Professional Documents
Culture Documents
GK99000353GB
GK99000353GB
Brief IR-Tester
operating instruction 9900.0353
English
!
NUR
NURFÜR
FÜR SERVICE-ZWECKE
SERVICE-ZWECKE
ONLY
ONLYFOR
FOR SERVICE PURPOSE
SERVICE PURPOSE
RESERVE
RESERVE AUX PERSONNEL
AUX PERSONNEL
TECHNIQUE
TECHNIQUE
FUNCT. FUNCT.
MENU MENU
ON/OFF
MAQUET
0 0
MAQUET AG Postfach 2162 D-76411 Rastatt Telefon (07222) 932-0
Telefax (07222) 932-... Geschäftsbereich: OP-Tische ...-640 Service ...-608
GK99000353GB
Switch on
ON/OFF
SELECT MENU
V1.X 023009940X
MENU MENU
MENU MENU
MENU MENU
MENU MENU
MENU MENU
MENU MENU
MENU MENU
MENU MENU
MENU MENU
MENU MENU
MENU MENU
Operating instructions
MAQUET IR-Tester 9900.0353
Datum:
Date: 02.99 Seite:
Page: 2
OP-TABLE 11.20
SYS: Y 1120
MENU MENU
IR-SYSTEM-CODE IR-SYSTEM-CODE
SYS: XY 1120 SYS: XY Sy- 1120 Setting of IR-System-Code
ON/OFF
MENU MENU
increase the right digit of the IR System Code
RIGHT DIGIT FUNCT.
SYS: XY Sy- 1120
decrease the right digit of the IR System Code
FUNCT.
MENU MENU
increase the left digit of the IR System Code
LEFT DIGIT FUNCT.
SYS: XY Sy- 1120
decrease the left digit of the IR System Code
FUNCT.
>ESCAPE<
ON/OFF SYS: XY Sy- 1120
MENU MENU
Move the height drive in an upward direction
HEIGHT FUNCT.
SYS: Y IR- 1120
Move the height drive downwards
FUNCT.
MENU MENU
Move the tilting drive to the right side
TILT FUNCT.
SYS: Y IR- 1120
Move the tilting drive to the left side
FUNCT.
MENU MENU
Move the trendelenburg drive to a "head down" direction
TRENDELENBURG FUNCT.
SYS: Y IR- 1120
Move the trendelenburg drive to a "feet down" direction
FUNCT.
MENU MENU
"Zero"- position adjustment
ZERO POSITION FUNCT.
SYS: Y IR- 1120
"Zero"- position adjustment
FUNCT.
MENU MENU
>ESCAPE<
ON/OFF SYS: Y IR- 1120
MENU MENU
>ESCAPE<
ON/OFF SYS: Y IR- 1120
>ESCAPE<
1120
ON/OFF
see on page 2
(main menu)
Operating instructions
MAQUET IR-Tester 9900.0353
Datum:
Date: 02.99 Seite:
Page: 3
OP-TABLE 11.50
SYS: XY 1150
MENU MENU
IR-SYSTEM-CODE IR-SYSTEM-CODE
SYS: XY 1150 SYS: XY Sy- 1150 Setting of IR-System-Code
ON/OFF
MENU MENU
increase the right digit of the IR System Code
RIGHT DIGIT FUNCT.
SYS: XY Sy- 1150
decrease the right digit of the IR System Code
FUNCT.
MENU MENU
increase the left digit of the IR System Code
LEFT DIGIT FUNCT.
SYS: XY Sy- 1150
decrease the left digit of the IR System Code
FUNCT.
>ESCAPE<
ON/OFF SYS: XY Sy- 1150
MENU MENU
Move the height drive in an upward direction
HEIGHT FUNCT.
SYS: XY IR- 1150
Move the height drive downwards
FUNCT.
MENU MENU
Move the tilting drive to the right side
TILT FUNCT.
SYS: XY IR- 1150
Move the tilting drive to the left side
FUNCT.
MENU MENU
Move the trendelenburg drive to a "head down" direction
TRENDELENBURG FUNCT.
SYS: XY IR- 1150
Move the trendelenburg drive to a "feet down" direction
FUNCT.
MENU MENU
Move the back plate drive up
BACK FUNCT.
