cannot be checked. 3.the item is
not set in analyze parameter.

Sampling Sample registered finished


41-1 Warning
finished sampling, stop to adding.

Automatic Timing calibration is not carried 1.Call up the system maintenance screen, first carry out the program "reset", then go on
51-1 Warning calibration out when reagent residual volume "mechanism check". 2.Various malfunctions or trouble can not be restored. Contact the service
disable cannot complete 10 tests. department.

230
User Manual of CS-400 Auto-Chemistry Analyzer

Code Level Name Describe Countermeasure

The mean electromotive force of


internal standard solution (EAV)
1.Carry out "ISE Check" on the Mechanisms Check screen. 2.If it is not easy for the user
is out of the following range at
60-1 Warning ISE LEVEL Error to remove the cause of trouble, contact the service department. 3.For details refer to
3 of 5 measuring points(internal
"electrolyte device maintenance" in "User manual".
standard solution)Na:-90.0mv≤
EAV≤-10mv

The mean electromotive force of


internal standard solution (EAV)
1.Carry out "ISE Check" on the Mechanisms Check screen. 2.If it is not easy for the user
is out of the following range at
60-2 Warning ISE LEVEL Error to remove the cause of trouble, contact the service department. 3.For details refer to
3 of 5 measuring points(internal
"electrolyte device maintenance" in "User manual".
standard solution)K:-90.0mv≤EAV
≤-10mv

The mean electromotive force of


internal standard solution (EAV)
1.Carry out "ISE Check" on the Mechanisms Check screen. 2.If it is not easy for the user
is out of the following range at
60-3 Warning ISE LEVEL Error to remove the cause of trouble, contact the service department. 3.For details refer to
3 of 5 measuring points(internal
"electrolyte device maintenance" in "User manual".
standard solution)Cl:80.0mv≤EAV
≤160mv

The difference between the


maximum and minimum electromotive
forces of internal standard
solution (FIV) is within the 1.Carry out "ISE Check" on the Mechanisms Check screen. 2.If it is not easy for the user
61-1 Warning ISE Noise Error following range at 3 of 5 to remove the cause of trouble, contact the service department. 3.For details refer to
measuring points.(internal "electrolyte device maintenance" in "User manual".
standard solution,
sample)Na:0.7mv≤
|FIV(2)-FIV(4)|

231
User Manual of CS-400 Auto-Chemistry Analyzer

Code Level Name Describe Countermeasure

The difference between the


maximum and minimum electromotive
forces of internal standard
1.Carry out "ISE Check" on the Mechanisms Check screen. 2.If it is not easy for the user
solution (FIV) is within the
61-2 Warning ISE Noise Error to remove the cause of trouble, contact the service department. 3.For details refer to
following range at 3 of 5
"electrolyte device maintenance" in "User manual".
measuring points.(internal
standard solution,
sample)K:1.0mv≤|FIV(2)-FIV(4)|

The difference between the


maximum and minimum electromotive
forces of internal standard
1.Carry out "ISE Check" on the Mechanisms Check screen. 2.If it is not easy for the user
solution (FIV) is within the
61-3 Warning ISE Noise Error to remove the cause of trouble, contact the service department. 3.For details refer to
following range at 3 of 5
"electrolyte device maintenance" in "User manual".
measuring points.(internal
standard solution, sample)
Cl:0.8mv≤|FIV(2)-FIV(4)|

1.Upon calibration, the slope


level is within the following
1.Carry out "ISE Check" on the Mechanisms Check screen. 2.If it is not easy for the user
ISE Prepare range. 2.The electrode response
62-1 Warning to remove the cause of trouble, contact the service department. 3.For details refer to
Error is degraded (the carryover rate A
"electrolyte device maintenance" in "User manual".
is as follows)Na:(1)32.0mv≤
SLOPE≤37.0mv or 68.1mv≤SLOPE

