Professional Documents
Culture Documents
Service-Manual ThermoMixer-C F0.5 F1.5 F2.0 FP ThermoStat-C Eng PDF
Service-Manual ThermoMixer-C F0.5 F1.5 F2.0 FP ThermoStat-C Eng PDF
ermoMixer
- English
manual
EN) C/F0.5/F1.5/F2.0/FP/ThermoStat C
- English
Eppendorf ThermoMixer
C/F0.5/F1.5/F2.0/FP/ThermoStat C
Service manual - English
Copyright© 2016 by Eppendorf AG. All rights reserved, including graphics and images. No part of this
publication may be reproduced without the prior permission of the copyright owner.
Eppendorf ThermoMixer® and Eppendorf ThermoTop® are registered trademarks of Eppendorf AG,
Germany.
Eppendorf® and the Eppendorf logo are registered trademarks of Eppendorf AG, Germany.
Registered trademarks and protected trademarks are not marked in all cases with ® or ™ in this manual.
5382 920.017-04/012016
Table of contents
Eppendorf ThermoMixer C/F0.5/F1.5/F2.0/FP/ThermoStat C 3
English (EN)
Table of contents
1 Operating instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.1 Using this manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.2 Danger symbols and danger levels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.2.1 Danger symbols. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.2.2 Danger levels. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.3 Symbols used . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2 Product description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.1 Device overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.2 Assemblies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.2.1 Chassis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.2.2 Upper part of housing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.2.3 Drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.2.4 PCB control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.2.5 PCB motor control. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.2.6 PCB temperature control. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.2.7 Heating/cooling plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.2.8 SmartBlocks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.2.9 Thermoblock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.2.10 Fan for cooling and for the electronic system. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.2.11 Switched-mode power supply. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
2.2.12 USB interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
2.2.13 ThermoTop . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3 Safety. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.1 User profile . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.2 Notes on liability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.3 Maintenance and repair hazards. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
4 Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
5 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
5.1 Error messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
5.2 Service functions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
5.2.1 Overview of service functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
5.2.2 Activating the service functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
5.2.3 Operating the service functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
5.2.4 SE 01 (error memory) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
5.2.5 SE 02 (operating times/runs) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
5.2.6 SE 03 (sensors) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
5.2.7 SE 04 (key test) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
5.2.8 SE 05 (motor) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
5.2.9 SE 06 (fan). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
5.2.10 SE 07 (memory reset) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Table of contents
4 Eppendorf ThermoMixer C/F0.5/F1.5/F2.0/FP/ThermoStat C
English (EN)
6 Disassembly/assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
6.1 Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
6.2 Separating the upper part of housing and the chassis (disassembling heating/cooling plate and
thermoblock)25
6.2.1 Eppendorf ThermoMixer C and Eppendorf ThermoStat C . . . . . . . . . . . . . . . . . . . . . . 25
6.2.2 Eppendorf ThermoMixer F0.5/F1.5/F2.0/FP. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
6.3 Disassembly of upper part of housing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
6.3.1 PCB control (incl. display). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
6.3.2 Switched-mode power supply, USB interface and fan for switched-mode power supply
31
6.3.3 Keypad and front clip . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
6.3.4 Centering pins (including PCB TT power) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
6.4 Disassembly from the chassis Eppendorf ThermoMixer C and Eppendorf ThermoMixer Fxx . 33
6.4.1 Orbital plate. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
6.4.2 Imbalance eccentric . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
6.4.3 Motor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
6.4.4 PCB motor control. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
6.4.5 Stabilizer links . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
6.4.6 Fan for cooling (Eppendorf ThermoMixer C only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
6.4.7 Chassis closing (Eppendorf ThermoMixer Fxx only) . . . . . . . . . . . . . . . . . . . . . . . . . . 39
6.5 Dismantling the Eppendorf ThermoStat C chassis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
6.5.1 PCB temperature control. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
6.5.2 Fan for cooling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
6.6 Device assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
7 Alignment/adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
7.1 Setting and adjusting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
7.1.1 Setting the electronics. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
7.1.2 Temperature adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
7.1.3 Mechanical adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
8 Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
8.1 Software update . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
8.2 Bootloader. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
9 Diagrams . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
9.1 Eppendorf ThermoMixer C . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
9.2 Eppendorf ThermoStat C . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
9.3 Eppendorf ThermoMixer F0.5/F1.5/F2.0 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
9.4 Eppendorf ThermoMixer FP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Table of contents
Eppendorf ThermoMixer C/F0.5/F1.5/F2.0/FP/ThermoStat C 5
English (EN)
10 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
10.1 Cleaning, disinfection, decontamination. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
10.2 Shipping the device. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
10.3 Service and inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
10.3.1 Essential check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
10.3.2 Advanced maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
10.3.3 Qualification services - operational qualification (OQ) . . . . . . . . . . . . . . . . . . . . . . . . . 54
10.4 Device inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
10.4.1 Temperature check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
10.4.2 Speed check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
10.4.3 Timer check. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
10.5 Electrical safety check. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
11 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
11.1 Power supply. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
11.2 Weight/dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
11.3 Ambient conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
11.4 Tolerances . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
11.4.1 Temperature accuracy. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
11.4.2 Temperature homogeneity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
11.4.3 Speed. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
11.4.4 Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
12 Ordering information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
12.1 Figures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
12.2 Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
12.2.1 Housing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
12.2.2 Drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
12.2.3 Electronics. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
12.2.4 Fan. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
12.2.5 Accessories/auxiliary aids . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
12.2.6 SmartBlocks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
12.3 Spare parts proposals – initial equipment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
13 Technical information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
Table of contents
6 Eppendorf ThermoMixer C/F0.5/F1.5/F2.0/FP/ThermoStat C
English (EN)
Operating instructions
Eppendorf ThermoMixer C/F0.5/F1.5/F2.0/FP/ThermoStat C 7
English (EN)
1 Operating instructions
1.1 Using this manual
Make sure that the service manual and the operating manual are available in the latest versions. To do
so, compare the version numbers. The latest versions of the manuals can be downloaded from
www.eppendorf-support.com and www.eppendorf.com.
