Professional Documents
Culture Documents
Service Manual Ergoselect 8 10 12 Rev02 en PDF
Service Manual Ergoselect 8 10 12 Rev02 en PDF
Service Manual Ergoselect 8 10 12 Rev02 en PDF
ergoselect 8 / 10 / 12
General Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Electrical Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Leakage currents. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Inspection according to German Medical Device Operator Ordinance. . . . . . . . . . . . . . . . 9
-3-
Mechanical Assembly ergoselect 8 / 10 / 12. . . . . . . . . . . . . . . . . . . . 55
Controls and indicators ergoselect 8. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Controls and indicators ergoselect 10. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Controls and indicators ergoselect 12. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Supine cushion - dismounting / mounting. . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Cushion dismounting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Cushion mounting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Load unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Removal of housing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Assembly of the load unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Exchange of Torque module. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Exchange of load unit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Swivelling Ergometry Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
Dismount pedal cranks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
Open the Swivelling Unit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Exchange Rubber Damper . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Position Sensor (Hallsensor) exchange. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
Cushion exchange (ergometry unit). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
Gasspring exchange. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Bowden cable exchange - swiveling ergometry unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
Exchange the Plastic Glider. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Exchange of power supply unit / Mains switch. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Motors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Motor „Inclination“ exchange (ergoselect 10/12). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Motor „Tilt“ Exchange (ergoselect 12). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
Control Units. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
Control unit D (integrated). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
Dismounting the control unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
Open control unit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
Control unit M (external) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
Dismounting the control unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
Control unit P/T (external). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
Dismounting the control unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
Open the control unit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
-4-
Alignment / Calibration control unit M (integrated / external). . . . . . . . . . 102
Dynamic Load Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102
Calibration of Blood pressure unit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
NIBP Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107
NIBP Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 108
Motor Calibration (ergoselect 10/12). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109
-5-
Alignment / Calibration control unit T. . . . . . . . . . . . . . . . . . . . . . 140
Dynamic Load Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 140
Calibration of Blood pressure unit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 144
NIBP Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 145
NIBP Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 146
Motor Calibration (ergoselect 10/12). . . . . . . . . . . . . . . . . . . . . . . . . . . . . .147
Addendum. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 155
Errorcodes Software ergoselect 8 / 10 / 12 . . . . . . . . . . . . . . . . . . . . . . . . . . .155
Error code bits DTU (Control terminal D / M - PCB). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 155
Error code bits MBU (Control terminal P/T - PCB) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 156
Error code bits LIU (Load regulation and interface PCB) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 157
Error code bits BPU (Blood pressure module). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 158
Display Errorcodes control unit D, M . . . . . . . . . . . . . . . . . . . . . . . . . . . . .159
-6-
General informations to the service manual
• ergoline GmbH considers itself responsible for the effects on safety,reliability, and performance of the equipment, only
if:
- assembly operations, extensions, readjustments, modifications, or repairs are carried out by ergoline GmbH or by
persons authorized by ergoline GmbH,
- the electrical installation of the relevant room complies with the applicable national and local requirements,
- and the instrument is used in accordance with the intended use and the instructions for use.
• This manual contains service information; operating instructions are provided in the operator’s manual of the
instrument.
ergoselect 8
ergoselect 10
ergoselect 12
• All rights are reserved for instruments, circuits, techniques, and names appearing in the manual.
• This manual was written with the utmost care. Should you still find details that do not correspond with the system,
please let us know and we will correct the issue as soon as possible.
• We reserve the right to modify the design and technical features of the device and are not bound by the information
and illustrations provided in this manual.
• All trademarks appearing in this document are trademarks of their respective owners. Their protection is acknowledged.
• No part of this manual may be reprinted, translated or reproduced without the manufacturer's written permission.
• This manual is not subject to any change order service. Please contact the manufacturer for the latest document
revision.
Printed in Germany
ergoline GmbH
Lindenstrasse 5
72475 Bitz
Germany
-7-
General Information Electrical Safety
• The products ergoselect bear the CE marking CE‑0123
(Notified Body: TÜV), indicating their compliance Warning
with the provisions of the Council Directive 93/42/ • Equipment Damage •
EEC about medical devices and fulfills the essential
requirements of Annex I of this directive. • The regulations about handling ESD sensitive components
The CE marking covers only the accessories listed in must be observed.
the Order Information chapter.
The ergometers are a MDD class IIa product.
-8-
Leakage currents
Measuring equipment
The measuring equipment used for the electrical safety
tests must be inspected and calibrated at regular intervals.
As per the German Medical Device Operator Ordinance, the
measurement results must be documented in the medical
device book.
ergoselect 8 / 10 / 12
These ergometers are protection class I equipment
(connection of ergometer frame to protective ground).
The necessary electrical safety tests for these devices
together with the limit values are listed in the chapters
„Preventive maintanance“.
-9-
- 10 -
Preventive Maintenance / Funtional tests
Hint
• Make sure that all attachement parts can be easily moved and reliable secured (e.g. handles, arm supports):
- check screws (star knobs)
- check easy movement on rails
- 11 -
• Check easy movement of ergometry unit:
- swivelling out
- safe and reliable fixation of both retaining bolts
- check for easy release of ergometry unit
- check that damper for ergometry unit is in place
• Power on
Using the power cord, connect the device to the power line and turn it on with the power switch.
Wait for self test to end:
- check LCD for error messages
- check the software version in the service menu for correct version
- check the “error log file“ in service menu
Expected result: Device powers on without errors, FW is correct version, no errors logged Pass/Fail
• Check speed indication (rpm) displayed on LCD and the additional speed indication on the patient display.
Check noise level of idling drive unit (for example, for grinding noises or noise from the bearing)
Expected result: Speed displayed on both displays and no noticeable noise Pass/Fail
• Set manual load (service menu > manual load), change load (for example, 50 W, 100 W)
- Check load level and load indication
- 12 -
Functional test ergoselect 10
• Check the setup for stability (all 6 feet mounted), adjust with leveling device if required.
• Make sure that all attachement parts can be easily moved and reliable secured (e.g. handles, arm supports):
- check screws (star knobs)
- check easy movement on rails
- 13 -
• Check cranks and pedalshoes:
- pedalshoes have no signs for damage / wear
- pedalshoes securely fixed on cranks
- crank screws securely fixed (torque 40 Nm) on axis
• Power on
Using the power cord, connect the device to the power line and turn it on with the power switch.
Wait for self test to end:
- check LCD for error messages
- check the software version in the service menu for correct version
- check the “error log file“ in service menu
Expected result: Device powers on without errors, FW is correct version, no errors logged Pass/Fail
• Check speed indication (rpm) displayed on LCD and the additional speed indication on the patient display.
Check noise level of idling drive unit (for example, for grinding noises or noise from the bearing)
Expected result: Speed displayed on both displays and no noticeable noise Pass/Fail
• Check speed indication (rpm) displayed on LCD and the additional speed indication on the patient display.