1.Upon calibration, the slope


level is within the following
1.Carry out "ISE Check" on the Mechanisms Check screen. 2.If it is not easy for the user
ISE Prepare range. 2.The electrode response
62-2 Warning to remove the cause of trouble, contact the service department. 3.For details refer to
Error is degraded (the carryover rate A
"electrolyte device maintenance" in "User manual".
is as follows)K:(1)32.0mv≤SLOPE
≤37.0mv or 68.1mv≤SLOPE

232
User Manual of CS-400 Auto-Chemistry Analyzer

Code Level Name Describe Countermeasure

1.Upon calibration, the slope


level is within the following
1.Carry out "ISE Check" on the Mechanisms Check screen. 2.If it is not a single malfunction,
range. 2.The electrode response
62-3 Warning ISE Slope Error contact the service department. 3.For details refer to "electrolyte device maintenance" in
is degraded (the carryover rate A
"User manual".
is as follows) Cl:(1)-30.0mv≤
SLOPE≤-25.0mv or -68.1mv≥SLOPE

1.Upon calibration, the slope


level is within the following
1.Carry out "ISE Check" on the Mechanisms Check screen. 2.If it is not a single malfunction,
range. 2.The electrode response
63-1 Warning ISE Slope Error contact the service department. 3.For details refer to "electrolyte device maintenance" in
is degraded (the carryover rate A
"User manual".
is as follows) Na:(1)SLOPE <
32.0mv

1.Upon calibration, the slope


level is within the following 1.Carry out "ISE Check" on the Mechanisms Check screen. 2.If it is not a single malfunction,
63-2 Warning ISE Slope Error range. 2.The electrode response contact the service department. 3.For details refer to "electrolyte device maintenance" in
is degraded (the carryover rate A "User manual".
is as follows) K:(1)SLOPE < 32.0mv

1.Upon calibration, the slope


level is within the following
1.Carry out "ISE Check" on the Mechanisms Check screen. 2.If it is not a single malfunction,
range. 2.The electrode response
63-3 Warning ISE Slope Error contact the service department. 3.For details refer to "electrolyte device maintenance" in
is degraded (the carryover rate A
"User manual".
is as follows)Cl:(1)SLOPE >
-25.0mv

The concentration of internal


ISE 1.Carry out "ISE Check" on the Mechanisms Check screen. 2.If it is not a single malfunction,
standard solution C(IS)is within
64-1 Warning concentration contact the service department. 3.For details refer to "electrolyte device maintenance" in
the following range: Na:C(IS)<
error "User manual".
120.0mmol/l or 160.0mmol/< C(IS)

233
User Manual of CS-400 Auto-Chemistry Analyzer

Code Level Name Describe Countermeasure

The concentration of internal


ISE 1.Carry out "ISE Check" on the Mechanisms Check screen. 2.If it is not a single malfunction,
standard solution C(IS)is within
64-2 Warning concentration contact the service department. 3.For details refer to "electrolyte device maintenance" in
the following range: K:C(IS)<
error "user manual".
3.0mmol/l or 7.0mmol/< C(IS)

The concentration of internal


ISE 1.Carry out "ISE Check" on the Mechanisms Check screen. 2.If it is not a single malfunction,
standard solution C(IS)is within
64-3 Warning concentration contact the service department. 3.For details refer to "electrolyte device maintenance" in
the following range: Cl:C(IS)<
error "user manual".
80.0mmol/l or 120.0mmol/< C(IS)

Carry out ISE calibration after


ISE require
65-1 Warning ISE maintenance(ISE rinsing、 Carry out ISE calibration.
calibration
rinsing all)

Time sync Time sync instruction 1.Call up the system maintenance menu and carry out the check program "reset". 2.Various
143-1 Stop
failure transmitting failure. malfunctions or trouble can not be restored . Contact the service department.

1.Check the supplying pipe whether air exists in.check the height of water tank whether it
Add water Water tank liquid path system is more than 1.5 meters (distance from ground)if it will be used.2.Check if water supplier,
143-2 Warning
overtime error, add water overtime. pipeline and filter is normal. 3.Various malfunctions or trouble can not be restored. Contact
the service department.

AD board reset Error occurs when AD board reset, 1.Call up the system maintenance menu and carry out the check program "reset". 2.Various
143-3 Warning
failure AD board reset failure. malfunctions or trouble can not be restored. Contact the service department.