Read the service manual before commencing work on the device.
Read the chapters "Installation" and "Operation" of the operating manual.
Observe the safety instructions in the operating manual.
The safety precautions in these instructions have the following danger symbols and danger levels.
Material damage
Depiction Meaning
1. Actions in the specified order
2.
Actions without a specified order
• List
Text Display text or software text
Additional information
Operating instructions
8 Eppendorf ThermoMixer C/F0.5/F1.5/F2.0/FP/ThermoStat C
English (EN)
Product description
Eppendorf ThermoMixer C/F0.5/F1.5/F2.0/FP/ThermoStat C 9
English (EN)
2 Product description
2.1 Device overview
Eppendorf ThermoMixer Fxx and TMFxx: abbreviation for Eppendorf ThermoMixer F0.5/
F1.5/F2.0/FP.
2.2 Assemblies
2.2.1 Chassis
All devices have the same metal chassis. It holds the drive and the fans (including the air duct) for cooling.
The Eppendorf ThermoStat C has an enlarged heat sink installed instead of the drive.
All devices have the same plastic upper part of housing. The device name is printed on an exchangeable
clip. The clip is inserted in the front of the housing. The upper part of housing and the clip are connected to
each other by the device-specific adhesive keypad.
The upper part of housing contains the PCB control, the mains input (switched-mode power supply), the
fan for the switched-mode power supply, the USB interface and the contact to the ThermoTop.
Product description
10 Eppendorf ThermoMixer C/F0.5/F1.5/F2.0/FP/ThermoStat C
English (EN)
2.2.3 Drive
The rotary movement of the motor is translated onto a circular orbit via an eccentric. A counterweight
compensates the resulting imbalance and ensures the stability of the device. The weight of the accessories
and their load is held by two stabilizer links. The stabilizer links are mounted on both sides of the motor.
The motor is controlled using the PCB motor control. A Hall sensor in the motor detects the speed and
reports it to the PCB control for checking.
The PCB control is the central control unit of the devices containing the firmware. All internal operating
modes are controlled and monitored here. The commands that are entered via the keypad are recorded and
forwarded to the corresponding assemblies (e.g., PCB motor control, PCB temperature control). All sensor
feedback signals are recorded and processed, e.g., speed and temperature. The firmware responds
independently to the values received and makes adjustments, if required.
The three thermal modules (heating foil, Peltier module, ThermoTop) are activated sequentially by a
multiplexer. In this way, each thermal module can be supplied with maximum power.
The PCB control also includes the display on which all information can be displayed that is relevant to the
operation.
The switched-mode power supply supplies the PCB control with 24V. The other voltages (5 V, 12 V, etc.)
are generated directly on the PCB control from the 24 V.
There are always two temperature sensors each. The two values that are reported back are constantly
compared with each other. If the deviation is too great, the unit concerned is switched off and an error
message appears on the display. The motor speed is monitored by a Hall sensor. If the deviation from the
set speed is too great, the motor switches off and a corresponding error message appears on the display. In
addition, the current flow to the heating and the motor is monitored constantly. If the maximum value is
exceeded, the corresponding unit is deactivated and a corresponding error message also appears on the
display. Fuses are used as a further safety measure, they interrupt the circuits irreversibly, if required (the
fuses cannot be replaced).
All devices have the same PCB control with the same firmware. The firmware recognizes the device model
via the connected keypad that is unique to the device.
The PCB motor control controls the motor. The parameters for controlling the motor of the device model
are stored on the PCB motor control. The heat sink fan is connected to the PCB motor control.
The PCB temperature control is only used with the Eppendorf ThermoStat C. The two heat sink fans are
connected to the PCB temperature control.