SYS: XY IR- 1150
Move the back plate drive down
FUNCT.
MENU MENU
Move the leg plate drive up
LEG PLATE FUNCT.
SYS: XY IR- 1150
Move the leg plate drive down
FUNCT.
MENU MENU
Move the table top longitudinally in head direction
LONGIT. SLIDE FUNCT.
SYS: XY IR- 1150
Move the table top longitudinally in a feet direction
FUNCT.
MENU MENU
"Zero"- position adjustment (table top only)
ALIGN TABLE TOP FUNCT.
SYS: XY IR- 1150
"Zero"- position adjustment (table top only)
FUNCT.
MENU MENU
Move the left leg plate drive up
LEG PLATE LEFT FUNCT.
SYS: XY IR- 1150
Move the left leg plate drive down
FUNCT.
MENU MENU
Move the right leg plate drive up
LEG PLATE RIGHT FUNCT.
SYS: XY IR- 1150
Move the right leg plate drive down
FUNCT.
MENU MENU
Gall position
FLEX / REFLEX FUNCT.
SYS: XY IR- 1150
Reverse of gall position
FUNCT.
MENU MENU
Move the height adjustment drive up with no table top on the column
H. WITHOUT T. TOP FUNCT.
(for purpose of cleaning)
SYS: XY IR- 1150
Move the height adjustment drive down
FUNCT.
MENU MENU
>ESCAPE<
ON/OFF SYS: XY IR- 1150
Operating instructions
MAQUET IR-Tester 9900.0353
Datum:
Date: 02.99 Seite:
Page: 4
FUNCTIONS CABLE FUNCTIONS CABLE
1150 BG- 1150 Operation of the tester as a cable connected hand control
ON/OFF
MENU MENU
Move the height drive in an upward direction
HEIGHT FUNCT.
BG- 1150
Move the height drive downwards
FUNCT.
MENU MENU
Move the tilting drive to the right side
TILT FUNCT.
BG- 1150
Move the tilting drive to the left side
FUNCT.
MENU MENU
Move the trendelenburg drive to a "head down" direction
TRENDELENBURG FUNCT.
BG- 1150
Move the trendelenburg drive to a "feet down" direction
FUNCT.
MENU MENU
Move the back plate drive up
BACK FUNCT.
BG- 1150
Move the back plate drive down
FUNCT.
MENU MENU
Move the leg plate drive up
LEG PLATE FUNCT.
BG- 1150
Move the leg plate drive down
FUNCT.
MENU MENU
Move the table top longitudinally in head direction
LONGIT. SLIDE FUNCT.
BG- 1150
Move the table top longitudinally in a feet direction
FUNCT.
MENU MENU
"Zero"- position adjustment (just table top)
ALIGN TABLE TOP FUNCT.
BG- 1150
"Zero"- position adjustment (just table top)
FUNCT.
MENU MENU
Move the left leg plate drive up
LEG PLATE LEFT FUNCT.
BG- 1150
Move the left leg plate drive down
FUNCT.
MENU MENU
Move the right leg plate drive up
LEG PLATE RIGHT FUNCT.
BG- 1150
Move the right leg plate drive down
FUNCT.
MENU MENU
Gall position
FLEX / REFLEX FUNCT.
BG- 1150
Reverse of gall position
FUNCT.
MENU MENU
Move the height adjustment drive up with no table top on the column
H. WITHOUT T. TOP FUNCT.
(for purpose of cleaning)
BG- 1150
Move the height adjustment drive down
FUNCT.
MENU MENU
>ESCAPE<
ON/OFF BG- 1150
MENU MENU
>ESCAPE<
ON/OFF SYS: XY IR- 1150
Operating instructions
MAQUET IR-Tester 9900.0353
Datum:
Date: 02.99 Seite:
Page: 5
ADJUSTMENT ADJUSTMENT
1150 Ju- 1150 Adjustment and setting of the drives
ON/OFF
MENU MENU
Activate the adjustment mode
ADJUSTMENT "ON" FUNCT.
Ju- 1150
Activate the adjustment mode
FUNCT.
MENU MENU
Increase the settingposition of motor 1
MOTOR 1 FUNCT.
Ju- 1150
Increase the settingposition of motor 1
FUNCT.