Error occurs when reaction disk 1.Check if there is waterdrop on the code wheel. 2.Call up the system maintenance menu and
Reaction disk
143-4 Warning resetting and reaction reset carry out the check program "reset". 3.Various malfunctions or trouble can not be restored.
reset failure
failure. Contact the service department.

Error occurs when sample disk


Sample disk 1.Check if there is substance in sample container. 2.Various malfunctions or trouble can
143-5 Warning resetting and sample disk reset
reset failure not be restored. Contact the service department.
failure.

234
User Manual of CS-400 Auto-Chemistry Analyzer

Code Level Name Describe Countermeasure

Error occurs when R1 disk


R1 disk reset 1.Check if there is substance in reagent R1 container. 2.Various malfunctions or trouble
143-6 Warning resetting and R1 disk reset
failure can not be restored. Contact the service department.
failure.

Error occurs when R2 disk


R2 disk reset 1.Check if there is substance in reagent R2 container. 2.Various malfunctions or trouble
143-7 Warning resetting and R2 disk reset
failure can not be restored. Contact the service department.
failure.

Water discharge Incubation bath discharge 1.Check if incubation filter and diluent waste pipeline is clog. 2.Various malfunctions or
143-8 Warning
failure failure. trouble can not be restored. Contact the service department.

Detergent is not added completely


Adding detergent 1.Call up the system maintenance menu and carry out the check program "reset" 2.Various
143-9 Warning in specified time by reagent probe
overtime malfunctions or trouble can not be restored. Contact the service department.
1 and reagent 2.

R1 liquid level R1 probe cannot detect the liquid 1.Call up the system maintenance menu and carry out the check program "reset" 2.Various
143-10 Warning
detect failure level when adding detergent. malfunctions or trouble can not be restored. Contact the service department.

R2 liquid level R2 probe cannot detect the liquid 1.Call up the system maintenance menu and carry out the check program "reset". 2.Various
143-11 Warning
detect failure level when adding detergent. malfunctions or trouble can not be restored. Contact the service department.

Incubation bath 1.Check if liquid level sensor of incubation bath is clean and if there are water in incubation
Incubation bath liquid level
143-12 Warning liquid level bath 2.Restart the instrument. 3.Various malfunctions or trouble can not be restored.
detect failure.
detect failure. Contact the service department.

Incubation bath 1.Check if liquid level sensor of incubation bath is clean and if there are water in incubation
Add water overtime in incubation
143-13 Warning liquid path bath 2.Restart the instrument. 3.Various malfunctions or trouble can not be restored.
bath.
error Contact the service department.

Barcode scanning 1.Call up the system maintenance menu and carry out the check program "reset". 2.Various
143-14 Warning Barcode scanning overtime.
overtime. malfunctions or trouble can not be restored. Contact the service department.

235
User Manual of CS-400 Auto-Chemistry Analyzer

Code Level Name Describe Countermeasure

1.Call up the system maintenance menu and carry out the check program "reset". 2.Various
143-15 Warning Degas overtime Degas overtime
malfunctions or trouble can not be restored. Contact the service department.

Reaction
1.Call up the system maintenance menu and carry out the check program "reset". 2.Various
143-16 Stop initialize Reaction initialize failure
malfunctions or trouble can not be restored. Contact the service department.
failure

Reaction disk 1.Call up the system maintenance menu and carry out the check program "reset". 2.Various
143-17 Stop Reaction disk stop failure
stop failure malfunctions or trouble can not be restored. Contact the service department.