Product description
Eppendorf ThermoMixer C/F0.5/F1.5/F2.0/FP/ThermoStat C 11
English (EN)
The heating/cooling plate includes a heating foil for heating and a Peltier module for cooling. An integrated
Hall sensor checks whether a SmartBlock has been mounted. The range of the Hall sensor has been
restricted so that the device can only be started if the SmartBlock has been mounted correctly.
2.2.8 SmartBlocks
The SmartBlocks have been actively integrated in the temperature control process. The SmartBlocks
contain the temperature sensors for temperature control and a memory which includes all the values that
are relevant for controlling the temperature. Each SmartBlock is individually adjusted during manufacture.
The SmartBlocks communicate with the device in a non-contact manner, via an optical-inductive interface
which is located on the bottom (front right).
The respective maximum speed of the SmartBlock is also stored in the SmartBlock and communicated to
the device via the optical inductive interface.
Power for the sensors and the memory is also being supplied in a non-contact manner via the
optical-inductive interface.
The SmartBlock plates has a light barrier on the right-hand side which is used to detect the height of the
loading used. If it has been loaded too high, the maximum speed is reduced.
2.2.9 Thermoblock
The Eppendorf ThermoMixer Fxx thermoblocks are tempered with a heating foil. The temperature sensors
and a memory with the parameters that are relevant for temperature and speed control are integrated into
the heating foil of the thermoblocks. They are individually adjusted during manufacture.
The thermoblock of the Eppendorf ThermoMixer FP has a light barrier on the right-hand side which is used
to detect the height of the loading used. If it has been loaded too high, the maximum speed is reduced.
The heatsink fans have a variable speed control which depends on the temperature of the heatsinks
(determined indirectly via the Peltier voltage) and the set target temperature. The fan for the
switched-mode power supply has a speed control which depends on the output (it is switched off < 50 W).
Product description
12 Eppendorf ThermoMixer C/F0.5/F1.5/F2.0/FP/ThermoStat C
English (EN)
The devices are supplied with power via a switched-mode power supply which outputs the internal supply
voltage (24 V) to the PCB control.
A USB interface is located at the rear. The firmware of the devices can be updated via this interface and an
external laptop/PC.
2.2.13 ThermoTop
The ThermoTop can be used with some SmartBlocks to prevent condensation. It holds a temperature which
is approx. 3 to 18 Kelvin above the set block temperature. Due to differences in height between some of the
SmartBlocks/thermoblocks the exact temperature depends on which SmartBlock/thermoblock is used. The
temperature is controlled via two integrated temperature sensors which communicate with the device via
an optical-inductive interface.
The heating foil is supplied with power via two centering pins on the upper part of the device housing. For
safety reasons, the current flow via the centering pins is only active when the ThermoTop has been
attached (however, a centering pin still has a variable potential!).
The centering pins and some magnets fix the ThermoTop at the rear. At the front, the ThermoTop is fixed
by two magnets.
Safety
Eppendorf ThermoMixer C/F0.5/F1.5/F2.0/FP/ThermoStat C 13
English (EN)
3 Safety
3.1 User profile
The specialist entrusted with maintenance, repair or testing must meet the following prerequisites:
• Successful participation in service trainings with certification by Eppendorf AG for the product that is to
be maintained, repaired or tested.
• Qualification as a specialist with knowledge of the applicable local and international standards.
• Qualification to evaluate the work assigned to him.
• Recognition of potential hazards and how to prevent or correct them.
The Eppendorf service organization entrusted with maintenance, servicing or testing is liable for ensuring
that all work is carried out in a professional manner.
Repair, maintenance and testing may not change the existing construction of the device. The device
must always be as safe as it is in its original state.
Only accessories and original spare parts recommended by Eppendorf, and measuring and test
equipment recommended by Eppendorf may be used for any maintenance, servicing or testing work.
Safety
14 Eppendorf ThermoMixer C/F0.5/F1.5/F2.0/FP/ThermoStat C
English (EN)
Do not operate the device in areas where work is completed with explosive substances.
Do not use this device to process any explosive or highly reactive substances.
Do not use this device to process any substances which could create an explosive
atmosphere.
4 Operation
Operation
16 Eppendorf ThermoMixer C/F0.5/F1.5/F2.0/FP/ThermoStat C
English (EN)
Troubleshooting
Eppendorf ThermoMixer C/F0.5/F1.5/F2.0/FP/ThermoStat C 17
English (EN)
5 Troubleshooting
5.1 Error messages
The error messages will be displayed in the XX.YYY format:
For the error analysis you need the digits YYY. The digits represent the error numbers
listed below.
All control commands, sensor signals and error signals are processed in the PCB Control.
Therefore the PCB Control and its supply lines such as cables and plug connections can
generally by the cause of every error.
• The operation of the service functions is the same as the menu operation in the normal operating mode.
• The service functions can be selected using the up/down arrow keys.
• Operate the menu/enter key to activate a service function or subfunction.
• To exit the service function or a subfunction, select the back menu item and operate the menu/enter
key.
Alternatively, for the Eppendorf ThermoMixer C and Eppendorf ThermoStat C, the left arrow
key can be used to exit the service function or subfunction, regardless of the menu item
selected.