MENU MENU
MENU MENU
Deactivate the adjustment mode
ADJUSTMENT "OFF" FUNCT.
Ju- 1150
Deactivate the adjustment mode
FUNCT.
MENU MENU
>ESCAPE<
ON/OFF Ju- 1150
MENU MENU
Activate the adjustment mode
ADJ. T. TOP "ON" FUNCT.
JL- 1150
Activate the adjustment mode
FUNCT.
MENU MENU
Increase the settingposition of motor 5
MOTOR 5 FUNCT.
JL- 1150
Increase the settingposition of motor 5
FUNCT.
MENU MENU
MENU MENU
Deactivate the adjustment mode
ADJ. T. TOP "OFF" FUNCT.
JL- 1150
Deactivate the adjustment mode
FUNCT.
MENU MENU
>ESCAPE<
ON/OFF JL- 1150
MENU MENU
Activate the testing mode
TEST "ON" FUNCT.
Ha- 1150
Activate the testing mode
FUNCT.
SENSORS
X000000000XXXXXX Display of the output state
MENU MENU
>ESCAPE<
ON/OFF Ha- 1150
Operating instructions
MAQUET IR-Tester 9900.0353
Datum:
Date: 02.99 Seite:
Page: 6
READ IN STATUS READ IN STATUS
1150 St- 1150 Controller status readings
ON/OFF
MENU MENU
Activate the status reading mode
STATUS "ON" FUNCT.
St- 1150
Activate the status reading mode
FUNCT.
wait for display
STATUS "ON"
DATA READ
MENU MENU
MENU MENU
MENU MENU
MENU MENU
MENU MENU
MENU MENU
MENU MENU
MENU MENU
Deactivate status reading
STATUS "OFF" FUNCT.
Ju- 1150
Deactivate status reading
FUNCT.
MENU MENU
>ESCAPE<
ON/OFF St- 1150
NEURO-LOCK NEURO-LOCK
1150 NL- 1150 Neuro-Lock
ON/OFF
MENU MENU
Activate neuro lock function
LOCK FUNCT.
NL- 1150
Activate neuro lock function
FUNCT.
MENU MENU
Deactivate neuro lock function
UNLOCK FUNCT.
MENU MENU NL- 1150
Deactivate neuro lock function
FUNCT.
MENU MENU
>ESCAPE<
ON/OFF NL- 1150
MENU MENU
Activate leg plate lock function
LOCK FUNCT.
BV- 1150
Activate leg plate lock function
FUNCT.
MENU MENU
Deactivate leg plate lock function
UNLOCK FUNCT.
MENU MENU BV- 1150
Deactivate leg plate lock function
FUNCT.
MENU MENU
>ESCAPE<
ON/OFF BV- 1150
>ESCAPE<
1150
ON/OFF
see on page 2
(main menu)
Operating instructions
MAQUET IR-Tester 9900.0353
Datum:
Date: 02.99 Seite:
Page: 7
OP-TABLE 11.30
SYS: XY 1130
MENU MENU
IR-SYSTEM-CODE IR-SYSTEM-CODE
SYS: XY 1130 SYS: XY Sy- 1130 Setting of IR-System-Code
ON/OFF
MENU MENU
increase the right digit of the IR System Code
RIGHT DIGIT FUNCT.
SYS: XY Sy- 1130
decrease the right digit of the IR System Code
FUNCT.
MENU MENU
increase the left digit of the IR System Code
LEFT DIGIT FUNCT.
SYS: XY Sy- 1130
decrease the left digit of the IR System Code
FUNCT.
>ESCAPE<
ON/OFF SYS: XY Sy- 1130
MENU MENU
Move the height drive in an upward direction
HEIGHT FUNCT.
SYS: Y IR- 1130
Move the height drive downwards
FUNCT.
MENU MENU
Move the tilting drive to the right side
TILT FUNCT.
SYS: Y IR- 1130
Move the tilting drive to the left side
FUNCT.
MENU MENU
Move the trendelenburg drive to a "head down" direction
TRENDELENBURG FUNCT.
SYS: Y IR- 1130
Move the trendelenburg drive to a "feet down" direction
FUNCT.
MENU MENU
Move the lower back plate drive up
LOWER BACK FUNCT.
SYS: Y IR- 1130
Move the lower back plate drive down
FUNCT.