Check noise level of idling drive unit (for example, for grinding noises or noise from the bearing)
Expected result: Speed displayed on both displays and no noticeable noise Pass/Fail
Expected result: motor function and movement range ok, no abnormal sound Pass/Fail
Expected result: motor function and movement range ok, no abnormal sound Pass/Fail
• Set manual load (service menu > manual load), change load (for example, 50 W, 100 W)
- Check load level and load indication
- 14 -
• Units with NIBP module:
- Check cuff tubing and tubing connections, microphone connectors
- Check blood pressure measurement on test subject at rest
- Check blood pressure readings on LCD
- 15 -
Functional test ergoselect 12
• Check the setup for stability (all 6 feet mounted), adjust with leveling device if required.
• Make sure that all attachement parts can be easily moved and reliable secured (e.g. handles, arm supports):
- check screws (star knobs)
- reliable latching and securing of hip and shoulder support
- check easy movement on rails
- 16 -
• Check saddle unit:
- safe and reliable fixation of saddle unit in saddle motor mount
- saddle fixation on mount
- no damage at saddle
• Using the power cord, connect the device to the power line and turn it on with the power switch.
Wait for self test to end:
- check LCD for error messages
- check the software version in the service menu for correct version
- check the “error log file“ in service menu
Expected result: Device powers on without errors, FW is correct version, no errors logged Pass/Fail
• Check speed indication (rpm) displayed on LCD and the additional speed indication on the patient display.
Check noise level of idling drive unit (for example, for grinding noises or noise from the bearing)
Expected result: Speed displayed on both displays and no noticeable noise Pass/Fail
• Check speed indication (rpm) displayed on LCD and the additional speed indication on the patient display.
Check noise level of idling drive unit (for example, for grinding noises or noise from the bearing)
Expected result: Speed displayed on both displays and no noticeable noise Pass/Fail
Expected result: motor function and movement range ok, no abnormal sound Pass/Fail
- 17 -
• Check supine motor „TILT“ with remote control (cable or wireless):
- abnormal noise of motor
- check zero position and maximum position (if not, perform motor calibration)
- check blocking function of hall sensor at ergometry unitn
Expected result: motor function and movement range ok, no abnormal sound Pass/Fail
Expected result: motor function and movement range ok, no abnormal sound Pass/Fail
• Set manual load (service menu > manual load), change load (for example, 50 W, 100 W)
- Check load level and load indication
- 18 -
Electrical Safety
Current leakage test results meet requirements?
Perform electrical safety checks when indicated.
All indicated electrical safety checks require a pass/fail indication for steps performed.
Record the measurement values in your debrief.
Step Bedingung
Condition UUT - EIN
ON Ergebnis
Result Grenzwert
Leakage
(1) (2) Current
Ableitstrom
limits
Erd-Ableitstrom
Earth Leakage Current
Gehäuse-Ableitstrom
Enclosure Leakage Current
Patienten-Ableitstrom
Patient Leakage Current
gegen
to Ground
Erde (3)
(3)
1. Forward Polarity NC __________ μA Pass/Fail 100 μA
Erd-Verbindung
Ground Continuity Resistance
- 19 -
Test procedure “RPM / Load“
Note
• Unless regulatory requirements of the country define a different test procedure or test intervall, it is recommended to perform the
test “RPM / load“ at least every 2 years..
Note
• An external calibration device for dynamic load testing is needed to perfom the following test procedures
(“ergoline ergotest 550“ - order direct from Ergoline P/N 705890) or equivalent device
• Please refer to the operator manual of the used calibration device.
Test preparation
• Follow the operator manual of the used calibration device to connect the calibration device with the axis
of the ergoselect ergometer.
• If supported by the calibration device, connect the ergoselect ergometer interface (RS-232 or USB) with the calibration
device.
• Connect the calibration device with the PC / notebook belonging to the device.
Start the test software of the calibration device.
• Perform the following RPM tests step by step. Note the results in the table or use the printed protocol of the
calibration device to document the test.
#1 30 ± 1 rpm
#2 40 ± 1 rpm
#3 50 ± 1 rpm
#4 100 ± 1 rpm
#5 120 ± 1 rpm
- 20 -
Load test protocol
• Perform the load test (RPM is set by the calibration device / load of the ergoselect ergometer is set either by command
of the test software or manually on the ergoselect ergometer (see „Manual load“) step by step.
• The actual load is measured and displayed by the calibration device software. Note the results in the table or use the
printed protocol of the calibration device to document the test.
Step RPM set Target Load Actual Load Deviation max. Pass / Fail
#1 40 25 ± 3 Watt
#2 40 50 ± 3 Watt
#3 40 100 ± 5 Watt
#4 70 200 ± 10 Watt
#5 90 300 ± 15 Watt
- 21 -
Test procedure “Blood pressure (NIBP)“
Note
• Unless regulatory requirements of the country define a different test procedure or test intervall, it is recommended to perform the
test “Blood pressure (NIBP)“ at least eevery 2 years.
Note
• A calibrated NIBP measure instrument set is needed to perfom the following test procedures
• Please refer to the operator manual of the used device.
Test preparation
• Power the ergoselect ergometer on and enter service menu / NIBP test (-see „Calibration of bloodpressure unit“)
• Connect the calibration equipment with the ergoselect ergometer cuff connector.
Leak-tightness test
• Use the connected test equipment to apply the target pressure (see table below) to the ergoselect ergometer cuff
connector, wait until the pressure in the system has stabilized (approx. 5 sec).
• After 1 min check the actual pressure and note the pressure drop in the table
Step Target Pressure Actual pressure drop Pressure drop max. Pass / Fail
Accuracy test
• Use the connected test equipment to apply the target pressure (see table below) to the ergoselect cuff connector, wait
until the pressure in the system has stabilized (approx. 5 sec).
• Check the actual pressure displayed on the ergoselect display and note the value in the table.
Step Target pressure Displayed pressure (eBike III) Deviation max. Pass / Fail
#1 0 ± 3 mmHg
#2 50 ± 3 mmHg
#3 100 ± 3 mmHg
#4 150 ± 3 mmHg
#5 200 ± 3 mmHg
#6 250 ± 3 mmHg
#7 300 ± 3 mmHg
- 22 -
Electronic modules / Assembly
Overview
Description of electronic and cabling
- 23 -
Electronic carrier
Description
All electronic boards (except for the motor control unit) are
mounted on a special metal carrier placed under the main
cushion.
Mounted on the carrier are the LIU PCB (load control and
interfaces) and (optional) the complete NIBP unit.
- 24 -
Block diagrams electronics ergoselect 8
- 25 -
ergoselect 8 M, complete with all options
- 26 -
ergoselect 8 P/T (external), complete with all options
- 27 -
Block diagrams electronics ergoselect 10
- 28 -
ergoselect 10 M, complete with all options
- 29 -
ergoselect 10 P/T (external), complete with all options
- 30 -
Block diagrams electronics ergoselect 12
- 31 -
ergoselect 12 M, complete with all options
- 32 -
ergoselect 12 P/T (external), complete with all optionsc
- 33 -
Wiring diagrams electronics ergoselect 8 D / M / P / T
ergoselect 8 D/M/P/T
- 34 -
Wiring diagrams electronics ergoselect 10 D / M / P / T
ergoselect 10 D/M/P/T
- 35 -
Wiring diagrams electronics ergoselect 12 D / M / P / T
ergoselect 12 D/M/P/T
- 36 -
Control unit electronicss
The ergoselect 8 / 10 / 12 supine ergometers can be
equipped with different control units.