Troubleshooting
Eppendorf ThermoMixer C/F0.5/F1.5/F2.0/FP/ThermoStat C 23
English (EN)
5.2.6 SE 03 (sensors)
5.2.8 SE 05 (motor)
Drive test
Display of the set speed and real speed.
Key functions:
• speed up/down: setting the speed.
• start/stop key: starting the drive.
Troubleshooting
24 Eppendorf ThermoMixer C/F0.5/F1.5/F2.0/FP/ThermoStat C
English (EN)
5.2.9 SE 06 (fan)
Fan test
Fan:
• Power supply fan: max. 3 100 rpm (all devices).
• Primary heat sink fan: max. 5 600 rpm
(Eppendorf ThermoMixer C and Eppendorf
ThermoStat C only).
• Secondary heat sink fan: max. 5 600 rpm
(Eppendorf ThermoStat C only).
Key functions:
• up/down menu arrow keys: selecting the fan.
• speed up/down: setting the speed.
• start/stop key: starting the fan.
6 Disassembly/assembly
6.1 Tools
6.2 Separating the upper part of housing and the chassis (disassembling
heating/cooling plate and thermoblock)
6.2.1 Eppendorf ThermoMixer C and Eppendorf ThermoStat C
Prerequisites
The SmartBlock has been removed.
Abb. 6-1: Heating/cooling plate with screws/housing screws Bottom of the device
3 3
Fig. 6-1: Heating/cooling plate with screws/housing screws Bottom of the device
The strain relief consists of two parts. When fitting a new heating/cooling plate, the cable
conduit must be broken off at the predetermined breaking point before assembly. When
the strain relief has been positioned and the ribbon cable has been connected, the cable
conduit must be placed on the strain relief pin and engaged (see figures below).
Push the strain relief fully back into the housing gap. It must fully rest on the front stop
(see figure below).
Make sure that the cables are positioned correctly, in particular, the ribbon cable to the
heating/cooling plate. They may not be jammed or damaged.
Pay attention to the ground contact at the front of the chassis. It must be pushed onto the
inside of the device housing (see figure below).
Disassembly/assembly
Eppendorf ThermoMixer C/F0.5/F1.5/F2.0/FP/ThermoStat C 27
English (EN)
Abb. 6-2: Strain relief with predetermined breaking point/positioning of the cable conduit
2 2 3
Fig. 6-2: Strain relief with predetermined breaking point/positioning of the cable conduit
3 3
37 °C 42 °C 56 °C 95 °C temp off
4
Assembly notes:
The strain relief consists of two parts. When fitting a new thermoblock, the cable conduit
must be broken off at the predetermined breaking points before installation. When the
strain relief has been positioned and the ribbon cable has been connected, the cable
conduit must be placed on the strain relief pin and engaged (Fig. 6-3 on p. 27).
Push the strain relief fully back into the housing gap. It must rest on the front stop.
Make sure that the cables are positioned correctly, particularly the ribbon cable to the
thermoblock. They may not be jammed or damaged.
Pay attention to the ground contact at the front of the chassis. It must be pushed onto the
inside of the upper part of housing (Fig. 6-4 on p. 27).
The plastic plate on the ribbon cable from the thermoblock must be positioned in the
thermoblock adapter guide (see figure below).
Abb. 6-6: Guide for ribbon cable
5 °C
temp
off
Prerequisites
The upper part of housing has been separated from the chassis.
5382 100.
28
PC/ABS
2 4
1 PCB control with display 3 Housing insert for IR interface with mounting
screws (2 x Phillips EJ 2.5 x 8)
2 Mounting screws (EJ 3 x 8/T10)
4 O-ring
1. Disconnect the plug connections of the cables (excluding the ribbon cable to the display).
2. Unscrew the mounting screws.
3. Remove the PCB control from the upper part of housing.
Note the IR-interface. It is firmly inserted in the upper part of housing.
The ribbon cable to the IR interface may tear if you pull too hard!
Carefully pull out the IR-interface together with the housing insert.
Assembly note:
Make sure that the IR-interface and the sealing ring are positioned correctly.
An initialization ("Factory Reset") must be performed after the PCB control has been replaced!
Disassembly/assembly
Eppendorf ThermoMixer C/F0.5/F1.5/F2.0/FP/ThermoStat C 31
English (EN)
6.3.2 Switched-mode power supply, USB interface and fan for switched-mode power
supply
Prerequisites
The upper part of housing has been separated from the chassis.
Abb. 6-8: Switched-mode power supply, USB interface and fan for switched-mode power supply
3
6
4 4
5
1 2
Fig. 6-8: Switched-mode power supply, USB interface and fan for switched-mode power supply
Prerequisites
The upper part of housing has been separated from the chassis.
Assembly note:
Clean the bonding surface thoroughly before mounting the new keypad.
Disassembly/assembly
32 Eppendorf ThermoMixer C/F0.5/F1.5/F2.0/FP/ThermoStat C
English (EN)
Prerequisites
The upper part of housing has been separated from the chassis.