MENU MENU
Move the leg plate drive up
LEG PLATE FUNCT.
SYS: Y IR- 1130
Move the leg plate drive down
FUNCT.
MENU MENU
"Zero"- position adjustment
ZERO POSITION FUNCT.
SYS: Y IR- 1130
"Zero"- position adjustment
FUNCT.
MENU MENU
>ESCAPE<
ON/OFF SYS: Y IR- 1130
MENU MENU
Move the height drive in an upward direction
HEIGHT FUNCT.
SYS: XY IR- 1130
Move the height drive downwards
FUNCT.
MENU MENU
Move the tilting drive to the right side
TILT FUNCT.
SYS: XY IR- 1130
Move the tilting drive to the left side
FUNCT.
MENU MENU
Move the trendelenburg drive to a "head down" direction
TRENDELENBURG FUNCT.
SYS: XY IR- 1130
Move the trendelenburg drive to a "feet down" direction
FUNCT.
MENU MENU
Move the lower back plate drive up
LOWER BACK FUNCT.
SYS: XY IR- 1130
Move the lower back plate drive down
FUNCT.
MENU MENU
Move the leg plate drive up
LEG PLATE FUNCT.
SYS: XY IR- 1130
Move the leg plate drive down
FUNCT.
MENU MENU
"Zero"- position adjustment
ZERO POSITION FUNCT.
SYS: XY IR- 1130
"Zero"- position adjustment
FUNCT.
MENU MENU
>ESCAPE<
ON/OFF SYS: XY IR- 1130
Operating instructions
MAQUET IR-Tester 9900.0353
Datum:
Date: 02.99 Seite:
Page: 8
FUNCTIONS CABLE FUNCTIONS CABLE
1130 BG- 1130 Operation of the tester as a cable connected hand control
ON/OFF
MENU MENU
Move the height drive in an upward direction
HEIGHT FUNCT.
BG- 1130
Move the height drive downwards
FUNCT.
MENU MENU
Move the tilting drive to the right side
TILT FUNCT.
BG- 1130
Move the tilting drive to the left side
FUNCT.
MENU MENU
Move the trendelenburg drive to a "head down" direction
TRENDELENBURG FUNCT.
BG- 1130
Move the trendelenburg drive to a "feet down" direction
FUNCT.
MENU MENU
Move the lower back plate drive up
LOWER BACK FUNCT.
BG- 1130
Move the lower back plate drive down
FUNCT.
MENU MENU
Move the leg plate drive up
LEG PLATE FUNCT.
BG- 1130
Move the leg plate drive down
FUNCT.
MENU MENU
"Zero"- position adjustment
ZERO POSITION FUNCT.
BG- 1130
"Zero"- position adjustment
FUNCT.
MENU MENU
>ESCAPE<
ON/OFF BG- 1130
MENU MENU
>ESCAPE<
ON/OFF SYS: XY IR- 1130
>ESCAPE<
1130
ON/OFF
see on page 2
(main menu)
Operating instructions
MAQUET IR-Tester 9900.0353
Datum:
Date: 02.99 Seite:
Page: 9
OP-TABLE 11.31. .
SYS: XY 1131
MENU MENU
IR-SYSTEM-CODE IR-SYSTEM-CODE
SYS: XY 1131 SYS: XY Sy- 1131 Setting of IR-System-Code
ON/OFF
MENU MENU
increase the right digit of the IR System Code
RIGHT DIGIT FUNCT.
SYS: XY Sy- 1131
decrease the right digit of the IR System Code
FUNCT.
MENU MENU
increase the left digit of the IR System Code
LEFT DIGIT FUNCT.
SYS: XY Sy- 1131
decrease the left digit of the IR System Code
FUNCT.
>ESCAPE<
ON/OFF SYS: XY Sy- 1131
MENU MENU
Move the height drive in an upward direction
HEIGHT FUNCT.
SYS: Y IR- 1131
Move the height drive downwards
FUNCT.
MENU MENU
Move the tilting drive to the right side
TILT FUNCT.
SYS: Y IR- 1131
Move the tilting drive to the left side
FUNCT.
MENU MENU
Move the trendelenburg drive to a "head down" direction
TRENDELENBURG FUNCT.