Control unit D
- 37 -
DTU - Control unit type M
Hint
• The housing of the control unit M CANNOT be opended!
• Upgrading of modules (e.g. bluetooth) is not possible in
the field
• For repairs / upgrades the complete control unit M has to
be sent to ergoline.
- 38 -
DTU - Patient display
The patient display for rpm together with the push buttons
for service menu control are internally connected by a flat 1 2 3
cable and screwed into the housing.
Hint
Control unit D uses the two push buttons to control the service
menu. DTU patient display / buttons for saddle height
The 7 segment displays and LEDs are not visible.
1 LED for rpm zones (control unit M only)
2 Display for rpm (control unit M only)
3 Push buttons for service menu control
DTU - Connectors
2
1
DTU
- 39 -
MBU - Control unit type P (external)
- 40 -
MBU - Control unit type T (external)
Hint
• Exchange of the touch display is not possible - the
complete upper part of the control unit housing has to be
replaced.
- 41 -
MBU-P/T - Connectors
1
2
- 42 -
BIU - Bluetooth module
BT - Bluetooth module
- 43 -
HFE - Receiverboard for digital chestbelt
- 44 -
OSU - SpO2 board
OSU board
- 45 -
Motor Control (ergoselect 10 / 12)
MCU - Motor Control Unit
1 2
Specification battery 1 : Pb 12 VDC 0,8 Ah
MCU - Connectors
- 46 -
[M1] [M2]
Motor Motor lateral tilt
Incline (ergoselect 12 only)
[S] [P1]
Saddle motor Connector LIU
- 47 -
Exchange of the rechargeable batteries
Place the housing cover back on the housing and fix the 4
screws.
- 48 -
Load control and Interfaces
LIU - Load regulation and interface board
The power supply (24 Volt) is connected to the LIU and fed
into the bus system. LIU board
2 3 4
1
5
10 9 8 7
- 49 -
ergoselect 10 / 12:
The connection between the motor control unit (MCU) and
LIU is made by a patch cable (RJ-45 jacks). The BPU unit is
connected to the LIU by direct connectors.
- 50 -
NIBP bloodpressure unit
Assembly
2
3
BPU Measuring electronic board
1 microphone connector (J 9)
2 measuring valve connector (J 6)
3 pump connector (J 4)
3
Foam carrier for measuring valve (1) and pump (2)
(with compensating container 3)
- 51 -
Placement of pressure tubes (schematic)
- 52 -
BPU - Bloodpressure board
1 6 5
The BPU electronics is responsible for analysing the
microfone signals and controlling the valves and the
pump.
2
3
BDE- bloodpressure measuring board
- 53 -
Cuff connector
ergoselect 8
The cuff connectors are mounted on both sides of the
frame by use of Y fittings, long pressure tubes and
microphone cables.
ergoselect 10
The cuff connector for the rihgt side is mounted directly
on the bottom of the electronic housing.
ergoselect 12
The cuff connector for the rihgt side is mounted directly
on the bottom of the electronic housing.
- 54 -
Mechanical Assembly ergoselect 8 / 10 / 12
14 1
2
13 3
12 4
11 6
9 8 7
10
2 Extension for lie down area (detachable) 9 Connector for NIBP cuff (on right and left side)
3 Shoulder supports with adjustment rails (detachable) 10 Release lever for swiveling ergometry unit
4 Release lever for head support (on right or left side) 11 Connectors for power cable and interface cables
6 Standard rail for accessories (on right and left side) 13 Pedalshoe
7 Leveling feet to adjust the ergometer to uneven floors 14 Patient rpm display
- 55 -
Controls and indicators ergoselect 10
11 12 13
10
9b
8
9a
6 5
2 Standard rail for accessories 9 Connector for NIBP cuff (on both sides)
4 Leveling feet to adjust the ergometer to uneven floors 11 Patient rpm display
- 56 -
Controls and indicators ergoselect 12
15
13 14
16 17
12
2
11 3
4
5
6
10
7
9 8
1 Control unit (type „D“), below main cushion 10 Swiveling ergometry unit
4 Release lever for echo flap 13 Arm rest for bloodpressure measurement
6 Saddle 15 Headrest
- 57 -
Supine cushion - dismounting / mounting
Cushion dismounting
Hint
• Always support the cushion in the middle during transport
to prevent it from braking or folding.
Cushion mounting
Carefully lift the cushion onto the metal frame and align
it. Use a Phillips screwdriver to screw the screws manually
into the existing holes in the cushion.
Hint
• Do not use a cordless screwdriver to fix the screws to
avoid overwinding!!
- 58 -
Load unit
Removal of housing
- 59 -
• carefully move the housing downwards
3 4 5
1 Galvanized steelhousing
2 Eddy current brake (450 W / 1000 W )
3 Calibrated torque modul
4 RPM sensor (inside)
5 Precision bearings
- 60 -
• unwind the lead through protection [1] off the
cable harnessv and pull the torque modul connector
through the frame hole
- 61 -
• lead the cable of the new module through the housing
gap
Hint
• Lead the cables on the outer edge of the load unit and fix
it with cable ties in a way that no contact with the belt
pulleys is possible.
• Shift the metal housing from below over the load unit
and fix it with 5 screws.
Hint
• After exchange of the torque module a dynamic load
calibration has to be performed (with calibration stand)!
- 62 -
Exchange of load unit
- 63 -
• disconnect (unscrew) the power connector from the
motor control unit
- 64 -
• move the service cart dense to the supine ergometer
and lift it exactly (stop bars) under the load unit
Hint
• Remove only the screws marked in the picture - the other
screws fix the load unit itself!
- 65 -
• lower the unit with the service cart, until the screw-on
flap are accessible
- 66 -
• subsequently drive the service cart with load units
sidewards, while releasing the cardan joint carefully
from the axis
• lift the new load unit with the cart, place it laterally
and connect the cardan joint with the axis
• lift the load unit until the fixing screws fit to the
screw-on flaps
- 67 -
Swivelling Ergometry Unit
Dismount pedal cranks
- 68 -
Open the Swivelling Unit
Hint
• Before reassembling the cover, check the exact position of
the 4 spring nuts.
- 69 -
• lower the ergometry unit
Hint
• The plastic clips are designed for single use - they should
not be reinserted!
- 70 -
Gasspring exchange
- 71 -
• place the new gas spring with the mounting tool aside
the rear bolt and shift it completely onto that bolt
- 72 -
Bowden cable exchange - swiveling ergometry unit
- 73 -
• Screw the bowden cable guide completely off the
mount.