Abb. 6-9: Centering pins and PCB TT power in the upper part of housing
1 2
Fig. 6-9: Centering pins and PCB TT power in the upper part of housing
2
1
3 4
5
5
3 Screws for upper locking ring (M4 x 10/T20) 5 Screws for orbital plate (M4 x 12/T20)
4 Orbital plate
Disassembly/assembly
34 Eppendorf ThermoMixer C/F0.5/F1.5/F2.0/FP/ThermoStat C
English (EN)
1. Only Eppendorf ThermoMixer Fxx: Undo the thermoblock adapter screws and remove the thermoblock
adapter.
2. Loosen the screws on the upper locking ring (only loosen, do not remove!).
3. Remove the screws from the ground cable and the orbital plate and lift off the orbital plate.
Assembly notes:
During assembly, tighten the locking ring screws last and evenly (in sequence). The
locking ring must not tilt.
Do not overtighten the screws for the locking ring and the orbital plate, as this could
damage the threads in the locking ring or in the plastic of the stabilizer links.
Prerequisites
The upper part of housing and the orbital plate have been removed.
1. Loosen the screws on the lower locking ring (only loosen, do not remove!).
2. Loosen the clamping screw of the imbalance eccentric slightly (1 turn max.).
3. Lift off the imbalance eccentric. Note: The bottom ball bearing of the imbalance eccentric is glued to the
motor casing (Fig. 6-13 on p. 36). If the imbalance eccentric is seated too firmly, it can be levered up
carefully and smoothly with two screwdrivers that are applied opposite each other.
Disassembly/assembly
Eppendorf ThermoMixer C/F0.5/F1.5/F2.0/FP/ThermoStat C 35
English (EN)
Assembly notes:
Apply a small amount of Loctite® 603™ around the bottom ball bearing of the imbalance
eccentric. Remove any glue residue first.
During assembly, tighten the locking ring screws evenly (in sequence). The locking ring
must not tilt.
Do not overtighten the locking ring screws. This could damage the thread in the locking
ring.
Tighten the clamping screw of the imbalance eccentric last.
Disassembly/assembly
36 Eppendorf ThermoMixer C/F0.5/F1.5/F2.0/FP/ThermoStat C
English (EN)
6.4.3 Motor
Prerequisites
The upper part of housing, the orbital plate and the imbalance eccentric have been removed.
2
2
6 6
Prerequisites
The upper part of housing, the orbital plate, the imbalance eccentric and the motor have been removed.
Abb. 6-15: PCB motor control in the Eppendorf ThermoMixer C
Pay attention to the ribbon cable and how it is laid! During assembly, it must be laid in the
same position again.
1. Eppendorf ThermoMixer C: Pull the PCB motor control out of the air duct (clamped).
2. Eppendorf ThermoMixer Fxx: Remove the retaining plate from the ribbon cable (Fig. 6-17 on p. 39) and
remove the PCB motor control.
On the Eppendorf ThermoMixer Fxx, the PCB motor control can also be removed without
disassembling the orbital plate, the imbalance eccentric and the motor (see Fig. 6-17 on
p. 39).
Pull off the circlips and remove the PCB motor control together with the cover.
Disassembly/assembly
38 Eppendorf ThermoMixer C/F0.5/F1.5/F2.0/FP/ThermoStat C
English (EN)
Prerequisites
The upper part of housing and the orbital plate have been removed.
3 2 1 3
3 1 2 3
Prerequisites
The upper part of housing, the orbital plate, the imbalance eccentric, the motor and the PCB motor control
have been removed.
Push the air duct and the fan downward out of the base plate.
Disassembly/assembly
Eppendorf ThermoMixer C/F0.5/F1.5/F2.0/FP/ThermoStat C 39
English (EN)
Prerequisites
The upper part of housing has been removed.
1 3 2 5
Pay attention to the ribbon cable and how it is laid! During assembly, it must be laid in the
same position again.
1. Undo the retaining plate screws (from the base, below the device feet).
2. Remove the retaining plate.
3. Remove the two circlips on the PCB motor control.
4. Pull the PCB motor control off with the cover.
5. Press the chassis closing downwards.
Disassembly/assembly
40 Eppendorf ThermoMixer C/F0.5/F1.5/F2.0/FP/ThermoStat C
English (EN)
Prerequisites
• The upper part of housing has been removed.
Abb. 6-18: PCB temperature control
1 2 3
Pay attention to the ribbon cable and how it is laid! During assembly, it must be laid in the
same position again.
Prerequisites
• The upper part of housing has been removed.
• The PCB temperature control has been removed.
Abb. 6-19: Air duct
1 2
With all assembly steps, install in reverse order. Also observe chapter 7 "Alignment/adjustment" and the
notes in the disassembly instructions during assembly.