SYS: Y IR- 1131
Move the trendelenburg drive to a "feet down" direction
FUNCT.
MENU MENU
"Zero"- position adjustment
ZERO POSITION FUNCT.
SYS: Y IR- 1131
"Zero"- position adjustment
FUNCT.
MENU MENU
>ESCAPE<
ON/OFF SYS: Y IR- 1131
MENU MENU
Move the height drive in an upward direction
HEIGHT FUNCT.
SYS: XY IR- 1131
Move the height drive downwards
FUNCT.
MENU MENU
Move the tilting drive to the right side
TILT FUNCT.
SYS: XY IR- 1131
Move the tilting drive to the left side
FUNCT.
MENU MENU
Move the trendelenburg drive to a "head down" direction
TRENDELENBURG FUNCT.
SYS: XY IR- 1131
Move the trendelenburg drive to a "feet down" direction
FUNCT.
MENU MENU
"Zero"- position adjustment
ZERO POSITION FUNCT.
SYS: XY IR- 1131
"Zero"- position adjustment
FUNCT.
MENU MENU
>ESCAPE<
ON/OFF SYS: XY IR- 1131
Operating instructions
MAQUET IR-Tester 9900.0353
Datum:
Date: 02.99 Seite:
Page: 10
FUNCTIONS CABLE FUNCTIONS CABLE
1131 BG- 1131 Operation of the tester as a cable connected hand control
ON/OFF
MENU MENU
Move the height drive in an upward direction
HEIGHT FUNCT.
BG- 1131
Move the height drive downwards
FUNCT.
MENU MENU
Move the tilting drive to the right side
TILT FUNCT.
BG- 1131
Move the tilting drive to the left side
FUNCT.
MENU MENU
Move the trendelenburg drive to a "head down" direction
TRENDELENBURG FUNCT.
BG- 1131
Move the trendelenburg drive to a "feet down" direction
FUNCT.
MENU MENU
"Zero"- position adjustment
ZERO POSITION FUNCT.
BG- 1131
"Zero"- position adjustment
FUNCT.
MENU MENU
>ESCAPE<
ON/OFF BG- 1131
MENU MENU
>ESCAPE<
ON/OFF SYS: XY IR- 1131
>ESCAPE<
1131
ON/OFF
see on page 2
(main menu)
Operating instructions
MAQUET IR-Tester 9900.0353
Datum:
Date: 02.99 Seite:
Page: 11
OP-TABLE 11.32
SYS: XY 1132
MENU MENU
IR-SYSTEM-CODE IR-SYSTEM-CODE
SYS: XY 1132 SYS: XY Sy- 1132 Setting of IR-System-Code
ON/OFF
MENU MENU
increase the right digit of the IR System Code
RIGHT DIGIT FUNCT.
SYS: XY Sy- 1132
decrease the right digit of the IR System Code
FUNCT.
MENU MENU
increase the left digit of the IR System Code
LEFT DIGIT FUNCT.
SYS: XY Sy- 1132
decrease the left digit of the IR System Code
FUNCT.
>ESCAPE<
ON/OFF SYS: XY Sy- 1132
MENU MENU
Steuerung einschalten
OFF ON FUNCT.
SYS: XY IR- 1132
Steuerung ausschalten
FUNCT.
MENU MENU
Move the height drive in an upward direction
HEIGHT FUNCT.
SYS: XY IR- 1132
Move the height drive downwards
FUNCT.
MENU MENU
Move the tilting drive to the right side
TILT FUNCT.
SYS: XY IR- 1132
Move the tilting drive to the left side
FUNCT.
MENU MENU
Move the trendelenburg drive to a "head down" direction
TRENDELENBURG FUNCT.
SYS: XY IR- 1132
Move the trendelenburg drive to a "feet down" direction
FUNCT.
MENU MENU
Move the back plate drive "up"
BACK FUNCT.
SYS: XY IR- 1132
Move the back plate drive "down"
FUNCT.
MENU MENU
"Zero"- position adjustment
ZERO POSITION FUNCT.
SYS: XY IR- 1132
"Zero"- position adjustment
FUNCT.
MENU MENU
Gall position
FLEX / REFLEX FUNCT.
SYS: XY IR- 1132
Reverse of gall position
FUNCT.
MENU MENU
Set castors to a drive position
LOCK / UNLOCK FUNCT.
