Professional Documents
Culture Documents
Biomedical Engineering Ktu Mod 2
Biomedical Engineering Ktu Mod 2
MODULE 2
• 120 is known as Systolic blood pressure and 80 is known as Diastolic blood pressure
• Diastolic blood pressure(Lower number): Pressure measured when the ventricles relax
Direct Measurement
Indirect Measurement
• When the needle is in place, a catheter is fed through the hollow needle.
• Catheter is filled with saline heparin solution and inserted into the patient. Heparin is
an anticoagulant. prevents blood clots.
• Catheter is a thin hollow flexible tube. The catheter is connected to a blood pressure
transducer, which generates corresponding electrical signal.
• The electronic pressure transducer is connected to the patient through the catheter.
• The diaphragm senses the pressure of the blood which is transmitted through the
fluid in the catheter
1. Sphygmomanometer 2. Stethoscope.
Sphygmomanometer
• The cuff consists of a rubber bladder inside an inelastic fabric covering that can be
wrapped around the upper arm
• Valve helps in the escape of air and reduction of pressure when required
Method of locating the systolic and diastolic pressure values by listening to the
Korotkoff sounds is called the auscultatory method of sphygmomanometry
Apply the stethoscope lightly over the brachial artery
Inflate the cuff to few mm of Hg above the pressure observed by palpatory method.
Deflate it slowly and listen to the appearance of a knocking sound called korotkoff
sound.
The pressure at which the sound appears is the systolic pressure.
The systolic number represents the pressure that blood exerts against the walls of an
artery following a heart beat or contraction.
The point at which the sound disappears is the diastolic blood pressure
During this method, Sound is heard in 4 different phases
1. Appearance of a knocking sound: Systolic pressure.
2. As it is deflated, gradually sound becomes murmurish
3. Then it becomes loud and clear
4. Then it becomes muffled( almost muted)
5. Finally disappears
6. Disappearance of sound indicates diastolic blood pressure.
3. Oscillometric measurement
As the cuff is deflated beloԝ the systolic pressure, the reducing pressure exerted on the artery
alloԝs blood to floԝ through it and sets up a detectable vibration in the arterial ԝall.
Ԝhen the cuff pressure falls beloԝ the patient’s diastolic pressure, blood floԝs smoothly
through the artery in the usual pulses, ԝithout any vibration being setup in the ԝall.
Vibrations occur at any point ԝhere the cuff pressure is sufficiently high that the blood has to
push the arterial ԝall open in order to floԝ through the artery.
The vibrations are transferred from the arterial ԝall, through the air inside the cuff, into a
transducer in the monitor that converts the measurements into electrical signals.
4. Ultrasonic method
Ultrasonic determination of blood pressure employs a Doppler sensor that detects the motion
of blood vessel walls in various states of occlusion.
Placement of compression cuff over two small transmitting and receiving ultra-sonic crystals.
Cuff pressure is increased above diastolic but below systolic the vessel opens and closes with
each heartbeat.
The Doppler ultra-sonic transmitted signal is focused on the vessel wall and the blood.
The reflected signal with frequency shift is detected by the receiving crystal and decoded.
The difference in frequency, in the range of 40 to 500 Hz between transmitted and received
signals. That frequency shift is proportional to the velocity of wall motion and blood velocity
Blood flow meters are used to monitor the blood flow in various blood vessels and
to measure cardiac output. Two Types are
E=BLV sinθ
Two types:
Blood velocity: v, Transmitted frequency: f1, Reflected frequency: f2, Doppler shift : f2-f1
Angle between ultrasound beam and flow direction: θ, Speed of sound in medium: c
Block diagram explanation
Function generator provides 5 MHz electrical signal for the transducer
Amplifier collects the electrical signal from the receiver and amplifies this
Mixer takes these 2 signals from the transmitter and receiver with frequency f1 and
f2 and outputs 2 signals.:
One with frequency f1-f2
Other with frequency f1+f2
These signals are passed to LPF
It removes the HF signal and keeps the LF signal
This is given to analog discovery kit where the doppler shift is calculated.
Transmitter emits ultrasonic energy which travels across to the receiver on the
opposite side .
These ultrasonic elements are piezoelectric oscillators of the same type as used in
Doppler shift flow meters.
Fluid flowing in the pipe causes a time difference between the transit times of the
beams travelling upstream and downstream
With no flow, the transmit time would be equal in both the upstream and
downstream directions.
With flow, sound will travel faster in the direction of flow and slower against the
flow.
3.3 PLETHYSMOGRAPHY
A plethysmograph is an instrument for measuring changes in volume within an organ or
whole body.
It measures these changes with blood pressure cuffs or other sensors.
1. Impedance Plethysmograph
This is based on the measurement of the electrical impedance (resistance) of a selected body
segment.
For the measurement of the electrical impedance usually 4 electrodes are applied to the body
surface.
The 2 outer electrodes (usually called current electrodes) are used to pass a very low and
constant alternating current through the body segment
The 2 inner electrodes (usually called measuring electrodes) are placed between the 2 current
electrodes and measure the voltage which is caused when the current flows through the body
segment.
This voltage depends on impedance of the body segment which changes depending on
venous and arterial blood volume variations.
The positioning of the measuring electrodes defines the segment which is analysed.
Limb is modelled as a uniform cylinder:
If limb volume increases by ΔV, cross sectional area increases by ΔA, L stays the same.
Impedance equation for blood is,
PPG shows the blood flow changes as a waveform with the help of a bar or a graph.
Heart sounds generally alters during cardiac diseases involving the valves of the heart. That’s
ԝhy heart sounds are having important diagnostic value.
There are three methods to study heart sounds:
1. By using stethoscope
2. By using microphone
3. By using phonocardiogram
3.5 PHONOCARDIOGRAPHY
Phonocardiography is the recording of the all the sounds made by the heart during a cardiac
cycle.
The sounds result from vibrations created by closure of the heart valves.
Two closure sounds:
1) Atrio ventricular valves (Tricuspid & Bicuspid valves).
2) Semilunar valves (Aortic valve & Pulmonary valve).
1. Right arm (RA) 2. Left arm (LA) 3. Left leg (LL) 4. Right leg (RL) 5. Chest(C)