- 74 -
• Unwind the bowden cable from the index plunger.
- 75 -
• The final position of the indexing plunger can be
adjusted by turning the nut at the end of the bowden
cable.
vertcal), the nut of the bowden cable need to have
a gap to the bushing of the indexing plunger (see
picture).
- 76 -
Accessories
Exchange the Plastic Glider
- 77 -
Power Supply
Exchange of power supply unit / Mains switch
• cut the cable tie fixing the power cable (from mains
switch)
- 78 -
Motors
Motor „Inclination“ exchange (ergoselect 10/12)
- 79 -
• Adjust the special supports at the front and rear side
to the frame of the ergometer and fix them tight with
the velcro straps.
- 80 -
• Remove the lower bolt screw wit a wrench and an
open-end wrench (19 mm).
- 81 -
Mounting of a New Motor
Warning
• Patient Hazard, Equipment Damage •
• Insert the bolt screw and fix it light with washer and
nut, if necessary position the plastic rings using a
small screwdriver
- 82 -
• Tightly fasten both bolt screws (top and bottom).
- 83 -
Motor „Tilt“ Exchange (ergoselect 12)
- 84 -
• Unwind the lower bolt screw with a wrench and open-
end wrench (19 mm) and remove it.
If necessary use a plastic hammer and a plastic or
wooden bar to eject the bolt
- 85 -
• Remove the motor by pulling it out off the upper
mounting flap
• Insert the bolt screw and fix it light with washer and
nut, if necessary position the plastic rings using a
small screwdriver
- 86 -
• Insert the 2 plastic rings at the upper motor bracket
an shift it onto the fixation flap.
If necessary change the position of the bracket by
turning the.motor spindle.
Use a plastic hammer to push the bracket carefully
onto the flap
• Insert the bolt screw and fix it light with washer and
nut, if necessary position the plastic rings using a
small screwdriver.
- 87 -
- 88 -
Control Units
Control unit D (integrated)
Dismounting the control unit
- 89 -
Control unit M (external)
Dismounting the control unit
Note
• It is not possible, to open the control unit M.
For repair or installing additional modules, the control
unit has to be sent to ergoline.
- 90 -
Open the control unit
The front and rear panel are plugged in the upper part of
the housing and can easily be removed.
Control unit P
Control unit T
- 91 -
First unlock the zero insertion force connector on the
MBU board: by shifting the 2 black sliders towards the
connecting cable.
Then insert the flat cable into the connector and move the
2 sliders back to fix the cable in the connector.
- 92 -
Servicemenü Control unit D (integrated) / M (external)
Activating service menu
Control unit D (integrated)
To activate the service menu the ergometer has to be
switched off and on again.
Note
• Do not pedal while the ergometer is being turned on and
during the self-test.
• No blood pressure cuff should be applied.
• If an error message appears immediately after the self-
test, please refer to the Addendum / Error Messages
section for advice.
- 93 -
To prevent settings from being changed in error, the
service menu is “protected“ with a password.
Service Code
Select “999“ with the arrow keys and confirm by pressing
NIBP.
999
The service menu together with setup functions displays. Service Menu
Contrast
ECG Type
Baudrate 8N2
Analog Calibration
Software update
Dynamic Load Calibr.
Saddle Calibration
- 94 -
Service menu functions
Hint Hint
• The service menu in the ergometers is available in english Modules
The order/ of
options
the function
have todescriptions
be activateddoes
in the
not
system
correspond to
only. configuration
the order in thebefore
service
themenu!
related functions appear in the service
menu.
• Contrast
• EKG-Typ
• NIBP Beep
• NIBP mmHg / hPa
• Baudrate 8N2
• Analog Calibration (not used)
• Software Update
• Dynamic Load Calibration
• Saddle Calibration (not used)
• Service Date
• NIBP Calibration (with option NIBP only)
• NIBP Test (with option NIBP only)
• System Config.
• Error log
• Handle Calibration (not used)
• Manual load
• Reset DMS values
• Ergometer type
• Drive unit
• Motor calibration (ergoselect 10/12 only)
• Remote control (ergoselect 10/12 only)
Contrast
The contrast of the LC Display can be adjusted between Contrast
0 to 100 %.
Confirm with NIBP.
50 %
- 95 -
EKG Type Note
• The EKG Type needs to be selected only when the ergom-
The selected EKG Type determines the communication eter is connected to an ECG unit. The selection is part of
method with the ECG recorder, PC-based ECG system, etc. the installation procedure.
NIBP Beep
The tone signal for the systolic pulse (during NIBP NIBP Beep
measurement) can be set to on or off.
On
Select the desired setting and confirm by pressing NIBP.
Off
Select the desired units and confirm by pressing NIBP. mmHg kPa
- 96 -
Baudrate 8N2
The baudrate can be selected between 1.200 Baud and Baudrate 8N2
115.000 Baud. 1.200
2.400
Default value for standard ergometers is 4.800 Baud.
4.800
Default value for REHAB ergometers is 38.400 Baud.
9.600
Select the desired parameter by use of the right and left 19.200
softkeys (↑ ↓) and confirm with ENTER. 38.400
76.800
Software Update
For a description of a software update of the ergometer
(firmware) see the separate chapter „software update“.
Service Date
The next inspection date can be set in this menu (the Service Date
inspection date is visible under „settings“ for the user.
08. 12. 2011
A hint „Service“ when switching the ergometer on can be Show message
activated. NO
Operating Hours
The operating hours represent the time, the ergometer was
127
powered on, independently from any load or activity on
the bike. EXIT
- 97 -
System Config.
The hardware configuration of the ergometer is defined in
this setup.
Bluetooth
This setting is used to activate an integrated Bluetooth Bluetooth Config.
module.
Default name used is ‚Ergo01‘ - the default channel
used is ‚1‘.
Bluetooth 0 I
Name and channel can be changed by using the ergoline Name Ergo01
PC service program (- the Bluetooth module has to be Channel 1
deactivated temporarily).
EXIT
Wifi
This setting is used to activate an integrated Wifi module.
Wifi Config.
IP - the actual configured IP-address IP: 192.168.200.010
of the Wifi module Wifi 0 I
Auto Reset - default setting is 1 Auto Reset 0 I
Reset Wifi Defaults - following an exchange of the Reset Wifi Defaults
Wifi module, the actual stored
Module Reset
configuration of the defect
module can be restored to the EXIT
new one
Module Reset - a module reset is performed
(for test purposes only)
- 98 -
Lean motor / Tilt motor
These settings are available for the different supine
ergometers ergoselect 10 / 12
RPM inverse
If the patient RPM display is mounted on the left side of
the device (ergoselect 8/10 only) this setting activates an
inverse display of the rpm values - the rpm module have
to be mounted in an upright position.
NIBP Dummy
Der Eintrag ist nur bei Ergometern OHNE integrierte
Blutdruckmessung vorhanden.
Manual Load
A manual load can be set between 1 - 999 Watt.