Sample probe
143-18 S .Stop Sample probe is block To remove the block please according to 12.4.1 in "user manual".
block

Add sample
143-19 S .Stop Last add sample failure After testing upon added sample, call up system maintenance menu, carry out "reset".
failure

143-20 S .Stop Add R1 failure Last add R1 failure. After testing upon added sample, call up system maintenance menu, carry out "reset"

143-21 S .Stop Add R2 failure Last add R2 failure After testing upon added sample, call up system maintenance menu, carry out "reset"

143-22 S .Stop Add R3 failure Last add R3 failure After testing upon added sample, call up system maintenance menu, carry out "reset"

143-23 S .Stop Add R4 failure Last add R4 failure

R1 stirring
143-24 S .Stop Last R1 stirrer failure After testing upon added sample, call up system maintenance menu, carry out "reset"
failure

R2 stirring
143-25 S .Stop Last R2 stirrer failure After testing upon added sample, call up system maintenance menu, carry out "reset"
failure

R3 stirring
143-26 S .Stop Last R3 stirrer failure After testing upon added sample, call up system maintenance menu, carry out "reset"
failure

236
User Manual of CS-400 Auto-Chemistry Analyzer

Code Level Name Describe Countermeasure

R4 stirring
143-27 S .Stop Last R4 stirrer failure After testing upon added sample, call up system maintenance menu, carry out "reset"
failure

Waste bottle
143-28 Warning The waste bottle is full Empty the waste bottle.
full

Error on float switch, high liquid


level float switch could detect
Float switch 1.Call up the system maintenance menu and carry out the check program "reset ". 2.Various
143-29 Stop the signal but the low liquid
error malfunctions or trouble can not be restored. Contact the service department.
level float switch cannot. is not
detect the signal.

Reagent
horizontal Reagent horizontal scanning 1.Call up the system maintenance menu and carry out the check program "reset ". 2.Various
143-30 Warning
scanning overtime. malfunctions or trouble can not be restored. Contact the service department.
overtime.

Vacuum pump Vacuum pump negative pressure too


143-31 Warning Please contact the maintenance department.
error low.

Barcode scanning Barcode scanning overtime when


143-32 Warning After testing upon added sample, call up system maintenance menu, carry out "reset".
overtime. testing.

Error occur in ISE resetting


ISE resetting 1.Call up the system maintenance menu and carry out the check program "reset ". 2.Various
143-33 Stop process, and the resetting do not
failure malfunctions or trouble can not be restored. Contact the service department.
accomplish.

ISE check
1.Call up the system maintenance menu and carry out the check program "reset ". 2.Various
143-34 Warning operation ISE check operation overtime.
malfunctions or trouble can not be restored. Contact the service department.
overtime

ISE pipeline 1.Call up the system maintenance menu and carry out the check program "reset ". 2.Various
143-35 Warning ISE pipeline rinsing overtime.
rinsing overtime malfunctions or trouble can not be restored. Contact the service department.

237
User Manual of CS-400 Auto-Chemistry Analyzer

Code Level Name Describe Countermeasure

ISE testing Error occur when ISE testing, ISE 1.Call up the system maintenance menu and carry out the check program "reset ". 2.Various
143-36 Warning
error testing stop. malfunctions or trouble can not be restored. Contact the service department.

Gear wheel pump


143-37 Stop Gear wheel pump pressure too low. Please contact the maintenance department.
error

R1 disk cover
143-38 Warning R1 disk cover is open.
open

R1 disk cover is close, instrument


R1 disk cover
143-39 Warning will automatically carry out
close
reagent horizontal scanning.

R2 disk cover
143-40 Warning R2 disk cover is open.
open.

R2 disk cover is close, instrument


R2 disk cover
143-41 Warning will automatically carry out
close.
reagent horizontal scanning.

Mapping Transmit reagent mapping


1.Call up the system maintenance menu and carry out the check program "reset". 2.Various
143-42 Stop information information failure and add
malfunctions or trouble can not be restored. Contact the service department
transmit failure reagent may failure too.

When water tank temperature


exceed 36.5 degree, infuse cold
1.Room temperature is too high or water supply equipment error occur. 2.Please contact the
143-43 Warning Cooling overtime water into water tack, and water
maintenance department.
tank temperature cannot reduce to
35.5 degree in 1 minute.

Instrument Error occurs between instrument 1.Pwer off and restart the instrument. 2.Various malfunctions or trouble cannot be restored.
143-44 Stop
module error modules. Contact the service department.

