Download as pdf or txt
Download as pdf or txt
You are on page 1of 11

GE Healthcare No.

10 - 2007
VividClub

Application
News
GE Healthcare
VividClub

Configure the measurements to your needs

Vivid 7
Vivid i
Vivid S6

-2-
GE Healthcare
VividClub

Content

CONFIGURE THE MEASUREMENTS TO YOUR NEEDS ..............................................................................4


Configure your measurements with existing parameters:................................................................................4
Selecting the mode ..............................................................................................................................................................4
Add/remove folders .............................................................................................................................................................4
Add/remove measurements............................................................................................................................................4
Sort the list...............................................................................................................................................................................5
Change the unit .....................................................................................................................................................................5

CREATE YOUR OWN MEASUREMENTS.......................................................................................................6


Add a folder .............................................................................................................................................................................6
Add a subfolder......................................................................................................................................................................6
Copy an existing measurement .....................................................................................................................................7
Create a new measurement............................................................................................................................................7
Delete a folder or measurement....................................................................................................................................9
Add a new formula...............................................................................................................................................................9
General information .........................................................................................................................................................11

NOTE
This hand out is additional training material.
For more information please refer to the user manual and/or reference manual.

-3-
GE Healthcare
VividClub

Configure the measurements to your needs

Configure your measurements with existing parameters:

Go to Config (F2) and select Meas/Text.

Selecting the mode


Click on the mode you want to work on.

Add/remove folders
You will see the list of folders on the right side in the same way as you are used to when opening
the Measurements.
If all folders are closed, you can select / deselect complete folders from the list by clicking into the
white box.

Add/remove measurements
Click on a folder on the right side to open the content.
By removing / adding checkmarks you can select the parameters that should be displayed. Here
an example for the doppler measurements of the mitral valve.

-4-
GE Healthcare
VividClub

If you select one measurement from the menu on the right side you will see all related
parameters. Here you can also select/deselect single parameters in the same way.

Sort the list


To change the order of your measurements in the list click on one parameter and move it up /
down by using the arrow keys.

The same can be done with the folder list.

Change the unit


Select your measurement from the list. In the window (in the middle of the screen) you will find all
related parameters for this measurement.
To change the unit of one parameter click on the unit and you will get a popup list. You can
choose the desired unit from the list.

-5-
GE Healthcare
VividClub

Create your own measurements

Add a folder
Select your mode and make sure all folders are closed.
Press the button Add folder.

Remove the USERDEF… and type in your preferred name.


The new folder will now appear in the list.

Add a subfolder
Open an existing folder i.e. Mitralvalve, press Add folder and type in a name.
This will appear now as a subfolder inside the Mitralvalve folder

-6-
GE Healthcare
VividClub

Copy an existing measurement


It might happen that you want to have an existing measurement in another folder.
Select the folder where you want to add the measurement.
Click on Add measurement.

Select the use copy of function.


With the arrow key you get a scroll down list. Choose your measurement from the list and click Ok.

Create a new measurement


Select your folder.
Press Add measurement
Select now blank and press Ok.

Enter your name for the measurement.


Click on the arrow in the tools bar. A list opens with all possible measurements.

-7-
GE Healthcare
VividClub

Some examples:
2D measurements
2D calliper = distance
2D double calliper = 2 times distance (can calculate the ratio)
2D circle = area

M-Mode
MM calliper = distance or time difference
MM time calliper = time difference
MM LV study, RV study, MV and LA/AO = as known from the M-Mode

Doppler
SD point calliper = 1 point for velocity and pressure
SD calliper = 2 points to mark velocitie and pressure also calculates the pressure difference, time
difference, dp/dt, pht…
SD manual trace = trace curve for maximum and mean velocity, pressure and VTI

These are only some examples, there are much more possibilities.

After selecting the tool there are all related parameters shown on the screen.
Each parameter that should be used needs to have a name.
Parameters without a name are not shown later on. In this way you can make a selection of your
useful parameters.

Here an example.
As measurement the SD calliper is used.

-8-
GE Healthcare
VividClub

Only the velocities and pressures are used here and the time difference.

As shown, all parameters with a name have the checkmark and will be used later on. All
parameters without changes to the default name will not be displayed during working later on.

Delete a folder or measurement


Only user defined folders or measurements can be deleted.
Select the folder/measurement you want to delete.

You see a button with a red cross.


If you are sure you want to delete press OK, after this the folder/measurement is gone.

Add a new formula


Select the measurement where you want to add the new formula.
Click in the window with the parameters and a new line will appear.

Type in a name, then press the = to insert the formula.

-9-
GE Healthcare
VividClub

A new window opens.

• All modes are present with a list of all parameters.


• Operators like + - / * ( )
• Functions with more advanced calculations like sin, cos, exp…
• One line for the Formula.

This means the formula needs to be set up in a way that all parameters and functions are in a row
following the mathematical rules.

Here an example:

Set up in one row:

(LVIDd – LVIDs) / LVIDd * 100

The operators and parameters can be chosen from the list, to avoid misunderstandings.
Parameters are displayed as followed:

- 10 -
GE Healthcare
VividClub

You see the parameter name and in bracket the folder where the measurement is placed.
Create your formula only by using these possibilities.
At the end you can make a check. The system will recognise missing brackets or mistakes of
mathematical principles. It can of course not check if the formula is right.
At the end type in your unit (can not be checked by the system).

General information
All factory parameters you can add to the formula are processed in the background in SI units.
SI units are the standard units for parameters like distance, time, weight and so on.
Some examples:
Distance = meter
Time = second
Weight = kilogram

Therefore you have to add the unit than the system is calculating from SI Unit to your desired unit.
If you don’t add the unit, the result of the formula is displayed in SI Units.

When you add a completely new parameter the system can’t now what is the SI Unit of this.
Therefore you need to add some multiplication factors as 100, 1000 or so on to make the
conversion.

One comment to the unit.


If you add a % to the unit field you don’t need to add the 100 to the formula.
Just only type in A / B, the system will do the rest for you.