SYS: XY IR- 1132
Set castors to a break position
FUNCT.
MENU MENU
>ESCAPE<
ON/OFF SYS: XY IR- 1132
Operating instructions
MAQUET IR-Tester 9900.0353
Datum:
Date: 02.99 Seite:
Page: 12
FUNCTIONS CABEL FUNCTIONS CABLE
1132 BG- 1132 Operation of the tester as a cable connected hand control
ON/OFF
MENU MENU
Activate the control
OFF ON FUNCT.
BG- 1132
Activate the control
FUNCT.
MENU MENU
Move the height drive in an upward direction
HEIGHT FUNCT.
BG- 1132
Move the height drive downwards
FUNCT.
MENU MENU
Move the tilting drive to the right side
TILT FUNCT.
BG- 1132
Move the tilting drive to the left side
FUNCT.
MENU MENU
Move the trendelenburg drive to a "head down" direction
TRENDELENBURG FUNCT.
BG- 1132
Move the trendelenburg drive to a "feet down" direction
FUNCT.
MENU MENU
Move the back plate drive "up"
BACK FUNCT.
BG- 1132
Move the back plate drive "down"
FUNCT.
MENU MENU
Move the leg plate drive "up"
LEG PLATE FUNCT.
BG- 1132
Move the leg plate drive "down"
FUNCT.
MENU MENU
Gall position
FLEX / REFLEX FUNCT.
BG- 1132
Reverse of gall position
FUNCT.
MENU MENU
Set castors to a drive position
LOCK / UNLOCK FUNCT.
BG- 1132
Set castors to a break position
FUNCT.
MENU MENU
Activate the motor drive unit forwards
MOTOR PWR. MOVE. FUNCT.
BG- 1132
Activate the motor drive unit backwards
FUNCT.
MENU MENU
>ESCAPE<
ON/OFF BG- 1132
MENU MENU
>ESCAPE<
ON/OFF SYS: XY IR- 1132
ADJUSTMENT ADJUSTMENT
1132 Ju- 1132 Adjustments
ON/OFF
MENU MENU
set the final adjustment of "zero" position
ZERO POSITION FUNCT.
Ju- 1132
set the final adjustment of "zero" position
FUNCT.
MENU MENU
presetting of a patient position
FLEX NORMAL FUNCT.
IR- 1132
presetting of a patient position
FUNCT.
MENU MENU
presetting of a patient position
REFLEX NORMAL FUNCT.
IR- 1132
presetting of a patient position
FUNCT.
MENU MENU
presetting of a patient position
REFLEX REVERSE FUNCT.
IR- 1132
presetting of a patient position
FUNCT.
MENU MENU
reset of every pre adjustments
SET DEFLT. VALUES FUNCT.
IR- 1132
reset of every pre adjustments
FUNCT.
MENU MENU
>ESCAPE<
ON/OFF SYS: XY IR- 1132
Operating instructions
MAQUET IR-Tester 9900.0353
Datum:
Date: 02.99 Seite:
Page: 13
READ DADTA-BUS READ DADTA-BUS
1132 DB- 1132 Read data information
ON/OFF
MENU MENU
CONTROL DATA
Controller data and nominal battery voltage
XXXXXXXXXX XX,XXV
MENU MENU
HAND-CONTR. DATA
Data of hand control (cable connected)
XXXXXXXXXX
FOOT-SWITCH DATA
Data of foot switch
XXXXXXXXXX
MENU MENU
>ESCAPE<
ON/OFF DB- 1132
MENU MENU
CONTROL : BYTES
X 00 00 00 00 00 Read bytes from controller
MENU MENU
HAND-CONT. : BYTES
X 00 00 00 00 00 Read bytes from cable connected hand control
MENU MENU
FOOT-SWS : BYTES
X 00 00 00 00 00 Read bytes from foot switch
MENU MENU
MOVING : BYTES
X 00 00 00 00 00 Read bytes from drive unit
TESTER : BYTES
X 00 00 00 00 00 Reserved for future applications
MENU MENU
EXP. 1 : BYTES
X 00 00 00 00 00 Reserved for future applications
MENU MENU
EXP. 2 : BYTES
X 00 00 00 00 00 Reserved for future applications
MENU MENU
>ESCAPE<
ON/OFF SYS: Y IR- 1120
>ESCAPE<
1120
ON/OFF
see on page 2
(main menu)
Operating instructions
MAQUET IR-Tester 9900.0353
Datum:
Date: 02.99 Seite:
Page: 14
TRANSMAQUET
SYS: Y TRNS
MENU MENU
IR-SYSTEM-CODE IR-SYSTEM-CODE
SYS: XY TRNS SYS: XY Sy- TRNS Setting of IR-System-Code
ON/OFF
MENU MENU
increase the right digit of the IR System Code
RIGHT DIGIT FUNCT.