238
User Manual of CS-400 Auto-Chemistry Analyzer

Code Level Name Describe Countermeasure

Continue
Continually 5 contaminated cup Carry out cell blank test to judge cup states. If cell blank value is abnormal, please change
143-45 Stop contaminated cup
occur. reaction cuvette cup, if cell blank value is normal please contact maintenance department.
occur
In resetting, if trouble cannot remove, or other malfunctions exist, please contact
143-46 Stop AD data error AD data error
maintenance department.

ISE test internal standard liquid In resetting, if trouble cannot remove, or other malfunctions exist, please contact
143-47 Warning Test ISE error
failure maintenance staff.

Read edition 1.Call up the system maintenance menu and carry out the check program "reset". 2.Various
143-48 Warning Read edition number overtime
number overtime malfunctions or trouble can not be restored. Contact the service department.

1.Check if reagent disk lid is covered. Check if environment temperature is confirm to


Cooling system
144-1 Warning Cooling time abnormal instrument requirement. 2.If trouble cannot be removed,or other malfunction exist, please
abnormal
contact the maintenance department.

Cooling system
144-2 Warning Cooling circuit abnormal Please contact the maintenance department.
abnormal
Cooling system
144-3 Warning Cooling chip abnormal Please contact the maintenance department.
abnormal
Cooling system
144-4 Warning Cooling liquid level abnormal Please add water in cooling system water take according to 12.4.8 in the "user manual".
abnormal

Cooling system
144-5 Warning Cooling status abnormal Please contact the maintenance department.
abnormal

Ambience Ambience temperature is over


144-6 Warning Please contact the service department.
temperature over 15-32 degree.

Path 1 cooling Path 1 cooling circuit is less


145-1 Warning Please contact the maintenance department.
chip error than 5A

Path 2 cooling Path 2 cooling circuit is less


145-2 Warning Please contact the maintenance department.
chip error than 5A

239
User Manual of CS-400 Auto-Chemistry Analyzer

Code Level Name Describe Countermeasure

Path 3 cooling Path 3 cooling circuit is less


145-3 Warning Please contact the maintenance department.
chip error than 5A

Path 4 cooling Path 4 cooling circuit is less


145-4 Warning Please contact the maintenance department.
chip error than 5A

Path 5 cooling Path 5 cooling circuit is less


145-5 Warning Please contact the maintenance department.
chip error than 5A

Path 6 cooling Path 6 cooling circuit is less


145-6 Warning Please contact the maintenance department.
chip error than 5A

Path 1 AD Path 1 AD collector value has


146-1 Warning Please contact the maintenance department.
collector error exceed normal range.

Path 2 AD Path 2 AD collector test value has


146-2 Warning Please contact the maintenance department.
collector error exceed normal range.

Path 3 AD Path 3 AD collector test value has


146-3 Warning Please contact the maintenance department.
collector error exceed normal range.

Path 4 AD Path 4 AD collector test value has


146-4 Warning Please contact the maintenance department.
collector error exceed normal range.

Path 5 AD Path 5 AD collector test value has


146-5 Warning Please contact the maintenance department.
collector error exceed normal range.

Path 6 AD Path 6 AD collector test value has


146-6 Warning Please contact the maintenance department.
collector error exceed normal range.

Path 7 AD Path 7 AD collector test value has


146-7 Warning Please contact the maintenance department.
collector error exceed normal range.

Path 8 AD Path 8 AD collector test value has


146-8 Warning Please contact the maintenance department.
collector error exceed normal range.

Path 9 AD Path 9 AD collector test value has


146-9 Warning Please contact the maintenance department.
collector error exceed normal range.

240
User Manual of CS-400 Auto-Chemistry Analyzer

Code Level Name Describe Countermeasure

Path 10 AD Path 10 AD collector test value


146-10 Warning Please contact the maintenance department.
collector error has exceed normal range.

Path 11 AD Path 11 AD collector test value


146-11 Warning Please contact the maintenance department.
collector error has exceed normal range.

Path 12 AD Path 12 AD collector test value


146-12 Warning Please contact the maintenance department.
collector error has exceed normal range.