Adjust the desired value with the arrow keys and set the Manual load
load with the NIBP. key
75 Watt
NIBP key at 0 Watt (EXIT) returns to the service menu. Manual load
0 Watt
EXIT
- 99 -
Error log
The software of the ergoselect II ergometers stores internal Error log
error messages in a special memory area (error log file). Err. Log Quantity: 2
These error messages can be checked and cleared in this
menu. Show Error Logs
Show NIBP Logs
Clear Error Logs
To display the stored error messages select „Show Error Clear NIBP Logs
Log“ and confirm with NIBP. EXIT
EXIT
By using the arrow keys, the following error messages can
be displayed. Displaying a single error
- 100 -
Ergometer Type
Ergometer Type
If in PC mode an „I“ command is received, the ergometer
sends an identification data string to the PC. / ECG
recorder. ER 900
Drive Unit
Drive Unit
The ergoselect III ergometers support two different load
units (AT-3 und AT-4).
Unit 3
For ergoselect 8 / 10 / 12 select type „Unit 3“ and confirm
with NIBP.
Unit 4
Motor calbration
The calibration of the tilt and lean motors is described in
the separate chapter „Calibration / alignment“.
Remote control
This function is used to configure the wireless remote
control („Pairing“) for the motor control of
ergoselect 10 / 12.
- 101 -
Alignment / Calibration control unit M (integrated / external)
Dynamic Load Calibration
• Dismount BOTH (!) pedal cranks of the ergometer.
- 102 -
Control unit M, integrated
Bedienteil M, extern
999
Navigate through the menus by using the arrow keys Service Menu
(saddle height) to position the cursor bar. Contrast
Confirm the selection by pressing NIBP (to the right ECG Type
of the LCD).
Baudrate 8N2
Analog Calibration
Software update
Dynamic Load Calibr.
Select „Dynamic load calibration“ and confirm with NIBP Saddle Calibration
- 103 -
• The line „DMS“ shows:
stored DMS value / actual DMS value Dynamic Load Cal.
Calibration offset
Gain: 1,00
DMS: 44 / 42
RPM: 0 B 0 P
Load: 0 SW 0 AW
• Start the offset calibration process with NIBP (START) START
• The actual offset value is averaged for 5 sec and Dynamic Load Cal.
stored Calibration offset
Gain: 1,00
DMS: 42 / 42
RPM: 0 B 0 P
Load: 0 SW 0 AW
WAITING...
• The line „RPM“ shows the measured rotations/min Dynamic load calibration step 1 - waiting for START
(issued by the calibrator unit)
[the value „P“ is not used in the ergometer)
- 104 -
• Set the calibrator unit to 110 rpm. Dynamic Load Cal.
Dyn. Calibration
• Press NIBP (START) to start calibration step #2
Gain: 1,00
DMS: 42 / 42
• Wait approx. 20 seconds, until the load display is RPM: 110 B 0 P
stable Load: 50 SW AW
START
• By using the arrow keys, input the actual value of the Dynamic Load Cal.
calibrator (e.g. 403,2 Watt) at the ergometer. At 110 /400 W 403,2 W
Gain: 1,00
• Store the value with SET. DMS: 42 / 42
RPM: 110 B 0 P
Load: 400 SW AW
SET
Dynamic load calibration step 2 - input load value
The successfull completion of the dynamic load calibration Dynamic Load Cal.
is confirmed.
EXIT
Dismount the calibration unit and reassemble the cranks /
pedals to the ergometer. Dynamic load calibration successfull
- 105 -
Calibration of Blood pressure unit
Recommended components:
• Hand pump
Hand pump
- 106 -
All the components are connected by tubes (one pressure
system) and connected to the cuff connector of the
ergoselect 8 / 10 / 12 (use cuff adaptor).
NIBP Test
123 mmHg
Use the hand pump to apply a fixed pressure to the
system.
Note
- 107 -
NIBP Calibration
Offset calibration
The cuff connector (self closing) has to be opened (use cuff Offset: 0 mmHg
connector).
SET
1. Remove cuff. EXIT
2. Connect the cuff connector with a short tube.
Pressure calibration - offset
3. The actual pressure value is displayed.
7. Wait several seconds until the pressure is stable; Pressure calibration - gain
correct the pressure if necessary.
- 108 -
Motor Calibration (ergoselect 10/12)
This function performs a calibration of the two motors
(„tilt“ and „lean“)
Press NIBP and hold the key to perform the automatic Motor calibration
calibration process..
Press and hold
START
EXIT
• The NIBP key has to be kept pressed during the complete ! PAUSED !
calibration procedure.
• If the key is released, the process is paused and can be
resumed by pressing and holding the NIBP key again. CONT
EXIT
! SUCCESSFULL !
CONT
- 109 -
After confirming [CONT by pressing the NIBP key, it is Motor zero adj.
possible to manually adjust the horizontal position of the
supine (this is not necessary in most cases).
LEAN 0
TILT 0
LEAN 2
TILT 0
EXIT
- 110 -
Service menu control unit P
The functions of the servicemenu are described in this Hint
chapter. • The service menu in the ergometers is available in english
only.
Depending on the actual configuration of the ergoselect
4/5 ergometer, some servce functions are not visible or • Leaving the service menu is only possible by switching the
activ, device off.
ergoline
Turn the ergometer on by pressing the power switch
marked I/O.
The ergometer runs a self-test. Subsequently, the standard
GmbH
display appears.
Self-test screen
Note
• Do not pedal while the ergometer is being turned on and
during the self-test.
• No blood pressure cuff should be applied.
• If an error message appears immediately after the self-
test, please refer to the Addendum / Error Messages
section for advice.
- 111 -
As soon as the self test appears on the ergometer display,
press and hold the two keys synchronously: Activating the service menu
ergoline
GmbH
Selftest
999
ENTER
Select the desired function with the softkeys (↑ ↓) and Service Menu
confirm with [ENTER]. Baudrate 8N2
Analog Calibration
ECG-Startpulse
NIBP Parameter
NIBP Test
ENTER
- 112 -
Service functions control unit P
Note
• Modules / options have to be activated in the system
configuration before the related functions appear in the
service menu.
• Baudrate 8N2
• Software Update
• Dynamic Load Calibration
• PC-Mode
• ECG-Startpulse
• System Config.
• Service date
• Error log
• Reset DMS values
• NIBP-Start
• Drive unit
• nalog Calibration
A with option COM module (analogue) only
• NIBP-Parameter with option blood pressure only
• Ergometer type without option blood pressure only
• NIBP Calibration with option blood pressure only
• NIBP Test with option blood pressure only
The sequence of the functional descriptions does not reflect the sequence in the service menu of the ergometer.
- 113 -
Baudrate 8N2
The baudrate can be selected between 1.200 Baud and Baudrate 8N2
115.000 Baud.
1.200
Default value for standard ergometers is 4.800 Baud. 2.400
Default value for REHAB ergometers is 38.400 Baud. 4.800
9.600
Select the desired parameter by use of the right and left 19.200
softkeys (↑ ↓) and confirm with ENTER.