SYS: XY Sy- TRNS
decrease the right digit of the IR System Code
FUNCT.
MENU MENU
increase the left digit of the IR System Code
LEFT DIGIT FUNCT.
SYS: XY Sy- TRNS
decrease the left digit of the IR System Code
FUNCT.
>ESCAPE<
ON/OFF SYS: XY Sy- Trns
MENU MENU
Programmunterstützte Umbettung (Ausschleusen)
TRANSFER FUNCT.
SYS: Y IR- TRNS
Patient transfer by programm support (
FUNCT.
MENU MENU
Speed up
SPEED FUNCT.
SYS: Y IR- TRNS
Speed up
FUNCT.
MENU MENU
Move the height drive in an upward direction
HEIGHT FUNCT.
SYS: Y IR- TRNS
MENU MENU
Move the height drive downwards
FUNCT.
MENU MENU
Skip button to the next programmstep
SKIP - BUTTON FUNCT.
SYS: Y IR- TRNS
Skip button to the next programmstep
FUNCT.
MENU MENU
>ESCAPE<
ON/OFF SYS: Y IR- TRNS
MENU MENU
>ESCAPE<
ON/OFF SYS: Y IR- TRNS
MENU MENU
>ESCAPE<
ON/OFF L- TRNS
>zurück<
TRNS
ON/OFF
see on page 2
(main menu)
Operating instructions
MAQUET IR-Tester 9900.0353
Datum:
Date: 02.99 Seite:
Page: 15
CHARGING UNIT CHARGING UNIT
CHARGER Testing of carging unit
ON/OFF
MENU MENU
CHARGING_TEST
CHARGER DEFECT/O.K. Tester indicates function of charging unit
>ESCAPE<
ON/OFF CHARGER
IR DETECT PA IR DETECT PA
IR- PA Tester receives all IR signals
ON/OFF
MENU MENU
TEST ACTIVE
IR- PA Interference signals can be searched for
>ESCAPE<
ON/OFF IR- PA
MENU MENU
TEST ACTIVE
IR- ASIC ASIC pulsed signal is being read (transmitter signals))
>ESCAPE<
ON/OFF IR- ASIC
LANGUAGE/Sprache LANGUAGE/Sprache
ON/OFF L/S
MENU MENU
GERMAN
L/S Set German as display language
ON/OFF
MENU MENU
ENGLISH
L/S Set English as display language
ON/OFF
FRENCH
L/S Set French as display language
ON/OFF
MENU MENU
>ESCAPE<
ON/OFF L/S
SWITCH OFF
Tester is switched off
ON/OFF
Operating instructions
MAQUET IR-Tester 9900.0353
Datum:
Date: 02.99 Seite:
Page: 16
MAQUET Operating instructions
IR-Tester 9900.0353
Datum:
Date: 02.99 Seite:
Page: 17
® MAQUET Eingetragenes Warenzeichen der MAQUET Aktiengesellschaft. Copyright © by MAQUET, Rastatt
MAQUET Aktiengesellschaft
Geschäftsbereich Service
Kehler Straße 31
76437 Rastatt/Germany
National
Medizintechnik 01 80/3 21 21 44
Sterilisation 01 80/3 21 21 66
Ersatzteile 0 72 22/932-707
International
Consulting +49/72 22/932-782
& Spare Parts
Fax +49/72 22/932-825
Technische und konstruktive Änderungen vorbehalten.
The right is reserved to make technical and design changes.
E-mail info.service@maquet.de Sous réserve de modifications dues au développement technique.
Internet www.maquet.com Nos reservamos el derecho de hacer modificaciones conformes al desarrollo técnico.
MAQUET