ISE Internal
Internal standard liquid shortage 1. Add ISE internal standard liquid. 2. Call up system maintenance screen, carry out ISE
38-1 Warning standard liquid
(Less than setup volume). internal standard liquid pipeline washing. 3. Carry out ISE calibration. .
shortage

ISE Diluent Diluent liquid shortage (Less


39-1 Warning Please check if ISE reagent volume is enough.
liquid shortage than set volume).

The remaining volume of Reference


ISE Reference
37-1 Warning Liquid is shortage (less than set Please check if ISE reagent volume is enough.
Liquid shortage
volume).

R1 reagent
10-5 Stop number R1 reagent number invalidity. Please contact the service department.
invalidity

R2 reagent
11-5 Stop number R2 reagent number invalidity. Please contact the service department.
invalidity

Sample probe Detergent is not detected when


5-17 Warning Check if detergent is placed at W1, W2, W3.
abnormal sample descending.

ISE exceed the


ISE QC exceed the lower limit of
0-31 Warning lower limit of Please check if Controls volume and reagent volume is enough, if the position is right.
measuring range.
measuring range

241
User Manual of CS-400 Auto-Chemistry Analyzer

Code Level Name Describe Countermeasure

ISE exceed the


ISE QC exceed the upper limit of
0-32 Warning upper limit of Please check if Controls volume and reagent volume is enough, if the position is right.
measuring range.
measuring range

ISE exceed the


ISE QC exceed the lower limit of
0-33 Warning lower limit of Please check if Controls volume and reagent volume is enough, if the position is right.
measuring range.
measuring range

ISE exceed the


ISE QC exceed the upper/lower
0-34 Warning upper limit of Please check if Controls volume and reagent volume is enough, if the position is right.
limit of measuring range.
measuring range

ISE exceed the ISE QC exceed the upper limit of


0-35 Warning Please check if sample volume and reagent volume is enough, if the position is right.
measuring range measuring range.

ISE calculated In calculation the denominator


0-12 Warning Please check if ISE reagent and sample are enough,and in right position.
error became zero

The R2 disk cannot detect the stop 1.Call up the mechanism check screen and carry out the check program "mechanism check".
11-1 Stop R2 disk abnormal
position 2.Various malfunctions or trouble can not be restored. Contact the service department.

Cooling status (1). Cooling system is not


144-7 Warning cannot be connected. (2). Cooling system Please contact maintenance staff.
confirmed. malfunction.

Cooling fan
144-8 Warning Cooling fan running abnormal. Please contact maintenance staff.
malfunction

New ISE calibration is required if


ISE request
0-13 Warning overtime calibration occurs in Execute ISE calibration
calibration
ISE calibration.

242
User Manual of CS-400 Auto-Chemistry Analyzer

Code Level Name Describe Countermeasure

Illegal test Illegal test number and reagent


143-49 Warning Please contact the service department.
item volume

Concentrated
waste liquid Concentrated waste liquid Call up the system maintenance menu, carry out the program "Auto Rinsing Concentrated waste
0-14 Warning
pipeline need to pipeline need to Rins liquid pipeline"
Rins

Note: Alarm code is composed of big sort and small sort.

××× — ×××

Small sort

Big sort

243
User Manual of CS-400 Auto-Chemistry Analyzer

Chapter 14 Instrument Transportation and Storage

14.1 Transportation requirement

Waterproof and moisture proof is required while transportation. Do not extrude or vibrate the instrument.
Handle it gently while carrying and loading.

14.2 Storage requirement

Instrument should be stored in clean room with no chemical, no aggressive gas.

14.3 Storage environment

Instrument should be stored in clean room with no chemical, no corrosive gas. Height above sea level within
2000 meter, temperature within -10℃~40℃, relative humidity within: 40% ~85% atmospheric
pressure within: 76kPa~106kPa.

244
User Manual of CS-400 Auto-Chemistry Analyzer

Addendum A Product Warranty


Dear customer:

Thank you for purchasing CS-400 Auto-Chemistry Analyzer of our company. We can offer you the following
service:

1. Technique consultation is provided at any time.

2. One year warranty from the day purchased. If instrument malfunction is caused by design defect,
manufacture defect, our company will repair them without payment.

