CALIBRATION CERTIFICATE
Page 1 of
CALIBRATION CERTIFICATE OF DIGITAL THERMO-HYGROMETER :
E-mail
Certificate No: ‘22100004764 Issued On: | 15/03/2022
[ULR No: C214622100004761F | Calibration Date: 14/03/2022
Job Identification No: _ | ICS/C/T/03/4761 Next Calibration Date: | 13/03/2023
[Ref. No: | SRF- Dated: 10/03/2022
‘Mis Manjula Laboratories,
CALIBRATED FOR: 13-8, Friends Industrial Estate,
Focal Point, Ludhiana ~ 141070.
a EQUIPMENT DESCRIPTION
Name Digital Thermo-Hygrometer
Make/Model | HTC-1 Visual inspection _| Ok
Serial No. - 1D. No. [-THM-02
[Range ~50°C to 70°C] 10 % to 90% (RH)_| Least Count 0.1°C/1% RH
Location Calibration Site | In-Lab
jact
E ENVIRONMENTAL CONDITIONS
[Temperature 2.0°C#3.0°C | Humidity 50£10%RH
‘STANDARD REFERENCE DETAILS
Traceable to National / International Standard
Name of the fi Calibration Dato of
| standard Make. Certificate No. | agencyiTraceability | Calibration |
Digital Temp: Pal
Indicator with “Lutron” 22100000001 Indian Calibration | oworvzo22 | 01012023
Sensor i
Temp. & Hum. ind. | ; ETL/ATOSRAICIFT sorar021
es Rotronie 506 FRI, Kerala. | S0/"2/202) "| oason/2023
USED EQUIPMENT DETAILS Environmental Chamber
PRINCIPLE/METHODOLOGY OF | As per Calibration Procedure No.: ICS/CALISOP-THOS (ITS-80)
CALIBRATION. (Comparison Method)
RESULTS: Thermal Calibration
(On Working Range) =
i : Corrected Value of Standard | Uncertainty at approx 95% C.L and
Ue, eee (ava.) ‘coverage factor k=2
[Temperature | Humidity [Temperature [Humidity Temperature Humidity
16.0°C 20% RH 14.93°C 20.74 %RH £037°C 16 %RH
40% RH 19.90°C 41.13 %RH £037°C 216% RH
75% RH 29.86°C 76.50 96RH $0.43°C = 1.6% RH
90% RH 39.83°C | 91.88 %RH #0436 | £1.6%RH
Valid up to
REMARKS: The recommended date for next calibration is mentioned, as asked by the customer.
**° END OF REPORT##*
The sabe rob ar atonal wane Cael Lav and oul nd nda Far and on borat
stint ete pr Indian Caliration Services
et ce srg nd aye pater Ensen opis aie erro inpd
"a ty cow eer ime Ube ce ovort
Sard it seo pe am a el arson Cte es dn pee os
ince on oct er Caren sean se Pen ote en Sat=4 Raises!
Tiectuto cease bred eC tne nr st einer, _ Checked by