ENTER
Software Update
For a description of a software update of the ergometer
(firmware) see the separate chapter „software update“.
System Config.
The hardware configuration of the ergometer is defined in
this setup.
- 114 -
Lean motor / Tilt motor / RPM inverse
These settings are reserved for the supine ergometers
ergoselect 8 / 10 / 12.
They have no function in bike ergometers.
NIBP Dummy
If the parameter NIBP-Dummy is set to I , a bloodpressure
value of 000/000 is sent from ergometers without
integrated NIBP module.
This enables the connection to ECG recorders which
support ergometers with NIBP module only
Bluetooth
This setting is used to activate an integrated Bluetooth Bluetooth Config.
module.
Default name used is ‚Ergo01‘ - the default channel
used is ‚1‘.
Bluetooth 0 I
Name and channel can be changed by using the ergoline Name Ergo01
PC service program (- the Bluetooth module has to be Channel 1
deactivated temporarily).
ENTER
Wifi
This setting is used to activate an integrated Wifi module.
Wifi Config.
IP - the actual configured IP-address IP: 192.168.200.010
of the Wifi module
Auto Reset - default setting is 1
Wifi 0 I
Reset Wifi Defaults - following an exchange of the Auto Reset 0 I
Wifi module, the actual stored Reset Wifi Defaults
configuration of the defect Module Reset
module can be restored to the
new one ENTER
Module Reset - a module reset is performed
(for test purposes only)
- 115 -
Heartrate belt
The type of the heartrate belt used can be selected. Heart rate belt
POLAR receiver 0 I
POLAR - POLAR chestbelt (e.g. T31) MM5-Modul 0 I
MM5-Modul - digital ergoline belt
MM6-Modul 0 I
MM6-Modul - reserved
ENTER
Ergometer Type
Ergometer Type
If in PC mode an „I“ command is received, the ergometer
sends an identification data string to the PC. / ECG
recorder. ER 900
- 116 -
Error log
The software of the ergoselect II ergometers stores internal Error log
error messages in a special memory area (error log file). Err. Log Quantity: 2
These error messages can be checked and cleared in this Show Error Logs
menu.
Show NIBP Logs
Clear Error Logs
Clear NIBP Logs
To display the stored error messages select „Show Error
Log“ and confirm with [ENTER]. ENTER
- 117 -
NIBP Parameter
If the ergometer load is controlled by the internal
programs, this parameter defines the start of the NIBP
module.
Leadtime
Defines the leadtime of the bloodpressure measurement to NIBP Parameter
the next load step
Leadtime
Setting range: 40 to 90 sec. in 1 sec. steps 60 sec
Resting NIBP
ON
Resting NIBP
Defines whether or not a resting bloodpressure
measurement is taken before starting the stress test. ENTER
PC-Mode
PC-Mode „Automatic“ PC Mode
If „YES“ is activated, the ergometer automatically switches
to remote mode (PC mode) as soon as a valid signal is
Automatic
received from the interface.
Yes
No
ENTER
Automatic Mode
Service Date
The next inspection date can be set in this menu (the Service Date
inspection date is visible under „settings“ for the user. 08. 12. 2011
A hint „Service“ when switching the ergometer on can be Show message
activated. NO
Operating Hours
The operating hours represent the time, the ergometer was 127
powered on, independently from any load or activity on
the bike. ENTER
- 118 -
NIBP Start
Function to initiate an automatic blood pressure
measurement from the servicemenu.
Drive Unit
Drive Unit
TThe ergoselect III ergometers support two different load
units (AT-3 und AT-4).
Unit 3
For ergoselect 8 / 10 / 12 select type „Unit 3“ and confirm
with ENTER.
Unit 4
ENTER
Motor calbration
The calibration of the tilt and lean motors is described in
the separate chapter „Calibration / alignment“.
Remote control
This function is used to configure the wireless remote
control („Pairing“) for the motor control of
ergoselect 10 / 12.
- 119 -
Alignment / Calibration control unit P
Dynamic Load Calibration
• Dismount BOTH (!) pedal cranks of the ergometer.
Service Code
999
ENTER
Navigate through the menus by using the arrow keys
(saddle height) to position the cursor bar. Password for service menu
Confirm the selection by pressing NIBP (to the right
of the LCD).
- 120 -
Select „Dynamic load calibration“ and confirm with ENTER Service Menu
Baudrate 8N2
Analog Calibration
Software update
Dynamic Load Calibr.
Saddle Calibration
ENTER
• Start the offset calibration process with START Dynamic Load Cal.
Calibration offset
Gain: 1,00
• The line „DMS“ shows:
stored DMS value / actual DMS value DMS: 44 / 42
RPM: 0 B 0 P
Load: 0 SW 0 AW
START
• By pressing START the actual value is averaged for 5 Dynamic Load Cal.
sec and stored
Calibration offset
Gain: 1,00
DMS: 42 / 42
RPM: 0 B 0 P
Load: 0 SW 0 AW
WAITING...
- 121 -
• Set the calibrator unit to 110 rpm. Dynamic Load Cal.
Dyn. Calibration
Gain: 1,00
• Press START to start calibration step #2
DMS: 42 / 42
RPM: 110 B 0 P
• Wait approx. 20 seconds, until the load display is Load: 50 SW AW
stable
START
• By using the arrow keys, input the actual value of the Dynamic Load Cal.
calibrator (e.g. 50,8 Watt) at the ergometer. At 110 /50 W 50,8 W
Gain: 1,00
DMS: 42 / 42
RPM: 110 B 0 P
Load: 50 SW AW
• The ergometer automatically sets the load to 400 Watt Dynamic Load Cal.
• Wait approx. 20 seconds, until the load display is
Dyn. Calibration
stable Gain: 1,00
DMS: 42 / 42
RPM: 110 B 0 P
Load: 400 SW AW
START
• By using the arrow keys, input the actual value of the Dynamic Load Cal.
calibrator (e.g. 403,2 Watt) at the ergometer. At 110 /400 W 403,2 W
Gain: 1,00
• Store the value with SET. DMS: 42 / 42
RPM: 110 B 0 P
Load: 400 SW AW
SET
Dynamic load calibration step 2 - input load value
- 122 -
Calibration of Blood pressure unit
Recommended components:
• Hand pump
Hand pump
- 123 -
All the components are connected by tubes (one pressure
system) and connected to the cuff connector of the
ergoselect 8 / 10 / 12 (use cuff adaptor).
NIBP Test
123 mmHg
Use the hand pump to apply a fixed pressure to the
system.
Note
- 124 -
NIBP Calibration
Offset calibration
The cuff connector (self closing) has to be opened (use cuff Offset: 0 mmHg
connector).
1. Remove cuff.
SET
2. Connect the cuff connector with a short tube.
Pressure calibration - offset
3. The actual pressure value is displayed.
7. Wait several seconds until the pressure is stable; Pressure calibration - gain
correct the pressure if necessary.