Disassembly/assembly
42 Eppendorf ThermoMixer C/F0.5/F1.5/F2.0/FP/ThermoStat C
English (EN)
Alignment/adjustment
Eppendorf ThermoMixer C/F0.5/F1.5/F2.0/FP/ThermoStat C 43
English (EN)
7 Alignment/adjustment
7.1 Setting and adjusting
7.1.1 Setting the electronics
After replacing the PCB control you have to perform an initialization (factory reset) (see
p. 24).
The SmartBlocks and thermoblocks are equipped with correction data during production. The correction
data cannot be changed later.
This also applies for spare parts. The temperature cannot and does not need to be adjusted after replacing
SmartBlocks and thermoblocks.
During assembly, make sure that the cables are positioned correctly, that the movable parts and assemblies
can move freely and, generally, that the individual parts and assemblies fit together.
8 Software
8.1 Software update
Printed circuit boards do not always have the latest software or firmware versions installed.
Check the software and firmware versions after replacing the PCBs.
Update the software and firmware if necessary.
You can update the device software with a computer via the USB interface of the device.
To enable communication between the devices and the computer you need to install the "Universal Device
Firmware Upgrade (UDFU)" software (version 1.1.7.0 or higher), the required drivers and Microsoft.NET
Framework on your computer.
The UDFU software with drivers and operating manual is available for download on
www.eppendorf-support.com (password required).
Read the UDFU software operating manual and the firmware installation instructions before
starting to download, install or update the software or firmware.
8.2 Bootloader
A faulty update or an interruption during the update can damage the firmware. It is possible that the device
can then no longer be operated and the display may be blank. In this case, the device can be switched to
bootloader mode and the firmware can be installed again.
9 Diagrams
9.1 Eppendorf ThermoMixer C
Diagrams
48 Eppendorf ThermoMixer C/F0.5/F1.5/F2.0/FP/ThermoStat C
English (EN)
≤
Diagrams
Eppendorf ThermoMixer C/F0.5/F1.5/F2.0/FP/ThermoStat C 49
English (EN)
10 Maintenance
10.1 Cleaning, disinfection, decontamination
Switch off the device and disconnect the mains/power plug before starting cleaning or
disinfection work.
Do not allow any liquids to penetrate the inside of the housing.
Use closed tubes and closed plates.
Do not spray clean/spray disinfect the housing.
Only plug the device back in if it is completely dry, both inside and outside.
Do not use any aggressive chemicals on the device or its accessories, such as strong and
weak bases, strong acids, acetone, formaldehyde, halogenated hydrocarbons or phenol.
If the device becomes contaminated with aggressive chemicals, clean it immediately using
a mild cleaning agent.
The device may only be stored and transported in its original packing.
Transport packing can be ordered via your Eppendorf service organization. The order number
can be found in the chapter Ordering information.
Visual check of exterior. Check all external parts for visible damages, e.g.
housing, chassis, device feet.
Visual check of accessories (if available). Check the exterior of the accessories for visible
• SmartBlocks damages.
• ThermoTop
• Lid
Check of SmartBlock detection (TMC and TSC). Attach each available SmartBlock one after the
other. Check whether all SmartBlocks are
correctly detected and indicated on the display.
Functional check (without measurement) Mixing: Set various arbitrary speeds and start
• Mixing (TMC and TMFxx) the drive. Check whether the device runs
• Heating (TMC, TMFxx and TSC) correctly and without unusual noises.
• Cooling (TMC and TSC). Heating: Set an arbitrary temperature higher
than the ambient temperature. Check whether
the temperature of the SmartBlock or the
Thermoblock is rising. Risk of burns: Do not set
the temperature too high!
Cooling: Set an arbitrary temperature lower than
the ambient temperature. Check whether the
temperature of the SmartBlock is reducing.
Maintenance
Eppendorf ThermoMixer C/F0.5/F1.5/F2.0/FP/ThermoStat C 53
English (EN)
Functional check with service functions: SE 03: Attach all available SmartBlocks. Check
• SE 03 (sensors) temperature sensors in all SmartBlocks and the
• SE 04 (key test) height sensor of the SmartBlock plates if
• SE 05 (motor) necessary.
• SE 06 (fan) SE 03: Attach ThermoTop (if available). Check
the temperature sensors in the ThermoTop.
SE 04: Actuate all buttons one after the other.
Check that all buttons are correctly displayed.
SE 05: Start the motor at different speeds. Check
that all selected speeds are attained properly.
SE 06: Start all fans at different speeds one after
the other. Check that all selected speeds are
attained properly.
Check of SmartBlock detection (TMC and TSC). Attach each available SmartBlock one after the
other. Check whether all SmartBlocks are
correctly detected and indicated on the display.
Functional check (without measurement) Mixing: Set various arbitrary speeds and start
• Mixing (TMC and TMFxx) the drive. Check whether the device runs
• Heating (TMC, TMFxx and TSC) correctly and without unusual noises.
• Cooling (TMC and TSC). Heating: Set an arbitrary temperature higher
than the ambient temperature. Check whether
the temperature of the SmartBlock or the
Thermoblock is rising. Risk of burns: Do not set
the temperature too high!