3. Paid service is provided in following condition:

(a) Product out of warranty period.

(b) Damage caused by accident or wrong operation.

(c) Operation is not according to the manual requirement.

(d) Repair the instrument without our permission.

4.Upgrading service is provided along with the technique improvement.

For technique support, contact the following address and telephone:

Manufacturer:Changchun Dirui Industrial CO. LTD.

Address:95,Yunhe Street, New & High Tech. Development Zone, Changchun, China

Sales department telephone:0431-85175023 85172600

After service telephone:8008468578 0431-85184809

Complain telephone: 0431-85177245

Fax:0431-85100405

Zip code:130012

E-mail:dirui@dirui.com.cn

Website: http://www.dirui.com.cn

245
User Manual of CS-400 Auto-Chemistry Analyzer

Addendum B Product Description


B.1 Product assortment

①According to medical equipment product assortment catalogue:

Belong to biochemical analyze system in clinic test analyze instrument (6840), type II in management type.

②According to electric shock protection assortment: I type

B.2 Accessory reagent

Following table explains the accessory reagent (often used biochemical reagent parameter table).

246
User Manual of CS-400 Auto-Chemistry Analyzer

Measurable Reagent
wavelength (nm) volume Blank
Photo Calibr
Item Assay Sample Span Reaction Deviation Discrete Sensitivity horizo
Time Main Sub metric Calibration method ation
name mode volume point direction rate check check check ntal
point point
wavel wavel R1 R2 check
ength ength
0.8-2
ALT Rate A 10 340 405 21-31 15 240 60 2-point linearity 2 2 Negative 3.3 0.05 0
.5
0.8-2
AST Rate A 10 340 405 21-31 15 240 60 2-point linearity 2 2 Negative 3.3 0.05 0
.5
ALP Rate A 10 405 505 21-31 4 200 50 2-point linearity 2 2 Positive 3.3 0.05 0 0-1.0

GGT Rate A 10 405 505 21-31 25 200 50 2-point linearity 2 2 Positive 3.3 0.05 0 0-1.2
2point
TBA rate 10 405 700 22-28 4 270 90 2-point linearity 2 2 Positive 3.3 0.05 0 0-1.2
assay
1point
TBIL 10 546 660 31 20 0 20 2-point linearity 2 2 Positive 3.3 0.05 0 0-0.5
assay
1 point
DBIL 10 546 600 31 20 0 20 2-point linearity 2 2 Positive 3.3 0.05 0 0-0.5
assay
1 point
TP 10 546 700 31 5 250 0 2-point linearity 2 2 Positive 3.3 0.05 0 0-0.5
assay
1 point
ALB 10 600 700 12 2 300 0 2-point linearity 2 2 Positive 3.3 0.05 0 0-0.5
assay
LAP Rate A 10 405 505 21-31 15 240 60 2-point linearity 2 2 Positive 3.3 0.05 0 0-1.2
SCH
Rate A 10 405 505 20-26 3 240 60 2-point linearity 2 2 Positive 3.3 0.05 0 0-3.3
E
ICDH Rate A 10 340 405 21-31 15 240 60 2-point linearity 2 2 Positive 3.3 0.05 0 0-0.8