- 125 -
Motor Calibration (ergoselect 10/12)
This function performs a calibration of the two motors
(„tilt“ and „lean“)
Press START and hold the key to perform the automatic Motor calibration
calibration process..
Press and hold
[START
START EXIT
RUN EXIT
! SUCCESSFULL !
CONT
- 126 -
After confirming [CONT], it is possible to manually adjust Motor zero adj.
the horizontal position of the supine (this is not necessary
in most cases).
LEAN 0
TILT 0
LEAN 2
TILT 0
ENTER
- 127 -
- 128 -
Service menu control unit T
The functions of the servicemenu are described in this Hint
chapter. • The service menu in the ergometers is available in english
only.
Depending on the actual configuration of the ergoselect
ergometer, some servce functions are not visible or activ, • Leaving the service menu is only possible by switching the
device off.
Manual
Setup
Main Menu
Date/Time Regulation
Setup menu
Press [Service] to activate the service menu.
- 129 -
Service
Drehzahl Ton
0 Service
- 130 -
Service functions control unit T
Note
Baudrate 8N2
Software Update
Dynamic Load Calibration
PC-Mode
ECG-Startpulse
System Config.
Service date
Error log
Reset DMS values
NIBP-Start
The sequence of the functional descriptions does not reflect the sequence in the service menu of the ergometer.
- 131 -
Baudrate 8N2
The baudrate can be selected between 1.200 Baud and
115.000 Baud.
Baudrate 8N1
Default value for standard ergometers is 4.800 Baud.
1.200 Baud 19.200 Baud
Default value for REHAB ergometers is 38.400 Baud.
2.400 Baud 38.400 Baud
Select the desired parameter and confirm with [ ].
. 4.800 Baud 57.600 Baud
Software Update
For a description of a software update of the ergometer
(firmware) see the separate chapter „software update“.
- 132 -
System Config.
The hardware configuration of the ergometer is defined in
this setup. System Config.
Heartrate belt ON
System Config.
NIBP Dummy
SpO2 Modul ON
If the parameter NIBP-Dummy is set to I , a bloodpressure 1.
value of 000/000 is sent from ergometers without SpO2 Messages OFF
3.
integrated NIBP module. 60 Hz Power Freq. OFF
This enables the connection to ECG recorders which
450 Watt brake OFF
support ergometers with NIBP module only
NIBP Dummy OFF
Handle engine ON
2.
Bluetooth ON
4.
WiFi OFF
No ECG in PC-Mode
The ECG display on the screen is suppressed during PC
mode. - not used in supine ergometers. System Config.
N/AF
N/A
N/A
N/A
- 133 -
Bluetooth
This setting is used to activate an integrated Bluetooth
module. Bluetooth Config.
Default name used is ‚Ergo01‘ - the default channel
used is ‚1‘.
Name and channel can be changed by using the ergoline
Bluetooth OFF
PC service program (- the Bluetooth module has to be
deactivated temporarily). Device Name Ergo01
Operating Channel 1
Wifi
This setting is used to activate an integrated Wifi module.
Heartrate belt
The type of the heartrate belt used can be selected.
System Config.
POLAR - POLAR chestbelt (e.g. T31)
MM5-Modul - digital ergoline belt
MM6-Modul - reserved
POLAR receiver OFF
MM5-Modul ON
MM6-Modul OFF
- 134 -
Ergometer Type
If in PC mode an „I“ command is received, the ergometer
sends an identification data string to the PC. / ECG
recorder.
Note
• This setting is only valid with ergometers without NIBP
module!
Drive Unit
The ergoselect III ergometers support two different load
units (AT-3 und AT-4).
Motor calbration
The calibration of the tilt and lean motors is described in
the separate chapter „Calibration / alignment“.
- 135 -
Service Date
The next inspection date can be set in this menu (the
inspection date is visible under „settings“ for the user. Service Date
A hint „Service“ when switching the ergometer on can be 116 Operating Hours
activated. 5 10 2019
The operating hours represent the time, the ergometer was Show Message
powered on, independently from any load or activity on
the bike. YES NO
NIBP Parameter
If the ergometer load is controlled by the internal
programs, this parameter defines the start of the NIBP NIBP Parameter
module.
Leadtime
Leadtime 60 sec
Defines the leadtime of the bloodpressure measurement to
the next load step Resting NIBP ON
Resting NIBP
Defines whether or not a resting bloodpressure
measurement is taken before starting the stress test.
PC-Mode
PC-Mode „Automatic“
If „YES“ is activated, the ergometer automatically switches PC Mode
to remote mode (PC mode) as soon as a valid signal is
received from the interface. Automatic
Yes No
- 136 -
ECG Startpulse
To configurate the remote start pulse for external ECG
recorders (ergometer is controlled by internal programs). ECG Startpulse
Leadtime
Remote pulse leadtime before the next load step
Pulsewidth
Remote pulse pulse width
- 137 -
Error log
The software of the ergoselect III ergometers stores
internal error messages in a special memory area Error Log
(error log file) and a statistical overview of the
bloodpressure measurements taken. Error Log Quantity: 6
To display the stored error messages select „Show Error Show NIBP Logs Clear NIBP Logs
Log“.
Mic. level 2: 6
Mic. level 3: 11
Mic. level 4: 22
Mic. level 5: 18
- 138 -
Chipcard-Mode
Ergoselect ergometers with control terminal „T“ are
equipped with an integrated chipcard device to read and Chipcard-Mode
write ergoline chipcards..
No chipcard Only chipcard
The settings define how the ergometer supports those Chipcard possible
chipcards.
Only chipcard
After power on the ergometer waits for an ergoline
chipcard to be inserted - no training is possible without a
chipcard.
Remote control
This function is used to configure the wireless remote
control („Pairing“) for the motor control of
ergoselect 10 / 12.
- 139 -
Alignment / Calibration control unit T
Dynamic Load Calibration
• Dismount BOTH (!) pedal cranks of the ergometer.
Manual
Setup
The following load / rpm combinations have to be
measured during the calibration process:
Laständerung
Load Change HF-Gurt
HR belt no.
Nummer
Datum/Uhrzeit
Date/Time Regelung
Regulation
• The [ Setup ] key in the main menu opens the
configuration menu. Drehzahl
RPM Ton
Beep Service
Setup menu
• To prevent settings from being changed by chance, the Standard Betrieb Programme
2.
service menu is „protected“ by a password.
Enter „999“ with the keyboard displayed and confirm EKG-Typ Training/Test
1 2 3
with [ ]
Laständerung HF-Gurt Nummer4 5 6
Service Code
Datum/Uhrzeit (0* ... 999*)
Regelung 7 8 9
Drehzahl Ton
0 Service
- 140 -
• The service menu displays.
Baudrate 8N1 PC-Mode
2.