Cooling: Set an arbitrary temperature lower than
the ambient temperature. Check whether the
temperature of the SmartBlock is reducing.
Check of ThermoTop detection and function (only if Attach the ThermoTop and set an arbitrary
the customer uses the ThermoTop). temperature higher than the ambient
temperature. Check whether the ThermoTop is
detected (symbol on the display and LED on the
ThermoTop) and if the temperature of the
ThermoTop increases. Risk of burns: Do not set
the temperature too high!
Eppendorf recommends checking the essential technical data and functions of the devices at least once per
year in accordance with the OQ checklist.
Detailed descriptions of how to perform the individual measurements can be found in the "Device
inspection" chapter.
The applicable tolerances and limiting values can be found in the "Technical data" chapter.
Perform the device test at an ambient temperature of 22°C ± 2°C.
1 2
A 3 4 5 6
B
Fig. 10-1: Test bore for temperature measurement, example thermoblock 1.5 mL
The temperature of the thermoblock is checked on all devices and SmartBlocks with the "Temperature
Verification System - single-channel" and the "Temperature sensor 384". All thermoblocks have a test bore
that matches the sensor.
Measuring device
• Temperature verification system - single-channel (0055 000.298 / 0056 000.003) with temperature
sensor 384 (5331 233.007 / 0056 002.006)
Measurement procedure:
1. Mount a SmartBlock (Eppendorf ThermoMixer C and Eppendorf ThermoStat C only).
2. Remove the stopper of the test bore.
3. Position the temperature sensor in the SmartBlock or thermoblock.
4. Equip SmartBlock or thermoblock with vessels.
5. Set the test temperature.
6. Wait until the test temperature has been reached and is stable.
7. Note the measuring result.
8. Repeat steps 4 to 6 for other temperatures.
Maintenance
56 Eppendorf ThermoMixer C/F0.5/F1.5/F2.0/FP/ThermoStat C
English (EN)
Measuring device
• Digital rotation meter, suitable for non-contact measurement.
• Measuring range: 1 to ≥ 9,999 rpm
• Accuracy: ≤ 0.1 %
Measurement procedure:
1. Mount a SmartBlock (Eppendorf ThermoMixer C only).
2. Mark the SmartBlock or thermoblock using a reflector (accessory for the measuring device).
3. Start the Eppendorf ThermoMixer with the desired rotational speed. Let the Eppendorf ThermoMixer
run briefly until the speed has stabilized.
4. Measure the speed (as described in the operating manual of the measuring device).
5. Note the measuring result.
6. If required, repeat steps 3 to 5 for other rotational speeds.
Measuring device
• Stopwatch
• Accuracy: ≤ 0.1 s
Measurement procedure:
1. Set a run time of > 5 minutes.
2. Start the device and the stopwatch simultaneously.
3. Let the device run for 5 minutes.
4. Compare the time displayed on the device and the time measured with the stopwatch.
Effective standard
• Pre-IEC 62638 (international)
• VDE 701/702 (Germany)
• NFPA 99 (USA)
11 Technical data
11.1 Power supply
11.2 Weight/dimensions
11.4 Tolerances
11.4.1 Temperature accuracy
The accuracy of the temperature verification system - single-channel with the temperature
sensor 384 (± 0,3 °C) has been taken into consideration in the tolerance data below. Therefore,
this data is different to the technical data in the operating manual.
11.4.3 Speed
Parameters Accuracy
Speed ±3%
11.4.4 Time
Parameters Accuracy
Timer 5 min ± 2 s
Technical data
60 Eppendorf ThermoMixer C/F0.5/F1.5/F2.0/FP/ThermoStat C
English (EN)
Ordering information
Eppendorf ThermoMixer C/F0.5/F1.5/F2.0/FP/ThermoStat C 61
English (EN)
12 Ordering information
12.1 Figures
Abb. 12-1: Front view
prog 5
prog 4
prog 3
prog 2
prog 1
short
menu
enter speed
temp
time
start
stop
1-2
2-3
4-1 2-4 2-2
2-2 2-1
3-5 3-2
5-2
4-3
1-5
1-1
3-1
1-4
1-3
4-1
Abb. 12-5: PCB motor control and fan for cooling (including air duct)
3-4 4-1
Fig. 12-5: PCB motor control and fan for cooling (including air duct)