247
User Manual of CS-400 Auto-Chemistry Analyzer

Reagent
Measurable(nm)
Photom volume Calibr Deviation Discrete Sensit Blank
Item Assay Sample Calibration Span Reaction
Time Main- Sub-w etric ation Rate rate ivity horizontal
name mode volume method point direction
wavel avelen point R1 R2 point check check check check
ength gth
Rate A 2-point
GLDH assay
10 340 405 21-31 15 240 60 2 2 Negative 3.3 0.05 0 0.8-2.5
linearity
Rate A 2-point
AMY assay
10 405 505 21-26 5 160 40 2 2 Positive 3.3 0.05 0 0-1.2
linearity
2-point 2-point
BUN assay
10 340 405 20-26 3 240 60 2 2 Negative 3.3 0.05 0 0.8-2.2
linearity
2-point 2-point
CRE assay
10 546 700 16-31 7.5 240 60 2 2 Positive 3.3 0.05 0 0-0.8
linearity
2-point 2-point
CRE assay
10 505 660 20-26 20 150 150 2 2 Positive 3.3 0.05 0 0-1.2
linearity
2-point 2-point
GLU assay
22 505 660 16-110 2 240 60 2 2 Positive 3.3 0.05 0 0-0.8
linearity
GLU 2-point 2-point
10 340 405 16-31 2 240 60 2 2 Positive 3.3 0.05 0 0-1.0
(HK) assay linearity
2-point 2-point
FMN assay
10 546 700 16-31 15 300 0 2 2 Positive 3.3 0.05 0 0-0.8
linearity
2-point 2-point
TC assay
10 505 660 16-31 3 240 60 2 2 Positive 3.3 0.05 0 0-0.5
linearity
2-point 2-point
TG assay
10 546 700 16-31 3 240 60 2 2 Positive 3.3 0.05 0 0-0.5
linearity
2-point 2-point
HDL-C assay
10 546 660 16-31 3 225 75 linearity 2 2 Positive 3.3 0.05 0 0-0.8

2-point 2-point
LDL-C assay
10 546 660 16-31 4 300 100 2 2 Positive 3.3 0.05 0 0-0.8
linearity
Rate A 2-point
CK-MB assay
10 340 405 25-36 10 200 50 2 2 Positive 3.3 0.05 0 0-0.8
linearity
Rate A 2-point
CK assay
10 340 405 21-31 5 200 50 2 2 Positive 3.3 0.05 0 0-1.2
linearity

248
User Manual of CS-400 Auto-Chemistry Analyzer

Reagent
Wavelength(nm)
Photom volume Calibr Deviati Sensit Blank
Item Assay Sample Calibration Span Reaction Discrete
Time Main-w Sub-w etric ation on rate ivity horizontal
name mode volume method point direction check
avelen avelen point R1 R2 point check check check
gth gth
Rate A 2-point
HBDH assay
10 340 405 21-31 6 240 60 2 2 Negative 3.3 0.05 0 0.8-2.5
linearity
Rate A 2-point
LDH assay
10 340 405 21-31 5 240 60 2 2 Positive 3.3 0.05 0 0-0.8
linearity
2-point 2-point
UA assay
10 546 700 16-31 4 200 50 2 2 Positive 3.3 0.05 0 0-1.5
linearity
1-point 2-point
P assay
10 340 405 10 4 200 0 2 2 Positive 3.3 0.05 0 0-0.8
linearity
2-point 2-point
Ca-CPC assay
10 570 660 16-31 5 150 150 2 2 Positive 3.3 0.05 0 0-0.8
linearity
1-point 2-point
Cl assay
10 505 660 31 3 300 0 2 2 Positive 3.3 0.05 0 0-0.5
linearity
1-point 2-point
Ca-ARS assay
10 660 750 31 3 300 0 2 2 Positive 3.3 0.05 0 0-0.8
linearity
1-point 2-point
Mg assay
10 546 750 10 3 300 0 2 2 Positive 3.3 0.05 0 0-1.2
linearity

Note: For specific parameter, please refer to reagent instruction. In order to add items, please set according to the parameters of reagent instruction. To calculate sample
result, calibration by calibration serum with the practical factor is advised if the items are tested by rate A assay.

249
User Manual of CS-400 Auto-Chemistry Analyzer

Statement
Dirui Co. LTD. has the final power of interpretation.
Dirui Co. LTD. is responsible for the security, reliability and capability of CS-240 B auto-Chemistry analyzer
under following circumstance.
1) Installation, adjustment, improvement and repair are proceeded by Dirui company professionals.
2) Relevant electric equipments are qualified according to state norms.
3) User Manual should be obeyed when operating instrument.

No extra announcement if interface changed.


Any question , please call 8008468578 or 0431-85184809.

250

You might also like