Analog Calibration Saddle calibration
- 141 -
• The line „Rotation (brake)“ shows the measured
rotations/min (issued by the calibrator unit)
[the value „pedal“ is not used in the ergometer) Dynamic Load Calibration at 50 rpm
• Wait approx. 20 seconds, until the load display is Dynamic Load Calibration at 110 rpm
stable
Calibration Offset :42
Calibration Gain :1.00
• Input the actual measured load value of the calibrator DMS Value :42
(e.g. 403,2 Watt) with the keyboard an confirm Rotation (brake/pedal) :110rpm / 0rpm
with [ .] LIU load (target/actual) :400W / 400W
Set
- 142 -
• The successfull completion of the dynamic load
calibration is confirmed.
- 143 -
Calibration of Blood pressure unit
Recommended components:
• Hand pump
Hand pump
- 144 -
All the components are connected by tubes (one pressure
system) and connected to the cuff connector of the
ergoselect 8 / 10 / 12 (use cuff adaptor).
NIBP Test
NIBP Test
Use the hand pump to apply a fixed pressure to the
system. NIBP pressure 0 mmHg
CLOSE VALVE
Note
- 145 -
NIBP Calibration
Offset calibration
NIBP Calibration
1. Remove cuff.
Offset 0 mmHg
- 146 -
Motor Calibration (ergoselect 10/12)
This function performs a calibration of the two motors
(„tilt“ and „lean“)
Hint
• During a motor calibration the swiveling ergometry unit Chipcard Mode Ergometer Type
1.
has to be fixed in the vertical position!
System Config. Motor Calibration
Touch Calibration
Calibration running !
Hint
Motor calibration
• The [Run calibration] key has to be kept pressed during
the complete calibration procedure.
• If the key is released, the process is paused and can be
Press and hold Continue
resumed by pressing and holding the CONTINUE key
again.
Calibration paused !
- 147 -
The successfull calibration is confirmed on the display.
Motor calibration
Calibration successfull !
LEAN motor 0
TILT motor 0
Using the arrow keys adjust the zero position and return to
the service menu with OK.
- 148 -
Software update D, M, P, T
Hint
The software update process is identical for all ergoselect
ergometers. • Always use the latest software version for the appropriate
ergometer.
To load a new version of the internal software into the The software can be downloaded from the ergoline
ergoselect III ergometers, a PC or Laptop with USB or a website (www.ergoline.com / infoportal/software) after
serial port (RS-232) is needed. LOGIN.
To use the ergoemter USB port, a special driver has to be • Use the PC update program included in the software
installed on the PC( virtual COM driver). package.
- 149 -
Control unit D / M
Selection of „Show Software rev.“ and confirmation Software Update
with NIBP displays the installed software versions of the
different electronic modules. EF 1.1
Update
Show Software rev.
EXIT
BDE 1.1 0 0
Press the „Arrow down“ button to display the installed Grp: Op-SW: Boot-Lo:
bootloader versions of the different electronic modules. DTU 1.0 0.1
LIU 1.1 0.3
Grp: name of the module
Op-SW: required software version
(has to be equal to the Ap-SW in previous display)
BootLo: actual bootloader version of PCB BDE 1.2 0.3
EXIT
Software Update
Start the software update by selecting „UPDATE“ and
confirm with NIBP. EF 1.1
Update
Show Software rev.
EXIT
- 150 -
Control unit P
Press (↓) to display the installed software versions of the Software Update
different electronic modules.
EF 1.1
UPDATE
Press (↓) to display the installed bootloader versions of the Grp: Op-SW: Bootlo:
different electronic modules.
DTU 1.0 0.1
Grp: name of the module
LIU 1.1 0.3
Op-SW: required software version
(has to be equal to the Ap-SW in previous display)
BootLo: actual bootloader version of PCB
- 151 -
Control unit T
The installed software versions of the different electronic
Update
modules are displayed.
Group: SetSW: AppSw: XCUT LP-No: Boot-Sw:
- 152 -
Update
Activate the PC- update program.
- 153 -
- 154 -
Addendum
Errorcodes Software ergoselect 8 / 10 / 12
Every electronic module of the ergoselect ergometers has its own error codes.
If an error occurs, a hexadecimal code is displayed on the ergometer display and internally stored with date/time of first
occurence and numbers of recurrences.
The displayed hexadecimal code is a combination of single error bits, each representing a special fault condition.
Bit 15
Bit 14
Bit 13
Bit 12
Bit 11
Bit 10
Bit 9
Bit 8
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
Error bits set: Bit 3 LCD communication error
Bit 4 +24V out of range
Note
• A service software to download Error codes and device settings from the ergoselect ergometers is available (downloadable) from
the ergoline website.
0 Illegal program jump Safety shut down - restart ergometer (OFF / ON)
- if error persists:
replace control terminal
2 not used
5 not used
6 not used
7 Safety state mode Safety shut down - restart ergometer (OFF / ON)
- if error persists:
replace control terminal
8 not used
- 155 -
Error code bits MBU (Control terminal P/T - PCB)
0 Illegal program jump Safety shut down - restart ergometer (OFF / ON)
- if error persists:
replace MBU
2 not used
7 Safety state mode Safety shut down - restart ergometer (OFF / ON)
- if error persists:
replace control terminal
8 RAM Test error Safety shut down - restart ergometer (OFF / ON)
- if error persists:
replace MBU
12 not used
12 not used
14 not used
15 not used
- 156 -
Error code bits LIU (Load regulation and interface PCB)
2 DMS Offs. invalid Safety shut down - perform static calibration OFFSET
- if error persists:
reset DMS default values
- if error persists:
replace strain gauge module
5 +5V out of range Safety shut down - restart ergometer (OFF / ON)
- if error persists: replace LIU
7 Safety state mode Safety shut down - restart ergometer (OFF / ON)
- if error persists:
replace LIU
9 CAN zykl. Timeout Safety shut down - restart ergometer (OFF / ON)
- if error persists:
check cable DTU <-> LIU
- check connection between
all PCB (e.g. LIU, SMU, BPU)
- replace LRE
- 157 -
Error code bits BPU (Blood pressure module)
3 Errror Calibration Module is switched off Internal error - perform NIBP test
- perform NIBP calibration
- replace BPU PCB
4 Error offset Module is switched off Internal error - perform NIBP test
- perform NIBP calibration
- replace BPU PCB
5 not used
6 not used
7 Safety state mode Module is switched off Internal error - restart ergometer (OFF / ON)
- if error persists:
replace BPU
8 Setup not ready Module is switched off Internal error - restart ergometer (OFF / ON)
- if error persists:
check connection LIU <-> BPU
- replace BPU
9 CAN zykl. Timeout Module is switched off Internal error - restart ergometer (OFF / ON)
- if error persists:
check cable DTU <-> LIU
check connection LIU <-> BPU
- replace BPU
10 not used
11 not used
12 not used
13 not used
14 not used
15 not used
- 158 -
Display Errorcodes control unit D, M
The control unit M shows errors in a coded form:
- 159 -
ergoline GmbH
Lindenstrasse 5
72475 Bitz
Germany
Fon.: +49 (0) 7431 - 98 94 0
Fax: +49 (0) 7431 - 98 94 128
email: info @ ergoline.com
http: www.ergoline.com