Ordering information
66 Eppendorf ThermoMixer C/F0.5/F1.5/F2.0/FP/ThermoStat C
English (EN)
3-10
3-4
4-4
Product numbers
• 5382: Eppendorf ThermoMixer C (TMC)
• 5383: Eppendorf ThermoStat C (TSC)
• 5384: Eppendorf ThermoMixer F1.5 (TMF1.5)
• 5385: Eppendorf ThermoMixer FP (TMFP)
• 5386: Eppendorf ThermoMixer F0.5 (TMF0.5)
• 5387: Eppendorf ThermoMixer F2.0 (TMF2.0)
12.2.1 Housing
5382
5383
5384
5385
5386
5387
(International)
Upper part of housing
w/o keypad
12-1 / 12-3 1-1 5382 850.019 TMC/TSC/TMF X X X X X X
Keypad
incl. Frontclip
12-1 1-2 5382 850.060 TMC X
12-1 1-2 5383 850.063 TSC X
12-1 1-2 5384 850.067 TMF1.5 X
12-1 1-2 5385 850.060 TMFP X
12-1 1-2 5386 850.064 TMF0.5 X
12-1 1-2 5387 850.068 TMF2.0 X
Spherical foot
12-4 1-3 5355 863.001 4 pieces X X X X X X
Spherical support
12-4 1-4 5353 861.001 4 pieces X X X X X X
Centering pin
12-1 / 12-3 1-5 5382 850.035 2 pcs., incl. PCB TT-Power X X X X X X
12.2.2 Drive
(International)
Motor
12-2 2-1 5382 850.108 TMC/TMF X X X X X
Stabilizer link
set = 2 pieces
12-2 2-2 5382 850.205 TMC/TMF X X X X X
Imbalance eccentric
incl. ball bearing
12-2 2-3 5382 850.302 TMC X
12-2 2-4 5384 850.300 TMF X X X X
Locking ring 2 pcs
12-10 2-6 5382 850.400 TMC/TMF X X X X X
Ordering information
72 Eppendorf ThermoMixer C/F0.5/F1.5/F2.0/FP/ThermoStat C
English (EN)
12.2.3 Electronics
In some cases, the software/firmware version of the PCB(s) does not match the current status.
After replacing the printed circuit board(s) check the software version and update it, if
required.
5382
5383
5384
5385
5386
5387
(International)
PCB Control
with display
12-3 3-1 5382 850.507 TMC/TSC/TMF X X X X X X
Switched-mode power supply
12-3 3-2 5382 850.604 incl. mains input module X X X X X X
PCB temperature control
12-7 3-3 5383 850.705 TSC X
PCB motor control
12-5 / 12-6 3-4 5382 850.701 TMC X
12-5 / 12-6 3-4 5384 850.105 TMF X X X X
USB interface
12-3 3-5 5382 850.906 TMC/TSC/TMF X X X X X X
Heating/cooling plate
12-1 3-6 5382 851.007 TMC X
12-7 3-7 5383 851.000 TSC X
Thermoblock
12-8 3-8 5384 851.004 TMF1.5 X
12-9 3-9 5385 851.008 TMFP X
(12-8) 3-11 5386 851.001 TMF0.5 X
(12-8) 3-12 5387 851.005 TMF2.0 X
Adapter for thermoblock
12-6 3-10 5384 851.209 TMF X X X X
12.2.4 Fan
(International)
Fan
incl. air duct
12-2 / 12-4 / 4-1 5382 851.104 TMC
12-5 X
12-7 4-2 5383 851.108 TSC X
Fan
for switched-mode power supply
12-3 4-3 5382 851.155 TMC/TSC/TMF X X X X X X
Chassis closing
12-6 4-4 5384 851.101 TMF X X X X
Ordering information
Eppendorf ThermoMixer C/F0.5/F1.5/F2.0/FP/ThermoStat C 73
English (EN)
5382
5383
5384
5385
5386
5387
(International)
Packing
---- 5-1 5382 851.406 TMC/TSC/TMF X X X X X X
Mounting elements
incl. cable
(12-3) 5-2 5382 851.201 TMC/TSC/TMF X X X X X X
Temperature Verification System USB –
Single channel
---- 5-3 0056 000.003 For Mastercycler nexus, Mastercycler pro X X X X X X
und Mastercycler ep, Eppendorf
ThermoMixer, Eppendorf ThermoStat
Temperature sensor
for Temperature Verification System
USB – Single channel
---- 5-4 0056 002.006 384 Well X X X X X X
USB software cable
---- 5-5 5428 850.280 USB-A/USB-B X X X X X X
12.2.6 SmartBlocks
5382
5383
5384
5385
5386
5387
(International)
SmartBlock cryo
Thermoblock for 24 Cryo tubes
---- 6-10 5367 000.025 1.5 mL – 2 mL, diameter max. 12,5 mm,
all base shapes X X
SmartBlock plates
Thermoblock for microplates and
deepwell plates
---- 6-11 5363 000.039 incl. Lid X X
SmartBlock PCR 96
Thermoblock for PCR plates 96
---- 6-12 5306 000.006 incl. Lid X X
SmartBlock PCR 384
Thermoblock for PCR plates 384
---- 6-13 5307 000.000 incl. Lid X X
Transfer Rack 1.5/2.0 mL
---- 6-14 3880 000.151 for 1.5/2.0 mL tubes X X X X
Transfer Rack 0.5 mL
---- 6-15 3880 000.305 for 0.5 mL tubes X X X
13 Technical information
Technical information
76 Eppendorf ThermoMixer C/F0.5/F1.5/F2.0/FP/ThermoStat C
English (EN)
Evaluate Your Manual